Professional Documents
Culture Documents
Zimmer® UNIVERSAL Power System Instruction Manual: Battery / Electric
Zimmer® UNIVERSAL Power System Instruction Manual: Battery / Electric
Instruction Manual
Battery / Electric
Contents
1. INDICATIONS FOR USE................................................................................................... 3
2. DESCRIPTION AND TECHNICAL DATA ......................................................................... 3
2.1 Battery / Electric Handpiece.........................................................................................................3
2.2 Electric Power Supply ..................................................................................................................4
2.3 Compact Battery Charger.............................................................................................................4
2.4 Battery and Aseptic Transfer Kit (ATK) Housing..........................................................................5
2.5 Flash Battery.................................................................................................................................6
3. OPERATING PROCEDURE.............................................................................................. 6
3.1 To connect and disconnect the attachments..................................................................................6
3.2 To connect and disconnect the battery..........................................................................................7
3.3 To connect and disconnect the power cord...................................................................................7
3.4 To turn on and off the handpiece...................................................................................................8
3.5 To start the motor Forward and Reverse......................................................................................8
3.6 To charge the Batteries..................................................................................................................9
3.7 To transfer Battery into sterile ATK Housing................................................................................9
3.8 To connect the UNIVERSAL Power Supply.................................................................................11
4. CLEANING, DECONTAMINATION AND STERILIZATION............................................. 11
5. ELECTRICAL SAFETY.................................................................................................... 18
5.1 Electromagnetic Emissions.........................................................................................................18
5.2 Electromagnetic Immunity..........................................................................................................19
5.3 Electromagnetic Immunity, RF Portable Equipment..................................................................20
5.4 Recommended Separation Distances..........................................................................................21
5.5 Cord (Cable) Length...................................................................................................................22
6. WARNING AND PRECAUTIONS.................................................................................... 23
7. TECHNICAL SERVICE.................................................................................................... 24
8. WARRANTY INFORMATION (U.S.A. only).................................................................... 25
9. WARRANTY (Outside U.S.A.)........................................................................................ 25
10. CONTACT INFORMATION.............................................................................................. 25
11. SYMBOLOGY.................................................................................................................. 26
2
1. INDICATIONS FOR USE
The Zimmer UNIVERSAL Power System is a power tool intended to be used for large and small
bone surgeries (such as knee, hip, foot, hand, and shoulder). It is to be used by surgeons in
the operating room environment.
The UNIVERSAL Handpiece can run on battery power or electric power when used with
the supplied power cord. It features forward and reverse functions and is offered in a single
trigger and double trigger model.
Universal locking
mechanism
Safety
lock
High-capacity lithium
rechargeable battery
Electric connection
Electric power cord
3
Power source Mains or Battery
Power supply Battery: 18V DC / 20A
Mains: 100-240V~, 50/60Hz, 4.6-1.6A
Operation Intermittent: 1 min on / 5 min off
Run time with battery Up to 15 min
Type of protection against electric shock Battery: Internally powered equipment
Mains: Class I
Degree of protection against electric shock Type BF
Degree of protection against water ingress IPXO (ordinary)
Operating conditions Operating room environment
Transport and storage Normal warehouse conditions
4
Output 4x 18V DC, 1.5A
Dimensions 321 mm x 206 mm x 93.65 mm
Weight 2.7 kg approximately (without battery)
Operating Conditions 5 - 25°C; Do not use in patient area
Transport and Storage Normal warehouse conditions
5
2.5 Flash Battery
REF
89-8510-440-50
3. OPERATING PROCEDURE
3.1 To connect and disconnect the attachments
Push
6
3.2 To connect and disconnect the battery
7
3.4 To turn on and off the handpiece
8
3.6 To charge the Batteries
Power Switch button
Indicator
Light
Remove batteries from charger when recharged.
Remove batteries before switching off charger.
Contacts
Contacts
Battery is not sterile; sterile person must not touch battery during transfer.
Housing is sterile; non-sterile person must not touch housing during transfer.
Seal between the lid and housing must be present and pressed to guarantee seal.
9
STERILE PERSON (A): Position locking system to and open lid of housing.
Unlocked
Locked
NON-STERILE PERSON (B): Insert battery into housing through funnel opening.
Push until fully inserted, and remove funnel.
Funnel B
A B B
A A A A
STERILE PERSON (A): Close lid. Position the locking system to to lock the
housing.
A
Unlocked Locked
Once a “click” has been heard, the battery pack is ready to use.
10
3.8 To connect the UNIVERSAL Power Supply Power Switch button
Indicator
- Indicator light will turn green when the power is on.
Lights Connector
Hold by the
connector to
take off the
cord
These instructions are applicable to the reusable components of the Zimmer UNIVERSAL
Power System, excluding the electronic power supply, battery, and battery charger.
11
4.1 PROCESS FLOW CHART
4.4. Pre-disinfection
4.11. Inspection
4.8. Disinfection
4.12. Sterilization
12
4.2 PREPARATION BEFORE CLEANING
The Zimmer UNIVERSAL Electric Power Supply (REF 89-8510-420-10), the Zimmer
UNIVERSAL Battery for Aseptic Transfer Kit (ATK), (REF 89-8510-440-20), and the
Zimmer UNIVERSAL Compact Battery Charger (REF 89-8510-421-00) must not be
cleaned in a washer disinfector or sterilized.
For additional information about the dismantling of the device, refer to the specific Zimmer
UNIVERSAL Power System Instruction for Use
4.4 PRE-DISINFECTION
Immediately after use, carefully rinse the device with sterile water in order to remove all
visible soil.
Remove any visible soil with a non-shedding soft cloth saturated with detergent (per hospital
protocol). Discard the soft cloth after use.
If the cleaning action cannot be done immediately, put the disconnected devices in a new
soft cloth strongly saturated with a detergent (per hospital protocol).
The Manual Cleaning process has been validated according to AAMI TIR Number 12
and ISO 17664 guidelines
Materials:
Do not utilize detergents with phenol, aldehyde and non alkaline type (pH should be
less than 9) active principles: quaternary ammonium and guanidine derivatives (process
validated with Deconex 53 plus, 4%).
Dilution and parameter temperature must follow the detergent manufacturer’s specification
Utilize Sterile and soft water (hardness lower than 120 mg/L of CaCO3)
Utilize Non abrasive brushes and non abrasive, non shedding soft cloth.
13
Zimmer UNIVERSAL Power System should not be immersed or cleaned in an ultrasonic
machine.
Carefully clean each part of the device with a brush and/or a soft cloth with an appropriate
detergent for a minimum of 2 minutes.
While cleaning, pay particular attention to critical sites: exposed parts, narrow cavities, small
openings and moving parts.
Let the detergent flow into the cavities and areas that are difficult to reach.
Dry the device utilizing a dry non-shedding cloth. Medical quality filtered air may be utilized if
available.
4.7 INSPECTION
Inspect all devices carefully to ensure all visible soil and detergent is removed.
If necessary, repeat the steps for disinfectant cleaning.
Note: If the devices are not disinfected immediately after rinsing, carefully dry the devices with
a soft cloth to avoid any microbial contamination due to humidity.
4.8 DISINFECTION
Place and cover the device in a soft cloth strongly saturated with the recommended
detergent/disinfectant for 30 minutes.
After the 30 minute contact time, wipe the device down with the saturated cloth paying
particular attention to joints, mated areas, crevices, and small openings.
14
Validated Manual Cleaning and Disinfection Procedure
Step Step Description Step Instruction Accessories Duration
WARNING:
Zimmer UNIVERSAL Power System should not be immersed or submerged in water
or cleaner.
15
4.9 AUTOMATIC DISINFECTANT CLEANING (Recommended)
The Automatic Cleaning and Disinfection process has been validated according to
AAMI TIR Number 30 and ISO 17664 guidelines
Materials:
Non abrasive and non shedding soft cloth.
Neutralizing agent.
(Validated with a neutralizing agent (at 20°C: pH: 2.6-3; specific gravity: 1.2g/cm3; viscosity:
<10mPas)).
Rinse aid
(Validated with a rinse liquid (at 20°C: pH: 5.9-6; specific gravity: 1.02g/cm3; viscosity: <50mPas)).
Washer disinfector
(Validated with a medical use washer disinfector according to the EN ISO 15883).
Critical sites must be manually precleaned with a non-shedding soft cloth and/or non-
abrasive brush before the washer disinfector cycle. While cleaning, pay particular attention to
critical sites such as narrow cavities, small openings and moving parts.
Put medical devices in the washer disinfector. Avoid all contact between the devices
(movement may damage devices and may jeopardize the cleaning operation)
Put implements so that holes are aimed downwards (to make the flow easier).
16
Automated Cleaning/Disinfection Procedure
Step Step Step Instruction Accessories Duration
Description
1 Contamination Rinse product with -Cold/ Until all visible soil
Removal of sterile water removing Roomtemperature is removed.
Critical Sites any visible organic sterile Water Minimum of two
material. Apply -Soft bristle brush (2) minutes is
detergent with a (Do not utilize metal recommended
non shedding cloth cleaning brushes)
and as necessary -Non-shedding cloth
with assistance of a -Recommended
soft bristle brush to Detergent/Disinfectant
remove any visible
soil.
2 Rinse Rinse product with -Cold/ Until all visible
sterile water. RoomTemperature detergent is
sterile Water removed
4.11 INSPECTION
Inspect each part of the devices to make sure all soil is removed.
If necessary, repeat the cleaning process.
4.12 STERILIZATION
Sterilization parameters have been validated according to AAMI TIR Number 12, ISO
17664 and ISO 17665 guidelines (warranty of a Security Assurance Level (SAL) of 10-6)
Remove all washing caps (REF 89-8510-455-50) on hoses and on pneumatic handpiece
air inlet before the sterilization process.
Do not insert any cap for the sterilization process.
17
The Zimmer handpieces and accessories have been designed to withstand the longest
sterilization process. Under this requirement the following Steam sterilization cycle has been
validated: 134°C with a holding time of 18 minutes (Prion cycle), minimum drying time of
5. ELECTRICAL SAFETY
This Zimmer UNIVERSAL Power System complies with the Electromagnetic
Compatibility Standard IEC 60601. The Zimmer UNIVERSAL Compact Battery
Charger complies with the Electromagnetic Compatibility Standard EN 55014.
Both devices need to be installed and put into service according to the section
“Electromagnetic compatibility” provided in this document.
This product should not be used adjacent to or stacked with other equipment. If
adjacent or stacked use is necessary, this product should be observed to verify
normal operation in the configuration in which it will be used.
The use of accessories and cables other than those specified or sold by the
manufacturer of this product as replacement parts, may result in increased
emissions or decreased immunity of this product.
18
5.2 Electromagnetic Immunity
The Zimmer UNIVERSAL Power System and Zimmer UNIVERSAL Compact Battery
Charger are intended for use in the electromagnetic environment specified below. The
customer or the user should assure that it is used in such an environment.
<5 % UT <5 % UT
(>95 % dip in UT) (>95 % dip in UT)
for 5 sec for 5 sec
Power frequency 3 A/m 3 A/m Power frequency magnetic fields
(50/60 Hz) should be at levels characteristic
magnetic field of a typical location in a typical
commercial or hospital environment.
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
19
5.3 Electromagnetic Immunity, RF Portable Equipment
The Zimmer UNIVERSAL Power System and Zimmer UNIVERSAL Compact Battery Charger
are intended for use in the electromagnetic environment specified below. The customer or the
user should assure that it is used in such an environment.
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio telephones (cellular/
cordless) and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the Zimmer Power System is used exceeds
the applicable RF compliance level above, the Zimmer Power System should be observed
to verify normal operation. If abnormal performance is observed, additional measures may
be necessary, such as re-orienting or relocating the Zimmer Power System.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
20
5.4 Recommended Separation Distances
The Zimmer UNIVERSAL Power System and Zimmer UNIVERSAL Compact Battery
Charger are intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Zimmer UNIVERSAL Power System or
Zimmer UNIVERSAL Compact Battery Charger as recommended below, according to the
maximum output power of the communications equipment.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
21
5.5 Cord (Cable) Length
Surge,
IEC 61000-4-5
Conducted RF
IEC 61000-4-6
Radiated RF,
IEC 61000-4-3
22
6. WARNING AND PRECAUTIONS
- Only medical professionals who are thoroughly familiar with the Zimmer UNIVERSAL
Power System function, application and instruction for use should operate any Zimmer
equipment.
-- Do not use devices if they appear to be damaged or defective.
-- Do not use in presence of flammable products.
-- Do not use accessories or cords other than those specified or provided by Zimmer.
-- The use of Flash Batteries (REF 89-8510-440-50) may not be authorised in some
countries.
-- The use of accessories and cords other than those specified or sold by Zimmer may
result in increased emissions.
-- Certain types of mobile telecommunication equipment could potentially interfere with this
product.
-- This product should not be used adjacent to or stacked with other equipment. If adjacent
or stacked use is necessary, this product should be observed to verify normal operation.
-- For intermittent (1 min on / 5 min off) operation only.
-- It is highly recommended to continuously moisten bones when drilling, cutting, or
reaming.
-- The handpieces have built-in software that protects the internal electronic components.
When this protection is activated, the handpiece may not be responsive. To reinitiate the
system, disconnect the battery and reconnect it.
-- When not in use, turn handpiece off (safety position “0”) and remove the Flash Battery or
ATK Housing.
-- Do not touch the contacts of the batteries nor put them in a “short circuit” situation.
-- Do not touch the contacts of the battery charger or Sterilizable Cord for Electric Power
when powered “on” nor put them in a “short circuit” situation.
-- Do not introduce any object into the Electric Power Supply or Compact Battery Charger.
-- Do not use alkaline cleaning agents on the devices.
-- Do not immerse.
-- Do not put the Battery (REF 89-8510-440-20) or Flash Battery (REF 89-8510-440-50) in
an automated washer.
-- Do not expose the Electronic Power Supply (REF 89-8510-420-00 / 89-8510-420-10),
Battery (REF 89-8510-440-20) and Compact Battery Charger (REF 89-8510-421-00) to
wet environment.
-- Do not sterilize the Electronic Power Supply (REF 89-8510-420-00 / 89-8510-420-10),
Battery (REF 89-8510-440-20) and Compact Battery Charger (REF 89-8510-421-00) to
wet environment.
-- The ATK Housing (REF 89-8510-440-10) must be sterilized with his lid open.
-- After sterilization, make sure the handpiece has reached temperature below 40°C before
use.
-- The use of a Zimmer UNIVERSAL Sterilization Case is strongly recommended.
-- Do not open the device as electrocution hazard may endanger health.
-- Device must be disconnected when servicing as electrocution hazard may seriously
damage health.
23
7. TECHNICAL SERVICE
Risk of electric shock. Do not remove cover.
Rough shocks/handling should be avoided.
No lubrication or maintenance needed.
Do not attempt to disassemble the handpiece. It is a factory sealed unit with no
user serviceable parts inside.
Zimmer is not liable for any device malfunction resulting from repairs or service
not performed by a Zimmer authorized service center.
Always decontaminate the system before returning to Zimmer.
The Zimmer UNIVERSAL Power System should be returned every 12 months for inspection
and preventive maintenance. Annual factory calibration checks are strongly recommended to
verify continued accuracy.
The handpieces have built-in software that protects the internal electronic components.
When this protection is activated, the handpiece may not be responsive. To reinitiate the
system, disconnect the battery and reconnect it.
When it is necessary to return the Zimmer UNIVERSAL Power System for preventive
maintenance or repair within the U.S.A. call 1-800-830-0970 to receive a Return Goods
Authorization (RGA) number. Outside the U.S.A., contact your local Zimmer representative.
The instrument must be properly packaged when sent in for repair. If the original packaging
is no longer available, proper packaging can be requested when the RGA is received. A
purchase order must accompany all equipment for repair. The customer will be responsible
for all shipping charges.
Fuse Replacement
Fuse replacement should be performed by a person familiar with electronic medical devices.
24
8. WARRANTY INFORMATION (U.S.A. only)
Zimmer Surgical warrants that the Zimmer UNIVERSAL Power System, along with all
parts and accessories, have been tested and inspected, and have left the factory in
proper working condition, free from visible defects.
Zimmer Surgical warrants to the first consumer purchaser of new Zimmer UNIVERSAL
Power System and accessories will, under normal and reasonable use, be free from
defects in material and workmanship for one (1) year after the date of shipment from the
factory. The warranty period for hoses is six (6) months. Zimmer UNIVERSAL Power
System consumables are warranted to be free from defects in material and workmanship
upon delivery. During the warranty period, Zimmer Surgical shall repair (or at its sole
option replace) the defective product or part without cost to the purchaser. Defective parts
replaced under this warranty shall become the property of Zimmer Surgical. This warranty
does not cover damage caused by misuse, abuse, accident, neglect, or any use not
prescribed in this manual. If the unit becomes defective because of misuse or abnormal
conditions of operation, repairs will be billed at our current rate.
Neither Zimmer Surgical nor the Zimmer distributor who sells the Zimmer UNIVERSAL
Power System is responsible for indirect, incidental, or consequential damages. Some
states do not allow the exclusion of incidental or consequential damages, so the above
limitations or exclusions may not apply to you.
25
11. SYMBOLOGY
Symbol Description
Caution
Do not immerse
Serial number
Can be recycled
Lot number
26