CR PHANTOM REPORT

LONG-TERM STABILITY TEST PROCEDURE FOR DÜRR NDT CR SYSTEMS

Report ID General Information

Company

Address

Scanner Country

Model

Serial No.

Firmware Version CR Phantom


Imaging Plate Model No.

Type Serial No.

Lot No. Procedure Rev.

Scanner Parameters X-ray Parameters


Resolution (µm) Tube Model

PMT Voltage (V) Tube Serial No.

Laser Power Tube Voltage (kV)

Laser Focus (µm) Tube Current (mA)

Beam Deflector Speed (rpm) Exposure Time (sec)

Filter (if any)

SDD (mm)

Conclusion

Place, date Signature


Rev. 02 / 2018-12-17

DÜRR NDT GmbH & Co. KG


Höpfigheimer Str. 22
74321 Bietigheim-Bissingen
Germany Page 1/2
CR PHANTOM REPORT
LONG-TERM STABILITY TEST PROCEDURE FOR DÜRR NDT CR SYSTEMS

Report ID

Test procedure Acceptance criteria

1 PV & BAM snail pass / fail pass

2 SRb (laser direction) µm , % –

SRb (transport direction) µm , % –

3 Signal-to-noise ratio ≥ 70

4 Geometric distortion (laser direction) % ≤ 2%

Geometric distortion (transport direction) % ≤ 2%

5 Laser jitter yes / no no

6 PMT non-linearity % –

7 Scanner slipping yes / no no

8 Scanner shading (EC - ER) ± % ± 15 %

Scanner shading (EC - EL) ± % ± 15 %

9 Contrast sensitivity (aluminium) % ≤ 2% step

Contrast sensitivity (copper) % ≤ 2% step

Contrast sensitivity (stainless steel) % ≤ 2% step

10 Erasure % ≤ 1%
Rev. 02 / 2018-12-17