Professional Documents
Culture Documents
RCT-implant Stability
RCT-implant Stability
Purpose: The primary goal of this stratified randomized controlled trial (SRCT) was to compare the stability of
dental implants placed under three different loading regimens during the first 16 weeks of healing following
implant placement. Implants were loaded immediately, early (6 weeks), or with conventional/delayed timing
(12 weeks). Secondary outcomes were to compare marginal bone adaptation for 3 years after placement.
Materials and Methods: Single posterior implant sites in the maxilla or mandible were examined. The insertion
torque value was the primary determinant of load assignment. Resonance frequency analysis was performed
at follow-up appointments for the first 16 weeks (with results provided as implant stability quotients [ISQs]).
Marginal bone levels were assessed via radiographs. Results: Forty patients each received a single 4.0-mm-
diameter dental implant between 2004 and 2007. One implant failure occurred in Lekholm and Zarb type 4
bone with insertion torque value (ITV) of < 8.1 Ncm; the cumulative success rate was 97.5%. All implants, when
classified by bone and loading type, increased in stability over time, with a minor reduction of 1.3 ISQ units
seen at 4 weeks in the immediate loading group. The mean marginal bone loss over 3 years was 0.22 mm.
The mean ITVs at implant placement for bone types 1 and 2 (grouped together), 3, and 4 were 32, 17, and 10,
respectively, and were significantly different (P < .05). Conclusions: ITV was a good objective measure of bone
type. Using an ITV of 20 Ncm as the determinant for immediate loading and an ITV of 10 Ncm or greater as
the determinant for early loading provided long-term success for this implant and led to no negative changes
in tissue response. All bone type groups and loading groups showed no reduction in stability during the first
4 months of healing. Int J Oral Maxillofac Implants 2012;27:945–956.
Key words: dental implants, immediate loading, implant stability, randomized controlled clinical trial,
resonance frequency analysis, single-tooth replacement
© 2012 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART OF MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Barewal et al
edentulous and partially edentulous patients.15 Twelve implant stability quotient (ISQ), an algorithm-derived
trials compared immediate with conventional loading, assessment of the damping of the harmonic frequency
three trials compared early with conventional loading, relative to the type of implant or abutment to which
and six trials compared immediate and early loading. it is connected. ISQ units range from 1 to 100 and are
The authors found that there were no statistically sig- derived from the stiffness (N/µm) of the transducer/
nificant differences in prosthesis success, implant suc- implant/bone system and the calibration parameters
cess, and marginal bone levels when different loading of the transducer. An increased ISQ value indicates
regimens were applied. However, the authors stated an increased stiffness of the implant and surrounding
that it was difficult to draw conclusions because of the bone. This device can provide prospective monitor-
small number of trials, low patient numbers, and short ing and shows fluctuations in stiffness of the implant
follow-up periods (4 months to 1 year). To date, there interface as bone matures from primary to secondary
are no RCTs comparing immediate loading to early and contact. The second-generation device, the Osstell
delayed loading for the single dental implant. With a Mentor (Integration Diagnostics) substitutes the use of
goal to expedite treatment without decreasing suc- the L-shaped transducer for a wireless receptor called
cess rates compared to conventional loading protocols a SmartPeg, which is excited by a set of “pulse trains”
for the single implant, studies are required to evaluate from a contact-free probe.25
the long-term predictability of outcomes. Clinically, RFA values vary based on three elements:
Biomechanically, the most challenging application the stiffness of an implant as a function of the geom-
of immediate loading is the single posterior dental etry and material composition; the stiffness of the
implant. However, the number of studies of this indi- implant-tissue interface, which is dependent on the
cation are small because of the restrictive selection bone-to-implant contact area and the height of the
criteria regarding implant length, bone quantity, and implant above the bone; and finally the stiffness of the
insertion torque.16–19 However, developing an imme- surrounding tissue, which is determined by the non-
diate loading protocol for the single posterior implant uniform ratio of cortical and cancellous bone and the
would be useful, as this is the most common indication inherent bone density.26,27
for implant dentistry today. Another measure of primary implant stability is
It is believed that the most important determinant cutting resistance. This was originally developed by
of success with immediate loading is primary implant Johansson and Strid28 and later improved by Friberg
stability.20–22 Without adequate primary implant sta- et al.29 It was observed that the energy required by an
bility, successful secondary stability caused by bone electric motor to cut bone during implant surgery cor-
regeneration and remodeling cannot occur, which relates to a degree with bone density and influences
would lead to failure of osseointegration. It is therefore implant stability.29 ITV is a numeric value given to the
of utmost importance to be able to quantify implant peak insertion torque reached by the surgical motor
stability upon placement and subsequent time points during the final stage of implant placement into the
during the early healing period. prepared site. ITV is a more objective, quantifiable as-
sessment of bone density than the clinician-dependent
evaluation of bone quality based on the Lekholm and
Stability Measurement of Implants Zarb classification.30 The use of ITV to determine opti-
mal healing periods prior to implant loading has been
Two well-recognized quantitative methods of assess- discussed.31
ing primary implant stability are insertion torque val- Although ISQ and ITV both provide quantifiable
ue (ITV) and resonance frequency analysis (RFA). RFA measures of implant stability, they assess different as-
offers a clinical, noninvasive measure of implant and pects of stability. ISQ measures the axial stability of the
bone stiffness and is presumed to be an indirect mea- implant, and ITV measures rotational stability. Both as-
sure of osseointegration.23,24 Meredith and coworkers sessments together provide the clinician with a better
reported on the use of a transducer that comprised two understanding of primary stability.
piezoceramic elements tightened to an implant body The aim of this stratified randomized controlled trial
or abutment with a screw. One of the piezoceramic el- (SRCT) was to compare the stability of dental implants
ements vibrates and the other serves as the receptor of placed in healed ridges in areas of bounded edentulous
the signal. The resonance peaks from the received sig- spaces using one of three loading regimens during the
nal indicate the first flexural (bending) resonance fre- first 16 weeks following implant placement. A second
quency of the measured object.24 Osstell (Integration aim is to assess the changes in bone crestal height over
Diagnostics) has combined the transducer, computer- the first 3 years in each loading category. Therefore the
ized analysis, and the excitation source into a single de- purpose of this SRCT was to compare the radiographic
vice. The unit of measure created for this device is the and tissue health outcomes of single-tooth implants
© 2012 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART OF MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Barewal et al
placed and loaded with one of three healing periods ment, the subject’s medical and dental history was re-
with a randomization criteria based on ITV. viewed, and defined inclusion/exclusion criteria were
applied (Table 1). Only nonsmoking patients requiring
one dental implant (4.0 mm in diameter, OsseoSpeed,
MATERIALS AND METHODS Astra Tech) in the posterior maxilla or mandible were
accepted (Fig 1). All sites had natural or restored teeth
Patient Selection mesial and distal to the planned site of interest (bound-
This clinical trial was designed as a prospective, strati- ed edentulous space). All patients had a restored stable
fied, randomized study. The study received local in- occlusion (ie, with canine or mutually protected disclu-
stitutional review board approval and the informed sion). Implants were 11 or 13 mm long. Clinical and ra-
consent of all subjects. At the initial screening appoint diographic examinations were used to limit the study
© 2012 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART OF MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Barewal et al
Excluded (n = 142)
• Not meeting inclusion criteria (n = 142)
Randomized (n = 40)
Allocation
Allocated to immediate Allocated to early Allocated to delayed
loading group (n = 8) loading group (n = 17) loading group (n = 15)
• Received allocated • Received allocated • Received allocated
intervention (n = 8) intervention (n = 17) intervention (n = 15)
Follow-up
Lost to follow-up (n = 1) after first Lost to follow-up (n = 0) Lost to follow-up (n = 1) after
year (moved) Discontinued intervention (n = 0) 1st year (moved)
Discontinued intervention (n = 0) Discontinued intervention (n = 1)
(early implant failure; replaced
implant and defaulted to 12 week
loading because of low ITV)
Analysis
Analyzed (n = 8) Analyzed (n = 17) Analyzed (n = 15)
• Excluded from analysis (n = 1) • Excluded from analysis (n = 0) • Excluded from analysis (n = 2)
for year 2 and 3 outcomes (rotational mobility at 6 weeks
after placement)
to patients with sufficient bone quantity to completely with poor primary stability, a stratified RCT was de-
encase the implant. This means there was sufficient signed with ITV as the primary determinant for alloca-
bone height such that the implant would not encroach tion to loading group. If the ITV was less than 10 Ncm,
on vital structures such as the inferior alveolar nerve the implant defaulted to the conventional loading
or the sinus floor. Sufficient width would exist so that (12-week) group.
the implant could be placed within the confines of
the existing bone without dehiscence or fenestrations Assignment and Randomization
requiring significant grafting at time of implant place- Since assignment and randomization were stratified by
ment. Typically, the space dimensions were: 6 mm or the ITV measured at implant placement, two random-
greater ridge width buccolingually and at least 6 mm ization lists were generated by the biostatistician: one
but less than 10 mm of ridge width mesiodistally. If the for ITV ≥ 20 Ncm (group A) and one for ITV < 20 but ≥
implant could be placed with only a few screw threads 10 Ncm (group B). Subjects were assigned to a loading
exposed, grafting was allowed to cover these threads group in a sequential manner from the appropriate list.
(freeze-dried bone allograft, Lifenet). In the occluso- For group A (≥ 20 Ncm ITV), for which all three loading
gingival dimension, there had to be at least 7 mm of groups were possible, loading group allocation was
space from the planned head of the implant to the oc- randomly assigned using alternating permuted blocks
clusal plane. Patients selected based on these criteria of size 6 or 9 to mask any pattern. For group B (10 to
were assigned a random numeric identifier to aid in < 20 Ncm ITV), allocation to either the 6-week or the
the blinded assignment of loading group. 12-week loading group was performed using alternat-
ing permuted blocks of size 4 or 8. In each instance,
Treatment Groups a randomization list twice the size anticipated to be
The loading groups were immediate, early (6 weeks), needed was prepared to accommodate unexpected
and conventional/delayed (12 weeks). Immediate variability in the distribution of ITVs. No randomization
loading was defined as provisionalization on the same list was needed for group C (0 to < 10 Ncm ITV), since
day as implant placement, and early and conventional all implants default to the delayed loading group. If an
loading were defined as provisionalization at 6 or 12 implant was rotationally mobile at the time of place-
weeks postplacement, respectively. Because of the risk ment, it was left undisturbed for 6 weeks, and the pa-
of failure involved in immediate loading of implants tient defaulted into the 12-week loading group.
© 2012 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART OF MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Barewal et al
50
45
40
35
Torque (Ncm)
30
25
20
15
10
5
0
0 6 12 18 24 30 36 42 48 54 60 66 72 78 84 90
Time (s)
© 2012 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART OF MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Barewal et al
4.0 coping, reference no. 22967, Astra Tech), an abut- the cone was the same for all radiographs. An indepen-
ment screw (reference no. 24132 Ti-Alloy, Astra Tech), dent radiologist masked to subject information deter-
and cold-curing acrylic resin (Jet Acrylic, Lang Dental). mined the distance from the mesial and distal crestal
The abutments were hand-tightened with finger pres- bone peaks to the outer aspect of the implant bevel
sure to approximately 10 to 15 Ncm. Off-axis loading to the nearest 0.1 mm. The changes in crestal bone
was minimized by narrowing the occlusal table and re- height from baseline to 3 years were calculated.
stricting occlusion to a single central contact in maxi-
mum intercuspation, which would allow dragging of a Definitive Crown Procedures
10-µm shim stock with no excursive contacts. Patients All implants were restored permanently following the
were instructed to chew predominantly on the oppo- 16-week healing period with a cement-retained all-
site side and to avoid hard foods. ceramic crown (Lava, 3M ESPE) supported by either a
titanium abutment (Ti Design 4.0, Astra Tech) or a pre-
Clinical and Radiographic Evaluations fabricated zirconium abutment (ZirDesign 4.0, Astra
Examinations were performed on the day of insertion, Tech). If the implant was in a molar location, a titanium
every 2 weeks for the first 16 weeks after implant place- abutment was used. If the implant was in a premolar lo-
ment, and at 1, 2, and 3 years. Implant stability was mea- cation, either the titanium abutment or the zirconium
sured at the implant level with the RFA device (Osstell, abutment was used, depending on implant angulation
Integration Diagnostics) at each visit by the first author and the availability of adequate thickness for the zirco-
up to 16 weeks. The transducer (SmartPeg Type 6, refer- nium abutment. An open-tray impression coping was
ence no. 100378, Integration Diagnostics) was calibrated used (Fixture Pick-up ST, Short, reference no. 22847,
prior to each use using an OsseoSpeed implant embed- Astra Tech) with polyvinyl siloxane impression material
ded in epoxy resin (Buehler) with a known ISQ. The heal- (Aquasil, Dentsply). All of the restorations were luted
ing abutment or provisional crown was removed and with the same cement (Relyx Unicem, 3M ESPE).
the SmartPeg placed via hand tightening 2 to 4 Ncm
onto the implant. RFA measurements were made twice Power Analysis and Sample Size Calculation
parallel to the implant and twice perpendicular to the Estimated samples sizes were based upon an estimated
implant in the arch owing to slight differences noted in within-treatment-group standard deviation of 5.0 for
a previous study caused by the differing densities of the the primary outcome variable—resonance frequency—
buccolingual plate of bone and the interradicular bone.32 measured using ISQ, two-sided hypothesis testing,
Previous recordings on the implant were not accessed and an overall level of type I error of .05 in conjunction
prior to RFA measurement to reduce observer bias. with a Bonferroni adjustment for three pairwise mul-
At each appointment, the implants were manually tiple comparisons of the loading groups. The number
tested for stability. The peri-implant marginal tissues of subjects needed to obtain 80% power to detect a
were evaluated using the Mombelli Index and the Apse difference of 6 ISQ between two subgroups is 80. The
score for inflammation levels, and the probing depth present study represents the 40 subjects treated at one
was measured in the mesiodistal and buccolingual di- of two centers. The remaining 40 are being evaluated
rections.33 The patient was asked about relative pain at another center, but because of changes in the im-
levels and, following placement of the provisional, the plant design, abutment connection, and drilling pro-
patient’s esthetic and functional satisfaction was de- tocols during the course of the study, the comparison
termined. Any implants that presented with pain, peri- group was not included in this analysis.
implant radiolucency, or clinical mobility were consid-
ered failures. If at any of the aforementioned visits, the Statistical Analysis
ISQ fell to 45 or lower, the implant was considered a Descriptive statistics were used to determine the distri-
potential failure and placed under unloaded healing bution of implants according to bone type, ITV, gender,
for the 12 weeks prior to repeat stability testing. and location. Mean ISQs and standard deviations were
calculated at all time points for the implants according
Radiographic Analysis to bone type and load type. The null hypothesis is that
Crestal bone height was assessed radiographically at the change in stability from baseline to 16 weeks is equal
baseline (implant placement), at 16 weeks, and at 1, between each pair of groups. A nonparametric statisti-
2, and 3 years postloading using standard periapical cal approach was applied, since the distribution of the
films and the long-cone paralleling technique. A Rinn data was unknown and could not be assumed to be nor-
posterior bite block (XCP, Dentsply) was indexed to the mal. A P value less than 5%, calculated by means of the
adjacent teeth and the opposing teeth with vinyl poly- Wilcoxon rank sum test (exact), was defined as statis-
siloxane (Regisil, Caulk). Each patient had their own tically significant and suggested a difference between
indexed Rinn holder to ensure that the angulation of groups, although adjustments for multiple comparisons
© 2012 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART OF MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Barewal et al
were not made. Baseline ISQ and ITV were compared Table 2 Implant Characteristics and Sites
for all implants, and correlation coefficients were as-
sessed using the Spearman rank test. Implants grouped Loading group
according to bone type were compared with respect Immediate Early Delayed Totals
to mean ITV using the Wilcoxon rank sum test, and a ITV
P value of .05 denoted a significant difference (Stat 0 to < 10 Ncm – – 7 7
Xact, version 6.2.0). 10 to < 20 Ncm – 11 2 13
20+ Ncm 8 6 6 20
Implant length
Results
11 mm 2 11 11 24
© 2012 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART OF MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Barewal et al
75
70
65 Types 1 and 2
Type 3
60
Type 4
55
50
Baseline 2 4 6 8 10 12 14 16
Time (wk)
75
70
65 Immediate loading
Early loading
60
Delayed loading
55
50
Baseline 2 4 6 8 10 12 14 16
Time (wk)
45
50
45
40
35 50
ITV (Ncm)
RFA (ISQ)
30
25
20 65
15
10
5 75
0
0 1 2 3 4
0 10 20 30 40 50
Bone type
ITV (Ncm)
Fig 5 Correlation of two quantitative measures of implant sta Fig 6 Correlation of two quantitative measures of implant sta
bility: ISQ and ITV. Spearman rank test: 0.4973; P = .0063. bility: ISQ and ITV with associated P value. Spearman rank test:
0.4973; P = .0063
© 2012 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART OF MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Barewal et al
Discussion
The current investigation sought to test the hypothesis old of 30 Ncm or more had been adopted for immedi-
that dental implant stability is minimally affected when ate loading, significant underpreparation of the surgical
physiologic loading is applied. The study was designed site would have been required; this would have created
as a prospective stratified randomized clinical trial with a subjective confounding variable that was controlled
strict inclusion and exclusion criteria to remove vari- in this protocol. As was shown in Table 4, type 3 bone
ables that would lead to uncertainty in the validity of showed the greatest variability in ITV, with 10 of the
the data. A single prosthodontist performed all stability 18 implants having a maximum ITV less than 20 Ncm
measurements with the Osstell device and all follow-up and two showing rotational mobility on placement.
examinations to control for observer bias. It is under- This is consistent with other studies.17,28,29 Determina-
stood that the primary stability of most implants in type tion of bone type is subject to interoperator variability
4 bone is unable to support occlusion in an unsplinted and difficulty in differentiation between intermediate
design. In this subject population, it was observed that bone types 2 and 3.35 The ITV, because it offers an ob-
the maximum ITV for implants in type 4 bone was 18 jective numeric representation of resistance to drilling,
Ncm and the minimum was 4 Ncm. An ITV of 20 Ncm— may therefore become the more relevant tool for com-
rather than bone type assessment by the surgeon— municating bone quality. A previous RCT using rough-
was used for inclusion in the immediate loading group, surfaced implants demonstrates that early loading leads
creating an ethical study-design threshold to allow im- to an acceptable survival rate regardless of the avail-
mediate loading of unsplinted implants in the posterior able bone type.36 This study confirmed that implants
region. There is a tremendous range in the suggested in all bone types were successfully loaded at 6 weeks
ITV cutoff for immediate loading of a single implant when ITV was used as the determinant for the timing
(from 30 to 60 Ncm).13,16,17,34 Ottoni et al suggested a of loading. In addition, two implants inserted in bone
torque value of 32 Ncm for immediate loading, as they classified as type 3 were immediately loaded success-
observed a high failure rate at 20 Ncm with the Frialit-2 fully when the ITV protocol was followed. The only im-
implant.34 The implant system used in the current study plant failure occurred in type 4 bone with ITV < 8.1 Ncm
has been noted to have lower ITV than other systems, (in the delayed loading group) and was removed at 10
likely as a result of its thread design. If a higher thresh- weeks following placement.
© 2012 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART OF MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Barewal et al
It is interesting to note that the ITV and baseline ISQ implants used in this study appeared to maintain sta-
measures were only weakly correlated. Da Cunha et bility during the peak of the resorptive phase of bone
al37 demonstrated similar findings with the TiUnite Mk healing (2 to 3 weeks). This is encouraging and may
III implant and no correlation of ITV and ISQ with the provide clinical support to earlier laboratory studies
machined Brånemark implant. Friberg et al29 found a regarding this device.41,42 This could also explain the
high degree of correlation between the ITV of the up- high success rate obtained in this study following early
per crestal third of the implant site and the resonance loading at 6 weeks of implants in type 4 bone.
frequency values upon insertion. The weak and incon- Underpreparation of the implant site, which is ac-
sistent correlation between ISQ and ITV can be attrib- complished by not following the standard drilling se-
uted to the fact that these tools measure two different quence indicated for a particular implant, has been
properties of the implant-to-bone connection. Inser- discussed as a means to improve primary stability.43
tion torque measures rotational resistance as the im- Although this may improve the ISQs and ITVs for an
plant is being placed and is dependent on mechanical implant and increase the confidence of the operator
properties of the bone such as density and hardness, in immediately loading the implant, it is not known
implant design, and site preparation.28 RFA, on the whether this will cause a greater resorptive effect,
other hand, measures the resonance of the implant in leading to a reduction in stability prior to secondary
bone after placement and is dependent on the axial bone formation. Maintenance of secondary stability
stiffness of the implant-to-bone area. In addition, it is derived from the remodeling of the implant interface
possible that the sensitivity of both instruments of the of an immediately loaded implant is equally important
implant-to-bone connection is not equal and therefore in reducing the risk of early implant failure.27 Further
the measures do not correlate strongly. Both rotational investigations are required to compare underprepara-
and axial stiffness are useful prognostic indicators of tion with standard preparation of implant sites and the
success with immediate loading and together provide effect of underpreparation on stability measurements
a better description of the primary stability of the im- and the success of immediately loaded single implants.
plant. It would be advantageous for the surgeon to A generalization from the results of this trial to clini-
have access to both ITVs and ISQs to assess risks of im- cal practice should be made with caution. In this trial,
mediate loading of an unsplinted implant. In this study, the inclusion criteria were strict (Table 1) and only pa-
the minimum ISQ for implants with an ITV ≥ 20 Ncm tients known to be ideal candidates for implant treat-
category was 67 and the maximum was 77 (mean ISQ ment were recruited, the clinical team was restricted to
of 72). It would seem reasonable based on the success one surgeon and one prosthodontist, and the opera-
achieved in this preliminary study that if the ITV was tors were highly experienced. On the other hand, a re-
at least 20 Ncm and the ISQ was 67 or higher, that a cent effectiveness-of-care study showed that minimal
standard-diameter implant of 11 or 13 mm in length complications with early and immediate loading oc-
could be immediately loaded. curred with the same implants supporting a range of
An interesting outcome of the study was that all im- prosthesis designs in a large effectiveness field trial.44
plant groups, when divided by time of loading or bone
type, showed a steady increase in stability (as mea-
sured by ISQ) over time. This is a remarkable difference, CONCLUSION
especially in type 4 bone, as compared to previous
studies. In a study of unloaded implants, a decrease in Following the protocol for this stratified randomized
the mean stability measurement occurred at 3 weeks clinical trial, no differences in bone levels were ob-
within each bone group, with the least stability seen in served after 3 years of loading for all implants in the
type 4 bone.32 Similar results were seen by Valderrama three loading groups. This indicates that a minimal in-
et al38 with the SLA Active implant (Institut Straumann), sertion torque of 20 Ncm may be an important thresh-
which features a roughened surface. Balshi and co- old determinant to consider immediate loading of
workers observed, with Brånemark System implants single-tooth implants in the posterior region. Limita-
(TiUnite Mk III and Mk IV), a decrease in ISQ for the tions of this study are the sample size and the com-
first 30 days.39 Al-Nawas et al,40 in looking at sand- plexity of the research design needed to address the
blasted/acid-etched, titanium plasma–sprayed, and research question. The observed lack of significant dif-
Mk III and IV implants, noted a decrease in ISQ during ference may become significant with a greater sample
the first 8 weeks of healing. Although primary stability size, but the trend of uniformity shown in this study
is assumed to be the most important determinant of suggests that any difference, while statistically signifi-
success with immediate loading, the maintenance of cant, may have limited impact from a clinical perspec-
implant stability during the transformation of primary tive. A second measure of stability is recommended
to secondary bone contact is equally important. The and may be provided with a measurement of implant
© 2012 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART OF MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Barewal et al
and bone stiffness (eg, implant stability quotient [ISQ] 14. Maló P, Rangert B, Dvarsater L. Immediate function of Brånemark
implants in the esthetic zone: A retrospective clinical study with 6
derived from the Osstell device). A baseline reading of months to 4 years of follow-up. Clin Implant Dent Relat Res 2000;2:
greater than 70 ISQ would increase the operator con- 138–146.
fidence in determination of immediately loading the 15. Esposito M, Grusovin MG, Willings M, Coulthard P, Worthington
HV, Esposito M. Interventions for replacing missing teeth: Different
dental implant. The maintenance of increasing stabil- times for loading dental implants. Cochrane Database Syst Rev
ity levels over time is encouraging and supports the 2009;1:CD003878.
hypothesis that the timing and method of load ap- 16. Testori T, Galli F, Capelli M, Zuffetti F, Esposito M. Immediate
non-occlusal versus early loading of dental implants in partially
plication to the implants was within their physiologic edentulous patients: 1-year results from a multicenter, randomized
capacity. Because of variations in in geometry and sur- controlled clinical trial. Int J Oral Maxillofac Implants 2007;22:
face technology, primary stability levels and loading 815–822.
17. Calandriello R, Tomatis M, Rangert B. Immediate functional loading
protocols will vary according to the implant type. of Brånemark system wide-platform TiUnite implants: An interim
report of a prospective open-ended clinical multicenter study. Clin
Implant Dent Relat Res 2003;5(suppl 1):10–20.
18. Güncü MB, Aslan Y, Tümer C, Güncü G, Uysal S. In-patient compari-
ACKNOWLEDGMENTS son of immediate and conventional loaded implants in mandibular
molar sites within 12 months. Clin Oral Implants Res 2008;19:
Many thanks to Mikael Åström, MSc, PhL, Statistical Science 335–341.
Director, StatCons, for his statistical analysis. The contributions 19. Schincaglia GP, Marzola R, Fazi G, Scapoli C, Scotti R. Replacement
of Ruth Bourke, CDT, RE Bourke Dental Lab, and Mike Gieseman, of mandibular molars with single-unit restorations supported by
CDT, Midwest Dental Lab, are greatly appreciated. This study wide-body implants: Immediate versus delayed loading. A random-
ized controlled study. Int J Oral Maxillofac Implants 2008;23:
was supported by a grant from AstraTech.
474–480.
20. Gapski R, Wang HL, Mascarenhas P, Lang N. Critical review of im-
mediate implant loading. Clin Oral Implants Res 2003;14:515–527.
REFERENCES 21. Roberts WE, Smith R, Zilberman Y, Mozsary P, Smith R. Osseous
adaptation to continuous loading of rigid endosseous implants.
Am J Orthod 1984;86:95–111.
1. Brånemark P-I, Hansson BO, Adell R, et al. Osseointegrated implants 22. Ericsson I, Nilson H, Lindh T, Nilner K, Randow K. Immediate
in the treatment of the edentulous jaw. Experience from a 10-year functional loading of Brånemark single tooth implants. Clin Oral
period. Scan J Plast Reconstr Surg 1977;16(suppl):1–132. Implants Res 2000;11:26–33.
2. Szmukler-Moncler S, Piattelli A, Favero GA, Dubruille J-H. Consid- 23. Meredith N, Alleyne D, Cawley P. Quantitative determination of the
erations preliminary to the application of early and immediate stability of the implant-tissue interface using resonance frequency
loading protocols in dental implantology. Clin Oral Implants Res analysis. Clin Oral Implants Res 1996;7:261–267.
2000;11:12–25. 24. Meredith N, Book K, Friberg B, Jemt T, Sennerby L. Resonance fre-
3. Lederman PD. Stegprothetische Versorgung des Zahnlosen Unter- quency measurements of implant stability in vivo. A cross-sectional
kierfers mit Hilfe Plasmabeschichteten Titanschraubimplantaten and longitudinal study of resonance frequency measurements on
[in German]. Dtsch Zahnärtzl Zeit 1979;34:907–911. implants in the edentulous and partially dentate maxilla. Clin Oral
4. Babbush CA, Kent J, Misiek D. Titanium plasma-sprayed (TPS) screw Implants Res 1997;8:226–233.
implants for the reconstruction of the edentulous mandible. J Oral 25. Atsumi M, Park S, Wang H. Methods used to assess implant stability:
Maxillofac Surg 1986;44:274–282. Current status. Int J Oral Maxillofac Implants 2007;22:743–754.
5. Tarnow DP, Emtiaz S, Classi A. Immediate loading of threaded 26. Meredith N. Assessment of implant stability as a prognostic deter-
implants at stage 1 surgery in edentulous arches: Ten consecutive minant. Int J Prosthodont 1998;11:491–501.
case reports with 1 to 5 year data. Int J Oral Maxillofac Implants 27. Stanford C, Brand R. Toward an understanding of implant occlusion
1997;12:319–324. and strain adaptive bone modeling and remodeling. J Prosthet
6. Ericsson I, Nilson H, Lindh T, Nilner K, Randow K. Immediate Dent 1999;81:553–561.
functional loading of Brånemark single tooth implants. Clin Oral 28. Johansson P, Strid K. Assessment of bone quality from cutting resis-
Implants Res. 2000;11:26-33. tance during implant surgery. Int J Oral Maxillofac Implants 1994;9:
7. Randow K, Ericsson I, Nilner K, Petersson A, Glantz PO. Immediate 279–288.
functional loading of Brånemark implants. An 18-month clinical 29. Friberg B, Sennerby L, Meredith N, Lekholm U. A comparison be-
follow-up study. Clin Oral Implants Res 1999;10:8–15. tween cutting torque and resonance frequency measurements of
8. Horiuchi K, Uchida H, Yamamoto K, Sugimura M. Immediate loading maxillary implants. A 20-month clinical study. Int J Oral Maxillofac
of Brånemark system implants following placement in edentulous Surg 1999;28:297–303.
patients: A clinical report. Int J Oral Maxillofac Implants 2000;12: 30. Lekholm U, Zarb G. Patient selection and preparation. In: Brånemark
319–324. P-I, Zarb GA, Albrektsson T (eds). Tissue-Integrated Prostheses: Os-
9. Mericske-Stern R, Zarb GA. Overdentures: An alternative im- seointegration in Clinical Dentistry. Chicago: Quintessence, 1985:
plant methodology for edentulous patients. Int J Prosthodont 199–209.
1993;6:203–208. 31. Friberg B, Sennerby L, Roos J, Johansson P, Strid CG, Lekholm U.
10. Şahin S, Çehreli M, Yalçin E. The influence of functional forces on Evaluation of bone density using cutting resistance measure-
the biomechanics of implant-supported prostheses—A review. ments and microradiography: An in vitro study in pig ribs. Clin Oral
J Dent 2002;30:271–282. Implants Res 1995;6:164–171.
11. Jaffin R, Berman C. The excessive loss of Brånemark fixtures in type 32. Barewal RM, Oates TW, Meredith N, Cochran DL. Resonance
IV bone: A 5-year analysis. J Periodontol 1991;62:2–4. frequency analysis of implant stability in vivo on implants with a
12. Donati M, La Scala V, Billi M, Di Dino B, Torrisi P, Berglundh T. Imme- sandblasted and acid-etched surface. Int J Oral Maxillofac Implants
diate functional loading of implants in single tooth replacement: A 2003;18:641–651.
prospective clinical multicenter study. Clin Oral Implants Res 2008; 33. Salvi G, Lang N. Diagnostic parameters for monitoring peri-implant
19:740–748. conditions. Int J Oral Maxillofac Implants 2004;19(suppl):116–127.
13. Wöhrle PS. Single tooth replacement in the aesthetic zone with 34. Ottoni JM, Olivieri ZF, Mansini R, Cabral AM. Correlation between
immediate provisionalization: Fourteen consecutive case reports. placement torque and survival of single-tooth implants. Int J Oral
Pract Periodontics Aesthet Dent 1998;9:1107–1114. Maxillofac Implants 2005;20:769–776.
© 2012 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART OF MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Barewal et al
35. Trisi P, Rao W. Bone classification: Clinical-histomorphometric com- 40. Al-Nawas B, Wagner W, Grötz KA. Insertion torque and resonance
parison. Clin Oral Implants Res 1999;10:1–7. frequency analysis of dental implant systems in an animal model
36. Ganeles J, Zöllner A, Jackowski J, ten Bruggenkate C, Beagle J, with loaded implants. Int J Oral Maxillofac Implants 2006;21:
Guerra F. Immediate and early loading of Straumann implants with 726–732.
a chemically modified surface (SLActive) in the posterior mandible 41. Berglundh T, Abrahamsson I, Albouy J-P, Lindhe J. Bone healing at
and maxilla: 1-year results from a prospective multicenter study. implants with a fluoride-modified surface: An experimental study
Clin Oral Implants Res 2008;19:1119–1128. in dogs. Clin Oral Implants Res 2007;18:147–152.
37. da Cunha HA, Francischone CE, Filho HN, de Oliveira RC. A compari- 42. Ellingsen JE, Johansson CB, Wennerberg A, Holmen A. Improved
son between cutting torque and resonance frequency in the as- retention and bone-to-implant contact with fluoride-modified
sessment of primary stability and final torque capacity of standard titanium implants. Int J Oral Maxillofac Implants 2004;19:659–666.
and TiUnite single-tooth implants under immediate loading. Int J 43. Norton M. The influence of insertion torque on the survival of
Oral Maxillofac Implants 2004;19:578–585. immediately placed and restored single-tooth implants. Int J Oral
38. Valderrama P, Oates T, Jones A, Simpson J, Schoolfield J, Cochran D. Maxillofac Implants 2011;26:1333–1343.
Evaluation of two different resonance frequency devices to detect 44. Stanford CM, Wagner W, Rodriguez Y, et al. Evaluation of the ef-
implant stability: A clinical trial. J Periodontol 2007;78:262–272. fectiveness of dental implant therapy in a practice-based network
39. Balshi SF, Allen FD, Wolfinger GJ, Balshi TJ. A resonance frequency (FOCUS). Int J Oral Maxillofac Implants 2010;25:367–373.
analysis assessment of maxillary and mandibular immediately
loaded implants. Int J Oral Maxillofac Implants 2005;20:584–594.
© 2012 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART OF MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.