Sop Yumizen H550
Sop Yumizen H550
Sop Yumizen H550
DIAGOMET
AUTOMATED BLOOD HAEMATOLOGY
LABORATORIES, DELHI
COUNT ANALYSER
It is the enumeration of cellular elements of blood, evaluation of red cells indices and
determination of cell morphology by means of stained smears. The various parameters generated
from the automated hematology analyzer are of clinical importance to differentiate between
diseases of hematological abnormalities. Reference values vary across the life cycle and between
sexes and differ in different age groups and in various disease conditions.
2. Principle:
WBC
The WBC detection principal is based on the Double Hydrodynamic Sequential System
“DHSS” which allows a linear flow of the cell through the light path (Horiba Medical Patent
technology)
Designation Signature
Prepared by: Senior Lab Technician
Reviewed & approved by: Quality Manager
Issued by: Technical Manager
YUMIZEN H550
DIAGOMET
AUTOMATED BLOOD HAEMATOLOGY
LABORATORIES, DELHI
COUNT ANALYSER
In the optical chamber during the acquisition of the LMNE matrix. (DHSS)
RBC/PLT
Measurement of impedance variation generated by the passage of RBC cells through a
calibrated micro aperture. The specimen is diluted in ABX Diluent (current conductor) and pulled
through the calibrated micro-aperture. Two electrodes are placed on either side of the aperture.
Electric current passes through the electrodes continuously. When the cell passes through the
aperture, electric resistance between the two electrodes increases proportionately to the cell
volume.
HB
HCT
All the RBC pulses are grouped into various sizes. Each group pulse height is then
averaged. All the pulse height averages are then averaged one final time for a mean average of
all the RBC pulse heights. This function is a numeric integration of the MCV, The HCT results
are given as a percentage of this integration
.
MCV: calculated from RBC and HCT
MCH: Calculated from Hb and RBC
MCHC: calculated from Hb and Hct
RDW: Calculated from deviation of RBC Distribution Width
Designation Signature
Prepared by: Senior Lab Technician
Reviewed & approved by: Quality Manager
Issued by: Technical Manager
YUMIZEN H550
DIAGOMET
AUTOMATED BLOOD HAEMATOLOGY
LABORATORIES, DELHI
COUNT ANALYSER
3. Performance specification:
Measurement Minimum detection
Linearity
Range range
WBC 0.0 – 300 x 103 0.5 – 15.0 x 103
0.3 (300 Cells / cu mm)
( X 103 Cells / μL) cells/μL cells/Μl
RBC 0.0 – 8.00 x 106 0.2 – 6.00 x 106 0.2 ( 0.2 million cells /
( X 106 Cells / μL) cells/μL cells/μL cumm)
Platelets 0.0 – 2500 x 103 10 – 500 x 103 10 (10,000 cells / cu
( X 103 Cells / μL) cells/μL cells/μL mm)
Hb (g/dL) 0.0 – 24.0 g/dL 2.5 -18.0 g/dL 2.5 g /dL
4. Primary Sample:
1. Use whole blood as sample collected in K3 EDTA
2. Collect 2 ml blood in lavender top vacutainer tube (Containing 3.6 mg K 2 EDTA) or
in tubes containing 40 μL of 10 % K2 EDTA per 2 ml of blood).
3. Mix the specimen thoroughly but gently in a hemomixer for 3 – 5 minutes
immediately prior to aspiration of the sample by the machine/ by 3- 5 gentle inversion
and rotation between the palms of hands.
4. Process the samples within 4 hours of collection.
5. Do not use samples which are clotted, containing fibrin microclots, decomposed or
contaminated sample.
6. Reject clotted samples and samples without ID.
5. Type of container and additive: Use a lavender color topped vacutainer tube
(containing 3.6 mg K3 EDTA)
6. Consumables/ Reagents:
1. Horiba Series reagent system,
2. ABX diluents 20L
3. Whitediff 1L
Designation Signature
Prepared by: Senior Lab Technician
Reviewed & approved by: Quality Manager
Issued by: Technical Manager
YUMIZEN H550
DIAGOMET
AUTOMATED BLOOD HAEMATOLOGY
LABORATORIES, DELHI
COUNT ANALYSER
4. ABX Cleaner 1L
7. Equipment: HORIBA YUMIZEN H500
8. Step by Step Procedure: HORIBA YUMIZEN – VI PART
Switch on CPU
“Login”
Startup
Control Run
Sample Analysis
START (Enter)
Designation Signature
Prepared by: Senior Lab Technician
Reviewed & approved by: Quality Manager
Issued by: Technical Manager
YUMIZEN H550
DIAGOMET
AUTOMATED BLOOD HAEMATOLOGY
LABORATORIES, DELHI
COUNT ANALYSER
2. Wait for Approximately 3 min for the initialisation process to be completed. (Only
Sart-up)
9. Quality Control Procedure:
1. Run all three levels of an assayed QC sample (Horiba Three Level Quality Control
samples - Low, Normal and High) and verify if the values are close to the Mean and
within the reference range mentioned in the QC product insert.
2. If the QC is not within the reference range even for any one level of QC, perform
another cleaning procedure and rerun that level of QC.
10. Reference Range:
MCHC 32.0-36.0%
11. Potential Sources of variability :
1. The Following factors will affect cell counting in an automated hematology analyzer
2. Visible clots Fibrin micro clots or cryoprecipitate in the specimen
3. Inadequate lysis of RBC
4. Inadequate dilution of specimen and inadequate calibration of the analyzer
5. Non Homogenous suspension of blood cells in the diluting fluid
Designation Signature
Prepared by: Senior Lab Technician
Reviewed & approved by: Quality Manager
Issued by: Technical Manager
YUMIZEN H550
DIAGOMET
AUTOMATED BLOOD HAEMATOLOGY
LABORATORIES, DELHI
COUNT ANALYSER
Designation Signature
Prepared by: Senior Lab Technician
Reviewed & approved by: Quality Manager
Issued by: Technical Manager