Download as pdf or txt
Download as pdf or txt
You are on page 1of 8

See

discussions, stats, and author profiles for this publication at: https://www.researchgate.net/publication/270003358

Lens Dose in Routine Head CT: Comparison of


Different Optimization Methods With
Anthropomorphic Phant....

Article in American Journal of Roentgenology · January 2015


DOI: 10.2214/AJR.14.12763 · Source: PubMed

CITATIONS READS

9 168

5 authors, including:

Ulla Nikupaavo Touko Kaasalainen


Helsinki University Central Hospital Helsinki University Central Hospital
1 PUBLICATION 9 CITATIONS 22 PUBLICATIONS 105 CITATIONS

SEE PROFILE SEE PROFILE

Vappu Reijonen Mika Kortesniemi


Helsinki University Central Hospital Helsinki University Central Hospital
23 PUBLICATIONS 216 CITATIONS 78 PUBLICATIONS 1,345 CITATIONS

SEE PROFILE SEE PROFILE

Some of the authors of this publication are also working on these related projects:

Organ segmentation for patient specific dose estimates based on Monte Carlo simulations View
project

Clinical and preclinical dosimetry in molecular radiotherapy View project

All content following this page was uploaded by Ulla Nikupaavo on 29 June 2015.

The user has requested enhancement of the downloaded file.


M e d i c a l P hy s i c s a n d I n f o r m a t i c s • O r i g i n a l R e s e a r c h

Nikupaavo et al.
Lens Dose in Head CT

Medical Physics and Informatics


Original Research
Downloaded from www.ajronline.org by HUS-KUVANTAMINEN on 05/18/15 from IP address 193.166.253.30. Copyright ARRS. For personal use only; all rights reserved

Lens Dose in Routine Head CT:


Comparison of Different
Optimization Methods With
Anthropomorphic Phantoms
Ulla Nikupaavo1,2 OBJECTIVE. The purpose of this study was to study different optimization methods for
Touko Kaasalainen1,3 reducing eye lens dose in head CT.
Vappu Reijonen1 MATERIALS AND METHODS. Two anthropomorphic phantoms were scanned with
Sanna-Mari Ahonen2 a routine head CT protocol for evaluation of the brain that included bismuth shielding, gan-
Mika Kortesniemi1,3 try tilting, organ-based tube current modulation, or combinations of these techniques. High-
sensitivity metal oxide semiconductor field effect transistor dosimeters were used to measure
Nikupaavo U, Kaasalainen T, Reijonen V, local equivalent doses in the head region. The relative changes in image noise and contrast
Ahonen SM, Kortesniemi M were determined by ROI analysis.
RESULTS. The mean absorbed lens doses varied from 4.9 to 19.7 mGy and from 10.8 to
16.9 mGy in the two phantoms. The most efficient method for reducing lens dose was gantry
tilting, which left the lenses outside the primary radiation beam, resulting in an approximately
75% decrease in lens dose. Image noise decreased, especially in the anterior part of the brain.
The use of organ-based tube current modulation resulted in an approximately 30% decrease in
lens dose. However, image noise increased as much as 30% in the posterior and central parts of
the brain. With bismuth shields, it was possible to reduce lens dose as much as 25%.
CONCLUSION. Our results indicate that gantry tilt, when possible, is an effective
method for reducing exposure of the eye lenses in CT of the brain without compromising im-
age quality. Measurements in two different phantoms showed how patient geometry affects
the optimization. When lenses can only partially be cropped outside the primary beam, or-
gan-based tube current modulation or bismuth shields can be useful in lens dose reduction.

I
n CT of the head, the eye lens is Analysis of data on occupational exposure
always exposed to scattered radi- and on patients undergoing diagnostic scan-
ation and, depending on the scan ning of the head has had mixed results. Klein
Keywords: anthropomorphic phantom, CT optimization, geometry, often to the direct et al. [17] reported a significant correlation
image quality, lens dose, MOSFET beam. Results of epidemiologic studies of between CT scanning and increased risk of
populations with low-dose radiation expo- nuclear sclerosis and posterior subcapsular
DOI:10.2214/AJR.14.12763
sure have suggested that the lens is more sen- opacity. Hourihan et al. [18], however, found
Received February 24, 2014; accepted after revision sitive to ionizing radiation than has previous- no evidence of the prevalence of any type of
March 31, 2014. ly been assumed and that the process of cataract in people with a history of head CT.
1
radiation-induced cataract formation may However, a 2013 study by Yuan et al. [19]
HUS Medical Imaging Center, Helsinki University Central
even be stochastic without a threshold dose corroborated the connection between in-
Hospital, POB 340 (Haartmaninkatu 4), 00290 Helsinki,
Finland. Address correspondence to U. Nikupaavo [1–3]. Therefore, the International Commis- creased risk of cataract formation and CT of
(ulla.nikupaavo@hus.fi). sion on Radiological Protection has reevalu- the head and neck region. The risk seems to
ated the equivalent dose limit of the eye lens to increase gradually with repeated CT studies.
2
University of Oulu, Institute of Health Sciences, ionizing radiation by lowering the recom- The radiation exposure of the lens and
Oulu, Finland.
mended acute dose threshold for lens effects of other selective radiation-sensitive tissues
3
Department of Physics, University of Helsinki, (cataracts and opacities) from 2–8 Gy to can be reduced by use of technical solutions
Helsinki, Finland. 0.50 Gy [4, 5]. Measurements in phantoms such as organ-based tube current modula-
and patients have varied considerably in ab- tion (OBTCM) [6–8], local exterior shield-
AJR 2015; 204:117–123 sorbed dose of the lens. The variation, ranging ing (bismuth shields) [7, 9, 10, 15, 16], and
0361–803X/15/2041–117
from a few to approximately 100 mGy per optimization of the gantry tilt angle [10, 11,
scan, depends on the scanner, imaging tech- 15, 20]. Users can also reduce radiation ex-
© American Roentgen Ray Society nique, and optimization methods used [6–16]. posure and improve image quality by taking

AJR:204, January 2015 117


Nikupaavo et al.

steps such as centering patients properly in


the scan isocenter [21, 22]. Iterative recon-
struction techniques have enabled major re-
ductions in total dose, 20–30% in the case of
Downloaded from www.ajronline.org by HUS-KUVANTAMINEN on 05/18/15 from IP address 193.166.253.30. Copyright ARRS. For personal use only; all rights reserved

head CT, while diagnostic image quality is


maintained [23–25].
Our aim was to compare a group of meth-
ods developed to reduce the dose to the eye
lens in head CT. We quantitatively assessed
their effect on dose distribution and image
quality in different anatomic locations by
performing measurements in two anthropo-
morphic phantoms.

Materials and Methods


Dose Measurements and Calculations
We imaged two tissue-equivalent anthropo-
morphic phantoms—ATOM model 702-D (CIRS)
and RANDO (The Phantom Laboratory)—in
the head-first supine position (Fig. 1) with a 128- A B
MDCT scanner (Somatom Definition AS+, Sie- Fig. 1—Photographs show anthropomorphic phantoms used in study.
mens Healthcare). We performed CT head scan- A, ATOM model 702-D (CIRS).
B, RANDO (Phantom Laboratory).
ning in helical mode with eight settings: reference
scan without an optimization method; gantry tilted cording to the clinical practice; and bismuth shield tor width of 38.4 mm). Scan range was set for a rou-
according to clinical practice (baseline from skull already set on the eyes in scout imaging (Fig. 2). tine head CT examination (from skull base to ver-
base to radix nasi); gantry tilted at an angle one The following scan parameters were identical for tex). Volume CT dose index varied from 46.06 to
half of that used in clinical practice; 0.06-mm lead all studied scan settings and defined as applied in 47.02 mGy for the ATOM phantom and from 44.38
equivalent bismuth shield (AttenuRad Radiation clinical practice: 120 kV, tube current modulation to 45.35 mGy for the RANDO phantom.
Protection, F&L Medical Products) on the eyes; with quality reference tube current–time product We measured local organ doses with high-sen-
both a bismuth shield and gantry tilted according of 410 mAs, 1-second rotation time, pitch of 0.6, sitivity TN-1002RD metal oxide semiconduc-
to clinical practice; OBTCM (X-CARE, Siemens and detector configuration of 128 × 0.6 mm (with tor field effect transistor (MOSFET) dosimeters
Healthcare); both OBTCM and gantry tilted ac- double z-sampling corresponding to a total detec- (TN-1002RD, Best Medical) with high bias set-
tings. Before the measurements were made, the
dosimeters were calibrated in the CT beam on an
axial scan obtained with 120-kVp tube voltage.
In calibration, the reference air kerma was mea-
sured with a CT pencil ionization chamber (Ray-
Safe Xi, Unfors RaySafe), and the calibration
factor was defined separately for each MOSFET
dosimeter. One of the dosimeters was discarded
because of observed high instability. Thus a total
of nine MOSFET dosimeters were inserted in the
phantom head in different locations: left mandi-
ble, both lenses, skull base, posterior right and left
hemispheres, occipital bone, anterior right hemi-
sphere, and anterior aspect of the central part of
the brain. The active part and epoxy bulb of the
MOSFET dosimeters were located slightly be-
low the midpoint of each phantom layer, the tips
pointing toward the anterior part of the phantom.
We read the MOSFETs after each acquisition, and
scanning was performed five times for each set-
ting. Average absorbed doses (in milligrays) were
calculated for each organ in each scanning setting.

Image Analysis
Fig. 2—Scan settings used for head CT and geometric profiles of phantoms. Top row, gantry tilting angles and
reference scan setting in two phantoms (left, ATOM, CIRS; right, RANDO, Phantom Laboratory). Bottom row, Because the MOSFET dosimeter lead wires
scan settings with bismuth shields. produce metal artifacts and interfere with image

118 AJR:204, January 2015


Downloaded from www.ajronline.org by HUS-KUVANTAMINEN on 05/18/15 from IP address 193.166.253.30. Copyright ARRS. For personal use only; all rights reserved

TABLE 1: Absorbed Organ Doses in Milligrays and Relative Change Compared With Reference Scanning of ATOM (CIRS) Phantom
Lens
Posterior Right Anterior Part of Posterior Left Anterior Right
Scan Setting Left Mandible Skull Base Hemisphere Central Brain Occipital Bone Hemisphere Hemisphere Right Left
Reference 3.1 ± 0.2 (0.0) 18.1 ± 0.3 (0.0) 20.1 ± 0.9 (0.0) 15.4 ± 0.3 (0.0) 16.9 ± 1.0 (0.0) 18.9 ± 0.4 (0.0) 17.5 ± 0.7 (0.0) 20.1 ± 1.3 (0.0) 19.2 ± 1.5 (0.0)
Bismuth shield after scout imaging 3.1 ± 0.2 (0.0) 18.0 ± 0.5 (–0.8) 20.3 ± 0.5 (1.1) 15.2 ± 0.6 (–1.3) 16.7 ± 0.8 (–0.8) 19.2 ± 1.3 (1.6) 17.0 ± 0.5 (–3.0) 14.9 ± 0.1 (–26.1) 15.2 ± 1.5 (–21.2)
Bismuth shield before scout imaging 3.5 ± 0.4 (13.8) 18.1 ± 0.5 (–0.1) 19.7 ± 1.0 (–2.2) 15.0 ± 0.5 (–2.5) 15.8 ± 0.8 (–6.2) 18.7 ± 0.7 (–1.1) 17.0 ± 0.8 (–3.0) 15.1 ± 1.2 (–24.8) 14.5 ± 1.2 (–24.4)
Bismuth shield with gantry tilt 2.2 ± 0.2 (–29.5) 18.3 ± 0.4 (1.2) 19.2 ± 0.5 (–4.5) 15.8 ± 0.8 (3.0) 15.5 ± 0.9 (–8.3) 18.9 ± 0.8 (0.4) 17.4 ± 0.8 (–0.5) 5.0 ± 0.4 (–74.9) 4.8 ± 0.6 (–74.9)
Gantry tilt 2.1 ± 0.2 (–31.1) 17.6 ± 0.7 (–2.7) 19.7 ± 1.1 (–2.2) 15.8 ± 0.5 (2.7) 15.5 ± 1.0 (–8.4) 19.2 ± 0.9 (1.6) 17.1 ± 0.5 (–2.0) 5.1 ± 0.2 (–74.5) 4.7 ± 0.2 (–75.6)
Gantry tilt with one-half angle 2.4 ± 0.2 (–20.9) 17.8 ± 0.6 (–1.6) 20.0 ± 0.8 (–0.8) 15.5 ± 0.3 (1.0) 15.3 ± 0.7 (–9.6) 18.3 ± 0.5 (–2.9) 17.3 ± 0.5 (–1.0) 16.0 ± 2.0 (–20.6) 15.3 ± 2.3 (–20.3)
OBTCM 3.5 ± 0.3 (14.5) 21.2 ± 0.4 (16.9) 22.8 ± 0.6 (13.5) 14.7 ± 0.2 (–4.2) 18.8 ± 1.2 (11.5) 20.2 ± 1.3 (7.2) 15.7 ± 0.5 (–10.3) 13.8 ± 0.4 (–31.3) 13.1 ± 1.1 (–32.1)
OBTCM with gantry tilt 2.4 ± 0.2 (–23.8) 20.4 ± 0.8 (12.8) 21.6 ± 0.7 (7.4) 14.1 ± 0.6 (–8.3) 17.7 ± 0.8 (5.1) 19.9 ± 0.6 (5.6) 14.9 ± 0.9 (–14.9) 6.2 ± 0.6 (–69.1) 5.4 ± 0.4 (–71.9)
Note—Values are mean ± SD. Values in parentheses are relative change as percentage. Negative relative change value indicates dose reduction compared with reference scan setting. OBTCM = organ-based
tube current modulation.

TABLE 2: Absorbed Organ Doses in Milligrays and Relative Change Compared With Reference Scan in RANDO Phantom (Phantom Laboratory)
Lens
Posterior Right Anterior Part of Posterior Left Anterior Right
Scan Setting Left Mandible Skull Base Hemisphere Central Brain Occipital Bone Hemisphere Hemisphere Right Left
Reference 3.8 ± 0.2 (0.0) 16.4 ± 0.7 (0.0) 17.6 ± 0.6 (0.0) 18.1 ± 0.7 (0.0) 15.0 ± 1.4 (0.0) 17.1 ± 0.9 (0.0) 18.0 ± 0.6 (0.0) 17.7 ± 1.8 (0.0) 16.1 ± 1.2 (0.0)
Bismuth shield after scout imaging 3.9 ± 0.5 (3.5) 16.9 ± 0.5 (2.9) 18.5 ± 0.4 (5.2) 17.9 ± 0.5 (–1.2) 14.8 ± 1.6 (–1.5) 17.1 ± 0.7 (0.0) 17.6 ± 0.6 (–2.6) 16.3 ± 1.3 (–7.6) 14.1 ± 1.1 (–12.3)
Bismuth shield before scout imaging 4.4 ± 0.4 (15.1) 16.8 ± 0.4 (2.8) 17.5 ± 1.0 (–0.3) 17.6 ± 0.5 (–2.8) 15.0 ± 1.0 (–0.4) 17.2 ± 0.9 (0.4) 17.4 ± 0.4 (–3.7) 14.5 ± 1.2 (–17.9) 15.2 ± 1.2 (–15.5)
Bismuth shield with gantry tilt 3.1 ± 0.3 (–18.9) 17.8 ± 0.4 (8.9) 18.3 ± 0.4 (4.2) 17.7 ± 0.5 (–2.4) 14.4 ± 1.2 (–4.5) 17.5 ± 0.6 (2.2) 18.3 ± 0.5 (1.4) 12.9 ± 0.7 (–27.2) 10.4 ± 0.7 (–35.3)
Gantry tilt 2.9 ± 0.1 (–22.0) 18.4 ± 0.3 (12.3) 18.0 ± 1.2 (2.5) 17.7 ± 0.7 (–2.2) 13.7 ± 1.2 (–8.9) 17.0 ± 1.0 (–0.9) 18.3 ± 0.7 (1.7) 14.8 ± 1.7 (–16.1) 12.9 ± 1.5 (–19.7)
Lens Dose in Head CT

Gantry tilt with one-half angle 3.1 ± 0.2 (–18.7) 17.7 ± 1.1 (7.8) 17.7 ± 0.5 (1.0) 17.9 ± 0.4 (–1.0) 13.9 ± 1.6 (–7.7) 17.0 ± 0.7 (–0.8) 17.9 ± 0.5 (–0.9) 16.5 ± 1.8 (–6.7) 14.9 ± 1.3 (–7.6)
OBTCM 4.6 ± 0.3 (21.4) 19.0 ± 0.7 (16.0) 19.3 ± 0.5 (9.9) 17.3 ± 0.3 (–4.5) 18.1 ± 1.6 (20.1) 19.6 ± 0.5 (14.5) 16.5 ± 0.4 (–8.5) 14.4 ± 0.4 (–18.6) 11.3 ± 0.5 (–30.0)
OBTCM with gantry tilt 3.5 ± 0.3 (–9.6) 20.9 ± 0.8 (27.6) 20.6 ± 0.6 (17.1) 17.7 ± 0.5 (–2.3) 15.9 ± 1.8 (5.6) 18.9 ± 1.0 (10.4) 17.3 ± 0.6 (–4.0) 11.3 ± 0.7 (–36.3) 10.4 ± 1.2 (–35.4)
Note—Values are mean ± SD. Values in parentheses are relative change as percentage. Negative relative change value indicates dose reduction compared with reference scan setting. OBTCM = organ-based
tube current modulation.

Results
Dosimetric Results
quality measurements.

National Institutes of Health).

to 22.8 mGy in the case of the ATOM

phantom and from 10.8 to 16.9 mGy


relative changes compared with the
from the ATOM phantom scans for image

interval. We performed the image analysis


el 2 (Siemens Healthcare), which is used at
ed the relative changes. We studied image
curacy of placement of the phantom for

AJR:204, January 2015 119


Depending on the scanning setting
absorbed organ doses varied from 2.2
reference scanning performed without
The size of ROI 1 was approximately 450
lobes, and ROIs 4 and 5 in the expected po-
right cerebellum, ROIs 2 and 3 in the ex-
particular clinical significance for ROIs:
of air gaps between the phantom layers,
the RANDO phantom was poor because
try scans was ± 2 mm in the x, y, and z di-

used, the mean lens dose varied


phantom and from 3.1 to 20.9 mGy
mm2. We compared image noise and con-
mm2, and the size of ROIs 2–5 was 300
sitions of the basal ganglia nuclei (Fig. 3).
pected positions of the anterior temporal
ROI 1 was located in the region of the
measuring 1 SD of the CT number in five
suring the mean CT number and noise by
We determined image contrast by mea-
the image quality as opposed to dosime-
the dosimeters. We estimated that the ac-
plugs. Each time, the scan parameters
quality analysis (noise, contrast), we re-

for the RANDO phantom. In the case


any optimization methods. The mean
in each scanning setting and their
results of the dose measurements
In Tables 1 and 2, we present the
with ImageJ software (version 1.48, U.S.
with 4-mm thickness and a 4-mm interslice
our clinic. Axial slices were reconstructed
and iterative reconstruction with Safire lev-
quality using both filtered back projection
trast of the reference images and calculat-
trast at each setting with the noise and con-
different ROIs. We selected locations of
and we decided to use only image data
rections. The image quality in the case of
simeter holes filled with tissue-equivalent
without the dosimeters and with the do-
peated all scans with different settings

from 4.9 to 19.7 mGy for the ATOM


in the case of the RANDO phantom.
were identical to the scans obtained with
Nikupaavo et al.

TABLE 3: Mean Image Noise (1 SD HU) and Relative Change Compared With Reference Setting in Five ROIs in
­Clinically Significant Areas
ROI 1 ROI 2 ROI 3 ROI 4 ROI 5
Scan Setting FBP Safire 2 FBP Safire 2 FBP Safire 2 FBP Safire 2 FBP Safire 2
Downloaded from www.ajronline.org by HUS-KUVANTAMINEN on 05/18/15 from IP address 193.166.253.30. Copyright ARRS. For personal use only; all rights reserved

Reference 5.0 (0.0) 4.1 (0.0) 5.8 (0.0) 4.9 (0.0) 5.8 (0.0) 4.9 (0.0) 4.7 (0.0) 3.9 (0.0) 4.5 (0.0) 3.7 (0.0)
Bismuth shield after scout imaging 5.0 (0.0) 4.1 (0.0) 5.4 (–7.4) 4.5 (–8.6) 6.0 (4.0) 5.1 (4.9) 4.9 (4.3) 4.0 (3.6) 5.1 (14.5) 4.3 (17.4)
Bismuth shield before scout imaging 5.4 (8.5) 4.6 (8.3) 5.5 (–5.8) 4.6 (–5.1) 5.4 (–7.4) 4.5 (–8.4) 4.9 (4.7) 4.1 (6.2) 4.7 (4.0) 3.8 (4.6)
Bismuth shield with gantry tilt 5.1 (3.2) 4.2 (2.9) 4.4 (–24.5) 3.6 (–26.0) 4.6 (–20.5) 3.8 (–22.2) 4.4 (–5.8) 3.7 (–6.2) 4.4 (–2.5) 3.6 (–2.7)
Gantry tilt 5.1 (2.4) 4.0 (–2.4) 4.5 (–22.6) 3.8 (–22.8) 4.7 (–19.8) 3.7 (–23.9) 4.8 (2.4) 3.9 (0.0) 4.7 (5.6) 3.9 (5.2)
Gantry tilt with half an angle 4.8 (–3.6) 3.9 (–6.1) 4.9 (–17.0) 4.0 (–18.7) 5.4 (–6.7) 4.5 (–9.0) 4.8 (2.2) 4.1 (4.1) 4.7 (5.4) 3.9 (5.7)
OBTCM 6.3 (27.9) 5.3 (29.7) 6.2 (5.8) 4.8 (–0.8) 6.5 (12.4) 5.2 (5.3) 5.7 (22.8) 4.7 (20.3) 5.8 (29.2) 4.7 (28.1)
OBTCM with gantry tilt 6.4 (30.1) 5.1 (24.8) 5.6 (–4.1) 4.8 (–0.8) 6.3 (7.9) 4.9 (0.0) 5.7 (22.8) 4.3 (9.7) 5.5 (22.0) 4.0 (7.1)
Note—Values in parentheses are relative change as percentage. Negative relative change value indicates noise reduction compared with reference scan setting. ROI 1 =
region of the right cerebellum, ROIs 2 and 3 = expected positions of the anterior temporal lobes, ROIs 4 and 5 = expected positions of the basal ganglia nuclei, FBP =
filtered back projection, Safire 2 = Safire level 2 iterative reconstruction (Siemens Healthcare). OBTCM = organ-based tube current modulation.

TABLE 4: Mean CT Numbers (HU) and Relative Change Compared With Reference Setting in Five ROIs in Clinically
Significant Areas
ROI 1 ROI 2 ROI 3 ROI 4 ROI 5
Scan Setting FBP Safire 2 FBP Safire 2 FBP Safire 2 FBP Safire 2 FBP Safire 2
Reference 51.0 (0.0) 50.8 (0.0) 51.9 (0.0) 51.7 (0.0) 53.2 (0.0) 53.3 (0.0) 49.4 (0.0) 49.2 (0.0) 49.7 (0.0) 49.7 (0.0)
Bismuth after scout imaging 51.6 (1.2) 51.3 (1.0) 51.8 (–0.2) 51.6 (–0.3) 53.9 (1.2) 53.8 (0.9) 49.6 (0.4) 49.5 (0.6) 50.1 (0.9) 50.2 (1.1)
Bismuth shield before scout imaging 51.7 (1.5) 51.8 (2.0) 53.8 (3.5) 53.4 (3.2) 55.4 (3.9) 54.4 (2.1) 49.3 (–0.2) 49.4 (0.4) 49.8 (0.3) 49.9 (0.6)
Bismuth shield with gantry tilt 48.3 (–5.3) 48.3 (–4.9) 52.3 (0.8) 52.3 (1.2) 53.2 (–0.2) 53.1 (–0.5) 49.4 (0.1) 49.3 (0.1) 49.7 (–0.1) 49.7 (0.1)
Gantry tilt 47.9 (–6.0) 48.0 (–5.5) 51.9 (–0.2) 52.1 (0.7) 53.0 (–0.5) 52.9 (–0.7) 49.3 (–0.2) 49.4 (0.3) 49.8 (0.2) 50.0 (0.7)
Gantry tilt with one-half angle 48.2 (–5.4) 47.9 (–5.7) 49.0 (–5.8) 49.9 (–3.4) 53.5 (0.5) 53.5 (0.4) 49.0 (–0.8) 48.8 (–0.9) 49.8 (0.2) 49.6 (–0.1)
OBTCM 50.5 (–1.0) 50.2 (–1.2) 53.4 (2.8) 53.7 (3.9) 51.5 (–3.3) 51.7 (–3.1) 48.7 (–1.4) 49.1 (–0.3) 50.2 (1.1) 50.2 (1.2)
OBTCM with gantry tilt 52.3 (2.7) 50.4 (–0.9) 51.4 (–1.1) 50.5 (–2.3) 55.2 (3.7) 55.9 (4.8) 48.2 (–2.5) 48.8 (–1.0) 49.9 (0.3) 49.0 (–1.3)
Note—Values in parentheses are relative change as percentage. Negative relative change value indicates noise reduction compared with reference scan setting. ROI 1 =
region of the right cerebellum, ROIs 2 and 3 = expected positions of the anterior temporal lobes, ROIs 4 and 5 = expected positions of the basal ganglia nuclei, FBP =
filtered back projection, Safire 2 = Safire level 2 iterative reconstruction (Siemens Healthcare). OBTCM = organ-based tube current modulation.

of the ATOM phantom, the most efficient (36%) was achieved with the RANDO decreased in the anterior tissues owing to re-
way to reduce eye lens dose appeared to be phantom with the combination of OBTCM duced radiation output in the anterior angles
gantry tilt according to our clinical practice and gantry tilt. The other methods produced (covering approximately 120°).
(baseline set from skull base to radix nasi), dose reductions of 31% (bismuth shield By tilting the gantry, we were able to re-
with or without bismuth shields, with an together with gantry tilt), 24% (OBTCM duce scan length by a few slices. Thus the
approximately 75% decrease in the absorbed alone), 18% (gantry tilt), 12% (bismuth shield dose-length product and therefore the effec-
dose compared with the reference setting. In already set on eyes in scout imaging), and tive dose calculated with International Com-
addition, using the combination of OBTCM 10% (bismuth shield alone). Using gantry tilt mission on Radiological Protection publi-
and gantry tilt produced a dose savings of at an angle one half of that used in clinical cation 103 conversion factors was reduced
70%, whereas OBTCM alone was associated practice produced only approximately 7% approximately 6% compared with the refer-
with a dose reduction of 32%. With the dose savings for the lens compared with the ence scan with no gantry tilt.
6-mm lead equivalent bismuth shield, it was results with the reference scan setting.
possible to reduce lens dose as much as 25%, When OBTCM was used, the dose distri- Image Analysis Results
whereas gantry tilted at an angle one half of bution differed substantially from the dis- Mean image noise and CT numbers are
that used in clinical practice produced 20% tributions observed with the other settings: shown in Tables 3 and 4. Image noise varied
dose savings for the lens compared with the The absorbed dose increased in the pos- in the range of 4.8–6.4 HU in ROI 1, 4.4–6.5
reference setting. In the case of the RANDO terior parts of the phantoms (up to 17% in HU in ROIs 2 and 3, and 4.4–5.7 HU in ROIs
phantom, the dose reduction was less and the case of the ATOM phantom and 27% in 4 and 5 when iterative reconstruction was not
occurred in different order than in the the case of the RANDO phantom) owing to used. When OBTCM was used with or with-
ATOM phantom owing to different phantom boosting of radiation output in the posteri- out gantry tilt, the image noise increased as
geometry. The highest lens dose reduction or angles (covering approximately 240°) and much as 30% in the bottom and posterior part

120 AJR:204, January 2015


Lens Dose in Head CT
Downloaded from www.ajronline.org by HUS-KUVANTAMINEN on 05/18/15 from IP address 193.166.253.30. Copyright ARRS. For personal use only; all rights reserved

A B C
Fig. 3—CT images show selected ROIs for measurement of image noise and contrast.
A, ROI 1 is in region of right cerebellum.
B, ROIs 2 and 3 are in expected positions of anterior temporal lobes.
C, ROIs 4 and 5 are in expected positions of basal ganglia nuclei.

of the brain (ROI 1) of the ATOM phantom. the eye lenses since results of several stud- ing up to 75%. In the case of the ­R ANDO
Similarly, OBTCM increased image noise ap- ies have indicated that the lens is more sen- phantom, however, the lens dose was re-
proximately 12% in the anterior part of the sitive to ionizing radiation than previously duced only up to 18%.
brain (ROIs 2 and 3) and 29% in the central believed. The aim of our study was to ex- We studied the effect of tilting the gantry
part of the brain (ROIs 4 and 5). The use of plore the lens dose reduction capability of at an angle one half of that used in clinical
gantry tilt decreased image noise approxi- different optimization methods in head CT practice (6.5° and 7°) for lens doses, because
mately a quarter in the anterior part of the of the brain by use of two tissue-equiva- tilting angles also vary remarkably in routine
brain compared with the reference scan set- lent anthropomorphic phantoms, MOSFET work owing to patient-specific physiologic
ting. In other ROIs, the change was more mod- dosimeters, and quantitative image quali- limitations [20]. Although the lens dose re-
erate. The use of a bismuth shield increased ty analysis. In all, we used eight different duction was considerably smaller (20% for
image noise approximately 17% in the central scanning settings. the ATOM phantom, 7% for the RANDO
part of the brain when the shield was set on In this study, we followed the gantry tilt phantom) than in the case of the original
the eyes after scout imaging. Image noise in- practice at our hospital, aligning the scan gantry tilt, using even a small tilting angle
creased less when the bismuth shield was set, plane along the skull base to the radix nasi. appears useful—because the scanning range
incorrectly, before scout imaging, because the The European guidelines call for a 10–12° may be shortened—with a corresponding de-
tube current modulation compensated the in- gantry tilt above the orbitomeatal line to re- crease in total radiation dose.
creased x-ray attenuation of the scan object by duce exposure of the eye lenses [26]. Be- Bismuth shields are used in CT exami-
increasing the tube current–time product. The cause of different geometric profiles of the nations to protect radiosensitive superficial
use of iterative reconstruction (Safire level 2) phantoms, tilting in the case of the ATOM tissues and organs such as the lens, thy-
decreased image noise approximately 20% in phantom was mainly comparable to the su- roid, and mammary tissue from primary
each ROI and for all scan settings. praorbital baseline, whereas in the case of radiation within the scanning range. In our
Image contrast was maintained for each the RANDO phantom, the gantry tilt fol- study, the reduction of lens dose due to bis-
scan setting when Safire level 2 was used for lowed closely the orbitomeatal baseline, al- muth shielding was approximately 10% in
image reconstruction. The mean CT num- though the tilting angles (13° and 11.5°) were the case of the RANDO phantom and ap-
ber, and thus the image contrast, in the brain about the same. Thus the lens dosimeter proximately 25% in the case of the ATOM
tissue of the ATOM phantom was approxi- points were left outside of the primary radi- phantom, which is moderate compared with
mately 50 HU (minimum and maximum, 48 ation beam in the ATOM phantom, where- earlier published results [7, 9, 10, 16, 27].
and 56 HU). The greatest variation (5–6%) as in the case of the RANDO phantom, the OBTCM can also be used for reduction
in image contrast was observed in the bot- lenses were partly inside the primary beam. of dose to superficial radiosensitive tissues.
tom posterior part of the brain when gantry According to results of previous studies, the In this study, we found that the lens doses
tilting was used. use of supraorbital gantry tilt reduces expo- were reduced as much as one third in the
sure of the eye lenses by 78–88%, depending case of the ATOM phantom and one fourth
Discussion on the CT scanner and technique used [10, for the RANDO phantom when we used the
CT scanner manufacturers and users 12, 13, 15]. Our results for the ATOM phan- OBTCM method. The results were similar
have made an effort to minimize the dose to tom were similar, the lens dose reduction be- to those in other studies of OBTCM tech-

AJR:204, January 2015 121


Nikupaavo et al.

niques, which have indicated dose reductions scout imaging in the case of automatic ex- both without (28%, 23%, and 29%) and with
of 26–59% to the lenses [6–8]. posure control, the patient dose will be in- combined gantry tilt (30%, 23%, and 22%).
We also studied whether combining bis- creased. These disadvantages have been ob- However, the absolute change in measured
muth shields or OBTCM with simultaneous served in the case of bismuth shielding of the noise level compared with the reference level
Downloaded from www.ajronline.org by HUS-KUVANTAMINEN on 05/18/15 from IP address 193.166.253.30. Copyright ARRS. For personal use only; all rights reserved

gantry tilt brings any benefit. When the lens- breast [32] and thyroid [33]. Authors of sev- was less than 1.5 HU in all of our study con-
es were located completely outside the pri- eral studies [34–37] have recommended the ditions. All the different imaging methods
mary beam owing to tilting, neither bismuth use of tube current modulation with lowered studied had only little effect on the measured
shielding nor OBTCM resulted in extra dose tube current–time values instead of bismuth contrast values (mean attenuation in HU
reduction. In fact, adding OBTCM result- breast shields. This method will not cause measured in the ROIs). The mean attenua-
ed in a slight increase in the lens dose, most image quality impairments due to streak ar- tion changed the most in the posterior fossa
probably due to higher scatter intensity origi- tifacts while resulting in a similar dose re- when gantry tilt was used. The decrease was
nating from the elevated primary beam dose duction. Nevertheless, in the use of bismuth 5–6% from the reference image, correspond-
distribution on the occipital side of the skull. shields on the eyes, artifacts seem to fall out- ing to an absolute difference of approximate-
On the other hand, with more gradual tilting, side of the diagnostic area of interest when- ly 3 HU. However, even such small changes
we observed that both the bismuth shield and ever visualization of the orbits and sinuses is in attenuation levels can be crucial in CT of
OBTCM resulted in clear dose reduction to not important [16, 38]. In earlier studies, use the brain, in which, for example, the contrast
the lenses. The primary beam dose contribu- of OBTCM techniques did not appear to af- between white and gray matter is approxi-
tion and close range scatter would otherwise fect image noise or CT numbers, or their in- mately only 4–5 HU.
become more prominent for the lenses. Wang fluence was scant [6–8]. Our study had limitations. First, we used
et al. [7] studied the simultaneous use of bis- In this study, we assessed image quality only one CT scanner from a single vendor. Be-
muth shielding and OBTCM and reported by using five ROIs located in the posterior, cause focus-detector and focus-isocenter dis-
considerable dose reduction (almost 50%) anterior, and central parts of the brain. The tances, tube current modulation techniques,
for the eye lenses. However, this combination RANDO phantom was not suitable for image and bowtie filters vary between scanners, the
should be used carefully because OBTCM quality evaluation because the air between effects of different dose reduction methods on
boosts radiation output in the posterior part its slices caused strong artifacts. Gantry tilt radiation dose distribution and image noise
of the brain, and the bismuth shielding scat- and partial gantry tilt reduced the noise level may vary between scanners. In addition, the
ters this back to the patient [28]. in the anterior parts of the brain (the former OBTCM technique is not available with most
Gantry tilting is not possible with all by almost one fourth), whereas in the case of of the scanners used today. Second, we used
available CT scanner models. In such cas- the posterior and central ROIs, the changes in only two anthropomorphic phantoms for dose
es, the patient’s head can be set on a head noise level were minor. The use of iterative re- measurements and only a single phantom for
support with the chin turned to the chest to construction (Safire level 2) resulted system- image quality assessment; the geometry and
mimic gantry tilt. In practice, however, tilt- atically in an approximately 20% decrease in size of real patients vary substantially. The
ing the patient’s head is not always conceiv- noise level. However, in this study we did not potential effect of this variation is seen in our
able because of anatomic properties and the consider other image quality aspects related dosimetry results with different dose savings
physiologic condition of the patient. In these to the use of iterative reconstruction. between the two phantoms.
cases, the use of bismuth shields or OBTCM According to our results, using bismuth
together with ordinary tube current modula- shielding of the eyes appears advantageous Conclusion
tion can be considered the primary means of with only little effect on image quality in the According to our study results, the most ef-
reducing the dose to the eye lenses. case of brain scans. No artifacts were ob- ficient way to reduce radiation exposure to the
The benefit of different dose reduction served, and the differences in noise level in lens appears to be gantry tilt. Compared with
methods must be evaluated against the re- the case of the reference scan and the scan the reference scan, using gantry tilt resulted in
quirements for clinically adequate image with a bismuth shield were measured to be a substantial decrease in the dose to the lens
quality. Gantry tilting has been reported to minor in all brain regions. Placing the bis- while preserving image contrast and reducing
cause beam-hardening and partial volume muth shield before scout imaging decreased image noise, especially in the anterior part of
artifacts in the base of the skull and to cause the exposure in all measuring points except the brain. If gantry tilting is not possible, or if
problems in visualization of the temporal for the mandible, where the increase in dose the eye lenses cannot be fully excluded from
lobes [15, 29]. However, in other reports [20, was 14% (resulting in a total increase in dose the exposed scan range, OBTCM or bismuth
30] it has been claimed that no substantial of 2.5% for the whole scan), whereas the shields can also be used, with some caution, to
differences in clinical image quality of head change in the image quality was small com- decrease the lens exposure.
CT can be observed with different gantry tilt pared with the reference scan. However, this
angles. The use of bismuth shielding has also feature is due to the real-time angular tube References
been questioned, because it can degrade im- current modulation technique and cannot be 1. Kleinman NJ. Radiation cataract. Ann ICRP
age quality and accuracy by causing streak generalized to automatic exposure control 2012; 41:80–97
and beam-hardening artifacts and waste of techniques in which the angular modulation 2. Shore RE, Neriishi K, Nakashima E. Epidemio-
image data by absorbing photons transmit- profile is based on the scout image. logical studies of cataract risk at low to moderate
ted through the patient before they would In this study, we found that OBTCM in- radiation doses: (not) seeing is believing. Radiat
appear on the detector [28, 31]. In addition, creased the noise level in the posterior and Res 2010; 174:889–894
if the shields are placed erroneously before central (right and left) regions of the brain 3. Ainsbury EA, Bouffler SD, Dörr W, et al. Radia-

122 AJR:204, January 2015


Lens Dose in Head CT

tion cataractogenesis: a review of recent studies. 15. Heaney DE, Norvill AJ. A comparison of reduc- 27. Catuzzo P, Aimonetto S, Fanelli G, et al. Dose re-
Radiat Res 2009; 172:1–9 tion in CT dose through the use of gantry angula- duction in multislice CT by means of bismuth
4. [No authors listed]. The 2007 recommendations tions or bismuth shields. Australas Phys Eng Sci shields: results of in vivo measurements and com-
of the International Commission on Radiological Med 2006; 29:172–178 puted evaluation. Radiol Med (Torino) 2010;
Downloaded from www.ajronline.org by HUS-KUVANTAMINEN on 05/18/15 from IP address 193.166.253.30. Copyright ARRS. For personal use only; all rights reserved

Protection: ICRP publication 103. Ann ICRP 16. Huggett J, Mukonoweshuro W, Loader R. A phan- 115:152–169
2007; 37:1–322 tom-based evaluation of three commercially 28. American Association of Physicists in Medicine
5. Authors on behalf of ICRP; Stewart FA, Akleyev available patient organ shields for computed to- website. AAPM position statement on the use of
AV, Hauer-Jensen M, et al. ICRP publication 118: mography x-ray examinations in diagnostic radi- bismuth shielding for the purpose of dose reduc-
ICRP statement on tissue reactions and early and late ology. Radiat Prot Dosimetry 2013; 155:161–168 tion in CT scanning. www.aapm.org/publicgeneral/
effects of radiation in normal tissues and organs— 17. Klein BE, Klein R, Linton KL, Franke T. Diag- BismuthShielding.pdf. February 7, 2012. Ac-
threshold doses for tissue reactions in a radiation nostic x-ray exposure and lens opacities: the Bea- cessed September 9, 2014
protection context. Ann ICRP 2012;41:1–322 ver Dam Eye Study. Am J Public Health 1993; 29. Hsieh J. Tomographic reconstruction for tilted he-
6. Reimann AJ, Davison C, Bjarnason T, et al. Or- 83:588–590 lical multislice CT. IEEE Trans Med Imaging
gan-based computed tomographic (CT) radiation 18. Hourihan F, Mitchell P, Cumming R. Possible as- 2000; 19:864–872
dose reduction to the lenses: impact on image sociation between computed tomography scan 30. van Straten M, Venema HW, Majoie CB, Freling
quality for CT of the head. J Comput Assist To- and cataract: the Blue Mountains Eye Study. Am J NJ, Grimbergen CA, den Heeten GJ. Image qual-
mogr 2012; 36:334–338 Public Health 1999; 89:1864–1866 ity of multisection CT of the brain: thickly colli-
7. Wang J, Duan X, Christner J, Leng S, Grant KL, 19. Yuan MK, Tsai DC, Chang SC, et al. The risk of mated sequential scanning versus thinly collimat-
McCollough GH. Bismuth shielding, organ-based cataract associated with repeated head and neck ed spiral scanning with image combining. AJNR
tube current modulation, and global reduction of CT studies: a nationwide population-based study. 2007; 28:421–427
tube current for dose reduction to the eye at head AJR 2013; 201:626–630 31. McCollough CH, Wang J, Gould RG, Orton CG.
CT. Radiology 2012; 262:191–198 20. Yeoman LJ, Howarth L, Britten A, et al. Gantry Point/counterpoint: the use of bismuth breast
8. Duan X, Wang J, Christner J, Leng S, Grant KL, angulation in brain CT: dosage implications, ef- shields for CT should be discouraged. Med Phys
McCollough GH. Dose reduction to anterior sur- fect on posterior fossa artifacts, and current inter- 2012; 39:2321–2324
faces with organ-based tube current modulation: national practice. Radiology 1992; 184:113–116 32. Coursey C, Frush DP, Yoshizumi T, Toncheva G,
evaluation of performance in a phantom study. 21. Kaasalainen T, Palmu K, Lampinen A, Kortesnie- Nguyen G, Greenberg SB. Pediatric chest MDCT
AJR 2011; 197:689–695 mi M. Effect of vertical positioning on organ dose, using tube current modulation: effect on radia-
9. Hopper KD, Neuman JD, King SH, Kunselman image noise and contrast in pediatric chest CT- tion dose with breast shielding. AJR 2008;
AR. Radioprotection to the eye during CT scan- phantom study. Pediatr Radiol 2013; 43:673–684 190:[web]W54–W61
ning. AJNR 2001; 22:1194–1198 22. Perisinakis K, Seimenis I, Tzedakis A, Papadakis 33. Leswick DA, Hunt MM, Webster ST, Fladeland
10. McLaughlin DJ, Mooney RB. Dose reduction to AE, Damilakis J. The effect of head size/shape, DA. Thyroid shields versus z-axis automatic tube
radiosensitive tissues in CT: do commercially miscentering, and bowtie filter on peak, patient tis- current modulation for dose reduction at neck CT.
available shields meet the users’ needs? Clin Ra- sue doses from modern brain perfusion 256-slice Radiology 2008; 249:572–580
diol 2004; 59:446–450 CT: how can we minimize the risk for determinis- 34. Colletti PM, Micheli OA, Lee KH. To shield or
11. Jaffe TA, Hoang JK, Yoshizumi TT, Toncheva G, tic effects? Med Phys 2013; 40:011911–011918 not to shield: application of bismuth breast shields.
Lowry C, Ravin C. Radiation dose for routine clini- 23. Korn A, Bender B, Fenchel M, et al. Sinogram af- AJR 2013; 200:503–507
cal audit brain CT: variability on different scanners firmed iterative reconstruction in head CT: im- 35. Foley SJ, McEntee MF, Rainford LA. An evalua-
at one institution. AJR 2010; 195:433–438 provement of objective and subjective image qual- tion of in-plane shields during thoracic CT. Radiat
12. Tan JS, Tan KL, Lee JC, Wan CM, Leong JL, ity with concomitant radiation dose reduction. Prot Dosimetry 2013; 155:439–450
Chan LL. Comparison of eye lens dose on neuro- Eur J Radiol 2013; 82:1431–1435 36. Servaes S, Zhu X. The effects of bismuth breast
imaging protocols between 16- and 64-section 24. Korn A, Fenchel M, Bender B, et al. Iterative recon- shields in conjunction with automatic tube current
multidetector CT: achieving the lowest possible struction in head CT: image quality of routine and modulation in CT imaging. Pediatr Radiol 2013;
dose. AJNR 2009; 30:373–377 low-dose protocols in comparison with standard 43:1287–1294
13. Suzuki S, Furui S, Ishitake T, et al. Lens exposure filtered back-projection. AJNR 2012; 33:218–224 37. Vollmar SV, Kalender WA. Reduction of dose to
during brain scans using multidetector row CT 25. Kilic K, Erbas G, Guryildirim M, Arac M, Ilqit E, the female breast in thoracic CT: a comparison of
scanners: methods for estimation lens dose. AJNR Coskun B. Lowering the dose in head CT using standard-protocol, bismuth-shielded, partial and
2010; 31:822–826 adaptive statistical iterative reconstruction. AJNR tube-current-modulated CT examinations. Eur
14. Diekmann S, Siebert E, Juran R, et al. Dose expo- 2011; 32:1578–1582 Radiol 2008; 18:1674–1682
sure of patients undergoing comprehensive stroke 26. Danish Society of Radiology website. European 38. Mukundan S, Wang P, Frush D, et al. MOSFET
imaging by multidetector-row CT: comparison of guidelines on quality criteria for computed tomogra- dosimetry for radiation dose assessment of bis-
320-detector row and 64-detector row CT scan- phy. EUR 16262 EN. www.drs.dk/­ guidelines/ct/ muth shielding of the eye in children. AJR 2007;
ners. AJNR 2010; 31:1003–1009 quality/mainindex.htm. Accessed September 9, 2014 188:1648–1650

AJR:204, January 2015 123

View publication stats

You might also like