AU680 Part G ANL Installation

G12 SMP Connection Kit Setup G12-1


Ver.7

SMP Connection Kit Setup


Outline
The setup procedure to connect the SMP connection kit to the DTS connection instrument is
indicated below. After this procedure follow the procedure of G5, and connect to the sampler.

SMP Connection kit Composition List


Part No. Name MW0330 Remarks
M541-OPK01
MF5328 R cover 2 1
MF5329 R cover 3 1
MF4983 Connection board 1
MW0376 M541-OPG01-DPC01 1 Packing material for
DTS connection kit

Setup Procedure
1) Turn OFF the power of the instrument.
2) Open the Top cover of the instrument.
3) Jack down the ANL.
4) Remove the connector which is connected to the DTS.
5) Separate the ANL from the DTS.
6) Take the cable connected to the DTS through the whole “A” and put it inside the ANL, and the
fix it with the SKB band
7) Remove screws fixing the right side cover, and remove the right side cover.

Screws

G12 SMP Connection Kit Setup e541g12.doc G12-1


AU680 Part G ANL Installation
G12 SMP Connection Kit Setup G12-2
Ver.7

Tip
• If you can access the cable connector and the Teflon tube of the FA unit from the right
side of the ANL when it is in the DTS connection, you do not perform the following steps
until the step 25).

8) Remove the Teflon tube from the syringe head for sample.
9) Remove the screw “A” fixing the M220-SA unit.
10) Pull the M220-SA unit forward and remove the connector [877] for the motor and connector
[878] for the sensor which are connected to the syringe and remove it.

Screws A

11) Remove the connector [502] and the Teflon tube which are from bottom of the FA unit and
connected to the MF6925 connection board.
12) Unscrew the 2 screws fixing the flame on the MF6925 connection board and remove the
connection board.

MF6925 Connection board

Screw
Screw

G12 SMP Connection Kit Setup e541g12.doc G12-2


AU680 Part G ANL Installation
G12 SMP Connection Kit Setup G12-3
Ver.7

13) Cut the tie wrap fixing the Fair-Rite core, and remove it from the connection board.
14) Remove the syringe head and the connected Teflon tube MF5159.

Tie wrap

15) Remove the faston terminal (Cable).


16) Mark direction on the connector which is connected the Fair-Rite cable.
17) Use the longnose pliers and pinch the nail of the MD7796 clamp and the marked connector
which is connected cable of the Fair-Rite core, and then remove them.
18) Unscrew the screws fixing the fitting and the terminal block of the Faston terminal, and remove
it from the connection board.

Mark the connector

Faston terminal

502

MD9212 Pipe joint

MD7796 Clamp

G12 SMP Connection Kit Setup e541g12.doc G12-3


AU680 Part G ANL Installation
G12 SMP Connection Kit Setup G12-4
Ver.7

19) Fix the removed the fitting and the faston terminal block with screws to the MF4983 connection
board which is enclosed the SMP connection kit.

MF4983 Connection board

Faston terminal

MD9212 Pipe joint

20) Fix the removed clamp and connector on the MF6925 connection board, at this time check the
right direction to connect.
21) Fix the faston terminal (cable) to the faston terminal block.

MD9212 Pipe joint

Faston terminal

Connector

MF4983 Connection board

MD7796 Clamp

Note
• When you fix the connector, the marking which is marked at the step 16) should be seen.
• Fix the clamp that the pipe joint side of the MD9212 can open.

G12 SMP Connection Kit Setup e541g12.doc G12-4


AU680 Part G ANL Installation
G12 SMP Connection Kit Setup G12-5
Ver.7

22) Remove the name plate of the connector No. [502] from the MF6925, and put it on the MF4983.
23) Fix the Fair-Rite core to the connection board with a tie wrap.

502

Tie wrap

24) Fix the MF4983 connection board to the ANL flame with 2 screws.

Screw
Screw

25) Fix the MF5159 Teflon tube which is connected with the syringe head to the top of the fitting.
26) Connect the Teflon tube of the FA unit and the cable connector [502] to the connection board.

G12 SMP Connection Kit Setup e541g12.doc G12-5


AU680 Part G ANL Installation
G12 SMP Connection Kit Setup G12-6
Ver.7

27) Fix the MF5328 R cover 2 and the MF5329 R cover 3 which is enclosed the SMP connection
kite to the ANL, and fix it with screws.

Screws

28) Take the grounding cable and the 24 V connector [300], which are fixed inside the ANL, out
from “A” to the outside the ANL.

29) Remove the terminator connected to the cuvette wash control PCB (MV3981) [COM2].
30) Follow the procedure of the Part G5 after this to connect to the sampler.

G12 SMP Connection Kit Setup e541g12.doc G12-6