Professional Documents
Culture Documents
ARIES - CH 06 - Settings&Adjustments - Rev.07 (1 of 2)
ARIES - CH 06 - Settings&Adjustments - Rev.07 (1 of 2)
ARIES - CH 06 - Settings&Adjustments - Rev.07 (1 of 2)
6.4 PHOTOMETER
6.4.1 Electronic adjustments of the Photometer
Make sure that the Photometer Halogen Lamp and the Reactions Plate cuvettes
are brand-new before applying the following electronic adjustments procedure.
1. Turn on the Ilab Aries system (instrument and computer) and launch the Diagnostic program.
2. Select "Miscellaneous” folder and request an Instrument General reset. Wait until the reset
instrument procedure has been completed.
3. Make sure that all the Home Sensors of the Sampling Arm, Reactions Plate, Reagents Plate,
Samples Plate and the Filters Wheel light up in green.
4. Remove the Reactions Plate cover (Fig. 7) and place 500 µL of distilled water in the
cuvette # 1 of the Reactions Plate.
5. Replace the Reactions Plate cover (Fig. 7) on its plate (in order to avoid external interferences
of light in the reading).
6. Select “Optic” folder and verify that the Photometer Halogen Lamp is in “ON” state.
7. Make sure that the voltage on the J2 connector of the “Photometer Lamp Board” (Fig. 52) is
equal to 5.8 Volts ± 0.1 . If it not be so, turn the R6 trimmer (Fig. 52) for adjusting the voltage.
Keep plate
Halogen lamp
J2 connector of the
Photometer Lamp
Board
R6 trimmer of the
Photometer Lamp Board
Make sure that the Reactions Plate cuvettes are perfectly clean. If it not be so,
perform the “Washing cuvettes” operation (form “Start Work”).
The electric light and the daylight can influence the readings of the Photometer.
Therefore, bring down the Raising Top and close the Reactions Plate with its
cover (Fig. 7) before to perform the reading operations.
8. Click the “Dark” button, click the "START Sample Channel Conversion" button and make
sure that the adjacent text box shows a counts value equal to about 100 ± 50. If it not be so,
turn the OFFSET trimmer of the “Pre-Ampl./ADC Board (Sample Channel)” in order to set
the correct counts value (Fig. 53) [important note: the setted value must not be 0].
Reactions Plate
Reaction
cuvette
Washing
Station
Fig. 53 – GAIN and OFFSET trimmers of the “Pre-Ampl./ADC Board” (Sample Channel)
9. After the Sample Channel OFFSET adjusting, click the "STOP Sample Channel Conversion"
button.
10. Click the "START Reference Channel Conversion" button (in the “Dark” position) and make
sure that the adjacent text box shows a counts value equal to about 100 ± 50. If it not be so,
turn the OFFSET trimmer of the “Pre-Ampl./ADC Board (Reference Channel)” in order to set
the correct counts value (Fig. 61) [important note: the setted value must not be 0].