Professional Documents
Culture Documents
MD Totco Series 2000 Instruments Installation, Calibration, & Operation
MD Totco Series 2000 Instruments Installation, Calibration, & Operation
MD Totco Series 2000 Instruments Installation, Calibration, & Operation
This document contains proprietary and confidential information which NOV M/D Totco
belongs to National-Oilwell Varco, L.P., its affiliates or subsidiaries (all
collectively referred to hereinafter as "NOV"). It is loaned for limited 1200 Cypress Creek Road
purposes only and remains the property of NOV. Reproduction, in whole Cedar Park, Texas 78613
or in part, or use of this design or distribution of this information to others
is not permitted without the express written consent of NOV. This
USA
document is to be returned to NOV upon request and in any event upon Phone 512-340-5000
completion of the use for which it was loaned. This document and the Fax 512-340-5219
information contained and represented herein is the copyrighted property
of NOV.
© National Oilwell Varco
DOCUMENT NUMBER REV
90-01 E
www.nov.com/mdtotco
Document number 90-01
Revision E
Page 1 of 158
Revision History
Change Description
www.nov.com/mdtotco
Document number 90-01
Revision E
Page 3 of 158
Table of Contents
Chapter 1: General Information
Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Notes, Cautions, and Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Illustrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Safety Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Personnel Training . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Recommended Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
General System Safety Practices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Replacing Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Routine Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Proper Use of Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Chapter 2: Overview
System Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
LM2000 Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Analog Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Digital Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Sensor Excitation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Analog Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Aux Comm (Half-duplex) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Industry standard plug-in relays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Environmental . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Input Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Back Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Chapter 3: Installation
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Mounting Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Standard Panel Mount . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Dashmount Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Heavy Duty Stainless Steel Bezel Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Remote Data Acquisition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Wire Connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
TB1-12/24VDC (Power Input) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
TB2-PWR OUT (Power Output) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
TB3-RS485 (Network Communication) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
TB4-ANLG1 (Analog Input #1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
TB5-ANLG2 (Not Used) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
TB6-ANLG3 (Not Used) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
www.nov.com/mdtotco
Document number 90-01
Revision E
Page 4 of 158
Table of Contents
TB7-ANLG4 (Not Used) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
TB8-POS PWR (Payout Sensor Power) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
+5 VDC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
+24 VDC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
+12 VDC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
TB9-POS SIG (Payout Sensor Signal) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
TTL (0 to +5 VDC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Open Collector (NPN) Sinking (Switch to Ground) . . . . . . . . . . . . . . . . . . . . . . . . . .26
Open Collector (PNP) Sourcing (Switch to +24 VDC) . . . . . . . . . . . . . . . . . . . . . . .27
NAMUR (Proximity) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
TB10 Through TB13 (Relay Option Terminal Blocks) . . . . . . . . . . . . . . . . . . . . . . . . . . .28
Terminal Blocks for Analog Output/Aux Comm Option Modules 1 to 4 . . . . . . . . . . . . .29
Analog Output Option Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
Aux Comm Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
Even Input Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
www.nov.com/mdtotco
Document number 90-01
Revision E
Page 5 of 158
Table of Contents
Big Bar Graph, Payout Operator’s Screen Layout . . . . . . . . . . . . . . . . . . . . . . . . . 45
Menu Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Chapter 6: Troubleshooting
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Navigating the Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Diagnostic Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Raw Counts (troubleshooting diagnostic) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
www.nov.com/mdtotco
Document number 90-01
Revision E
Page 6 of 158
Table of Contents
Power Supplies (run time background) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .79
Error Logs (troubleshooting diagnostic) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .80
Local (or Sensor) Network Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .81
Local (or Sensor) Hardware Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .82
Self Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .83
Self Test Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .84
Perform Self Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85
Load Factory Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85
Relays-Alarms Disabled . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .86
Options (troubleshooting diagnostic) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .86
SPI Diagnostics (run-time background) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .87
Software Update Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .87
Error Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .89
Appendix A: Glossary
www.nov.com/mdtotco
Document number 90-01
Revision E
Page 7 of 158
Table of Contents
Product Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
www.nov.com/mdtotco
Document number 90-01
Revision E
Page 8 of 158
Table of Contents
Aux Comm Response Times . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .128
SS (single scan) Command . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .128
C1 (Cal1) Command . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .129
CD (Configuration Dump) Command . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .129
Conclusions on Response Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .129
Appendix G: Networking
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .131
Sensor Input Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .131
Sensor Input Unit With Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .131
Sensor Input Unit Without Display (DAQ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .131
Remote Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .131
Sensor Input Unit Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .132
Remote Display Unit Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .132
Network Examples. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .133
Network #1—Remote DAQ to Remote Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .133
Network #2—Sensor/Display to Remote Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . .133
Network #3—DAQ to Multiple Remote Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .134
Network #4—Sensor/Display to Multiple Remote Displays . . . . . . . . . . . . . . . . . . . . .135
Network #5—Remote Display to Multiple DAQs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .136
Network #6—Remote Display to Multiple Sensor/Displays . . . . . . . . . . . . . . . . . . . . .137
Network #7—Multiple Remote DAQs to Multiple Remote Displays . . . . . . . . . . . . . . .138
Network #8—Multiple Sensor/Display Units to Multiple Remote Displays . . . . . . . . . .139
Networking and Aux Comm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .140
Network/Aux Comm Examples. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .140
Network #9—Multi. Sensor/Display to Multi. Remote Displays w/ Aux Comm . . . . . . .140
Network #10—Multi. Sensor/Display w/ Aux Comm to Multi. Remote Displays . . . . . .142
www.nov.com/mdtotco
Document number 90-01
Revision E
Page 9 of 158
1: General Information
This manual contains installation, operation, maintenance and parts information. Information in this
manual should enable qualified personnel to install, operate and troubleshoot this system. Every
effort has been made to ensure the accuracy of the information contained herein. National Oilwell
Varco’s M/D Totco, will not be held liable for errors in this material, or for consequences arising
from misuse of this material.
Conventions
Notes, Cautions, and War nings
Notes, cautions, and warnings provide readers with additional information and advise the reader to
take specific action to protect personnel from potential injury or lethal conditions. They may also
inform the reader of actions necessary to prevent equipment damage. Please pay close attention
to these advisories
Note: The note symbol indicates that additional information is provided about the
current topics.
Caution: The caution symbol indicates that potential damage to equipment or injury to
personnel exists. Follow instructions explicitly. Extreme care should be taken
when performing operations or procedures preceded by this caution symbol.
Warning: The warning symbol indicates a definite risk of equipment damage or
danger to personnel. Failure to observe and follow proper procedures
could result in serious or fatal injury to personnel, significant property
loss, or significant equipment damage.
ESD The ESD (Electrostatic Discharge) warning symbol indicates that static
Warning: control precautions are needed.
Illustrations
Illustrations (figures) provide a graphical representation of equipment components or screen
snapshots for use in identifying parts or establishing nomenclature, and may or may not be drawn
to scale.
For component information specific to your application, see the technical drawings included with
your M/D Totco documentation.
www.nov.com/mdtotco
Document number 90-01
Revision E 1: General Information
Page 10 of 158
Safety Requirements
M/D Totco equipment is installed and operated in a controlled drilling rig environment involving
hazardous situations. Proper maintenance is important for safe and reliable operation. Procedures
outlined in M/D Totco manuals are the recommended methods of performing operations and
maintenance.
Personnel Training
All personnel performing installation, operations, repair, or maintenance procedures on the
equipment, or those in the vicinity of the equipment, should be trained on rig safety, tool operation,
and maintenance to ensure their safety.
Personnel should wear protective gear during installation, maintenance, and certain
operations.
Recommended Tools
Service operations may require the use of tools designed specifically for the purpose described.
M/D Totco recommends that only those tools specified be used when stated. Ensure that
personnel and equipment safety are not jeopardized when following service procedures or using
tools not specifically recommended by M/D Totco.
Read and follow the guidelines below before installing equipment or performing
maintenance to avoid endangering exposed persons or damaging equipment.
12
www.nov.com/mdtotco
Document number 90-01
1: General Information
Revision E
Page 11 of 158
Routine Maintenance
Equipment must be maintained on a routine basis. See the service manual for maintenance
recommendations.
www.nov.com/mdtotco
Document number 90-01
Revision E 1: General Information
Page 12 of 158
12
www.nov.com/mdtotco
Document number 90-01
Revision E
Page 13 of 158
2: Overview
System Overview
The M/D TOTCO LM2000 is an instrument that displays the tension, payout, and speed of a cable
controlled by a winch. Information is displayed in digital and graphical form on a liquid-crystal
display (LCD). Information is presented on various screens which are user-selectable. Some
screens include bar graphs, while others include large digital readings. The most typical screen is
the Operator’s Screen in Figure 2.1.
High Alarm
Low-Low Alarm Low Alarm
High-High Alarm
L H
Numerical TENSION
2762LBS 0 5000 Bar Graph
L H
Value
PAYOUT
1000
FEET 0 5000
L H
SPEED
0
/MIN - 300 500
Keypad Label Bar
ACK PAYOUT CAL SCREEN MENU
Keypad
S/C + -
Figure 2.1: Typical Operator’s Screen
Values for tension, payout, and speed are shown both as numbers and horizontal bar graphs. The
bar graphs also show up to four alarm setpoints for low-low, low, high, and high-high alarms. More
information about alarms is in Chapter 4, titled "Normal Operation".
The instrument receives signals from sensors that measure tension, payout, and speed of a cable.
The tension sensor must have a 4-20mA output. Payout is measured by a phase-sensitive
(quadrature) position sensor. Speed is calculated by the instrument from the payout signal.
The sections that follow describe the instrument’s features and functions.
www.nov.com/mdtotco
Document number 90-01
Revision E 2: Overview
Page 14 of 158
LM2000 Specifications
Display
Graphic LCD (240 x 128 pixel)
Viewing area of 114mm wide x 64mm high
Adjustable backlight color (red/green) via front panel
Adjustable backlight brightness via front panel
Adjustable LCD contrast via front panel
Field scalable (full floating point) -999999 to 9999999 (max 7 characters) including “-” and “.”
Engineering units selectable from predefined list
Display resolution 1, 10, 100, 1000, 0.1, 0.01, 0.001, and 0.0001
Tri-color alarm status LED (green, amber, red)
Analog Inputs
One each 4–20 mA input; (2 for dual-axis option)
100 input impedance (0.4 2.0 V loop drop)
Common mode range: ±200 V operational; 024 V specified accuracy
Accuracy: 0.01% FS typical (1 part in 10,000) at 25° C
Temperature drift: 10ppm/C typical (50ppm/C max)
Common mode error: 0.016%/V (virtually eliminated if calibrated in place)
A/D Conversion:
Technique16 bit (02.5V = 065535 counts)
Resolution >40,000 counts for 420 mA
Input update rate16 Hz
Output update rate4 Hz
Digital Inputs
One quadrature position/speed input activated by:
-switch closure pulling to ground (< 3 V)
-external voltage applied (12–24 V)
-NAMUR type proximity sensor
Bandwidth 10 kHz
Sensor Excitation
24 V available for each analog input and each analog output (max 170 mA total for all 8 I/O
combined)
Position sensor power supply of 5–24 V at 100 mA
18
www.nov.com/mdtotco
Document number 90-01
2: Overview
Revision E
Page 15 of 158
Outputs
Analog Outputs
one for tension, one for payout, and one for speed
individually selectable module types for each parameter
three types of modules:
Non-isolated 4–20 mA sourced (1 k max load)
Non-isolated ±10 VDC (5 mA max)
Ground isolated 2-wire 420 mA transmitter (828 V), the voltage between the + and -
signal of this module must remain between 8 and 28 VDC at all times for proper
operation.
Aux Comm (Half-duplex)
RS232
RS485
Alar ms
Four alarm setpoints (2 high, 2 low) for each displayed parameter
Industry standard plug-in relays
various contact configurations and ratings available
solid state or dry contacts
latching or non-latching (front panel programmable)
fail-safe or de-energized operation
Environmental
Operating: -20 to +55° C (slow display below -10° C)
Storage: -30 to 80° C
Input Power
12–24 VDC (10 W typical, 20 W max)
Front Panel
The front panel (Figure 1-2) has five parts:
A power-on light in the upper left corner
The power-on light is green when the instrument is powered on and working normally. It
also lights up to indicate CPU self-test.
An alarm light just to the right of the power-on light
The alarm light is green when there are no alarm conditions, amber when there is at
least one low or high alarm condition, and red when there is at least one low-low or high-
high alarm condition. It also lights up to indicate CPU self-test.
www.nov.com/mdtotco
Document number 90-01
Revision E 2: Overview
Page 16 of 158
The operator can read the LCD in sunlight; it is backlit for use at night. Moreover, the
brightness and color of the backlight can be adjusted. Color ranges from green to red in
small steps (see Chapter 5, titled "Configuration and Calibration").
18
www.nov.com/mdtotco
Document number 90-01
2: Overview
Revision E
Page 17 of 158
The display operates in temperature ranges from -20° C to +70° C (-4° F to +158°
F).Contrast can be adjusted via the front panel controls. The unit must be stored within
the temperature range from
-30° C to +80° C (-22° F to +176° F).
Back Panel
TB1-12/24VDC TB2-PWR OUT TB3-RS485 TB10-RLY1 TB11-RLY2
ON/OFF CHASSIS A B A B M2 M1 G2 G1 M2 M1 G2 G1
V+ S+ S- V- V+ S+ S- V- V+ S+ S- V- V+ S+ S- V-
TB4-ANLG1 TB5-ANLG2 TB6-ANLG3 TB7-ANLG4
ADJUST
CONTRAST
RESET
ADDR + A - CHS A+ A- B+ B- M2 M1 G2 G1 M2 M1 G2 G1
TB8-POS PWR TB9-POS SIG TB13-RLY4 TB12-RLY3
Figure 2.3 shows the back panel of the unit. Along the top and bottom edges of the back panel are
rows of terminal blocks. These terminal blocks are for connecting power and sensor wires to the
instrument, and for sending signals or power out to other instruments. The black chassis housing
has white letters to identify the terminal blocks. Each terminal block has a TB-XX label, where XX
is the terminal-block number. A short, printed phrase indicates the purpose of each terminal.
For example, the chassis near terminal block number one, in the far upper left corner, has the label
TB1-12/24 VDC and ON/OFF + -. The top half of the label, 12/24 VDC, means that TB1 is for
connecting 12 or 24 VDC power. The bottom half of the label, ON/OFF + -, means that the
corresponding terminals are for connecting an external on/off switch and for connecting positive
input voltage and a common.
At the right edge of the chassis is a row of four receptacles that can accommodate option relays,
labeled RLY1 to RLY4 (Figure 1-3). Various relay contact ratings and configurations are available.
Caution: When a Series 2000 instrument is used in Class I, Division 2, Group C & D
hazardous locations, only solid-state relay options 0, 6, 7, and 8 may be used.
www.nov.com/mdtotco
Document number 90-01
Revision E 2: Overview
Page 18 of 158
The purpose of the instrument’s relay options is to initiate external warning devices such as horns,
sirens and/or lights, or commence other external relays for remote calibration and individual
customer applications. Refer to Appendix B for more information.
In the middle of the panel is a row of four receptacles used for installing option modules. Option
Modules send scaled analog signals to external instrumentation, such as a strip-chart recorder or
a data logger; these option module receptacles are also used for interfacing auxiliary
communication option modules that use RS-232 or RS-485 connections.
There are two green screws for chassis grounding: one is at the top middle edge of the back
panel, and the other is at the bottom-middle edge.
In the lower-left corner of the back panel are three holes allowing access for adjusting the screen
contrast, setting the address of the unit on a network, and reset. Once the unit is installed, the
back panel will rarely need attending.
18
www.nov.com/mdtotco
Document number 90-01
Revision E
Page 19 of 158
3: Installation
Overview
The standard instrument is designed to be mounted in a customer supplied control panel. Connect
the wires for power, sensors, and options to the terminal blocks on the back panel of the unit.
This manual does not address sensor mounting. Install all sensors according to the manufacturer’s
recommendations.
Mounting Information
Standard Panel Mount
The standard instrument is sealed only to the front panel. This configuration is used only when the
external enclosure provides adequate sealing to the rear of the S-2000 display assembly.
Locate the desired area in the operator’s console instrument panel and cut a hole to the
dimensions shown in Figure 3.1
.075” 19mm
7.01”
178mm
130mm
5.12”
0.75”
19mm
www.nov.com/mdtotco
Document number 90-01
Revision E 3: Installation
Page 20 of 158
Note: Do not over tighten the screws; the brackets, chassis, and bezel might be damaged.
Dashmount Option
An optional dashmount enclosure is available. This option provides NEMA-4 style sealing to the
rear of the S-2000 display assembly. Refer to drawing 220898 at the end of Appendix I for more
detail.
32
www.nov.com/mdtotco
Document number 90-01
3: Installation
Revision E
Page 21 of 158
Wire Connection
All wire connections to the instrument are made using the 13 or more terminal blocks on the back
panel (Figure 3.2). These terminal blocks are marked TB1 through TB13. There can be an
additional four terminal blocks installed on the unit, each a part of an option module. Connect the
wires by stripping off 0.25" (6 mm) of insulation, inserting the wire into the proper terminal, and
tightening the screw firmly. Barrel-type crimp ferrules should be used on stranded wires. If
significant electrical noise is present, twisted shielded pairs are required.
TB1-12/24VDC TB2-PWR OUT TB3-RS485 TB10-RLY1 TB11-RLY2
ON/OFF CHASSIS A B A B M2 M1 G2 G1 M2 M1 G2 G1
V+ S+ S- V- V+ S+ S- V- V+ S+ S- V- V+ S+ S- V-
TB4-ANLG1 TB5-ANLG2 TB6-ANLG3 TB7-ANLG4
ADJUST
CONTRAST
RESET
ADDR + A - CHS A+ A- B+ B- M2 M1 G2 G1 M2 M1 G2 G1
TB8-POS PWR TB9-POS SIG TB13-RLY4 TB12-RLY3
Power, input and output wiring shall be in accordance with Class I Division 2 wiring methods,
Article 501-4 (b) of the National Electric Code, NFPA 70 for installations in the United States, or as
specified in Section 18-1J2 of the Canadian Electrical Code for installations within Canada, and in
accordance with the authority having jurisdiction.
www.nov.com/mdtotco
Document number 90-01
Revision E 3: Installation
Page 22 of 158
The instrument has input protection for both reverse polarity and over-voltage.
External or Jumper TB1
ON/OFF
Switch
ON
OFF
12 VDC @ 2A max + +
or
24 VDC @ 1A max - -
+ + + +
PWR
IN
- - - -
32
www.nov.com/mdtotco
Document number 90-01
3: Installation
Revision E
Page 23 of 158
A A A
B B B
A A A
B B B
Line Line Line
Monitor 1 Monitor 2 Monitor 3
Figure 3.5: Networking Series 2000 Instruments
+
V+ (+24V Output)
2-Wire
XMTR
S+ (+Signal In)
-
S- (-Signal In)
Jumper
V- (+24V Return)
www.nov.com/mdtotco
Document number 90-01
Revision E 3: Installation
Page 24 of 158
TB4
+5 VDC
To supply +5V to the sensor, the “A” terminal must be open (not connected), as shown in Figure
3.8.
TB8
+ +5 VDC Output
A N.C.
- Common
CHAS
+24 VDC
To supply +24V to the sensor, connect the “A” terminal to PWR common, as shown in Figure 3.9.
TB8
- Common
CHAS
32
www.nov.com/mdtotco
Document number 90-01
3: Installation
Revision E
Page 25 of 158
+12 VDC
To supply +12V to the sensor, install a 1000 ohm resistor, as shown in Figure 3.10.
TB8
A
1K Ω
- Common
CHAS
TTL (0 to +5 VDC)
To connect a payout sensor that provides 0 to +5 VDC signals, connect the wires as shown in
Figure 3.11. The TTL signals must be referenced to the common terminal, labeled -, of TB8.
TB9
A+ TTL Phase A
A- N.C.
B+ TTL Phase B
B- N.C.
www.nov.com/mdtotco
Document number 90-01
Revision E 3: Installation
Page 26 of 158
A+
A-
B+
B-
TB8
CHAS
32
www.nov.com/mdtotco
Document number 90-01
3: Installation
Revision E
Page 27 of 158
V+
TB9
A+
A-
B+
B-
NAMUR (Proximity)
To connect a payout sensor using NAMUR-type proximity sensing, connect the wires as shown in
Figure 3.14
TB8
A
Jumper +
- NAMUR
-
CHAS
TB9
A+ +
NAMUR
A- -
B+
B-
www.nov.com/mdtotco
Document number 90-01
Revision E 3: Installation
Page 28 of 158
N/A M2
N/A M1 1250 VA G-5AMP5 927003-008
Form-A G2 10 VA ODC5R 440-00080
G1
M2
Form-B
M1 1250 VA 5AMP5-B 927003-009
N/A G2 (no LED Indicator)
N/A G1
N/A M2
N/A M1 125 VA G-1AMP5-B 927003-010
G2
Form-B
G1
M2
Form-B M1 125 VA GX2-1AMP5-AB 927003-007
G2 125 VA GX2-1AMP5-AB-LC 927003-007
Form-A
G1 GX2-DRY5-AB 927003-006
Mechanical
Solid-State
N/A M2
N/A M1 2 AMPS G4 0AC5A 927003-011
G2 2 AMPS G4 0DC5 4400-00073
Form-A
G1
N/A M2
N/A M1 2 AMPS G4 0AC5A5 (NC) 4400-00073
G2
Form-B
G1
Caution: If Series 2000 Instruments are to be installed in Class I, Division 2, Group C and D
hazardous locations, use only the solid state relays listed above.
In Figure 3.15, Form A is normally open. If controlled in “Fail-safe” mode, then it will be closed
during normal operation, and open during alarm or power off.
Form B is normally closed. If controlled in "Fail-safe" mode, then it will be open during normal
operation, and closed during alarm or power off. Refer to Chapter 5, titled "Configuration and
Calibration" and Appendix B for more information.
32
www.nov.com/mdtotco
Document number 90-01
3: Installation
Revision E
Page 29 of 158
I I
OR
M V
V Load I Load
Com Com
Wiring for the ground-isolated analog output module is shown in Figure 3.17.
+24 Customer
(+24 V Not Isolated) Power Supply
vdc
4-20
Isolated + + ma
Internal
XMTR - 4-20
- ma Load
(Power Common
Not Isolated) Com
www.nov.com/mdtotco
Document number 90-01
Revision E 3: Installation
Page 30 of 158
RS - 232
RTS (Optional)
Tx
Rx
Com
Wiring for the RS-485 Aux Comm module is shown in Figure 3.19.
RS - 485
Talk (Optional)
Data (A)
Data (B)
Com
32
www.nov.com/mdtotco
Document number 90-01
3: Installation
Revision E
Page 31 of 158
SW1
SW1
SW2
SW2
Event input modules are designed to connect external switches to Series 2000 instruments. The
signal is activated by the external switch closing. SW1 is typically used as a push-to-print button.
Pressing this button causes a single transmission (SS command) out of the Aux Comm port. This
button must be a momentary switch (normally open).
SW2 is typically used to reset a value to zero (tare function). In the LM2000 product, payout gets
reset to zero by closing this switch.
www.nov.com/mdtotco
Document number 90-01
Revision E 3: Installation
Page 32 of 158
32
www.nov.com/mdtotco
Document number 90-01
Revision E
Page 33 of 158
4: Normal Operation
Power Up
When the unit is powered on, it performs a short self-test and then shows the Operator’s Screen. If
the unit is working properly, the power-on light in Figure 4.1 turns green and there is a short,
audible beep. The alarm light in Figure 4.1 turns green if there are no values in alarm; it turns
amber or red if there are one or more values in an alarm condition.
www.nov.com/mdtotco
Document number 90-01
Revision E 4: Normal Operation
Page 34 of 158
Operator’s Screen
The functions of the keys on the Operator’s Screen are shown in "Operator’s Screen Keypad
Function":
Operator’s Screen Keypad Function
Key Function
46
www.nov.com/mdtotco
Document number 90-01
4: Normal Operation
Revision E
Page 35 of 158
Keypad Functionality
During setup and calibration, the keys have the functions shown in Table 3-2. The exception to this
rule is the Operator’s Screen (Figure 4.1 on page 33) and the MORE key on the MenuLabel Bar
(Figure 4.4).
Keypad Functionality in Menu Screens
Move cursor
Enter/Select/Take action
Label Bars
The label bar is dynamic, meaning it changes to indicate the purpose of each key when the
operator navigates among the various screens and menus. There are five different label bars used
in the screens and menus:
www.nov.com/mdtotco
Document number 90-01
Revision E 4: Normal Operation
Page 36 of 158
EXIT ENTER
46
www.nov.com/mdtotco
Document number 90-01
4: Normal Operation
Revision E
Page 37 of 158
www.nov.com/mdtotco
Document number 90-01
Revision E 4: Normal Operation
Page 38 of 158
Use the DIGIT> key to move the cursor to the digit space to be modified. The digit that the
operator selects highlights in reverse video.
Use the INC (+) and DEC (-) keys to increase or decrease the highlighted digit. DEC can
also insert a negative sign (-) before a number if the cursor is placed in the space to the left
of the numbers.
Repeat steps 4 and 5 if needed.
Press ENTER when the desired payout number appears on the screen (or press CANCEL
to return to the original value). The system stores the number the operator entered and
redisplays the Screen Label Bar on the Payout Screen (Figure 4.9).
Press EXIT to return to the Operator’s Screen (Figure 4.1 on page 33). The payout value
that the operator entered appears as a digital value and is indicated on the bar graph.
46
www.nov.com/mdtotco
Document number 90-01
4: Normal Operation
Revision E
Page 39 of 158
CAL Key
Use the Cal Check Screen to engage remote calibration circuits through optional relays in order to
verify tension sensor calibration. Refer to Chapter 5, titled "Configuration and Calibration" for more
information.
www.nov.com/mdtotco
Document number 90-01
Revision E 4: Normal Operation
Page 40 of 158
Note: The TARGET1, TARGET2 and SERIAL NUMBER fields are for user convenience
only. The instrument has no “intelligence” regarding these numbers.
Figure 4.12: Data Entry Label Bar on the Cal Check Screen
Use the DIGIT> key to move the cursor to the digit space (one of seven) to be modified. The
digit that the operator selects highlights in reverse video.
Use the INC and DEC keys to increase or decrease the highlighted digit.
Repeat steps 4 and 5 as needed.
Press ENTER to store the value or CANCEL to abort the selection. Pressing either ENTER
or CANCEL causes the Cal Check Screen to redisplay the Item Label Bar as shown in
Figure 4.11.
Use NEXT to highlight another TARGET field or press EXIT to redisplay the Operator’s
Screen.
46
www.nov.com/mdtotco
Document number 90-01
4: Normal Operation
Revision E
Page 41 of 158
Screen Key
www.nov.com/mdtotco
Document number 90-01
Revision E 4: Normal Operation
Page 42 of 158
46
www.nov.com/mdtotco
Document number 90-01
4: Normal Operation
Revision E
Page 43 of 158
www.nov.com/mdtotco
Document number 90-01
Revision E 4: Normal Operation
Page 44 of 158
46
www.nov.com/mdtotco
Document number 90-01
4: Normal Operation
Revision E
Page 45 of 158
Menu Key
Press the MENU key to bring up the Main Menu in Figure 4.21. This menu is the gateway to all
configuration and calibration. For a description of the Main Menu, see Chapter 5, titled
"Configuration and Calibration".
www.nov.com/mdtotco
Document number 90-01
Revision E 4: Normal Operation
Page 46 of 158
46
www.nov.com/mdtotco
Document number 90-01
Revision E
Page 47 of 158
This chapter shows how to navigate through and use the Configuration and Calibration menus and
screens for changing system values and presets. The operator begins from the Operator’s Screen.
For a complete description of the Operator’s Screen, refer to Chapter 4, titled "Normal Operation".
Figure 5.1 on page 48 is a menu tree that shows the navigation routes from the Operator’s Screen.
The Configuration and Calibration menus and screens are available from the Main Menu.
www.nov.com/mdtotco
Page
Revision
Operator’s Screen
Document number
48 of 158
Main
Menu
76
Setpoints Adjust Scales Calibration Setup
Calibration 2 Point Data Analog Out Operator Units-of- Relay Network Aux Comm
Operator Units-of-
Privileges Measure
5: Configuration and Calibration
www.nov.com/mdtotco
Document number 90-01
5: Configuration and Calibration
Revision E
Page 49 of 158
To select an item in the Main Menu, use the NEXT key to highlight the desired item; then press the
SELECT key.
www.nov.com/mdtotco
Document number 90-01
Revision E 5: Configuration and Calibration
Page 50 of 158
Figure 5.4: Data Entry Label Bar on the Alarm Setpoints Screen
Use the DIGIT> key to move the digit cursor and highlight the digit to be modified.
Use the INC or DEC keys to increment or decrement the selected digit.
Press the ENTER key to accept the desired value or Cancel to return the original value. The
number field containing the modified digit displays the new value in reverse video. The Data-
Entry Label Bar changes back to the Item Label Bar (Figure 5.3).
76
www.nov.com/mdtotco
Document number 90-01
5: Configuration and Calibration
Revision E
Page 51 of 158
Repeat Step 1 through Step 7 above to change other alarm setpoints, if necessary.
Press the EXIT key. The Main Menu appears (Figure 5.2).
www.nov.com/mdtotco
Document number 90-01
Revision E 5: Configuration and Calibration
Page 52 of 158
Use the DIGIT> key to move the digit cursor and highlight the digit to be modified.
Use the INC or DEC keys to increment or decrement the selected digit.
Press the ENTER key to accept the desired value. The number field containing the modified
digit displays the new value in reverse video. The Backlight Brightness now conforms to the
new setting. The Item Label Bar is redisplayed (Figure 5.5).
Press the EXIT key. The Main Menu appears (Figure 5.2).
Adjusting the Backlight Color
Choose Display Adjust on the Main Menu to bring up the Display Adjust Screen (Figure
5.5). The Display Adjust Screen appears.
Use the NEXT key to select Backlight Color. The number field to the right of the Backlight
Color item is highlighted in reverse video to indicate it is active.
Press MODIFY to change the Backlight color value. The Display Adjust Screen goes into
configuration mode, the Data-Entry Label Bar appears (Figure 5.7), and the digit cursor is
displayed.
Press ENTER when complete
Use the DIGIT> key to move the digit cursor and highlight the digit to be modified.
Use the INC or DEC keys to increment or decrement the selected digit.
Press the ENTER key to accept the desired value. The number field containing the modified
digit displays the new value in reverse video. The LCD color conforms to the new screen
setting. The Item Label Bar is redisplayed (Figure 5.5).
Press the EXIT key. The Main Menu appears (Figure 5.2).
76
www.nov.com/mdtotco
Document number 90-01
5: Configuration and Calibration
Revision E
Page 53 of 158
Note: Exchange the Left Scale and Right Scale values to display graph scale changes in
the opposite direction.
www.nov.com/mdtotco
Document number 90-01
Revision E 5: Configuration and Calibration
Page 54 of 158
Press the MODIFY key. The Graph Scales Screen goes into configuration mode, the Data-
entry Label Bar appears (Figure 5.9), and the digit cursor is displayed.
Press ENTER when complete
Use the DIGIT> key to move the digit cursor and highlight the digit space to be modified.
Use the INC or DEC keys to increment or decrement the selected digit.
Press the ENTER key to accept the desired value or CANCEL to return the original value.
The number field containing the modified digit displays the new value in reverse video. The
Data-entry Label Bar changes back to the Screen Label Bar (Figure 4-8).
Repeat the above procedure to change other Graph Scales, if necessary.
Press the EXIT key. The Main Menu appears (Figure 5.2).
Calibration Submenu
Use the Calibration submenu to check sensor calibration, calibrate sensors, and to calibrate
analogoutput modules. Highlight one of the following options using the NEXT button:
Cal check
Two-point calibration
Data calibration
Analog-output calibration
Each of the above options is explained in the subsections that follow.
Choose Calibration on the Main Menu to bring up the Calibration Submenu (Figure 5.10).
76
www.nov.com/mdtotco
Document number 90-01
5: Configuration and Calibration
Revision E
Page 55 of 158
Note: The TARGET1, TARGET2 and SERIAL NUMBER fields are for user convenience
only. The instrument has no “intelligence” regarding these numbers.
www.nov.com/mdtotco
Document number 90-01
Revision E 5: Configuration and Calibration
Page 56 of 158
Use the DIGIT> key to move the cursor to the digit space (one of seven) to be modified. The
digit that the operator selects highlights in reverse video.
Use the INC and DEC keys to increase or decrease the highlighted digit.
Repeat Steps 4 and 5 as needed.
Press ENTER to store the value or CANCEL to abort the selection. Pressing either ENTER
or CANCEL causes the Cal Check Screen to redisplay the Item Label Bar as in Figure 5.11
Use NEXT to highlight another TARGET field or press EXIT to redisplay the
Operator’s Screen.
76
www.nov.com/mdtotco
Document number 90-01
5: Configuration and Calibration
Revision E
Page 57 of 158
Use the DIGIT> key to move the digit cursor to the digit space (one of seven) to be modified.
Use the INC and DEC keys to increment or decrement the highlighted digit.
Press ENTER to store the value or CANCEL to abort the selection. At this point, load on
sensor is read (captured) and assigned to display the value entered.
The Item Label Bar reappears.
Use the NEXT key to highlight the High Cal Point character field
Apply a known maximum load to the tension sensor.
www.nov.com/mdtotco
Document number 90-01
Revision E 5: Configuration and Calibration
Page 58 of 158
Press MODIFY.
The Data-Entry Label Bar (Figure 5.14) appears, and the digit cursor is displayed.
Use the DIGIT> key to move the digit cursor to the digit space (one of seven) to be modified.
Use the INC and DEC keys to increment or decrement the highlighted digit.
Press ENTER to store the value or CANCEL to abort the selection. At this point, load on
sensor is read (captured) and assigned to display the value entered.
The Item Label Bar reappears.
Press EXIT to redisplay the Operator’s Screen.
Note: It is very important to record the numbers shown for the offset and slope or each
sensor, after they are calibrated. Keep the recorded numbers in a secure place. They can
be used to restore calibration if the unit’s data gets corrupted.
Shunt Calibration (Remote Calibration)
Use the Two-Point Calibration Screen to do Shunt Calibration as follows:
Navigate to the Two-Point Calibration screen: Main Menu > Calibration > Two-Point
Calibration. The Two-Point Calibration Screen (Figure 5.13) appears.
Use the + or - keys to select TENSION.
Choose SENSOR next to the Signal From item.
Use the NEXT key to highlight the Low Cal Point character field.
Apply a zero load to the tension sensor.
Press MODIFY. The Two-Point Calibration Screen will show the Data-Entry Label Bar
(Figure 5.15), and the digit cursor is displayed.
Press ENTER when complete
Key in the minimum value (usually zero) for the Low Cal Point on the screen; press ENTER.
Use the NEXT key to highlight the Signal From character field.
Use the + or - keys to select CAL 1. This engages a Shunt Calibration circuit across the
tension sensor.
Use the NEXT key to highlight the High Cal Point character field.
Press MODIFY.
The Two-Point Calibration Screen will show the Data-Entry Label Bar (Figure 5.15), and the
digit cursor is displayed.
Use the DIGIT> key to move the digit cursor to the digit space (one of seven) to be modified.
Use the INC and DEC keys to increment or decrement the highlighted digit.
Press ENTER to store the value or CANCEL to return the original value. At this point, load
on sensor is read (captured) and assigned to display the value entered.
76
www.nov.com/mdtotco
Document number 90-01
5: Configuration and Calibration
Revision E
Page 59 of 158
To key in the appropriate tension value of the low tension standard for the Low Cal Point on
the screen, use the DIGIT> key to move the digit cursor to the digit space (one of seven) to
be modified.
Use the INC and DEC keys to increment or decrement the highlighted digit.
Press ENTER to store the value and capture the Low Cal Point; press CANCEL to abort the
selection. The Item Label Bar reappears.
Use the NEXT key to highlight the Signal From character field.
Use the + or - keys to select CAL 2.
Use the NEXT key to highlight the High Cal Point character field.
Press MODIFY. The Two-Point Calibration Screen will show the Data-Entry Label Bar
(Figure 5.15), and the digit cursor is displayed.
To key in the maximum tension value of the high load standard for the High Cal Point on the
screen.
Use the DIGIT> key to move the digit cursor to the digit space (one of seven) to be modified,
use the INC and DEC keys to increment or decrement the highlighted digit.
Press ENTER to store the value and capture the High Cal Point; press CANCEL to abort the
selection. At this point, load on sensor is read (captured) and assigned to display the value
entered. The Item Label Bar reappears.
After the calibration is completed, be sure to enter Sensor in the Signal From character
field.
www.nov.com/mdtotco
Document number 90-01
Revision E 5: Configuration and Calibration
Page 60 of 158
To enter the “offset” calibration factor, use DIGIT> to move the digit cursor and highlight the
desired whole number digit position. Use the INC and DEC keys to increment or decrement
the number in the highlighted digit space.
Once the desired number is entered, press ENTER to store the value or CANCEL to return
the original value.
Press Next to select Slope.
76
www.nov.com/mdtotco
Document number 90-01
5: Configuration and Calibration
Revision E
Page 61 of 158
Press Modify if changes are required. The Data-Entry Label bar appears (Figure 4-19), and
the cursor changes to the digital cursor.
Press ENTER when complete
To enter the “slope” calibration factor, use DIGIT> to move the digit cursor and highlight the
desired whole number digit position. Use the INC and DEC keys to increment or decrement
the number in the highlighted digit space.
Once the desired number is entered, press ENTER to store the value or CANCEL to return
the original value.
Repeat Steps 2 - 10 for tension and payout until all settings are correct.
Press Exit to return to the Cal Submenu.
Enter this number as the slope in the data calibration screen. With regard to payout, the offset
value is overridden by the front panel operator keys during normal operation. Therefore, set the
offset to zero.
www.nov.com/mdtotco
Document number 90-01
Revision E 5: Configuration and Calibration
Page 62 of 158
Use the DIGIT> key to move the digit cursor to the digit space (one of seven) to be modified.
Use the INC and DEC keys to increment or decrement the highlighted digit until the actual
payout value is displayed (typically zero).
Press ENTER to store the value or CANCEL to abort the selection. At this point, the payout
counter is read (captured) and assigned to display the value entered.
The Item Label Bar reappears.
Use the NEXT key to highlight the High Cal Point character field
Payout a measured amount of line.
Press MODIFY.
The Data-Entry Label Bar (Figure 5.14) appears, and the digit cursor is displayed.
Use the DIGIT> key to move the digit cursor to the digit space (one of seven) to be
modified.
Use the INC and DEC keys to increment or decrement the highlighted digit to enter the
amount of line paid out. This is the actual value, not the change ().
Press ENTER to store the value or CANCEL to abort the selection.
The Item Label Bar reappears.
76
www.nov.com/mdtotco
Document number 90-01
5: Configuration and Calibration
Revision E
Page 63 of 158
Note: It is very important to record the numbers shown for the offset and slope for each
sensor after they are calibrated. Keep the recorded numbers in a secure place. They can
be used to restore calibration if the calibration data gets corrupted.
In this screen, the left column represents display values; the right column represents the signal
level at the output. The screen in Figure 5.21 shows that an output signal level of 47.9% is
transmitted when the tension display value is 0 lbs. The output signal level of 87.2% is transmitted
when the tension display value is 20,000 lbs. In general, the% output value is adjusted until the
external device matches the display value (in the left column of Figure 5.21).
The input requirements of the external device will determine what type of analog output module
you need and how to calibrate it. The analog output signal value is represented on the unit’s
display as a percentage. The relationship between this percent value and the value of the output
signal (voltage or milliampere) depends on the type of module installed. Also, the exact value of
the output will differ slightly from unit to unit. Prior to connecting the output signal to the external
device, you should set the output signal value (percent) to the nominal value. For example, if a
display of measured load from 0 to 20000 lbs must create an output signal of 0–10V, then set the
analog output calibration screen as shown above in Figure 5.21. Refer to the following tables for
nominal values for isolated and non-isolated analog output modules.
www.nov.com/mdtotco
Document number 90-01
Revision E 5: Configuration and Calibration
Page 64 of 158
-12.18 0
-10 8.6
-5 28.2
0 47.9
+5 67.5
+10 87.2
+13.27 100
0 0
4 18.3
12 55.0
20 91.7
21.8 100
3.13 0
4 4.3
12 43.5
20 82.8
23.5 100
76
www.nov.com/mdtotco
Document number 90-01
5: Configuration and Calibration
Revision E
Page 65 of 158
The procedure for using the Analog Output Calibration Screen is as follows:
Navigate to the Analog Output Calibration Screen: Main Menu > Calibration > Analog Output
Cal. The Analog Output Calibration Screen (Figure 5.21) appears.
Use the + or - keys to select TENSION, PAYOUT or SPEED.
Use the Next key to highlight the top value in the left column of the screen. This number
represents the display value for the point being calibrated.
Press Modify. The Data-entry Label Bar appears (Figure 5.22), and the digit cursor is
displayed.
Press ENTER when complete
Figure 5.22: Data-entry Label Bar on the Analog Output Cal Screen
Use DIGIT> to move the digit cursor and highlight the desired whole number digit position.
Use the INC and DEC keys to increment or decrement the number in the highlighted digit
space.
Once the desired number is entered, press ENTER to store the value or Cancel to return the
original value.
Use the Next key to highlight the top value in the right column of the screen. This number
represents the value of the output signal.
Note: During this procedure, the output changes in real-time as you change the %-output
value on the display. Change the value until the external device reads the same value in
the analog output calibration screen’s left column (if the external device is a display unit).
Alternatively, change the “%output” value to achieve an accurate reading on an external
current meter or volt meter.
Press Modify. The Data-entry Label Bar appears (Figure 5.22), and the digit cursor is
displayed.
Use DIGIT> to move the digit cursor and highlight the desired whole number digit position.
Use the INC and DEC keys to increment or decrement the number in the highlighted digit
space.
Once the desired number is entered and the external device exhibits the correct value, press
ENTER to store the value or Cancel to return the original value.
Use the Next key to highlight the bottom value in the left column of the screen. This number
represents the display value for the point being calibrated.
Press Modify. The Data-entry Label Bar appears (Figure 5.22), and the digit cursor is
displayed.
Use DIGIT> to move the digit cursor and highlight the desired whole number digit position.
Use the INC and DEC keys to increment or decrement the number in the highlighted digit
space. Make number selection based on the external display device or a digital mulitmeter.
Once the desired number is entered, press ENTER to store the value or Cancel to return the
original value.
www.nov.com/mdtotco
Document number 90-01
Revision E 5: Configuration and Calibration
Page 66 of 158
Use the Next key to highlight the bottom value in the right column of the screen. This
number represents the value of the output signal.
Note: During this procedure, the output changes in real-time as you change the %-output
value on the display. Change the value until the external device reads the same value in
the analog output calibration screen’s left column (if the external device is a display unit).
Alternatively, change the “%output” value to achieve an accurate reading on an external
current meter or volt meter.
To change this output value to match the unit’s corresponding displayed value, press
Modify. The Data-entry Label Bar appears (Figure 5.22), and the digit cursor is displayed.
Use DIGIT> to move the digit cursor and highlight the desired whole number digit position.
Use the INC and DEC keys to increment or decrement the number in the highlighted digit
space. Make number selection based on the external display device or a digital mulitmeter.
Once the desired number is entered, press ENTER to store the value or Cancel to return the
original value.
Press EXIT to redisplay the Calibration Submenu.
76
www.nov.com/mdtotco
Document number 90-01
5: Configuration and Calibration
Revision E
Page 67 of 158
Note: All privileges are accessible with the use of the product PASSWORD (even those set
to N).
Select the desired privilege using the Next key. The corresponding privilege (Y or N)
highlights.
Use the + or - keys to change the selected privilege from Y to N as required. The highlighted
privilege, Y or N, changes to its opposite.
Repeat the above procedure for each privilege to be changed.
Once the desired privileges have been entered, press EXIT.
Note: If the operator chooses N for System Setup, he will not be able to enter the Operator
Privileges Screen without a password.
www.nov.com/mdtotco
Document number 90-01
Revision E 5: Configuration and Calibration
Page 68 of 158
Caution: Units of measure are displayed text only. The calibration is NOT adjusted
automatically to follow the units. If units are changed, recalibration is required.
76
www.nov.com/mdtotco
Document number 90-01
5: Configuration and Calibration
Revision E
Page 69 of 158
Press Modify (if a change is needed). Otherwise, go to Step 10. The Analog Units of
Measure Submenu goes into configuration mode, the Data-entry Label Bar appears (Figure
5.26), and the cursor changes to the digit cursor.
Press ENTER when complete
Use DIGIT> to move the digit cursor and highlight the desired whole number digit position.
Use the INC and DEC keys to increment or decrement the number in the highlighted digit
space.
Once the desired number is entered, press ENTER to store the value or Cancel to return the
original value.
Press the Next key to highlight TENSION, PAYOUT, or SPEED.
Repeat Steps 2 through 8 until Tension, Payout, and Speed are configured.
Press Exit. The System Setup Submenu reappears (Figure 5.23).
www.nov.com/mdtotco
Document number 90-01
Revision E 5: Configuration and Calibration
Page 70 of 158
Use the + or - keys to select Alarm Assignment or Relay Type. The operator’s selection
highlights with a dark field to indicate it is active.
Press Select. The submenu corresponding with the operator’s selection, Alarm
Assignments or Relay Type, appears.
Alarm Assignments Submenu
There are 12 alarm sources:
Four setpoints for tension
Four setpoints for payout
Four setpoints for speed
There are four relays (alarm destinations). The Alarm Assignments Screen is used to assign each
alarm source to control any or all relays (or no relays).
76
www.nov.com/mdtotco
Document number 90-01
5: Configuration and Calibration
Revision E
Page 71 of 158
Use the + or - keys until the desired relay type appears highlighted. Possible relay type
selections are:
Latched Horn
Latched Relay
Unlatched Horn
Unlatched Relay
Remote Cal 1
Remote Cal 2
Horn relays are normally deenergized. General purpose relays are normally energized so
that they will fail-safe in their alarm mode if power is lost.
www.nov.com/mdtotco
Document number 90-01
Revision E 5: Configuration and Calibration
Page 72 of 158
Press the NEXT key to go to another relay if necessary. The corresponding Relay Type field
highlights.
Repeat Steps 2 and 3 until all settings are correct.
Press the EXIT key.
The System Setup Submenu reappears (Figure 5.23).
Network Setup Screen
Use the Network Setup Screen to designate a unit as a sensor input or remote display unit.
For units designated as Sensor Input Units, it is necessary to set the rotary switch on the back of
the unit until the first unit reads “1” on the Rotary Switch field of the NetworkSetup screen. Then
use the rotary switch to set each succeeding unit such that each unit has a unique number.
For units designated as Remote Displays, it is necessary to select the Remote Display number in
the first field of the Network Setup Screen. Then use the rotary switch to match the winch number
that will most often be displayed by this remote. In the Operator’s Screen, select the winch number
to be displayed now. For more information on network setup, refer to Appendix G.
Use the + or - keys until the top field reads SENSOR INPUT UNIT.
The top field reads Sensor Input Unit highlighted in reverse video.
Adjust the rotary switch on the back of the unit to the appropriate setting.
The number to the right of the Rotary Switch label reflects the current rotary switch setting.
76
www.nov.com/mdtotco
Document number 90-01
5: Configuration and Calibration
Revision E
Page 73 of 158
www.nov.com/mdtotco
Document number 90-01
Revision E 5: Configuration and Calibration
Page 74 of 158
Note: The default number in the Data Interval (1/4 Sec) character field is 4. (A setting of 1
means that data is transmitted every 1/4 sec; a setting of 4 means that data is transmitted
every second; 8 = every 2 sec, etc.).
Press Modify.
The Data-entry Label Bar appears (Figure 5.32), and the digit cursor is displayed.
Figure 5.32: Data-entry Label Bar on the Aux. Comm. Setup Screen
To change the data interval, use DIGIT> to move the digit cursor and highlight the desired
whole number digit position.
Use the INC and DEC keys to increment or decrement the number in the highlighted digit
space.
Once the desired number is entered, press ENTER to store the value or CANCEL to return
to the original value.
Press the NEXT key to highlight the PROTOCOL character field.
Use the + or - keys to scroll through the protocol choices (1, 2, or continuous mode).
Press the NEXT key to highlight the DELIMITER character field.
Use the + or - keys to scroll through the delimiter choices (comma, space, or tab).
Press the NEXT key to highlight the BAUD character field.
9600 baud is the default.
76
www.nov.com/mdtotco
Document number 90-01
5: Configuration and Calibration
Revision E
Page 75 of 158
Use the + or - keys to scroll through the baud-rate choices (9600, 19.2K, 300, 600, 1200,
2400, 4800).
Use the NEXT key to highlight the Parity Enabled? character field.
Use the + or - keys to choose Y for Yes or N for No.
N is the default.
Use the NEXT key to move the highlight to the Odd Parity? character field.
www.nov.com/mdtotco
Document number 90-01
Revision E 5: Configuration and Calibration
Page 76 of 158
76
www.nov.com/mdtotco
Document number 90-01
Revision E
Page 77 of 158
6: Troubleshooting
Overview
There are three types of diagnostics available on Series 2000 Instruments:
Power-On Self Test—test routines that the unit runs to test its internal subsystems. These
tests are performed at power up and after reset during normal operation.
Run Time Background Tests—test routines that run automatically during normal operation
that detect and record errors in the entire system. This includes creating error logs.
Troubleshooting Diagnostics—tools that allow you to view the raw sensor data. By changing
the value of the signal that is input to the unit, you can look for the expected changes in the
raw data. These diagnostics also control output devices such as relays and analog outputs.
www.nov.com/mdtotco
Page
Revision
Document number
Main
Menu
press the + key
90
Navigating the Menus
www.nov.com/mdtotco
Document number 90-01
6: Troubleshooting
Revision E
Page 79 of 158
Diagnostic Menus
The following subsections describe the individual diagnostic menus and examples of their use in
troubleshooting.
-15 V 115-145
+15 V 90-115
+24 V 150-180
+3.84 215-240
www.nov.com/mdtotco
Document number 90-01
Revision E 6: Troubleshooting
Page 80 of 158
Notice that the remote display’s screen is slightly different than the sensor display. It includes
information on sensor unit errors as well as remote unit errors.
Each unit keeps a count of errors it senses during operation. If there seems to be some kind of
problem during operation, check the error logs to narrow down where the problem might be. Select
an error log by pressing the:
90
www.nov.com/mdtotco
Document number 90-01
6: Troubleshooting
Revision E
Page 81 of 158
Recv CRC errors—number of times the system has detected errors in CRC of records
coming in from the network.
Recv sync errors—number of times the system has detected an unexpected sync
character in a record.
Buffers full—number of times the buffer was full when it needed to receive a record from
the network.
UART Errors—number of errors generated by the UART processor interface.
Search Mode—number of times the system has been put into search mode.
Timeouts—number of times the system has not received aresponse from a device it has
sent a request to.
www.nov.com/mdtotco
Document number 90-01
Revision E 6: Troubleshooting
Page 82 of 158
ADC Data Ready Errors—number of times the ADC has failed to convert data within its
allotted time.
ADC Cal Errors—number of times the ADC failed to calibrate a channel.
-15 VDC—number of failures recorded from this power supply.
+15 VDC—number of failures recorded from this power supply.
+24 VDC—number of failures recorded from this power supply.
+3.84 VDC—number of failures recorded from this power supply.
90
www.nov.com/mdtotco
Document number 90-01
6: Troubleshooting
Revision E
Page 83 of 158
Self Tests
In the diagnostic menu, select self test. The display will present you with the following menu:
www.nov.com/mdtotco
Document number 90-01
Revision E 6: Troubleshooting
Page 84 of 158
90
www.nov.com/mdtotco
Document number 90-01
6: Troubleshooting
Revision E
Page 85 of 158
www.nov.com/mdtotco
Document number 90-01
Revision E 6: Troubleshooting
Page 86 of 158
Relays-Alar ms Disabled
This diagnostic allows you to activate and deactivate relays to ensure they are working properly.
Once you have selected this option, use the key to select the relay you want to toggle on/off. Once
this relay is highlighted, press the key, and then the + or - keys to toggle the relay.
Warning: This test procedure disables system alarm checking. This means that
alarm conditions sensed while toggling the relays on and off will not generate an
alarm in the system.
90
www.nov.com/mdtotco
Document number 90-01
6: Troubleshooting
Revision E
Page 87 of 158
www.nov.com/mdtotco
Document number 90-01
Revision E 6: Troubleshooting
Page 88 of 158
Refer to view A of the 220870 drawing and locate the following EPROM’s: U10, U27, and
U30. Carefully pry the EPROM’s from their sockets.
Warning: This unit can be damaged by static electricity. If a static controlled work
station is not available, always keep one hand touching one of the six metal hex-
shaped standoffs the entire time the rear cover is removed. This will help avoid
static build-up.
Gently place each new EPROM into its correct socket. Each EPROM is labeled U10, U27,
or U30. The chip MUST be oriented properly so that pin #1 is in the correct location. Note
that one end has a dot over pin #1 and a notched or chamfered corner. If there is any
question, remember...IF THE CHIP IS INSTALLED BACKWARDS, IT WILL BE
DESTROYED. Once properly aligned, press firmly to seat the chip in place.
Replace the field wiring board, rear cover and wiring. Be sure that the connector is properly
aligned and press it into place. Replace the six black screws.
Power the unit up and allow it to go through selftest.
Initialize the software as follows:
Press the F1 key. Wait one second. (The alarm screen is displayed.)
Press the F2 key. Wait one second. (The display does not change.)
Press the F3 key. Wait one second. (The display does not change.)
Press the F4 key. Wait one second. (An initialization screen is displayed.)
Press the NEXT key. (The cursor moves to the number field.)
Press MODIFY. (The unit enters the data entry mode.)
Press DIGIT>. (The cursor moves one space to the right.)
Enter the product initialization code using the DIGIT>, INC, or DEC buttons as required.
When the correct number is displayed, then press CONFIG. (Consult the factory for the
initialization code.) Standard and Single-Axis line tension-payout-speed systems have an
initialization code of 6538.
Wait 10 seconds.
Turn the power off for 5 seconds.
Turn the power back on. The unit should now be ready for operation. Verify that all
calibration constants and configuration parameters are correct.
90
www.nov.com/mdtotco
Document number 90-01
6: Troubleshooting
Revision E
Page 89 of 158
Error Symbols
The following table lists the various error symbols that an S-2000 instrument can display, their
meaning, and their possible causes.
Error Symbols
www.nov.com/mdtotco
Document number 90-01
Revision E 6: Troubleshooting
Page 90 of 158
90
www.nov.com/mdtotco
Document number 90-01
Revision E
Page 91 of 158
A: Glossary
Label Bar A display just above the keypad which shows the function of each key. The keys
in the keypad can have different functions in different levels of the menu and
screen hierarchy.
Load Cell A type of sensor
Menu A display of options that allows the operator to change system values or to
navigate to a screen or a submenu.
N.C. Normally Closed or Not Connected
Navigation Moving through the hierarchy of menus, submenus, and screens by pressing
the keys below the display.
Networking Connectiong two or more instrument together with RS-485 communications
links so that the operator can view the information from a remote display.
Operator’s This is the default operating screen. The operator can choose from several
Screen different screen layouts to optimize parameter viewing.
RS-232 Recommended Standard 232. Describes the signal function and voltage levels
of a communication standard. Typically used by PCs.
RS-485 Recommended Standard 485. Describes the signal function and voltage levels
of a communication standard. Typically used in party-line (or multi-drop)
communications.
Reverse Video A convention to indicate a menu item is selected and active. In reverse video,
what is normally white on the LCD changes to black, and what is normally black
changes to white
Screen Displayed image format on the front of the instrument.
www.nov.com/mdtotco
Document number 90-01
Revision E A: Glossary
Page 92 of 158
92
www.nov.com/mdtotco
Document number 90-01
Revision E
Page 93 of 158
Overview
The purpose of this Appendix is to describe the alarm and relay characteristics for the M/D TOTCO
Series 2000 instrument family. Items discussed include contact types, software control of relays,
alarm operation, and functions accessible from the front panel of the instrument.
Series 2000 instruments are equipped with 4 relay sockets. Optional relays can be installed in
these sockets for alarm or calibration purposes. The sockets can accept various types of relays, as
described below. Programming of the relays is achieved using the keys below the display.
de-energized energized
Form B contacts are normally closed (N.C.); energizing the coil opens the contacts.
Form-B
de-energized energized
www.nov.com/mdtotco
Document number 90-01
Revision E B: Alarms and Relays
Page 94 of 158
Form-C contacts act like a single pole, double throw (SPDT) switch. Internal to this type of
contact is both a normally open and a normally closed contact, and a common. The Series
2000 does not have a Form-C contact; instead, a relay can be installed that has both a
Form-A and a Form-B contact (Form-1A/1B). This is more flexible because you can wire it to
behave like a Form-C, or use it as two separate contacts.
Form-C
normally closed (N.C.)
common
Form A Form A
Form B Form B
de-energized energized
Note: When choosing a solid state relay, you must know what kind of power the contact will
be switching. Solid state relays for AC power are NOT interchangeable with solid state
relays for DC power.
Note: Fail-safe describes the way the coil of the relay is controlled; it has nothing to do with
the contacts of the relay being normally open or normally closed.
98
www.nov.com/mdtotco
Document number 90-01
B: Alarms and Relays
Revision E
Page 95 of 158
Cal Relays
These are special purpose relays meant for controlling calibration circuits in external signal
conditioners. Up to two cal relays can be programmed via the front panel. In applications that use
shunt calibration, only one cal relay is needed. This relay is normally installed in relay socket 4 on
the back of a Series 2000 instrument. In applications using voltage substitution, both cal relays
must be used.
Note: Cal relays take up space that could otherwise be used for alarm relays. If a relay
socket is defined as a cal relay, that socket cannot be used for any other purpose until the
unit is reprogrammed.
On strain gauge input versions of Series 2000 instruments, there is one cal relay built into the
system internally, in addition to the 4 external relays on the back of the unit. This built in relay has
two normally open contacts. When these contacts are closed, they can activate two shuntcal
resistors located on the terminal blocks on the rear of the unit.
Alarm Operation
Each measured or calculated channel has four alarm setpoints (2 high & 2 low). Each setpoint is
individually programmed via the front panel. Each alarm source can be individually assigned to
one or multiple relays. It is also possible to assign multiple alarm sources to a single relay.
TRI-COLOR LED—located on the front of the display, second from the left. The different colors of
the LED represent differing levels of alarm.
Green (normal operation)= no alarm
Yellow (caution)= at least one alarm source is in low or high alarm
Red (overload)= at least one alarm source is in high-high or low-low alarm
www.nov.com/mdtotco
Document number 90-01
Revision E B: Alarms and Relays
Page 96 of 158
FLASHING DIGITS—when a particular parameter goes into alarm condition, the digits displaying
the value of the parameter on the display start flashing. They flash even if the alarm is
acknowledged, and only stop flashing when the alarm condition goes away.
AUDIBLE BEEP—sounds when at least one alarm source is in low or high alarm condition. The
beep is silenced by pressing the ACK button, or when all of the alarm conditions are gone.
Caution: The audible beep on Series 2000 instruments is not meant to be used as a
warning device in industrial environments because it is not loud enough to serve that
purpose. Its primary purpose is to provide audio feedback for key presses.
AUX COMM ALARM STATUS CHARACTERS—the alarm status of each parameter is also
transmitted on the aux comm data package. Refer to the aux comm protocol for more information.
RELAYS—react as programmed to indicate alarm conditions.
ACK & SILENCE—front panel button that ACKnowledges alarms. Pressing this button during
alarm conditions silences the audible beep and resets all relays programmed as horns. The
operator has two choices: continue with no further change or clear all general purpose relays and
continue.
Alar m Assignments
Each alarm source can be assigned to control any or all alarm destinations (in any combination
desired):
Low Relay 3
Low-Low Relay 4
A setpoint value for each of these alarm sources can be entered via the front panel.
98
www.nov.com/mdtotco
Document number 90-01
B: Alarms and Relays
Revision E
Page 97 of 158
Relay Types
Relay functions must be selected via the front panel of Series 2000 instruments, with six relay
types to choose from. This selection determines the way the software controls the relay; it does
NOT refer to the type of physical relay installed. The choices are as follows:
UNLATCHED HORN—meant to drive external horns
LATCHED HORN—meant to drive external horns
UNLATCHED RELAY—fail-safe control of general purpose alarm relay
LATCHED RELAY—fail-safe control of general purpose alarm relay
CAL 1—meant to control external calibration circuits
CAL 2—meant to control external calibration circuits
High
Setpoint
Value
Load
Low
Setpoint
Low-Low
Setpoint
Time
The diagram above shows how a Series 2000 instrument reacts to changes in load value
overtime. The darkened bars represent the “on” or “activated” condition of a component.
www.nov.com/mdtotco
Document number 90-01
Revision E B: Alarms and Relays
Page 98 of 158
98
www.nov.com/mdtotco
Document number 90-01
Revision E
Page 99 of 158
Overview
There are four recommended techniques for calibrating strain gauge sensors:
Physical Calibration—calibrates the sensor with an accurately known load
Shunt Calibration (Shuntcal)—shunts one leg of the strain gauge bridge with a precision
resistor
Voltage Substitution—substitutes the entire strain gauge bridge with a precision resistor
divider in order to produce a specific mV/V value
Bridge Substitution—same as voltage substitution, only the strain gauge bridge is used as
part of the resistor divider
Physical calibration is the most desirable technique because it takes into account all possible parts
of an installation by directly loading the sensor. However, many applications will not allow for a
physical calibration, which means that the sensor will have to be calibrated using simulated loads.
Each of the calibration techniques described in the following sections assumes the sensor reacts
in a linear fashion to increasing load. This means that the calibration techniques are all of a 2-point
variety, with the two points defining the linear response of the sensor.
As with any 2-point calibration, the farther apart the data points are, the more accurate the
calibration. As a general rule, try to take one data point from somewhere less than 25% of total
capacity, and a data point from somewhere above 75% capacity.
Physical Calibration
Calibration is done by applying an accurately known load to the system, and adjusting the display
to match the known load value. It is recommended that every installation get a physical calibration
at least once, preferably at commissioning.
www.nov.com/mdtotco
Document number 90-01
Revision E C: Strain Gauge Sensor
Page 100 of 158
-Sig +Sig
Shuntcal Resistor
106
www.nov.com/mdtotco
Document number 90-01
C: Strain Gauge Sensor
Revision E
Page 101 of 158
Shuntcal is a relative calibration, creating a known (change) in the output. For example, assume
that the shuntcal value for a given application is 80%. If the load displayed prior to activating the
shuntcal switch is 12%, then the display should read 92% after activating the shuntcal circuit. This
change depends on the value of the resistor and the resistance of the bridge itself. Refer to the
following table:
Shuntcal Value Approximations
Note: The output of each individual load cell is recorded during its factory calibration.
These output values are supplied on the Calibration Certificate that accompanies each
load cell from the factory. Shuntcal resistor and data are not automatically supplied with all
sensors and must be specifically ordered with the system.
Shuntcal Procedure
Remove all load so that the system is in the condition where the measured value is zero (or
as close to zero as possible).
Adjust the display to read zero (or the known minimum load).
Activate shuntcal and adjust the display to read the value that was recorded when the
shuntcal circuit was activated during physical calibration (plus any minimum load present
from the step above). If no previous physical calibration was performed, use the shuntcal
value from the load cell calibration certificate.
Note: The value on the calibration certificate refers to the actual load applied to the load
cell, not necessarily the measured load shown on the display.
Deactivate shuntcal and verify that the display reads zero (or the known minimum load
again).
www.nov.com/mdtotco
Document number 90-01
Revision E C: Strain Gauge Sensor
Page 102 of 158
Voltage Substitution
This method of calibration replaces the load cell’s strain gauge bridge with a precision resistor
divider. The divider consists of three resistors in series with excitation applied to the endpoints, and
the signal read off the center resistor. See Figure C.2
+Ex +Ex
+500R +500R
+Sig +Sig
R 0.1 R
-Sig -Sig
+500R +500R
-Ex -Ex
High Cal Point Divider Low Cal Point Divider
Figure C.2: Resistor Divider for Voltage Substitution Calibration (resistor values shown as
examples only)
Depending on the values of the divider’s resistors, any "mV/V" value can be created. The center
resistor has a much smaller value than the other two, and the low cal point center resistor has a
small value compared to the high cal point center resistor. If the low cal point center resistor is
actually a short between +Sig and -Sig, then a true 0.0 mV/V value is applied. Two separate
dividers give you the high and low calibration points.
Voltage substitution creates a specific mV/V value that is independent of the load on the load cell.
This makes it a popular calibration technique for those applications where the load cell is a
structural load-bearing member, when the load cell is monitoring mooring lines for large permanent
structures, or any time load cannot be removed for calibration.
106
www.nov.com/mdtotco
Document number 90-01
C: Strain Gauge Sensor
Revision E
Page 103 of 158
Caution: The values on the calibration certificate pertain to load on the load cell, and do not
take into account any tare load (zero offset).
Note: Zero offset cal factors are not routinely recorded on the calibration certificate, except
when specifically requested on special "turn-key" systems.
In some cases, the values substituted are zero and full scale. This can create problems if the
instrumentation does not have sufficient headroom below zero and above full scale. This
headroom is required as a result of tare loads and load cell output tolerances. In cases where this
is a concern, mV/V values can be chosen that will prevent this from being a problem. For example,
the substituted values can be approximately 15% and 90% of full scale. This creates sufficient
headroom even in cases where the instrumentation cannot display values below zero or above full
scale.
As a troubleshooting tool, this method verifies all circuits from the signal conditioner input through
the display. It does not verify the load cell or the load cell wiring.
www.nov.com/mdtotco
Document number 90-01
Revision E C: Strain Gauge Sensor
Page 104 of 158
Bridge Substitution
This calibration method is similar to voltage substitution, except that the center resistor in the
resistor divider is the strain gauge bridge itself. See Figure C.3
+Ex +Ex
175KΩ 1750KΩ
+Sig +Sig
Strain Strain
350Ω Gauge 350Ω Gauge
Bridge Bridge
-Sig -Sig
175KΩ 1750KΩ
-Ex -Ex
High Cal Point Divider Low Cal Point Divider
Figure C.3: Resistor Divider for Bridge Substitution Calibration
The advantages of this method are that it checks the two signal wires to the bridge and checks that
the input resistance of the bridge is reasonably close to what it should be. The disadvantages of
this method are that it is more difficult and costly to implement, calibration constants become more
difficult to deal with, and it can actually make troubleshooting the system more complicated.
Application Notes
When a system is defined, the decision of what kind of electrical calibration to use should be
settled during system definition prior to sale. As a rule of thumb:
If the load can be removed routinely, such as a simple crane application, then choose shunt
calibration.
If the load cannot be removed easily, then choose voltage substitution.
At this time, one should also start planning for a physical calibration at the commissioning of the
system. This physical calibration must be performed at least once so that accurate cal constants
can be acquired. These values become the calibration constants for future maintenance and
routine calibrations.
106
www.nov.com/mdtotco
Document number 90-01
C: Strain Gauge Sensor
Revision E
Page 105 of 158
Routine Calibration
Provided you have calibration constants from a good physical calibration, frequent checks of
system calibration can be accomplished by closing a switch and making sure the displayed value
matches the recorded value.
This means that in most applications, daily calibration checks can be performed by the operator.
Maintenance personnel are not needed except to resolve situations where the displayed value
does NOT match the recorded value.
Product Features
2077 Motherboard
Shuntcal 1 - standard
Shuntcal 2 - optional
Voltage Substitution - optional, full scale configure per job
Voltage Substitution - optional, shorted for zero
SC103
Voltage Substitution (0.2 mV/V) for zero
Voltage Substitution (0.4 mV/V) for 0.5 mV/V sensor (Dipswitch selectable)
Voltage Substitution (1.2 mV/V) for 1.5 mV/V sensor (Dipswitch selectable)
Voltage Substitution (1.6 mV/V) for 2.0 mV/V sensor (Dipswitch selectable)
Shuntcal or other values of Voltage Substitution available on special order
Series 2000 Instruments
Optional calibration relays for driving remote calibration circuits—shuntcal, voltage
substitution, bridge substitution, or others
Strain gauge input version has on-board shuntcal relays standard—shuntcal resistor(s)
mounted on the terminal blocks on the rear of the unit
www.nov.com/mdtotco
Document number 90-01
Revision E C: Strain Gauge Sensor
Page 106 of 158
106
www.nov.com/mdtotco
Document number 90-01
Revision E
Page 107 of 158
Overview
This chapter provides technical information to assist field personnel in troubleshooting load cells.
The first part of the chapter discusses background information about load cells. The last part of the
chapter deals with the techniques used to troubleshoot load cells.
Caution: The following information is not meant to be a substitute for proper training. This
information is provided as a guideline for technicians.
The exact absolute value of the excitation voltage is typically not the primary concern. The primary
concern is the stability of the voltage level over time and changes in temperature. Exceptions to
this are instruments which "ratiometrically" track change in excitation voltage over time or change
in temperature. These instruments internally compensate for the drift in excitation voltage. M/D
TOTCO (Metrox) instruments typically are configured for stable, absolute voltage supplies.
The excitation voltage must also be as free of electrical noise as possible. Excitation voltage must
be measured at the load cell if possible. There will be a voltage drop in the interconnect cables
between the signal conditioner and the load cell itself. In applications using long cable runs (>50')
or requiring higher accuracy, remote sense techniques compensate for this voltage drop, which
will be discussed later in this appendix.
www.nov.com/mdtotco
Document number 90-01
Revision E D: Load Cell Troubleshooting
Page 108 of 158
-Exitation
Voltage (-Ex)
Figure D.1: Signal Detection Points
The value of +signal and -signal have meaning only in reference to each other. The signal is not
measured with respect to ground. The signal must be measured with equipment having high
impedance inputs, otherwise current will be pulled through the signal wires causing errors. Most
modern digital volt meters (DVM’s) have sufficiently high impedance inputs so as not to degrade
the signal.
The value of the signal output from a strain gauge bridge is very low, typically 0–10 mV or 0–20
mV. Therefore, any volt meter used to measure these signals must have the capability to measure
these low signal levels. For troubleshooting in the field, the minimum voltmeter requirement is
typically a 4 ½ digit DVM.
The value of the signal is proportional to both the excitation voltage and the applied load.
Therefore, the unit of measure for these signals is expressed in millivolts of signal per volt of
excitation (mV/V). That is, for each volt of excitation supplied, you get some millivolt of signal. For
example, if a load cell has an output of 2.0mV/V, and it has an excitation voltage of 10VDC applied
to it, then the signal level should be 20mV (2mV for every volt of excitation; 2 x 10=20).
112
www.nov.com/mdtotco
Document number 90-01
D: Load Cell Troubleshooting
Revision E
Page 109 of 158
Bridge Resistance
Generally, Metrox sensors from M/D TOTCO are configured in a Wheatstone bridge as shown in
Figure D.1. Total bridge input resistance is measured from (+)Ex to (-)Ex. Bridge output resistance
is measured from (+) Signal to (-) Signal.
Note: When measuring bridge resistance, the bridge must be disconnected from the
instrumentation and there must be absolutely NO electrical connection to the bridge except
the ohmmeter.
Typical bridge output resistance values are 350 or 700 ohms . Input resistance should be the same
as the output resistance, however, there are systems where the input resistance is greater due to
internal compensation, which is usually less than 100 ohms . Refer to the specification control
drawing for the nominal bridge resistance of a particular load cell.
Insulation Resistance
The body of each load cell is electrically isolated from the bridge. To measure the insulation
resistance of the load cell, the bridge must be disconnected from its instrumentation, and there
must be NO power applied. Resistance can be measured from any or all bridge leads to the body
of the load cell. This resistance must be greater than 100M ohms. Typically, values are greater
than 5000M ohms.
Warning: DO NOT use a high-voltage Hypot tester! Test equipment must apply no
more than 50 VDC. If internal varisters are used for lightning protection, the test
equipment must apply no more than 20 VDC.
The average DVM applies less than 9 VDC, so it shouldn’t cause any damage. However, many of
these DVM’s are often unable to measure as high as 100M ohms.
Interconnect Techniques
All full bridge load cells require a minimum of four interconnect wires:
+E x
+Sig Signal Conditioner,
Display,
-Ex Recorder,
+Sig Etc.
Interconnect cables need to be twisted, shielded pairs. The two excitation lines are one twisted
pair, and the signal lines are a second twisted pair. Shields are normally connected to ground only
at the instrumentation end and not at the sensor end.
If even a small amount of current flows through the excitation wires, it creates a voltage drop. This
drop means that the excitation voltage at the load cell is less than the excitation voltage at the
instrument, which causes errors in the measurements being taken. To overcome this, you can run
an additional twisted pair of wires from the load cell to the instrument, so the instrument can
www.nov.com/mdtotco
Document number 90-01
Revision E D: Load Cell Troubleshooting
Page 110 of 158
remotely sense the voltage at the load cell. The instrument can then compensate for the voltage
drop in the excitation wires. Remote sense is recommended for all applications with long cable
runs, or for applications that require a high degree of accuracy. Cable runs of less than 50 feet that
don’t require high accuracy typically do not need remote sense.
When remote sense and shunt calibration are included together in the same application, a seventh
wire is usually added to provide a current carrying path that is separate from the -Signal wire. This
information is used by the instrument to overcome error induced by the current flow. Refer to
Appendix C for more information on shuntcal.
Troubleshooting Techniques
The following are functional tests for strain gauge load cells. The tests described below should be
performed before requesting field support. Use the page at the end of this appendix to record the
results so they will be easily accessible when working with field support.
Insulation Resistance
Disconnect the load cell from the interconnect cable if attached by a connector or terminal
block.
Measure the resistance from any or all leads to the body of the load cell.
Warning: DO NOT use a Hypot tester! Test equipment must apply no more than 50
VDC. If internal varisters are used for lightening protection, the test equipment
must apply no more than 20 VDC.
If the resistance is less than 100M ohms, the load cell is faulty.
Cable Insulation Resistance
(Does Not Apply To Integral Cables)
Disconnect the interconnect cable from both the load cell and the instrument.
Verify that the insulation between each wire is good by measuring the resistance between
each wire, making sure it meets spec.
Verify that the insulation between each wire and the shield is to spec.
Bridge Resistance
Disconnect the load cell from the interconnect cable.
Measure the bridge resistance and verify it is to spec.
112
www.nov.com/mdtotco
Document number 90-01
D: Load Cell Troubleshooting
Revision E
Page 111 of 158
Reconnect the interconnect cable to the load cell, and disconnect the cable from the
instrument.
Measure the bridge resistance through the cable. It should be within a few ohms of the
bridge resistance measured at the load cell.
Note: Measuring the bridge resistance through the interconnect cable only verifies that the
cable makes connection. There can still be wiring errors, such as swapped polarities on the
signal or excitation wires. Therefore, in all new installations, a pin-to-pin verification must
be performed.
Excitation Voltage
Disconnect the load cell from the interconnect cable.
Connect the instrument to the interconnect cable, and power the instrument.
Verify that the excitation voltage at the load cell end of the cable is to spec.
Signal Voltage
Connect load cell to instrument.
Verify that the voltage between (-) excitation and (-) signal is ½ of the excitation voltage.
Verify that the voltage between (-) excitation and (+) signal is ½ of the excitation voltage.
Using a digital volt meter that can measure small voltages, verify that the voltage between
(+) signal and (-) signal is less than 2mV when there is no load applied to the load cell.
Apply some load to the load cell and verify that the signal voltage increases with the right
polarity.
Activate the shunt calibration and/or the voltage substitution circuits and verify their
operation.
www.nov.com/mdtotco
Document number 90-01
Revision E D: Load Cell Troubleshooting
Page 112 of 158
112
www.nov.com/mdtotco
Document number 90-01
Revision E
Page 113 of 158
E: Option Modules
Overview
Series 2000 instruments can accept up to four plug-in option modules, which are used for
interfacing with external equipment. Examples include serial communications, analog output, etc.
Located on the back of Series 2000 instruments, under the break-away tabs labeled OPT1, OPT 2,
OPT 3, and OPT 4, are four 28-pin sockets for option modules.
TB1-12/24VDC TB2-PWR OUT TB3-RS485 TB10-RLY1 TB11-RLY2
ON/OFF + - + - CHASSIS A B A B M2 M1 G2 G1 M2 M1 G2 G1
RLY1
RLY1
RLY2
RLY2
RLY3
RLY3
OPT1 OPT2 OPT3 OPT4 RLY4
RLY4
V+ S+ S- V- V+ S+ S- V- V+ S+ S- V- V+ S+ S- V-
TB4-ANLG1 TB5-ANLG2 TB6-ANLG3 TB7-ANLG4
+ -
ADJUST
CONTRAST
RESET + A - CHS A+ A- B+ B- M2 M1 G2 G1 M2 M1 G2 G1
ADDR TB8-POS PWR TB9-POS SIG TB13-RLY4 TB12-RLY3
Electrically, all sockets are identical with the exception of pin 1, which is an “address” or “select”
line. As a result, some module types work the same in all four sockets, and some have different
functions depending on which position they are installed. The model code reflects which module is
installed in which socket at the factory. Option modules may also be field installed.
www.nov.com/mdtotco
Document number 90-01
Revision E E: Option Modules
Page 114 of 158
Note: Only one aux comm module can be installed per instrument at any one time.
Note: Only one event input module can be installed per instrument at any one time.
116
www.nov.com/mdtotco
Document number 90-01
E: Option Modules
Revision E
Page 115 of 158
Installation
Locate the four rectangular tabs on the back of the instrument labeled OPT 1, OPT 2,OPT 3,
and OPT 4.
TB1-12/24VDC TB2-PWR OUT TB3-RS485 TB10-RLY1 TB11-RLY2
ON/OFF + - + - CHASSIS A B A B M2 M1 G2 G1 M2 M1 G2 G1
RLY1
RLY1
RLY2
RLY2
RLY3
RLY3
OPT1 OPT2 OPT3 OPT4 RLY4
RLY4
V+ S+ S- V- V+ S+ S- V- V+ S+ S- V- V+ S+ S- V-
TB4-ANLG1 TB5-ANLG2 TB6-ANLG3 TB7-ANLG4
+ -
ADJUST
CONTRAST
RESET + A - CHS A+ A- B+ B- M2 M1 G2 G1 M2 M1 G2 h G1
ADDR TB8-POS PWR TB9-POS SIG TB13-RLY4 TB12-RLY3
Figure E.2: Series 2000 Back Cover (Option Module Socket Covers Highlighted)
Remove the rear cover by loosening the six, black retaining screws securing the cover to the
unit.
Break out the appropriate tabs to expose the required option module sockets.
Verify that all of the pins on the bottom of the option module are straight. If necessary,
carefully bend them back straight. All pins must be straight for the module to fit in the
instrument.
Using the four long guide pins, carefully position the module over the socket.
www.nov.com/mdtotco
Document number 90-01
Revision E E: Option Modules
Page 116 of 158
Place your fingers on the terminal block and the edges of the housing and press firmly until
the module is fully seated in the socket. DO NOT apply excessive force to the center of the
housing. Refer to Figure E.3.
Press here
+ + - R
2 4-20mA E
4 Isolate T
V
Press here
116
www.nov.com/mdtotco
Document number 90-01
Revision E
Page 117 of 158
F: Auxillary Communications
Overview
The purpose of auxiliary communication output is to provide real-time values of measured
parameters to external equipment, such as computers, data loggers, etc. Some operator
commands can also be transmitted from an external device into a Series 2000 instrument.
Examples of these are ACKnowlege alarms, activate CAL1 or CAL 2 relays, perform or report
SELFTEST, and RESET certain values to zero.
There are two types of aux comm modules that can be installed in one of the option module slots
on the back of Series 2000 instruments:
RS232
RS485
Although an aux comm module can be installed in any of the option module slots, typically it is
installed in slot 4.
Note: Only one aux comm module can be installed in a Series 2000 instrument at one time.
Protocol Selection
There are three different protocols to choose from for aux comm:
PROTOCOL #1—the original, simplified comm protocol which is limited to the three basic
commands (SS, CS, and CD). This protocol is still a good choice when connecting only one
Series 2000 instrument to a computer. This protocol also allows for support of older, existing
field installations.
PROTOCOL #2—features of protocol #1 plus an expanded instruction set and more efficient
protocol for party line applications.
Continuous Mode—forces the instrument to automatically power up in the continuous scan
mode. The instrument acts like it has received a CS command. This is typically used when
the instrument is connected to a serial printer. The transmission interval is programmable via
the front panel, from 0.25 sec to 60 sec (in 0.25 sec increments).
Command List
SS (single scan)—request for a single transmission of all present values.
CS (continuous scan)—same as SS, only the transmissions will continue at a programmed interval
until an "ESCape" character is received or the instrument is reset.
CD (configuration dump)—request for a list of all configuration information. The answer returned
includes calibration constants, a channel list for the product (this list matches the channels
transmitted for SS and CS), alarm limits, and device numbers.
www.nov.com/mdtotco
Document number 90-01
Revision E F: Auxillary Communications
Page 118 of 158
ACK (acknowledge)—acknowledges alarms the same as pressing the front panel ACK button.
This silences all horns (internal and external), but does NOT clear the general purpose relays.
There is no response transmitted back from the instrument.
C1 (CAL 1)—turns on the "CAL 1" relay on the back of the instrument the same as activating the
relay from the front panel. The relay is cleared by transmitting an "ESCape" character to the
instrument. There is no response transmitted back from the instrument for either the C1 or the
"ESCape" signal.
C2 (CAL 2)—turns on the "CAL 2" relay on the back of the instrument the same as activating the
relay from the front panel. The relay is cleared by transmitting an "ESCape" character to the
instrument. There is no response transmitted back from the instrument for either the C2 or the
"ESCape" signal.
RS (reset)—resets the present value to zero. In the case of an LM2000, it resets the payout to
zero. In the case of the CI2000, the main net load and whip net load are set to zero (new "TARE"
value). There is no response transmitted back from the instrument.
ST (perform self test)—causes the instrument to perform the power up self test sequence. The
instrument is forced to reset and automatically runs the power-on self tests. There is no response
transmitted back from the instrument.
RT (report self test)—requests a transmission of the self test results that were stored at the last
self test. This does not trigger the instrument to perform a self test.
Delimiter Selection
In the SS and CS commands, the values returned are separated by a programmable character.
The choices are:
Comma
Space
Tab
130
www.nov.com/mdtotco
Document number 90-01
F: Auxillary Communications
Revision E
Page 119 of 158
Protocol # 1
The basics of protocol #1 involve:
signing on to particular unit
receiving a response of self-test results from the unit
sending one of three commands (SS, CS, or CD)
receiving the response back from the instrument until you log onto another unit.
Once the communications wiring is properly connected, aux comm is initiated as follows:
Sign on to a particular Series 2000 unit by sending the following command:
where
A = one digit result of sensor unit CPU ram test
0 if passed
1 if failed
B = one digit result of sensor unit CPU PROM test
0 if passed
1 if failed
C = one digit result of sensor unit CPU EEPROM test
0 if passed
1 if failed
DDD = three digit error count for -15V power supply
000 if no errors
001–999 if any errors
www.nov.com/mdtotco
Document number 90-01
Revision E F: Auxillary Communications
Page 120 of 158
SS Command
When you send the SS command to a Series 2000 instrument, you are asking it to respond with a
Single Scan of current values and alarm status information.
To send the SS command from a terminal or computer, simply type "SS" at the keyboard. The
instrument will respond with information in the following format:
130
www.nov.com/mdtotco
Document number 90-01
F: Auxillary Communications
Revision E
Page 121 of 158
where
nn = sensor unit number (drop address) of the sensor unit sending this data
FF...F = current value in floating point format, including sign and decimal, right justified with
decimal position depending on channel precision. Current value and status information
provided for as many channels as listed in the response to the CD command, and in the
same order.
S = one character alarm status response
H for high-high alarm
h for high alarm
l for low alarm
L for low-low alarm
Space (blank) if no alarm condition
cr = carriage return
lf = line feed
CS Command
When you send the CS command to a Series 2000 instrument, you are asking it to respond with a
Continuous Scan of current values and alarm status information, at an interval you determine
when configuring the instrument.
To send the CS command from a terminal or computer, type "CS" at the keyboard. The instrument
will respond with information in the same format as a SS, only it will continue updating the
information until you press the Esc key on the keyboard.
www.nov.com/mdtotco
Document number 90-01
Revision E F: Auxillary Communications
Page 122 of 158
CD Command
When you send the CD command to a Series 2000 instrument, you are asking it to respond with a
Configuration Dump of the currently selected instrument. The instrument will respond with data in
the following format:
130
www.nov.com/mdtotco
Document number 90-01
F: Auxillary Communications
Revision E
Page 123 of 158
Protocol # 2
Protocol #2 is the preferred protocol for “party line” applications. In protocol #2, the sign on and the
request for data are part of the same command, thereby not wasting transmission bandwidth on
sending expanded self-test results over and over. Also note that a four character system status/
selftest status is added to the SS and CS response. If these results show a self-test failure, then
the more detailed self-test results can be requested. In addition, protocol #2 has a device address
wild card. For commands that require NO response, the device address can be replaced with “$$”,
which will cause all connected units to take the action the command specifies. Typical uses are
activating cal relays (C1) or acknowledging alarms (AK).
Once the communications wiring is properly connected, aux comm protocol #2 is initiated as
follows:
Sign on to a particular Series 2000 unit by sending the following command:
www.nov.com/mdtotco
Document number 90-01
Revision E F: Auxillary Communications
Page 124 of 158
CD Command
When you send the CD command to a Series 2000 instrument, you are asking it to respond with a
Configuration Dump of the currently selected instrument. The instrument will respond with data in
the following format:
130
www.nov.com/mdtotco
Document number 90-01
F: Auxillary Communications
Revision E
Page 125 of 158
When you send the SS command to a Series 2000 instrument, you are asking it to respond with a
Single Scan of current values and alarm status information.
To send the SS command, simply type:
at the keyboard. The instrument will respond with information in the following format:
where:
nn = sensor unit number (drop address) of the sensor sending this data
FF...F = current value in floating point format, including sign and decimal, right justified with
decimal position depending on channel precision. Current value and status information
provided for as many channels as listed in the response to the CD command, and in the
same order.
S = one character alarm status response
H for high-high alarm
h for high alarm
l for low alarm
L for low-low alarm
Space (blank) if no alarm condition
XXXX = four character system status
CAL1 if the CAL1 relay is active, affecting current values
CAL2 if the CAL2 relay is active, affecting current values
0000 if no cal relay is active and all self tests passed
0001–9999 if no cal relay is active and there are any self test failures
www.nov.com/mdtotco
Document number 90-01
Revision E F: Auxillary Communications
Page 126 of 158
cr = carriage return
lf = line feed
CS Command
When you send the CS command to a Series 2000 instrument, you are asking it to respond with a
Continuous Scan of current values and alarm status information, at an interval you determine
when configuring the instrument.
To send the CS command, type:
at the keyboard. The instrument will respond with information in the same format as an SS, only it
will continue updating the information until you press the Esc key on the keyboard.
AK Command
C1 Command
C2 Command
RS Command
ST Command
Sending the ST command causes the system to reset, which then causes a self test to occur.
Series 2000 instruments send back no reply to this command.
130
www.nov.com/mdtotco
Document number 90-01
F: Auxillary Communications
Revision E
Page 127 of 158
RT Command
As a response to the "RT" command, the Series 2000 instrument will answer with self test results
as follows:
where:
A = one digit result of sensor unit CPU ram test
0 if passed
1 if failed
B = one digit result of sensor unit CPU PROM test
0 if passed
1 if failed
C = one digit result of sensor unit CPU EEPROM test
0 if passed
1 if failed
DDD = three digit error count for -15V power supply
000 if no errors
001–999 if any errors
EEE = three digit error count for +15V power supply
000 if no errors
001–999 if any errors
FFF = three digit error count for +24V power supply
000 if no errors
001–999 if any errors
GGG = three digit error count for 3.84V reference
000 if no errors
001–999 if any errors
HHH = three digit error count for A/D calibration errors
000 if no errors
001–999 if any errors
JJJ = three digit error count for A/D data ready errors
000 if no errors
001–999 if any errors
www.nov.com/mdtotco
Document number 90-01
Revision E F: Auxillary Communications
Page 128 of 158
130
www.nov.com/mdtotco
Document number 90-01
F: Auxillary Communications
Revision E
Page 129 of 158
C1 (Cal1) Command
Process time for this command ranges from 3ms to about 400ms. This command does not require
any response to the host. Also, the process time for this command is applicable to all other
commands that do not require any response (C2, AK, RS). Because this type of command is
transmitted infrequently, it is recommended to wait >500ms after transmitting before transmitting
any additional command.
www.nov.com/mdtotco
Document number 90-01
Revision E F: Auxillary Communications
Page 130 of 158
130
www.nov.com/mdtotco
Document number 90-01
Revision E
Page 131 of 158
G: Networking
Overview
Networking is used to operate a group of up to 16 Series 2000 units together for the purpose of
having remote displays, multiple displays, etc. This communication port is standard on all Series
2000 instruments through the TB3 connections on the back of each unit.
There are two different types of stations in a network:
Sensor input unit
Remote display unit
Remote Display
Remote displays cannot read signals directly from sensors. Remote displays must get their
information from the network. They can, however, drive relays and option modules. Every remote
display can be configured (via the front panel) to act as a display for any sensor input unit in the
network. Remote displays have access to all calibration and configuration features just like a local
display on a sensor input unit. There can be a maximum of 15 remote displays in a network.
www.nov.com/mdtotco
Document number 90-01
Revision E G: Networking
Page 132 of 158
Rotary
Address V+ S+ S- V- V+ S+ S- V- V+ S+ S- V- V+ S+ S- V-
Switch TB4-ANLG1 TB5-ANLG2 TB6-ANLG3 TB7-ANLG4
ADJUST
CONTRAST
RESET
ADDR + A - CHS A+ A- B+ B- M2 M1 G2 G1 M2 M
TB8-POS PWR TB9-POS SIG TB13-RLY4 TB
Power up the unit, and verify that the unit is configured as a “sensor input unit” in the
Network Setup screen. See Chapter 5, titled "Configuration and Calibration" for more
information.
For units with displays, look in the Network Setup screen to see what switch setting the unit
has. Always use this number and ignore what is printed on the switch itself.
For remote data acquisition units (sensor input units without displays), you must read the
numbers printed on the rotary switch body. The switch is printed with hexadecimal numbers,
so refer to the following table to determine what hexadecimal number corresponds to what
switch setting:
The setting of the rotary switch becomes the unit number when referenced by any remote
displays. For example, on LM2000 systems, this rotary switch sets the WINCH#. On CI2000
systems, this rotary switch sets the CRANE#.
142
www.nov.com/mdtotco
Document number 90-01
G: Networking
Revision E
Page 133 of 158
Select the number of the sensor unit you chose in step 2 above.
Note: If the data is ever corrupted, the remote display will have to load all new default
values. The rotary switch setting of the remote display will be used as the default unit
number for that unit. Therefore, it is suggested that the switch setting of the remote display
match that of the sensor unit to be displayed remotely.
Network Examples
The following examples are diagrams of various network examples. The switch settings and
network setup parameters are shown in each example.
Note: Please note that the switch setting listed in the diagram refers to the switch setting as
displayed in the network setup screen, not what is printed on the switch body itself. For
more information on switch settings, refer to "Sensor Input Unit Setup" on page 132.
www.nov.com/mdtotco
Document number 90-01
Revision E G: Networking
Page 134 of 158
142
www.nov.com/mdtotco
Document number 90-01
G: Networking
Revision E
Page 135 of 158
www.nov.com/mdtotco
Document number 90-01
Revision E G: Networking
Page 136 of 158
142
www.nov.com/mdtotco
Document number 90-01
G: Networking
Revision E
Page 137 of 158
www.nov.com/mdtotco
Document number 90-01
Revision E G: Networking
Page 138 of 158
142
www.nov.com/mdtotco
Document number 90-01
G: Networking
Revision E
Page 139 of 158
www.nov.com/mdtotco
Document number 90-01
Revision E G: Networking
Page 140 of 158
Note: Please note that the switch setting listed in the diagram refers to the switch setting as
displayed in the network setup screen, not what is printed on the switch body itself.
142
www.nov.com/mdtotco
Document number 90-01
G: Networking
Revision E
Page 141 of 158
www.nov.com/mdtotco
Document number 90-01
Revision E G: Networking
Page 142 of 158
Figure G.11: Multiple Sensor/Display with Aux Comm to Multiple Remote Displays
142
www.nov.com/mdtotco
Document number 90-01
Revision E
Page 143 of 158
Overview
The following drawings show wiring examples for various types of enclosures and equipment
setups. Use these drawings as a reference when wiring Series 2000 systems.
Sheet 1 (221117-001)—SC103 with shuntcal connected to a load cell and an SP102 payout
sensor.
Sheet 2 (221117-002)—2078 signal conditioner (and 2077 motherboard) with voltage
substitution connected to an SP102 payout sensor.
Sheet 3 (221117-003)—2078 signal conditioner (and 2077 motherboard) with shuntcal
connected to an SP102 payout sensor.
Sheet 4 (221117-004)—SC103 and prox sensors in a hazardous environment, and the
display in a safe area.
Sheet 5 (221117-005)—SC103 without shuntcal connected to a load cell and an SP102
sensor.
www.nov.com/mdtotco
Document number 90-01
Revision E H: Interconnect Examples
Page 144 of 158
144
www.nov.com/mdtotco
Document number 90-01
Revision E
Page 145 of 158
Overview
The appendix contains miscellaneous drawings and parts lists for LM2000 products. The drawings
are listed in the order that they appear at the end of this appendix.
Series 2000 Catalog Drawing (drawing # 220921)
Main Assembly Drawing (drawing # 220870)
Dashmount Enclosure Kit (drawing # 220898)
Remote Data Acquisition Enclosure Kit (drawing # 220930)
Installation Kit (drawing # 220925)
The following sections describe how to choose the correct part numbers for any configuration of
LM2000.
www.nov.com/mdtotco
Document number 90-01
Revision E I: Illustrated Parts Lists
Page 146 of 158
RLY1
left side of the cover.
RLY2
RLY2
RLY3
RLY3
OPT1 OPT2 OPT3 OPT4 RLY4
RLY4
V+ S+ S- V- V+ S+ S- V- V+ S+ S- V- V+ S+ S- V-
TB4-ANLG1 TB5-ANLG2 TB6-ANLG3 TB7-ANLG4
+ -
ADJUST
CONTRAST
RESET + A - CHS A+ A- B+ B- M2 M1 G2 G1 M2 M1 G2 G1
ADDR TB8-POS PWR TB9-POS SIG TB13-RLY4 TB12-RLY3
158
www.nov.com/mdtotco
Document number 90-01
I: Illustrated Parts Lists
Revision E
Page 147 of 158
Model Code
Each LM2000 has a model code associated with it. Once this model code has been located, you
can use it to determine the part numbers of each field replaceable part.
The model code will look like the following example:
LM2-1A-1234-AABC where
what’s in option socket 1
what’s in option socket 2
enclosure
what’s in option socket 3
type
what’s in option socket 4
LM2-
Once you have located the model code, refer to the catalog drawing as follows to determine the
necessary part numbers. The catalog drawing is the first of the large Z-folded pages at the end of
this appendix. The catalog drawing images on the following pages are too small to read and are for
mapping purposes only. Refer to the full size drawing at the end of this appendix when necessary.
www.nov.com/mdtotco
Document number 90-01
Revision E I: Illustrated Parts Lists
Page 148 of 158
158
www.nov.com/mdtotco
Document number 90-01
I: Illustrated Parts Lists
Revision E
Page 149 of 158
Find numbers listed below and refer to the circled numbers in drawing 220870-1XXX.
Core Assembly Components on Display Units (220870-1XXX)
www.nov.com/mdtotco
Document number 90-01
Revision E I: Illustrated Parts Lists
Page 150 of 158
Find numbers listed below and refer to the circled numbers in drawing 220870-2XXX.
Core Assembly Components on Remote Data Acquisition Units (220870-
2XXX)
158
www.nov.com/mdtotco
Document number 90-01
I: Illustrated Parts Lists
Revision E
Page 151 of 158
Find numbers listed below and refer to the circled numbers in drawing 220870-3XXX.
Core Assembly Components on Stainless Steel Display Units (220870-
3XXX)
www.nov.com/mdtotco
Document number 90-01
Revision E I: Illustrated Parts Lists
Page 152 of 158
158
www.nov.com/mdtotco
Document number 90-01
I: Illustrated Parts Lists
Revision E
Page 153 of 158
www.nov.com/mdtotco
Document number 90-01
Revision E I: Illustrated Parts Lists
Page 154 of 158
158
www.nov.com/mdtotco
Document number 90-01
I: Illustrated Parts Lists
Revision E
Page 155 of 158
www.nov.com/mdtotco
Document number 90-01
Revision E I: Illustrated Parts Lists
Page 156 of 158
158
www.nov.com/mdtotco
Document number 90-01
I: Illustrated Parts Lists
Revision E
Page 157 of 158
www.nov.com/mdtotco
Document number 90-01
Revision E I: Illustrated Parts Lists
Page 158 of 158
158
www.nov.com/mdtotco