Model 28-5 and 28.6 are manually operated, Bears
Calibrators, for calibrating all types of radiation
detection instruments. Model 28-5 incorporates
tether 2 120 or 2 200 mCi '°Cs source, Model
28.6 incorporates either 2 600 mCi or a L2 Ci
BCs source. A slip-on attenuator (X-10) is pro-
vided with all units. For additional details on Model
285 with loadings up to 130 Ci. Cs, see the
Beam Calbrator/lradiator Catalog
LOW EXTERNAL RADIATION LEVELS: <5
mR/he (typically <2 mR/he) at one foot from the
surface, with the source in the “Off” positon,
behind the Calrator with the source in the
position
OPERATION: The source i fixed to the end of
the shielded operating rod: moved from the com
pletely shielded "Of" position to the exposed "On!
position by means of an operating handle, located
‘n top of the unit. Source position indicating lights
and a padlock to lock the source inthe “Of?” posi
tion, are provided for all units.
STAND: All units include a stand to provide a 36”
beam height,
CC-6 COLLIMATOR OPTION: For use wth
Model 28-68 Calibrator. Provides constancy checks
for 0.6 cc ionization therapy chambers placed in
‘contact with source-tube. Dose rate s approximately
10 R/rminSS ee
= ee | ep
as To |_Seee =
~- SEES =
- = a =
“Al sources cated “Special For
‘Other beam angles: 15.20, and 30 are available upon request Beam pot. as sted, wil be supped unless otherwise
insted
Model 28 with IT1 Calibration Table
CERTIFICATION: Units are calibrated using Na
tional Bureau of Standards traceable Roentgen
Meters, accuracy 5%. Calibration, Extemal
Radiation Level and Leak Test Certificates are pro
vided with each unit
LICENSING: These units appear in the "Approv-
‘ed Sources and Devices” Catalog of the USNRC
MANUALS: Provided with each unt
‘SHIPMENT: All nits moet DOT
7A specifications and are
shipped with source installed,
ready for operation.
WARRANTY: Fee ports and
service for three months
following delivery, with
replacement of faulty com
ponents for an additional
fine months.Ze
Model 28 with Slip-On Attenuators
‘MODEL 311 SLIP-ON ATTENUATORS: Set of
three. Values: X-2, Kd and X10, to provide
cumulative attenuation factors of 2, 4, 10 and 100,
(using the X10 provided with the unit)
CC SERIES, REMOVABLE COLLIMATORS:
To reduce beam angle as required.
RS SERIES, ROLLING STANDS: 60 cr square,
with casters and position stops.
ET SERIES, ELAPSED TIMERS:
Modet X, range 9999.9 seconds or
Model ¥! range 9999.9 minutes
Features pushbutton reset and s engaged whenever
the source isin the “radiate” position, Mounted
fon the operating tower of the Calibrator.
MODEL IT-1 CALIBRATION TABLE: Avaliable
in 1 wide x 8” long modules. One end bolts to
the base ofthe Calibrator; the other end is mounted
‘on a hinged stand with leveling jacks fr storage and
rapid table assembly disassembly, Modules may be
bolted together to form a table 16, 24° or 32”
long. They are constructed of light weight (0.91
T./I). ultra strong (572 pst), lowZ* (27) material
to minimize scattered radiation.
SERIES 10 PORTABLE BEAM CALIBRATORS.
WITH REMOVABLE SOURCE
Model 108 are light weight, portable, Calibrators
designed for the calibration of low or intermediate
range survey meters, yemote area monitor probes
for other applications, when its destrable to use a
‘gamma source, with handler, for fee air exposures
These units consis of a "Cs source, mounted to
the end of a shielded plug, which fis into an DOT
7A shipping/storage container. An 18” handler
which threads into the shielded end of the plug, is
provided for manipulation of the source so thatthe
Source may be removed from the shield for free ar
exposure. A second, removable plug is provided in
line withthe source, providing a 20° beam port for
the calbration of survey instruments, Both plugs are
equipped with padlocks for storage. A 5 meter
retractable tape (not shown) is mounted at the
source centerine for calibrations using the beam
port
“EXTERNAL RADIATION LEVEL: <5 mR/he
at one foot from the surface
CERTIFICATION: All sources are calibrated with
‘National Bureau of Standards traceable Roentgen
Meter, accuracy +5%. Dose Rate [source
removed-free air and 20° beam port), External
Radiation Level and Leak Test Cerilicates are pro-
vided with each unt
LICENSING: Series 10 Calibrators appear in the
“Approved Sources and Devices” Catalog of the
USNRC.
dale, CA. Allusits are DOT 7A shipping containers
‘and are shipped complete, ready for operation.
WARRANTY: Free parts and service for three
‘months folowing delivery. with replacement of faul=
ty components for an additional nine months,
ATTENUATOR: A X-10 attenuator which fis over
the 20° beam port is available as an accessory.
soccceniess, /