Ak 96 Preset List For External Use Program Version 3.50: Gambro Lundia AB

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 28

Gambro Lundia AB

AK 96 PRESET LIST

FOR EXTERNAL USE

Program version 3.50

MDREC-132304 AK 96 program version 3.50 1 (28)


Gambro Lundia AB

Table of contents
0 START UP..................................................................................................................................................3
1 CONFIGURATION .....................................................................................................................................3
2 SET VALUES AND LIMITS .......................................................................................................................5
3 OVERRIDE TIMES...................................................................................................................................10
4 ASSISTED PRIMING ...............................................................................................................................11
5 RINSE BACK ...........................................................................................................................................13
6 DISINFECTION ........................................................................................................................................13
7 SINGLE NEEDLE.....................................................................................................................................18
8 ACETATE CONCENTRATES..................................................................................................................18
9 BICARBONATE CONCENTRATES........................................................................................................21
10 BPM SETTINGS.....................................................................................................................................23
11 PROFILING ............................................................................................................................................24
12 DIASCAN SETTINGS ............................................................................................................................26
13 USER INTERFACE ................................................................................................................................27
14 RECIRCULATION..................................................................................................................................27

This document lists the presets parameters that are available in the AK 96 dialysis machine together with
the defalt values and the possible settings for the preset parameters.

MDREC-132304 AK 96 program version 3.50 2 (28)


Gambro Lundia AB

Group Name Parameter Default Minimum Maximum Comment


0 Start up
0-0 Acetate/Bicarbonat Acetate/Bicarb. Bicarb. Acetate Bicarb. Acetate or Bicarbonate at start up
e mode
0-1 Double/single DN/SN DN DN SN Double or Single needle at start up
needle mode
0-3 Priming type Priming type NORMAL NORMAL Assisted Normal = Machine starts up in normal priming mode
Assisted = Machine starts up in Assisted priming mode
1 Configuration
1-0 Language Language English See See English
comment comment French
Czech
Portuguese
Spanish
Norwegian
Finnish
Swedish
Danish
Slovak
Greek
Polish
Turkish
Russian,
Hungarian
Simplified Chinese
Traditional Chinese
Bulgarian
1-1 Arterial pressure Arterial preset Not Inst Not Inst Inst Fixed high arterial pressure alarm limit.
preset The high arterial pressure limit will be set to a fixed value
(Arterial pressure upper limit, 2-18) during a centralization
request (pressing the flashing arterial pressure button).
1-2 Heater power Heater power (W) 3*580 See See 1000 (100V mains voltage)
comment comment 1300 (115V mains voltage)
1750 (230V mains voltage)
3*580 (230V mains voltage)
1-3 pH cell pH cell Not Inst Not Inst Inst Machine equipped with pH cell
1-4 BiCart holder BiCart holder Inst Not Inst Inst Machine equipped with BiCart holder

MDREC-132304 AK 96 program version 3.50 3 (28)


Gambro Lundia AB

Group Name Parameter Default Minimum Maximum Comment


1-5 BiCart emptying BiCart emptying Fluid Fluid Powder Fluid or powder emptying of BiCart.
If this preset is set to powder, the BiCart will be completely
emptied.
1-6 BiCart priming, Flush (sec) DISABLED 10 600 Extra flush time during BiCart priming.
extra flush
1-7 Degassing Degass pressure -610 -300 -700 Degassing pressure (negative sign omitted)
pressure The value must be changed depending on altitude.

Altitude (m): Recommended pressure:


0 -610 mmHg
200 -590 mmHg
400 -570 mmHg
600 -550 mmHg
800 -530 mmHg
1000 -510 mmHg
1400 -480 mmHg
1800 -450 mmHg
2200 -415 mmHg
2400 -400 mmHg
3000 -350 mmHg
3800 -300 mmHg
1-8 Blood pump
segment diameters
1-8-0 Blood pump tube Tube diam (mm) 7.90 2.00 8.50 Blood tube set no 1
ALT1
1-8-1 Blood pump tube Tube diam (mm) 6.35 2.00 8.50 Blood tube set no 2
ALT2
1-8-2 Blood pump tube Tube diam (mm) 4.00 2.00 8.50 Blood tube set no 3
ALT3
1-9 Heparin pump
1-9-0 Heparin pump Heparin pump Inst Not Inst Inst Machine equipped with heparin pump
install
1-9-1 Heparin, inner Diameter (mm) 20.0 10.0 30.0 Heparin, inner diameter.
diameter This value is used by the machine to calculate the infusion
volume.
1-9-2 Heparin syringe Label Not defined N/A N/A Heparin syringe label
type

MDREC-132304 AK 96 program version 3.50 4 (28)


Gambro Lundia AB

Group Name Parameter Default Minimum Maximum Comment


1-12 Clin Automation Clin aut com sup Not Inst Not Inst Inst
Com Superv
1-13 Isol UF Handling Isol UF Handling Acc Acc Diff Time handling in isolated UF (isol. UF time added/not added to
total treatment time)
ACC = The UF volume during Isol UF added
Diff = The UF volume during Isol UF not added
1-14 Serial Serial Comm. RS232 RS232 RS422 Serial communication type
Communication
type
1-15 Flush tube attention Flush tube attn Not Inst Not Inst Inst Attention to remind operator to flush fluid tubes before
treatment
1-16 HPG Max Limit HPG Max Limit 750 450 800 Maximum HPG pressure limit in mmHg
1-17 UFD UFD Not Inst Not Inst Inst Machine equipped with UFD-kit.
1-18 Negative TMP Negative TMP Attention False False True Attention if TMP is negative
Attention True = A back filtration warning will be given the first time the
TMP low limit is set to a negative value.
The operator has to confirm the warning by pressing the TMP
button.
False = This must be confirmed when exiting the preset mode.
No back filtration warning will be given when setting a negative
TMP low limit during treatment.
1-19 Blood pump stop in Blood pump stop in priming Inst Not Inst Inst Blood pump stops when machine detects blood.
priming When this preset is set to Inst, the blood pump will
automatically stop when blood is detected by the priming
detector.
1-20 Blood pressure BPM Not Inst Not Inst Inst BPM installed.
monitor
1-21 Profiling Profile Inst Not Inst Inst Profiling installed.
1-22 Inlet conductivity In cnd check False False True False = No inlet cond
check True = Max 2.5 mS/cm
True = It is possible to, under the control and confirmation of
the operator, run the machine with higher inlet conductivity. If
the inlet conductivity is above 2.5 mS/cm the error code C
CFM 028 003 will occur.
1-23 Variable fluid flow Var. flow inst Inst Not Inst Inst Variable flow kit installed.
1-24 No water No water consump False False True False = No water saving during concentrate standby mode.
consumption mode True = INVA close, pumps stop, signal to turn of WRO when

MDREC-132304 AK 96 program version 3.50 5 (28)


Gambro Lundia AB

Group Name Parameter Default Minimum Maximum Comment


concentrate standby mode is activated.
1-25 Concentrate Standby Fch (sec) 0 0 3600 0 = No automatic concentrate standby mode after Fch.
standby Fch 1-3600 = Seconds after Fch until concentrate standby mode is
activated.
1-26 Concentrate Standby prim (sec) 0 0 3600 0 = No automatic concentrate standby mode after priming.
standby priming 1-3600 = Seconds after priming until concentrate standby
mode is activated.
1-27 DiaScan installed DiaScan Not Inst NotInst Inst Diascan installed.
1-28 WRO 300H WRO 300H inst Not Inst Not Inst Inst Inst = Possible to perform an integrated heat with a WRO 300
installed H
Not Inst = Not possible to perform an integrated heat with a
WRO 300 H
1-30 Arterial clamp Arterial clamp Inst Not Inst Inst Arterial clamp installed.
installed
1-31 Verification of Verify conc Inst Not Inst Inst Not Inst = No confirmation of concentrate needed
concentrate Inst = Concentrate must be confirmed before treatment
1-32 Loudspeaker
1-32-0 Loudspeaker Speaker volume 6 2 8 Sound level of the loudspeaker
volume
1-33 Ven clamp open on Ven clmp Open False False True False = The venous clamp closes when the blood pump is
blood detection stopped
True = The venous clamp remains open when the blood pump
is stopped for the first time after blood detection.
2 Set values and limits
2-0 Blood flow
2-0-0 Blood flow set Blood flow 100 0 500 Blood flow value at start-up.
(ml/min)
2-0-1 Blood flow low limit Blood fl Low Lim 100 0 500 Blood flow low limit at start-up.
(ml/min)
2-0-2 Blood flow low LL alarm delay (m) 5 0 60 Delay time before attention regarding low blood flow.
alarm delay (min)
2-0-3 Blood flow bypass Blood fl bypass 50 0 500 Blood flow during override
(ml/min) The speed of the blood pump during override of a blood leak
alarm.
2-0-4 Blood flow stop Stoptime max (s) 60 20 180 Delay time before attention regarding blood pump stopped
time max

MDREC-132304 AK 96 program version 3.50 6 (28)


Gambro Lundia AB

Group Name Parameter Default Minimum Maximum Comment


2-0-5 Blood flow high Blood fl High lim 500 0 500 Maximum allowed Blood flow setting.
limit (ml/min)
2-1 Blood pump coeff. Bl. pump P coeff 2 2 5 P coefficient in the blood pump regulation during high venous
high vpr test pressure test (ramping speed of the blood pump)
2-2 Priming volume
limit
2-2-0 PrimingVol (Adult) Priming vol (l) 1.5 0.5 100.0 Priming volume for adult blood set (segment > 4 mm)
(l)
2-2-1 PrimingVol Priming vol (l) 0.5 0.31 5.00 Priming volume for pediatric blood set (segment <= 4 mm)
(Pediatrics) (l)
2-2-2 ExtraPrimVol Extra pr vol (l) 0.2 0.0 10.0 Extra priming volume for adult blood set (segment > 4 mm)
(Adult) (l) 0 = No limit for the Extra Priming and no volume achieved
attention will be generated.
2-2-3 ExtraPrimVol Extra pr vol (l) 0.05 0.0 1.0 Extra priming volume for adult blood set (segment <= 4 mm)
(Pediatrics) (l) 0 = No limit for the Extra Priming and no volume achieved
attention will be generated.
2-3 Connect Patient
volume limit
2-3-0 Connect Patient ConPat vol (l) 0.35 0.0 0.75 Largest acceptable blood volume between activation of
volume limit (l) CONNECT PATIENT and blood detection. If the limit is
exceeded an alarm is generated.
2-4 Heparin flow Flow (ml/min) 0.0 0.0 10.0 Heparin flow value at start-up.
2-5 Heparin stop limit Heparin stoptime 00:20 00:00 09:59 Automatic stop time for Heparin pump (in minutes before end
of treatment).
2-6 Temperature Temp Set temperature value and operator LL and HL (low/high level)
LL 35.0 33.0 HL
SET 37.0 LL HL
HL 39.0 LL 40.0
2-7 pH pH pH low and high limits.
LL 6.00 5.00 HL
HL 8.00 LL 9.00
2-8 TMP limits TMP TMP start-up limits.
LL -100 -200 HL
HL 400 LL 500
2-9 TMP window Width (mmHg) 200 10 300 Width = TMP window size during treatment (diffusion)
TMP low part 25 20 80 TMP low part = Part (%) of the TMP window below the set
point (diffusion)

MDREC-132304 AK 96 program version 3.50 7 (28)


Gambro Lundia AB

Group Name Parameter Default Minimum Maximum Comment


2-10 TMP window isol. Width (mmHg) 200 10 300 Width = TMP window size during treatment (isolated UF)
UF TMP isol low part 75 20 80 TMP isol low part = Part (%) of the TMP window below the set
point (isolated UF)
2-11 Venous pressure Pressure (mmHg) Venous pressure start-up limits.
limits LL -100 -100 HL Note that the low and high start up limits must be set more
HL 300 LL 500 than 15 mmHg apart.
2-12 Venous pressure Width (mmHg) 100 10 200 Width = Venous window size during treatment.
window Ven. pr. low part 30 20 80 Ven. pr. low part = Part (%) of the venous window below the
set point.
2-13 Ven.pr.auto limits Auto S/N (mmHg) Venous pressure start-up limits during single needle (S/N).
S/N LL 150 -100 HL
HL 350 LL 500
2-14 Ven.pr.auto limit delay time (s) 600 120 1200 Attention delay time for venous pressure low limit < 10mmHg
alarm delay
2-15 Arterial pressure Limits (mmHg) Arterial pressure start-up limits.
limits LL -250 -700 HL Note that the low and high start up limits must be set more
HL 250 LL 750 than 15 mmHg apart.
2-16 Arterial pressure Width (mmHg) 100 10 2000 How much lower than the current arterial pressure the the low
low margin arterial pressure alarm limit will be set to during a centralization
request (pressing the flashing arterial pressure button). Only
effective if Arterial pressure preset (1-1) is set to Inst.
2-17 Arterial pressure Width (mmHg) 200 10 300 Width = Arterial window size during treatment.
window Art.pr. low part 50 20 80 Art.pr. low part = Part (%) of the arterial window that is below
the set point.
2-18 Arterial pressure Limits (mmHg) 150 -200 250 The high arterial pressure alarm limit will be set to this value
upper limit during a centralization request (pressing the flashing arterial
pressure button). Only effective if Arterial pressure preset (1-1)
is set to Inst.
2-19 Na set acetate Acetate Na set Sodium value (acetate mode) and operator LL and HL
(mmol/l) LL 130.0 130.0 HL (low/high level).
SET 140.0 LL HL
HL 160.0 LL 160.0
2-20 Ac conc acetate Ac conc. select ALT1 ALT1 ALT3 Active Acetate bank during start-up
2-21 Na set bicarbonate Bicarb Na set Sodium value (bicarbonate mode) and operator LL and HL
(mmol/l) LL 130.0 130.0 HL (low/high level).
SET 140.0 LL HL
HL 160.0 LL 160.0

MDREC-132304 AK 96 program version 3.50 8 (28)


Gambro Lundia AB

Group Name Parameter Default Minimum Maximum Comment


2-22 HCO3 set Bicarb HCO3 set Bicarbonate value and operator LL and HL (low/high level).
bicarbonate LL 20.0 20.0 HL
(mmol/l) SET 34.0 LL HL
HL 40.0 LL 40.0
2-23 Bc conc. Bc conc. select ALT1 ALT1 ALT3 Active Bicarbonate bank during start-up.
bicarbonate
2-24 Concentrate pump
windows
2-24-0 Acetate pump B Ac pumpB window % 10 0 30 Maximum allowed pump speed deviation (acetate).
window (%)
2-24-1 Bicarb. pump A Bc pumpA window % 10 0 30 Maximum allowed pump speed deviation (A-concentrate).
window (%)
2-24-2 Bicarb. pump B Bc pumpB window % 20 0 40 Maximum allowed pump speed deviation (Bicarb-concentrate).
window (%)
2-24-3 Composition Comp. dev. % 15 0 30 maximum relative pump speed deviation.
deviation (%)
2-25 UF volume and UF vol Diff. (l) 0.00 0.00 10.00 UF volume, start-up value
time/Diffusion Time diff. 04:00 00:00 09:59 Treatment time, start-up value
2-26 UF volume and UF vol Isol. UF 0.00 0.00 10.00 UF volume (isolated UF), start-up value
time/Seq. -isol Time Isol. UF 00:00 00:00 09:59 Treatment time (isolated UF), start-up value
2-27 UF stop timeout Timeout 00:05 00:01 01:00 Attention delay for stopped ultrafiltration
2-28 UF rate
2-28-0 Minimun UF rate UF rate (l/hour)
LL 0.00 N/A N/A LL is a fixed value, not possible to configure.
SET 0.00 0.00 HL Start-up value for minimum UF rate.
HL 0.00 LL 4.00 Minimum allowed UF rate.
2-28-1 Maximum UF rate UF rate (l/hour) 4.00 0.00 5.00 Maximum allowed UF rate.
2-28-2 Priming UF rate UF rate (l/hour) 0.20 0.00 4.00 Start-up value for priming UF rate.
2-28-3 UF gain coefficient UF Gain Max gain 1.20 1.00 2.00 Maximum allowed deviation between actual and calculated UF
rate
(UF rate limit = “UF Gain Max gain” x calculated UF rate).
2-29 DialysisFlow Limits (ml/min) 425 200 800 Limit for low flow attention. If the machine is equipped with
adjustable flow this value will be set to the set value -75 ml/min
automatically.
2-30 Max conductivity Limits (mS/cm) 0.3 0 1.0 Maximum allowed conductivity difference between cond A and
difference cond B during Function check.
2-31 Max temperature Temp limits Maximum allowed temperature limits.
MDREC-132304 AK 96 program version 3.50 9 (28)
Gambro Lundia AB

Group Name Parameter Default Minimum Maximum Comment


limits LL 33.0 33.0 HL
HL 40.0 LL 40.0
2-32 Max pH limits pH limits Maximum allowed pH limits.
LL 5.00 5.00 HL
HL 9.00 LL 9.00
2-33 Max venous Limits (mmHg) Maximum allowed venous pressure limits.
pressure limits LL -100 -100 HL
HL 500 LL 500
2-34 Max arterial Limits (mmHg) Maximum allowed arterial pressure limits.
pressure limits LL -700 -700 HL
HL 750 LL 750
2-35 Variable fluid flow Fluid flow
value LL 300 300 HL Minimum flow allowed
SET 500 LL HL Start-up value
HL 700 LL 700 Maximum flow allowed0
2-36 Initial INVA open Delay (seconds) 0 0 3600
delay
2-37 UFD replace. day Nbr Days 90 0 32767 Number of days between a change of UFD filter until an
count attention is generated.
0 = no attention.
2-38 UFD replace. disinf Max count 150 0 32767 Number of disinfections between a change of UFD filter until
count an attention is generated.
0 = no attention.
2-39 UFD replace. hypo Max count 12 0 32676 Number of hypochlorite disinfections between a change of
count UFD filter until an attention is generated.
0 = no attention.
2-40 Dialysis pressure Limits (mmHg) 900 700 1000 Maximum allowed dialysis pressure.
Sup
3 Override times
3-0 AirDetOverrideTim AirDetT Time (s) 0 0 15 Air detector override time during treatment (0 = inactivated).
eTmt In alarm situations during treatment it can be possible to
override the alarm actions for a limited time.
3-1 AprOverrideTime Apr Time (s) 15 5 120 Arterial pressure override time.
In alarm situations it is possible to override the alarm actions
for a limited time.
3-2 BlLeakOverrideTim BlLeak Time (s) 15 5 120 Blood leak attention override time.

MDREC-132304 AK 96 program version 3.50 10 (28)


Gambro Lundia AB

Group Name Parameter Default Minimum Maximum Comment


e In alarm situations it is possible to override the alarm actions
for a limited time.
3-3 VprOverrideTime Vpr Time (s) 15 5 120 Venous pressure override time.
In alarm situations it is possible to override the alarm actions
for a limited time.
3-4 AirDetOverrideTim AirDetP Time (s) 3 0 15 Air detector override time during priming (0 = inactivated).
ePrim In alarm situations during priming it can be possible to override
the alarm actions for a limited time.
4 Assisted priming
4-0 Assisted priming Default ALT ALT1 ALT1 ALT3 Active Assisted priming bank during start-up.
default ALT
4-1 Assisted priming Label DIA 1 <text> <text> Label = in ALT1
ALT1 FillFlow (ml/min) 20 0 500 FillFlow = Blood pump speed during fill of dialyzer. This is the
Fill time (s) 180 0 3600 initial flow rate when filling the dialyzer and blood lines.
Pr. build time (s) 0 0 30 Fill time = Blood pump running time during fill of dialyzer. This
Pr. drop time (s) 0 0 60 is the time for filling the dialyzer and blood lines com--pletely.
Pr. drop (mmHg) 0 0 1000 Pr. build time = Time-out if venous pressure high limit not
PrimFlow (ml/min) 100 0 500 reached. The venous clamp closes. This is the time for building
Priming time (s) 120 0 3600 up the venous pressure alarm limit.
Cycle time (s) 3 0 60 Pr. drop time = Pressure drop measuring time (only active if Pr.
Level time (s) 0 0 60 build time > 0). The measurement time for the pressure drop
Flushing time1 (s) 120 0 3600 when performing the leakage test of the dialyzer.
Flushing time 2 (s) 120 0 3600 Pr. drop = Pressure drop limit (only active if Pr. build time >0).
UF Rate (l/hour) 1.00 0.00 4.00 The acceptable pressure drop when performing the leakage
Ven line time (s) 0 0 120 test of the dialyzer.
Early FM start Not Inst Not Inst Inst PrimFlow = Blood pump speed during priming of dialyzer. This
is the flow rate during the priming phase as well as during the
flushing phases.
Priming time = Blood pump running time during priming of
dialyzer. This is the required time for the priming phase.
Cycle time = Venous clamp toggle time. This time specifies the
opening and closing time of the venous clamp during the
priming phase.
Level time = Time during fill of venous line. This is the time
necessary to fill up the venous drip chamber to a correct level.
Flushing time1 = Time during flush of dialyzer, part 1. During
this time, the remaining air in the dialyzer and blood lines will
MDREC-132304 AK 96 program version 3.50 11 (28)
Gambro Lundia AB

Group Name Parameter Default Minimum Maximum Comment


be expelled.
Flushing time 2 = Time during flush of dialyzer, part 2. If a
second flush is requested, select this time.
UF Rate = Assisted priming UF rate.
Ven line time = Time during flush of venous line (only active if
Level time > 0). The time when the venous clamp is open. In
order to expell any air bubbles in the venous line.
Early FM start = Inst: Dialysis fluid must be achieved before
assisted priming can start. Not Inst: Assisted priming can start
without dialysis fluid.
4-2 Assisted priming Label DIA 1 <text> <text> See Assisted priming ALT 1 (4-1)
ALT2 FillFlow (ml/min) 20 0 500
Fill time (s) 180 0 3600
Pr. build time (s) 0 0 30
Pr. drop time (s) 0 0 60
Pr. drop (mmHg) 0 0 1000
PrimFlow (ml/min) 100 0 500
Priming time (s) 120 0 3600
Cycle time (s) 3 0 60
Level time (s) 0 0 60
Flushing time1 (s) 120 0 3600
Flushing time 2 (s) 120 0 3600
UF Rate (l/hour) 1.00 0.00 4.00
Ven line time (s) 0 0 120
Early FM start Not Inst Not Inst Inst
4-3 Assisted priming Label DIA 1 <text> <text> See Assisted priming ALT 1 (4-1)
ALT3 FillFlow (ml/min) 20 0 500
Fill time (s) 180 0 3600
Pr. build time (s) 0 0 30
Pr. drop time (s) 0 0 60
Pr. drop (mmHg) 0 0 1000
PrimFlow (ml/min) 100 0 500
Priming time (s) 120 0 3600
Cycle time (s) 3 0 60
Level time (s) 0 0 60
Flushing time1 (s) 120 0 3600
Flushing time 2 (s) 120 0 3600
UF Rate (l/hour) 1.00 0.00 4.00
MDREC-132304 AK 96 program version 3.50 12 (28)
Gambro Lundia AB

Group Name Parameter Default Minimum Maximum Comment


Ven line time (s) 0 0 120
Early FM start Not Inst Not Inst Inst
5 Rinse back
5-1 Rinseback flow Rinsebck flow 100 25 500 Blood pump speed during rinse back.
(ml/min)
5-2 Rinseback volume Rinsebck vol (l) 0.50 0.00 5.00 Accumulated volume during rinse back.
5-4 Rinseback fluid Rinseb. Fluidbyp Not Inst Not Inst Inst Fluid flow in bypass during rinse back.
bypass
5-5 Rinseback no blood No blood detect True False True False = Blood pump stop when priming detector stop sensing
detect blood or when rinseback volume is achieved which ever comes
first.
True = Blood pump will run until rinseback volume is achieved.
5-6 Extra Rinseback Extra Rb vol (l) 0.20 0.00 1.00 Accumulated volume during extra rinse back.
volume
5-7 Rinseback ven. Open dur. (s) 2 0.5 2300 During Rinse-back the venous clamp can open and close
clamp open time intermediatly.
This value specifies the duration of the clamp open phase.
5-8 Rinseback ven Close dur. (s) 0 0 2300 During Rinse-back the venous clamp can open and close
clamp close time intermediatly.
This value specifies the duration of the clamp close phase.
0 = No intermediate open and closing of the venous clamp
during Rinse-back.
5-9 Gravity rinseback Gravity rinse Not Inst Not Inst Inst Not Inst = The arterial clamp is closed when the blood pump is
stopped.
Inst = The arterial clamp remains open when the blood pump is
stopped during Rinse-back.
6 Disinfection
6-0 Chemical
disinfection
6-0-0 Chem ALT1 Chemical disinfection bank 1.
6-0-0-0 Chem ALT1 label Label Perace <text> <text> Label for chemical disinfection bank 1.
6-0-0-1 Chem ALT1 Central chem False False True Central chemical disinfection installed.
Central chem
6-0-0-2 Chem ALT1 Concentrate Conc Conc Work sol Concentrated or Working (mixed) solution (only active if 6-0-0-
Conc./Work 1 = FALSE)

MDREC-132304 AK 96 program version 3.50 13 (28)


Gambro Lundia AB

Group Name Parameter Default Minimum Maximum Comment


solution
6-0-0-3 Chem ALT1 mixing Mixing ratio 1+34 1+9 1+39 Mixing ratio of the concentrated solution (only active if 6-0-0-2
ratio = Conc and 6-0-0-10 = 0).
6-0-0-4 Chem ALT1 dwell Dwell time 00:10 00:00 09:55, Chemical disinfection dwell time. If dwell time FOREVER and
time FOREVER logical OFF, then machine will switch off after fill-up phase.
6-0-0-5 Chem ALT1 rinse Rinse time 00:11 00:11 09:55 Chemical disinfection rinse time.
time Rinse time if UFD is NOT installed (with UFD see 6-0-0-9)
6-0-0-6 Chem ALT1 Rinse temp 60 40 70 Chemical disinfection rinse temperature.
temperature
6-0-0-7 Chem ALT1 Cond. Cond. limits Conductivity limits for the disinfecting solution.
check LL 0.0 0.0 HL
HL 50.0 LL 50.0
6-0-0-8 Chem ALT1 OVERNIGHT False False True Overnight disinfection allowed.
Overnight dis. If this preset is True and the machine is switched off (logical
allowed off), the machine will turn off after fill up phase and stay off until
it is switched on either by logical on or by the auto heat / rinse
function independent of the dwell time settings.
6-0-0-9 Chem ALT1 rinse Rinse time UFD 00:39 00:21 09:59 Chemical disinfection rinse time.
time UFD Rinse time if UFD is installed (without UFD see 6-0-0-5)
6-0-0-10 Chem ALT 1 conc Conc percent (%) 0.0 0.0 100.0 0.0 = Not activated (6-0-0-3 is used).
percent 0.1-100.0 = The percent marked on the container, machine is
calculating the ratio automatically. (6-0-0-3 is NOT used). Note:
Do not set below 0.9%
6-0-1 Chem ALT2 Chemical disinfection bank 2.
6-0-1-0 Chem ALT2 label Label Hypochl <text> <text> Label for chemical disinfection bank 2.
10%
6-0-1-1 Chem ALT2 Central chem False False True See comment for ALT1 (6-0-0-1).
Central chem
6-0-1-2 Chem ALT2 Concentrate Conc Conc Work sol See comment for ALT1 (6-0-0-2).
Conc./Work
solution
6-0-1-3 Chem ALT2 mixing Mixing ratio 1+19 1+9 1+39 See comment for ALT1 (6-0-0-3).
ratio
6-0-1-4 Chem ALT2 dwell Dwell time 00:10 00:00 09:55, See comment for ALT1 (6-0-0-4).
time FOREVER
6-0-1-5 Chem ALT2 rinse Rinse time 00:11 00:11 09:55 See comment for ALT1 (6-0-0-5).
time

MDREC-132304 AK 96 program version 3.50 14 (28)


Gambro Lundia AB

Group Name Parameter Default Minimum Maximum Comment


6-0-1-6 Chem ALT2 Rinse temp 60 40 70 See comment for ALT1 (6-0-0-6).
temperature
6-0-1-7 Chem ALT2 Cond. Cond. limits See comment for ALT1 (6-0-0-7).
check LL 0.0 0.0 HL
HL 50.0 LL 50.0
6-0-1-8 Chem ALT2 OVERNIGHT False False True See comment for ALT1 (6-0-0-8).
Overnight dis.
allowed
6-0-1-9 Chem ALT2 rinse Rinse time UFD 00:28 00:21 09:59 See comment for ALT1 (6-0-0-9).
time UFD
6-0-1-10 Chem ALT 2 conc Conc percent (%) 0.0 0.0 100.0 See comment for ALT1 (6-0-0-10).
percent
6-0-2 Chem ALT3 Chemical disinfection bank 3.
6-0-2-0 Chem ALT3 label Label Formal <text> <text> Label for chemical disinfection bank 3.
37%
6-0-2-1 Chem ALT3 Central Chem False False True See comment for ALT1 (6-0-0-1).
Central chem
6-0-2-2 Chem ALT3 Concentrate Conc Conc Work sol See comment for ALT1 (6-0-0-2).
Conc./Work
solution
6-0-2-3 Chem ALT3 mixing Mixing ratio 1+9 1+9 1+39 See comment for ALT1 (6-0-0-3).
ratio
6-0-2-4 Chem ALT3 dwell Dwell time 00:20 00:00 09:55, See comment for ALT1 (6-0-0-4).
time FOREVER
6-0-2-5 Chem ALT3 rinse Rinse time 01:30 00:11 09:55 See comment for ALT1 (6-0-0-5).
time
6-0-2-6 Chem ALT3 Rinse temp 70 40 70 See comment for ALT1 (6-0-0-6).
temperature
6-0-2-7 Chem ALT3 Cond. Cond. limits See comment for ALT1 (6-0-0-7).
check LL 0.0 0.0 HL
HL 50.0 LL 50.0
6-0-2-8 Chem ALT3 OVERNIGHT False False True See comment for ALT1 (6-0-0-8).
Overnight dis.
allowed
6-0-2-9 Chem ALT3 rinse Rinse time UFD 02:00 00:21 09:59 See comment for ALT1 (6-0-0-9).
time UFD
6-0-2-10 Chem ALT 3 conc Conc percent (%) 0.0 0.0 100.0 See comment for ALT1 (6-0-0-10).
MDREC-132304 AK 96 program version 3.50 15 (28)
Gambro Lundia AB

Group Name Parameter Default Minimum Maximum Comment


percent
6-0-3 Active Chem ALT Chem dis Alt ALT1 ALT1 ALT3 Active chemical disinfection bank during start-up.
6-0-4 Central rinse delay Delay minutes (min) 0 0 1440 0 = Not activated
5 - 1440 = Time in minutes to start up in Auto rinse after shut
down from Central Chem. (Cental chem. and overnight
disinfection allowed must be set to True).
6-1 Heat disinfection
6-1-0 Heating (ALT1) Heat disinfection (water)
6-1-0-0 Heating Heating temp 93 70 95 Disinfection temperature for ALT1, ALT2 and ALT3
temperature Height above sea level: Water boiling temp:
0 100ºC
500 98.1ºC
1000 96.2ºC
1500 94.4ºC
2000 92.6ºC
2500 90.7ºC
6-1-0-1 Heating cycles Heating cycles 12 5 255 Number of heat cycles. 1cycle = 1 minute.
6-1-0-2 Heating extra cool Cool off (sec) DISABLED DISABLED 600 Cool off phase at end of disinfection.
off , 10
6-1-1 Heat decalcification Heat disinfection (liquid citric solution).
(ALT2)
6-1-1-0 Heat decalc. label Label Citric 20% <text> <text> Label for heat disinfection bank 2.
6-1-1-1 Heat decalc. Concentrate Conc Conc Worl sol Concentrated or Working (mixed) solution.
concentrate
6-1-1-2 Heat decalc. mixing Mixing ratio 1+9 1+9 1+39 Mixing ratio of the concentrated solution (only active if 6-1-1-1
ratio = Conc and 6-1-1-9 = 0)
6-1-1-3 Heat decalc. dwell Dwell time 00:10 00:00 09:55, Heat disinfection dwell time. If dwell time FOREVER and
time FOREVER logical OFF, then machine will switch off after fill-up phase.
6-1-1-4 Heat decalc. Heating cycles 12 12 255 Number of heat cycles. 1cycle = 1 minute.
heating cycles
6-1-1-5 Heat decalc. extra Cool off (sec) DISABLED DISABLED 600 Cool off phase at end of disinfection.
cool off , 10
6-1-1-6 Heat decalc. cond. Cond limits Conductivity limits.
check LL 0.0 0.0 HL
HL 50.0 LL 50.0
6-1-1-7 Post chem. rinse Rinse Time 00:11 00:11 09:55 Rinse time for Post Chem program without UFD.
time
MDREC-132304 AK 96 program version 3.50 16 (28)
Gambro Lundia AB

Group Name Parameter Default Minimum Maximum Comment


6-1-1-8 Post chem. rinse Rinse Time UFD 00:21 00:11 09:55 Rinse time for Post Chem program with UFD.
time UFD
6-1-1-9 Heat decalc. conc Conc percent (%) 0.0 0.0 100.0 0.0 = Not activated (6-1-1-2 is used)
percent 0.1-100.0 = The percent marked on the container, machine is
calculating the ratio automatically. (6-1-1-2 is NOT used). Note:
Do not set below 18%
6-1-2 Heat decalc Heat disinfection (CleanCart)
CleanCart (ALT3)
6-1-2-0 Heat CleanCart Dwell Time 00:10 00:00 09:55, Heat disinfection dwell time. If dwell time FOREVER and
dwell time FOREVER logical OFF, then machine will switch off after fill-up phase
6-1-2-1 Heat CleanCart Heating cycles 12 12 255 Number of heat cycles. 1cycle = 1 minute.
heating cycles
6-1-2-2 Heat CleanCart Cool off (sec) DISABLED DISABLED 600 Cool off phase at end of disinfection.
extra cool off , 10
6-1-2-3 Post chem. rinse Rinse Time 00:13 00:11 09:55 Rinse time for Post Chem program without UFD.
time
6-1-2-5 Post chem. rinse Rinse Time UFD 00:21 00:11 09:55 Rinse time for Post Chem program with UFD.
time UFD
6-1-3 Active Heat ALT Heat Dis Alt ALT1 ALT1 ALT3 Active heat disinfection bank during start-up.
6-1-4 Pump control
values
6-1-4-0 Conc. pump flush A+B pump speed (%) 10 10 100 Conc. Pump speed during dwell.
recirc. Speed
6-1-4-1 Flush recirc. HPG Pressure (mmHg) -100 -200 200 Suction pump regulation during dwell.
pressure
6-1-4-2 Recirc. ZEVA press Pressure (mmHg) 500 400 600 Determines the limit for ZEVA to open in order to release
release lim pressure during heat recirculation.
6-1-4-3 Recirc. Suc Pump 20% speed True False True Suction pump ON/OFF during heat recirculation.
20% duty cycle
6-1-5 WRO 300 H
6-1-5-0 Auto. disinfection of With WRO 300 H True False True Determines the default value for whether WRO 300 H shall be
WRO 300 H included in heat disinfection or not.
6-1-5-1 Cold preparation Prepare cold tme 00:01 00:00 01:00 Describes the estimated time it should take for WRO 300 H to
time be able to start producing cold water. The inlet pressure
attention is activated after this time. This time is also used
when the heat time is estimated.
6-1-5-2 LFH preparation Prepare LFH time 00:05 00:00 01:00 Describes the estimated time it should take for WRO 300 H to
MDREC-132304 AK 96 program version 3.50 17 (28)
Gambro Lundia AB

Group Name Parameter Default Minimum Maximum Comment


time be able to start producing warm water. This time is also used
when the heat time is estimated.
6-1-5-3 Heat + cool off time Heat + cool time 01:00 00:00 01:00 Describes the estimated time to do the self-heating of the
WRO 300 H + the time for cool off. This preset is used in the
calculation of the auto heat time.
6-1-5-5 Low Flow Heat time Low flow heat time 00:15 00:00 02:00 Describes for how long the low flow heating shall be done.
6-1-5-6 Max startup LFH LFH mode Timeout 00:05 00:01 01:00 Determines the maximum allowable time before the LFH mode
mode time starts. This time affects the attention for “LOW FLOW HEAT
ABORTED”.
6-1-5-7 Communication Comm mode Automat. Manual Automat. Determines if low flow heat mode shall be started manually or
mode if the machine shall communicate with WRO 300 H in order to
start it.
6-1-5-8 Low Flow Heat Pressure (mmHg) 100 0 200 Determines the set value of the degassing pressure and the
pressure HPG pressure during low flow heat. The flow pump and suction
pump run in order to obtain this pressure.
6-2 Rinse /
Decalcification
6-2-0 Rinse/Drain (ALT1) Rinse, bank 1 (water)
6-2-0-0 Rinse/Drain total Total time 00:10 00:10 09:59 Total rinse time
time
6-3 Remove CleanCart Remove CleanCrt False False True Attention to remind operator to remove CleanCart after
attention priming.
If FALSE then CleanCart mounted during all rinse/disinfection
program
6-4 Disinf/Decalc attn.
and lim.
6-4-0 Time between two Time Hour 120 0 500 Time, in hours, between disinfections, concerns calendar time.
disinfections The time flag is stored in Saferam C and cleared by both the
heat programs and the chemical programs.
6-4-1 Time between two Time Hour 0 0 500 Time, in hours, between decalcifications, concerns time in
decalcificat. Bicarbonate mode. The time flag is stored in Saferam C and
cleared by either CleanCart C or liquid citric.
To clear the flag by CleanCart C, there is a conductivity
measuring during CleanCart heat (<15mS/cm).
6-4-2 Time between two Time Hour 50 0 200 Time, in hours, between CleanCart A / hypochlorite cleanings
cleanings concerns total time in fluid mode (Bicarbonate or Acetate).
The time flag is stored in Saferam C and cleared by either

MDREC-132304 AK 96 program version 3.50 18 (28)


Gambro Lundia AB

Group Name Parameter Default Minimum Maximum Comment


Cleancart A or hypochloride.
To clear the flag by CleanCart A, there is a conductivity
measuring during CleanCart heat (>15mS/cm for
approximately 2sec).
To clear the flag by hypochloride it is of importance that
hypochloride is placed in bank 2 since in this case it is the
bank that clears the flag.
6-5 AUTO OFF after Time Minute DISABLED DISABLED 100 Time in minutes before the machine automatically turns off with
Rinse/Drain ,1 blinking Rinse/Drain after a disinfection program.
6-6 HPG Max Limit HPG Max Disinf 750 450 1500 Maximum allowed pressure, measured by HPG and PD, during
During Disinf disinfection. If this pressure is exceeded an attention will be
generated and the disinfection is aborted.
7 Single needle
7-0 Arterial phase time Art time S/N (s) 10.0 2.0 20.0 Arterial phase in single needle (start-up value)
7-1 Venous phase time Ven time S/N (s) 10.0 2.0 20.0 Venous phase in single needle (start-up value)
7-2 Stroke volume limit Stroke Vol (ml) 0 0 200 Stroke volume limit in single needle (start-up value)
There are three options to enter the values for each Acetate
8 Acetate concentrates alternative,
Preset using gram/litre, molecules
Preset using mmol/litre, molecules
Preset using mmol/litre, ions

When mmol/litre, ions is used to enter the values the


chemical checksum is calculated by the machine automatically.

Note: only one of these presets needs to be / can be used to


enter the values for each Acetate concentrate
8-0 Acetate alt. 1 Acetate concentrate, bank 1
8-0-0 Acetate 1 Ac 1 Label Label for bank 1 (text)
(gram/l, molecule) CH3CooNa Sodium Acetate
H2O with CH3COONa Sodium Acetate, hydrated
NaCl Sodium Chloride
KCl Potassium Chloride
MgCl2 Magnesium Chloride
H2O with MgCl2 Magnesium Chloride, hydrated
CaCl2 Calcium Chloride
H2O with CaCl2 Calcium Chloride, hydrated
MDREC-132304 AK 96 program version 3.50 19 (28)
Gambro Lundia AB

Group Name Parameter Default Minimum Maximum Comment


8-0-1 Acetate 1 Ac 1 label Label for bank 1 (text)
(mmol/l, molecule CH3COONa Sodium Acetate
NaCl Sodium Chloride
KCI Potassium Chloride
MGCl2 Magnesium Chloride
CaCl2 Calcium Chloride
8-0-2 Acetate 1 Ac 1 label Label for bank 1
(mmol/l, ion) Na Sodium ion
K Potassium ion
Mg Magnesium ion
Ca Calcium ion
Cl Chloride ion
CH3COO Acetate ion
8-1 Acetate alt. 2 Acetate concentrate, bank 2
8-1-0 Acetate 2 See bank 1 (8-0-0) See comment for bank 1 (8-0-0).
(gram/l, molecule)
8-1-1 Acetate 2 See bank 1 (8-0-1) See comment for bank 1 (8-0-1).
(mmol/l, molecule
8-1-2 Acetate 2 See bank 1 (8-0-2) See comment for bank 1 (8-0-2).
(mmol/l, ion)
8-2 Acetate alt. 3 Acetate concentrate, bank 3
8-2-0 Acetate 3 See bank 1 (8-0-0) See comment for bank 1 (8-0-0).
(gram/l, molecule)
8-2-1 Acetate 3 See bank 1 (8-0-1) See comment for bank 1 (8-0-1).
(mmol/l, molecule
8-2-2 Acetate 3 See bank 1 (8-0-2) See comment for bank 1 (8-0-2).
(mmol/l, ion)

MDREC-132304 AK 96 program version 3.50 20 (28)


Gambro Lundia AB

Group Name Parameter Default Minimum Maximum Comment


Presets for liquid Bicarbonate concentrate:
9 Bicarbonate concentrates 9-0-0-
Presets for BiCart concentrate:
9-0-1-

There are three options to enter the values for each


Bicarbonate alternative,
Preset using gram/litre, molecules
Preset using mmol/litre, molecules
Preset using mmol/litre, ions

When mmol/litre, ions is used to enter the values the


chemical checksum is calculated by the machine automatically.

Note: only one of these presets needs to be / can be used to


enter the values for each Bicarbonate concentrate
9-0 Bicarbonate alt. 1 Bicarbonate concentrate, bank 1.
9-0-0 Liquid bicarbonate Liquid bicarbonate for bank 1.
1
9-0-0-0 Bicarbonate 1 Bicarb 1 Label Label for bank 1 (text)
(gram/l, molecule) A:CH3COOH Acetic Acid
A:CH3COONa Sodium Acetate
H2O with A:CH3COONa Acetic Acid, hydrated
A:NaCl Sodium Chloride
A:KCl Potassium Chloride
A:MgCl2 Magnesium Chloride
H2O with A:MgCl2 Magnesium Chloride, hydrated
A:CaCl2 Calcium Chloride
H2O with A:CaCl2 Calcium Chloride, hydrated
B:NaHCO3 Sodium Bicarbonate
B:NaCl Sodium Chloride
B:CH3COONa Sodium Acetate
H2O with B:CH3COONa Acetic Acid, hydrated
9-0-0-1 Bicarbonate 1 Bicarb 1 Label Label for bank 1 (text)
(mmol/l, molecule) A:CH3COOH Acetic Acid
A:CH3COONa Sodium Acetate
A:NaCl Sodium Chloride
A:KCl Potassium Chloride
MDREC-132304 AK 96 program version 3.50 21 (28)
Gambro Lundia AB

Group Name Parameter Default Minimum Maximum Comment


A:MgCl2 Magnesium Chloride
A:CaCl2 Calcium Chloride
B:NaHCO3 Sodium Bicarbonate
B:NaCl Sodium Chloride
B:CH3COONa Sodium Acetate
9-0-0-2 Bicarbonate 1 Bicarb 1 Label Label for bank 1 (text)
(mmol/l, ion) A:Na Sodium ion
A:K Potassium ion
A:Mg Magnesium ion
A:Ca Calcium ion
A:Cl Chloride ion
A:CH3COO Acetate ion
A:H Hydrogen ion
B:Na Sodium ion
B:HCO3 Bicarbonate ion
B:Cl Chloride ion
B:CH3COO Acetate ion
9-0-1 BiCart 1 BiCart bicarbonate for bank 1.
9-0-1-0 BiCart 1 See bank 1 (9-0-0-0) See comment for liquid bicarbonate bank 1 (9-0-0-0).
(gram/l, molecule)
9-0-1-1 BiCart 1 See bank 1 (9-0-0-1) See comment for liquid bicarbonate bank 1 (9-0-0-1).
(mmol/l, molecule)
9-0-1-2 BiCart 1 See bank 1 (9-0-0-2) See comment for liquid bicarbonate bank 1 (9-0-0-2).
(mmol/l, ion)
9-1 Bicarbonate alt. 2 Bicarbonate concentrate, bank 2.
9-1-0 Liquid bicarbonate Liquid bicarbonate for bank 2.
2
9-1-0-0 Bicarbonate 2 See bank 1 (9-0-0-0) See comment for liquid bicarbonate bank 1 (9-0-0-0).
(gram/l, molecule)
9-1-0-1 Bicarbonate 2 See bank 1 (9-0-0-1) See comment for liquid bicarbonate bank 1 (9-0-0-1).
(mmol/l, molecule)
9-1-0-2 Bicarbonate 2 See bank 1 (9-0-0-2) See comment for liquid bicarbonate bank 1 (9-0-0-2).
(mmol/l, ion)
9-1-1 BiCart 2 BiCart bicarbonate for bank 2.
9-1-1-0 BiCart 2 See bank 1 (9-0-0-0) See comment for liquid bicarbonate bank 1 (9-0-0-0).
(gram/l, molecule)
9-1-1-1 BiCart 2 See bank 1 (9-0-0-1) See comment for liquid bicarbonate bank 1 (9-0-0-1).

MDREC-132304 AK 96 program version 3.50 22 (28)


Gambro Lundia AB

Group Name Parameter Default Minimum Maximum Comment


(mmol/l, molecule)
9-1-1-2 BiCart 2 See bank 1 (9-0-0-2) See comment for liquid bicarbonate bank 1 (9-0-0-2).
(mmol/l, ion)
9-2 Bicarbonate alt. 3 Bicarbonate concentrate, bank 2.
9-2-0 Liquid bicarbonate Liquid bicarbonate for bank 3
3
9-2-0-0 Bicarbonate 3 See bank 1 (9-0-0-0) See comment for liquid bicarbonate bank 1 (9-0-0-0).
(gram/l, molecule)
9-2-0-1 Bicarbonate 3 See bank 1 (9-0-0-1) See comment for liquid bicarbonate bank 1 (9-0-0-1).
(mmol/l, molecule)
9-2-0-2 Bicarbonate 3 See bank 1 (9-0-0-2) See comment for liquid bicarbonate bank 1 (9-0-0-2).
(mmol/l, ion)
9-2-1 BiCart 3 BiCart bicarbonate for bank 3.
9-2-1-0 BiCart 3 See bank 1 (9-0-0-0) See comment for liquid bicarbonate bank 1 (9-0-0-0).
(gram/l, molecule)
9-2-1-1 BiCart 3 See bank 1 (9-0-0-1) See comment for liquid bicarbonate bank 1 (9-0-0-1).
(mmol/l, molecule)
9-2-1-2 BiCart 3 See bank 1 (9-0-0-2) See comment for liquid bicarbonate bank 1 (9-0-0-2).
(mmol/l, ion)
10 BPM settings
10-0 BPM systolic BPM sys limits Lower and upper alarm limit for Systolic blood pressure
pressure limits LL 100 60 HL (mmHg).
HL 150 LL 250
10-1 BPM diastolic BPM dia limits Lower and upper alarm limit for Diastolic blood pressure
pressure limits LL 60 40 HL (mmHg).
HL 100 LL 200
10-2 BPM mean BPM mean limits Lower and upper alarm limit for Mean blood pressure (mmHg).
pressure limits LL 75 45 HL
HL 115 LL 235
10-3 BPM pulse limit BPM pulse Lower and upper alarm limit for Pulse rate.
LL 40 40 HL
HL 150 LL 200
10-4 BPM measurement BPM meas interval 30 5 60 Measurement interval for BPM measurements during treatment
interval if the BPM is in auto mode.
10-5 BPM alarm mode BPM alarm OFF OFF ON BPM alarm on/off. Enable /Disable BPM alarms according to
10-0, 10-1, 10-2 and 10-3.
MDREC-132304 AK 96 program version 3.50 23 (28)
Gambro Lundia AB

Group Name Parameter Default Minimum Maximum Comment


10-6 BPM initial cuff BPM init. cuff pr 180 120 250 Initial cuff pressure.
pressure
10-7 BPM inflation BPM infl pr mode Measured Preset Measured Preset = Inflation pressure set by the preset 10-6, is used.
pressure mode Measured = First measurement: inflation pressure is set to
initial value.
Thereafter: inflation pressure = last systolic BP value + α (α =
30mmHg) Note! Preset 10-6 not used.
10-8 BPM meas ready BPM rdy attn OFF OFF ON OFF = Do not generate an attention when a new BPM
attention measurement is available.
ON = Generate an attention when a new BPM measurement is
available.
11 Profiling
11-0 Model 1
11-0-0 Na+ Acetate Na+ prof type OFF OFF LINEAR Na+ profiling type in Acetate.
profiling type Profiling curve for Na+ in Acetate mode. Can be set to either:
OFF
LINEAR
11-0-1 Na+ profiling type Na+ prof type OFF OFF LINEAR Na+ profiling type in Bicarbonate.
Profiling curve for Na+ in Bicarbonate mode. Both Na+ and
HCO3- profiling can be active in Bicarbonate mode. Can be set
to either:
OFF
LINEAR
11-0-2 HCO3- profiling HCO3- prof type OFF OFF LINEAR HCO3- profiling type.
type Profiling curve for HCO3-. Both Na+ and HCO3- profiling can
be active in Bicarbonate mode. Can be set to either:
OFF
LINEAR
11-0-3 UF profiling type UF prof type OFF See See UF profiling type.
comment comment Profiling curve for UF rate. Can be set to either:
OFF
LINEAR
STEP
INTERVAL.
11-0-4 Na+ Acetate start Start value 140.0 130.0 160.0 Start value for Na+ in Acetate mode.
value

MDREC-132304 AK 96 program version 3.50 24 (28)


Gambro Lundia AB

Group Name Parameter Default Minimum Maximum Comment


11-0-5 Na+ Acetate stop Stop value 140.0 130.0 160.0 Stop value for Na+ in Acetate mode.
value
11-0-6 Na+ start value Start value 140.0 130.0 160.0 Start value for Na+ in Bicarbonate mode.
11-0-7 Na+ stop value Stop value 140.0 130.0 160.0 Stop value for Na+ in Bicarbonate mode.
11-0-8 HCO3- start value Start value 34.0 20.0 40.0 Start value for HCO3-.
11-0-9 HCO3- stop value Stop value 34.0 20.0 40.0 Stop value for HCO3-.
11-0-10 UF rate diff (l/h) UF rate diff 0.00 -5.00 5.00 Difference between UF start value and UF stop value. The
machine uses this preset to calculate the start and stop UF
rate if UF profiling curve, based on UF volume and treatment
time.
11-0-11 UF profiling UF interval 2 2 10 Number of steps/intervals during the treatment.
steps/intervals Total treatment time / the preseted value = time in each
step/interval
11-1 Model 2
11-1-0 Na+ Acetate Na+ prof type OFF OFF LINEAR See comment for Model 1 (11-0-0).
profiling type
11-1-1 Na+ profiling type Na+ prof type OFF OFF LINEAR See comment for Model 1 (11-0-1).
11-1-2 HCO3- profiling HCO3- prof type OFF OFF LINEAR See comment for Model 1 (11-0-2).
type
11-1-3 UF profiling type UF prof type OFF See See See comment for Model 1 (11-0-3).
comment comment
11-1-4 Na+ Acetate start Start value 140.0 130.0 160.0 See comment for Model 1 (11-0-4).
value
11-1-5 Na+ Acetate stop Stop value 140.0 130.0 160.0 See comment for Model 1 (11-0-5).
value
11-1-6 Na+ start value Start value 140.0 130.0 160.0 See comment for Model 1 (11-0-6).
11-1-7 Na+ stop value Stop value 140.0 130.0 160.0 See comment for Model 1 (11-0-7).
11-1-8 HCO3- start value Start value 34.0 20.0 40.0 See comment for Model 1 (11-0-8).
11-1-9 HCO3- stop value Stop value 34.0 20.0 40.0 See comment for Model 1 (11-0-9).
11-1-10 UF rate diff (l/h) UF rate diff 0.00 -5.00 5.00 See comment for Model 1 (11-0-10).
11-1-11 UF profiling UF interval 2 2 10 See comment for Model 1 (11-0-11).
steps/intervals
11-2 Model 3
11-2-0 Na+ Acetate Na+ prof type OFF OFF LINEAR See comment for Model 1 (11-0-0).
profiling type
11-2-1 Na+ profiling type Na+ prof type OFF OFF LINEAR See comment for Model 1 (11-0-1).
11-2-2 HCO3- profiling HCO3- prof type OFF OFF LINEAR See comment for Model 1 (11-0-2).
MDREC-132304 AK 96 program version 3.50 25 (28)
Gambro Lundia AB

Group Name Parameter Default Minimum Maximum Comment


type
11-2-3 UF profiling type UF prof type OFF See See See comment for Model 1 (11-0-3).
comment comment
11-2-4 Na+ Acetate start Start value 140.0 130.0 160.0 See comment for Model 1 (11-0-4).
value
11-2-5 Na+ Acetate stop Stop value 140.0 130.0 160.0 See comment for Model 1 (11-0-5).
value
11-2-6 Na+ start value Start value 140.0 130.0 160.0 See comment for Model 1 (11-0-6).
11-2-7 Na+ stop value Stop value 140.0 130.0 160.0 See comment for Model 1 (11-0-7).
11-2-8 HCO3- start value Start value 34.0 20.0 40.0 See comment for Model 1 (11-0-8).
11-2-9 HCO3- stop value Stop value 34.0 20.0 40.0 See comment for Model 1 (11-0-9).
11-2-10 UF rate diff (l/h) UF rate diff 0.00 -5.00 5.00 See comment for Model 1 (11-0-10).
11-2-11 UF profiling UF interval 2 2 10 See comment for Model 1 (11-0-11).
steps/intervals
12 Diascan settings
12-0 Conductivity step Step height 50 10 50 Step height is how much the conductivity is increased or
height decreased during the conductivity step of the Diascan
measurement.
12-1 Conductivity step Step max height 50 10 50 Step maximum is how much the conductivity alarm limit is
maximum increased in protective system during the conductivity step.
12-2 Conductivity step Step duration (sec) 120 1 900 Step duration is the duration of the conductivity step.
duration
12-3 Sodium load Dose max limit 60 0 100 Sodium load maximum limit is the maximum sodium dose
maximum limit (conductivity step area) allowed during a treatment. The
sodium dose is calculated during the conductivity step and 60
seconds after the step is completed.
12-4 Sodium load Dose min limit -60 -100 0 Sodium load minimum limit is the minimum sodium dose
minimum limit (conductivity step area) allowed during a treatment. The
sodium dose is calculated during the conductivity step and 60
seconds after the step is completed
12-5 Patient total body Water volume 40 0 150 Patient total body water is the urea distribution volume in the
water Patient.
12-6 Clearance alarm K alarm limit 0 0 500 Clearance alarm limit is the lowest acceptable clearance
limit measurement.
12-7 Clearance target Kt/V target 0.00 0.00 2.00 Target is the desired Kt/V dose to be achieved during the
treatment.
MDREC-132304 AK 96 program version 3.50 26 (28)
Gambro Lundia AB

Group Name Parameter Default Minimum Maximum Comment


12-8 Clearance target Kt/V limit 0.00 0.00 1.95 Target alarm is the lowest acceptable Kt/V dose at the end of
alarm limit treatment.
12-9 Measuring mode Meas mode OFF See See The Diascan start up mode, can be set to:
comment comment OFF
Single
15 min
30 min
60 min
12-10 Minimum Cdi for Min Cdi neg step 1300 1100 1700 Minimum Cdi for negative step is the limit for the conductivity
negative step during a negative conductivity step.
If the conductivity will be lower than the minimum Cdi during
the conductivity step, a positive step is done instead.
12-11 Diascan ready Diascan rdy attn Not Inst Not Inst Inst Not Inst = Do not generate an attention when a new Diascan
attention measurement is available.
Inst = Generate an attention when a new Diascan
measurement is available.
13 User interface
13-0 Open setting Setting reminder Not Inst Not Inst Inst Not Inst = The operator will not be notified when a changed
reminder setting is left open.
Inst = If a changed setting is left open for more than 30
seconds without any buttons being pressed, the operator will
be notified.
13-1 Menu closed Closed MenuAttn Not Inst Not Inst Inst Not Inst = No attention will appear when leaving the menu
attentions while a setting is open. The changed value is still used.
Inst = If a changed setting is not closed when leaving the
menu, an attention will inform the operator that the changed
value is being used.
14 Recirculation
14-0 Blood flow (ml/min) Blood flow 150 20 500 Default blood flow when starting recirculation.
14-1 Dialysis flow Dia flow (ml/min) 500 300 700 Default dialysis flow when starting priming and recirculation.
(ml/min)
14-2 UF volume and Recirc
time UF vol (l) 0.20 0.00 4.00 Set UF volume during recirculation.
Time (min) 00:10 00:00 01:00 Set Time during recirculation.
14-3 Recirculation CSBM False False True False = Do not start concentrate stand by mode during

MDREC-132304 AK 96 program version 3.50 27 (28)


Gambro Lundia AB

Group Name Parameter Default Minimum Maximum Comment


CSBM recirculation.
True = Start concentrate stand by mode when the recirculation
UF volume has been achieved.
14-4 Extra UF volume Extra Recirc
and time UF vol (l) 0.10 0.00 2.00 Set UF volume during extra recirculation.
Time (min) 00:05 00:00 00:30 Set Time during extra recirculation.

MDREC-132304 AK 96 program version 3.50 28 (28)