Moxi All SpecSheet

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

Moxi All ™

T Moxi All Pro, T Moxi All 800, T Moxi All 700, T Moxi All 600, T Moxi All 500
13 receiver in canal (RIC) direct connectivity hearing instrument series
Moxi All

SoundCore T Pro T 800 T 700 T 600 T 500 Fitting guides


SoundNav Music • • 250 500 1000 2000 4000 8000 Hz
-10
Noise • • • 0
Conversation in a crowd • • • 10
20
Conversation in a small group • • • 30
40
Conversation in quiet • • • • 50
Conversation in noise • • • • 60
70
Quiet • • • • 80
90
Total environments 7 7 6 3 AutoMic 100
SpeechPro SpeechPro • SpeechZone 2 SpeechZone 110
120
Speech Locator • • •
Speech Focus • Standard receiver (xS)
Dynamic Spatial Awareness •
Sound Conductor Speech enhancement • • • • • -10
250 500 1000 2000 4000 8000 Hz

Noise reduction • • • • • 0
10
Adaptive directionality Multiband Multiband Multiband Multiband • 20
30
Spatial Awareness Spatial Awareness Dynamic Personalized • 40
Pinna Effect • • • • • 50
60
70
Sound Stabilization 80
90
AntiShock 2 • • • • • 100
Wind control • • • • • 110
120
Feedback manager • • • • •
Natural Sound Balance • • • • • Power receiver (xP)

Experience innovations 250 500 1000 2000 4000 8000 Hz


-10
Patient insights Log It All • • • • • 0
10
Patient Ratings • • • • • 20
Data logging • • • • • 30
40
Flex Flex:trial • • • • • 50
60
Flex:upgrade • • • • 70
80
Convenience 90
100
Direct connectivity • • • • • 110
120
Rechargeable option • • • • •
DuoLink • • • • • Super power plus receiver
(xSP plus)
Fitting
Automatic Adaptation Manager • • • • • Open dome
Closed dome
MyMusic Binaural Binaural • • • P
 ower dome or
Frequency compression • • • • • sleeve mold
Tinnitus masker • • • • •
IntelliVent • • • • •
Streaming programs • • • • •
Manual programs • • • • •
NAL-NL2/NL1 and DSLv5 • • • • •
Fitting channels 20 20 16 10 6

Moxi All is rated IP 68

Not all technology levels are available in all markets.

2017-10 027-6250-02
A Sonova brand ©2018 Unitron. All rights reserved.
Standard receiver Power receiver Super power plus
T Moxi All (xS) (xP) receiver (xSP plus)

ANSI 3.22 2014/IEC 60118-0: 2015 2cc coupler technical data


140
Pout
BSPL
130
OSPL90
120

110
Maximum (dB SPL) 111 124 127
100 HFA - OSPL90 (dB SPL) 106 119 122
90

80 Hz
100 1000 10000

70
Gain
dB60 Full on gain (input 50 dB SPL)
50
Maximum (dB) 47 57 66
40

30 HFA - FOG (dB) 40 50 59


20

10 Hz
100 1000 10000

120
Pout
BSPL
110
Reference test setting (RTS)
100 Frequency range (Hz) <100 - 8500 <100 - 7300 <100 - 6000
90

80 Reference test gain (dB) 29 42 45


70

60 Hz
Current drain at RTS (mA) 1.7 1.7 1.7
100 1000 10000
Typical battery life (h) 180 180 180
Equivalent input noise at RTS (dB SPL) 19 18 19
Total harmonic distortion at 500 Hz/800 Hz/1600 Hz (%) 1.0/1.0/1.0 1.5/1.0/0.5 1.0/1.5/1.0

Electromagnetic compatibility
EMC immunity by ANSI c63.19-2011 EMC, omni M4 M4 M4

IEC 60118-0: 1994 OES coupler technical data


Reference test frequency - IEC 60118-0 (kHz) 1.6 1.6 1.6
OSPL90
140
Pout
BSPL
130

120
Maximum (dB SPL) 122 133 138
110

100 at RTF (dB SPL) 114 130 136


90

80 Hz
100 1000 10000

80
Gain
dB
70
Full on gain (input 50 dB SPL)
60
Maximum (dB) 58 67 74
50

40 at RTF (dB) 48 62 71
30

20 Hz
100 1000 10000

130
Pout
BSPL
120
Basic frequency response
110
Frequency range (DIN 45605) (Hz) <100 - 9500 <100 - 6700 <100 - 5500
100

90 Reference test gain (dB) 39 55 61


80

70 Hz Current drain at RTG (mA) 1.8 1.8 2.0


100 1000 10000

Typical battery life (h) 170 170 160


Equivalent input noise at RTG (dB SPL) 19 19 19
Total harmonic distortion at 500 Hz/800 Hz/1600 Hz (%) 1.0/1.5/1.5 1.5/1.5/1.0 1.5/1.5/1.0
Electromagnetic compatibility
EMC immunity by IEC 60118-13, 2016 user compatibility, field 17/16/17 11/2/14 37/9/13
strength 90/50/35 V/m, omni. IRIL low/medium/high band (dB SPL)

Legend Test conditions


xS receiver Battery: size 13 Zinc-air; Source: voltage 1.3 V
xP receiver The measurements obtained with a closed configuration using an HA-1 coupler (ANSI-3.7-1995) or occluded ear simulator
xSP plus receiver (EN 60711, coupling arrangement according to fig. 4 in the test standard). The hearing instrument set to Unitron Truefit
test settings. LLE is applied at an approximate level of 35 dB SPL.
Domes should never be fit on patients with perforated eardrums, exposed middle ear cavities, or surgically altered
ear canals. In the case of such a condition, we recommend use of a customized earmold.
Sound pressure level of these hearing instruments exceeds 132 dB SPL.
We reserve the right to change specification data without notice as improvements are introduced.

unitron.com

You might also like