Professional Documents
Culture Documents
Insuquant Kit Manual: Capture The Potential
Insuquant Kit Manual: Capture The Potential
Insuquant Kit Manual: Capture The Potential
Protocol for using the InsuQuantTM Kit with the FinnpipetteTM Novus i Multichannel
Electronic Pipette
991INSK96C
991INSK96
991SP12
Revision Date: May 19, 2017
Table of Contents
System Overview 4
Required Equipment 5
Procedural Guidelines 5
InsuQuant Protocol 6
Time Allotment 6
Step 6 – Elution 11
Performance Characteristics 14
Ordering Information 15
www.thermofisher.com/MSIA
InsuQuantTM Kit Protocol
The kit contains the reagents and consumables necessary to perform insulin MSIA. The
primary component is the 96-pack of Insulin MSIA D.A.R.T.’S. Within each of the D.A.R.T.’S is a
monolithic microcolumn that is densely coated with anti-insulin antibody. Each microcolumn is
porous throughout its entire volume, allowing solution to be repeatedly passed through its
microfluidic channels. This repeated driving interaction between the immobilized antibody and
the solution allows for immunocapture to take place, capturing the analyte(s) of interest on the
surface of the microcolumn. After removing the extraneous particulates, the microcolumn is
incubated with an elution solution to release the analytes back into solution for analysis by LC-
MS.
5 6
1
3
4
2
www.thermofisher.com/MSIA
InsuQuantTM Kit Protocol
System Overview
The InsuQuant Kit provides the necessary reagents and consumables to perform the
immunoaffinity purification of insulin analogues from biological samples for subsequent analyses
by LC-MS. The InsuQuant Kit is compatible with the Thermo Scientific™ Finnpipette™ Novus i
Multichannel Electronic pipette and the Thermo Scientific Versette™ Automated Liquid Handler.
With these liquid handling solutions, parallel processing of 12 and 96 samples is achievable,
respectively. Both can be programmed to repetitively pipette bidirectionally (aspirate and
dispense) to enable the immunocapture of the target analytes.
There are three main steps that constitute the immunopurification of insulin (Figure 2). In Step 1,
the incubation of the Insulin MSIA D.A.R.T.’S with the diluted biological sample occurs through
repetitive pipetting, driving the immunocapture of insulin analogues. In Step 2, the Insulin MSIA
D.A.R.T.’S are rinsed with a wash buffer and water, removing extraneous components of the
sample. In the Step 3, the Insulin MSIA D.A.R.T.’S are incubated briefly with an Elution Solution
to release the captured analytes of interest into solution form. The eluates are then ready for
analysis by LC-MS.
Figure 2. An overview of the immunoaffinity purification of insulin and the bovine insulin internal standard
(ISTD) using the InsuQuant™ Kit.
www.thermofisher.com/MSIA
InsuQuantTM Kit Protocol
Required Equipment
The following equipment and materials are required for the Insulin MSIA protocol and are not
provided in the kit
Procedural Guidelines
Cover or cap all reagents when not in use
Store all components of the kit at 2 – 8 °C when not in use
Acclimate all buffers to room temperature before use
Stock solutions of insulin standard and internal standard (ISTD, bovine insulin)
should be kept at 2 – 8 °C to preserve integrity
Do not use any reagent or consumable after the kit expiration date, printed on the
label of the kit
Do not mix or interchange different reagent lots from various kit lots
Avoid multiple freeze/thaw cycles of frozen samples
Thaw frozen samples completely before use
Centrifuge thawed samples at low speed to pellet particulates, carefully remove
aliquots from centrifuged samples to avoid disruption of particulates, reducing the
risk of particulates clogging the microcolumns during sample incubation
Upon diluting samples in sample buffer, mix well prior to analysis
Run controls with every assay – controls in this protocol are optional only if the
circumstances call for a more ideal control sample
www.thermofisher.com/MSIA
InsuQuantTM Kit Protocol
Time Allotment
Table 1 – Time allotment per step of InsuQuant Protocol
(Based on parallel processing of 12 samples; 8 calibration points and 4 controls)
6 Elution 5 min
Wash 150 1 1 20
Capture 250 1 1 500
Elute 50 1 1 100
www.thermofisher.com/MSIA
InsuQuantTM Kit Protocol
Figure 3. 96-Well Color Coded Plate Map for InsuQuant with Finnpipette Novus i
Key to Figure 3
Representative
Reagent Volume (µL) Rows
Texture
Wash Buffer Lines 200 A, C, D
1. Label four (4) 2 mL and one (1) 15 mL centrifuge tubes with “internal standard (ISTD)” and
the concentrations listed in the right most column of Table 2.1.
2. Add the corresponding volume of diluent (Sample Buffer) to each of the labeled centrifuge
tubes represented under the column “Diluent: Sample Buffer” in Table 2.1.
3. Sequentially add the specified volume and concentration of internal standard (ISTD) to each
corresponding centrifuge tube beginning with Step 1 in Table 2.1. Please note that the
internal standard stock provided in the kit is used for the first step of the dilution only.
www.thermofisher.com/MSIA
InsuQuantTM Kit Protocol
Table 2.1. Serial dilutions for preparation of the Internal Standard (ISTD). The final volume prepares enough
for 12 samples (8 point calibration curve and 4 samples, or 12 samples), 500 µL/sample, maximizing the
throughput of a 12 channel Novus i.
4. Label ten (10) 2 mL centrifuge tubes with “Insulin Standard” and the concentrations listed in
the right most column of Table 2.2.
5. Add the corresponding volume of diluent (Matrix) to each of the labeled centrifuge tubes
represented under the column “Diluent: Matrix” in Table 2.2.
6. Sequentially add the specified volume and concentration of insulin standard to each
corresponding centrifuge tube beginning with Step 1 in Table 2.2. Please note that the
insulin stock provided in the kit is used for the first step of the dilution only.
Table 2.2 – Serial dilutions for the preparation of the Recombinant Human Insulin 8-point calibration curve.
The curve samples mimic a 500 µL sample volume and should be prepared in an appropriate *Matrix such
as analyte stripped plasma or 50 g/L BSA in 10mM PBS.
The following control helps test the performance of the assay. The controls may be replaced with
an in-house set of controls to further optimize the assay.
7. Label five (5) 2 mL and one (1) 15 mL centrifuge tubes with “Control” and the concentations
listed in the right most column in Table 2.3.
www.thermofisher.com/MSIA
InsuQuantTM Kit Protocol
8. Add the corresponding volume of diluent (Matrix) to each of the labeled centrifuge tubes
represented under the column “Diluent: Matrix” in Table 2.3.
9. Sequentially add the specified volume and concentration of insulin standard to each
corresponding centrifuge tube beginning with Step 1 in Table 2.3. Please note that the
insulin stock provided in the kit is used for the first step of the dilution only.
Table 2.3 – Dilutions for preparation of the Recombinant Human Insulin 50 pM control samples. The final
volume prepares 4 samples (500 µL/sample) and should be prepared in an appropriate *matrix such as
analyte stripped plasma or 50 g/L BSA in 10mM PBS.
2. Follow Table 3 to add 500 µL of each of the Insulin Standard Curve Calibrants (Table 2.2)
and the controls/samples (Table 2.3) to their corresponding wells in Row B of the Sample
Plate.
Curve - 7.5 pM B1
Curve - 15 pM B2
Curve - 30 pM B3
Curve - 60 pM B4
Curve - 240 pM B5
Curve - 480 pM B6
Curve - 720 pM B7
Curve - 960 pM B8
Control / Sample – 1 B9
Control / Sample – 2 B10
Control / Sample – 3 B11
Control / Sample – 4 B12
www.thermofisher.com/MSIA
InsuQuantTM Kit Protocol
Note: The Novus i is capable of processing up to 12 samples in parallel. If more than 12 samples (or 4
samples and an 8 point calibration curve) are to be analyzed, then all samples should be prepared at the
same time before proceeding to step 5. Prepare samples in sets of 12, depositing each set of 12
samples in row B, along with fresh wash buffer and water, in a new microtiter plate as depicted in Figure
3. Sequentially perform step 5 for each set of 12 samples, completing step 5 for each sample before
proceeding to step 6.
For larger numbers of samples, it is recommended that the samples be processed using a Thermo
Scientific™ Versette™ Automated Liquid Handler, which enables the processing of up to 96 samples in
parallel.
2. Place the Sample Plate containing the MSIA reagents (Reference Figure 4) onto the Novus i
stand platform. Position the plate as reflected in Figure 4.
Well
A1
3. Load 12 MSIA Insulin D.A.R.T.’S onto the 12-channel Novus i. Twist the D.A.R.T.’S into
place at the base to ensure they are securely attached. Position the D.A.R.T.’S directly
above the corresponding reagent located in Row A of the Sample Plate.
www.thermofisher.com/MSIA
InsuQuantTM Kit Protocol
Note: When lowering the D.A.R.T’S into the appropriate wells, ensure that the tips are 2 - 3 mm
from the bottom of the well.
Upon completion of a Novus i MSIA program, perform a “Blow Out” by raising the distal end of
the D.A.R.T’S to the surface of the reagent in the well and pressing the trigger button located on
the Novus i. Pull fully out of the liquid prior to the formation of the new air gap. If any liquid is
pulled up into the D.A.R.T.’S perform the program again, but press cancel immediately to
interrupt it. Then perform the second Blow Out to ensure no liquid remains.
4. Follow Table 4, which outlines the necessary order that each well and MSIA program should
be utilized to perform the affinity purification of insulin from the samples. For each step,
align the D.A.R.T.’S and the appropriate reagent by sliding the Sample Plate in the platform.
Table 4. Protocol to perform insulin affinity purification with the Novus i – step through each row with the
corresponding solution, volume, and program.
Reagent Volume in
Plate Row Assay Step Assay Solution Novus i Program
Well
A Buffer Pre-Rinse Wash Buffer 200 µL Wash
Calibrants, Controls and
B Insulin Capture 1000 µL Capture
Samples
Note: If more than 12 samples (or 8 point calibration curve and 4 samples) are being analyzed, then
process all samples up to step 6, Elution. Once all samples have been processed up to step 6, then
proceed with the elution step for each. The elution should be performed in a single microtiter plate to
consolidate all the eluates within a single plate to be loaded into the autosampler for LC-MS analyses.
Before proceeding with the elution step, install and condition the analytical LC column following the
additional instructions provided specific for the column.
Step 6 – Elution
1. Prepare the elution solution by adding 1 mL of the elution solvent to the vial of leucine
enkephalin, mixing to ensure the entire contents is dissolved into solution.
2. Add the 1 mL solution, containing the dissolved leucine enkephalin, back into the bottle of the
remaining elution solvent, effectively diluting the solution of leucine enkephalin into the entire
www.thermofisher.com/MSIA
InsuQuantTM Kit Protocol
volume of the orginal elution solvent. Gently mix the solution by either pipetting solution or
gently vortexing. The elution solution is now ready for use in the elution step.
Note: Store the prepared elution solution with the kit at 2 – 8 °C. The elution solution may be used up to 6
months from the date of its preparation.
4. The volume of Elution Solution to be used is dependent on the volume to be injected on the
LC system and the number of injections needed per each eluate. Therefore, it is
recommended to add 65 - 100 µL of the Elution Solution to each of the 12 wells in Row A of
the Elution Plate to accommodate your specific requirements.
5. Remove the Sample Plate from the Novus i stand and place the Elution Plate onto the stand
with the same positioning as shown in Figure 4.
6. Lower the D.A.R.T.’S into the wells of Row A and perform the Elution program outlined in
Table 5.
Plate Row Assay Step Assay Solution Total Well Volume Novus i Program
A Elution Elution Solution 65 - 100 µL Elute
7. Upon completion of the elution protocol, the eluate is ready to be injected onto the LC-MS
for analysis of intact insulin. Load eluate directly from elution plate onto the LC system. For
optimal sensitivity, ensure the majority of the eluate is loaded. However, less of eluate may
be loaded, but should be tested beforehand to ensure required sensitivity of the assay is
reached.
Note that sudden increases in flow rates may damage the monolithic column. Therefore, use gradual
increases in flow rate to prevent column damage. Furthermore, note the maximum pressure limit is 3000
psi (20.7 MPa) for the 1 x 250 mm column and exceeding this should be avoided.
www.thermofisher.com/MSIA
InsuQuantTM Kit Protocol
The optimization of the method for insulin analyses was performed on a Thermo Scientific™ Q Exactive™ Mass
Spectrometer. Using alternative instruments to the Q Exactive may require further optimization of the MS
parameters.
All data for determining performance characteristics were acquired using a Thermo Scientific™
Q Exactive Orbitrap mass spectrometer operated with the parameters listed in Table 7.
Acquiring HRAM MS data provides the benefit of utilizing the most abundant charge states and
the most abundant isotopes per each charge state to provide both qualitative validation, as well
as quantitative measurement for each insulin analog (Proteomics 2014, 14, 1445–1456). The
summation of the five most abundant isotopes per each of the +5 and +6 charge states were
combined to provide the total AUC (Area Under the Curve). The +5 and +6 charge states of the
rHum Insulin were normalized to the +5 and +6 charge states for the internal standard (ISTD),
bovine insulin, respectively, to generate the Peak Area Ratio plotted in the calibration curve of
Figure 5. Note that the prominent charge states of rHum and bovine insulins may vary from one
MS instrument to another depending on instrument tuning. We, therefore, highly recommend
that you determine the most prominent charge states of both rHum and bovine insulins on your
MS instrument and use those charge states for quantification.
For recomended parameters for other mass spectrometers or assistance, please contact your
local sales representative.
www.thermofisher.com/MSIA
InsuQuantTM Kit Protocol
Performance Characteristics
Figure 5. Recombinant Human Insulin Standard curve obtained for concentration range 7.5 -
960 pM. This study was performed using 50g/L BSA as sample matrix.
15.0
0.0
0 200 400 600 800 1000 1200
Recombinant Human insulin Concentration (pM)
Table 8. Intra-day assay data obtained from five recombinant human insulin standard curves (n=5)
performed on the same day.
Table 9. Inter-day assay data obtained from five recombinant human insulin standard curves (n=5)
performed on different days.
www.thermofisher.com/MSIA
InsuQuantTM Kit Protocol
Linearity Plot
Concentration (pM)
800
Insulin
600
Ordering Information
MSIA Kits for Immunoaffinity Capture
Compatible with the Thermo Scientific Versette Automated Liquid Handler and Thermo Scientific Finnpipette ® Novus i
Multichannel Electronic Pipette
Cat. No. Description Packaging
991INSK96 InsuQuant™ Kit 1 Kit
991INSK96C InsuQuant™ Kit with LC Analytical Column 1 Kit
MSIA D.A.R.T.’S for Immunoaffinity Capture
Compatible with the Thermo Scientific Versette Automated Liquid Handler and Thermo Scientific Finnpipette ® Novus i
Multichannel Electronic Pipette
www.thermofisher.com/MSIA
InsuQuantTM Kit Protocol
If the plasma is to be analyzed at a later date, apportion it into aliquots in polypropylene microcentrifuge tubes and
store at -80°C. Avoid multiple freeze-thaw cycles. When you are ready for analysis, allow the samples to thaw
completely on ice. All plasma samples should be clarified by centrifugation at 2,000 rpm for 10 min at 4°C in a
refrigerated microcentrifuge, and the supernatant transferred to the microtiter plate to be prepared for immediate
analysis following the protocol provided with the kit.
Why are the values of my calibration curve not matching the values in the example?
Common reasons for poor calibration curves include:
Not using the recommended method for solubilizing or diluting – standards should be reconstituted
according to the directions indicated on the manual
Using diluent not provided in the kit – no other diluent should be used
Improper mixing – the vials should be mixed gently and given 10 minutes to acclimate to room temperature
to ensure complete solubilization. They should then be used within 1 hr and mixed again before preparing
the dilutions according to the manual
I ran out of insulin standard that came in the kit. Can I get more?
The insulin standard conjugate that is supplied in this kit is not available as a stand-alone product.
www.thermofisher.com/MSIA