Pl M

PlazMax
Advantages of Plasma Technology

•Low Temperature Sterilization

•Can sterilize heat sensitive loads that cannot be subjected to steam
j

•Low energy requirements

•Low utility requirements (only sterilizing agent and electricity)
L tilit i t ( l t ili i t d l t i it )

•No drying stage

•No ventilation needed (like EtO)

Back Forward

1
Pl M
PlazMax
•Cost‐Effective

•After sterilization loads are dry and available for immediate use or storage which
reduces the need for expensive backup inventories.
It is recommended to wait for a successful biological test
before using the sterilized loads.

•Sterilized loads can remain in packaging (no need for repackaging)

•Fast turnaround means increased inventory utilization

•Gentle sterilization extends your instrument’s useful life span by
minimizing damage and reducing costly repair or replacement

•Environmentally friendly due to no water consumption, no toxic exhaust,
and low electric power needs
and low electric power needs
Back Forward

2
Back Forward

3
What is Plasma?
What is Plasma?
Plasma was named after an American physicist, Irving Langmuir.
When gas is applied with a greater energy it turns into
When gas is applied with a greater energy, it turns into
plasma with ionized elements, which are different from the prior phases. 
As the number of positive and negative ions compensates each other, the overall 
electrical property of plasma. is neutrali ed
electrical property of plasma. is neutralized
Plasma is known as the fourth state of matter and 99% of all the materials in 
space is known to be plasma. Solid has the lowest energy amongst all types of 
matter. When solid acquires heat (energy), it transforms to phases of liquid 
q ( gy) p q
and gas, sequentially. 

Back Forward

4
The application of Plasma in PlazMax
How is sterilization accomplished with H2O2?

The PlazMax utilizes the microbicidal (bacterium killing)  power of free radicals
The PlazMax utilizes the microbicidal (bacterium killing) power of free radicals
released by applying energy (heating liquid into gas via the vaporizer) to hydrogen peroxide.
By definition, a free radical is any atom (e.g. oxygen, nitrogen) with at least one
p , p p
unpaired electron in the outermost shell, and is capable of independent existence

Back Forward

5
The application of Plasma in PlazMax
The application of Plasma in PlazMax Hydroxyl (OHˉ) 
Free radical

A free radical is easily formed when a covalent
A free radical is easily formed when a covalent
bond between entities is broken (due to heating in our case) and one electron 
remains with each newly formed atom. When free radicals steal an electron from a 
surrounding compound or molecule a new free radical is formed in its place.
surrounding compound or molecule a new free radical is formed in its place.
In turn the newly formed radical then looks to return to its ground state by
stealing electrons with anti parallel spins from cellular structures or molecules.
Thus, the chain reaction continues and can be "thousand of events long”…
, g
Hydrogen peroxide is unique in that it can be converted to the 
Hydroxyl radical (OHˉ) or be catalyzed (with the “Burner” in our case) and excreted 
harmlessly as water and/or oxygen

Back Forward

6
The application of Plasma in PlazMax
The application of Plasma in PlazMax
The sterilizing agent in PlazMax is H2O2,
namely Hydrogen Peroxide
namely, Hydrogen Peroxide. 
Hydrogen peroxide was first isolated in 1818 by Louis Jacques Thénard ,
and H2O2 has long been widely used in the pharmaceutical industry. 
Hydrogen peroxide has been utilized for a decade or more in a wide variety of
Hydrogen peroxide has been utilized for a decade or more in a wide variety of 
facilities and has proven compatible with materials including sensitive electronics, 
microscopes, stainless steel, and HEPA filters. 

Back Forward

7
The application of Plasma in PlazMax
The application of Plasma in PlazMax

Applying energy causes alteration of OHˉ (Hydroxyl) from a bonder between the 
molecules in the liquid state into a free radical in the gas state   

Back Forward

8
Free
Radical

Example of an Electron “stealing” from the bacteria by the free radical

Bacteria Free
Radicals

After the alteration, the Bacteria ceases to function as Bacteria and becomes a free radical  of its own. As one, 
“steals” an Electron from another Bacteria, and alters it to a free radical and so forth. A chain reaction occurs 
effecting and destroying all Bacterium and thus, sterilizing the material.

Back Forward

9
The application of Plasma in PlazMax
The application of Plasma in PlazMax

The Plasma application is used in the Plazmax , expressed by utilizing the 
The Plasma application is used in the Plazmax expressed by utilizing the “Burner”
Burner
Its purpose is to “break” the H2O2 molecules down after the sterilizing phase.
The system consists of a high‐voltage generator, which applies high 
voltage to two electrodes located a few millimetres apart from each other
voltage to two electrodes located a few millimetres apart from each other
Inside the Burner.

When high voltage is applied to the electric plates facing each other,
g g pp p g ,
the H2O2 that passes through the space between the plates is ionized and
Plasma is generated and “breaks” down the gas into molecules of water and oxygen.

Back Forward

10
The application of Plasma in PlazMax
The application of Plasma in PlazMax

Vaporized Hydrogen Peroxide (H2O2) (A)

High Plasma
Voltage
o tage
(B)

(C)
GND
2H2O (Water) + O2 (Oxygen)

Back Forward

11
The application of Plasma in PlazMax
The application of Plasma in PlazMax

Back Forward

12
13
Cycle graphs
Cycle graphs
Average cycle temp. 50 C

Normal cycle

1000 mBar
Aeration
Preparation
Preparation 
Preparation 
For
For
Exhaustion diffusion
diffusion
Diffusion Plasma Diffusion Plasma
1 1 2 2

0
Vaporization

Back Forward

14
Cycle graphs
Cycle graphs
Average cycle temp. 50 C
Advanced cycle

1000 mBar
Aeration
Preparation
Preparation 
Preparation 
For
For
Exhaustion diffusion
diffusion
Diffusion Plasma Diffusion Plasma
1 1 2 2

0
Vaporization

Back Forward

15
Cycle graphs
y g p
Average cycle temp. 40 C
Endoscope cycle
Preparation 
For
diffusion

1000 mBar

Aeration

Preparation  Diffusion Plasma Diffusion Plasma


For 1 1 2 2
Exhaustion
diffusion

0
Vaporization

Back Forward

16
Cycle graphs
y g p
Leakage test

1000 mBar

Achieving deep vacuum Chamber is sealed 


and monitored for leaks

The line represents
the allowed limit of
pressure increase
p
0

Back Forward

17
Cycle graphs
y g p
Penetration test

1000 mBar

Aeration

Exhaustion Preparation  Diffusion Plasma


For
diffusion

0
Vaporization

Back Forward

18
Cycle graphs
y g p
Penetration & leak test combined

Diffusion

Preparation  Plasma Aeration


1000 mBar
Achieving deep Chamber is sealed  For
vacuum and monitored for leaks diffusion

0
Vaporization

Back Forward

19
Indicators & consumables

Biological indicator
The Biological indicator contains a defined quantity of 
inoculated spores in population of 1.9 x 10⁶, placed on a 
stainless steel disc. 
The BI (biological indicator) has a breakable glass ampoule
The BI (biological indicator)  has a breakable glass ampoule 
which includes a culture media to allow the growth of the 
spores. The disc and the ampoule are both placed inside a 
plastic tube.
The BI has a chemical indicator which will indicate if the BI
The BI has a chemical indicator , which will indicate if the BI 
was exposed to H2O2.
The BI is designed for hydrogen peroxide plasma sterilization 
processes, and utilizes Geobacillus Stearothermophilus spores.
As the spores are highly resistant sterilizing them serves as a
As the spores are highly  resistant, sterilizing them serves as a 
good indication of the PlazMax abilities. 

Back Forward

20
Indicators & consumables

Biological indicator

Cap

Plastic tube
Plastic tube

Breakable glass ampoule

Exposed to H2O2 Exposed to H2O2
and sterilized and  was not
Disc  inoculated with spores sterilized

Back Forward

21
Indicators & consumables
Incubators

The incubator is designed to contain the BI’s and supply the correct amount
Th i b t i d i dt t i th BI’ d l th t t
of heat needed for the spores to grow. 
We currently have 24 hours results using the standard Brown incubator,
During the coming weeks we should have the Smart well incubator
During the coming weeks we should have the Smart‐well incubator,
Which includes printer, touch screen and results gathered in 6‐10 hours.
Brown incubator
Smart Well incubator
Smart‐Well incubator

Back Forward

22
Indicators & consumables

Chemical indicator

The chemical indicator
Indicates that the sterilizing agent underwent exposure.
If it was exposed to H2O2 the color will change from purple to pink.
.

H2 O 2

Back Forward

23
Indicators & consumables

Chemical Tape
The chemical tape has two applications:
1. Indicates exposure to H2O2.
2. Tapes are used to close polypropylene Packaging materials

After Before

Back Forward

24
Indicators & consumables
Lumens PCD kit
Monitoring  the sterilization process can be performed with flexible lumen kits.
Th PCD (P
The PCD (Process Challenge Device) kit contains 4 flexible lumens;
Ch ll D i ) ki i 4 fl ibl l
2 with openings on both sides, 
And 2 with openings on one side:
10 meter Lumen:
Consists of a cylinder for positioning the indicators.
On each side of the cylinder is a lumen of 5 meters 
long with 1 mm diameter.
g
5 meter Lumen:
y p g
Consists of a cylinder for positioning the indicators . On
each side of the cylinder is a lumen of 2,5 meters long with 1 mm diameter.

Back Forward

25
Indicators & consumables
Lumens PCD kit

3 meter Lumen: 
Consists of a cylinder for positioning the indicators.
It is sealed on one end with a lumen of 3 meters
It is sealed on one end with a lumen of 3 meters
long with 1 mm diameter on the other side.
2 meter Lumen:
Consists of a cylinder for positioning the indicators .
It is sealed on one of the sides with a lumen of 2 meters long with
1 mm diameter on the other side.

Back Forward

26
Indicators & consumables
Lumens PCD kit

PCD’
PCD’s usage distribution according to cycles: 
di t ib ti di t l

Advanced Cycle 4 PCD’s (10 meters, 5 meters, 3 meters and 2 meters)
Normal Cycle
Normal Cycle 2 PCD’ss (5 meters and 2 meters)
2 PCD (5 meters and 2 meters)
Endoscopes Cycle 4 PCD’s (10 meters, 5 meters, 3 meters and 2 meters)
Penetration Cycle 2 PCD’s (5 meters and 2 meters) 

Back Forward

27
Indicators & consumables

Packaging materials

Tyvec Reels Tyvec Pouches

Back Forward

28
Pl M
PlazMax
Sterilizing agent consumption

H2O2 Bottle PlazMax


Supplier cat. No. No. of cycles
cat. no. Model

PLZ094-0001 AS040400 P50 12

PLZ094-0002 AS080400 P80 7

PLZ094-0003 AS100400 P110 5

PLZ094-0004 AS150400 P160 4

Back Forward

29
Pl M
PlazMax advantages in comparison to J&J
d i i J&J

Cycle duration:
1. Shorter cycle durations compared to both 100s & NX systems:
Cycle name
Cycle name Tuttnauer J&J
Tuttnauer J&J PlazMax NX system 100s
Normal Standard 26 Min. 28 Min. 51 Min.
Advanced Advanced 34 Min. 38 Min. 72 Min.
Not  25 Min.
Endoscope NA NA
Applicable

Back Forward

30
Pl M
PlazMax advantages in comparison to J&J
d i i J&J

Test cycles:
3 test cycles in comparison to none on J&J,
They claim to do it as part of the cycle‐ we reach deeper vacuum on a test cycle 
Than the operation cycles set points. 
h h i l i

Installation:
V
Very easy to install; only 1 phase 230v required, no special requirements needed 
t i t ll l 1 h 230 i d i l i t d d
from the client. 

Back Forward

31
Pl M
PlazMax advantages in comparison to J&J
d i i J&J

Endoscope cycle:
We offer a specific cycle for Endoscopes, J&J perform it on their advanced cycle
at a higher temperature and exposure to H2O2.

The short cycle time helps insure that no damaging method is applied

Add d l l (4th cycle):


Added value cycle (4 l )
We give the ability to suit a special needed program to specific instruments that will allow
The client to monitor them in accordance to the strict manufacturer limitations.

Back Forward

32
Pl M
PlazMax advantages in comparison to J&J
d i i J&J

Flexibility:
The machines can be changed from single door into a 2 door machine due to the 
Design applied and the upgrade can be performed on site.

Comfortable operation:
Unlike J&J the chamber is not sensitive to touch.
O
Operating is very easy and does not obliges
ti i dd t bli th li t t t k
the client to take special
i l
care in the load positioning.

Back Forward

33
Pl M
PlazMax advantages in comparison to J&J
d i i J&J

H2O2 decomposing method:
J&J perform the decomposing in the chamber. It is less reliable than our
system that force all particles through the burner and decompose.

Maintenance costs:
Unlike J&J that charge big amounts on maintenance kits and service agreements
O
Our maintenance system requires less parts and labor hours.
i t t i l t dl b h

Back Forward

34
Pl M
PlazMax advantages in comparison to J&J
d i i J&J

Complete CSSD solution:
Tuttnauer has the ability to provide full CSSD solution and turn key projects
Unlike J&J that has no other technologies.

Useable Volume:

IIn comparison to 100s model that has rounded chambers,
i t 100 d l th t h d d h b
resulting in a loss of about 30% in volume, we offer a rectangular 
chamber with full usable volume.
In comparison to NX system that has an inner casing we have also an advantage on
In comparison to NX system that has an inner casing, we have also an advantage on
of 15% usable volume. 

Back Forward

35
s

PlazMax
Materials that can be sterilized by PlazMax
Common materials to sterilize with Plasma:
•Stainless steel
•Polyethylene
•Teflon
Teflon
Metals
Stainless steel 300 series
Aluminum 6000 series
Titanium

Back Forward

36
Non-Metals

PlazMax
Materials that can be sterilized by PlazMax

Non‐Metals
Acryl nitrite butadiene styrene (ABS) Polycarbonate
Delrin Polyethylene terephthalate (PET)
EVA (Etheyl vinyl acetate) Polymethyl methacrylate (PMMA)
Fluorinated ethylene propylene (PTFE)
Fluorinated ethylene propylene (PTFE) Polyphenyl sulfone
Polyphenyl sulfone
Glass, USP type I borosilicate Polypropylene
Low density polyethylene Polystyrene
Latex Polysulfone
Low density polyethylene
Low density polyethylene Teflon (PTFE)
Teflon (PTFE)
Monel Polyurethane
Neoprene rubber (polychloroprene) Polyvinyl chloride (PVC)
Nylon Silicone
Phenolic resin
Phenolic resin

Back Forward

37
PlazMax

Materials that cannot be sterilized by the PlazMax

• Paper (but very small amounts of paper are possible. e.g: small barcode label).
• Absorbent materials (cellulose)
• Wood linen paper spongy materials
• Wood, linen, paper, spongy materials
• Liquids
• Powder
• Oil
• Oil
• Instruments with lumen, having less than 1 mm diameter and/or over 10 m length

Back Forward

38
39
40
41
42
43
44
45
46
47
48
49
50
Whatever your application may be, Tuttnauer makes the right
At l
Autoclave with
ith th
the right
i ht capacity
it !

Tuttnauer Europe b.v. Tuttnauer USA Co. Ltd.


Paardeweide 36 P.O.B. 7191 25 Power Drive
4800GD Breda, The Netherlands Hauppauge, NY,11788,USA
Tel: (31) 76 5423510, Tel: (631) 737 48 50, (800) 624 5836
Fax: (31) 76 5423540 Fax: (631) 737 07 20
Email: tuttnauer@tip.nl Email: info@tuttnauer.com
www.tuttnauer.com
98 www.tuttnauer.com

51

You might also like