Professional Documents
Culture Documents
6 Emission Spectroscopy - Manual
6 Emission Spectroscopy - Manual
6 Emission Spectroscopy - Manual
EMISSION SPECTROSCOPY
JHU INTRO. CHEM. LAB II
SPRING 2019
You will have one lab period to complete this experiment. There will be Chem21labs prelab and postlab
assignments due as shown on the schedule. You will need to upload some of your spectra to Chem21labs. There is a
prelab video on Blackboard that will help you understand the experiment and the analysis of the spectra.
When an atomic or molecular species absorbs energy, the energy has different effects. Some goes into translation,
making the species move faster. Some goes into vibration, causing the distance between atoms in the molecule to
change rapidly. Some goes into rotation, causing the atoms in the molecule to rotate around bonds. Some of the energy
goes into the electrons, causing them to jump to higher energy levels. The separation between these energy levels are
different for all atoms and molecules. Excited electrons then return to lower energy levels, emitting their excess energy
as photons. Given that different molecules have distinct energy levels, the pattern of light absorption and emission can
be used to identify species. Absorption spectroscopy is the measurement of the energy (or wavelength) that is absorbed
as electrons move to higher electronic energy levels. Emission spectroscopy is the measurement of the energy released
as the electrons fall back to lower levels. In this experiment, you will measure emission spectra from atoms and
molecules that are excited to higher electronic energy levels by heat, energy released in chemical reactions, and
voltages. You will see how emission spectroscopy can be used as a tool to identify species. In the following
experiment, you will measure absorption spectroscopy and see how it can be used to not only identify species but also
to determine their concentrations.
You experience the excitement of atomic emission when you watch fireworks with their bright displays of red,
yellow, green, and blue. Each color is characteristic of a metal. Although fireworks have a much longer history, it was
only in the 19th century that metal salts were added for color effects. And it has only been in the last century that this
process has been understood. In this experiment you will determine what colors are characteristic of specific metals
in flames and use this to determine some of the cations present in sparklers. You will also analyze the atomic emission
lines from hydrogen and relate your observations to atomic structure.
The human eye cannot separate mixtures of light into their component colors. A prism or a diffraction grating is
needed to do this. You will use a spectrophotometer (see diagram below) with a diffraction grating. A
spectrophotometer is a device that measures both the wavelength and intensity of light. A fiber optic cable will be used
to bring the light from the flame and sparkler to the entrance slit of the spectrophotometer.
Light is brought to the spectrophotometer slit by a fiber optic cable. It is separated into its component colors by
the diffraction grating. This causes different colors of light to be reflected at different angles back to the linear charge-
coupled device (CCD) detector, which records the intensity of the light at the different wavelengths. Our eyes measure
only one region of the spectrum, from ~400 nm to ~700 nm (namely the visible region). The SpectroVis Plus
spectrophotometers measure a slightly greater range, from 380 - 900 nm.
You will measure atomic spectra from a hydrogen lamp, from flames, and from sparklers. You will work with
your partner and a second group to record the flame spectra. Your TA will call each group of 4 up to measure the
atomic hydrogen lines and the sparklers. You may need to do Part 3 before finishing Parts 1 and 2. You will have one
lab period to complete this experiment. Each group of 4 students will need one Vernier LabQuest2 interface, one
SpectroVis Plus spectrophotometer, one fiber optic cable, and the appropriate power cord and cables from the
stockroom. You will get applicators to be used for the flames from your TA, since they will be pre-soaked in water to
make it less likely that they will catch on fire.
Objectives
• Use a spectrophotometer to measure emission spectra of atoms and molecules.
• Predict the spectral lines for hydrogen atoms and compare them with experimental values.
• Observe the difference between atomic and molecular spectra.
• Determine what species are present in unknowns using the spectra of known atoms and molecules.
Safety First!
Fire Safety
Keep your desk area clean. Only your lab notebook and manual should be on the lab bench. Backpacks and coats
should be in the front of the room.
Keep a large beaker with water in your work area that you can use to put out applicator fires.
If you have long hair, make sure it is tied back.
Make sure your hood is set up at least 18 inches above the flame (24 inches above the top of the Bunsen burner) to
avoid melting the plastic shields on the hood.
Be sure to not look directly into any bright lights generated by the flame test and sparklers.
Chemical Safety
Barium chloride (BaCl2) is a toxic salt, and can cause skin irritation, eye damage, and respiratory irritation.
Lithium chloride (LiCl2) is harmful if swallowed, and can also cause skin, eye, and respiratory irritation.
Copper(II) chloride (CuCl2) is toxic if swallowed or in contact with skin, and can produce toxic fumes when
exposed to heat. It can also cause skin irritation and serious eye damage.
All other salts can cause skin irritation, as well as serious eye irritation. Keep experiment setup under the fume
hood.
Always wear goggles, gloves, and apron during lab.
SPECTROPHOTOMETERS
The following section describes how to set up the SpectroVis spectrophotometer, The spectrophotometer plugs
into the Vernier LabQuest2 via the USB port on the upper left side of the unit. The rectangular box at the end of the
optical fiber is inserted into the SpectroVis, with the white arrow on the optical fiber on the same side as the white
arrow on the spectrophotometer.
Set up the LabQuest2 as in previous experiments and as described in Section III at the beginning of the manual.
The default for the spectrophotometer is absorbance (Abs), which will have to be changed into emission. To do this,
from the top toolbar select “Sensors” → “Change Units” → “USB: Spectrometer” → “Intensity”. Click on “Mode” to
change the settings on the mode to 200 ms for the “Sample time” and the “Samples to Average” should be set at 1.
The remainder of the default settings are correct. Record all the settings in your notebook. To avoid needing to reset
the SpectroVis any time you click on “File” → “New”, you should store each run in the same file using the file cabinet
icon. Record the run number, species, and observations (color and wavelengths of the major peaks) in your lab
notebook based on the spectra you observe. You will find it helpful to record this in a table in your notebook.
First familiarize yourself with the operation of the spectrometer. Your first scan should be of the room lights.
Press the green start arrow. The spectrometer will start taking spectra continuously at this time, but it won’t save any
spectra until you press the red stop button when you are satisfied with the spectra you observe. This is similar to using
a camera – you move the camera around until you get the best picture, but the picture isn’t saved until you press the
shutter button.
For the room light spectrum, point the yellow fiber optic cable tip at the room lights. Move the cable around,
observing the effect on the peak intensities. As you move the fiber tip closer to the lights, you may observe peaks that
look flattened at the top because they are off-scale. You need to move the tip further from the lights so that all peaks
are on-scale and not flattened. You should observe that the relative intensities of the lines remains the same and the
width of the lines and bands remains the same. However, the absolute intensity is only a function of how close the
cable is to the light. When you are satisfied with your spectrum, hit the red stop square. (Note the quickness of the
response – it only takes 200ms for the spectrophotometer to collect the light.) You should record the wavelengths of
the major peaks in the room light spectrum in your notebook so that you will know if you are observing room lights
instead of the desired spectra in the rest of the experiment. This scan should be stored and later e-mailed to all the
members of your group because you will need to compare it to your spectra to see if any bands can be identified as
due to the room lights. You will be able to append additional spectra without needing to open a new file, although you
will need to periodically save your file as you go along. You will also be able to look at this spectrum any time during
your experiment using the file cabinet icon and choosing the correct run so you can see if room lights are a major
contaminant in your spectra.
Spectrophotometer accuracy: The LabQuest records wavelengths in the files as xxx.x or xxx.xx nm. However, the
resolution of the spectrometers we use are less than that. For the Spectrovis Plus spectrometer (used for the metal
flames and the sparkler) the wavelength resolution is wavelength dependent. It ranges from 3 to 7 nm, so, for this
experiment, you should consider the absolute accuracy of the wavelength to be approximately ± 5 nm. However, the
repeatability is higher than that, so the same instrument can be expected to measure the peak from the same molecule
to within 1 or 2 nm. For the Red Tide spectrometer (used for the hydrogen spectrum) the resolution is ± 1 nm.
PROCEDURE
There is a technique video on Blackboard on flame spectra that will help you with this procedure.
Safety Concerns for Part 1:
• Some of the salts are toxic. Wear gloves and goggles at all times. Set up the experiment under a hood.
However, the hood needs to be at least 18 inches above the flame (24 inches above the top of the Bunsen
burner) to avoid melting the plastic shields on the hood.
• To prevent fires: Keep your desk area clean. Only your lab manual and your lab notebook should be on the
lab bench. Backpacks and coats should be in the front of the room. Keep a large beaker with water in your
work area that you can use to put out applicator fires.
Note: It is important to avoid contaminating your samples. Be sure all glassware that you use is clean and labeled.
Keep unused applicators separated from the used ones. Do not reuse applicators on different salts. You will, however,
reuse the same applicator on the same salt.
• Make sure you have all of the relevant equipment on hand.
• Label the 6 small test tubes in the rack NaCl, LiCl, KCl, SrCl2, BaCl2, and CuCl2. Keep them as far apart
from each other as possible in the test tube rack to avoid cross-contamination.
• Get about 0.15 g of each compound using the labeled spatula – NOT your own – to
1-2 mm of Salt
avoid contaminating the chemicals. You only need enough to fill the bottom of the
small test tubes with 1 or 2 mm of solid. It should look as follows:
• Adjust the air intake on your Bunsen burner so that you have an inner cone as shown
in the diagram below.
Inner Cone
where is the frequency of the transition, which is equal to c/, where c is the speed of light and is the wavelength
of the light. The value 3.28805 x 1015 s-1 comes from the best available values of the fundamental constants included
in Equation 3 and is the value you should use in your calculations in Chem21labs. As you can see, Equation 5 is the
same as Equation 2, with the empirical Rydberg constant replaced with a value derived from fundamental constants.
You will compare your experimental values of the frequencies to those calculated using this theoretical model. The
unit s-1 can also be called the hertz, abbreviated Hz.
n=∞ An energy level diagram showing the emission
n=6 lines for hydrogen atoms is shown left. This diagram
n=5 shows the energy levels predicted by the Bohr model
n=4 and the transitions between them. The levels are labeled
n=3
Paschen Series
with the quantum number n, which ranges from the
lowest energy level of 1 to infinity, which would be for
Balmer Series
n=2 the separated electron and proton.
Since emission lines are observed, the initial levels,
Energy →
ASSIGNMENT
Complete the Chem21labs assignment. You will also need to upload a total of four spectra including:
• A spectrum from two of your metal salts,
• the hydrogen lamp,
• and the sparkler.
You will be assigned specific metal salts on Chem21labs when you do the assignment. Students will randomly
be assigned one of the spectra from the sodium, lithium, or potassium salts and one of the spectra from the copper,
barium or strontium salts. Format your spectra as lines, not showing individual data points. This is an option in the
scatter plots on Excel. An example spectrum is shown on Blackboard so that you can see an example of good
formatting.
Each spectrum should be labeled with a title that contains the name of what you were measuring. You should
identify any extraneous lines due to sodium, potassium, or room lights on the spectra. You should also identify the
lines on your sparkler spectrum due to sodium, lithium, potassium, strontium, barium, and/or copper salts.
All students should make their own spectra. There will be a significant loss of points if the spectra are copied!