AK 96 - Technical Data 4:1
AK 96 - Technical Data 4:1
AK 96 - Technical data
Contents
Double Needle
Single Needle
Blood pressure
Venous Pressure
The alarm limits below can be preset. The value put in brackets and in
italics is the default value.
Temperature
Flow rate
Degassing
Ultrafiltration control
Note
• Besides the ultrafiltration, patient weight change during treatment is
also affected by other factors. These include factors such as fluid
intake, food intake, perspiration, drug administration, infusion priming
and rinse-back volumes amongst others. In addition, precise pre-
and post-treatment weight are critical for adequate assessment of
the ultrafiltration during the treatment. If these measurements are not
accurate a discrepancy between the ultrafiltration achieved during
the treatment and the changes in body weight will occur.
Note
Profiling
UF-rate 0.0 to 4.0 l/h
Na+, Acetate/Bicarbonate 130 to 160 mmol/l
mode
HCO3-, Bicarbonate mode 20 to 40 mmol/l
Diascan
Clearance measuring is only available if the optional Diascan is installed.
Concentration of 37 % formaldehyde
disinfectant
Concentration in machine 3.7 %; i.e. diluted 1 + 9 (1:10)
Volume Approx. 275 ml (with UFD-kit approx.
290 ml)
Contact time between 20 minutes
treatments
Contact time overnight Minimum 6h dwell time
or when not in use
(recommended with
UFD-kit)
Total time 119 minutes (with UFD-kit 150 minutes)
including 20 min contact time
Note
• The temperatures are verified at nominal values for the mains voltage
and at 20 °C ambient temperature.
Heat disinfection
Heat 33 37
Without UFD Heat Citric acid 20 % 50 55
Heat CleanCart 46 51
Rinse/Drain 10 min.
Drain 4 min.
External equipment
Extern alarm
Max voltage 24 V AC or DC
Max current 100 mA AC or DC
RS-232C
RS-422/Current loop
Typ current 20 mA DC
Max current 50 mA DC
Battery Back-up
Battery back-up of power 24 V, 7.0 Ah
supply
Battery type Sealed Lead Acid, 2 x 12 V
Running time >30 minutes
Fuse 12 AT
Arterial pressure
Air Detection
Detection method Ultrasonic detector placed at the venous
drip chamber. The detector has a two
channel structure and the function of the
detector is tested at the function test made
by the microcomputers.
Drip chamber size Diameter 22 mm
Sensitivity Bubbles larger than 1 μl will be trapped by
the drip chamber. An alarm will be issued
if the blood level falls below the middle of
the air detector.
Conductivity
Ultrafiltration Supervision
TMP TMP is defined as the difference, Pb out -
Pd out, where Pb out is the venous pressure
and Pd out is the pressure measured in the
dialysis fluid, where it enters the machine
after the dialyzer.
Alarm limits -100 to 500 mmHg
Accuracy ±10 mmHg or ±5 %, whichever is largest
(within range ±500 mmHg)
pH supervision
pH supervision is only available if the optional pH kit is installed.
Infusion stand
Maximum total load 2 kg
Polymers
Silicon rubber
PVC (Polyvinylchloride)
PEEK (Polyetherketone)
PEX (Polyethylene)
PP (Polypropylene)
PSU (Polysulphone)
PVDF (Polyvinylidene fluoride)
PTFE (Polytetrafluoro ethylene)
Metals
Stainless steel SS2343
Stainless steel SS2353
Stainless steel SS2562
Titanium
Platinum
Others
Carbon
Ceramic, Steatite 221
Ceramic, Aluminium oxide (Al2O3)
Glass
Operation
If condensation occur when moving the equipment between locations
with different temperatures and high relative humidity ( e.g. outdoor and
indoor locations ), the inside of the equipment shall be allowed to dry
before switching on the equipment.
Electromagnetic environment
The AK 96 is intended for use in the electromagnetic environment
specified below. The customer or the user of the AK 96 should assure
that it is used in such an environment.
Note
• UT is the a.c. mains voltage prior to application of the test level.
Note
Note
• At 80 MHz and 800 MHz, the higher frequency range applies.
• These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects and people.
Note
2 Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic
site survey should be considered. If the measured field strength in the location in
which the AK 96 dialysis machine is used exceeds the applicable RF compliance level
above, the AK 96 dialysis machine should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the AK 96 dialysis machine.
3 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than
3 V/m.
100 12 12 23
Rated maximum - - ⎡7⎤
d=⎢ ⎥ P
output power of ⎣ 30 ⎦
mobile phone
2W - - 0.33
GSM/3G
For transmitters rated at a maximum output power not listed above, the recommended separation distance d
in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note
• At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.