Product Conformity Certificate: Model APMA 370 CO Analyser
Product Conformity Certificate: Model APMA 370 CO Analyser
Certification Ranges :
3
CO 0 to 60 mg/m
0 to 100 mg/m3
This certificate may only be reproduced in its entirety and without change
Registered Office: Rake Lane, Eccleston, Chester, UK CH4 9JN
To authenticate the validity of this certificate please visit www.siracertification.com/mcerts
Page 1 of 5
Approved Site Application
On the basis of these tests this certificate is valid when the instrument is used on urban air
quality and similar applications.
Any potential user should ensure, in consultation with the manufacturer, that the air monitoring
system is suitable for the process on which it will be installed.
Basis of Certification
This certification is based on the following Test Report(s) and on Sira’s assessment and
ongoing surveillance of the product and the manufacturing process:
Product Certified
This certificate applies to all instruments fitted with software version P1000878001C onwards
(serial number 432866-10031 onwards).
This certificate may only be reproduced in its entirety and without change
To authenticate the validity of this certificate please visit www.siracertification.com/mcerts
Page 2 of 5
Certified Performance
The instrument was evaluated for use under the following conditions:
0.08 µmol/mol/K
This certificate may only be reproduced in its entirety and without change
To authenticate the validity of this certificate please visit www.siracertification.com/mcerts
Page 3 of 5
Test Results expressed as % of measured Other results MCERTS
value specification
<0.5 <1 <2 <5
Interference by NO (at concentration of
0.35 µmol/mol <0.5 µmol/mol
1µmol/mol)
Interference by CO2 (at concentration of
0.16 µmol/mol <0.5 µmol/mol
500µmol/mol)
Interference by N2O (at concentration of
50nmol/mol) 0.16 µmol/mol <0.5 µmol/mol
Short term zero drift (over 12h) 0.04 µmol/mol <0.1 µmol/mol
Short term span drift (over 12h) 0.06 µmol/mol <0.6 µmol/mol
Long term zero drift (over 3 months) Note 1 0.17 µmol/mol <0.5 µmol/mol
Long term span drift (over 3 months) Note 1 4.5 <5% of the max of
certification range
Note 1: Field test: The field test was performed at an urban site for 3 months.
This certificate may only be reproduced in its entirety and without change
To authenticate the validity of this certificate please visit www.siracertification.com/mcerts
Page 4 of 5
Description
The APMA-370 ambient air analyser measures CO using a combination of cross flow
modulation and the non-dispersive infrared absorption (NDIR) principle and allows continuous
measurement.
The APMA-370 uses a solenoid valve to alternately inject fixed amounts of sample and
reference gas into the measurement cell. If the same gas is used for both sample and
reference gas (e.g. zero gas) no modulation signal is generated. This allows the measurement
of low levels of the sample without zero drift. The cross flow technique reduces the number of
components in the system and improves stability and reliability.
The analyser also has internal data storage capacity for three different mean value periods.
General Notes
1. This certificate is based upon the equipment tested. The Manufacturer is responsible for
ensuring that on-going production complies with the standard(s) and performance
criteria defined in this Certificate. The Manufacturer is required to maintain an approved
quality management system controlling the manufacture of the certified product. Both
the product and the quality management system shall be subject to regular surveillance
according to ‘Regulations Applicable to the Holders of Sira Certificates’. The design of
the product certified is defined in the Sira Design Schedule for certificate
No. Sira MC 070089/03.
2. If certified product is found not to comply, Sira Certification Service should be notified
immediately at the address shown on this certificate.
3. The Certification Marks that can be applied to the product or used in publicity material
are defined in ‘Regulations Applicable to the Holders of Sira Certificates’.
4. This document remains the property of Sira and shall be returned when requested by
the company.
This certificate may only be reproduced in its entirety and without change
To authenticate the validity of this certificate please visit www.siracertification.com/mcerts
Page 5 of 5