COMPANY CONFIDENTIAL

Category I MSA Instrument Division Product Group 0700 ATO Order Form

Ultima X Series Gas Monitors


A-ULTIMAX __ __ __ __ __ __ __ __ __ __ __ __ __ __
1 2 3 4 5 6 7 8 9 10 11 12 13 14

1 Model Type 5 Sensor Mounting Style


A Nema 4X Plastic Enclosure + S = Single Condulet
E 316 Stainless Steel Enclosure * D = Dual Condulet (Remote)

6 Gas Sampling
2 Gas Type 0 = Diffusion
11 Carbon Monoxide 0-100 PPM
12 Carbon Monoxide 0-500 PPM
7 LED / Relays Options
13 Oxygen 0- 10% 0 = No LED’s and No Relays
14 Oxygen 0-25%
1 = LED’s and No Relays
15 Hydrogen Sulfide 0-10 PPM
16 Hydrogen Sulfide 0-50 PPM 2 = Relays and No LED’s
17 Hydrogen Sulfide 0-100 PPM 3 = LED’s and Relays
# 18 Chlorine 0-5 PPM
8 Language
19 Sulfur Dioxide 0-25 PPM
20 Nitric Oxide 0-100 PPM 0 = English
# 21 Nitrogen Dioxide 0-10 PPM
22 Hydrogen Cyanide 0-50 PPM 9 Power Supply
# 23 Hydrogen Chloride 0-50 PPM 0 = None
# 24 Chlorine Dioxide 0-3 PPM
31 Combustible gas 0-100% LEL - Natural Gas & H2 10 Gas Type – Second Sensor
32 Combustible gas 0-100% LEL – Petroleum Vapors 0 = None
33 Combustible gas 0-100% LEL – Solvents
38 IR combustible 0-100% LEL – Methane 11 Gas Type – Third Sensor
39 IR combustible 0-100% LEL – Non-Methane 0 = None
41 Phosphine 0-2 PPM
42 Arsine 0-2 PPM 12 Calibration Options
43 Silane 0-25 PPM
0 = None
44 Germane 0-3 PPM
1 = Calibrator (P/N 809997)
45 Diborane 0-50 PPM
2 = Controller (P/N 809086)
# 46 Bromine 0-5 PPM
3 = Calibrator and Controller
# 47 Flourine 0-10 PPM
# 48 Ammonia 0–50 PPM Note: Calibrator not Intrinsically Safe in Europe.
49 Hydrogen 0-1000 PPM 13 Installation Hardware
# ^ 50 Ethylene Oxide 0-10 PPM 0 = None
1 = Strap assembly (XE only)
3 Sensor Output
B Milliamp Output 4-20, 2 wire toxic 3 = NEMA 4X installation kit (XA only)
And oxygen (without LED’s or relays)
14 Custom Features
E Milliamp Output 4-20, 3 wire toxic
and oxygen sensors (with LED’s and/or relays) 0 = None
AND all combustible sensors C = Custom material
T = Tagging – Aluminum tagging
4 Enclosure Type - Stainless Steel tagging
0 = GP Enclosure
Key: + Explosion-proof single condulet versions are approved for:
+ * 1 = XP Enclosure – cUL approved: NPT cUL Class I, Div. 1, Groups B, C, D
threads (CSA combustible performance pending)
* Explosion proof dual condulet versions are approved for:
cUL Class I, Div. 1, Groups A, B, C, D
(CSA combustible performance pending)
# Not available with explosion-proof approvals
^ 10 ppm SO2 cal gas = 4 ppm EtO
^ 10 ppm SO2 cal gas = 4 ppm Eto
COMPANY CONFIDENTIAL
Category I MSA Instrument Division Product Group 0700 ATO Order Form

Accessories
809997 Calibrator
809086 Controller
697281 Strap Assembly (XE only)
10041893 NEMA 4X Installation kit (XA only)
10041866 Flow cap – Ultima XE
10028904 Sensorgard – Ultima XE/XA
10041265 Sensorgard – Ultima XIR
10042600 Flow Cap – Ultima XIR
10041533 Calibration Cap – Ultima XIR
10036101 Ultima X Series Manual (printed version)
10042446 Ultima X Series product CD
10042803 Ultima X Controller Upgrade Kit

SELECTION GUIDE FOR


ULTIMA XIR COMBUSTIBLE
CATEGORY 38: Methane Calibration

Controller Cal cylinder cal


Code cylinder P/N Span Value
Methane 1 2.5% Methane 10028032 50% LEL

CATEGORY 39: Non-Methane Calibration

Controller Cal cylinder cal


Code cylinder P/N Span Value
Propane 2 0.6% Propane 10028032 29% LEL
Ethane 3 0.6% Propane 10028034 25% LEL
Butane 4 0.6% Propane 10028034 28% LEL
Pentane 5 0.6% Propane 10028034 33% LEL
Hexane 6 0.6% Propane 10028034 41% LEL
Cyclopentane 7 0.6% Propane 10028034 30% LEL
Ethylene 8 0.1% Propane 711054 28% LEL

For sensing multiple gases always calibrate for the least sensitive
gas or vapor expected to be measured (highest response factor
within a category).

All other combustible gas span values available upon request.