User Guide
P/N: 19196-U
Rev. 6/23/2016
HYDRAFACIAL® TOWER User Guide
Contact Info:
2
CONTENTS
HYDRAFACIAL® TOWER
Contents
User Guide
3
BEFORE YOU START
HYDRAFACIAL® TOWER User Guide
INTRODUCTION
Thank you for purchasing the HydraFacial® System, a revolutionary
breakthrough in aesthetic technology! The HydraFacial® is the next
generation of exfoliation and skin rejuvenation done simultaneously.
The process removes dead skin cells and impurities while simultane-
ously Vortex-Fusing™ the underlying new skin with cleansing, hydrat-
ing and moisturizing skin solutions and antioxidants. The procedure is
soothing, refreshing, non-irritating and very effective. You can have the
utmost confidence in this leading-edge system manufactured in the USA
by Edge Systems LLC.
Operating Conditions:
Temperature: Normal (10-40º C)
Relative Humidty: 30-75% RH
Altitude: 700-1060 hPa
4
BEFORE YOU START
HYDRAFACIAL® TOWER User Guide
SAFETY INFORMATION
PRODUCT MARKING SYMBOLS
Symbol Description
Attention: Refer to accompanying documentation.
Aufmerksamkeit: Beziehen Sie sich begleitend auf Dokumentation.
Attention : Consultez la documentation d’accompagnement.
Separate collection for electronic.
Trennen Sie Sammlung für elektronisch.
Collecte sélective pour l’électronique.
Direct Current
Gleichstrom
Courant continu
Alternating Current
Wechselstrom
Courant alternatif
ON (Power)
EIN (Stromversorgung)
ALLUME (Alimentation)
OFF (Power)
AUS (Stromversorgung)
ETEINT (Alimentation)
STANDBY (Power)
STANDBY (Stromversorgung)
PAUSE (Alimentation)
Operating Instructions
Bedienungsanleitung
Mode d'emploi
Fuse
Sicherung
Fusible
Pushing Prohibited
Drücken ist untersagt
Interdiction de pousser
Certified by CSA Group
Zertifiziert von der CSA-Gruppe
Certifié par CSA Group
Connect to supply mains with protective earth (ground)
Zum Versorgungsnetz durch eine Schutzerdung anschließen
Connectez au secteur, avec protection mise à la terre
5
BEFORE YOU START
HYDRAFACIAL® TOWER User Guide
Symbol Description
Calls attention to a procedure, practice, or condition that could
possibly cause bodily injury or death.
Ruft Aufmerksamkeit zu einem Verfahren, Praxis, oder Bedingung,
die vielleicht körperliche Verletzung oder Tod verursachen könnte.
Une attention particulière doit être portée à une procédure, pratique
ou condition qui pourrait causer des blessures corporelles ou la
mort.
Calls attention to a procedure, practice, or condition that could
possibly cause damage to equipment or permanent loss of data.
Ruft Aufmerksamkeit zu einem Verfahren, Praxis, oder Bedingung,
die vielleicht Schaden an Ausrüstungen oder permanentem Verlust
der Daten verursachen könnte.
Une attention particulière doit être portée à une procédure, pratique
ou condition qui pourrait causer des dommages à l’appareil ou la
perte permanente de données.
Do Not Re-use
Nicht wieder verwenden
Ne pas réutiliser
Manufactured by
Hergestellt von
Fabriqué par
Keep dry
Bleib trocken
Garder au sec
6
BEFORE YOU START
HYDRAFACIAL® TOWER User Guide
SAFETY GUIDELINES
1. Ensure that all operators of the HydraFacial® System are trained and
licensed as required by the state or country. Do not operate the unit
before being trained. For any questions regarding training, call your
salesperson or Edge Systems.
3. Always use a clean tip for each procedure. The HydroPeel® tips are
for single use only. Use for more than one procedure may result in
infection.
4. Reusing contaminated skin solution can cause harm to the client and
will void all warranties.
5. Skin solutions should not exceed the “Use By” date on the bottle.
Should any products (skin solutions) settle, please contact your dis-
tributor or Edge Systems.
7
BEFORE YOU START
HYDRAFACIAL® TOWER User Guide
11. The client should have a minimum sunscreen SPF 30 applied after
the procedure and should use sunscreen on an ongoing basis. Use of
Edge Systems’ UV Smart™ Daily sunscreen is highly recommended.
12. Empty the waste container after each service according to your
waste handling protocol. Follow the system cleaning instructions in
this guide to clean your machine and handpiece after each
procedure.
13. Only Edge Systems skin solutions should be suctioned through the
HydraFacial® handpiece. Use of other, non-approved skin solutions
will void the warranty and clog the machine. Always use solution
when performing abrasion.
L’appareil n’est pas destiné à être utilisé en présence d’un mélange anes-
thésique inflammable avec de l’air, de l’oxygène ou de l’oxyde nitreux.
8
BEFORE YOU START
HYDRAFACIAL® TOWER User Guide
DO NOT suction liquids through the dry port for crystal-free procedure
located on the side of the unit. This will damage the unit and will void
the warranty.
9
BEFORE YOU START
HYDRAFACIAL® TOWER User Guide
10
BEFORE YOU START
HYDRAFACIAL® TOWER User Guide
11
GETTING STARTED
HYDRAFACIAL® TOWER User Guide
GETTING STARTED
Unpack the unit, accessories and all documents. Retain the original
packaging in case you need to ship the unit for any reason. Verify that
all appropriate items have been received.
The unit is large & heavy and should not be lifted alone.
Das Gerät ist groß und schwer und sollte nicht ohne Hilfe aufgehoben
werden.
L'appareil est gros et lourd, et ne doit pas être soulevé par une seule
personne.
12
DGETTING STARTED
HYDRAFACIAL® TOWER User Guide
Package Contents
Console Unit
Wet (HydraFacial®) Handpiece Assembly
Tool Holder
HydroPeel® Tips
Skin Solutions
Puncture Tool
Waste Container Assembly
Cleaning Cap
Profit Pack
Success Marketing Kit
Options:
Lymphatic Drainage Therapy Glass
Handpiece Cups
13
GETTING STARTED
HYDRAFACIAL® TOWER User Guide
Device Components
14
DGETTING STARTED
HYDRAFACIAL® TOWER User Guide
INSTALLATION
TOOL HOLDER
WASTE CONTAINER
HydraFacial® HANDPIECE
15
GETTING STARTED
HYDRAFACIAL® TOWER User Guide
5. Insert the gray cartridge into the handpiece. Align the arrow on the
cartridge with the unlock symbol on the handpiece. Push firmly until
it snaps into place, then twist the cartridge up until the arrow aligns
with the lock symbol.
SPECIALTY VIAL:
1. To access specialty vials, remove the gray cartridge from the hand-
piece by twisting until the arrow aligns with the unlock symbol, then
pull the cartridge straight out.
16
DGETTING STARTED
HYDRAFACIAL® TOWER User Guide
3. Insert the vial into the handpiece. Push firmly until it snaps into place.
Note: Keep all the toggle switches off (see below) while using spe-
cialty vials.
4. To re-access the skin solution bottles, remove the specialty vial and
insert the gray cartridge into the handpiece. Push firmly until it snaps
into place, then twist the cartridge up into the locked position.
SKIN SOLUTION BOTTLES
1. Open the front door. Remove the cap of the skin solution bottle and
use the puncture tool to pierce the seal inside the bottle.
2. Press and hold each manifold button as you insert each bottle into
the manifold, and then release the button.
Note: The Rinseaway™ System Cleaning Solution should be installed
in the manifold with the gray toggle switch labeled “SYSTEM RINSE”.
3. The toggle switches control which skin solution bottle is being used.
When a switch is in the ON position, fluid will be drawn from that
bottle. When switched to the OFF position, no fluid will be drawn
from the corresponding bottle.
Note: Always keep the “SYSTEM RINSE” toggle switch (gray) off while
performing procedures.
17
GETTING STARTED
HYDRAFACIAL® TOWER User Guide
OPTIONAL
If you have purchased the optional kits, please see below.
1. Place the diamond pad inside the tip and push it in.
2. Place the tip on the HydraFacial® handpiece and press-in until it is
fully seated.
diamond tip
pad
1. Plug the metal tube coupling attached to the Dry Handpiece Tubing
into the port on the side of the machine.
2. Connect the plastic coupling on the other end of the tubing into the
Lymphatic Drainage Therapy handpiece. To connect the tubing, sim-
ply press the two halves together until you hear them click into place.
To disconnect the coupling halves, press the thumb latch button and
gently pull the two halves apart.
3. Place the handpiece(s) in the handpiece holder.
4. Place the tips and accessories on the holder when they are not in use.
18
DGETTING STARTED
HYDRAFACIAL® TOWER User Guide
Red and Blue LED Light Therapy mode (optional): This accessory is
not currently available for EU countries.
1. Plug in the light handpieces into the ports on the other side of the
unit. The connectors must be aligned properly. Line up the notch on
the cable end connector of the handpiece with the tab on the port.
Note: The ports are color-coded (2 blue ports above; 2 red ports be-
low). Make sure the handpieces are connected into the correct ports.
2. Attach the LED/Lymphatic universal tool holder.
3. Insert the cable end connector and twist its lock ring clockwise to
lock the connector into the port. You should feel the connector click
as it locks into place.
4. Place the light handpieces on the handpiece holder when not in use.
19
OPERATING INSTRUCTIONS
HYDRAFACIAL® TOWER User Guide
OPERATION
TURN ON THE SYSTEM
20
SOPERATING INSTRUCTIONS
HYDRAFACIAL® TOWER User Guide
TOUCHSCREEN OPERATION
21
OPERATING INSTRUCTIONS
HYDRAFACIAL® TOWER User Guide
RED LED
BLUE LED
TIMER OPERATION
22
S SYSTEM CLEANING
HYDRAFACIAL® TOWER User Guide
WEEKLY CLEANING
23
SYSTEM CLEANING
HYDRAFACIAL® TOWER User Guide
HydraFacial® Handpiece
The HydraFacial® handpiece tubing chambers must be cleaned and dis-
infected using the cleaning brush every week to prevent any bacterial
contamination.
Vacuum
Opening
24
GSYSTEM CLEANING
HYDRAFACIAL® TOWER User Guide
4. Using a compressed air canister with the straw attached, blow air
first into the vacuum (smaller opening) on the face of the handpiece,
and then into the opening in the gray cartridge on the back of the
handpiece.
5. When the brush and air cleaning are completed, make sure all the
blue toggle switches are in the OFF position. Turn the gray toggle
switch “SYSTEM RINSE” up to the ON position and the vacuum set-
ting on “26”. Press on the cleaning cap and flush through the hand-
piece and tubing for 1 minute.
25
SYSTEM CLEANING
HYDRAFACIAL® TOWER User Guide
Other
WASTE CONTAINER
1. Wash, dry, and wipe/spray germicide over waste canister and lid.
Leave open overnight to air-dry.
Note: Always remove the waste container before transporting the
unit.
TOUCHSCREEN
26
GSYSTEM CLEANING
HYDRAFACIAL® TOWER User Guide
Options
WET DIAMOND AGGRESSIVE ABRASION (MEDICAL USE ONLY)
3. Place abrasive pad under running water and brush away all the debris
with the included brush. Rinse away any debris very thoroughly.
5. For maximum sterility, abrasive pads, tip, and pad holder are
autoclavable.
27
SYSTEM CLEANING
HYDRAFACIAL® TOWER User Guide
1. Clean the equipment by wiping the handpiece and focusing cone with
a germicide or antibacterial.
Note: Check with the Board of Cosmetology for approved products in
your state.
28
G TROUBLESHOOTING
HYDRAFACIAL® TOWER User Guide
TROUBLESHOOTING
Problem Possible Cause Solution
Make sure power cord is properly
Power cord connected to back of the unit and to
a working wall outlet.
Unplug unit from main wall out-
No Power let prior to this. Use a small flat
head screwdriver to remove the fuse
Fuses holder box located in between the
ON/OFF switch and power cord, and
check that fuses are in good working
conditions (not blown).
If the main power is “ON” and the
touchscreen is blank, press the
Blank screen / No
power button located on the right
video signal
hand side of touchscreen. Check
video cable connection in the rear.
Touchscreen
No image, or
inverted image
on the screen Call your local representative or
Edge Systems.
Screen does not
respond to touch
Toggle switch Make sure the switch is on.
Make sure your skin solution bottle
Empty bottle
is not empty.
Make sure that your HydraFacial®
handpiece is properly connected to
Improper the side of the skin solution station
connection block, and to the waste container.
Also, make sure that the lid is firmly
attached to the waste container.
Low or No
Fluid Flow Make sure the tip is firmly installed
Tip is improperly onto the handpiece. Angled tips
installed should face downwards in order to
facilitate flow.
Make sure the flow control
valve on the handpiece
Flow control is open. To increase fluid
valve is closed flow, turn the flow
control valve towards
the + sign.
29
TROUBLESHOOTING
HYDRAFACIAL® TOWER User Guide
30
TROUBLESHOOTING
HYDRAFACIAL® TOWER User Guide
OPTIONS
Note: Circuit diagrams & component part lists are available upon
request for service technicians.
31
LIMITED WARRANTY
HYDRAFACIAL® TOWER User Guide
LIMITED WARRANTY
Form of Product Warranty
LIMITED WARRANTY
(HydraFacial® SYSTEM)
What is Covered:
The warranty for the HydraFacial® System (the “System”) covers defects
in material or workmanship in the System.
Who is Covered:
This limited warranty applies and extends only to the original Customer
purchasing the System from Edge Systems in the country where the sale
occurred. This limited warranty is non-transferable and non-assignable
by the original Customer. Edge Systems shall have no obligations under
this limited warranty in the event that an original Customer attempts
an unpermitted assignment of the original Customer’s rights under this
limited warranty. Notwithstanding the foregoing, Edge Systems agrees
to extend the benefits of this limited warranty to a leasing company pro-
viding financing to the original Customer for the System, provided the
System is leased back to and operated by the original Customer and any
such leasing company has no greater rights than the original Customer.
How Long:
The System warranty runs for a period of one (1) year from the date of
delivery of the System to the original Customer. We warrant any replace-
ment part for the balance of the original warranty period. Extended war-
ranties are available.
Service Loaner:
During the first year of warranty Edge Systems, to the extent available,
will provide to Customer a loaner System (the “Loaner”) upon determi-
nation that the failure is a warrantied event, and following execution
of Edge Systems’ Loaner Agreement. Loaner systems are not available
outside the United States.
32
LIMITED WARRANTY
HYDRAFACIAL® TOWER User Guide
33
APPENDICES
HYDRAFACIAL® TOWER User Guide
Name: _______________________________________________________________
Phone: (______)_______________________________________________________
Address: ____________________________________________________________
City: ___________________________ State: ______ Zip: __________________
Email: ______________________________________________________________
Relative Contraindications
Anticoagulants therapy (use lower settings)
Very thin skin
Other Aesthetic Treatments: Botox: wait 5-7 days; Fillers: wait
7-10 days; Peels: wait 30 days
Laser Treatments: wait until lesions heal & swelling & redness is
resolved
Other Concerns
Keloids: avoid direct contact
Rosacea, telangiectasia (use lower vacuum)
Unrealistic expectations
By signing below, I certify that I have read and fully understood the con-
tents of this consent form, and that the information I provided above are
complete, accurate, and up-to-date to my knowledge.
35
APPENDICES
HYDRAFACIAL® TOWER User Guide
APPENDIX II:
Certificates
36
APPENDICES
HYDRAFACIAL® TOWER User Guide
37
APPENDICES
HYDRAFACIAL® TOWER User Guide
38
APPENDICES
HYDRAFACIAL® TOWER User Guide
39
2277 Redondo Ave, Signal Hill,
CA 90755, USA
Phone: (800) 603-4996
(562) 597-0102
Fax: (562) 597-0148
www.edgeforlife.com