Electromagnetic Flowmeter FSM4000: Data Sheet
Electromagnetic Flowmeter FSM4000: Data Sheet
D184S073U02
FSM4000
Pos: 1 /Titelblätter / Copyright/DB/Durchfluss/FSM4000 @ 3\mod_1157697105687_3101.doc @ 40413
Function
- Electromagnetic flowmeters can be used to
accurately measure the flowrate of liquids, slurries,
pastes and sludges with an electrical conductivity
of at least 20 µS/cm (opt. 5/0.5 µS/cm).
Applications
- The system is specially designed to measure
liquids in the pulp & paper and food & beverage
industries. It can be used to measure fast
changing processes, two phase liquids, continuous
and pulsating flows (piston pump application).
Benefits
- Digital signal processing (DSP) and zero stability
ensure long-term stability and accuracy in both
flow directions.
- Enhanced diagnostic functions and monitoring of
coil and electrode circuits as well as magnetic field
for increased operational stability. Verification via
fingerprint.
Key features
- Flowmeter sensor designed acc. to PED Directive
97/23 EC
- Meter size DN 1 (1/25") ... DN 40 (4")
- Accuracy < ± 0.5% of rate (> DN 2)
- Fluid temperature -40 … 130 °C (-40 ... 266 °F) P R O F I
®
B U S
- Hygienic certification: 3A, EHEDG
- Liners, PFA, PTFE, hard or soft rubber
- Standardized installation lengths for DIN, ASME
B16.5 or JIS B2210-10K flanges The all-purpose flowmeter
- Power supply AC/DC
for conductive fluids
Contents
1 Overview of flowmeter sensor and transmitter designs .................................................................................3
2 General information ............................................................................................................................................4
2.1 Measuring accuracy .......................................................................................................................................4
2.2 Grounding.......................................................................................................................................................4
2.3 Installation Requirements...............................................................................................................................4
2.4 Flowmeter Sizes, Pressure Ratings, Flow Range..........................................................................................7
2.5 Flowmeter Sizes, Pressure Ratings, Flow Range..........................................................................................7
2.6 Flowrate nomograph ......................................................................................................................................8
3 Model SE41F ......................................................................................................................................................10
3.1 Technical data ..............................................................................................................................................10
3.2 Dimensional drawings ..................................................................................................................................13
3.3 Ordering information.....................................................................................................................................19
4 Model SE21_ ......................................................................................................................................................26
4.1 Technical data ..............................................................................................................................................26
4.2 Dimensional drawings ..................................................................................................................................28
4.3 Ordering information.....................................................................................................................................39
5 Mounting accessories.......................................................................................................................................42
6 Transmitter.........................................................................................................................................................43
6.1 Technical data ..............................................................................................................................................43
6.2 Inputs/outputs...............................................................................................................................................45
6.3 Digital communication ..................................................................................................................................45
6.4 Terminal connection diagrams .....................................................................................................................47
6.5 Connection examples for peripherals (incl. HART)......................................................................................51
6.6 Dimensional drawings ..................................................................................................................................54
6.7 Ordering information.....................................................................................................................................55
7 Order information: Wafer accessories (table H) ............................................................................................56
8 Order information: Flowmeter sensor simulator FXC 4000 ..........................................................................57
2
Electromagnetic Flowmeter FSM4000 D184S073U02
DN 3 ... 40 (1/10 ... 1 1/2") DN 3 ... 40 (1/10 ... 1 1/2”) DN 50 ... 100 DN 3 ... 40
(2 ... 4") (1/10 ... 1 1/2")
1) 2)
3) 4)
DN 50 ... 1000 (2 ... 4") DN 3 ... 100 (1/10 ... 4") DN 50 ... 100 (2 ... 4”)
G00211 G00210 G00209 G00208
3
Electromagnetic Flowmeter FSM4000 D184S073U02
Pulse output
2.3 Installation Requirements
− DN 1 ... DN 2 (1/25 ... 1/12"): Pos: 11.2 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Einbaubedingungen/Allgemeine Hinweise @ 5\mod_1163075963593_3101.doc @ 46893
± 1% of measured value, ± 0.001 QmaxDN The following points must be observed for the installation:
− DN 3 ... DN 1000 (1/10 ... 40"): • The meter tube must always be completely full.
Q > 0.05 QmaxDN ± 0.5% of measured value • The flow direction must correspond to the identification if present.
− Q < 0.05 QmaxDN ± 0.00025 QmaxDN • The maximum torque for all flange connections must be complied
QmaxDN = maximum flowrate for the flowmeter size 10 m/s with.
• The devices must be installed without mechanical tension
(torsion, bending).
Analog Output Effects • Install flange and wafer units with coplanar counter flanges and
Same as pulse output plus ±0.1% of rate ±0.01 mA use only appropriate gaskets.
6,0 % • Use only flange seals made from a compatible material for the
fluid and fluid temperatures.
5,0 % • Gaskets must not extend into the flow area since possible
turbulence could influence the device accuracy.
4,0 %
• The pipeline may not exert any unallowable forces or torques on
Y 3,0 % the device.
• Do not remove the plugs in the cable connectors until you are
2,0 % DN 1...2 ready to install the electrical cable.
(1/25...1/12”)
1,0 % • Install transmitter at a largely vibration-free location.
DN 3...1000 • Do not expose the transmitter to direct sunlight or provide for
0,0 % (1/10...40”) appropriate sun protection where necessary.
0 2 4 6 8 10 m/s Pos: 11.3 /======= Spaltenumbruch ======== @ 0\mod_1132937966324_3101.doc @ 3831
Fig. 1
Y Accuracy ± of rate
X Flow velocity
Pos: 8 /======= Spaltenumbruch ======== @ 0\mod_1132937966324_3101.doc @ 3831
4
Electromagnetic Flowmeter FSM4000 D184S073U02
Pos: 11.4 /Technische Daten / Datenblatt/Durchfluss/FXE4000/Einbaubedingungen/Elektrodenachse @ 2\mod_1155119847515_3101.doc @ 38596 Pos: 11.10 /Technische Daten / Datenblatt/Durchfluss/FXE4000/Einbaubedingungen/Horizontale Leitungen @ 2\mod_1155120108937_3101.doc @ 38659
3°
max. 45° G00038
Fig. 5
Pos: 11.11 /==== Leeres Modul mit einer Absatzmarke, DS, 1-spaltig ==== @ 2\mod_1153381574375_0.doc @ 35553
1
Pos: 11.12 /Technische Daten / Datenblatt/Durchfluss/FXE4000/Einbaubedingungen/Freier Ein- bzw. Auslauf @ 2\mod_1155120144781_3101.doc @ 38680
• Provide for a siphon fluid intake for free inlets or outlets so that
Pos: 11.6 /Technische Daten / Datenblatt/Durchfluss/FXE4000/Einbaubedingungen/Ein- und Auslaufstrecke (BA + IA) @ 2\mod_1155134311203_3101.doc @ 38877
outlet pipe section (2). Pos: 11.14 /Technische Daten / Datenblatt/Durchfluss/FXE4000/Einbaubedingungen/Stark verschmutzte Messstoffe @ 2\mod_1155120187750_3101.doc @ 38701
• For compliance with the measuring accuracy, observe the inlet 2.3.6 Strongly contaminated fluids
and outlet pipe sections.
• For strongly contaminated fluids, a bypass connection according
to the figure is recommended so that operation of the system can
continue to run without interruption the during the mechanical
1 2 cleaning.
3xDN 2xDN
G00037
Fig. 3
Pos: 11.7 /==== Leeres Modul mit einer Absatzmarke, DS, 1-spaltig ==== @ 2\mod_1153381574375_0.doc @ 35553
Pos: 11.16 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Einbaubedingungen/Montage in der Nähe von Pumpen @ 3\mod_1157631958985_3101.doc @ 39844
G00206
Fig. 4
Pos: 11.9 /======= Spaltenumbruch ======== @ 0\mod_1132937966324_3101.doc @ 3831
G00207
Fig. 8
Pos: 11.17 /======= Spaltenumbruch ======== @ 0\mod_1132937966324_3101.doc @ 3831
5
Electromagnetic Flowmeter FSM4000 D184S073U02
Pos: 11.18 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Einbaubedingungen/Einbau in Rohrleitungen größerer Nennweiten @ 3\mod_1157632081771_3101.doc @ 39865
2.3.8 Installation in larger size pipelines Nomograph for pressure drop calculations
Determine the resulting pressure loss when using reduction For flange transition piece with α/2 = 8°
pieces (1):
1. Calculate the diameter ratio d/D.
2. Determine the flow velocity based on the flow range nomograph
(Fig. 10).
3. Read the pressure drop on the Y-axis in Fig. 10.
8°
D
G00213
Fig. 9
1 = Flange reducer
d = Inside diameter of the flowmeter
V = flow velocity [m/s]
Δp = pressure drop [mbar]
D = Inside diameter of the pipeline
Fig. 10
Pos: 12 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830
6
Electromagnetic Flowmeter FSM4000 D184S073U02
7
Electromagnetic Flowmeter FSM4000 D184S073U02
0
4
80 10
6 10
or steel 10 bar JIS
DN DN
5 3 8
80 SS 1.4571 40 bar DIN 4
2 6
65
(3") (316 Ti) CL150, CL300 ASME 3 5
DN
or steel 10 bar JIS 4
2
100 SS 1.4571 16, 40 bar DIN 3
50
10
1
DN
(4") (316 Ti) CL150, CL300 ASME 8
2
102
40
or steel 10 bar JIS 6
DN
8 5
125 SS 1.4571 16, 40 bar DIN
32
4
6 1
DN
(5") (316 Ti) CL150, CL300 ASME 5 3 8
or steel 4
6
2
15 20 25
150 DIN SS 1.4571 16, 40 bar 3 5
DN DN DN
4
(6") ASME (316 Ti) CL150, CL300 2
1 3
or steel
8
200 DIN SS 1.4571 10, 16, 25, 40 bar 2
10 6
(8") ASME (316 Ti) CL150, CL300 8 5
10
or steel 6
4 DN
10-1
250 DIN SS 1.4571 10, 16, 25, 40 bar 5 3 8
8
DN
4
(10") ASME (316 Ti) CL150, CL300 2
6
6
3 5
or steel
DN
4
300 DIN SS 1.4571 10, 16, 25, 40 bar 2
3
10-1
(12") ASME (316 Ti) CL150, CL300 8
4
2
or steel
DN
1 6
350 ... 1000 DIN SS 1.4571 10, 16, 25 bar 8 5
4
(14 ... 40") ASME (316 Ti) CL150 10-2
3
6
DN
5 3 8
or steel
4
1) Connecting dimensions for flange acc. to DIN2501 / EN1092-1 or 2 6
2
3 5
ASME or JIS.
DN
4
2
10-2 3
5
5
DN
8
4
6 10-3
5 3 8
4
2 6
3 5
4
2
10-3 3
8
Electromagnetic Flowmeter FSM4000 D184S073U02
Example:
l/min m3/h l/s
Flowrate = 7 m3/h (max. value = range end value). Flowmeter sizes
DN 20 ... DN 65 (3/4 ... 2 1/2“) are suitable for a flow velocity between
0.5 ... 10 m/s.
3 Pos: 19 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830
4 6
2 5
3
4
2 3
104
8 2
10 5 6
0 0 0 00 0 0 0 0 00
8 5
25 30 35 40 4 5 5 60 70 80 90 10
3
4 10
DN DN DN DN DN D DN DN DN DN DN
6
5 3 8
4 6
3 2 5
4
2 3
103
N
8 2
10 4 6
8 5
0 0
4 2
6 10
5 3 8
4
2 6
0
20
3 5
DN
4
2
102 3
0
12 15
DN
8
5
2
10 3 6
DN
8 5
4
6 10
5 3 8
4
2 6
3 5
4
2
10 3
9
Electromagnetic Flowmeter FSM4000 D184S073U02
3 Model SE41F
Pos: 22 /Überschriften/1.1/1-spaltig/S - U/Technische Daten @ 7\mod_1157704677334_3101.doc @ 40581
10
Electromagnetic Flowmeter FSM4000 D184S073U02
Storage Temperature
PS PS
-20 ... 70 °C (-4 ... 158 °F) [bar] [psi]
45 652.5
IP 67 35 507.5
IP 68 (option) 30 435.0
25 362.5
Pipeline Vibration Following EN 60068-2-6 PN 25
20 290.0
Transmitter
15 217.5
- In the range of 10…55 Hz, max. deflection 0.15 mm PN 16
10 145.0
Flowmeter sensor 5
PN 10
72.5
- In the range of 10…55 Hz, max. deflection 0.15 mm 0 0
- In the range of 10…55 Hz, max. acceleration 2 g -30 -10 10 30 50 70 90 110 130 150 170 190 [°C]
TS
-22 14 50 86 122 158 194 230 266 302 338 374 [°F]
G00215
Designs
The flanged flowmeters comply with the installation lengths specified Fig. 14: DIN flanges SS 1.4571 (316 Ti) to DN 600 (24")
in VDI / VDE 2641, ISO 13359 or according to DVGW (process sheet
W420, design WP; ISO 4064 short).
PS PS
[bar] [psi]
55 797.5
50 725.5
Material load 45
CL 300
652.5
Limits for allowable fluid temperature (TS) and allowable pressure 40 580.0
(PS) are a function of the liner and flange material used (see the 35 507.5
SS 1.4571 -15 °C 90 °C − Fig. 15: ASME flange SS 1.4571 (316 Ti) to DN 300 (CL150/300)
(316 Ti) (5°F) (194 °F) to DN 1000 (CL150)
Soft rubber Steel -10 °C 60 °C −
(14 °F) (140 °F)
PS PS
SS 1.4571 -15 °C 60 °C − [bar] [psi]
45 652.5
(316 Ti) (5 °F) (140 °F)
40 580.0
PTFE / PFA / Steel -10 °C 130 °C − PN 40
ETFE (14 °F) (266 °F) 35 507.5
0 0
-30 -10 10 30 50 70 90 110 130 150 170 190 [°C]
-22 14 50 86 122 158 194 230 266 302 338 374 [°F] TS
G00217
11
Electromagnetic Flowmeter FSM4000 D184S073U02
35 507.5
32 … 300 Steel 10 -10 … +180 °C 10
30 435.0
(1¼ … 12“) (14 … 266 °F)
25 362.5
20 290.0
CL 150
15 217.5 PS PS
[bar] [psi]
10 145.0 17 246.5
5 72.5 16 232.0
DN 700 PN 16
0 0 15 217.5
-30 -10 10 30 50 70 90 110 130 150 170 190 [°C]
-22 14 50 86 122 158 194 230 266 302 338 374 [°F] TS 14 203.0
13 188.5
G00218
DN 900 PN 16
12 DN 800 PN 16 174.0
Fig. 17: ASME flange steel to DN 300 (CL150/300) 11 159.5
DN 1000 PN 16
to DN 1000 (CL150) 10 145.0
7 DN 1000 PN 10 101.5
6 87.0
-30 -20 -10 0 10 20 30 40 50 60 70 80 90 [°C]
32 TS
-22 -4 14 50 68 86 104 122 140 158 176 194 [°F]
G00219
Fig. 18: DIN flange SS 1.4571 (316 Ti), DN 700 ... DN 1000
(28 ... 40“)
PS PS
[bar] [psi]
17 246.5
16 232.0
15 217.5
14 203.0
13 DN 700 PN 16 188.5
12 174.0
11 DN 900 PN 16 159.5
DN 800 PN 16
10 145.0
DN 1000 PN 16
9 130.5
8 DN 900 PN 10 116.0
DN 800 PN 10
7 DN 700 PN 10 101.5
6 DN 1000 PN 10 87.0
-10 0 10 20 30 40 50 60 70 80 90 [°C]
TS
-14 32 50 68 86 104 122 140 158 176 194 [°F] G00220
Fig. 19: DIN flange steel, DN 700 ... DN 1000 (28 ... 40“)
Pos: 25 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830
12
Electromagnetic Flowmeter FSM4000 D184S073U02
1 88 (3,46) 88 (3,46)
35 (1,38) 35 (1,38)
1
2
ØD2
d2
E
G
E
2
Ød4
ØD
F
F
k
A b k
L* H
G00227
Tolerance: L* DN 3 ... DN 200: +0/-3 (0.12"); DN 250 ... DN 300: +0/-5 (0.20")
Tolerance: L* 1/10 ... 8": +0/-3 (0.12“); 10 ... 12“: +0/-5 (0.20“)
Fig. 20: Dimensions in mm (inch)
1 Cable gland M20 x 1.5 or NPT
2 Number of holes N
13
Electromagnetic Flowmeter FSM4000 D184S073U02
14
Electromagnetic Flowmeter FSM4000 D184S073U02
Pos: 27.4 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Maßzeichnungen/Durchflussmesser SE41/Maßzeichnung Flansch DN3-DN300 (ASME) @ 3\mod_1157638358417_3101.doc @ 40096
Meter dimensions acc. to ASME Flange dimensions acc. to ASME Flange dimensions acc. to ASME Weight
CL 150 / CL 300 CL 150 CL 300
mm mm mm
A L1) 2) E
3)
F G D/H d4 b k d2 N D/H d4 b k d2 N CL 150/
Inch CL 300
ISO old
kg
13359 Inst.
lngth.
1/10...5/16 67 130 130 101 62 129 89 35 13 60 16 4 95 35 16 67 16 4 2.5
3/8. 1/2 67 200 270 101 62 129 89 35 13 60 16 4 95 35 16 67 16 4 3
3/4 87 200 270 112 73 140 99 43 14 70 16 4 117 43 18 83 19 4 3.5
1 87 200 270 112 73 140 108 51 16 80 16 4 124 51 20 89 19 4 4
1 1/4 95 200 280 117 78 145 117 64 18 89 16 4 134 64 21 99 19 4 5
1 1/2 100 200 280 121 82 149 127 73 20 99 16 4 156 73 23 115 23 4 6
2 116 200 280 129 90 157 152 92 21 121 19 4 165 92 24 127 19 8 8
2 1/2 100 200 330 153 104 171 178 105 25 140 19 4 191 105 27 150 23 8 12
3 100 200 340 159 110 177 191 127 26 153 19 4 210 127 30 168 23 8 16
4 130 250 400 179 130 197 229 157 26 191 19 8 254 157 34 200 22 8 15
5 124 250 450 169 127 195 254 186 28 216 22 8 280 186 39 235 22 8 36/38
6 170 300 450 190 148 216 279 216 30 241 22 8 318 216 41 270 22 12 38
8 195 350 500 222 179 248 343 270 34 298 22 8 381 270 47 330 25 12 66
10 250 450 550 250 207 276 406 324 35 362 25 12 445 324 53 387 28 16 98
12 250 500 620 293 250 319 483 381 37 432 26 12 521 381 56 451 32 16 124/181
inch inch inch
A L1) 2) E
3)
F G D/H d4 b k d2 N D/H d4 b k d2 N CL 150/
Inch CL 300
ISO old
lb
13359 Inst.
lngth.
1/10...5/16 2.64 5.12 5.12 3.98 2.44 5.08 3.50 1.38 0.51 2.36 0.63 0.16 3.74 1.38 0.63 2.64 0.63 0.16 5.5
3/8. 1/2 2.64 7.87 10.63 3.98 2.44 5.08 3.50 1.38 0.51 2.36 0.63 0.16 3.74 1.38 0.63 2.64 0.63 0.16 6.6
3/4 3.43 7.87 10.63 4.41 2.87 5.51 3.90 1.69 0.55 2.76 0.63 0.16 4.61 1.69 0.71 3.27 0.75 0.16 7.7
1 3.43 7.87 10.63 4.41 2.87 5.51 4.25 2.01 0.63 3.15 0.63 0.16 4.88 2.01 0.79 3.50 0.75 0.16 8.8
1 1/4 3.74 7.87 11.02 4.61 3.07 5.71 4.61 2.52 0.71 3.50 0.63 0.16 5.28 2.52 0.83 3.90 0.75 0.16 11.0
1 1/2 3.94 7.87 11.02 4.76 3.23 5.87 5.00 2.87 0.79 3.90 0.63 0.16 6.14 2.87 0.91 4.53 0.91 0.16 13.2
2 4.57 7.87 11.02 5.08 3.54 6.18 5.98 3.62 0.83 4.76 0.75 0.16 6.50 3.62 0.94 5.00 0.75 0.31 17.6
2 1/2 3.94 7.87 12.99 6.02 4.09 6.73 7.01 4.13 0.98 5.51 0.75 0.16 7.52 4.13 1.06 5.91 0.91 0.31 26.5
3 3.94 7.87 13.39 6.26 4.33 6.97 7.52 5.00 1.02 6.02 0.75 0.16 8.27 5.00 1.18 6.61 0.91 0.31 35.3
4 5.12 9.84 15.75 7.05 5.12 7.76 9.02 6.18 1.02 7.52 0.75 0.31 10.00 6.18 1.34 7.87 0.87 0.31 33.1
5 4.88 9.84 17.72 6.65 5.00 7.68 10.00 7.32 1.10 8.50 0.87 0.31 11.02 7.32 1.54 9.25 0.87 0.31 79.4/83.8
6 6.69 11.81 17.72 7.48 5.83 8.50 10.98 8.50 1.18 9.49 0.87 0.31 12.52 8.50 1.61 10.63 0.87 0.47 83.8
8 7.68 13.78 19.69 8.74 7.05 9.76 13.50 10.63 1.34 11.73 0.87 0.31 15.00 10.63 1.85 12.99 0.98 0.47 145.5
10 9.84 17.72 21.65 9.84 8.15 10.87 15.98 12.76 1.38 14.25 0.98 0.47 17.52 12.76 2.09 15.24 1.10 0.63 216.1
12 9.84 19.69 24.41 11.54 9.84 12.56 19.02 15.00 1.46 17.01 1.02 0.47 20.51 15.00 2.20 17.76 1.26 0.63 273.4/399.0
1) With a grounding plate mounted on one side of flange in SS 1.4571 (316 Ti), 1/10 ... 5/16" L + 3 mm (0.12"), 5 ... 12" L + 5 mm (0.20").
Other materials upon request. For additional information. see "Grounding".
2) With protective flanges. The protective flanges provide the grounding function, grounding plate is not required, 1/10 ... 5/16“
L + 6 mm (0.24"), 5 ... 12" L + 10 mm (0.39").
Also available with protective flanges in 1000 series. Installation length ≤ 3" L + 20 mm (0.79"), for 4" or higher L + 25 mm (0.98").
3) For high temperature designs < 5" + 20 mm (0.79"), > 4" + 120 mm (4.72").
4) Connection flange 1/2".
Pos: 27.5 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830
15
Electromagnetic Flowmeter FSM4000 D184S073U02
Pos: 27.6 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Maßzeichnungen/Durchflussmesser SE41/Maßzeichnung Flansch DN350-DN1000 (DIN) @ 3\mod_1157638408849_3101.doc @ 40117
Flange acc. to DIN / EN 1092-14). DN 350 ... DN 1000; ASME B16.5 / ASME B16.47 Series B. 14 ... 40"
1
2
d2
G
E
d4
F
b A k
L* D
G00228
Tolerance: L* DN 350 ... DN 500: +0/-5 (0.20"); DN 600 ... DN 1000: +0/-10 (0.39“)
Tolerance: L* 14 ... 20“: +0/-5 (0.20“); 24 ... 40“: +0/-10 (0.39“)
Fig. 21: Dimensions in mm (inch)
1 Cable gland M20 x 1.5 or NPT
2 Number of holes N
16
Electromagnetic Flowmeter FSM4000 D184S073U02
17
Electromagnetic Flowmeter FSM4000 D184S073U02
Pos: 27.8 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Maßzeichnungen/Durchflussmesser SE41/Maßzeichnung Flansch DN350-DN1000 (ASME) @ 3\mod_1157638448817_3101.doc @ 40138
Meter dimensions acc. to ASME CL 150 Flange dimensions acc. to ASME CL 1501) Weight
mm mm
A L2) 3) E F G D k d4 d2 N b
Inch appr. kg
ISO old
13359 Install.
length
14 322 550 650 313 250 341 534 476 413 28.6 12 40 178
16 370 600 700 329 275 367 597 540 470 28.6 16 42 225
18 407 686 – 375 310 403 635 578 533 31.7 16 45 230
20 416 762 780 375 310 403 699 635 584 31.7 20 48 263
24 469 914 850 423 361 454 813 749 692 34.9 20 53 327
28 537 – 910 469 405 495 837 749 762 34.9 20 50 385
32 605 – 1040 519 455 545 942 900 864 22.2 48 51 513
36 671 – 1170 569 505 595 1057 1010 972 25.4 44 57 662
40 739 – 1300 619 555 645 1380 1210 1140 56.0 28 63 1132
inch inch
A L2) 3) E F G D k d4 d2 N b
Inch appr. lb
ISO old
13359 Install.
length
14 12.68 21.65 25.59 12.32 9.84 13.43 21.02 18.74 16.26 1.13 0.47 1.57 392.4
16 14.57 23.62 27.56 12.95 10.83 14.45 23.50 21.26 18.50 1.13 0.63 1.65 496.0
18 16.02 27.01 – 14.76 12.20 15.87 25.00 22.76 20.98 1.25 0.63 1.77 507.1
20 16.38 30.00 30.71 14.76 12.20 15.87 27.52 25.00 22.99 1.25 0.79 1.89 579.8
24 18.46 35.98 33.46 16.65 14.21 17.87 32.01 29.49 27.24 1.37 0.79 2.09 720.9
28 21.14 – 35.83 18.46 15.94 19.49 32.95 29.49 30.00 1.37 0.79 1.97 848.8
32 23.82 – 40.94 20.43 17.91 21.46 37.09 35.43 34.02 0.87 1.89 2.01 1.131.0
36 26.42 – 46.06 22.40 19.88 23.43 41.61 39.76 38.27 1.00 1.73 2.24 1.459.5
40 29.09 – 51.18 24.37 21.85 25.39 54.33 47.64 44.88 2.20 1.10 2.48 2.495.6
1) Other pressure stages upon request. Dimensions 28" or larger acc. to ASME B16.47 Series B.
2) Grounding plate mounted on one side of flange in SS 1.4571 (316 Ti). Other materials upon request, L + 5 mm (0.20").
3) With protection plates attached to flange on both sides. Grounding plate is not required L + 10 mm (0.39").
Also available with protective flanges in 1000 series. Installation length ≤ 3" L + 20 mm (0.79"), for 4" or higher L + 25 mm (0.98").
Pos: 28 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830
18
Electromagnetic Flowmeter FSM4000 D184S073U02
19
Electromagnetic Flowmeter FSM4000 D184S073U02
20
Electromagnetic Flowmeter FSM4000 D184S073U02
21
Electromagnetic Flowmeter FSM4000 D184S073U02
22
Electromagnetic Flowmeter FSM4000 D184S073U02
23
Electromagnetic Flowmeter FSM4000 D184S073U02
24
Electromagnetic Flowmeter FSM4000 D184S073U02
25
Electromagnetic Flowmeter FSM4000 D184S073U02
4 Model SE21_
Pos: 33 /Überschriften/1.1/1-spaltig/S - U/Technische Daten @ 7\mod_1157704677334_3101.doc @ 40581
40 (104)
Gasket material (internal)
Process Standard Option
TMeasuring Connection
-40 -10 90 (194) 120 130 Wafer design None –
(14F) (248) (266) °C (°F)
-10 DN 1 ... DN 2 Weld stubs EPDM (Ethylene- Silicon with FDA
(14) Propylene) with approval (optional,
Threaded pipe
connection FDA approval, resistant to oils and
-25 (-13) G00224 silicone with FDA grease)
Tri-Clamp
approval (CIP- PTFE with
Fig. 22: Temperature graph External threads resistant, no oils or FDA approval
grease) (DN 3 ... 8)
The flowmeter sensor may not be isolated
1/8" sanitary PTFE Viton (only in
connector combination with
Fluid temperature DN 1 … DN 2 (1/25 ... 1/12") PVC process
-10 … 120 °C (14 ... 248 °F), max. allowable cleaning temperature, connection)
see table Flat housing gaskets Silicon (resistant –
to oil, grease)
26
Electromagnetic Flowmeter FSM4000 D184S073U02
35 507.5
5 72.5
0 0
40 580.0
35 507.5
PN 40/CL 300
30 435.0
25 362.5
20 PN 25 290.0
15 217.5
PN 16/CL 150
10 145.0
PN 10
5 72.5
0 0
-60 -50 -40 -30 -20 -10 0 10 20 30 40 50 60 70 80 90 100 110 120 130 *) [°C]
TS
-76 -58 -40 -22 -4 14 32 50 68 86 140 122 140158 176194 212 230 248 266 *) [°F] G00223
Fig. 25
Higher temperatures are allowed for CIP/SIP cleaning for limited time
periods, see Table "Maximum Allowable Cleaning Temperature".
Pos: 36 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830
27
Electromagnetic Flowmeter FSM4000 D184S073U02
G
2
Ø
d2
E
E
Ød4
ØD
F
F
A b
k
L D
G00229
28
Electromagnetic Flowmeter FSM4000 D184S073U02
29
Electromagnetic Flowmeter FSM4000 D184S073U02
Pos: 38.4 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Maßzeichnungen/Durchflussmesser SE21/Maßzeichnung Flansch SE21F, DN50-DN100 @ 12\mod_1189600221421_3101.doc @ 121613
Stainless steel housing, flange acc. to DIN / EN 1092-1, DN 50 ... DN 100; ASME B16.5, 2 ... 4"
80 (3,15)
41 (1,61)
1
71 (2,80)
2
G
G
d2
E
E
Ød4
ØD
A
b
L
G00230
30
Electromagnetic Flowmeter FSM4000 D184S073U02
31
Electromagnetic Flowmeter FSM4000 D184S073U02
Pos: 38.6 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Maßzeichnungen/Durchflussmesser SE21/Maßzeichnung Zwischenflansch SE21W, DN3-DN100 @ 3\mod_1157639606021_3101.doc @ 40264
1 80 (3,15)
41 (1,61)
71 (2,80)
G
E
ØD
ØC
DN 3 ... DN 8 (1/10 ... 5/16”)
A J
F
ØDi
37 (1,46)
B C
4 x Ø5,5 (0,22”) 4
2
ØDi
Ø13.6
A*
3
L G00231
1 80 (3,15)
41 (1,61)
71 (2,80)
G
E
Ø Di
ØC
ØD
F
A*
L
G00459
32
Electromagnetic Flowmeter FSM4000 D184S073U02
33
Electromagnetic Flowmeter FSM4000 D184S073U02
Pos: 38.8 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Maßzeichnungen/Durchflussmesser SE21/Maßzeichnung Variabler Prozessanschluss SE21_, DN3-DN100 @ 3\mod_1157639609055_3101.doc @ 40285
E71 (2,80)
G
ØD
3 (0,12)
3
F
A 4 x Ø5.5
B C
(1,46)
2 5
37
4 G00232
3 (0,12)
G
E
ØD
F
10,5 (0,41)
B 2 C 3
4 x Ø5.5 5 DN 65...100 4 x Ø5.5 DN 50 (2)
(2 1/2 ... 4) 4
G00233
Fig. 29: Dimensions in mm (inch)
1 Cable gland Pg 13.5 or NPT 4 Hole pattern
2 Pattern for axial holes 5 LPipe
3 Mounting bracket (optional) 6 Number of holes N
34
Electromagnetic Flowmeter FSM4000 D184S073U02
35
Electromagnetic Flowmeter FSM4000 D184S073U02
Pos: 38.10 /Technische Daten / Datenblatt/Durchfluss/FXE4000/Maßzeichnungen/Durchflussmesser DE21-23/Maßzeichnung Adapter für variable Anschlüsse @ 8\mod_1175156265359_3101.doc @ 75146
4.2.4 Adapter for variable process connections DN 3 ... DN 100 (1/10 ... 4")
Da
Di
L L
1 2
a
Da
Di
R
L L
3 4 G00072
Fig. 30
1 Weld stubs 3 Tri-Clamp
2 Threaded pipe connection DIN 11851 4 External threads
36
Electromagnetic Flowmeter FSM4000 D184S073U02
37
Electromagnetic Flowmeter FSM4000 D184S073U02
Pos: 38.12 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Maßzeichnungen/Durchflussmesser SE21/Maßzeichnung 1/8"-Sanitäranschluss DS21B, DN1-DN2 @ 3\mod_1157639612500_3101.doc @ 40306
2 (0,07)
4
10 (0,73)
26
(1,02) 4 x Ø5.5
40
60 (2,36) (1,57)
25
90 (3,54) (0,98)
135 (5,31)
69 (2,71)
G00236
Fig. 31: Dimensions in mm (inch)
1 Gaskets 3 Cable gland Pg 13.5 or NPT
2 Connecting dimensions for gland G 1/8" internal thread 4 Mounting bracket, standard
Important
The flowmeter sensor is equipped as standard with grounding electrodes in the same material as the measuring
electrodes. The flowmeter sensor can only be operated on S4 transmitters with software version B.30 or higher.
Preamplifier A B C D E F
Without / type A 119 (4.68) 71 (2.79) 64 (2.51) 40 (1.57) 80 (3.14) 41 (1.61)
Type B 104 (4.09) 56 (2.20) 163 (6.41) 71 (2.79) 142 (5.59) 70 (2.75)
38
Electromagnetic Flowmeter FSM4000 D184S073U02
Fixed and wafer flange, 1/8 in. Sanitary connectors, variable process connections, liner Torlon, PEEK, PFA
Electromagnetic Flowmeter Variant digit No. 1-5 6 7 8 9 10 11 12 13
FSM4000-SE21 Catalog No. SE21-
remote standard DN 1 ... DN 100
Process Connection
Wafer type 1) W
Welded flange DIN 2501 / EN 1092-1 / ASME (DN 3 ... DN 100) F
Weld stubs DIN 11850 R
Weld stubs ISO 2037 (only DN 10 ... DN 100) P
Weld stubs DIN 2463 Q
Weld stubs ISO 1127 J
Weld stubs SMS (only DN 25, DN 40 ... 100) X
Without adapter Spare sensor, for replacement only V
Pipe connection DIN 11851 S
Tri-Clamp DIN 32676 T
Tri-Clamp ASME BPE (DN 3 ... DN 25, DN 40 ... DN 100) K
Male thread ISO 228 / DIN 2999 conical (only DN 3 ... DN 25) E
1/8 in. pipe thread (only DN 1 ... DN 2) B
Others Z
Liner Material Meter Size
PEEK DN 1 1/25 in. 2) 3) K 0 1
PEEK DN 1.5 1/17 in. 2) 3) K 1 S
PEEK DN 2 1/12 in. 2) 3) K 0 2
Torlon DN 1 1/25 in. 2) 3) N 0 1
Torlon DN 1.5 1/17 in. 2) 3) N 1 S
Torlon DN 2 1/12 in. 2) 3) N 0 2
PFA DN 3 1/10 in. P 0 3
PFA DN 4 5/32 in. P 0 4
PFA DN 6 1/4 in. P 0 6
PFA DN 8 5/16 in. P 0 8
PFA DN 10 3/8 in. P 1 0
PFA DN 15 1/2 in. P 1 5
PFA DN 20 3/4 in. P 2 0
PFA DN 25 1 in. P 2 5
PFA DN 32 1-1/4 in. P 3 2
PFA DN 40 1-1/2 in. P 4 0
PFA DN 50 2 in. P 5 0
PFA DN 65 2-1/2 in. P 6 5
PFA DN 80 3 in. P 8 0
PFA DN 100 4 in. P 1 H
39
Electromagnetic Flowmeter FSM4000 D184S073U02
40
Electromagnetic Flowmeter FSM4000 D184S073U02
41
Electromagnetic Flowmeter FSM4000 D184S073U02
5 Mounting accessories
Pos: 44 /==== Wechsel ein- auf zweispaltig ==== @ 0\mod_1130421847171_3101.doc @ 3828
Wechsel ein-auf zweispaltig
Pos: 45 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Allgemein/Montagezubehör @ 3\mod_1157702194544_3101.doc @ 40434
Protective flange
Protective flanges protect the liner from abrasive fluids and are
centered in the pipeline using a groove or spring. Protective flanges
are available in two thicknesses and are mounted on the flowmeter.
Installation length for Series 1000 and DIN / ANSI flanges ≤ DN 80
(3") L+20 mm, from DN 100 (4") L+25 mm.
Grounding plate is installed loose between flange and counter flange. Chemical applications (grounding)
For plastic pipelines or pipelines with insulating liners,
protection/grounding plates made of conductive PTFE (patent
Available in DN 3 ... DN 600 (1/10 ... 24"), Mat. 1.4571 (316 Ti),
pending) are used instead of expensive special materials (platinum,
Other materials upon request.
tantalum, etc.). These plates have tested and used successfully in
DN Inch PN S = Thickness numerous applications.
3 ... 80 1/10 ... 3 10 ... 40 1.5
100 ... 150 4 ... 6 25 ... 40 1.5 di
1
200 ... 250 8... 10 10 ... 16 1.5
d2
300 ... 600 12 ... 24 10 ... 16 3 R8 R2
k D
16
6,6
Ground connection1)
C l1 4±0,5
R2 R2 2
6,4
G00238
R8 B R8 Fig. 34: PTFE protection/grounding plates
ØB
20
ØA
16
16
10
6,6
1) Wafer accessories
Attach grounding cable with M6 screw, spring washer and nut
to the washer The following accessories are available based on meter size and
pressure rating: Bolts, nuts, lockwashers. Gaskets are not included in
ØB = Internal diameter of liner the accessories. For order information, see page 56 .
2) Attach the grounding plate directly on the flowmeter
DN 10 ... DN 600 (3/8 ... 24") Weld connections
Material: Stainless steel 1.4571 (316 Ti) For flowmeters with welded connections, welding adapters are
Thickness ≤ DN 100 (4") = 3 mm; ≥ DN 125 (5") = 5 mm available. For order information, see page 56 .
ØB = Internal diameter of liner Pos: 46 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830
42
Electromagnetic Flowmeter FSM4000 D184S073U02
6 Transmitter
Pos: 49 /Überschriften/1.1/1-spaltig/S - U/Technische Daten @ 7\mod_1157704677334_3101.doc @ 40581
Design
Field-mount housing made of cast aluminum per DIN 1725, painted.
Paint coat is 80 µm thick. Lower section (RAL 7012), upper section
(RAL 9002). See dimensions, page 54. Weight, approx. 3.3 kg.
Electrical connections
Cable entry M20 x 1.5, NPT, PF with adapter (upon request), screw
terminals
Damping (1τ)
Configurable for 0.07 ... 20 s.
Signal cable
Fig. 35: Transmitter FSM4000-S4
Max. cable length between flowmeter sensor and transmitter is 50 m
for the standard design and versions with empty pipe detection, from
DN 10 (3/8") and from 20 µS/cm. 200 m for designs with preamplifier.
Flow range Any flow range whose 100% value A 10 m signal cable is included with the flowmeter. If a longer cable is
corresponds to a flow velocity between 0.5 m/s required, see the order information for the transmitter on page 55.
and 10 m/s can be selected.
Minimum ≥ 20 µS/cm standard DN 10 ... 1000 Magnet coil supply cable
conductivity (3/8 ... 40“) Is required to connect the flowmeter sensor to the transmitter. For
≥ 20 µS/cm with preamplifier DN 1 ... 8 EMC reasons, the 2-wire cable is shielded. A 10 m cable is included
with the flowmeter. If a longer cable is required, see order information
(1/25 ... 5/16“) on page 55.
≥ 5 µS/cm with preamplifier DN 1 ... 1000
(1/25 ... 40“) Forward/reverse flow metering
≥ 0.5 µS/cm with preamplifier DN 10 ... 1000 Flow direction is indicated by direction arrows in the display and over
(3/8 ... 40“) the contact for external signaling. The contact output can also
transmit this information.
Reproducibility DN 1 ... 2 (1/25...1/12”):
(measurement ≤ ± (0.3 % of meas. value + 0.04 % of QmaxDN) Display
period = 100 s) DN 3 ... 1000 (1/10...40"): 4-line, illuminated graphic display. To improve readability, the flow
≤ ± (0.1 % of meas. value + 0,01 % of QmaxDN) information can be displayed on two lines with larger characters. This
allows the individual flowrate and output values to be displayed.
Response time 1τ = 70 ms (0 ... 66 %) "Fast" mode distinctive dot-matrix display with max. 4x16 characters. A plaintext
1τ = 200 ms (0 ... 66 %) "Standard" mode / diagnostic display opens automatically when the system detects an
piston pump error. An alarm condition can also be signaled via contact output.
Supply power U = 100 ... 230 V, 50/60 Hz
Urat = 85 ... 253 V, 50/60 Hz Data security
Via FRAM (ferroelectric non-volatile random access memory) all data
50/60 Hz ± 6% is stored for a period of 10 years (even without power) at shut-off or
20.4 ... 26.4 V AC, during a power outage. Additional security is provided by a serial
20.4 ... 31.2 V DC, ripple ≤ 5 % FRAM installed in the transmitter and on the external connection
Power S ≤ 45 VA (flowmeter sensor incl. transmitter) board in which the parameter settings and process information are
stored. This makes it easy to replace a transmitter – without
Ambient -20 ... 60 °C (-4 ... 140 °F)
reconfiguration – by simply uploading the stored data from the
temperature
external FRAM to the new transmitter.
Storage -20 ... 80 °C (-4 ... 176 °F)
Temperature Flow totalization
The flow is totalized in engineering units. The limits for the pulse
Relative humidity factor are configurable and depend on the measuring range (Qmax)
as well as the totalizer units selected. The limits are a min.
Acc. to IEC 60068-2-30 classification of environmental conditions,
(0.00016 Hz) and max. pulse frequency (5 KHz). Totalizer values for
natural factors, temperature and air humidity. No effect under the
the forward and reverse flow directions can be displayed simultan-
following conditions: Temperature in range from 25 ... 55 °C
eously by selecting the "Display" menu item.
(77 ... 131 °F) and a relative humidity of 94 ... 97%.
Vibration
Acc. to IEC 60068-2-6 (03/95), grouping of devices according to table
C2 for general industry applications. No additional effect on the
following levels of vibration. Frequency: 10 ... 55 Hz; amplitude max.
0.15 mm.
43
Electromagnetic Flowmeter FSM4000 D184S073U02
Parameter adjustment The parameters can be checked for min.-max. errors. If the value is
The display supports several languages (German, English, French, above or below the specified limit, a diagnostic error or warning,
Finnish, Swedish, Italian, Spanish, Dutch or Turkish), and information depending on configuration, is triggered. The fingerprint database
can be entered by using the three keys in the dialog or via integrated in the transmitter allows you to compare the values at the
communication interface with PC, HART protocol, PROFIBUS or time of factory calibration or commissioning with the currently
FOUNDATION Fieldbus. The magnet stick can also be used to recorded values.
configure the flowmeter with the front door closed. Position the Users can store up to three different fingerprint data records
magnet stick over the magnet icons on the display cover. (commissioning, product1, product2) in the fingerprint database. The
default fingerprint (optional) cannot be overwritten.
Alarm signals
A max-min alarm can be used with the transmitter. The alarm limits Detectable error states
can be set between 0 ... 103% of the flow range. When the flowrate is Monitoring of ... Detection of ...
outside the alarm limits, an error message is displayed in the upper
line and a contact is triggered. The contact output can also transmit Coil current / coil Changes in the driver or coil circuit.
this information. voltage / DAC value
Coil resistance Changes in coil circuit, e.g., fine short.
Wiring error Coil temperature Risk of overheating due to fluid.
Monitoring of magnet coil or electrode circuits for short circuits or wire Coil insulation Problems with the coil insulation. Can be
breaks. resistance caused, e.g., by dampness in the
flowmeter sensor or in the connection
Upgrades box.
The transmitter with all its functions can also be used with older Linearity of magnetic External magn. field interference on the
model electromagnetic flowmeters. The appropriate model number circuit EMF, resulting in loss of accuracy.
must be selected in the transmitter. It is also possible to use the Signal quality (signal- Changes in fluid, e.g., gas bubbles,
transmitter with model 10D1422 flowmeters. A special FRAM (Cs = to-noise ratio) increased/reduced amount of solids.
100%, Cz = 0%) is required when updating an older model (see order
information for transmitter) and applies for meter sizes DN 3 ... DN Electrode balance Distorted flow profile, e.g., due to
1000 (1/10 ... 40") in low voltage design only. improper installation.
Detect interference in electrode circuit,
Isolation of input/output e.g., failure of electrode due to insulating
The current output, pulse output, contact input and contact output are coating or a collapsed liner (vacuum
galvanically isolated from the input circuit and from each other. shock).
Recorded parameters/fingerprints
Cyclical (10 s ... 7 days) or manual acquisition of parameters.
The data records are stored in the transmitter FRAM (with time
stamp / operating hours counter). A maximum of 10 records can be
stored. When the eleventh record is stored, the oldest data record is
overwritten automatically.
44
Electromagnetic Flowmeter FSM4000 D184S073U02
Pos: 53 /Überschriften/1.1/1-spaltig/D - F/Ein-/Ausgänge @ 11\mod_1157704418793_3101.doc @ 40560 Pos: 56 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Technische Daten/Messumformer/Digitale Kommunikation @ 3\mod_1157705112080_3101.doc @ 40602
DC output (terminals 31, 32) The transmitter has the following options for digital communication:
The following functions can be selected in the software:
0 ... 20 mA load ≤ 560 Ω 6.3.1 HART protocol
4 ... 20 mA load ≤ 560 Ω
0 ... 10 mA load ≤ 1120 Ω The system is registered with the HART Communication Foundation.
2 ... 10 mA load ≤ 1120 Ω
45
Electromagnetic Flowmeter FSM4000 D184S073U02
100
HSE-Bus B 1 F
100
H2-Bus A 1 F
System integration
The following is required:
• DD file (Device Description), which includes the device
description.
• The CFF file (common file format) is required for the segment
engineering. The engineering can be undertaken online or offline.
You can find both files, as well as the interface description on the
supplied CD (part no.: D699D002U01). This can be re-ordered, if
necessary, from ABB at no cost.
The files needed for operation can also be downloaded at
http://www.fieldbus.org.
Pos: 57 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830
46
Electromagnetic Flowmeter FSM4000 D184S073U02
1
1 1 2- 1+ 1
1
3 2S 2 1 1S 3 16 6 7 51 52 81 82 41 42 31 32 M3 M1 N L
8 9 10 11 12 13 11 11 14 15 2 3 4 5 6 7
SE 3 2S 2 1 1S 3A 16 M3 M1 SE
1 G00595
Flowmeter sensor
Fig. 39: Connection diagram: Flowmeter sensor standard DN 10 ... DN 1000 (3/8 ... 40")
1) see chapter “connection examples for peripherals“ in the operating instructions and/or data sheet
Comment:
We recommend that shielded output cables be used with the shields connected to the functional ground at one end.
Pos: 59.3 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830
47
Electromagnetic Flowmeter FSM4000 D184S073U02
Pos: 59.4 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Elektrischer Anschluss/mit Vorverstärker, DN1 ... DN1000 @ 19\mod_1206005940562_3101.doc @ 171683
1
1 1 2- 1+ 1
1
3 U+ 2 1 U- 3 16 6 7 51 52 81 82 41 42 31 32 M3 M1 N L
8 9 10 11 12 13 11 11 14 15 2 3 4 5 6 7
16
SE 3 U+ 2 1 U- 3A 16 M3 M1 SE
1 G00596
Flowmeter sensor
Fig. 40: Connection diagram: Flowmeter sensor with preamplifier DN 1 ... DN 1000 (1/25 ... 40"), transmitter field-mount housing
1) see chapter “connection examples for peripherals“ in the operating instructions and/or data sheet
Comment:
We recommend that shielded output cables be used with the shields connected to the functional ground at one end.
Important
If the flowmeter sensor is equipped with a preamplifier for low conductivity or in meter size DN 1 ... DN 8
(1/25 ... 5/16“), the shieldings of the signal wires must be connected to terminals U+ and U- on both the flowmeter
sensor and the transmitter.
Pos: 59.5 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830
48
Electromagnetic Flowmeter FSM4000 D184S073U02
Pos: 59.6 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Elektrischer Anschluss/Nachrüstung für Modell 10D1422_, DN3 ... DN1000 @ 19\mod_1206006044656_3101.doc @ 171708
Retrofitting for model 10D1422: DN 3 ... DN 1000 (1/10 ... 40“); model 10DI1425 and 10DS3111A-E:
DN 500 ... DN 1000 (20 ... 40“)
10
1 8
24 V
3 2S 2 1 1S 3 16 6 7 51 52 81 82 41 42 31 32 M3 M1 N L 1
1
7 2 3 4 5 1 6
SE 3 2 1 6 7 L N
1
G00241
Flowmeter sensor
Fig. 41: Flowmeter sensor standard DN 3 ... DN 1000 (1/10 ... 40"), transmitter field-mount housing
1 Functional ground (busbar) 7 Shielded signal cable: Use ABB order no. D173D025U01
2 Pulse output1) or incorporate in existing wiring
3 Contact input1) 8 Reference cable:
4 Contact output1) Only for connection to model 10D1422
5 Current output1) Terminals: close 6,7 and the hook switch S903
6 Supply power 9 Magnet coil supply:
Low voltage: 100 ... 230 V AC, terminals L, N, Magnet coil supply via supply power
Frequency: 47 Hz ≤ f ≤ 53 Hz; 50 Hz supply power 10 Anschlussplatte: D685A1020U03
Frequency: 56 Hz ≤ f ≤ 64 Hz; 60 Hz supply power
1) see chapter “connection examples for peripherals“ in the operating instructions and/or data sheet
Comment:
We recommend that shielded output cables be used with the shields connected to the functional ground at one end.
49
Electromagnetic Flowmeter FSM4000 D184S073U02
Pos: 59.8 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Elektrischer Anschluss/PROFIBUS PA / FOUNDATION Fieldbus, DN1 ... DN1000 @ 19\mod_1206006116187_3101.doc @ 171733
6 6 3c
FF+ FF-
3b
1 3a
U+ U- PA+ PA- 1 2- 1+ 1
1
3 2S 2 1 1S 3 16 41 42 97 98 M3 M1 N L
1
7 8 9 10 11 12 10 10 13 14 2 3 4 5
15
SE 3 2S 2 1 1S 3A 16 M3 M1 SE
16
SE 3 U+ 2 1 U- 3A 16 M3 M1 SE
1 G00597
Flowmeter sensor
50
Electromagnetic Flowmeter FSM4000 D184S073U02
I E I = internal
E = external
+31
0/4 ... 20 mA
0/2 ... 10 mA
-32
G00243
Current output: selectable in software
Function: Active
0/4 ... 20 mA (0 Ω ≤ RB ≤ 560 Ω)
0/2 ... 10 mA (0 Ω ≤ RB ≤ 1120 Ω) (with HART only 4 ... 20 mA)
Terminals: 31, 32
Fig. 43: Current output active with/without HART protocol (4 ... 20 mA)
Contact output
I E Note
R B*
+U Horn and alarm light are shown only
41 0V as examples. Other suitable devices
24 V DC such as bells, sirens, buzzers, etc.,
42 can also be used.
* R B UCE
ICE G00244
Fig. 44: Contact output for system monitoring, Max.-Min. alarm for empty tube or forward/reverse signal
Contact input
I E
81 + 16 ... 30 V DC
Ri
82 0V
G00245
Fig. 45: Contact input for external totalizer reset and external zero return
51
Electromagnetic Flowmeter FSM4000 D184S073U02
Pulse output
52 -
* RB UCE
ICE G00247
Fig. 47
52
Electromagnetic Flowmeter FSM4000 D184S073U02
Important
If the transmitter module is removed, bus termination is also canceled.
3 U+ 2S 2 1 1S U- 3 16 6 7 41 42 97 98 M3 M1
G00249
Hook switches for
Connection board for field housing unit
bus termination must be closed
Fig. 48
Pin assignment
4 3 (Front view showing pin insert and pins)
PIN 1 = PA+
1 2
PIN 2 = nc
PIN 3 = PA-
PIN 4 = shield
G00250
Fig. 49
Pos: 60 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830
53
Electromagnetic Flowmeter FSM4000 D184S073U02
265 (10,43)
249 (9,80)
14,5 (0,57)
1
3
2
38 (1,49)
66 (2,59)
132 (5,19) 83,5 (3,28)
M20 x 1,5
min.175
(6,88)
min. 62 (2,44)
4
G00073
54
Electromagnetic Flowmeter FSM4000 D184S073U02
Accessories
Catalog No.
Shielded signal cable (10 m included in price of the sensor) D173D025U01 price per meter
Shielded coil cable (10 m included in price of the sensor) D173D147U01 price per meter
Accessories for 2 in. pipe mounting 612B091U07
55
Electromagnetic Flowmeter FSM4000 D184S073U02
Note:
Depending on the nominal width and nominal pressure, the following accessories are available: bolts, nuts,
lock washers, centering elements. Seals are not included in the accessories.
This fitting piece is an aid for primaries with weld stubs process connection
It allows these weld stubs to be welded coplanar into the pipe
Pos: 68 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830
56
Electromagnetic Flowmeter FSM4000 D184S073U02
57
ABB has Sales & Customer Support The Company’s policy is one of continuous product
expertise in over 100 countries worldwide. improvement and the right is reserved to modify the
information contained herein without notice.
www.abb.com/flow