Data Sheet Electromagnetic Flowmeter

D184S073U02
FSM4000
Pos: 1 /Titelblätter / Copyright/DB/Durchfluss/FSM4000 @ 3\mod_1157697105687_3101.doc @ 40413

Function
- Electromagnetic flowmeters can be used to
accurately measure the flowrate of liquids, slurries,
pastes and sludges with an electrical conductivity
of at least 20 µS/cm (opt. 5/0.5 µS/cm).

Applications
- The system is specially designed to measure
liquids in the pulp & paper and food & beverage
industries. It can be used to measure fast
changing processes, two phase liquids, continuous
and pulsating flows (piston pump application).

Benefits
- Digital signal processing (DSP) and zero stability
ensure long-term stability and accuracy in both
flow directions.
- Enhanced diagnostic functions and monitoring of
coil and electrode circuits as well as magnetic field
for increased operational stability. Verification via
fingerprint.

Key features
- Flowmeter sensor designed acc. to PED Directive
97/23 EC
- Meter size DN 1 (1/25") ... DN 40 (4")
- Accuracy < ± 0.5% of rate (> DN 2)
- Fluid temperature -40 … 130 °C (-40 ... 266 °F) P R O F I
®

(opt. 180 °C [356 °F]) PROCESS FIELD BUS

B U S
- Hygienic certification: 3A, EHEDG
- Liners, PFA, PTFE, hard or soft rubber
- Standardized installation lengths for DIN, ASME
B16.5 or JIS B2210-10K flanges The all-purpose flowmeter
- Power supply AC/DC
for conductive fluids

Pos: 2 /Inhaltsverzeichnis/Inhaltsverzeichnis für alle Dokumente @ 0\mod_1138710310890_3101.doc @ 3129


Contents
Electromagnetic Flowmeter FSM4000 D184S073U02

Contents
1 Overview of flowmeter sensor and transmitter designs .................................................................................3
2 General information ............................................................................................................................................4
2.1 Measuring accuracy .......................................................................................................................................4
2.2 Grounding.......................................................................................................................................................4
2.3 Installation Requirements...............................................................................................................................4
2.4 Flowmeter Sizes, Pressure Ratings, Flow Range..........................................................................................7
2.5 Flowmeter Sizes, Pressure Ratings, Flow Range..........................................................................................7
2.6 Flowrate nomograph ......................................................................................................................................8
3 Model SE41F ......................................................................................................................................................10
3.1 Technical data ..............................................................................................................................................10
3.2 Dimensional drawings ..................................................................................................................................13
3.3 Ordering information.....................................................................................................................................19
4 Model SE21_ ......................................................................................................................................................26
4.1 Technical data ..............................................................................................................................................26
4.2 Dimensional drawings ..................................................................................................................................28
4.3 Ordering information.....................................................................................................................................39
5 Mounting accessories.......................................................................................................................................42
6 Transmitter.........................................................................................................................................................43
6.1 Technical data ..............................................................................................................................................43
6.2 Inputs/outputs...............................................................................................................................................45
6.3 Digital communication ..................................................................................................................................45
6.4 Terminal connection diagrams .....................................................................................................................47
6.5 Connection examples for peripherals (incl. HART)......................................................................................51
6.6 Dimensional drawings ..................................................................................................................................54
6.7 Ordering information.....................................................................................................................................55
7 Order information: Wafer accessories (table H) ............................................................................................56
8 Order information: Flowmeter sensor simulator FXC 4000 ..........................................................................57

2
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 3 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Übersichten/Übersicht Aufnehmer und Messumformer @ 3\mod_1157623996796_3101.doc @ 39718

1 Overview of flowmeter sensor and transmitter designs


Housing Material Alumin. Hous. Series 4000 Stainless Steel Housing Series 2000

DN 3 ... 40 (1/10 ... 1 1/2") DN 3 ... 40 (1/10 ... 1 1/2”) DN 50 ... 100 DN 3 ... 40
(2 ... 4") (1/10 ... 1 1/2")

1) 2)

3) 4)
DN 50 ... 1000 (2 ... 4") DN 3 ... 100 (1/10 ... 4") DN 50 ... 100 (2 ... 4”)
G00211 G00210 G00209 G00208

Fixed Flange Fixed Flange Wafer design Variable Connections


Flowmeter sensor
Model Number SE41F SE21F SE21W SE21_*
Accuracy 0.5 % of rate (> DN 2)
DN PN DN PN DN PN DN PN *
- - 3 … 50 10 … 40
Wafer design
65 … 100 10 … 16
Flange DIN 2501/EN 1092-1 3 ... 1000 10 ... 40 3 … 100 10 … 40 - -
Flange ASME B16.5 / B16.47 1/10 ... 40" CL 150 1/10“ … CL150 … 1/10“ … 2“ CL300/JIS -
JIS B2210-10K 1/10 … 12“ CL 300 4“ CL300/JIS 3“ … 4“ CL150/JIS
- - - 3…40 (1/10...1 1/2“) 40 S
Pipe connection
50, 80 (2“, 3“) 16 S
DIN 11851
65, 100 (2 1/2“, 4“) 10 S
- - - 3 … 40 (1/10...1 1/2“) 40 R
Weld stubs DIN 11850 50, 80 (2“, 3“) 16 R
65, 100 (2 1/2“, 4“) 10 R
- - - 3 … 40 (1/10...1 1/2“) 40 Q/J
Weld stubs DIN 2463 /
50, 80 (2“, 3“) 16 Q/J
ISO 1127
65, 100 (2 1/2“, 4“) 10 Q/J
- - - 25…40 (1...1 1/2“) 40 P/X
Weld stubs ISO 2037 / SMS 50, 80 (2“, 3“) 16 P/X
65, 100 (2 1/2“, 4“) 10 P/X
Tri-Clamp DIN 32676 / - - - 3 … 50 (1/10...2“) 16/10 T/K
ASME BPE 65, 100 (2 1/2“, 4“) 10/10 T/K
External threads ISO 228 / - - - 3 …25 (1/10...1”) 16 E
DIN 2999
1/8" sanitary connectors - - - 1 …2 (1/25...1/12”) 10 B
Hard/soft rubber, PTFE, PFA, PEEK, Torlon (< DN 3)
Liner PFA (vacuum tight) PFA (vacuum tight)
ETFE, other PFA (> DN 2)
Conductivity ≥ 20 µS/cm (optional ≥ 5/0.5 µS/cm)
Electrodes SS 1.4571 (316 Ti), 1.4539 (904 L), Hastelloy B-3/C-4, Platinum-Iridium, Tantalum, Titanium
> DN 2: SS 1.4404 (316 L)
Process Connection Material Steel, SS 1.4571 (316 Ti) 1.4404 (316 L) -
< DN 3: 1.4571 (316 Ti), PVC, POM
Protect. Class per EN 60529 IP 67 / IP 68 IP 67 / IP 68 IP 67 / IP 68 IP 67 / IP 68
-25 ... 130 °C / 180 °C -40 ... 130 °C -25 ... 130 °C -25 ... 130 °C
Fluid Temperatures
-13 ... 266 °F / 356 °F -40 ... 266 °F -13 ... 266 °F -13 ... 266 °F
Approvals
Hygienic + sterile requirem. CIP/SIP-enabled 3A, EHEDG, CIP/SIP-enabled
Transmitter
Model Number S4
Supply power 85 ... 253 V AC, 24 V AC/DC
Current output 0/2 ... 10 mA, 0/4 ... 20 mA
Pulse output active (24 V), optocoupler (220 mA)
Ext. zero return yes
Ext. totalizer reset yes
Forward/Rev. Flow Metering yes
Communication HART protocol, PROFIBUS PA, FOUNDATION Fieldbus
Pipe empty detection std. yes, DN 10 or higher and ≥ 20 µS/cm
Self-monitoring, enhanced yes , expanded diagnostic functions / fingerprint only in connection with flowmeters SE21,
diagnostic functions SE21F and SE41F for DN 10 or higher
On-site display/totalization yes
Density correction yes, manual entry (totalize and display in mass units)
G00212
Protect. Class per EN 60529 IP 67, NEMA 4X
Housing Field housing unit
1) Weld stubs 2) Tri-Clamp 3) Threaded pipe connection 4) External threads
Pos: 4 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

3
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 5 /==== Wechsel ein- auf zweispaltig ==== @ 0\mod_1130421847171_3101.doc @ 3828


Wechsel ein-auf zweispaltig
Pos: 6 /Überschriften/1/A - C/Allgemeine Daten @ 0\mod_1132927923465_3101.doc @ 3133 Pos: 9 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Allgemein/Erdung @ 11\mod_1184752201296_3101.doc @ 110913

2 General information 2.2 Grounding


Pos: 7 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Allgemein/Messgenauigkeit @ 3\mod_1157629438861_3101.doc @ 39739

The grounding of the flowmeter sensor is necessary not only for


2.1 Measuring accuracy safety reasons, but also to as sure proper operation of the electro-
magnetic flowmeter. The ground screws on the flowmeter sensor are
2.1.1 Reference conditionsper EN 29104 to be connected to earth potential. For technical reasons, this
Fluid Temperatures 20 °C (68 °F) ± 2 K potential should be identical to the potential of the metering fluid, if
Ambient temp. 20 °C (68 °F) ± 2 K possible.
Supply power Line voltage per type plate UN ± 1% For plastic or insulated lined pipelines, the fluid is grounded by
installing ground plates. When there are stray potentials present in
Installation Conditions − upstream > 10 x DN the pipeline, a grounding plate is recommended on both ends of the
straight section meter sensor.
− downstream > 5 x DN For devices with hard rubber lining for DN 125 (5") and larger, the
straight section liner contains a conductive element. This ensures the measured
medium is grounded.
DN = Flowmeter sensor size
To comply with the EMC- and Low Voltage Regulations the
Warm Up Phase 30 min
connection box/transmitter must be grounded in addition to the meter
tube of the flowmeter sensor.

2.1.2 Maximum Measurement Error


Pos: 10 /==== Leeres Modul mit einer Absatzmarke, DS, 1-spaltig ==== @ 2\mod_1153381574375_0.doc @ 35553

Pos: 11.1 /Überschriften/1.1/2-spaltig/Einbaubedingungen @ 2\mod_1155119709593_3101.doc @ 38554

Pulse output
2.3 Installation Requirements
− DN 1 ... DN 2 (1/25 ... 1/12"): Pos: 11.2 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Einbaubedingungen/Allgemeine Hinweise @ 5\mod_1163075963593_3101.doc @ 46893

± 1% of measured value, ± 0.001 QmaxDN The following points must be observed for the installation:
− DN 3 ... DN 1000 (1/10 ... 40"): • The meter tube must always be completely full.
Q > 0.05 QmaxDN ± 0.5% of measured value • The flow direction must correspond to the identification if present.
− Q < 0.05 QmaxDN ± 0.00025 QmaxDN • The maximum torque for all flange connections must be complied
QmaxDN = maximum flowrate for the flowmeter size 10 m/s with.
• The devices must be installed without mechanical tension
(torsion, bending).
Analog Output Effects • Install flange and wafer units with coplanar counter flanges and
Same as pulse output plus ±0.1% of rate ±0.01 mA use only appropriate gaskets.
6,0 % • Use only flange seals made from a compatible material for the
fluid and fluid temperatures.
5,0 % • Gaskets must not extend into the flow area since possible
turbulence could influence the device accuracy.
4,0 %
• The pipeline may not exert any unallowable forces or torques on
Y 3,0 % the device.
• Do not remove the plugs in the cable connectors until you are
2,0 % DN 1...2 ready to install the electrical cable.
(1/25...1/12”)
1,0 % • Install transmitter at a largely vibration-free location.
DN 3...1000 • Do not expose the transmitter to direct sunlight or provide for
0,0 % (1/10...40”) appropriate sun protection where necessary.
0 2 4 6 8 10 m/s Pos: 11.3 /======= Spaltenumbruch ======== @ 0\mod_1132937966324_3101.doc @ 3831

0 6,5 13,1 19,7 26,2 32,8 ft/s


X G00456

Fig. 1
Y Accuracy ± of rate
X Flow velocity
Pos: 8 /======= Spaltenumbruch ======== @ 0\mod_1132937966324_3101.doc @ 3831

4
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 11.4 /Technische Daten / Datenblatt/Durchfluss/FXE4000/Einbaubedingungen/Elektrodenachse @ 2\mod_1155119847515_3101.doc @ 38596 Pos: 11.10 /Technische Daten / Datenblatt/Durchfluss/FXE4000/Einbaubedingungen/Horizontale Leitungen @ 2\mod_1155120108937_3101.doc @ 38659

2.3.1 Electrode axis 2.3.4 Horizontal connections


Electrode axis (1) should be horizontal if at all possible or no more • Meter tube must always be completely full.
that 45° from horizontal. • Provide for a slight incline of the connection for degassing.


max. 45° G00038

Fig. 5
Pos: 11.11 /==== Leeres Modul mit einer Absatzmarke, DS, 1-spaltig ==== @ 2\mod_1153381574375_0.doc @ 35553

1
Pos: 11.12 /Technische Daten / Datenblatt/Durchfluss/FXE4000/Einbaubedingungen/Freier Ein- bzw. Auslauf @ 2\mod_1155120144781_3101.doc @ 38680

2.3.5 Free inlet or outlet


G00041 • Do not install the flowmeter at the highest point or in the draining-
off side of the pipeline, flowmeter runs empty, air bubbles can
Fig. 2 form (1).
Pos: 11.5 /==== Leeres Modul mit einer Absatzmarke, DS, 1-spaltig ==== @ 2\mod_1153381574375_0.doc @ 35553

• Provide for a siphon fluid intake for free inlets or outlets so that
Pos: 11.6 /Technische Daten / Datenblatt/Durchfluss/FXE4000/Einbaubedingungen/Ein- und Auslaufstrecke (BA + IA) @ 2\mod_1155134311203_3101.doc @ 38877

the pipeline is always full (2).


2.3.2 In- and outlet pipe sections
Straight inlet section Straight outlet section 2
≥ 3 x DN ≥ 2 x DN
DN = Flowmeter sensor size 1
• Do not install fittings, manifolds, valves etc. directly in front of the
meter tube (1).
G00040
• Butterfly valves must be installed so that the valve plate does not
extend into the flowmeter sensor. Fig. 6
• Valves or other turn-off components should be installed in the Pos: 11.13 /==== Leeres Modul mit einer Absatzmarke, DS, 1-spaltig ==== @ 2\mod_1153381574375_0.doc @ 35553

outlet pipe section (2). Pos: 11.14 /Technische Daten / Datenblatt/Durchfluss/FXE4000/Einbaubedingungen/Stark verschmutzte Messstoffe @ 2\mod_1155120187750_3101.doc @ 38701

• For compliance with the measuring accuracy, observe the inlet 2.3.6 Strongly contaminated fluids
and outlet pipe sections.
• For strongly contaminated fluids, a bypass connection according
to the figure is recommended so that operation of the system can
continue to run without interruption the during the mechanical
1 2 cleaning.

3xDN 2xDN
G00037

Fig. 3
Pos: 11.7 /==== Leeres Modul mit einer Absatzmarke, DS, 1-spaltig ==== @ 2\mod_1153381574375_0.doc @ 35553

Pos: 11.8 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Einbaubedingungen/Vertikale Leitungen @ 3\mod_1157631734081_3101.doc @ 39823

2.3.3 Vertical connections G00042

• Vertical installation for measuring abrasive fluids, preferably with Fig. 7


flow in upward direction. Pos: 11.15 /==== Leeres Modul mit einer Absatzmarke, DS, 1-spaltig ==== @ 2\mod_1153381574375_0.doc @ 35553

Pos: 11.16 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Einbaubedingungen/Montage in der Nähe von Pumpen @ 3\mod_1157631958985_3101.doc @ 39844

2.3.7 Installation in the vicinity of pumps


• For flowmeter primaries that are installed near pumps or other
vibration-causing fixtures, the use of mechanical vibration control
components is mandatory.

G00206

Fig. 4
Pos: 11.9 /======= Spaltenumbruch ======== @ 0\mod_1132937966324_3101.doc @ 3831

G00207

Fig. 8
Pos: 11.17 /======= Spaltenumbruch ======== @ 0\mod_1132937966324_3101.doc @ 3831

5
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 11.18 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Einbaubedingungen/Einbau in Rohrleitungen größerer Nennweiten @ 3\mod_1157632081771_3101.doc @ 39865

2.3.8 Installation in larger size pipelines Nomograph for pressure drop calculations
Determine the resulting pressure loss when using reduction For flange transition piece with α/2 = 8°
pieces (1):
1. Calculate the diameter ratio d/D.
2. Determine the flow velocity based on the flow range nomograph
(Fig. 10).
3. Read the pressure drop on the Y-axis in Fig. 10.


D

G00213

Fig. 9
1 = Flange reducer
d = Inside diameter of the flowmeter
V = flow velocity [m/s]
Δp = pressure drop [mbar]
D = Inside diameter of the pipeline

Fig. 10
Pos: 12 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

6
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 13 /==== Wechsel zwei- auf einspaltig ==== @ 0\mod_1130421955859_3101.doc @ 3829


Wechsel ein-auf zweispaltig
Pos: 14 /Überschriften/1.1/1-spaltig/M - O/Nennweite, Nenndruck, Messbereich @ 17\mod_1201168918704_3101.doc @ 154023

2.4 Flowmeter Sizes, Pressure Ratings, Flow Range


Pos: 15 /Technische Daten / Datenblatt/Durchfluss/FXE4000/Messbereich @ 7\mod_1175088848531_3101.doc @ 74665

2.5 Flowmeter Sizes, Pressure Ratings, Flow Range


Meter Size Std. Press. Min. Flow Range Flow Velocity Max. Flow Range Flow Velocity
Rating PN 0 … 0.5 m/s 0 … 10 m/s
3 1/10 40 0 … 0.2 l/min 0.1 US gal/min 0 … 4 l/min 1.1 US gal/min
4 5/32 40 0 … 0.4 l/min 0.1 US gal/min 0 … 8 l/min 2.1 US gal/min
6 1/4 40 0 … 1 l/min 0.3 US gal/min 0 … 20 l/min 5.3 US gal/min
8 5/16 40 0 … 1.5 l/min 0.4 US gal/min 0 … 30 l/min 7.9 US gal/min
10 3/8 40 0 … 2.25 l/min 0.6 US gal/min 0 … 45 l/min 12 US gal/min
15 1/2 40 0 … 5.0 l/min 1.3 US gal/min 0 … 100 l/min 36 US gal/min
20 3/4 40 0 … 7.5 l/min 2.0 US gal/min 0 … 150 l/min 40 US gal/min
25 1 40 0 … 10 l/min 2.6 US gal/min 0 … 200 l/min 53 US gal/min
32 1 1/4 40 0 … 20 l/min 5.3 US gal/min 0 … 400 l/min 106 US gal/min
40 1 1/2 40 0 … 30 l/min 7.9 US gal/min 0 … 600 l/min 159 US gal/min
3 3
50 2 40 0 … 3 m /h 13 US gal/min 0 … 60 m /h 264 US gal/min
3 3
65 2 1/2 40 0 … 6 m /h 26 US gal/min 0 … 120 m /h 528 US gal/min
3 3
80 3 40 0 … 9 m /h 40 US gal/min 0 … 180 m /h 793 US gal/min
3 3
100 4 16 0 … 12 m /h 53 US gal/min 0 … 240 m /h 1057 US gal/min
3 3
125 5 16 0 … 21 m /h 92 US gal/min 0 … 420 m /h 1849 US gal/min
3 3
150 6 16 0 … 30 m /h 132 US gal/min 0 … 600 m /h 2642 US gal/min
3 3
200 8 10/16 0 … 54 m /h 238 US gal/min 0 … 1080 m /h 4755 US gal/min
3 3
250 10 10/16 0 … 90 m /h 396 US gal/min 0 … 1800 m /h 7925 US gal/min
3 3
300 12 10/16 0 … 120 m /h 528 US gal/min 0 … 2400 m /h 10567 US gal/min
3 3
350 14 10/16 0 … 165 m /h 726 US gal/min 0 … 3300 m /h 14529 US gal/min
3 3
400 16 10/16 0 … 225 m /h 991 US gal/min 0 … 4500 m /h 19813 US gal/min
3 3
450 18 10/16 0 … 300 m /h 1321 US gal/min 0 … 6000 m /h 26417 US gal/min
3 3
500 20 10 0 … 330 m /h 1453 US gal/min 0 … 6600 m /h 29059 US gal/min
3 3
600 24 10 0 … 480 m /h 2113 US gal/min 0 … 9600 m /h 30380 US gal/min
3 3
700 28 10 0 … 660 m /h 2906 US gal/min 0 … 13200 m /h 58118 US gal/min
3 3
800 32 10 0 … 900 m /h 3963 US gal/min 0 … 18000 m /h 79252 US gal/min
3 3
900 36 10 0 … 1200 m /h 5283 US gal/min 0 … 24000 m /h 105669 US gal/min
3 3
1000 40 10 0 … 1350 m /h 5944 US gal/min 0 … 27000 m /h 118877 US gal/min
Pos: 16 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

7
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 17 /==== Wechsel ein- auf zweispaltig ==== @ 0\mod_1130421847171_3101.doc @ 3828


Wechsel ein-auf zweispaltig
Pos: 18 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Allgemein/Durchflussnomogramm @ 3\mod_1157633105784_3101.doc @ 39886

Flange design and pressure rating 2.6 Flowrate nomograph


1)
Meter size Material PN PED Flange The volume flowrate is a function of the flow velocity and the
3 ... 25 SS 1.4571 40 bar SEP DIN diameter of the flowmeter. The flowrate nomograph indicates the
(1/10 ... 1") (316 Ti) CL150, CL300 Art.3 ASME flowrate range for a specific flowmeter size and which flowmeter
sizes are suitable for a specific flowrate.
or steel 10 bar para.3 JIS
32 SS 1.4571 40 bar DIN
(1 1/4") (316 Ti) CL150, CL300 ASME l/min m3/h l/s

or steel 10 bar JIS


40 SS 1.4571 40 bar DIN
5 3 8
(1 1/2") (316 Ti) CL150, CL300 ASME 4

Conformity assessment in accordance with category III, module B1 + D, fluid group 1


6
or steel 10 bar JIS 3
2
5
50 SS 1.4571 40 bar DIN 4
2
(2") (316 Ti) CL150, CL300 ASME 102 3
or steel 10 bar JIS 8
2
65 SS 1.4571 16, 40 bar DIN 103 6
8 5
(2 1/2") (316 Ti) CL150, CL300 ASME

0
4

80 10
6 10
or steel 10 bar JIS

DN DN
5 3 8
80 SS 1.4571 40 bar DIN 4
2 6

65
(3") (316 Ti) CL150, CL300 ASME 3 5

DN
or steel 10 bar JIS 4
2
100 SS 1.4571 16, 40 bar DIN 3

50
10
1

DN
(4") (316 Ti) CL150, CL300 ASME 8
2
102

40
or steel 10 bar JIS 6

DN
8 5
125 SS 1.4571 16, 40 bar DIN

32
4
6 1

DN
(5") (316 Ti) CL150, CL300 ASME 5 3 8
or steel 4
6
2

15 20 25
150 DIN SS 1.4571 16, 40 bar 3 5

DN DN DN
4
(6") ASME (316 Ti) CL150, CL300 2
1 3
or steel
8
200 DIN SS 1.4571 10, 16, 25, 40 bar 2
10 6
(8") ASME (316 Ti) CL150, CL300 8 5

10
or steel 6
4 DN
10-1
250 DIN SS 1.4571 10, 16, 25, 40 bar 5 3 8
8
DN

4
(10") ASME (316 Ti) CL150, CL300 2
6
6

3 5
or steel
DN

4
300 DIN SS 1.4571 10, 16, 25, 40 bar 2
3
10-1
(12") ASME (316 Ti) CL150, CL300 8
4

2
or steel
DN

1 6
350 ... 1000 DIN SS 1.4571 10, 16, 25 bar 8 5
4
(14 ... 40") ASME (316 Ti) CL150 10-2
3

6
DN

5 3 8
or steel
4
1) Connecting dimensions for flange acc. to DIN2501 / EN1092-1 or 2 6
2

3 5
ASME or JIS.
DN

4
2
10-2 3
5

Other meter sizes, pressure stages and temperature classes are


1.
DN

available upon request. 8


2
10-1 6
1

5
DN

8
4
6 10-3
5 3 8
4
2 6
3 5
4
2
10-3 3

0,5 0,6 0,8 1 2 3 4 5 6 8 10


m/s
G00225

Fig. 11: Flowrate nomograph DN 1 ... DN 100 (1/25 ... 4")


1 Example

8
Electromagnetic Flowmeter FSM4000 D184S073U02

Example:
l/min m3/h l/s
Flowrate = 7 m3/h (max. value = range end value). Flowmeter sizes
DN 20 ... DN 65 (3/4 ... 2 1/2“) are suitable for a flow velocity between
0.5 ... 10 m/s.
3 Pos: 19 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

4 6
2 5
3
4
2 3
104
8 2
10 5 6
0 0 0 00 0 0 0 0 00

8 5
25 30 35 40 4 5 5 60 70 80 90 10

3
4 10
DN DN DN DN DN D DN DN DN DN DN

6
5 3 8
4 6
3 2 5
4
2 3
103
N

8 2
10 4 6
8 5
0 0

4 2
6 10
5 3 8
4
2 6
0
20

3 5
DN

4
2
102 3
0
12 15
DN

8
5

2
10 3 6
DN

8 5
4
6 10
5 3 8
4
2 6
3 5
4
2
10 3

0,5 0,6 0,8 1 2 3 4 5 6 8 10


m/s G00226

Fig. 12: Flowrate nomograph DN 125 ... DN 1000 (5 ... 40")

9
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 20 /==== Wechsel zwei- auf einspaltig ==== @ 0\mod_1130421955859_3101.doc @ 3829


Wechsel ein-auf zweispaltig
Pos: 21 /Überschriften/1/M - O/Modell SE41F @ 3\mod_1157708869843_3101.doc @ 40854

3 Model SE41F
Pos: 22 /Überschriften/1.1/1-spaltig/S - U/Technische Daten @ 7\mod_1157704677334_3101.doc @ 40581

3.1 Technical data


Pos: 23 /==== Wechsel ein- auf zweispaltig ==== @ 0\mod_1130421847171_3101.doc @ 3828
Wechsel ein-auf zweispaltig
Pos: 24 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Technische Daten/Messaufnehmer/Modell SE41F @ 3\mod_1157634558052_3101.doc @ 39949

Temperature graph Flowmeter sensor material


Parts Standard Options
1
Liner PTFE, PFA, hard/soft −
rubber, ETFE
Signal and ground SS 1.4571 (316 Ti) Hastelloy B-3 (2.4600),
electrode for Hastelloy C-4 (2.4610),
- Hard rubber Titanium, tantalum,
- Soft rubber platinum iridium,
SS 1.4539 (904 L)
2 - PTFE, PFA, Hastelloy C-4 SS 1.4571 (316 Ti)
ETFE (2.4610) Hastelloy B-3 (2.4600),
Titanium, tantalum,
platinum iridium,
3 5
SS 1.4539 (904 L)
4 Grounding plate SS 1.4571 (316 Ti) upon request
6
G00214 Protection plate SS 1.4571 (316 Ti) upon request
Fig. 13: Fluid temperature as function of ambient
temperature Process connection material
1 Ambient temperature °C Parts Standard Options
2 Fluid temperature °C Flanges
3 Standard flange (steel): Hard/soft rubber max. 90/60 °C DN 3 ... DN 15 SS 1.4571 (316 Ti), standard
(194 ... 140 °F)
(1/10 ... 1/2“)
4 with stainless steel flange
DN 20 ... DN 300 Galvanized:
5 Standard flange (steel): PTFE/PFA/ETFE max.130 °C (266°F)
(3/4 ... 10“) DIN/EN: 1.0038/1.0570/1.0460 SS 1.4571
6 High temperature: Thick PTFE / PFA max. 180 °C (356 °F)
ASME: A105/C21 (316 Ti)
DN 350 ... DN 1000 Painted:
Max. allowable cleaning temperature PTFE-, PFA-design
(18 ... 40“) DIN/EN: 1.0038/1.0570/1.0460
CIP-Cleaning Liner Tmax tmax Tamb. ASME: A105/C21
Flowmeter Min.
sensor Housing
Steam cleaning PTFE, PFA 150 °C (302 °F) 60 25 °C (77 °F) DN 3 ... 300 Dual-shell casing Al
Liquid cleaning PTFE, PFA 140 °C (284 °F) 60 25 °C (77 °F) (1/10 ... 10“) casting, painted, paint coat, −
80 µm thick, RAL 9002
If the ambient temperature is > 25 °C, then the difference must be
subtracted from the max. cleaning temperature. DN 350 ... DN 1000 Welded steel design,
Tmax - Δ °C.Δ °C = (Tamb. - 25 °C) (18 ... 40“) painted, paint coat, −
80 µm thick, RAL 9002
Weight Connection box Al alloy, painted,
Refer to dimensional drawings 80 µm dick, frame:

dark gray, RAL 7012
Min. allowable pressure as factor of fluid temperature Cover: light gray, RAL 9002
Liner Nominal size POperation at TOperating Meter tube SS 1.4301 (304) −
DN mbar abs PG-Connector Polyamide −
Hard rubber 15 … 1000 0 < 90 °C (194 °F)
(1/2 … 40“)
Soft rubber 50 … 1000 0 < 60 °C (140 °F)
(2 … 40“)
PTFE 10 … 600 270 < 20 °C (68 °F)
(3/8 … 24“) 400 < 100 °C (212 °F)
500 < 130°C (266 °F)
Thick PTFE 25 … 80 0 < 180 °C (356 °F)
High (1 … 3“)
temperature 100 … 250 67 < 180 °C (356 °F)
Design (4 … 10“)
300 (12“) 27 < 180 °C (356 °F)
PFA 3 … 200 0 < 130°C (266 °F)
(1/10 … 8“) 0 < 180 °C (356 °F)
ETFE 25 ... 1000 100 < 130 °C (266 °F)
(1 ... 40“)

10
Electromagnetic Flowmeter FSM4000 D184S073U02

Storage Temperature
PS PS
-20 ... 70 °C (-4 ... 158 °F) [bar] [psi]
45 652.5

Protection Class per EN 60529 40


PN 40
580.0

IP 67 35 507.5

IP 68 (option) 30 435.0

25 362.5
Pipeline Vibration Following EN 60068-2-6 PN 25
20 290.0
Transmitter
15 217.5
- In the range of 10…55 Hz, max. deflection 0.15 mm PN 16
10 145.0
Flowmeter sensor 5
PN 10
72.5
- In the range of 10…55 Hz, max. deflection 0.15 mm 0 0
- In the range of 10…55 Hz, max. acceleration 2 g -30 -10 10 30 50 70 90 110 130 150 170 190 [°C]
TS
-22 14 50 86 122 158 194 230 266 302 338 374 [°F]
G00215
Designs
The flanged flowmeters comply with the installation lengths specified Fig. 14: DIN flanges SS 1.4571 (316 Ti) to DN 600 (24")
in VDI / VDE 2641, ISO 13359 or according to DVGW (process sheet
W420, design WP; ISO 4064 short).
PS PS
[bar] [psi]
55 797.5
50 725.5
Material load 45
CL 300
652.5

Limits for allowable fluid temperature (TS) and allowable pressure 40 580.0

(PS) are a function of the liner and flange material used (see the 35 507.5

builder's and model plate of the unit). 30 435.0


25 362.5
20 290.0
Temperature limits 15
CL 150
217.5

Liner Flange Min. Max. temperature 10 145.0

material temperature Standard High 5 72.5


0
temperature 0
-30 -10 10 30 50 70 90 110 130 150 170 190 [°C]
-22 14 50 86 122 158 194 230 266 302 338 374 [°F] TS
Hard rubber Steel -10 °C 90 °C −
(14 °F) (194 °F) G00216

SS 1.4571 -15 °C 90 °C − Fig. 15: ASME flange SS 1.4571 (316 Ti) to DN 300 (CL150/300)
(316 Ti) (5°F) (194 °F) to DN 1000 (CL150)
Soft rubber Steel -10 °C 60 °C −
(14 °F) (140 °F)
PS PS
SS 1.4571 -15 °C 60 °C − [bar] [psi]
45 652.5
(316 Ti) (5 °F) (140 °F)
40 580.0
PTFE / PFA / Steel -10 °C 130 °C − PN 40
ETFE (14 °F) (266 °F) 35 507.5

SS 1.4571 -25 °C 130 °C − 30 435.0

(316 Ti) (-13 °F) (266 °F) 25 362.5


PN 25
Thick PTFE / Steel -10 °C 130 °C 180 °C 20 290.0
PFA (14 °F) (266 °F) (356 °F) 15 217.5
PN 16
SS 1.4571 -25 °C 130 °C 180 °C
10 145.0
(316 Ti) (-13 °F) (266 °F) (356 °F) PN 10
5 72.5

0 0
-30 -10 10 30 50 70 90 110 130 150 170 190 [°C]
-22 14 50 86 122 158 194 230 266 302 338 374 [°F] TS

G00217

Fig. 16: DIN flange SS 1.4571 (316 Ti) to DN 600 (24“)

11
Electromagnetic Flowmeter FSM4000 D184S073U02

JIS 10K-B2210 Flange


PS PS
[bar] [psi] Nominal Material PN TS PS
55 797.5
size DN [bar]
50 725.0
CL 300 32 … 300 SS 1.4571 10 -25 … +180 °C 10
45 652.5
(1¼ … 12“) (316 Ti) (-13 … +356 °F)
40 580.0

35 507.5
32 … 300 Steel 10 -10 … +180 °C 10
30 435.0
(1¼ … 12“) (14 … 266 °F)
25 362.5

20 290.0
CL 150
15 217.5 PS PS
[bar] [psi]
10 145.0 17 246.5
5 72.5 16 232.0
DN 700 PN 16
0 0 15 217.5
-30 -10 10 30 50 70 90 110 130 150 170 190 [°C]
-22 14 50 86 122 158 194 230 266 302 338 374 [°F] TS 14 203.0

13 188.5
G00218
DN 900 PN 16
12 DN 800 PN 16 174.0
Fig. 17: ASME flange steel to DN 300 (CL150/300) 11 159.5
DN 1000 PN 16
to DN 1000 (CL150) 10 145.0

9 DN 900 PN10 130.5


DN 800 PN 10
8 DN 700 PN 10 116.0

7 DN 1000 PN 10 101.5
6 87.0
-30 -20 -10 0 10 20 30 40 50 60 70 80 90 [°C]
32 TS
-22 -4 14 50 68 86 104 122 140 158 176 194 [°F]
G00219

Fig. 18: DIN flange SS 1.4571 (316 Ti), DN 700 ... DN 1000
(28 ... 40“)

PS PS
[bar] [psi]
17 246.5

16 232.0

15 217.5

14 203.0

13 DN 700 PN 16 188.5

12 174.0

11 DN 900 PN 16 159.5
DN 800 PN 16
10 145.0
DN 1000 PN 16
9 130.5

8 DN 900 PN 10 116.0
DN 800 PN 10
7 DN 700 PN 10 101.5

6 DN 1000 PN 10 87.0
-10 0 10 20 30 40 50 60 70 80 90 [°C]
TS
-14 32 50 68 86 104 122 140 158 176 194 [°F] G00220

Fig. 19: DIN flange steel, DN 700 ... DN 1000 (28 ... 40“)
Pos: 25 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

12
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 26 /==== Wechsel zwei- auf einspaltig ==== @ 0\mod_1130421955859_3101.doc @ 3829


Wechsel ein-auf zweispaltig
Pos: 27.1 /Überschriften/1.1/1-spaltig/M - O/Maßzeichnungen @ 11\mod_1157707121639_3101.doc @ 40791

3.2 Dimensional drawings


Pos: 27.2 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Maßzeichnungen/Durchflussmesser SE41/Maßzeichnung Flansch DN3-DN300 (DIN) @ 3\mod_1157638227388_3101.doc @ 40075

3.2.1 Model SE41F


Flange acc. to DIN / EN 1092-15), DN 3 … DN 300; ASME B16.5, 1/10 ... 12"

1 88 (3,46) 88 (3,46)
35 (1,38) 35 (1,38)
1

2
ØD2
d2

E
G
E
2
Ød4
ØD

F
F

k
A b k
L* H

G00227

Tolerance: L* DN 3 ... DN 200: +0/-3 (0.12"); DN 250 ... DN 300: +0/-5 (0.20")
Tolerance: L* 1/10 ... 8": +0/-3 (0.12“); 10 ... 12“: +0/-5 (0.20“)
Fig. 20: Dimensions in mm (inch)
1 Cable gland M20 x 1.5 or NPT
2 Number of holes N

13
Electromagnetic Flowmeter FSM4000 D184S073U02

Flange dimensions acc. to DIN / EN Meter dimensions acc. to DIN / EN Wht.


mm mm
DN PN1) kg
D/H d4 b k d2 N A L* L2) L3) E4) F G4)
3…86) 10 … 40 90 40 18 60 14 4 67 130 133 136 101 62 129 2.5
10 10 … 40 90 40 18 60 14 4 67 200 203 206 101 62 129 3
15 10 … 40 95 45 18 65 14 4 67 200 203 206 101 62 129 3
20 10 … 40 105 58 20 75 14 4 87 200 203 206 112 73 140 3.5
25 10 … 40 115 68 20 85 14 4 87 200 203 206 112 73 140 4
32 10 … 40 140 78 20 100 18 4 95 200 203 206 117 78 145 5
40 10 … 40 150 88 20 110 18 4 100 200 203 206 121 82 149 6
50 10 … 40 165 102 21 125 18 4 116 200 203 206 129 90 157 8
655) 10 … 16 185 122 21 145 18 4 100 200 203 206 153 104 171 12
25 … 40 185 122 25 145 18 8 100 200 203 206 153 104 171 12
80 10 … 40 200 138 27 160 18 8 100 200 203 206 159 110 177 16
100 10 … 16 220 158 23 180 18 8 130 250 253 256 179 130 197 15
25 … 40 235 162 27 190 22 8 130 250 253 256 179 130 197 15
125 10 … 16 250 188 25 210 18 8 124 250 255 260 170 127 195 23
25 … 40 270 188 29 220 26 8 124 250 255 260 170 127 195 30
150 10 … 16 285 212 25 240 22 8 170 300 305 310 190 148 216 28
25 … 40 300 218 31 250 26 8 170 300 305 310 190 148 216 34
200 10 340 268 28 295 22 8 195 350 355 360 222 179 248 54
16 340 268 28 295 22 12 195 350 355 360 222 179 248 54
250 10 395 320 30 350 22 12 250 450 455 460 250 207 276 80
16 405 320 30 355 26 12 250 450 455 460 250 207 275 80
300 10 445 370 31 400 22 12 250 500 505 510 293 250 318 83
16 460 378 33 410 26 12 250 500 505 510 293 250 318 92
inch inch
DN PN1) lb
D/H d4 b k d2 N A L* L2) L3) E4) F G4)
3…86) 10 … 40 3.54 1.57 0.71 2.36 0.55 0.16 2.64 5.12 5.24 5.35 3.98 2.44 5.08 5.51
10 10 … 40 3.54 1.57 0.71 2.36 0.55 0.16 2.64 7.87 7.99 8.11 3.98 2.44 5.08 6.61
15 10 … 40 3.74 1.77 0.71 2.56 0.55 0.16 2.64 7.87 7.99 8.11 3.98 2.44 5.08 6.61
20 10 … 40 4.13 2.28 0.79 2.95 0.55 0.16 3.43 7.87 7.99 8.11 4.41 2.87 5.51 7.72
25 10 … 40 4.53 2.68 0.79 3.35 0.55 0.16 3.43 7.87 7.99 8.11 4.41 2.87 5.51 8.82
32 10 … 40 5.51 3.07 0.79 3.94 0.71 0.16 3.74 7.87 7.99 8.11 4.61 3.07 5.71 11.02
40 10 … 40 5.91 3.46 0.79 4.33 0.71 0.16 3.94 7.87 7.99 8.11 4.76 3.23 5.87 13.23
50 10 … 40 6.50 4.02 0.83 4.92 0.71 0.16 4.57 7.87 7.99 8.11 5.08 3.54 6.18 17.64
655) 10 … 16 7.28 4.80 0.83 5.71 0.71 0.16 3.94 7.87 7.99 8.11 6.02 4.09 6.73 26.46
25 … 40 7.28 4.80 0.98 5.71 0.71 0.31 3.94 7.87 7.99 8.11 6.02 4.09 6.73 26.46
80 10 … 40 7.87 5.43 1.06 6.30 0.71 0.31 3.94 7.87 7.99 8.11 6.26 4.33 6.97 35.27
100 10 … 16 8.66 6.22 0.91 7.09 0.71 0.31 5.12 9.84 9.96 10.08 7.05 5.12 7.76 33.07
25 … 40 9.25 6.38 1.06 7.48 0.87 0.31 5.12 9.84 9.96 10.08 7.05 5.12 7.76 33.07
125 10 … 16 9.84 7.40 0.98 8.27 0.71 0.31 4.88 9.84 10.04 10.24 6.69 5.00 7.68 50.71
25 … 40 10.63 7.40 1.14 8.66 1.02 0.31 4.88 9.84 10.04 10.24 6.69 5.00 7.68 66.14
150 10 … 16 11.22 8.35 0.98 9.45 0.87 0.31 6.69 11.81 12.01 12.20 7.48 5.83 8.50 61.73
25 … 40 11.81 8.58 1.22 9.84 1.02 0.31 6.69 11.81 12.01 12.20 7.48 5.83 8.50 74.96
200 10 13.39 10.55 1.10 11.61 0.87 0.31 7.68 13.78 13.98 14.17 8.74 7.05 9.76 119.05
16 13.39 10.55 1.10 11.61 0.87 0.47 7.68 13.78 13.98 14.17 8.74 7.05 9.76 119.05
250 10 15.55 12.60 1.18 13.78 0.87 0.47 9.84 17.72 17.91 18.11 9.84 8.15 10.87 176.37
16 15.94 12.60 1.18 13.98 1.02 0.47 9.84 17.72 17.91 18.11 9.84 8.15 10.83 176.37
300 10 17.52 14.57 1.22 15.75 0.87 0.47 9.84 19.69 19.88 20.08 11.54 9.84 12.52 182.98
16 18.11 14.88 1.30 16.14 1.02 0.47 9.84 19.69 19.88 20.08 11.54 9.84 12.52 202.83
1) Other pressure stages upon request.
2) Standard with a grounding plate mounted on one side of flange in SS 1.4571 (316 Ti). Other materials upon request.
For additional information, see "Grounding". DN 3 ... DN 100 L + 3 mm (0.12"), DN 125 ... DN 300 L + 5 mm (0.20").
3) With protective flanges. The protective flanges provide the grounding function, grounding plate is not required.
DN 3 ... DN 100, DN 125 ... DN 300 L + 10 mm (0.39").
Also available with protective flanges in 1000 series. Installation length ≤ DN 80 L+ 20 mm (0.79"), for DN 100 or higher L+ 25 mm (0.98").
4) For high temperature designs < DN 125 + 20 mm (0.79"), > DN 100 + 120 mm (4.72").
5) Connecting dimensions acc. to EN1092-1. For DN 65 / PN 16 acc. to EN1092-1 please order PN 40.
6) Connection flange DN10.
Pos: 27.3 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

14
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 27.4 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Maßzeichnungen/Durchflussmesser SE41/Maßzeichnung Flansch DN3-DN300 (ASME) @ 3\mod_1157638358417_3101.doc @ 40096

Meter dimensions acc. to ASME Flange dimensions acc. to ASME Flange dimensions acc. to ASME Weight
CL 150 / CL 300 CL 150 CL 300
mm mm mm
A L1) 2) E
3)
F G D/H d4 b k d2 N D/H d4 b k d2 N CL 150/
Inch CL 300
ISO old
kg
13359 Inst.
lngth.
1/10...5/16 67 130 130 101 62 129 89 35 13 60 16 4 95 35 16 67 16 4 2.5
3/8. 1/2 67 200 270 101 62 129 89 35 13 60 16 4 95 35 16 67 16 4 3
3/4 87 200 270 112 73 140 99 43 14 70 16 4 117 43 18 83 19 4 3.5
1 87 200 270 112 73 140 108 51 16 80 16 4 124 51 20 89 19 4 4
1 1/4 95 200 280 117 78 145 117 64 18 89 16 4 134 64 21 99 19 4 5
1 1/2 100 200 280 121 82 149 127 73 20 99 16 4 156 73 23 115 23 4 6
2 116 200 280 129 90 157 152 92 21 121 19 4 165 92 24 127 19 8 8
2 1/2 100 200 330 153 104 171 178 105 25 140 19 4 191 105 27 150 23 8 12
3 100 200 340 159 110 177 191 127 26 153 19 4 210 127 30 168 23 8 16
4 130 250 400 179 130 197 229 157 26 191 19 8 254 157 34 200 22 8 15
5 124 250 450 169 127 195 254 186 28 216 22 8 280 186 39 235 22 8 36/38
6 170 300 450 190 148 216 279 216 30 241 22 8 318 216 41 270 22 12 38
8 195 350 500 222 179 248 343 270 34 298 22 8 381 270 47 330 25 12 66
10 250 450 550 250 207 276 406 324 35 362 25 12 445 324 53 387 28 16 98
12 250 500 620 293 250 319 483 381 37 432 26 12 521 381 56 451 32 16 124/181
inch inch inch
A L1) 2) E
3)
F G D/H d4 b k d2 N D/H d4 b k d2 N CL 150/
Inch CL 300
ISO old
lb
13359 Inst.
lngth.
1/10...5/16 2.64 5.12 5.12 3.98 2.44 5.08 3.50 1.38 0.51 2.36 0.63 0.16 3.74 1.38 0.63 2.64 0.63 0.16 5.5
3/8. 1/2 2.64 7.87 10.63 3.98 2.44 5.08 3.50 1.38 0.51 2.36 0.63 0.16 3.74 1.38 0.63 2.64 0.63 0.16 6.6
3/4 3.43 7.87 10.63 4.41 2.87 5.51 3.90 1.69 0.55 2.76 0.63 0.16 4.61 1.69 0.71 3.27 0.75 0.16 7.7
1 3.43 7.87 10.63 4.41 2.87 5.51 4.25 2.01 0.63 3.15 0.63 0.16 4.88 2.01 0.79 3.50 0.75 0.16 8.8
1 1/4 3.74 7.87 11.02 4.61 3.07 5.71 4.61 2.52 0.71 3.50 0.63 0.16 5.28 2.52 0.83 3.90 0.75 0.16 11.0
1 1/2 3.94 7.87 11.02 4.76 3.23 5.87 5.00 2.87 0.79 3.90 0.63 0.16 6.14 2.87 0.91 4.53 0.91 0.16 13.2
2 4.57 7.87 11.02 5.08 3.54 6.18 5.98 3.62 0.83 4.76 0.75 0.16 6.50 3.62 0.94 5.00 0.75 0.31 17.6
2 1/2 3.94 7.87 12.99 6.02 4.09 6.73 7.01 4.13 0.98 5.51 0.75 0.16 7.52 4.13 1.06 5.91 0.91 0.31 26.5
3 3.94 7.87 13.39 6.26 4.33 6.97 7.52 5.00 1.02 6.02 0.75 0.16 8.27 5.00 1.18 6.61 0.91 0.31 35.3
4 5.12 9.84 15.75 7.05 5.12 7.76 9.02 6.18 1.02 7.52 0.75 0.31 10.00 6.18 1.34 7.87 0.87 0.31 33.1
5 4.88 9.84 17.72 6.65 5.00 7.68 10.00 7.32 1.10 8.50 0.87 0.31 11.02 7.32 1.54 9.25 0.87 0.31 79.4/83.8
6 6.69 11.81 17.72 7.48 5.83 8.50 10.98 8.50 1.18 9.49 0.87 0.31 12.52 8.50 1.61 10.63 0.87 0.47 83.8
8 7.68 13.78 19.69 8.74 7.05 9.76 13.50 10.63 1.34 11.73 0.87 0.31 15.00 10.63 1.85 12.99 0.98 0.47 145.5
10 9.84 17.72 21.65 9.84 8.15 10.87 15.98 12.76 1.38 14.25 0.98 0.47 17.52 12.76 2.09 15.24 1.10 0.63 216.1
12 9.84 19.69 24.41 11.54 9.84 12.56 19.02 15.00 1.46 17.01 1.02 0.47 20.51 15.00 2.20 17.76 1.26 0.63 273.4/399.0
1) With a grounding plate mounted on one side of flange in SS 1.4571 (316 Ti), 1/10 ... 5/16" L + 3 mm (0.12"), 5 ... 12" L + 5 mm (0.20").
Other materials upon request. For additional information. see "Grounding".
2) With protective flanges. The protective flanges provide the grounding function, grounding plate is not required, 1/10 ... 5/16“
L + 6 mm (0.24"), 5 ... 12" L + 10 mm (0.39").
Also available with protective flanges in 1000 series. Installation length ≤ 3" L + 20 mm (0.79"), for 4" or higher L + 25 mm (0.98").
3) For high temperature designs < 5" + 20 mm (0.79"), > 4" + 120 mm (4.72").
4) Connection flange 1/2".
Pos: 27.5 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

15
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 27.6 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Maßzeichnungen/Durchflussmesser SE41/Maßzeichnung Flansch DN350-DN1000 (DIN) @ 3\mod_1157638408849_3101.doc @ 40117

Flange acc. to DIN / EN 1092-14). DN 350 ... DN 1000; ASME B16.5 / ASME B16.47 Series B. 14 ... 40"

1
2

d2
G
E

d4
F

b A k
L* D

G00228

Tolerance: L* DN 350 ... DN 500: +0/-5 (0.20"); DN 600 ... DN 1000: +0/-10 (0.39“)
Tolerance: L* 14 ... 20“: +0/-5 (0.20“); 24 ... 40“: +0/-10 (0.39“)
Fig. 21: Dimensions in mm (inch)
1 Cable gland M20 x 1.5 or NPT
2 Number of holes N

16
Electromagnetic Flowmeter FSM4000 D184S073U02

Flange dimensions acc. to DIN / EN Meter dimensions acc. to DIN / EN Weight


mm mm
DN PN1) appr. kg
D k d4 d2 N b A L L2) L3) G E F
350 10 505 460 430 22 16 31 322 550 555 560 341 313 250 126
350 16 520 470 438 26 16 35 322 550 555 560 341 313 250 140
400 10 565 515 482 26 16 31 370 600 605 610 367 339 275 155
400 16 580 525 490 30 16 37 370 600 605 610 367 339 275 175
500 10 670 620 585 26 20 33 407 650 655 660 403 375 310 191
500 16 715 650 610 33 20 39 407 650 655 660 403 375 310 235
600 10 780 725 685 30 20 33 469 780 785 790 454 426 361 243
600 16 840 770 725 36 20 41 469 780 785 790 454 426 361 302
700 10 895 840 800 30 24 35 537 910 915 920 495 469 405 315
700 16 910 840 795 36 24 41 537 910 915 920 495 469 405 340
800 10 1015 950 905 33 24 37 605 1040 1045 1050 545 519 455 405
800 16 1025 950 900 39 24 43 605 1040 1045 1050 545 519 455 467
900 10 1115 1050 1005 33 28 39 671 1170 1175 1180 595 569 505 483
900 16 1125 1050 1000 39 28 45 671 1170 1175 1180 595 569 505 567
1000 6 1175 1120 1080 30 28 31 739 1300 1305 1310 645 619 555 475
1000 10 1230 1160 1110 36 28 39 739 1300 1305 1310 645 619 555 575
1000 16 1255 1170 1115 42 28 47 739 1300 1305 1310 645 619 555 764
inch inch
DN PN1) appr. lb
D k d4 d2 N b A L L2) L3) G E F
350 10 19.88 18.11 16.93 0.87 0.63 1.22 12.68 21.65 21.85 22.05 13.43 12.32 9.84 277.8
350 16 20.47 18.50 17.24 1.02 0.63 1.38 12.68 21.65 21.85 22.05 13.43 12.32 9.84 308.6
400 10 22.24 20.28 18.98 1.02 0.63 1.22 14.57 23.62 23.82 24.02 14.45 13.35 10.83 341.7
400 16 22.83 20.67 19.29 1.18 0.63 1.46 14.57 23.62 23.82 24.02 14.45 13.35 10.83 385.8
500 10 26.38 24.41 23.03 1.02 0.79 1.30 16.02 25.59 25.79 25.98 15.87 14.76 12.20 421.1
500 16 28.15 25.59 24.02 1.30 0.79 1.54 16.02 25.59 25.79 25.98 15.87 14.76 12.20 518.1
600 10 30.71 28.54 26.97 1.18 0.79 1.30 18.46 30.71 30.91 31.10 17.87 16.77 14.21 535.7
600 16 33.07 30.31 28.54 1.42 0.79 1.61 18.46 30.71 30.91 31.10 17.87 16.77 14.21 665.8
700 10 35.24 33.07 31.50 1.18 0.94 1.38 21.14 35.83 36.02 36.22 19.49 18.46 15.94 694.5
700 16 35.83 33.07 31.30 1.42 0.94 1.61 21.14 35.83 36.02 36.22 19.49 18.46 15.94 749.6
800 10 39.96 37.40 35.63 1.30 0.94 1.46 23.82 40.94 41.14 41.34 21.46 20.43 17.91 892.9
800 16 40.35 37.40 35.43 1.54 0.94 1.69 23.82 40.94 41.14 41.34 21.46 20.43 17.91 1.029.6
900 10 43.90 41.34 39.57 1.30 1.10 1.54 26.42 46.06 46.26 46.46 23.43 22.40 19.88 1.064.8
900 16 44.29 41.34 39.37 1.54 1.10 1.77 26.42 46.06 46.26 46.46 23.43 22.40 19.88 1.250.0
1000 6 46.26 44.09 42.52 1.18 1.10 1.22 29.09 51.18 51.38 51.57 25.39 24.37 21.85 1.047.2
1000 10 48.43 45.67 43.70 1.42 1.10 1.54 29.09 51.18 51.38 51.57 25.39 24.37 21.85 1.267.7
1000 16 49.41 46.06 43.90 1.65 1.10 1.85 29.09 51.18 51.38 51.57 25.39 24.37 21.85 1.684.3
1) Other pressure stages upon request.
2) Grounding plate mounted on one side of flange in SS 1.4571 (316 Ti). Other materials upon request, L + 5 mm (0.20").
3) With protection plates attached to flange on both sides. Grounding plate is not required L + 10 mm (0.39").
Also available with protective flanges in 1000 series. Installation length ≤ DN 80 L + 20 mm (0.79"), for DN 100 or higher L + 25 mm (0.98").
4) Connecting dimensions acc. to EN1092-1.
Pos: 27.7 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

17
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 27.8 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Maßzeichnungen/Durchflussmesser SE41/Maßzeichnung Flansch DN350-DN1000 (ASME) @ 3\mod_1157638448817_3101.doc @ 40138

Meter dimensions acc. to ASME CL 150 Flange dimensions acc. to ASME CL 1501) Weight
mm mm
A L2) 3) E F G D k d4 d2 N b
Inch appr. kg
ISO old
13359 Install.
length
14 322 550 650 313 250 341 534 476 413 28.6 12 40 178
16 370 600 700 329 275 367 597 540 470 28.6 16 42 225
18 407 686 – 375 310 403 635 578 533 31.7 16 45 230
20 416 762 780 375 310 403 699 635 584 31.7 20 48 263
24 469 914 850 423 361 454 813 749 692 34.9 20 53 327
28 537 – 910 469 405 495 837 749 762 34.9 20 50 385
32 605 – 1040 519 455 545 942 900 864 22.2 48 51 513
36 671 – 1170 569 505 595 1057 1010 972 25.4 44 57 662
40 739 – 1300 619 555 645 1380 1210 1140 56.0 28 63 1132
inch inch
A L2) 3) E F G D k d4 d2 N b
Inch appr. lb
ISO old
13359 Install.
length
14 12.68 21.65 25.59 12.32 9.84 13.43 21.02 18.74 16.26 1.13 0.47 1.57 392.4
16 14.57 23.62 27.56 12.95 10.83 14.45 23.50 21.26 18.50 1.13 0.63 1.65 496.0
18 16.02 27.01 – 14.76 12.20 15.87 25.00 22.76 20.98 1.25 0.63 1.77 507.1
20 16.38 30.00 30.71 14.76 12.20 15.87 27.52 25.00 22.99 1.25 0.79 1.89 579.8
24 18.46 35.98 33.46 16.65 14.21 17.87 32.01 29.49 27.24 1.37 0.79 2.09 720.9
28 21.14 – 35.83 18.46 15.94 19.49 32.95 29.49 30.00 1.37 0.79 1.97 848.8
32 23.82 – 40.94 20.43 17.91 21.46 37.09 35.43 34.02 0.87 1.89 2.01 1.131.0
36 26.42 – 46.06 22.40 19.88 23.43 41.61 39.76 38.27 1.00 1.73 2.24 1.459.5
40 29.09 – 51.18 24.37 21.85 25.39 54.33 47.64 44.88 2.20 1.10 2.48 2.495.6
1) Other pressure stages upon request. Dimensions 28" or larger acc. to ASME B16.47 Series B.
2) Grounding plate mounted on one side of flange in SS 1.4571 (316 Ti). Other materials upon request, L + 5 mm (0.20").
3) With protection plates attached to flange on both sides. Grounding plate is not required L + 10 mm (0.39").
Also available with protective flanges in 1000 series. Installation length ≤ 3" L + 20 mm (0.79"), for 4" or higher L + 25 mm (0.98").
Pos: 28 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

18
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 29 /Überschriften/1.1/1-spaltig/A - C/Bestellinformationen @ 2\mod_1139828409218_3101.doc @ 3181

3.3 Ordering information


Pos: 30 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Bestellinformationen/SE41F @ 3\mod_1157640360285_3101.doc @ 40348

Information on 3A conformity: If 3A conformity is desired, please order model SE21


Fixed flange, liner thick PTFE, PTFE, PFA, ETFE
Electromagnetic Flowmeter Variant digit No. 1-6 7 8 9 10 11 12 13 14
FSM4000-SE41F Catalog No. SE41F-
remote sensor with welded flange
Liner Material Meter Size
PFA DN 3 1/10 in. P 0 3
PFA DN 4 5/32 in. P 0 4
PFA DN 6 1/4 in. P 0 6
PFA DN 8 5/16 in. P 0 8
PTFE DN 10 3/8 in. T 1 0
PTFE DN 15 1/2 in. T 1 5
PTFE DN 20 3/4 in. T 2 0
PTFE DN 25 1 in. T 2 5
PTFE DN 32 1-1/4 in. T 3 2
PTFE DN 40 1-1/2 in. T 4 0
PTFE DN 50 2 in. T 5 0
PTFE DN 65 2-1/2 in. T 6 5
PTFE DN 80 3 in. T 8 0
PTFE DN 100 4 in. T 1 H
PTFE DN 125 5 in. T 1 Q
PTFE DN 150 6 in. T 1 F
PTFE DN 200 8 in. T 2 H
PTFE DN 250 10 in. T 2 F
PTFE DN 300 12 in. T 3 H
PTFE DN 350 14 in. T 3 F
PTFE DN 400 16 in. T 4 H
PTFE DN 450 18 in. (only ASME CL 150) T 4 F
PTFE DN 500 20 in. T 5 H
PTFE DN 600 24 in. T 6 H
Thick-PTFE DN 25 1 in. F 2 5
Thick-PTFE DN 32 1-1/4 in. F 3 2
Thick-PTFE DN 40 1-1/2 in. F 4 0
Thick-PTFE DN 50 2 in. F 5 0
Thick-PTFE DN 65 2-1/2 in. F 6 5
Thick-PTFE DN 80 3 in. F 8 0
Thick-PTFE DN 100 4 in. F 1 H
Thick-PTFE DN 125 5 in. F 1 Q
Thick-PTFE DN 150 6 in. F 1 F
Thick-PTFE DN 200 8 in. F 2 H
Thick-PTFE DN 250 10 in. F 2 F
Thick-PTFE DN 300 12 in. F 3 H

Continued on next page


1) Grounding electrodes available only for meter size DN 3 ... DN 300

19
Electromagnetic Flowmeter FSM4000 D184S073U02

Electromagnetic Flowmeter Variant digit No. 1-6 7 8 9 10 11 12 13 14


FSM4000-SE41F Catalog No. SE41F-
remote sensor with welded flange
Liner Material Meter Size
ETFE DN 25 1 in. E 2 5
ETFE DN 32 1-1/4 in. E 3 2
ETFE DN 40 1-1/2 in. E 4 0
ETFE DN 50 2 in. E 5 0
ETFE DN 65 2-1/2 in. E 6 5
ETFE DN 80 3 in. E 8 0
ETFE DN 100 4 in. E 1 H
ETFE DN 125 5 in. E 1 Q
ETFE DN 150 6 in. E 1 F
ETFE DN 200 8 in. E 2 H
ETFE DN 250 10 in. E 2 F
ETFE DN 300 12 in. E 3 H
ETFE DN 350 14 in. E 3 F
ETFE DN 400 16 in. E 4 H
ETFE DN 450 18 in. (only ASME CL 150) E 4 F
ETFE DN 500 20 in. E 5 H
ETFE DN 600 24 in. E 6 H
ETFE DN 700 28 in. E 7 H
ETFE DN 800 32 in. E 8 H
ETFE DN 900 36 in. E 9 H
ETFE DN 1000 40 in. E 1 T
Measuring Electrodes Material Grounding Electrodes 1)
Hastelloy C-4 (2.4610) Without H
Hastelloy B-3 (2.4600) Without B
SST 1.4571 (316Ti) Without S
Titanium Without M
Tantalum Without T
SST 1.4539 (904L) Without F
Platinum-Iridium Without P
Hastelloy C-4 (2.4610) With O
Hastelloy B-3 (2.4600) With N
SST 1.4571 (316Ti) With E
Titanium With I
Tantalum With Q
SST 1.4539 (904L) With R
Platinum-Iridium With G

Continued on next page


1) Grounding electrodes available only for meter size DN 3 ... DN 300

20
Electromagnetic Flowmeter FSM4000 D184S073U02

Electromagnetic Flowmeter Variant digit No. 1-6 11 12 13 14 15 16 17 18


FSM4000-SE41F Catalog No. SE41F-
remote sensor with welded flange
Pressure Rating 2)
PN 10 ISO laid length C
PN 16 ISO laid length 3) D
PN 25 ISO laid length E
PN 40 Standard for DN 3 ... DN 80 ISO laid length F
JIS K10 ISO laid length K
ASME CL 150 <= DN 600 old laid length, only for replacement P
ASME CL 300 <= DN 300 old laid length, only for replacement Q
ASME CL 150 <= DN 600 ISO laid length R
ASME CL 300 <= DN 400 ISO laid length S
Material of Process Connection
Steel from DN 20 1
SST 1.4571 (316Ti) Standard for DN 3 ... DN 15, option from DN 20 3
Accessories
Without A
Protection rings 1.4571 (316Ti) 4) B
Grounding ring 1.4571 (316Ti) 4) C
Protection flanges 1.4571 (316Ti) 5) D
Temperature Range
Standard design <= 130 °C (266 °F) S
High temperature design <= 180 °C (356 °F) only with Thick-PTFE DN 25 ... DN 100 H
High temperature design <= 180 °C (356 °F) only with Thick-PTFE DN 125 ... DN 300 H
Certificates 6)
PED Standard A
Material certificate per EN 10204 3.1 + pressure test per AD 2000 D
Pressure test per AD 2000 G
Inspection certificate per EN 10204 3.1 F
Calibration Certificates
Standard A
Fingerprint saved at factory > DN 8 F
Others Z
Type of Protection Cable Entry
IP 67 Thread for screw-type conduit fitting M20 x 1.5 2
IP 67 Thread for screw-type conduit fitting NPT 1/2 in. 4
IP 68 Thread for screw-type conduit fitting M20 x 1.5 7) 7
IP 68 Cable connected and Connection box potted 8
Design
Without pre-amplifier from DN 10 >= 20 μS/cm 1
With pre-amplifier type A 2
Standard up to DN 8 >= 20 μS/cm, option > DN 8 >= 5 μS/cm and laid length max. 200 m
With pre-amplifier type B 4
Option up to DN 8 >= 5 µS/cm, > DN 8 >= 0,5 µS/cm and laid length max. 200 m

Continued on next page


2) Dimensions of the flanges per DIN 2501 / EN 1092-1, or ASME B16.5 or JISB2210-10K
3) DN 65 / PN 16 with connection dimensions acc. EN 1092-1: Please order PN 40
4) Protection rings mounted on both sides and grounding ring
mounted on one side to flange, SST 1.4571 (316Ti)
5) Only with laid length 1000 series, ASME / DIN flanges,
installation length <= DN 80: L + 20 mm, >= DN 100: L + 25 mm
6) Material certificate for meter tube and flanges
7) Sealing compound (optionally) D141B038U01

21
Electromagnetic Flowmeter FSM4000 D184S073U02

Electromagnetic Flowmeter Variant digit No. 1-6 15 16 17 18 19 20 21 22


FSM4000-SE41F Catalog No. SE41F-
remote sensor with welded flange
Name Plate
German adhesive foil G
English adhesive foil E
French adhesive foil F
German SST plate J
English SST plate K
French SST plate L
Design Level (Specified by ABB) *
Laid Length
Standard A
Electrode Design
Standard 1
Conical head, SST 1.4539 (904L) (>= DN 10) 8) 2
Swedish design, material 1.4571 (316Ti) (DN 50 ... DN 100) 9) 3
Swedish design, material 1.4571 (316Ti) (DN 125 ... DN 250) 9) 3
Swedish design, material Hastelloy C-4 (DN 50 ... DN 100) 9) 3
Swedish design, material Hastelloy C-4 (DN 125 ... DN 250) 9) 3

8) For application e.g. with high fat contents


9) For pulp applications with high resin contents

Related Transmitter FSM4000-S4

22
Electromagnetic Flowmeter FSM4000 D184S073U02

Fixed flange, liner hard/soft rubber


Electromagnetic Flowmeter Variant digit No. 1-6 7 8 9 10 11 12 13 14
FSM4000-SE41F Catalog No. SE41F-
remote sensor with welded flange
Liner Material Meter Size
Hard rubber DN 15 1/2 in. H 1 5
Hard rubber DN 20 3/4 in. H 2 0
Hard rubber DN 25 1 in. H 2 5
Hard rubber DN 32 1-1/4 in. H 3 2
Hard rubber DN 40 1-1/2 in. H 4 0
Hard rubber DN 50 2 in. H 5 0
Hard rubber DN 65 2-1/2 in. H 6 5
Hard rubber DN 80 3 in. H 8 0
Hard rubber DN 100 4 in. H 1 H
Hard rubber DN 125 5 in. H 1 Q
Hard rubber DN 150 6 in. H 1 F
Hard rubber DN 200 8 in. H 2 H
Hard rubber DN 250 10 in. H 2 F
Hard rubber DN 300 12 in. H 3 H
Hard rubber DN 350 14 in. H 3 F
Hard rubber DN 400 16 in. H 4 H
Hard rubber DN 450 18 in. (only ASME CL 150) H 4 F
Hard rubber DN 500 20 in. H 5 H
Hard rubber DN 600 24 in. H 6 H
Hard rubber DN 700 28 in. H 7 H
Hard rubber DN 800 32 in. H 8 H
Hard rubber DN 900 36 in. H 9 H
Hard rubber DN 1000 40 in. H 1 T
Soft rubber DN 50 2 in. S 5 0
Soft rubber DN 65 2-1/2 in. S 6 5
Soft rubber DN 80 3 in. S 8 0
Soft rubber DN 100 4 in. S 1 H
Soft rubber DN 125 5 in. S 1 Q
Soft rubber DN 150 6 in. S 1 F
Soft rubber DN 200 8 in. S 2 H
Soft rubber DN 250 10 in. S 2 F
Soft rubber DN 300 12 in. S 3 H
Soft rubber DN 350 14 in. S 3 F
Soft rubber DN 400 16 in. S 4 H
Soft rubber DN 450 18 in. (only ASME CL 150) S 4 F
Soft rubber DN 500 20 in. S 5 H
Soft rubber DN 600 24 in. S 6 H
Soft rubber DN 700 28 in. S 7 H
Soft rubber DN 800 32 in. S 8 H
Soft rubber DN 900 36 in. S 9 H
Soft rubber DN 1000 40 in. S 1 T

Continued on next page

23
Electromagnetic Flowmeter FSM4000 D184S073U02

Electromagnetic Flowmeter Variant digit No. 1-6 10 11 12 13 14 15 16 17


FSM4000-SE41F Catalog No. SE41F-
remote sensor with welded flange
Measuring Electrodes Material Grounding Electrodes 1)
SST 1.4571 (316Ti) Without S
Hastelloy C-4 (2.4610) Without H
Hastelloy B-3 (2.4600) Without B
Titanium Without M
Tantalum Without T
SST 1.4539 (904L) Without F
Platinum-Iridium Without P
Hastelloy C-4 (2.4610) With O
Hastelloy B-3 (2.4600) With N
SST 1.4571 (316Ti) With E
Titanium With I
Tantalum With Q
SST 1.4539 (904L) With R
Platinum-Iridium With G
Pressure Rating 2)
PN 10 ISO laid length C
PN 16 ISO laid length 3) D
PN 25 ISO laid length E
PN 40 Standard for DN 3 ... DN 80 ISO laid length F
JIS K10 ISO laid length K
ASME CL 150 <= DN 600 old laid length, only for replacement P
ASME CL 300 <= DN 300 old laid length, only for replacement Q
ASME CL 150 <= DN 600 ISO laid length R
ASME CL 300 <= DN 400 ISO laid length S
Material of Process Connection
Steel from DN 20 1
SST 1.4571 (316Ti) Standard for DN 15, option from DN 20 3
Accessories
Without A
Temperature Range
Standard design <= 130 °C (266 °F) S
High temperature design <= 180 °C (356 °F) (only for DN 25 ... DN 300) H
Certificates
PED Standard A
Material certificate per EN 10204 3.1 + pressure test per AD 2000 D
Pressure test per AD 2000 G
Inspection certificate per EN 10204 3.1 F
Calibration Certificates
Standard A
Fingerprint saved at factory > DN 8 F
Others Z
Type of Protection Cable Entry
IP 67 Thread for screw-type conduit fitting M20 x 1.5 2
IP 67 Thread for screw-type conduit fitting NPT 1/2 in. 4
IP 68 Thread for screw-type conduit fitting M20 x 1.5 4) 7
IP 68 Cable connected and Connection box potted 8

Continued on next page


1) Grounding electrodes available only for meter size DN 15 ... DN 100.
Instruments equipped with hard rubber liner (DN 125 ... DN 1000) incorporate a conductive element
that grounds the fluid. Grounding electrodes are not required.
2) Connection dimensions for flanges acc. to DIN 2501 / EN 1092-1 or ASME B16.5 or JIS B2210-10K
3) DN 65 / PN 16 with connection dimensions acc. EN 1092-1: Please order PN 40
4) Sealing compound (optionally) D141B038U01

24
Electromagnetic Flowmeter FSM4000 D184S073U02

Electromagnetic Flowmeter Variant digit No. 1-6 15 16 17 18 19 20 21 22


FSM4000-SE41F Catalog No. SE41F-
remote sensor with welded flange
Design
Without pre-amplifier >= 20 μS/cm 1
With pre-amplifier type A 2
Option >= 5 μS/cm and laid length max. 200 m
With pre-amplifier type B 4
Option >= 0,5 µS/cm and laid length max. 200 m
Name Plate
German adhesive foil G
English adhesive foil E
French adhesive foil F
German SST plate J
English SST plate K
French SST plate L
Design Level (Specified by ABB) *
Laid Length
Standard A
Electrode Design
Standard 1
Conical head, SST 1.4539 (904L) 5) 2

5) For application e.g. with high fat contents

Related Transmitter FSM4000-S4


Pos: 31 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

25
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 32 /Überschriften/1/M - O/Modell SE21_ @ 3\mod_1157709057953_3101.doc @ 40875

4 Model SE21_
Pos: 33 /Überschriften/1.1/1-spaltig/S - U/Technische Daten @ 7\mod_1157704677334_3101.doc @ 40581

4.1 Technical data


Pos: 34 /==== Wechsel ein- auf zweispaltig ==== @ 0\mod_1130421847171_3101.doc @ 3828
Wechsel ein-auf zweispaltig
Pos: 35 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Technische Daten/Messaufnehmer/Modell SE21_ @ 3\mod_1157640761823_3101.doc @ 40390

Minimum Allowable Absolute Pressure Flowmeter material


Liner Nominal size POperation at TOperating* Lining Electrode material Electrode design
mbar abs DN Standard Options Standard Options
°C
PFA, Hast.-C4 Hast.-B3 Flat Pointed
PFA 3 … 100 0 ≤ 130 °C
Peek, (2.4610) (2.4600) head head
(1/10 … 4“) (266 °F)
Torlon (1.4539 SS 1.4539 (≥ DN 10)
Peek/Torlon 1 …2 0 ≤ 120 °C
[904 L] (904 L) SS 1.4539
(1/25 … 1/12“) (248 °F)
for pipe SS 1.4571 (904 L)
1) Higher temperatures are allowed for CIP/SIP cleaning for limited (316 Ti)
time periods, see Table "Maximum Allowable Cleaning conn. and
Temperature". Tri-Clamp) Titanium,
tantalum,
platinum-
iridium
Maximum Allowable Cleaning Temperature
CIP-Cleaning Liner Tmax Tmax Tamb. 1/8" sanitary connectors always with 2 grounding electrodes in
material for signal electrodes, standard.
Minutes
Steam cleaning 150 °C 60 25 °C PFA / Peek
(302 °F) (77 °F) Process Connection Materials
Liquid cleaning PFA / Peek / 140 °C 60 25 °C
Process Connection Standard Option
Torlon (284 °F) (77 °F)
Flanges SS 1.4571 (316 Ti) on request
If the ambient temperature is > 25 °C (77 °F), then the difference
Wafer design None
must be subtracted from the max. cleaning temperature.
Tmax – Δ °C, Δ °C = (TAmb - 25 °C) Weld stubs SS 1.4404 (316 L) on request
Threaded pipe connection SS 1.4404 (316 L) on request
Tri-Clamp SS 1.4404 (316 L) on request
Maximum Allowable Temperature Shock External threads SS 1.4404 (316 L) on request
Liner Temp Shock max. Temp. gradient 1/8" sanitary connectors SS 1.4571 (316 Ti) POM, brass,
Temp. Diff. °C °C/min PVC
PFA any any Connection box
Peek, Torlon any any - without/with SS 1.4301 (304) –
preamplifier, type A
- with preamplifier, Al alloy, painted, paint –
type B coat frame: dark gray,
Maximum allowable ambient temperature as function of
RAL 7012 cover: light
fluid temperature gray, RAL 9002
Tamb °C (°F) Meter tube SS 1.4301 (304) –
Cable gland Polyamide PVDF
60 (140) Flowmeter sensor SS 1.4301 (304) –
housing
DN 1 ... DN 2

40 (104)
Gasket material (internal)
Process Standard Option
TMeasuring Connection
-40 -10 90 (194) 120 130 Wafer design None –
(14F) (248) (266) °C (°F)
-10 DN 1 ... DN 2 Weld stubs EPDM (Ethylene- Silicon with FDA
(14) Propylene) with approval (optional,
Threaded pipe
connection FDA approval, resistant to oils and
-25 (-13) G00224 silicone with FDA grease)
Tri-Clamp
approval (CIP- PTFE with
Fig. 22: Temperature graph External threads resistant, no oils or FDA approval
grease) (DN 3 ... 8)
The flowmeter sensor may not be isolated
1/8" sanitary PTFE Viton (only in
connector combination with
Fluid temperature DN 1 … DN 2 (1/25 ... 1/12") PVC process
-10 … 120 °C (14 ... 248 °F), max. allowable cleaning temperature, connection)
see table Flat housing gaskets Silicon (resistant –
to oil, grease)

26
Electromagnetic Flowmeter FSM4000 D184S073U02

Storage Temperature Material load for flange model SE21F


-25 … 70 °C (-13 ... 158 °F) Lining: PFA
PS PS
Protection Class per EN 60529 [bar] [psi]
45 652.5
• IP 67
• IP 68 (option) 40
PN 40
580.0

35 507.5

Pipeline Vibration Following EN 60068-2-6 30 435.0


Transmitter
25 362.5
- In the range of 10…55 Hz, max. deflection 0.15 mm PN 25

Flowmeter sensor 20 290.0

- In the range of 10…55 Hz, max. deflection 0.15 mm 15


PN 16
217.5

- In the range of 55…150 Hz, max. acceleration 2 g 10


PN 10
145.0

5 72.5

0 0

Materialload for meters with variable process -30 -20 -10 0


-22 -4 14 32
10 20 30 40 50 60 70 80 90 100 110 120 130 *) [°C]
50 68 86 104 122 140 158 176 194 212 230 248 266 *) [°F] TS G00221
connections / wafer flange SE21 DN 1 ... DN 100
Fig. 23: DIN-Flanges SS 1.4571 (316 Ti) to DN 100 (4")
(1/25 ... 4")
Process connect. Nominal size PSmax TSmin TSmax Liner PFA
Liner PFA DN [bar] PS PS
[bar] [psi]
Wafer design 3 … 50 40 55 697.5
(1/10 … 2“) -25 °C 130 °C
50 725.0
65 … 100 16 (-13 °F) (266 °F) CL 300
(2 1/2 … 4“) 45 652.5

Weld stubs 3 … 40 40 40 580.0

(1/10 … 1 1/2“) 35 507.5

50; 80 16 -25 °C 130 °C 30 435.0


(2“, 3“) (-13 °F) (266 °F) 25 362.5
65, 100 10 20 290.0
(2 1/2 … 4“) 15
CL 150
217.5
Threaded pipe 3 … 40 40 10 145.0
connection (1/10 … 1 1/2“)
5 72.5
DIN 11851 50; 80 16 -25 °C 130 °C
0
(2“, 3“) (-13 °F) (266 °F) 0
-30 -10 10 30 50 70 90 110 130 *) [°C]
65, 100 10 -22 14 50 86 122 158 194 230 266 *) [°F] TS
G00222
(2 1/2 … 4“)
3 … 50 16 Fig. 24: ASME flange SS 1.4571 (316 Ti) to DN 100 (4")
Tri-clamp acc. to
DIN 32676 (1/10 … 2“) -25 °C 121 °C
65 … 100 10 (-13 °F) (250 °F)
(2 1/2 … 4“) JIS B2210-10K wafer flange
Tri-Clamp per 3 ... 100 10 -25 °C 130 °C
Meter size Material PN TS PS
ASME BPE 1 (1/10 ... 4“) (-13 °F) (266 °F)
DN [bar]
External threads 3 … 25 16 -25 °C 130 °C
32 … 100 SS 1.4404 (316 L)
ISO 228 (1/10 … 1“ (-13 °F) (266 °F) -40 … 130 °C
(1¼ … 4“) SS 1.4435 (316 L) 10 10
1/8" sanitary 1…2 10 -10 °C 120 °C (-40 … 266 °F)
(1/25 … 1/12“) (14 °F) (248 °F) SS 1.4301 (304)
connector

Material load for flange model SE21W


Lining: PFA wafer flange
PS PS
[bar] [psi]
45 652.5

40 580.0

35 507.5
PN 40/CL 300
30 435.0

25 362.5

20 PN 25 290.0

15 217.5
PN 16/CL 150
10 145.0
PN 10
5 72.5

0 0
-60 -50 -40 -30 -20 -10 0 10 20 30 40 50 60 70 80 90 100 110 120 130 *) [°C]
TS
-76 -58 -40 -22 -4 14 32 50 68 86 140 122 140158 176194 212 230 248 266 *) [°F] G00223

Fig. 25
Higher temperatures are allowed for CIP/SIP cleaning for limited time
periods, see Table "Maximum Allowable Cleaning Temperature".
Pos: 36 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

27
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 37 /==== Wechsel zwei- auf einspaltig ==== @ 0\mod_1130421955859_3101.doc @ 3829


Wechsel ein-auf zweispaltig
Pos: 38.1 /Überschriften/1.1/1-spaltig/M - O/Maßzeichnungen @ 11\mod_1157707121639_3101.doc @ 40791

4.2 Dimensional drawings


Pos: 38.2 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Maßzeichnungen/Durchflussmesser SE21/Maßzeichnung Flansch SE21F, DN3-DN40 @ 3\mod_1157639602746_3101.doc @ 40243

4.2.1 Model SE21F


Stainless steel housing, flange acc. to DIN / EN 1092-1, DN 3 ... DN 40; ASME B16.5, 1/10 ... 1 1/2"

104 (4,09) 80 (3,15)


41 (1,61)
1
71 (2,80)
G

G
2

Ø
d2
E

E
Ød4

ØD
F

F
A b
k
L D
G00229

Fig. 26: Dimensions in mm (inch)


1 Cable gland M20 x 1.5 or NPT
2 Number of holes N

Flange dimensions acc. to DIN / EN, liner PFA Weight


mm
DN PN appr. kg
L1) A D k d4 d2 b E F G
3 ... 83) 130 37 90 60 42 14 18 62 39 133 2.5
10, 15 200 37 90, 95 65 36 14 18 62 39 133 2.5
20 200 42 105 75 41 14 20 66 43 137 2.5
10 ... 40
25 200 54 115 85 54 14 20 73 48 144 3
32 200 62 140 100 64 14 20 78 53 149 4
40 200 67 150 110 74 14 20 82 57 153 4.5
inch
DN PN appr. lb
L1) A D k d4 d2 b E F G
3 ... 83) 5.12 1.46 3.54 2.36 1.65 0.55 0.71 2.44 1.54 5.24 5.5
10, 15 7.7 1.46 3.54, 3.74 2.56 1.42 0.55 0.71 2.44 1.54 5.24 5.5
20 7.87 1.65 4.13 2.95 1.61 0.55 0.79 2.60 1.69 5.39 5.5
10 ... 40
25 7.87 2.13 4.53 3.35 2.13 0.55 0.79 2.87 1.89 5.67 6.6
32 7.87 2.44 5.51 3.94 2.52 0.55 0.79 3.07 2.09 5.87 8.8
40 7.87 2.64 5.91 4.33 2.91 0.55 0.79 3.23 2.24 6.02 9.9
1) If a grounding plate is required, L + 3 mm (0.12"). With 2 grounding plates (protection plates) L + 6 mm (0.24"), material upon request.
2)
Connecting dimensions acc. to EN1092-1
3) Connection flange DN 10 (3/8")
4) Connection flange 1/2"

28
Electromagnetic Flowmeter FSM4000 D184S073U02

Flange dimensions acc. to ASME B16.5, liner PFA Weight


mm
Inch PN appr. kg
L1) A D k d4 d2 b E F G
1/10 ... 5/164) 130 37 88.9 60.3 42.0 15.9 18.0 62 39 133 2.5
3/84), 1/2 200 37 88.9 60.3 34.8 15.9 12.6 62 39 133 2.5
3/4 CL 150 200 42 98.4 69.8 42.9 15.9 14.2 66 43 137 2.5
1 200 54 108.0 79.2 50.8 15.9 15.8 73 48 144 3
1 1/4 200 62 117.5 88.9 63.5 15.9 17.4 78 53 149 3
1 1/2 200 67 127.0 98.6 73.0 15.9 19.0 82 57 153 3.5
1/10 ... 5/164) 130 37 95.2 66.7 42.0 15.9 18.0 62 39 133 2.5
3/84), 1/2 200 37 95.2 66.7 34.8 15.9 15.8 62 39 133 2.5
3/4 CL 300 200 42 117.5 82.5 42.9 19.0 17.4 66 43 137 2.5
1 200 54 123.8 88.9 50.8 19.0 19.0 73 48 144 3
1 1/4 200 62 133.3 98.4 63.5 19.0 20.5 78 53 149 4
1 1/2 200 67 155.6 114.3 73.0 22.2 22.1 82 57 153 4.5
inch
Inch PN appr. lb
L1) A D k d4 d2 b E F G
1/10 ... 5/164) 5.12 1.46 3.50 2.37 1.65 0.63 0.71 2.44 1.54 5.24 5.5
3/84), 1/2 7.87 1.46 3.50 2.37 1.37 0.63 0.50 2.44 1.54 5.24 5.5
3/4 CL 150 7.87 1.65 3.87 2.75 1.69 0.63 0.56 2.60 1.69 5.39 5.5
1 7.87 2.13 4.25 3.12 2.00 0.63 0.62 2.87 1.89 5.67 6.6
1 1/4 7.87 2.44 4.63 3.50 2.50 0.63 0.69 3.07 2.09 5.87 6.6
1 1/2 7.87 2.64 5.00 3.88 2.87 0.63 0.75 3.23 2.24 6.02 7.7
1/10 ... 5/164) 5.12 1.46 3.75 2.63 1.65 0.63 0.71 2.44 1.54 5.24 5.5
3/84), 1/2 7.87 1.46 3.75 2.63 1.37 0.63 0.62 2.44 1.54 5.24 5.5
3/4 CL 300 7.87 1.65 4.63 3.25 1.69 0.75 0.69 2.60 1.69 5.39 5.5
1 7.87 2.13 4.87 3.50 2.00 0.75 0.75 2.87 1.89 5.67 6.6
1 1/4 7.87 2.44 5.25 3.87 2.50 0.75 0.81 3.07 2.09 5.87 8.8
1 1/2 7.87 2.64 6.13 4.50 2.87 0.87 0.87 3.23 2.24 6.02 9.9
1) If a grounding plate is required, L + 3 mm (0.12"). With 2 grounding plates (protection plates) L + 6 mm (0.24"), material upon request.
2) Connecting dimensions acc. to EN1092-1
3) Connection flange DN 10 (3/8")
4) Connection flange 1/2"
Pos: 38.3 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

29
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 38.4 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Maßzeichnungen/Durchflussmesser SE21/Maßzeichnung Flansch SE21F, DN50-DN100 @ 12\mod_1189600221421_3101.doc @ 121613

Stainless steel housing, flange acc. to DIN / EN 1092-1, DN 50 ... DN 100; ASME B16.5, 2 ... 4"

80 (3,15)
41 (1,61)
1

71 (2,80)
2
G

G
d2
E

E
Ød4
ØD
A
b
L

G00230

Fig. 27: Dimensions in mm (inch)


1 Cable gland M20 x 1.5 or NPT
2 Number of holes N

Flange dimensions acc. to DIN / EN, liner PFA Weight


mm
DN PN appr. kg
L1) A D k d4 d2 N b E G
50 10 ... 40 200 100 165 125 104 18 4 24 90 161 8
2)
65 10 ... 16 200 107 185 145 124 18 4 26 105 176 10
25 ... 40 200 107 185 145 124 18 8 26 105 176 10
80 10 ... 40 200 107 200 160 139 18 8 28 114 185 12
100 10 ... 16 250 159 220 180 161 18 8 24 128 199 18
25 ... 40 250 159 235 190 167 22 8 28 128 199 18
inch
DN PN appr. lb
L1) A D k d4 d2 N b E G
50 10 ... 40 7.87 3.94 6.50 4.92 4.09 0.71 0.16 0.94 3.54 6.34 17.6
2)
65 10 ... 16 7.87 4.21 7.28 5.71 4.88 0.71 0.16 1.02 4.13 6.93 22.0
25 ... 40 7.87 4.21 7.28 5.71 4.88 0.71 0.31 1.02 4.13 6.93 22.0
80 10 ... 40 7.87 4.21 7.87 6.30 5.47 0.71 0.31 1.10 4.49 7.28 26.5
100 10 ... 16 9.84 6.26 8.66 7.09 6.34 0.71 0.31 0.94 5.04 7.83 39.7
25 ... 40 9.84 6.26 9.25 7.48 6.57 0.87 0.31 1.10 5.04 7.83 39.7
1) If a grounding plate is required, L + 3 mm (0.12"). With 2 grounding plates (protection plates) L + 6 mm (0.24"), material upon request.
2) For DN 65/PN 16 acc. to EN1092-1 please order PN 40.

30
Electromagnetic Flowmeter FSM4000 D184S073U02

Flange dimensions acc. to ASME B16.5, liner PFA Weight


mm
Inch PN appr. kg
L1) A D k d4 d2 N b E G
2 200 100 152 121 99 19 4 23 90 161 8
2 1/2 200 107 178 140 118 19 4 26 105 176 10
CL 150
3 200 107 191 152 131 19 4 28 114 185 12
4 250 159 229 190 171 19 8 28 128 199 18
2 200 100 165 127 102 19 8 26 90 161 8
2 1/2 200 107 191 149 124 22 8 29 105 176 10
CL 300
3 200 107 210 168 143 22 8 32 114 185 12
4 250 159 254 200 177 22 8 36 128 199 18
inch
Inch PN appr. lb
L1) A D k d4 d2 N b E G
2 7.87 3.94 5.98 4.76 3.90 0.75 0.16 0.91 3.54 6.34 17.6
2 1/2 7.87 4.21 7.01 5.51 4.65 0.75 0.16 1.02 4.13 6.93 22.0
CL 150
3 7.87 4.21 7.52 5.98 5.16 0.75 0.16 1.10 4.49 7.28 26.5
4 9.84 6.26 9.02 7.48 6.73 0.75 0.31 1.10 5.04 7.83 39.7
2 7.87 3.94 6.50 5.00 4.02 0.75 0.31 1.02 3.54 6.34 17.6
2 1/2 7.87 4.21 7.52 5.87 4.88 0.87 0.31 1.14 4.13 6.93 22.0
CL 300
3 7.87 4.21 8.27 6.61 5.63 0.87 0.31 1.26 4.49 7.28 26.5
4 9.84 6.26 10.00 7.87 6.97 0.87 0.31 1.42 5.04 7.83 39.7
1) If a grounding plate is required. L + 3 mm (0.12"). With 2 grounding plates (protection plates) L + 6 mm (0.24"), material upon request.
Pos: 38.5 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

31
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 38.6 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Maßzeichnungen/Durchflussmesser SE21/Maßzeichnung Zwischenflansch SE21W, DN3-DN100 @ 3\mod_1157639606021_3101.doc @ 40264

4.2.2 Model SE21W


Stainless steel housing, wafer flange DN 3 ... DN 100 (1/10 ... 4")
DN 3 ... DN 40 (1/10 ... 1½")

1 80 (3,15)
41 (1,61)

71 (2,80)
G
E

ØD
ØC
DN 3 ... DN 8 (1/10 ... 5/16”)
A J
F

ØDi
37 (1,46)

B C
4 x Ø5,5 (0,22”) 4
2

ØDi
Ø13.6
A*
3
L G00231

DN 50 ... DN 100 (2 ... 4")

1 80 (3,15)
41 (1,61)
71 (2,80)
G
E

Ø Di
ØC
ØD
F

A*
L
G00459

Fig. 28: Dimensions in mm (inch)


1 Cable gland Pg 13.5 or NPT 3 Hole pattern
2 Pattern for axial holes 4 Mounting bracket (optional)

32
Electromagnetic Flowmeter FSM4000 D184S073U02

Wafer flange dimensions Weight


mm
DN (inch) PN appr. kg
L1) A* A B C ØC ØDi ØD E F G J
3 (1/10) 3
4 (5/32) 4
6 (1/4) 68 64 37 28 50 42 6 45 62 39 172 28 1.5
8 (5/16) 8
10 (3/8) 10 ... 40 10
15 (1/2) CL150 13 32
20 (3/4) CL300 78 74 42 28 50 50 18 54 66 43 205 32 2.0
JIS
25 (1) 90 86 42 46 70 59 24 63 73 48 219 32 2.0
32 (1 1/4) 98 94 42 46 70 69 30 73 78 53 229 32 2.5
40 (1 1/2) 103 99 42 46 70 77 36 82 82 57 237 32 3.0
50 (2) 117 112 – 60 110 95 47 100 90 50 213 – 4.0
65 (2 1/2) 10 ... 16 103 99 – 60 110 111 62 116 105 58 235 – 4.5
80 (3) CL150 103 99 – 60 110 128 74 133 114 67 252 – 6.5
JIS
100 (4) 133 129 – 60 110 155 96 160 128 80 279 – 8.5
inch
DN (inch) PN appr. lb
L1) A* A B C ØC ØDi ØD E F G J
3 (1/10) 0.12
4 (5/32) 0.16
6 (1/4) 2.68 2.52 1.46 1.10 1.97 1.65 0.24 1.77 2.44 1.54 6.77 1.10 3.3
8 (5/16) 0.31
10 (3/8) 10 ... 40 0.39
15 (1/2) CL150 0.51 1.26
20 (3/4) CL300 3.07 2.91 1.65 1.10 1.97 1.97 0.71 2.13 2.60 1.69 8.07 1.26 4.4
JIS
25 (1) 3.54 3.39 1.65 1.81 2.76 2.32 0.94 2.48 2.87 1.89 8.62 1.26 4.4
32 (1 1/4) 3.86 3.70 1.65 1.81 2.76 2.72 1.18 2.87 3.07 2.09 9.02 1.26 5.5
40 (1 1/2) 4.06 3.90 1.65 1.81 2.76 3.03 1.42 3.23 3.23 2.24 9.33 1.26 6.6
50 (2) 4.61 4.41 – 2.36 4.33 3.74 1.85 3.94 3.54 1.97 8.39 – 8.8
65 (2 1/2) 10 ... 16 4.06 3.90 – 2.36 4.33 4.37 2.44 4.57 4.13 2.28 9.25 – 9.9
80 (3) CL150 4.06 3.90 – 2.36 4.33 5.04 2.91 5.24 4.49 2.64 9.92 – 14.3
JIS
100 (4) 5.24 5.08 – 2.36 4.33 6.10 3.78 6.30 5.04 3.15 10.98 – 18.7
1) Installation lengths with 2 ground plates L + 3 mm (0.12")
Mounting bracket (optional)
With mounting bracket, combined height + 37 mm (1.46") DN 3 ... DN 40 (1/10 ... 1½"), + 10.5 mm (0.41") DN 50 ... DN 100 (2 ... 4").
Pos: 38.7 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

33
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 38.8 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Maßzeichnungen/Durchflussmesser SE21/Maßzeichnung Variabler Prozessanschluss SE21_, DN3-DN100 @ 3\mod_1157639609055_3101.doc @ 40285

4.2.3 Model SE21_


Variable process connection
DN 3 ... DN 40 (1/10 ... 1½")
104 (4,09) 80 (3,15)
80 (3,15) 1 41 (1,61)

E71 (2,80)
G

ØD
3 (0,12)
3
F

A 4 x Ø5.5

B C
(1,46)

2 5
37

4 G00232

DN 50 ... DN 100 (2 ... 4")


104 (4,09)
1 80 (3,15)
80 (3,15) 41 (1,61)
71 (2,80)

3 (0,12)
G
E

ØD
F
10,5 (0,41)

B 2 C 3
4 x Ø5.5 5 DN 65...100 4 x Ø5.5 DN 50 (2)
(2 1/2 ... 4) 4
G00233
Fig. 29: Dimensions in mm (inch)
1 Cable gland Pg 13.5 or NPT 4 Hole pattern
2 Pattern for axial holes 5 LPipe
3 Mounting bracket (optional) 6 Number of holes N

34
Electromagnetic Flowmeter FSM4000 D184S073U02

Variable process connection Weight


mm
DN (inch) appr. kg1)
A ØD B C E F G LPipe N
3 ... 10 (1/10 ... 3/8) 37 44 28 50 62 39 172 85 – 1.5
15 (1/2) 37 44 28 50 62 39 172 85 – 1.5
20 (3/4) 42 63 28 50 66 43 180 90 – 2.0
25 (1) 42 63 46 70 73 48 192 105 – 2.0
32 (1 1/4) 42 78 46 70 78 53 202 120 – 2.5
40 (1 1/2) 42 78 46 70 82 57 210 125 – 3.0
50 (2) – 100 60 110 90 50 211 128 8 4.0
65 (2 1/2) – 116 60 110 105 58 234 114 10 4.5
80 (3) – 133 60 110 114 67 252 114 6 6.5
100 (4) – 160 60 110 128 80 279 114 6 9.0
inch
DN (inch) appr. lb1)
A ØD B C E F G LPipe N
3 ... 10 (1/10 ... 3/8) 1.46 1.73 1.10 1.97 2.44 1.54 6.77 3.35 – 3.3
15 (1/2) 1.46 1.73 1.10 1.97 2.44 1.54 6.77 3.35 – 3.3
20 (3/4) 1.65 2.48 1.10 1.97 2.60 1.69 7.09 3.54 – 4.4
25 (1) 1.65 2.48 1.81 2.76 2.87 1.89 7.56 4.13 – 4.4
32 (1 1/4) 1.65 3.07 1.81 2.76 3.07 2.09 7.95 4.72 – 5.5
40 (1 1/2) 1.65 3.07 1.81 2.76 3.23 2.24 8.27 4.92 – 6.6
50 (2) – 3.94 2.36 4.33 3.54 1.97 8.31 5.04 0.31 8.8
65 (2 1/2) – 4.57 2.36 4.33 4.13 2.28 9.21 4.49 0.39 9.9
80 (3) – 5.24 2.36 4.33 4.49 2.64 9.92 4.49 0.24 14.3
100 (4) – 6.30 2.36 4.33 5.04 3.15 10.98 4.49 0.24 19.8
1) For process connection weights see page 36
Mounting bracket (optional)
With mounting bracket, combined height +37 mm (1.46") DN 3 ... DN 40 (1/10 ... 1½"), +10.5 mm (0.41") DN 50 ... DN 100 (2 ... 4").
Pos: 38.9 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

35
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 38.10 /Technische Daten / Datenblatt/Durchfluss/FXE4000/Maßzeichnungen/Durchflussmesser DE21-23/Maßzeichnung Adapter für variable Anschlüsse @ 8\mod_1175156265359_3101.doc @ 75146

4.2.4 Adapter for variable process connections DN 3 ... DN 100 (1/10 ... 4")

Da
Di
L L
1 2
a

Da
Di

R
L L
3 4 G00072

Fig. 30
1 Weld stubs 3 Tri-Clamp
2 Threaded pipe connection DIN 11851 4 External threads

Weld stubs Weight


mm
DN (inch) ISO 2037 DIN 11850 SMS DIN 2463 ISO 1127 kg
Ø Di Ø Da Ø Di Ø Da Series Ø Di Ø Da Ø Di Ø Da Ø Di Ø Da Series L
3 ... 10 (1/10 ... 3/8) 10.0 12.0 10 13 2 - - 10.3 13.5 10.3 13.5 1 127 0.4
15 (1/2) 15.2 17.2 16 19 2 - - 18.1 21.3 18.1 21.3 1 127 0.4
20 (3/4) 19.3 21.3 20 23 2 - - 23.7 26.9 23.7 26.9 1 132 0.7
25 (1) 22.6 25.0 26 29 2 22.6 25.0 25.0 28.0 23.7 26.9 1 149 0.7
32 (1 1/4) 31.3 33.7 32 34 1 - - 32.0 35.0 30.5 33.7 1 166 1.0
40 (1 1/2) 35.6 38.0 38 41 2 35.6 38.0 36.8 40.0 39.0 42.2 1 171 1.0
50 (2) 48.6 51.0 50 54 3 48.6 51.0 49.0 52.0 47.8 51.0 2 173 1.0
65 (2 1/2) 60.3 63.5 66 70 2 60.3 63.5 66.0 70.0 66.0 70.0 2 165 1.4
80 (3) 72.9 76.1 81 85 2 72.9 76.1 81.0 85.0 72.9 76.1 1 169 2
100 (4) 97.6 101.6 100 104 2 100.0 104.0 100.0 104.0 97.6 101.6 2 199 3
inch
DN (inch) ISO 2037 DIN 11850 SMS DIN 2463 ISO 1127 lb
Ø Di Ø Da Ø Di Ø Da Series Ø Di Ø Da Ø Di Ø Da Ø Di Ø Da Series L
3 ... 10 (1/10 ... 3/8) 0.39 0.47 0.39 0.51 2 - - 0.41 0.53 0.41 0.53 1 5.00 0.9
15 (1/2) 0.60 0.68 0.63 0.75 2 - - 0.71 0.84 0.71 0.84 1 5.00 0.9
20 (3/4) 0.76 0.84 0.79 0.91 2 - - 0.93 1.06 0.93 1.06 1 5.20 1.5
25 (1) 0.89 0.98 1.02 1.14 2 0.89 0.98 0.98 1.10 0.93 1.06 1 5.87 1.5
32 (1 1/4) 1.23 1.33 1.26 1.34 1 - - 1.26 1.38 1.20 1.33 1 6.54 2.2
40 (1 1/2) 1.40 1.50 1.50 1.61 2 1.40 1.50 1.45 1.57 1.54 1.66 1 6.73 2.2
50 (2) 1.91 2.01 1.97 2.13 3 1.91 2.01 1.93 2.05 1.88 2.01 2 6.81 2.2
65 (2 1/2) 2.37 2.50 2.60 2.76 2 2.37 2.50 2.60 2.76 2.60 2.76 2 6.50 3.1
80 (3) 2.87 3.00 3.19 3.35 2 2.87 3.00 3.19 3.35 2.87 3.00 1 6.65 4.4
100 (4) 3.84 4.00 3.94 4.09 2 3.94 4.09 3.94 4.09 3.84 4.00 2 7.83 6.6

36
Electromagnetic Flowmeter FSM4000 D184S073U02

Threaded pipe connection Tri-Clamp


DIN 11851 DIN 32676 ASME BPE
mm mm mm
DN (inch)
Weight Weight Weight
Rd. thd. L /kg Ø Di Ø Da Series L /kg Tri-Clamp Ø Di Ø Da L kg
3 ... 10 (1/10 ... 3/8) 28 x 1/8" 169 0.5 10 34.0 3 163 0.5 1/2" 9.4 25.0 143 0.5
15 (1/2) 34 x 1/8" 169 0.5 16 34.0 3 163 0.5 3/4" 15.7 25.0 143 0.5
20 (3/4) 44 x 1/6" 180 0.9 20 34.0 3 168 0.7 1" 22.1 50.4 143 0.7
25 (1) 52 x 1/6" 207 0.9 26 50.5 3 192 0.8 1" 22.1 50.4 143 1.2
32 (1 1/4) 58 x 1/6" 230 1.4 32 50.5 3 209 1.5 - - - - -
40 (1 1/2) 65 x 1/6" 237 1.4 38 50.5 3 214 1.4 1 1/2" 34.8 50.4 277 1.8
50 (2) 78 x 1/6" 243 1.4 50 64.0 3 216 1.2 2" 47.5 63.9 277 1.8
65 (2 1/2) 96 x 1/6" 245 2.2 66 91.0 1 221 1.6 2 1/2" 60.2 77.4 277 2.0
80 (3) 110 x 1/4" 259 3.2 81 106.0 1 225 2.4 3" 72.9 90.9 337 3.6
100 (4) 130 x 1/4" 307 4.4 100 119.0 1 255 3.1 4" 97.4 118.9 337 4.1
inch inch inch
DN (inch) Weight Weight Weight
Rd. thd. L /lb Ø Di Ø Da Series L /lb Tri-Clamp Ø Di Ø Da L lb
3 ... 10 (1/10 ... 3/8) 28 x 1/8" 6.65 1.1 0.39 1.34 3 6.42 1.1 1/2" 0.37 0.98 5.63 1.1
15 (1/2) 34 x 1/8" 6.65 1.1 0.63 1.34 3 6.42 1.1 3/4" 0.62 0.98 5.63 1.1
20 (3/4) 44 x 1/6" 7.09 2.0 0.79 1.34 3 6.61 1.5 1" 0.87 1.98 5.63 1.5
25 (1) 52 x 1/6" 8.15 2.0 1.02 1.99 3 7.56 1.8 1" 0.87 1.98 5.63 2.6
32 (1 1/4) 58 x 1/6" 9.06 3.1 1.26 1.99 3 8.23 3.3 - - - - -
40 (1 1/2) 65 x 1/6" 9.33 3.1 1.50 1.99 3 8.43 3.1 1 1/2" 1.37 1.98 10.91 4.0
50 (2) 78 x 1/6" 9.57 3.1 1.97 2.52 3 8.50 2.6 2" 1.87 2.52 10.91 4.0
65 (2 1/2) 96 x 1/6" 9.65 4.9 2.60 3.58 1 8.70 3.5 2 1/2" 2.37 3.05 10.91 4.4
80 (3) 110 x 1/4" 10.20 7.1 3.19 4.17 1 8.86 5.3 3" 2.87 3.58 13.27 7.9
100 (4) 130 x 1/4" 12.09 9.7 3.94 4.69 1 10.04 6.8 4" 3.83 4.68 13.27 9.0

External threads ISO 228 / DIN 2999, tapered Weight


mm
DN (inch) kg
R a L
3 ... 10 (1/10 ... 3/8) 3/8" 18 139 0.4
15 (1/2) 1/2" 18 139 0.4
20 (3/4) 3/4" 25 164 0.8
25 (1) 1" 25 179 0.8
inch
DN (inch) lb
R a L
3 ... 10 (1/10 ... 3/8) 3/8" 0.71 5.47 0.88
15 (1/2) 1/2" 0.71 5.47 0.88
20 (3/4) 3/4" 0.98 6.46 1.76
25 (1) 1" 0.98 7.05 1.76
Pos: 38.11 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

37
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 38.12 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Maßzeichnungen/Durchflussmesser SE21/Maßzeichnung 1/8"-Sanitäranschluss DS21B, DN1-DN2 @ 3\mod_1157639612500_3101.doc @ 40306

4.2.5 Model SE21B


1/8" sanitary connectors, DN 1 ... DN 2 (1/25 ... 3/32")
C D
E
F
3
B 1
A
48 (1,89) 48 (1,89)

2 (0,07)
4
10 (0,73)

26
(1,02) 4 x Ø5.5
40
60 (2,36) (1,57)
25
90 (3,54) (0,98)
135 (5,31)
69 (2,71)
G00236
Fig. 31: Dimensions in mm (inch)
1 Gaskets 3 Cable gland Pg 13.5 or NPT
2 Connecting dimensions for gland G 1/8" internal thread 4 Mounting bracket, standard

Important
The flowmeter sensor is equipped as standard with grounding electrodes in the same material as the measuring
electrodes. The flowmeter sensor can only be operated on S4 transmitters with software version B.30 or higher.

Preamplifier A B C D E F
Without / type A 119 (4.68) 71 (2.79) 64 (2.51) 40 (1.57) 80 (3.14) 41 (1.61)
Type B 104 (4.09) 56 (2.20) 163 (6.41) 71 (2.79) 142 (5.59) 70 (2.75)

Process Connection DN PN Weight kg


1/8" sanitary connectors1) 1 ... 2 (1/25 ... 1/12") 10 1.5 (3.3 lb)
1) For hose 6 x 4 mm
Pos: 39 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

38
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 40 /Überschriften/1.1/1-spaltig/A - C/Bestellinformationen @ 2\mod_1139828409218_3101.doc @ 3181

4.3 Ordering information


Pos: 41 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Bestellinformationen/SE21_ @ 3\mod_1157640488830_3101.doc @ 40369

Fixed and wafer flange, 1/8 in. Sanitary connectors, variable process connections, liner Torlon, PEEK, PFA
Electromagnetic Flowmeter Variant digit No. 1-5 6 7 8 9 10 11 12 13
FSM4000-SE21 Catalog No. SE21-
remote standard DN 1 ... DN 100
Process Connection
Wafer type 1) W
Welded flange DIN 2501 / EN 1092-1 / ASME (DN 3 ... DN 100) F
Weld stubs DIN 11850 R
Weld stubs ISO 2037 (only DN 10 ... DN 100) P
Weld stubs DIN 2463 Q
Weld stubs ISO 1127 J
Weld stubs SMS (only DN 25, DN 40 ... 100) X
Without adapter Spare sensor, for replacement only V
Pipe connection DIN 11851 S
Tri-Clamp DIN 32676 T
Tri-Clamp ASME BPE (DN 3 ... DN 25, DN 40 ... DN 100) K
Male thread ISO 228 / DIN 2999 conical (only DN 3 ... DN 25) E
1/8 in. pipe thread (only DN 1 ... DN 2) B
Others Z
Liner Material Meter Size
PEEK DN 1 1/25 in. 2) 3) K 0 1
PEEK DN 1.5 1/17 in. 2) 3) K 1 S
PEEK DN 2 1/12 in. 2) 3) K 0 2
Torlon DN 1 1/25 in. 2) 3) N 0 1
Torlon DN 1.5 1/17 in. 2) 3) N 1 S
Torlon DN 2 1/12 in. 2) 3) N 0 2
PFA DN 3 1/10 in. P 0 3
PFA DN 4 5/32 in. P 0 4
PFA DN 6 1/4 in. P 0 6
PFA DN 8 5/16 in. P 0 8
PFA DN 10 3/8 in. P 1 0
PFA DN 15 1/2 in. P 1 5
PFA DN 20 3/4 in. P 2 0
PFA DN 25 1 in. P 2 5
PFA DN 32 1-1/4 in. P 3 2
PFA DN 40 1-1/2 in. P 4 0
PFA DN 50 2 in. P 5 0
PFA DN 65 2-1/2 in. P 6 5
PFA DN 80 3 in. P 8 0
PFA DN 100 4 in. P 1 H

1) Accessories see table H Continued on next page


2) Grounding electrodes are mandatory
3) Operating only with S4 transmitter starting from software version B.30

39
Electromagnetic Flowmeter FSM4000 D184S073U02

Electromagnetic Flowmeter Variant digit No. 1-5 7 8 9 10 11 12 13 14


FSM4000-SE21 Catalog No. SE21-
remote sensor standard DN 3 ... DN 100
Measuring Electrodes Material Grounding Electrodes 4)
Hastelloy C-4 (2.4610) Without H
Hastelloy B-3 (2.4600) Without B
SST 1.4571 (316Ti) Without S
Titanium Without M
Tantalum Without T
SST 1.4539 (904L) Without for food and beverage applications F
Platinum-Iridium Without (< DN 10) P
Platinum-Iridium Without (DN 10 ... DN 32) P
Platinum-Iridium Without (>= DN 40) P
Hastelloy C-4 (2.4610) With O
Hastelloy B-3 (2.4600) With N
SST 1.4571 (316Ti) With E
Titanium With I
Tantalum With Q
SST 1.4539 (904L) With for food and beverage applications R
Platinum-Iridium With (< DN 10) G
Platinum-Iridium With (DN 10 ... DN 32) G
Platinum-Iridium With (>= DN 40) G
Pressure Rating 5)
PN 10 Tri-Clamp (DN 65 ... DN 100), pipe connection / welded stub ends / C
w/o adapter (DN 65, DN 100), 1/8 in. pipe thread
PN 16 Wafer type (DN 65 ... DN 100), Tri-Clamp (DN 3 ... DN 50), 6) D
pipe connection / welded stub ends / w/o adapter (DN 50, DN 80),
welded flange (DN 100), male thread
PN 40 Wafer type (DN 3 ... DN 50), pipe connection / welded stub ends / F
w/o adapter (DN 3 ... DN 40), welded flange (DN 3 ... DN 80)
ASME CL 150 Wafer type (DN 3 ... DN 100), welded flange (DN 3 ... DN 100) ISO laid length P
ASME CL 300 Wafer type (DN 3 ... DN 50), welded flange (DN 3 ... DN 100) ISO laid length Q
Material of Process Connection
Without only with wafer type 0
SST 1.4571 (316Ti) standard for 1/8 in. sanitary connection and welded flange 3
SST 1.4404 (316L) 4
PVC DN 1 ... DN 2 see Gasket 7
POM only 1/8 in. sanitary connection 8
Others 9
Accessories not with welded flange
Without 7) A
With fastening element C
Others Z
Accessories only with welded flange
Without 7) A
Protection rings SST 1.4571 (316Ti) 8) B
Grounding ring SST 1.4571 (316Ti) 8) C
Temperature Range
Standard design <= 130 °C (266 °F) (DN 1 ... DN 2: max. 120 °C / 248 °F) S

Continued on next page


4) Grounding electrodes are mandatory for DN 1 ... DN 2, optional from DN 3
5) Connection dimensions for flanges acc. to DIN 2501 / EN 1092-1 or ASME B16.5 or JIS B2210-10K
6) DN 65 / PN 16 with connection dimensions acc. EN 1092-1: Please order PN 40
7) Fastening element is mandatory for DN 1 ... DN 2
8) Protection rings on both and grounding ring fastened on one side to flange

40
Electromagnetic Flowmeter FSM4000 D184S073U02

Electromagnetic Flowmeter Variant digit No. 1-5 15 16 17 18 19 20 21 22


FSM4000-SE21 Catalog No. SE21-
remote sensor standard DN 3 ... DN 100
Certificates
PED Standard A
Material certificate per EN 10204 3.1 + pressure test per AD 2000 (> DN 2) D
Pressure test per AD 2000 G
Inspection certificate per EN 10204 3.1 F
Calibration Certificates
Standard A
Fingerprint saved at factory > DN 8 F
Type of Protection Cable Entry
IP 67 Thread for screw-type conduit fitting M20 x 1.5 2
IP 67 Thread for screw-type conduit fitting NPT 1/2 in. 4
IP 68 Thread for screw-type conduit fitting M20 x 1.5 9) 7
IP 68 Cable connected and Connection box potted 8
Others 9
Design
Without pre-amplifier from DN 10 >= 20 μS/cm 1
With pre-amplifier type A 2
Standard up to DN 8 >= 20 μS/cm, option > DN 8 >= 5 μS/cm and laid length max. 200 m
With pre-amplifier type B 4
Option up to DN 8 >= 5 µS/cm, > DN 8 >= 0,5 µS/cm and laid length max. 200 m
Name Plate
German adhesive foil G
English adhesive foil E
French adhesive foil F
Design Level (Specified by ABB) *
Electrode Design
Standard 1
Conical head, SST 1.4539 (904L)(from DN 10) 10) 2
Gasket Material
Without A
EPDM (with FDA certificate) E
Silicone (with FDA certificate) S
PTFE (only DN 1 ... DN 2) T
Viton (only DN 1 ... DN 2 and PVC process connection) V
Others Z

9) Sealing compound (optionally) D141B038U01


10) For application e.g. with high fat contents
Pos: 42 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

41
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 43 /Überschriften/1/M - O/Montagezubehör @ 3\mod_1157702574331_3101.doc @ 40455

5 Mounting accessories
Pos: 44 /==== Wechsel ein- auf zweispaltig ==== @ 0\mod_1130421847171_3101.doc @ 3828
Wechsel ein-auf zweispaltig
Pos: 45 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Allgemein/Montagezubehör @ 3\mod_1157702194544_3101.doc @ 40434

Protective flange
Protective flanges protect the liner from abrasive fluids and are
centered in the pipeline using a groove or spring. Protective flanges
are available in two thicknesses and are mounted on the flowmeter.
Installation length for Series 1000 and DIN / ANSI flanges ≤ DN 80
(3") L+20 mm, from DN 100 (4") L+25 mm.

Gaskets for protection plates / grounding plate


Only one gasket at the mating flange is required for PTFE/PFA-lined
flowmeters installed with metal grounding plates. A gasket should be
placed on either side of the grounding plate for hard/soft rubber lined
G00457
flowmeters. Commercially available gaskets made of material
(rubber, lt, PTFE, etc.) that is compatible with the fluid and the
Fig. 32: Protection plates / Grounding plate operating temperatures should be used.

Grounding plate is installed loose between flange and counter flange. Chemical applications (grounding)
For plastic pipelines or pipelines with insulating liners,
protection/grounding plates made of conductive PTFE (patent
Available in DN 3 ... DN 600 (1/10 ... 24"), Mat. 1.4571 (316 Ti),
pending) are used instead of expensive special materials (platinum,
Other materials upon request.
tantalum, etc.). These plates have tested and used successfully in
DN Inch PN S = Thickness numerous applications.
3 ... 80 1/10 ... 3 10 ... 40 1.5
100 ... 150 4 ... 6 25 ... 40 1.5 di
1
200 ... 250 8... 10 10 ... 16 1.5

d2
300 ... 600 12 ... 24 10 ... 16 3 R8 R2
k D

16
6,6
Ground connection1)
C l1 4±0,5
R2 R2 2
6,4
G00238
R8 B R8 Fig. 34: PTFE protection/grounding plates
ØB

20
ØA

16
16

10
6,6

Nominal size Inch PN Thickness


S

G00237 DN 10 ... DN 150 3/8 ... 6 10 ... 40 / CL 150 4 mm


Grounding plate, loose (centered Protection plates / < DN 200 upon request
by mounting bolts) Grounding plate attached to
flowmeter2)

1) Wafer accessories
Attach grounding cable with M6 screw, spring washer and nut
to the washer The following accessories are available based on meter size and
pressure rating: Bolts, nuts, lockwashers. Gaskets are not included in
ØB = Internal diameter of liner the accessories. For order information, see page 56 .
2) Attach the grounding plate directly on the flowmeter
DN 10 ... DN 600 (3/8 ... 24") Weld connections
Material: Stainless steel 1.4571 (316 Ti) For flowmeters with welded connections, welding adapters are
Thickness ≤ DN 100 (4") = 3 mm; ≥ DN 125 (5") = 5 mm available. For order information, see page 56 .
ØB = Internal diameter of liner Pos: 46 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

Fig. 33: Protection plates / Grounding plate

42
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 47 /==== Wechsel zwei- auf einspaltig ==== @ 0\mod_1130421955859_3101.doc @ 3829


Wechsel ein-auf zweispaltig
Pos: 48 /Überschriften/1/M - O/Messumformer @ 0\mod_1130331200101_3101.doc @ 3158

6 Transmitter
Pos: 49 /Überschriften/1.1/1-spaltig/S - U/Technische Daten @ 7\mod_1157704677334_3101.doc @ 40581

6.1 Technical data


Pos: 50 /==== Wechsel ein- auf zweispaltig ==== @ 0\mod_1130421847171_3101.doc @ 3828
Wechsel ein-auf zweispaltig
Pos: 51 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Technische Daten/Messumformer/Technische Daten @ 3\mod_1157703433676_3101.doc @ 40476

Design
Field-mount housing made of cast aluminum per DIN 1725, painted.
Paint coat is 80 µm thick. Lower section (RAL 7012), upper section
(RAL 9002). See dimensions, page 54. Weight, approx. 3.3 kg.

Electrical connections
Cable entry M20 x 1.5, NPT, PF with adapter (upon request), screw
terminals

Damping (1τ)
Configurable for 0.07 ... 20 s.

Low flow cutoff


G00458
Configurable for 0 ... 10% of end value range

Signal cable
Fig. 35: Transmitter FSM4000-S4
Max. cable length between flowmeter sensor and transmitter is 50 m
for the standard design and versions with empty pipe detection, from
DN 10 (3/8") and from 20 µS/cm. 200 m for designs with preamplifier.
Flow range Any flow range whose 100% value A 10 m signal cable is included with the flowmeter. If a longer cable is
corresponds to a flow velocity between 0.5 m/s required, see the order information for the transmitter on page 55.
and 10 m/s can be selected.
Minimum ≥ 20 µS/cm standard DN 10 ... 1000 Magnet coil supply cable
conductivity (3/8 ... 40“) Is required to connect the flowmeter sensor to the transmitter. For
≥ 20 µS/cm with preamplifier DN 1 ... 8 EMC reasons, the 2-wire cable is shielded. A 10 m cable is included
with the flowmeter. If a longer cable is required, see order information
(1/25 ... 5/16“) on page 55.
≥ 5 µS/cm with preamplifier DN 1 ... 1000
(1/25 ... 40“) Forward/reverse flow metering
≥ 0.5 µS/cm with preamplifier DN 10 ... 1000 Flow direction is indicated by direction arrows in the display and over
(3/8 ... 40“) the contact for external signaling. The contact output can also
transmit this information.
Reproducibility DN 1 ... 2 (1/25...1/12”):
(measurement ≤ ± (0.3 % of meas. value + 0.04 % of QmaxDN) Display
period = 100 s) DN 3 ... 1000 (1/10...40"): 4-line, illuminated graphic display. To improve readability, the flow
≤ ± (0.1 % of meas. value + 0,01 % of QmaxDN) information can be displayed on two lines with larger characters. This
allows the individual flowrate and output values to be displayed.
Response time 1τ = 70 ms (0 ... 66 %) "Fast" mode distinctive dot-matrix display with max. 4x16 characters. A plaintext
1τ = 200 ms (0 ... 66 %) "Standard" mode / diagnostic display opens automatically when the system detects an
piston pump error. An alarm condition can also be signaled via contact output.
Supply power U = 100 ... 230 V, 50/60 Hz
Urat = 85 ... 253 V, 50/60 Hz Data security
Via FRAM (ferroelectric non-volatile random access memory) all data
50/60 Hz ± 6% is stored for a period of 10 years (even without power) at shut-off or
20.4 ... 26.4 V AC, during a power outage. Additional security is provided by a serial
20.4 ... 31.2 V DC, ripple ≤ 5 % FRAM installed in the transmitter and on the external connection
Power S ≤ 45 VA (flowmeter sensor incl. transmitter) board in which the parameter settings and process information are
stored. This makes it easy to replace a transmitter – without
Ambient -20 ... 60 °C (-4 ... 140 °F)
reconfiguration – by simply uploading the stored data from the
temperature
external FRAM to the new transmitter.
Storage -20 ... 80 °C (-4 ... 176 °F)
Temperature Flow totalization
The flow is totalized in engineering units. The limits for the pulse
Relative humidity factor are configurable and depend on the measuring range (Qmax)
as well as the totalizer units selected. The limits are a min.
Acc. to IEC 60068-2-30 classification of environmental conditions,
(0.00016 Hz) and max. pulse frequency (5 KHz). Totalizer values for
natural factors, temperature and air humidity. No effect under the
the forward and reverse flow directions can be displayed simultan-
following conditions: Temperature in range from 25 ... 55 °C
eously by selecting the "Display" menu item.
(77 ... 131 °F) and a relative humidity of 94 ... 97%.

Vibration
Acc. to IEC 60068-2-6 (03/95), grouping of devices according to table
C2 for general industry applications. No additional effect on the
following levels of vibration. Frequency: 10 ... 55 Hz; amplitude max.
0.15 mm.

Protection Class per EN 60529


IP 67 for field-mount housing and NEMA 4X

43
Electromagnetic Flowmeter FSM4000 D184S073U02

Parameter adjustment The parameters can be checked for min.-max. errors. If the value is
The display supports several languages (German, English, French, above or below the specified limit, a diagnostic error or warning,
Finnish, Swedish, Italian, Spanish, Dutch or Turkish), and information depending on configuration, is triggered. The fingerprint database
can be entered by using the three keys in the dialog or via integrated in the transmitter allows you to compare the values at the
communication interface with PC, HART protocol, PROFIBUS or time of factory calibration or commissioning with the currently
FOUNDATION Fieldbus. The magnet stick can also be used to recorded values.
configure the flowmeter with the front door closed. Position the Users can store up to three different fingerprint data records
magnet stick over the magnet icons on the display cover. (commissioning, product1, product2) in the fingerprint database. The
default fingerprint (optional) cannot be overwritten.
Alarm signals
A max-min alarm can be used with the transmitter. The alarm limits Detectable error states
can be set between 0 ... 103% of the flow range. When the flowrate is Monitoring of ... Detection of ...
outside the alarm limits, an error message is displayed in the upper
line and a contact is triggered. The contact output can also transmit Coil current / coil Changes in the driver or coil circuit.
this information. voltage / DAC value
Coil resistance Changes in coil circuit, e.g., fine short.
Wiring error Coil temperature Risk of overheating due to fluid.
Monitoring of magnet coil or electrode circuits for short circuits or wire Coil insulation Problems with the coil insulation. Can be
breaks. resistance caused, e.g., by dampness in the
flowmeter sensor or in the connection
Upgrades box.
The transmitter with all its functions can also be used with older Linearity of magnetic External magn. field interference on the
model electromagnetic flowmeters. The appropriate model number circuit EMF, resulting in loss of accuracy.
must be selected in the transmitter. It is also possible to use the Signal quality (signal- Changes in fluid, e.g., gas bubbles,
transmitter with model 10D1422 flowmeters. A special FRAM (Cs = to-noise ratio) increased/reduced amount of solids.
100%, Cz = 0%) is required when updating an older model (see order
information for transmitter) and applies for meter sizes DN 3 ... DN Electrode balance Distorted flow profile, e.g., due to
1000 (1/10 ... 40") in low voltage design only. improper installation.
Detect interference in electrode circuit,
Isolation of input/output e.g., failure of electrode due to insulating
The current output, pulse output, contact input and contact output are coating or a collapsed liner (vacuum
galvanically isolated from the input circuit and from each other. shock).

Extended diagnostic functions Pos: 52 /======= Spaltenumbruch ======== @ 0\mod_1132937966324_3101.doc @ 3831

Recorded parameters/fingerprints
Cyclical (10 s ... 7 days) or manual acquisition of parameters.
The data records are stored in the transmitter FRAM (with time
stamp / operating hours counter). A maximum of 10 records can be
stored. When the eleventh record is stored, the oldest data record is
overwritten automatically.

Magnet coil circuit AC voltage


AC
DC resistance (primary value for
temperature)
Temperature
Insulation resistance (ground)
DAC value (primary value for coil current)
Electrode circuit Magnetic linearity
Electrode voltage E1 (primary value for
electrode balance)
Electrode voltage E2 (primary value for
electrode balance)
Electrode balance
Quality (signal-to-noise ratio)

44
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 53 /Überschriften/1.1/1-spaltig/D - F/Ein-/Ausgänge @ 11\mod_1157704418793_3101.doc @ 40560 Pos: 56 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Technische Daten/Messumformer/Digitale Kommunikation @ 3\mod_1157705112080_3101.doc @ 40602

6.2 Inputs/outputs 6.3 Digital communication


Pos: 54 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Technische Daten/Messumformer/Ein-/Ausgänge @ 3\mod_1157703875982_3101.doc @ 40539

DC output (terminals 31, 32) The transmitter has the following options for digital communication:
The following functions can be selected in the software:
0 ... 20 mA load ≤ 560 Ω 6.3.1 HART protocol
4 ... 20 mA load ≤ 560 Ω
0 ... 10 mA load ≤ 1120 Ω The system is registered with the HART Communication Foundation.
2 ... 10 mA load ≤ 1120 Ω

Contact output (terminals 41, 42)


The following functions can be selected in the software:
– Forward/reverse flow direction signal
– Flowrate alarm (max. or min.)
– General alarm
– Empty pipe
The contact output can be configured as normally open or normally
closed.
Optocoupler Fig. 36
"closed" 0 V ≤ UCEL ≤ 2 V, 2 mA ≤ ICEL ≤ 220 mA
"open" 16 V ≤ UCEH ≤ 30 V, 0 mA ≤ ICEH ≤ 2 mA
HART protocol
Configuration directly on the unit
Scaled pulse output (terminals 51, 52)
Software DSV401 (+ HART-DTM)
Maximum totalizer frequency 5 kHz. Pulse factor per engineering unit
from 0.001 ... 1000 / unit selectable in the software. The pulse width Transmission FSK modulation on voltage output
is configurable from 0.100 ... The version (active, passive) can be 4 ... 20 mA acc. to Bell 202 standard
changed via jumpers. Max. signal 1.2 mAss
amplitude
Active Current output min. 250 Ω, max. = 560 Ω
Voltage pulse 24 V rectangular load
– Load ≥ 150 Ω Cable AWG 24 twisted
– Pulse width ≤ 50 ms, max. pulse frequency ≤ 3 Hz, Max. cable length
1500 m
– Load ≥ 500 Ω Baud rate 1200 Baud
– Pulse width ≥ 0.1 ms, max. pulse frequency: 5 kHz. Display Log. 1: 1200 Hz
Log. 0: 2200 Hz
Passive
For additional information, see the separate data link description.
Optocoupler
"closed": 0 V ≤ UCEL ≤ 2 V, 2 mA ≤ ICEL ≤ 220 mA System integration
"open": 16 V ≤ UCEH ≤ 30 V, 0 mA ≤ ICEH ≤ 2 mA In conjunction with the DTM (Device Type Manager) to the device,
the communication (configuration, parametrization) can occur with
the corresponding framework applications according to FDT 0.98 or
Contact input (terminals 81/82)
1.2 (DSV401 R2), respectively. Other desired tool/system
Passive based on normally open contact (closed). The following integrations (e.g., AMS or Siemens S7) are available upon request.
functions can be selected in the software: DSV401 communication tool for HART®, free 90-day test version
also available upon request. DTMs are included in DSV401.
• External output signal switch-off
When the meter tube empties the output signals can be turned
off. DC output, e.g., 0 mA or 2/4 mA, depending on current output
configuration.
• External totalizer reset
For totalizer overflows displayed internally and pulse counters for
both flow directions.
• External system zero position
Starts the external zero adjustment. To run a zero adjustment,
the metering tube must be full and flowrate must be zero. The
adjustment lasts approx. 1 minute.
• Design passive (optocoupler)
"ON": 16 V ≤ UKL ≤ 30 V
"OFF" 0 V ≤ UKL ≤ 2 V
Ri: 2 kΩ
Pos: 55 /======= Spaltenumbruch ======== @ 0\mod_1132937966324_3101.doc @ 3831

45
Electromagnetic Flowmeter FSM4000 D184S073U02

6.3.2 PROFIBUS PA protocol 6.3.3 FOUNDATION Fieldbus (FF)


The interface conforms to the Profile 3.0 (Standard PROFIBUS, EN FF interface compliant with FF standard 890/ 891
50170, DIN 19245 [PRO91]). and FF-902/90
PROFIBUS PA ID no.: 078C hex. Interoperability Test IT 027200
Alternative standard ID no.: 9700 or 9740 campaign no.
Configuration directly on the unit Manufacturer ID 0x000320
Software DSV401 Device ID 0x0017
(+ PROFIBUS PA-DTM) Configuration • directly on the unit
Transmission signal acc. to IEC 61158-2 • via services integrated in the system
Cable shielded, twisted cable (acc. to • National configurator
IEC 61158-2, types A or B are Transmission signal acc. to IEC 61158-2
preferred).

Ethernet FOUNDATION Fieldbus H1


PROFIBUS DP PROFIBUS PA

100
HSE-Bus B 1 F
100
H2-Bus A 1 F

FF+ FF- FF+ FF- FF+ FF-


PA+ PA- PA+ PA- PA+ PA-
G00112
G00111 B = Linking device (incl. bus supply and termination)
A = Segment coupler (incl. bus supply and termination) Fig. 38: Example for FOUNDATION Fieldbus interface connection
Fig. 37: Example for PROFIBUS PA interface connection
Bus topology
• Tree and/or line structure
Bus topology • Bus termination: passive at both wire ends of the bus main wire
• Tree and/or line structure (RC-element R = 100 Ω, C = 1 µF)
• Bus termination: passive at both wire ends of the bus main wire
(RC-element R = 100 Ω, C = 1 µF)
Voltage/current consumption
• Medium current consumption: 10 mA
Voltage/current consumption • In the event of an error, the integrated FDE function (=Fault
Disconnection Electronic) integrated in the device is ensures that
• Medium current consumption: 10 mA the current consumption can rise to a maximum of 13 mA.
• In the event of an error, the integrated FDE function (=Fault • Upper current limit: electronically limited.
Disconnection Electronic) integrated in the device is ensures that
the current consumption can rise to a maximum of 13 mA. • The voltage on the bus line must lie in the range of 9 ... 32 V DC.
• The upper limit of the voltage is electronically limited.
• The voltage on the bus line must lie in the range of 9 ... 32 V DC. Bus address
The bus address is automatically assigned or can be set in the
For additional information, see the separate data link description. system manually.
The address detection occurs via an explicit combination of
manufacturer ID, device ID and device serial number.

System integration
The following is required:
• DD file (Device Description), which includes the device
description.
• The CFF file (common file format) is required for the segment
engineering. The engineering can be undertaken online or offline.
You can find both files, as well as the interface description on the
supplied CD (part no.: D699D002U01). This can be re-ordered, if
necessary, from ABB at no cost.
The files needed for operation can also be downloaded at
http://www.fieldbus.org.
Pos: 57 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

46
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 58 /==== Wechsel zwei- auf einspaltig ==== @ 0\mod_1130421955859_3101.doc @ 3829


Wechsel ein-auf zweispaltig
Pos: 59.1 /Überschriften/1.1/1-spaltig/A - C/Anschlusspläne @ 0\mod_1139844727625_3101.doc @ 3179

6.4 Terminal connection diagrams


Pos: 59.2 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Elektrischer Anschluss/Standard DN10 ... DN1000 @ 19\mod_1206005623203_3101.doc @ 171628

Standard DN 10 ... DN 1000 (3/8 ... 40")

S4 transmitter (wall mount housing)

1
1 1 2- 1+ 1
1
3 2S 2 1 1S 3 16 6 7 51 52 81 82 41 42 31 32 M3 M1 N L

8 9 10 11 12 13 11 11 14 15 2 3 4 5 6 7

SE 3 2S 2 1 1S 3A 16 M3 M1 SE

1 G00595

Flowmeter sensor

Fig. 39: Connection diagram: Flowmeter sensor standard DN 10 ... DN 1000 (3/8 ... 40")

1 Functional ground (busbar) 8 Steel shielding


2 Pulse output1) 9 Aluminum foil
3 Contact input1) 10 Yellow
4 Contact output1) 11 Shield
5 Current output1) 12 Blue
6 Magnet coil cable: 13 Red
shielded 2 x 1 mm2 CE Typ 227 TEC 74 14 white
ABB order no. D173D147U01, 10 m included in shipment, 15 Shielded signal cable: ABB order no. D173D025U01, 10 m
standard included in shipment
7 Supply power
Low voltage: 100 ... 230 V AC, terminals L, N,
Low voltage: 20.4 ...26.4 V AC;
Low voltage: 20.4 ... 31.2 V DC
Low voltage: Terminals 1+, 2-,
Frequency: 47 Hz ≤ f ≤ 53 Hz; 50 Hz supply power
Frequency: 56 Hz ≤ f ≤ 64 Hz; 60 Hz supply power

1) see chapter “connection examples for peripherals“ in the operating instructions and/or data sheet

Comment:
We recommend that shielded output cables be used with the shields connected to the functional ground at one end.
Pos: 59.3 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

47
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 59.4 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Elektrischer Anschluss/mit Vorverstärker, DN1 ... DN1000 @ 19\mod_1206005940562_3101.doc @ 171683

with preamplifier DN 1 ... DN 1000 (1/25 ... 40")

Transmitter S4 (wall mount housing)

1
1 1 2- 1+ 1
1
3 U+ 2 1 U- 3 16 6 7 51 52 81 82 41 42 31 32 M3 M1 N L

8 9 10 11 12 13 11 11 14 15 2 3 4 5 6 7

16
SE 3 U+ 2 1 U- 3A 16 M3 M1 SE

1 G00596

Flowmeter sensor

Fig. 40: Connection diagram: Flowmeter sensor with preamplifier DN 1 ... DN 1000 (1/25 ... 40"), transmitter field-mount housing

1 Functional ground (busbar) 8 Steel shielding


2 Pulse output1) 9 Aluminum foil
3 Contact input1) 10 Yellow
4 Contact output1) 11 Shield
5 Current output1) 12 Blue
6 Magnet coil cable: 13 Red
shielded 2 x 1 mm2 CE type 227 TEC 74 14 white
ABB order no. D173D147U01, 10 m included in shipment, 15 Shielded signal cable: ABB order no. D173D025U01,
standard 10 m, included in shipment
7 Supply power 16 with preamplifier (always with DN 1 ... DN 8 [1/25 ... 5/16"])
Low voltage: 100 ... 230 V AC, terminals L, N,
Low voltage: 20.4 ...26.4 V AC;
Low voltage: 20.4 ... 31.2 V DC
Low voltage: Terminals 1+, 2-,
Frequency: 47 Hz ≤ f ≤ 53 Hz; 50 Hz supply power
Frequency: 56 Hz ≤ f ≤ 64 Hz; 60 Hz supply power

1) see chapter “connection examples for peripherals“ in the operating instructions and/or data sheet

Comment:
We recommend that shielded output cables be used with the shields connected to the functional ground at one end.

Important
If the flowmeter sensor is equipped with a preamplifier for low conductivity or in meter size DN 1 ... DN 8
(1/25 ... 5/16“), the shieldings of the signal wires must be connected to terminals U+ and U- on both the flowmeter
sensor and the transmitter.
Pos: 59.5 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

48
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 59.6 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Elektrischer Anschluss/Nachrüstung für Modell 10D1422_, DN3 ... DN1000 @ 19\mod_1206006044656_3101.doc @ 171708

Retrofitting for model 10D1422: DN 3 ... DN 1000 (1/10 ... 40“); model 10DI1425 and 10DS3111A-E:
DN 500 ... DN 1000 (20 ... 40“)

S4 transmitter (wall mount housing)

10

1 8
24 V
3 2S 2 1 1S 3 16 6 7 51 52 81 82 41 42 31 32 M3 M1 N L 1
1

7 2 3 4 5 1 6

SE 3 2 1 6 7 L N

1
G00241
Flowmeter sensor

Fig. 41: Flowmeter sensor standard DN 3 ... DN 1000 (1/10 ... 40"), transmitter field-mount housing

1 Functional ground (busbar) 7 Shielded signal cable: Use ABB order no. D173D025U01
2 Pulse output1) or incorporate in existing wiring
3 Contact input1) 8 Reference cable:
4 Contact output1) Only for connection to model 10D1422
5 Current output1) Terminals: close 6,7 and the hook switch S903
6 Supply power 9 Magnet coil supply:
Low voltage: 100 ... 230 V AC, terminals L, N, Magnet coil supply via supply power
Frequency: 47 Hz ≤ f ≤ 53 Hz; 50 Hz supply power 10 Anschlussplatte: D685A1020U03
Frequency: 56 Hz ≤ f ≤ 64 Hz; 60 Hz supply power

1) see chapter “connection examples for peripherals“ in the operating instructions and/or data sheet

Comment:
We recommend that shielded output cables be used with the shields connected to the functional ground at one end.

Pos: 59.7 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

49
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 59.8 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Elektrischer Anschluss/PROFIBUS PA / FOUNDATION Fieldbus, DN1 ... DN1000 @ 19\mod_1206006116187_3101.doc @ 171733

DN 1 ... DN 1000 (1/25 ... 40") with PROFIBUS PA / FOUNDATION Fieldbus


S4 transmitter (wall mount housing)

6 6 3c
FF+ FF-
3b
1 3a
U+ U- PA+ PA- 1 2- 1+ 1
1
3 2S 2 1 1S 3 16 41 42 97 98 M3 M1 N L
1

7 8 9 10 11 12 10 10 13 14 2 3 4 5

15
SE 3 2S 2 1 1S 3A 16 M3 M1 SE
16
SE 3 U+ 2 1 U- 3A 16 M3 M1 SE
1 G00597

Flowmeter sensor

Fig. 42: Connection diagram: transmitter S4 with PROFIBUS PA / FOUNDATION Fieldbus

1 Functional ground (busbar) 5 Supply power


2 Contact output (see chapter “connection examples for Low voltage: 100 ... 230 V AC, terminals L, N,
peripherals” in the operating instructions and/or data sheet) Low voltage: 20.4 ...26.4 V AC;
3 Digital communication Low voltage: 20.4 ... 31.2 V DC
3a: PROFIBUS PA design accord. to IEC 61158-2 (Profile 3.0) Low voltage: Terminals 1+, 2-,
3a: U = 9 ... 32 V, I = 10 mA (normal operation) Frequency: 47 Hz ≤ f ≤ 53 Hz; 50 Hz supply power
3a: U = 9 ... 32 V, I = 13 mA (in the event of error/FDE) Frequency: 56 Hz ≤ f ≤ 64 Hz; 60 Hz supply power
3a: Terminals: 97/98, PA+/PA- 6 Shielded signal cable:
3a: (see chapter “connection via M12 plug” in the operating Power supply for flowmeter sensor with preamplifier
3a: instructions and/or data sheet) Terminals U+, U- instead of 2S and 1S in standard unit
3b: FOUNDATION Fieldbus design according to IEC 61158-2 7 Steel shielding
3a: U = 9 ... 32 V, I = 10 mA (normal operation) 8 Aluminum foil
3a: U = 9 ... 32 V, I = 13 mA (in the event of error/FDE) 9 Yellow
3a: Terminals: 97/98, FF+/FF- 10 Shield
3a: (see chapter “connection via M12 plug” in the operating 11 Blue
3a: instructions and/or data sheet) 12 Red
3c: Bus termination with installed bus termination 13 white
3a: components with hook switches closed 14 Shielded signal cable: ABB order no. D173D025U01,
4 Magnet coil cable: 10 m, included in shipment
shielded 2 x 1 mm2 CE Typ 227 TEC 74 15 Without preamplifier
ABB order no. D173D147U01, 10 m included in shipment, 16 With preamplifier (always with DN 1 ... DN 8 [1/25 ... 5/16"])
standard
Comment:
We recommend that shielded output cables be used with the shields connected to the functional ground at one end.
Important
If the flowmeter sensor is equipped with a preamplifier for low conductivity or in meter size DN 1 ... DN 8
(1/25 ... 5/16“), the shieldings of the signal wires must be connected to terminals U+ and U- on both the flowmeter
sensor and the transmitter.
Pos: 59.9 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

50
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 59.10 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Elektrischer Anschluss/Anschlussbeispiele (BA) @ 17\mod_1200932005063_3101.doc @ 153602

6.5 Connection examples for peripherals (incl. HART)


Current output

I E I = internal
E = external
+31
0/4 ... 20 mA
0/2 ... 10 mA
-32

G00243
Current output: selectable in software
Function: Active
0/4 ... 20 mA (0 Ω ≤ RB ≤ 560 Ω)
0/2 ... 10 mA (0 Ω ≤ RB ≤ 1120 Ω) (with HART only 4 ... 20 mA)
Terminals: 31, 32

Fig. 43: Current output active with/without HART protocol (4 ... 20 mA)

Contact output

I E Note
R B*
+U Horn and alarm light are shown only
41 0V as examples. Other suitable devices
24 V DC such as bells, sirens, buzzers, etc.,
42 can also be used.

* R B UCE
ICE G00244

Contact output: selectable in software


Function: Passive
"closed" 0 V ≤ UCEL ≤ 2 V, 2 mA ≤ ICEL ≤ 220 mA
"open" 16 V ≤ UCEH ≤ 30 V, 0 mA ≤ ICEH ≤ 2 mA
Terminals: 41, 42

Fig. 44: Contact output for system monitoring, Max.-Min. alarm for empty tube or forward/reverse signal

Contact input

I E

81 + 16 ... 30 V DC
Ri
82 0V
G00245

Contact input: selectable in software


Function: Passive
"On" 16 V ≤ UKL ≤ 30 V
"Off" 0 V ≤ UKL ≤ 2 V
Ri = 2 kΩ
Terminals: 81, 82

Fig. 45: Contact input for external totalizer reset and external zero return

51
Electromagnetic Flowmeter FSM4000 D184S073U02

Pulse output

Pulse output active Pulse output, passive, optocoupler


I E
R B*
51 + 16 ... 30 V DC

52 -

* RB UCE
ICE G00247

Function: active/passive, selectable via jumper (see "Commissioning" in operating instructions)


fmax: 5 kHz,
fmin: 0.00016 Hz
Setting range: Pulses/unit, pulse width (observe dynamic limits)
Active: 150 Ω ≤ Load < 10 kΩ Pulse width ≤ 50 ms, max. pulse frequency ≤ 3 Hz,
500 Ω ≤ Load < 10 kΩ Pulse width ≥ 0.1 ms, max. pulse frequency: 5 kHz
Passive: "closed": 0 V ≤ UCEL ≤ 2 V, 2 mA ≤ ICEL ≤ 220 mA
"open": 16 V ≤ UCEH ≤ 30 V, 0 mA ≤ ICEH ≤ 2 mA
Terminals: 51, 52
Fig. 46: Pulse output, active and passive, optocoupler

PROFIBUS PA / FOUNDATION Fieldbus

I E I E The resistance R and condenser C form the bus termination. They


must be installed when the device is connected to the
PA- FF- end of the entire bus cable.
98 98 R = 100 Ω; C = 1 µF
R R
PA+ FF+ 1 PROFIBUS PA
C C
97 97 2 FOUNDATION Fieldbus
I = Internal
1 2 G00248 E = external

Fig. 47

52
Electromagnetic Flowmeter FSM4000 D184S073U02

Bus termination for S4 transmitter


To terminate the bus if the instrument is at the end of the bus cable, the termination components in the S4 transmitter can be used. To do so,
close both hook switches in the connection area of the transmitter.

Important
If the transmitter module is removed, bus termination is also canceled.

3 U+ 2S 2 1 1S U- 3 16 6 7 41 42 97 98 M3 M1

G00249
Hook switches for
Connection board for field housing unit
bus termination must be closed

Fig. 48

Connection via M12 plug (for PROFIBUS PA only)


As an option, the bus can also be connected via an M12 plug instead of the cable gland (see order information for device). The device can be
shipped completely prewired. For information about suitable connectors (type EPG300) and other accessories, refer to the data sheet
10/63.6.44 DE.

Pin assignment
4 3 (Front view showing pin insert and pins)
PIN 1 = PA+
1 2
PIN 2 = nc
PIN 3 = PA-
PIN 4 = shield

G00250

Fig. 49
Pos: 60 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

53
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 61 /Überschriften/1.1/1-spaltig/M - O/Maßzeichnungen @ 11\mod_1157707121639_3101.doc @ 40791

6.6 Dimensional drawings


Pos: 62 /Technische Daten / Datenblatt/Durchfluss/FXE4000/Maßzeichnungen/Messumformer MAG-XE/Messumformergehäuse und Montagevorschlag @ 0\mod_1131704576853_3101.doc @ 3671

6.6.1 Transmitter housing and mounting arrangements

198 (7,79) 139,7 (5,50) 167 (6,57) 7)


10 (0,39) (0,2
Ø7

265 (10,43)

249 (9,80)
14,5 (0,57)
1
3

2
38 (1,49)
66 (2,59)
132 (5,19) 83,5 (3,28)
M20 x 1,5

min.175
(6,88)
min. 62 (2,44)

4
G00073

Fig. 50 (Dimensions in mm)

1 Field mount housing with window


2 Cable connector M20 x 1.5
3 Mounting holes for pipe mounting set for a 2” – pipe installation; mounting set upon request (order no. 612B091U07)
4 Protection class IP 67
Pos: 63 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

54
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 64 /Überschriften/1.1/1-spaltig/A - C/Bestellinformationen @ 2\mod_1139828409218_3101.doc @ 3181

6.7 Ordering information


Pos: 65 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Bestellinformationen/Messumformer S4 @ 3\mod_1157707575712_3101.doc @ 40812

Accuracy 0.5% of rate


External Transmitter Variant digit No. 1-3 4 5 6 7 8 9 10 11 12 13 14
FSM4000-S4 Catalog No. S4-
Housing
Field housing (cable gland M20 x 1.5) A
(cable gland 1/2 in. NPT) B
Others Z
Power Supply
100 ... 230 V AC G
20.4 ... 26.4 V AC / 20.4 ... 31.2 V DC K
Display
Magnetic stick operation + display illuminated A
Input / Output Options
Current output active + Pulse output active + Contact input + Contact output 1) 0 1
Current output active + Pulse output active + Contact input + Contact output + HART 0 2
Current output active + Pulse output passive + Contact input + Contact output 1) 0 3
Current output active + Pulse output passive + Contact input + Contact output + HART 0 4
Contact output + PROFIBUS PA 0 7
Contact output + FOUNDATION Fieldbus 0 8
Contact output + PROFIBUS PA with M12 plug 1 0
Application
Standard 0
Extended diagnostics (only with sensor SE21/SE41F) 1
For Sensor
SE2_, SE4_ 2) 1
DS2_, DS4_, 10DS3111, 10DI1425 (>= DN 500), 10D1422, 10D1418, 10D1430 3) 2
10D1462, 101472, 10DS3111 (Design Level A) > 12 in. 4) 4
Others 9
Certificates
Standard 0
Name Plate
German adhesive foil G
English adhesive foil E
French adhesive foil F
German SST plate J
English SST plate K
French SST plate L
Design Level (Specified by ABB) *
Software Level (Specified by ABB) *

Accessories
Catalog No.
Shielded signal cable (10 m included in price of the sensor) D173D025U01 price per meter
Shielded coil cable (10 m included in price of the sensor) D173D147U01 price per meter
Accessories for 2 in. pipe mounting 612B091U07

1) Not with low voltage power supply (code K)


2) The external FRAM is in the terminal box of the associated sensor on delivery
3) In this variant, the external FRAM already is plugged into the transmitter with the values Cz = 0 % and Cs = 100 %.
With sensor type 10D1422, 10DI1425 (>= DN 500), only Power Supply = 100 ... 230 V AC
and Input / Output Options 01 ... 04 available
4) In this variant, the external FRAM already is plugged into the transmitter with the values Cz = 0 % and Cs = 100 %
Only Power Supply = 100 ... 230 V AC and Input / Output Options 01 ... 04 available
A special reference network (1000 Ohm) is delivered with the transmitter. Details see Operating Instruction
Pos: 66 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

55
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 67 /Technische Daten / Datenblatt/Durchfluss/FXE4000/Bestellinformationen/FXE4000 - Zwischenflanschzubehör Tabelle H @ 15\mod_1195656143437_3101.doc @ 142263

7 Order information: Wafer accessories (table H)


FX / FSM Meter Size Pressure Rating Catalog No.
Material: SST DN 3 ... DN 10 1/10 ... 3/8 in. PN 10 ... PN 40 D614L265U03
ASME CL 150 D614L265U03
ASME CL 300 D614L265U04
DN 15 1/2 in. PN 10 ... PN 40 D614L265U03
ASME CL 150 D614L266U05
ASME CL 300 D614L266U06
DN 20 3/4 in. PN 10 ... PN 40 D614L267U04
ASME CL 150 D614L267U05
ASME CL 300 D614L267U06
DN 25 1 in. PN 10 ... PN 40 D614L268U04
ASME CL 150 D614L268U05
ASME CL 300 D614L268U06
DN 32 1-1/4 in. PN 10 ... PN 40 D614L269U04
ASME CL 150 D614L269U05
ASME CL 300 D614L269U06
DN 40 1-1/2 in. PN 10 ... PN 40 D614L270U04
ASME CL 150 D614L270U05
ASME CL 300 D614L270U06
DN 50 2 in. PN 10 ... PN 40 D614L296U04
ASME CL 150 D614L296U05
ASME CL 300 D614L296U06
DN 65 2-1/2 in. PN 10 ... PN 16 D614L297U08
PN 25 ... PN 40 D614L297U09
ASME CL 150 D614L297U10
ASME CL 300 D614L297U11
DN 80 3 in. PN 10 ... PN 40 D614L298U08
ASME CL 150 D614L298U09
ASME CL 300 D614L298U10
DN 100 4 in. PN 10 ... PN 16 D614L299U07
PN 25 ... PN 40 D614L299U08
ASME CL 150 D614L299U09

Note:
Depending on the nominal width and nominal pressure, the following accessories are available: bolts, nuts,
lock washers, centering elements. Seals are not included in the accessories.

Fitting Piece for welding


FX / FSM Meter Size Catalog No.
DN 3 ... DN 10 1/10 ... 3/8 in. D413C470U01
DN 15 1/2 in. D413C471U01
DN 20 3/4 in. D413C472U01
DN 25 1 in. D413C473U01
DN 32 1-1/4 in. D413C474U01
DN 40 1-1/2 in. D413C475U01
DN 50 2 in. D413C488U03
DN 65 2-1/2 in. D413C461U09
DN 80 3 in. D413C496U03
DN 100 4 in. D413C498U03

This fitting piece is an aid for primaries with weld stubs process connection
It allows these weld stubs to be welded coplanar into the pipe
Pos: 68 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830

56
Electromagnetic Flowmeter FSM4000 D184S073U02

Pos: 69 /Technische Daten / Datenblatt/Durchfluss/FXE4000/Bestellinformationen/FXC4000 - Aufnehmersimulator @ 15\mod_1195657516484_3101.doc @ 142289

8 Order information: Flowmeter sensor simulator FXC 4000


Flowmeter Sensor Simulator Variant digit No. 1-6 7 8 9 10 11
FXC4000 1) Catalog No. 55XC4-
Flow Signal Setting
Without (adapter only) 0
3-position digitswitch with 1000 steps 1
Others 9
Power Supply 2)
Without (adapter only) 0
With Schuko plug 100 ... 230 V AC 50 / 60 Hz 1
With 4 mm plug 24 ... 48 V AC / DC 2
With US plug 110 ... 240 V AC 50 / 60 Hz 3
Others 9
Accessories
Without 0
Adapter for transmitter type FXE4000-E4, FXM2000-XM2, FXF2000-DF23 1
Adapter plate for transmitter FSM4000-S4 5
Adapter plate for transmitter FET321 6
Others 9
Design Level (Specified by ABB) *
Name Plate
German 1
English 2
French 3
Others 9

1) Also older product models 50XM1000, 50XH1000, 50XO1000, 50SM1000


can be tested with test simulator type FXC4000
2) Power supply is for transmitter supplying
Pos: 70 /Rückseiten/Göttingen @ 0\mod_1138784575421_3101.doc @ 3125
A

===== Ende der Stückliste =====

57
ABB has Sales & Customer Support The Company’s policy is one of continuous product
expertise in over 100 countries worldwide. improvement and the right is reserved to modify the
information contained herein without notice.
www.abb.com/flow
Printed in the Fed. Rep. of Germany (04.2008)
D184S073U02 Rev. 07

© ABB 2008

ABB Limited ABB Inc. ABB Automation Products GmbH


Oldends Lane, Stonehouse 125 E. County Line Dransfelder Str. 2
Gloucestershire, GL10 3TA Road 37079 Goettingen
UK Warminster, PA 18974 Germany
Tel: +44 (0)1453 826661 USA Tel: +49 551 905-534
Fax: +44 (0)1453 829671 Tel: +1 215 674 6000 Fax: +49 551 905-555
Fax: +1 215 674 7183 CCC-support.deapr@de.abb.com

You might also like