CARD1XP1X: XP Series Main Unit Program
CARD1XP1X: XP Series Main Unit Program
XP Series main unit program
CARD1XP1X:
CARD1XP1X: Ver.00-13 Program Upgrade Procedure
<Applicable to S/N> <Release Date>
XP-100 S/N: Not fixed yet May 1, 2014
XP-300 S/N: Not fixed yet
<Part No./Name>
BD347455: CARD_ASSY NO.25(1XP1X)
Changed Specifications
(2) A function has been added to output DP printing paper from the front or the rear in accordance
with setting made on the “DP Output” screen.
(3) “DP Output” has been added to service mode setting value printing format.
Resolved issues
None
Remarks
None
ECR:314B016
XP Series S/M
XP Series > Service Manual > Version History 1/11 ページ
Version History
CARD1XP1X: XP Series main unit program
CARD1XP1X: Ver.00-12 Program Upgrade Procedure
Changed Specifications
Latin American Portuguese setting values have been added to language settings.
1 Previously, even though the instrument was shut down and was then switched OFF and ON
after a replace reagent error had occurred, the same error occurred again when the
instrument was restarted. This issue has now been resolved.
Remarks
・ Main Ver.: Ver.00-12 build0003
・ BIOS Ver.: Ver.00-01
・ Seq. Ver.: Ver.0005
ECR: 313C032-R
[Back to Top]
Changed Specifications
file://tsc-serv02/cs-tech2/COMMON/TS_English/SM/Web_Manual/XP-300/SM_Cntnt/AppenD/CAR... 2013/10/03
XP Series > Service Manual > Version History 2/11 ページ
1 As authentication methods for sending e-mails to SNCS, SMTP(LOGIN) authentication and SMTP(PLAIN)
authentication have been added in addition to POP authentication.
(1) SMTP(LOGIN) authentication and SMTP(PLAIN) authentication have been added as authentication method
setting values.
The following changes have been made to SNCS [Online QC] Network Settings screen, [Maintenance] Network
Settings screen and [Screen Image] Network Settings screen. The changes have also been included in service
settings.
The item name for authentication method has been changed from "POP" to "Authentication".
Options for authentication method have been changed to "Disable", "POP", "SMTP(LOGIN)" and "SMTP
(PLAIN)".
The item name for authentication server name has been changed from "POP Server Name" to "Auth
Server Name".
The item name for authentication server port number has been changed from "POP Server Port" to "Auth
Server Port".
The item name for user name used for authentication has been changed from "POP User Name" to "Auth
User Name".
The item name for password used for authentication has been changed from "POP Password" to "Auth
Password".
(2) E-mails can now be sent via SMTP(LOGIN) authentication and SMTP(PLAIN) authentication.
E-mails will be sent via POP authentication, SMTP(LOGIN) authentication or SMTP(PLAIN) authentication, in
accordance with authentication method setting values.
None
file://tsc-serv02/cs-tech2/COMMON/TS_English/SM/Web_Manual/XP-300/SM_Cntnt/AppenD/CAR... 2013/10/03
XP Series > Service Manual > Version History 3/11 ページ
Remarks
・ Main Ver.: Ver.00-11 build0004
・ BIOS Ver.: Ver.00-01
・ Seq. Ver.: Ver.0005
ECR:313C031
[Back to Top]
Changed Specifications
1 Due to request from the Chinese markets, only CELLPACK CPK303A can now be used on
XP-300 instruments for the Chinese markets. Other types of CELLPACK; PK-30L, CPK-300A
and CPK-310A can no longer be used in the Chinese markets.
1 Previously, on the external printer settings screen, [LX-300+] printer was displayed as a
printer type option, even though it was not supported by XP Series. This issue has now been
resolved.
As shown in the diagrams below, [LX-300+] has been deleted from the options to choose as a
printer type.
NOTE:
If the program is upgraded to version 00-10 from version 00-09 or lower when [LX-300+]
printer is set, [LX-300+] printer will still be set for the printer type. However, this would not be
a problem as [LX-300+] printer is controlled in the same manner as ESC/P.
Before change After change
file://tsc-serv02/cs-tech2/COMMON/TS_English/SM/Web_Manual/XP-300/SM_Cntnt/AppenD/CAR... 2013/10/03
XP Series > Service Manual > Version History 4/11 ページ
2 Previously, if any of the [pocH], [KX-21N] and [K-1000] formats are selected in Host Output
settings and pre-diluted mode analysis results are output to the host computer, volume
distribution analysis parameters were masked as "*0000" except for CBC8 parameters.
This issue has now been resolved.
Remarks
・ Main Ver.: Ver.00-10 build0008
・ BIOS Ver.: Ver.00-01
・ Seq. Ver.: Ver.0005
ECR:312J086
[Back to Top]
Changed Specifications
1 Steps have been taken to resolve the issue where QC chart was not displayed depending on
how the instrument was restarted after the instrument had been shut down.
Previously, if the instrument was restarted by pressing "Restart" button without switching
OFF the power after a shutdown sequence was executed, the following problems occurred.
Past QC data were not displayed on the QC chart screen.
file://tsc-serv02/cs-tech2/COMMON/TS_English/SM/Web_Manual/XP-300/SM_Cntnt/AppenD/CAR... 2013/10/03
XP Series > Service Manual > Version History 5/11 ページ
If QC analysis was performed when there were less than 60 plots of QC data, only
analysis results were displayed on the QC chart screen and QC data analyzed after the
instrument was restarted were not displayed. (However, if the instrument was restarted
by switching OFF and ON the XP main unit, all QC data were displayed.)
If QC analysis was performed when there were 60 plots of old QC data, QC analysis
was not completed successfully and the instrument locked up, displaying a system
error.
Remarks
・ Main Ver.: Ver.00-09 build0003
・ BIOS Ver.: Ver.00-01
・ Seq. Ver.: Ver.0005
ECR:312I034
[Back to Top]
Changed Specifications
1 Characters "." and "_" have been added to the alpha/numeric keypads.
1. For upper case, "-" key has been changed to " .- _ ".
2. For lower case, "-" key has been changed to " .- _".
Before change After change
file://tsc-serv02/cs-tech2/COMMON/TS_English/SM/Web_Manual/XP-300/SM_Cntnt/AppenD/CAR... 2013/10/03
XP Series > Service Manual > Version History 6/11 ページ
2 The number of characters which can be set for POP user name has been increased.
1. POP user name input screens for QC data, maintenance data and image data have been
changed so that up to 50 characters can be set for POP user name.
Before change After change
Note:
If using more than 15 characters for POP user name, make sure to upgrade CDR2XP1X
program version to [Main Ver.:Ver.1.10].
3 Duration of SNCS time out has been changed.
To reduce the number of SNCS errors, the timeout duration for an SNCS connection error
has been changed from 30 seconds to 3 minutes.
4 Date and time information has been added to file names for SNCS maintenance reports.
Of files with the same name, only 1 file can be imported per day in SNCS systems except for
file://tsc-serv02/cs-tech2/COMMON/TS_English/SM/Web_Manual/XP-300/SM_Cntnt/AppenD/CAR... 2013/10/03
XP Series > Service Manual > Version History 7/11 ページ
Online QC. Therefore, time and date information [YYYYMMDDhhmmss] is added to file
names for instruments which support SNCS.
Previously, as no date and time information was included in files sent from the XP instrument,
only 1 file was imported per day and subsequent files were deleted. To resolve this issue,
date and time information has been added to the names of files attached to the following 4
maintenance reports:
Startup report
Shutdown report
Event history
Error report
Remarks
file://tsc-serv02/cs-tech2/COMMON/TS_English/SM/Web_Manual/XP-300/SM_Cntnt/AppenD/CAR... 2013/10/03
XP Series > Service Manual > Version History 8/11 ページ
Print Server program has been added in order to provide the field with a program that can be
operated on a PC when printing on a Windows compatible printer via a PC using the LAN port
(for connecting to the host computer).
ECR: 312I034
[Back to Top]
Changed Specifications
1 "@" key has been added to the alphabet and numeric keypads to allow "@" to be used
for POP user name, etc.
For upper case, "QZ" key has been changed to "QZ@".
For lower case, "qz" key has been changed to "qz@".
1 Previously, on instruments for China, some characters were garbled in "Settings Error"
screen. This issue has now been resolved.
2 Previously, when there was no paper in a graphic printer (GP) and multiple sample
data were sent to the printer from stored data, "GP paper empty" error occurred, and
there were cases where only data for the first sample were printed and data for the
remaining samples were not printed even though paper was subsequently refilled in the
printer and [Retry] button was pressed in the help screen.
Also, when the XP instrument was not connected to a graphic printer (GP) and multiple
sample data were sent to the printer from stored data, "GP printout error" occurred, and
there were cases where only data for the first sample were printed and data for the
remaining samples were not printed even though the printer was subsequently
connected to the instrument and [Retry] button was pressed in the help screen.
These issues have now been resolved.
file://tsc-serv02/cs-tech2/COMMON/TS_English/SM/Web_Manual/XP-300/SM_Cntnt/AppenD/CAR... 2013/10/03
XP Series > Service Manual > Version History 9/11 ページ
Remarks
ECR:312G004
[Back to Top]
Changed Specifications
1 A function has been added to send print data to a PC connected via LAN.
NOTE:
With this specification change, it will be possible to print with a commercially-
available printer, using a PC connected to an XP instrument via a LAN cable as
shown in the diagram below. However, this function is not available at this stage,
as it is necessary to install designated software (to be released at the end of
August 2012) into the PC.
None
Remarks
file://tsc-serv02/cs-tech2/COMMON/TS_English/SM/Web_Manual/XP-300/SM_Cntnt/AppenD/CAR... 2013/10/03
XP Series > Service Manual > Version History 10/11 ページ
ECR:312F014
[Back to Top]
Changed Specifications
1 When a calibration error occurs during automatic calibration, it is necessary to clear the
error to perform automatic calibration once again. Previously, however, a calibration
error could only be cleared when calibration values are confirmed and the Quit button
was pressed in the manual calibration screen. Therefore, when a calibration error
occurred during automatic calibration, it was difficult to figure out how to clear the error.
To allow automatic calibration to be performed smoothly once again, a change has
been made to clear the calibration error even from the automatic calibration screen.
2 To reduce the frequency of SNCS connection errors occurring even in a poor network
environment, the following changes have been made:
1. The duration of timeout for SNCS connection errors has been extended from 6
seconds to 30 seconds.
2. When there is a connection problem with the SNCS server, up to 5 attempts will
be made to connect the SNCS server.
3 Start-up report files will now be sent even when the instrument is started using Hot
Start.
(However, background check service data will not be sent, as background check is not
performed when the instrument is started using Hot Start.)
1 Previously, when data of more than 2001 samples were output at the same time to
several devices (Printing on internal printer (IP), printing on graphic printer (GP),
printing on list printer (LP), and output to host computer), a dialog box appeared,
stating that "Please wait for a moment." and data output could not be started. This
issue has now been resolved.
2 Previously, if a "Waste not drained" error occurred during the preprocessing sequence
(a preparation sequence for TD cleaning) when TD cleaning was executed from the
maintenance menu screen, there were cases where "Waste not drained" errors
occurred consecutively and the instrument locked up after the preprocessing sequence
was completed. To resolve this issue, a change has been made so that until the initial
"Waste not drained" error is cleared, the same error will not occur.
3 Previously, when the analysis sequence was stopped at Sequence No.7 in the HGB
rough adjustment screen, HGB value was always displayed as 0.0, failing to display the
correct value. This issue has now been resolved.
4 Previously, when QC (X-Bar) was performed, "-" was displayed in the [Judgment]
column even though the results were within the acceptable range. This issue has now
been resolved.
5 Previously, when analysis results of samples with sample ID attribute "M" were
analyzed manually, sample ID attribute "A" was incorrectly displayed for manual
file://tsc-serv02/cs-tech2/COMMON/TS_English/SM/Web_Manual/XP-300/SM_Cntnt/AppenD/CAR... 2013/10/03
XP Series > Service Manual > Version History 11/11 ページ
Remarks
ECR:312E028
[Back to Top]
CARD1XP1X: Ver.00-04
<Parts Information>
Changed Specifications
Remarks
This is the initial version in the field.
[Back to Top]