Technical Bulletin: Updating The CELLCHECK-400 Set Values Used For WBC Sensitivities
Technical Bulletin: Updating The CELLCHECK-400 Set Values Used For WBC Sensitivities
TECHNICAL BULLETIN
No: 201379-R
TITLE: Updating the CELLCHECK-400 Set Values used for WBC Sensitivities
PRIORITY: Information only
DATE RELEASED: October 18, 2013
MODEL: KX-21, pocH-100i, K-1000, K-4500, XP-Series
Revision Information:
201379-R: Added K-4500 Service Manual revised page.
Applicable To:
CELLCHECK-400 Lot No.A3006 and thereafter
Summary:
The lot of latex particles utilized in CELLCHECK-400 has been changed.
Detailed Information:
The lot of latex particles utilized in CELLCHECK-400 has been changed from Lot No.A3006 and thereafter.
In accordance with this change, the WBC Sensitivity Adjustment Target Values have been changed as listed
below.
Product WBC
K-1000, K-4500 133.3 +/- 2 fL
pocH-100i 133.3 x 1.451= 193.4 +/- 3.5 fL
KX-21, XP Series 133.3 x 1.354 = 180.5 +/- 2 fL
In addition, the wrong description on XP-Series Service Manual page 4-5 has been corrected.
In Step (13) W-FMV has been corrected to W-MFV.
Action To Be Taken :
1.Please refer to the following pages of updated SYSMEX documentation (attached) for applicable information.
Reference Document :
None
file://tsc-serv02/cs-tech3/TS_Japan/CSO_SM_Eng/KX-21_Series_E_2013.11.22/SM_Cntnt/AppenC/2... 2013/11/22
201379-R 2/2 ページ
Attachments :
KX-21 Series Service Manual page: 4-1
pocH-100i Series Service Manual page: 4-11
XP-Series Service Manual page: 4-3, 4-5
K-4500 Service Manual page: 10-2