No: 314B016
Title: Program Update to Version 00-13
Included ECR No:
Priority: Information Only
Date Released: May 1, 2014
Model: XP Series
Revision Information:
N/A
Applicable To:
This change will be applied to instruments assembled in April 2014 and thereafter. The serial
number information will be provided in ECR revision when confirmed.
Summary:
The XP Series main unit program has been updated from version 00-12 to version 00-13. In
addition, the XP Series program for SNCS communication PCB (SNCS Firmware) has been
updated from version 00-03 to version 00-04.
Detailed Information:
<Changed Specifications>
It is now possible to select the direction to output printing paper on DP printer.
(1) DP Output setting function has been added to [Settings] in Service Menu.
The fourth screen has been added to [Parameters] in [Settings] to display setting
parameter [DP Output].
On this screen there are two options to choose from.
- Backward (Printing paper is output from the rear. Default)
- Forward (Printing paper is output from the front.)
2. [CDR2XP1X (SNCS Firmware): Program for SNCS communication PCB (PCB NO.90043)]
Changes are as follows:
<Changed Specifications>
As MAC addresses for LAN port expansion module (XPort04R) were running out, an
alternate module (XPort05R) is now supported.
(1) Firmware has been changed to support MAC addresses of both XPort04R and
XPort05R.
Previously, only lower MAC addresses were referenced. However, the program has
been changed to reference upper addresses as well.
3. XP Series Service Manual Chapter 5: Service Program has been revised as listed below.
Action To Be Taken:
1. This program will not be provided free of charge for instruments in the field. If you need to
update the program version, an order should be placed using the part number listed below.
4. Please refer to the attached pages of updated SYSMEX documentation for applicable
information. These updated pages will be included in the next Manual update.
Reference Document:
None
CDR2XP1X XP Series Program for SNCS Communication PCB (PCB NO.90043) Version
History
XP Series Service Manual Chapter 5 Service Program pages 5-31, 5-32, 5-58