System Evaluation Siemens Somatom Emotion 16
System Evaluation Siemens Somatom Emotion 16
System Specifications
The status of the system is very good, it looked well maintained and very clean.
The system did pass all tests without any problems. The following photo shows the results
of the last Constancy Check
IDTM GmbH System Evaluation - Siemens Emotion 16 Page 4
X-Ray Tube
The tube has been replaced on 15. February 2019 and had 50385 scan seconds on the
date of inspection.
IDTM GmbH System Evaluation - Siemens Emotion 16 Page 5
System Components
LCB
IDTM GmbH System Evaluation - Siemens Emotion 16 Page 12
Console
ICS Tower 11B, Celsius M470 – 2 Power IRS Tower 10G, Celsius M470 – 2 Power
IDTM GmbH System Evaluation - Siemens Emotion 16 Page 13
Console
UPS
IDTM GmbH System Evaluation - Siemens Emotion 16 Page 14
Accessories
Technical Documentation
System Software
IDTM GmbH System Evaluation - Siemens Emotion 16 Page 15
Phantoms
CCR-Combi Phantom
IDTM GmbH System Evaluation - Siemens Emotion 16 Page 16
Patient Positioning
IDTM GmbH System Evaluation - Siemens Emotion 16 Page 17
DVD_SOM5_IRS_VB42B_V3.0_64bit
DVD_SOM5_ICS/IES_VB42B_1 of 2_32bit
DVD_SOM5_ICS/IES_VB42B_2 of 2_32bit
syngo Expert-i
CD_SOM5_ICS/IES_VB42B_32bit_BARE
Update_Licenses_VB42B
E-Manual, syngo CT 2014A
ICS 11B Recovery DVD, Disc1 of 2
ICS 11B Recovery DVD, Disc1 of 2
QA Check
Check Up including Tube Warm Up and Defective Channels Correction, Air Calibration
Constancy Check including Phantom Check, Slice, Homogeneity/Water, Noise and MTF
The system is located inside the Radiology Clinic, which in the ground level of the main
hospital building. The control room is on the right side of the scan room. The LCB is
located inside the scan room.
The removal route is not too long and not overly complicated.
IDTM GmbH System Evaluation - Siemens Emotion 16 Page 25
2.30m
IDTM GmbH System Evaluation - Siemens Emotion 16 Page 27
View to door no. 3 (2.04m * >2m) and door no. 4 (2.20m * >2m)
IDTM GmbH System Evaluation - Siemens Emotion 16 Page 28
This is the emergency entrance and it should not be blocked during loading of the
equipment.
Cable Routing
The cables from the console to the gantry are routed through an on-wall cable tray into
the suspended ceiling and out of the suspended ceiling through another on-wall cable tray
into floor cable ducts
IDTM GmbH System Evaluation - Siemens Emotion 16 Page 29