Professional Documents
Culture Documents
XP PM Standard Procedure en
XP PM Standard Procedure en
Maintenance
Report No.
Facility Name
Instrument
XP-100/XP-300 Serial No.
Name
Maintenance
Date
Name of Service
Engineer
Operatio
Items to be checked Procedure
n time
Go to [Menu] -> [Service] -> [Settings] -> [Print
Print Setting Value
Checking setting Settings] to print all setting values.
5 min.
values Go to [Menu] -> [Settings] -> [Host Output
HOST Output Setting OFF
Settings] to change "Connection" to "Disconnect".
Analyze normal sample or control material and
Data analysis Check background count level print average values or record them using stored
before 10 min. data/QC chart.
maintenance Check data using control material
Perform background check.
or samples
Go to [Menu] -> [Service] -> [Service Data]. Print
HGB background convert
Check out and check service data for the latest samples.
WBC clog level
adjustment
RBC clog level
values before Go to [Menu] -> [Maint.] -> [Status Display] to
0.05MPa
maintenance check pressure.
-0.0333MPa
Removing covers Remove front and side covers.
Disassemble and clean SRV Use CELLCLEAN and distilled water to perform
cleaning.
SRV Clean the tray 15 min. (Refer to Instructions for Use Chapter 12.)
Replace PIPETTE NO.57 using FITTING NO. 9-
Replace the aspiration pipette
A.
Use CELLCLEAN and distilled water to perform
Clean Rinse Cup 5 min. cleaning.
Rinse Cup (Refer to Instructions for Use Chapter 12.)
Replace Tube Pharmed for Waste Replace by using Tube Pharmed 1/4 IN X 1/8 IN.
5 min.
Line (Refer to Service Manual "Schematics".)
Replace Waste Chamber (Every 5
5 min. Replace WASTE CHAMBER ASSY NO.35.
years)
Disassemble and clean Waste Wipe the interior of the chamber and the float
Waste Chamber Chamber (when not replacing 5 min. switch using gauze impregnated with
Waste Chamber). CELLCLEAN.
Replace Waste Chamber Drain Replace silicon tube F739-8.
5 min.
Line silicon tubes. (Refer to Service Manual "Schematics".)
Wipe the unit and the needle tip using gauze
Disassemble and rinse Bellows Unit 5 min.
impregnated with ethanol.
Inject ethanol using an injection syringe to clean.
Bellows Unit If the tube is extremely dirty, replace the whole
Clean orifice tube 5 min. tube.
TUBE POLYURETHANE 1.8MMX3.4MM
(1400mm)
Rinse Trap Chamber Clean Trap Chamber using distilled water.
Pneumatic Unit 30 min. Replace by using KX-21 Pneumatic Unit (XP
Replace Compressor series).
(Refer to Service Manual Parts List.)
Replace Pneumatic Unit protection
Regulator Unit B 2 min B Replace PU PROTECTION FILTER NO.1. B
filter
0.5kg/cm2(0.05MPa) Adjust pressure to 0.05Mpa using Regulator.
Pressure and 10 min.
Pressure adjustment (Refer to Instructions for Use Chapter 12.)
vacuum
250 mmHg (-0.0333MPa) Adjust vacuum to -0.0333MPa using the bellows
adjustment 5 min.
Vacuum adjustment unit. (Refer to Instructions for Use Chapter 12.)
Clean WBC transducer aperture Apply CELLCLEAN to a brush for clog removal
and clean the aperture. (Refer to Instructions for
Transducer Clean RBC transducer aperture 10 min. Use Chapter 12.)
Remove dust, etc. and wipe with gauze
Clean inside of Transducer
impregnated with ethanol. Then, wipe with dry