Om Combiscan 500-5

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

Operation

4.5.1.2.  Strip options


There are three possible options to be set: units, pad sensitivity and pad sequence on
printout.

4.5.1.2.1.  Measurement units


Results can be displayed in different units. Use the  control keys in MENU/SETUP/
STRIP/UNIT submenu to select the desired unit. The CombiScan® 500 supports the fol-
lowing measuring units:
 Conventional
 SI
 Arbitrary
 Conventional + Arbitrary
 SI + Arbitrary

4.5.1.2.2.  Sensitivity adjustment


The CombiScan® 500 provides semiquantitative results. In special cases, it may be nec-
essary to adjust the sensitivity of single parameters to special requirements of the user,
although it is not recommended by ANALYTICON. To adjust the sensitivity, enter MENU/
SETUP/STRIP/SENS submenu.
Pad measurement sensitivity can be set in two levels in both directions (+/-). The set val-
ues on the display are as follows:

– – 0    – 0    0    0 +    0 + +

2004/08/09 19:19:26
MENU/SETUP/STRIP

UNIT SENS. PADSEQ.

PAD SENSITIVITY
DEFAULT/USER DEFINE

DEFAULT SET QUIT

PAD SENSITIVITY
PAD NAME: BIL
SENSITIVITY: 0
  NEXT

When SET option is selected the first PAD name with the actual setting is displayed. After
selecting the SET option the proper sensitivity of each test pad can be selected, using the
 control keys. To enter the selected sensitivity of a parameter and to jump to the next
parameter, press the NEXT key. After having set the sensitivity, the program will automati-
cally return to the previous function menu. It is not possible to escape until all pad sensi-
tivities are set. To skip a pad without any modifications just select NEXT.
To get back to MAIN MENU without changing the actual settings, just press the QUIT
function key.
To reset the settings adjusted by the manufacturer for all pads select the DEFAULT option.
Going back to MAIN MENU without changing the actual settings, just press QUIT.

15
Operation

IMPORTANT! The adjustment of the pad sensitivity effects the results of the
measurement! When the reader is adjusted different from manufacturer setting
it will be indicated by asterisks in the footer of the printout!

4.5.1.2.3.  Pad sequence in printout


By default, the reader prints out the result in the sequence of the pads physical position
but it is also possible to customize the layout.
2004/08/09 19:19:26
MENU/SETUP/STRIP

UNIT SENS. PADSEQ.

PAD PRINTING SEQUENCE


DEFAULT/USER DEFINE

DEFAULT SET QUIT

PAD PRINTING SEQUENCE


REPORT LINE: 01
PAD NAME: BIL
  NEXT

This setup option effects only the printout and data transfer!

If the PADSEQ option is selected, you can restore either the default sequence or set up
your own sequence by selecting the SET option. Selecting SET you can choose pads in
a desired sequence of the printout lines. In the second row the line number is displayed
and the third line indicates the actual parameter name. Using the  control keys you can
perform changes. Pressing the NEXT option will display the next printout line. If you select
“End rep.” only the previously selected parameters will be printed out. When the last line
is reached, the program returns to PADSEQ submenu.

NOTE: You can not define empty printout lines or the same parameter twice or
more times.

4.5.1.3.  Clock and date settings


The built-in digital clock enables to store the date and time of the measurement. It is
working even if the reader is turned off or the power supply is not connected. The reader
supports different date and time formats.
2004/08/09 19:19:26
MENU/SETUP

OUTPUT STRIP CLOCK

2004/08/09 19:19:26
MENU/SETUP/CLOCK

SET 12/24 SEQUENCE

2004/08/09 19:19:26
ENTER NEW DATE & TIME

>>> � �

16