Professional Documents
Culture Documents
ALUP ABAC MC2 Controller Manual
ALUP ABAC MC2 Controller Manual
“<<<< mc >>>>” 2
CONTROL UNIT
USER’S MANUAL
1.0 INTRODUCTION
The “mc2” controller is designed specifically to be used in machinery for generating
compressed air.
Use of the controller for other reasons or its improper usage will void any and all
responsibility attributable to the manufacturer.
P1 P9
P2 P8
DY
P3 P7
P5 P6 L4
L1 L2
P4
L3 L5
COLUMN 1 COLUMN 20
20 characters can thus be displayed on each line.
<<<<mc2>>>>
Soft. Vers. : reference number relating to the software loaded into the memory
S/N : product serial number .
LINES III and IV also indicate the menus selected using the P7 key in ENTER
Pressing the P8 and P9 keys provides access to information relating to the main settings
and total number of hours, loaded, and maintenance.
At any point in the menu, press the P1 key or wait 60 seconds without pressing
any keys to return to the main menu, or previous menu.
>Time to mainten.<
Components life
Compon.Replacement
$
Hours of work
Time to maint.
Oil filter: 98 hrs
time remaining until next maintenance
e.g. oil filter change
Cyclically, all components and their
respective remaining hours are displayed.
page 9
CRG 001 PAGE 8
page 8
components life
replace components
Time remaining for
Replacement m a d e for air filter:1000hrs
time remaining for air filter replacement
air filter: OK?
oil filter:1000hrs
oil filter: OK? time remaining for oil filter replacement
Confirm replacement
air filter: OK? for example
work hours
Operating time
total: 2 hrs operating hours for motor
>Language <
Select language:
V ITALIAN
ENGLISH
etc...
Once the dialogue language has been
selected, confirm with ENTER (P7); for a few
seconds the tick mark will appear in column 1
>Measur.Unit <
Pressure unit:
unit: bar
unit: psi
confirm selection with P7;
OPERATION CARRIED OUT will appear,
page 11 returning to the “Unit of measure” menu
CRG 001 PAGE 10
page.10
Temperature unit:
unit: Celsius
unit: Fahr.
confirm selection with P7;
OPERATION CARRIED OUT will appear,
returning to the “Unit of measure” menu
>Date and time <
$
Set date and time
20:58 MAR 13/02/2000
The P8 and P9 keys set flashing values,
while the P2 and P3 keys select values to
be modified that are flashing.
>Display settings<
The P8 and
P9 keys
Contrast
modify Intensity : 68%
values
that are min 0 max 100%
confirmed with P7
(enter) while the P2 and Illumination ON
P3 keys select the
appropriate heading. Brightness: 100%
min 50 max 100%
Confirm selection with
P7; OPERATION
CARRIED OUT will Illumination OFF
appear, returning to
the “Display settings” Brightness: 12%
page 12 menu min 0 max 30%
CRG 001 PAGE 11
page 11
CONTINUOUS
Keeps the compressor always running
without taking empty operating times into
account .
>Local/rem/prog.ctrl
$
Selection of the Compress. control
operating mode is
carried out using the V LOCAL
P7 key (enter),
which will indicate Manages operation of the compressor in
the tick symbol in local mode, enabling functions on the local
column 1 of the keyboard.
display.
Return to
“Local contr./rem/
prog” menu
REMOTE
The compressor accepts load/empty or
L5 LED will light up motor on/off commands from units external
when the REMOTE to the machine.
or PROGRAM
modes are inserted.
PROGRAM
The compressor’s start-stop is managed by
weekly or daily programs.
page 13
CRG 001 PAGE 12
page 12
$
Selection of the Type remote control
operating mode is
carried out using the V ON/OFF
P7 key (enter),
which will indicate External control signals control the
the tick symbol in machine’s start and stop
column 1 of the
display.
Return to “Type of
remote control”
LOAD/ULOAD
menu External control signals control the intake
valve - load/empty
ON/OFF-LOAD/UNLOAD
Press ESC at any heading in the
External control signals control both the
present menu to return to selection
motor’s start-stop and the intake valve -
menu
load/empty
>Pressures <
Operating pressure
Max: 8.0 bar
The P8 and P9 The maximum setable pressure value, defined by the
keys modify machine’s base parameters: the limit beyond which
values. the compressor goes from loaded to empty.
Confirm
selection with
P7; OPERATION
CARRIED OUT will Min: 7.0 bar
appear, returning to
The minimum setable pressure value, defined by
the “Pressures” menu
machine’s base parameters: the limit beyond which
the compressor goes from empty to loaded.
page14
CRG 001 PAGE 13
page 13
>Timings <
Unload run time
val: 5 min
min 2 max 240 min
The time, in minutes, the compressor
remains in operation, after which it is in
pause mode. Starts up again when the
The P8 and pressure is less than the minimum value set
P9 keys
modify
values.
Confirm Time drain open
selection with P7;
OPERATION val: 5 sec
CARRIED OUT min 1 max 10sec
will appear, The opening time, in seconds, of the
returning to condensation discharge valve
the “Timings”
menu
>Programming mode<
Programming mode
V DAILY
P7 to confirm selection:
OPERATION Allows for activation of the same programing
CARRIED OUT for every day of the week
appears, returning to
“Reset prog.” menu
Selection of the
operating mode is
Programming mode
carried out using the WEEKLY
P7 key (enter), which
will indicate the tick Allows for activation of distinct programing
symbol in column 1 of for every day of the week
the display.
Return to “Program-
ing mode” menu
>Daily programming<
Every day
Start 1: -----
-
Set the first start hour using the P8
and P9 keys.
---- appear after reseting
The time is expressed in hrs-min
Monday
Stop 1: ----
For each day of -
Set the first stop hour using the P8
the week, there and P9 keys. The hour displayed will be the
are 3 “Starts” and same as the previous start.The time is
3 “Stops.” expressed in hrs-min
Tuesday
Wednesday
Thursday
Friday
Saturday
Sunday
Stop 3: -----
P7 to confirm selection: Set the last stop hour using the P8
OPERATION CARRIED and P9 keys.
OUT appears, returning The hour displayed will be the same as the
to “Weekly prog.” menu previous start
The time is expressed in hrs-min
>Diagnostics TEST<
Run diagnostic
$
TEST : OK?
Allows access to the display test (switches the
entire DY matrix on and off) and LED test, in
sequence.The L1 LED remains always lighted
page18
CRG 001 PAGE 17
page 17
Use this
menu to DIAGNOSTICS TEST
check if the Rotation dir. TEST
motor is
turning in
the correct Press < for TEST
direction
DIAGNOSTICS TEST
Rotation dir. TEST
Press < if OK
Press ESC if WRONG
This view allows the operator to check the
rotation direction again without the external
control units.
>Software version<
Software version:
<<<mc2>>> 1.0.0.1
13.0 MICROINTERRUPTIONS AND LINE POWER SHORTAGES
13.1 MICROINTERRUPTIONS
Microinterruptions occur when power is interrupted for very brief intervals
(typically for periods exceeding 40msec.), which can cause false switchovers
of the remote control switches. In this case, the equipment identifies the
malfunction and orders the compressor to shut down with the relative message.
13.2 LINE POWER SHORTAGES
If the line voltage is interrupted decisively during operation, the malfunction is
identified and a machine block message is then indicated
ANALOG. INPUT
EXCESSIVE PRESSURE PRESSURE X
SENSOR INTERR. OR SHORT. PRESSURE X
SENSOR POWER ABSENT PRESS. / TEMP. X
TEMPERATURE <0°C TEMP. X
TEMPERATURE <5°C TEMP X
TEMP <10° MAX SET TEMP X
TEMP >= MAX SET TEMP X
SENSOR INTERR. OR SHORT. TEMP X
No. starts/hour
excessive - block
To reset this error two solutions can be adopted:
solution a) wait for the subsequent hour to elapse, reset using the RESET command;
then press the START key to re-start the compressor.
FORCED START
procedure
The flashing L3 LED will remain lighted; press the RESET key
to return to normal operating conditions.
An error message will remain in the MESSAGES menu.