ElemeNtS®

Combustion UV/Chemiluminescence Analyzer for Total Sulfur and Nitrogen

Technical Offer
Version 2020.5

The global market leader for Elemental Combustion analysis of total Sulfur and total Nitrogen.

• Unrivaled Performance: under 30 ppbw LOD according ISO11843 and Excellent Precision
• Robust & reliable: 4 orders of LDR, Long term Stability
• Easiest to use: easy access, 10” touchscreen user interface & versatile software
• Lab bench space saver: only 50 cm wide.
Copyright © 2020 by Petroleum Analyzer Company, L.P.

All text, graphics, design, and other works contained herein are the copyrighted works of
Petroleum Analyzer Company, L.P (PAC). All rights reserved. Any redistribution or reproduction of any materials contained herein is
strictly prohibited without the express written permission of the copyright holder.

The PAC trademark, the PAC service mark, PAC’s logo trademark, the ElemeNtS trademark, and Petroleum Analyzer Company, L.P.’s
distinctive color trademark as used on its instruments and in these materials, are trademarks, service marks, registered trademarks,
and/or registered service marks of Petroleum Analyzer Company, L.P. in the U.S. and other countries and are its sole and exclusive
property.

Other trademarks contained within these materials are the property of their respective owners.

Information

The information contained within these materials is subject to change without notice. PAC makes no warranties or guarantees,
either expressed or implied in these materials, including the warranty of merchantability or fitness for a particular purpose. At any
time, PAC may modify these materials, its instruments or its programs without notice and subsequent versions of these materials may
contain different information. These materials could contain technical inaccuracies and/or typographical errors.

PAC does not assume responsibility for the accuracy of any translation of these materials.

About PAC

PAC is an international manufacturing and service organization with a portfolio that spans
petroleum, petrochemical, biofuels, environmental, food and beverage, pharmaceutical and industrial analysis solutions. PAC
provides advanced testing equipment for laboratory, process on-line and field use for entities ranging from the smallest to the largest
enterprise businesses. PAC offers analytical solutions for a wide variety of applications, including chromatographic systems and
detectors, elemental, laboratory, and on-line process analyzers, software applications, and spectroscopy. More information about
PAC is available at http://www.paclp.com.

Rotterdam, The Netherlands

Publication Date: 23-07-2020

Document Number: 2020.5

Page 2 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


The symbol indicates PAC’s compliance with the European Union (EU) directives and standards listed below:

• EMC Directive 2014/30/EU

• EN 61326-1:2013 “Electrical equipment for measurement, control and laboratory use. EMC requirements. Particular

requirements. Test configurations, operational conditions and performance criteria for sensitive test and measurement

equipment for EMC unprotected applications”

• This system classified for use in Class B commercial environment and is not designed or intended for use in other EMC

environments. The user of this system is obliged to properly use and install the system and to take all steps necessary to

remove sources of interference to telecommunications or other devices.

• Low Voltage Directive 2014/35/EU

• EN 61010-1: 2010 “Safety requirements for electrical equipment for measurement, control and laboratory use”

• The ElemeNtS has been designed and tested according to 61010-01 (Safety requirements for electrical equipment for

measurement, control and laboratory use), tested by TUV Certification B.V., Leek, the Netherlands

• Electrical safety tests in accordance with EN61010-1:2010

• (Report form J with the US and Canadian national deviations)

• Safety tests in accordance with EN61010-2-081:2015

• (Report form F version with the US and Canadian national deviations)

• Safety tests in accordance with EN61010-2-010:2014

• (Safety requirements for electrical equipment for measurement, control and laboratory use)

• cTUVus certification, safety testing applicable to US and Canadian standards

• EMC, EN55011:2016 (Industrial, scientific and medical equipment)

• FCC: 47 CFR Part 15 RF Devices

• Canada ICES-003 Issue 6 (Validation only in combination with FCC Part 15B testing

• EMC test in accordance with IEC 61326-2014

Page 3 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


CONTENT :

1. Introduction ................................................................................................................. 5
2. System description ....................................................................................................... 6
3. Theory of Operation ..................................................................................................... 7
Sulfur: principle of operation ......................................................................................... 7
Nitrogen: principle of operation ...................................................................................... 7
General system flow diagram ......................................................................................... 8
4. Methods & Compliance ................................................................................................. 9
5. System Configurations ................................................................................................ 12
ElemeNtS Sulfur ....................................................................................................... 12
ElemeNtS Nitrogen ................................................................................................... 13
ElemeNtS Sulfur & Nitrogen ......................................................................................... 14
6. Accessories / Options ................................................................................................. 15
Model 749 for Liquid Sampling ...................................................................................... 15
Options 749 sampler .................................................................................................. 16
Accura for Gas & LPG Compressed Gas Sampling ................................................................ 17
Single Shot Autosampler for Liquid Injections .................................................................... 18
Nitrogen Interference kit for ElemeNtS ........................................................................... 19
PC and Software ...................................................................................................... 20
7. Consumable & Spare part Kits ..................................................................................... 21
ElemeNtS Consumables kits ......................................................................................... 21
ElemeNtS Spares kit Sulfur .......................................................................................... 23
ElemeNtS Spares Kit Nitrogen ....................................................................................... 24
ElemeNtS Spares Kit Sulfur + Nitrogen ............................................................................. 25
749 Autosampler Kits ................................................................................................. 26
SSA Kits ................................................................................................................. 27
8. Consumables ElemeNtS & Accesories ........................................................................... 28
ElemeNtS consumables............................................................................................... 28
749 Autosampler Consumables ...................................................................................... 28
9. Calibration and Reference Standards ........................................................................... 29
Liquid standards ...................................................................................................... 29
Calibration Gasses .................................................................................................... 31
10. Technical Requirements ............................................................................................. 32
Gas supply & connections ........................................................................................... 32
Power requirements .................................................................................................. 33
Physical specifications ............................................................................................... 33
Environmental conditions ............................................................................................ 34
Compliance & certifications ......................................................................................... 35
Computer Requirements ............................................................................................. 36

Page 4 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


1. Introduction
There is an increasing demand for fast and accurate analysis of unwanted contaminants such as Sulfur and
Nitrogen in the worldwide Hydrocarbon Processing Industry (HPI).

The overall Sulfur content of crude oil coming into the refineries has been increasing since 1985. Sulfur
analysis in crude oil is very important because it causes difficulties, such as corrosion of metals, in processing
the oil and because of air pollution from the burning of fuels processed from crude oils having high Sulfur
content. Sulfur compounds can also poison catalysts causing downtime and loss of revenue. In order to
process higher Sulfur crude oils, equipment must be built from more expensive alloys to provide higher
corrosion resistance.

The past decade new emission regulations (air quality improvement programs) have passed to decrease the
amount of Sulfur in finished products. In a huge number of countries the Sulfur content in Gasoline & Diesel
has been reduced to a level of less than 10 ppm. Due to these new Sulfur regulations, there have been
several techniques that have appeared on the market to measure individual Sulfur compounds and total
Sulfur at trace level in fuels.

There is a need for Nitrogen testing in refineries as some Nitrogen compounds can form salts that can cause
heat exchanger fouling, blocking filters within the refinery process. Nitrogen compounds can also
temporarily deactivate catalysts used in fluid catalytic cracking (FCC) units. Production may have to run at
lower rates (less profit) if the FCC unit is offline. Wastewater may contain high levels of Nitrogen. Excessive
release to the environment can lead to build up of nutrients, affecting the water quality.

The ElemeNtS analyzer combines the testing of Sulfur and Nitrogen for monitoring and protecting key
processes in the refining industry, examples of processes are listed below:

▪ Corrosion protection
→ Protect refinery parts and electronics against corrosive effects of Sulfur

▪ Catalyst protection
→ Sulfur and Nitrogen may act as a catalyst poison in the refinery industry

▪ Environmental protection
→ Protect people, animals, plant and aquatic life from the harmful effects of Sulfur & Nitrogen
in the environment

▪ Employee protection (reduced sampling handling)


→ Automates sample preparation, reducing exposure potential to hazardous material

Page 5 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


2. System description
PAC ElemeNtS is the most advanced, analytical system available for the detection of total Nitrogen and/or
total Sulfur content in liquid, or gaseous materials. Three configurations are available: ElemeNtS-N for
Nitrogen analysis, ElemeNtS-S for Sulfur analysis and ElemeNtS N-S for simultaneous Nitrogen and Sulfur
analysis.

The ElemeNtS detects these substances through a uniquely efficient process that includes Ultra Violet
Fluorescence (UVF) and Chemiluminescence (CLD). The ElemeNtS includes numerous benefits, including:

• Analyzes a Wide Application Range all in One Instrument


o Quick Analysis of Various Sample Types for a variety of industries, including Hydrocarbon
fractions, chemicals, foods and beverages, etc.
o ElemeNtS can analyze liquids, LPG’s and gases quickly and accurately
o Long term stability for continuous use

• Automation and Stable Operation Reduce Operator Intervention


o Factory tuned and tested analyzer; the analyzer will be factory tested with a number of
Reference standards (like gasoline & diesel).
o High capacity liquid autosampler (324 samples) reduces need for operators to be present
when running samples
o Long term stability reduces calibrations by operators for any measurement range
o Wide dynamic range means user can run samples with a broad concentration range (10 4)
utilizing a single calibration and method, which minimizes running samples under
multiple methods/instruments
o Control and processing are accomplished by an embedded PC and microprocessor
controller.

Page 6 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


3. Theory of Operation
The principle of operation for Nitrogen and Sulfur analysis begins with the complete, high-temperature
oxidation of the entire sample matrix illustrated in Equation (1). The sample is combusted with Oxygen at
a temperature of 1050ºC. Oxidation products include CO 2, H2O, NO, SO2, and various other oxides (designated
MOX below). The combustion gases are routed through a membrane drying system to remove all the water
and then to the Sulfur and/or Nitrogen detector module(s) for quantitation.

R-N + R-S + O2 → CO2 + H2O + NO + SO2 + MOX (1)

Nitrogen and Sulfur calibration standards are analyzed to produce calibration curves. When samples of
unknown Nitrogen and/or Sulfur content are analyzed, the ElemeNtS software compares the raw sample
data to the calibration curve to generate and report Nitrogen and/or Sulfur concentrations.

Sulfur: principle of operation

SO2 in the combustion gases is exposed to ultraviolet radiation of a specific wavelength as shown in Equation
(2) and excited to a higher energy state. Decay to ground state emit radiation in the form of Sulfur
fluorescence. This fluorescence is detected by a photomultiplier tube (PMT) and is proportional to the
amount of Sulfur in the original sample.

SO2 + hνex → SO2* → SO2 + hνem (2)

Nitrogen: principle of operation

In Equation (3), the NO is reacted with O3 (ozone), produced by an onboard ozone generator, to form NO 2*
(excited Nitrogen dioxide). As the metastable species decays, a photon of light is emitted at specific
wavelengths and detected by a photomultiplier tube (PMT). This chemiluminescent emission is specific for
Nitrogen and is proportional to the amount of Nitrogen in the original sample. Only chemically bound
Nitrogen is detected — atmospheric, diatomic Nitrogen (N2) is not.

NO + O3 → NO2* + O2 → NO2 + hνem + O2 (3)

Page 7 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


General system flow diagram

Figure 3-1: General gas supply flow diagram

Note: The ozone generator and scrubber are only used in combination with the Nitrogen detector option.

Page 8 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


4. Methods & Compliance
The ElemeNtS system is a modular analytical system based on UV-fluorescence and Chemiluminescence
analysis techniques for detecting (only or simultaneously) Sulfur and Nitrogen in LPG, gaseous or liquid
matrices.

Linear
Instrument Application Range LOD Dynamic Repeatability
Range
ElemeNtS Sulfur Trace ppb to % level 20 ppbW 10e4 <1% at 1 ppm

ElemeNtS Nitrogen Trace ppb to % level 30 ppbW 10e3 <1% at 1 ppm

S: 20 ppbW
ElemeNtS Sulfur & Nitrogen Trace ppb to % level <1% at 1 ppm
N: 30 ppbW

NOTE: The above referenced LOD’s have been determined under optimized instrument conditions, using hydrocarbon
matrices and following the procedure as described in ISO 11843. For specific applications LOD’s need to be validated.

Specifications on specific methods are listed in the test documentation provided with the complete system
and dedicated Application notes.

STANDARD TEST METHOD COMPLIANCY

ASTM EN & DIN & IP & ISO Others

Sulfur

D5453 EN 15486 UOP 989


D6667 EN 17178 GOST R 56866
D7551 (EN) ISO 20846 GOST R 56342
D7183 ISO 20729 UOP 987 part A
IP 554 SH/T 0689
IP 490 JIS K 2541

Nitrogen

D4629 EN 12260 UOP 936


D5176 DIN 51444 UOP 981
D7184 UOP 971
SH/T 0657
GB/T 17674
JIS K 2609

Page 9 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


In below table all methods are specified with the advised system configuration.

Product Method Detection Vertical


SULFUR &
SULFUR NITROGEN
NITROGEN
(Natural) Gas ASTM D7551 UVF ● ●
LPG ASTM D5453 UVF ● ●
ASTM D6667 UVF ● ●
ASTM D7551 UVF ● ●
ISO 20729 UVF ● ●
EN 17178 UVF ● ●
UOP 936 CLD ● ●
UOP 989 UVF ● ●
GOST R 56866 UVF ● ●
Gasoline ASTM D5453 UVF ● ●
EN ISO 20846 UVF ● ●
GOST R 56342 UVF ● ●
ASTM D4629 CLD ● ●
IP490 UVF ● ●
DIN 51444 CLD ● ●
UOP 981 CLD ● ●
UOP 987 part A UVF ● ●
SH/T 0657 CLD ● ●
SH/T 0689 UVF ● ●
Ethanol EN 15486 UVF ● ●
IP554 UVF ● ●
ASTM D5453 UVF ● ●
ASTM D4629 CLD ● ●
UOP 987 part A UVF ● ●
Jet ASTM D5453 UVF ● ●
DIN 51444 CLD ● ●
EN ISO 20846 UVF ● ●
UOP 987 part A UVF ● ●
Diesel ASTM D4629 CLD ● ●
ASTM D5453 UVF ● ●
IP 490 UVF ● ●
DIN 51444 CLD ● ●
UOP 987 part A UVF ● ●
Aromatics ASTM D7183 UVF ● ●
ASTM D7184 CLD ● ●
DIN 51444 CLD ● ●
UOP 971 CLD ● ●
UOP 981 CLD ● ●
UOP 987 part A UVF ● ●
Heating Oil DIN 51444 CLD ● ●
ASTM D4629 CLD ● ●
Water EN 12260 CLD ● ●
ASTM D5176 CLD ● ●

Page 10 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


Common use method details:

Repeatability Reproducibility
Method Element Matrix Range
(r) (R)
<400 mg/kg: r =
<400 mg/kg: R =
0.1788X0.75
Liquid 0.5797X0.75
ASTM D5453 S 1 – 8000 mg/kg
Hydrocarbons
>400 mg/kg: r =
>400 mg/kg:R = 0.1267 X
0.02902 X

Automotive
EN ISO 20846 S 3 – 500 mg/kg Table, see method Table, see method
fuels

EN ISO 15486 S Ethanol 5 – 20 mg/kg r = 0.02X + 0.25 R = 0.16X + 1.67

1 – 100 mg/kg 1 – 100 mg/kg


r = 0.1152X R =0.3130X
ASTM D6667 S LPG 1 – 196 mg/kg
1 – 196 mg/kg 1 – 196 mg/kg
r = 0.2707X0.7 R =2.3205X0.7

ASTM D7183 S Aromatics 0.5 – 100 mg/kg See method See method

LPG/ gaseous
ASTM D7551 S 1 – 200 mg/kg See method See method
hydrocarbons

Liquid
UOP 987 – part A S 100 – 1500 ng/g See method See method
Hydrocarbons

Liquid
ASTM D4629 N 0.3 – 100 mg/kg r = 0.1825X0.5149 R = 0.8094X0.5149
Hydrocarbons

ASTM D5176 N Water 0.5 – 1000 mg/l See method See method

ASTM D7184 N Aromatics 0.1 – 1.2 mg/kg See method See method

0.1 ppmW –
UOP 936 N LPG 5.9 ± 0.08 ppmW Insufficient data
3 W%

Liquid
UOP 981 N 50 – 300 ng/g See method See method
Hydrocarbons

Page 11 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


5. System Configurations
ElemeNtS Sulfur

Ordering No. Description

ElemeNtS Combustion Analyzer, Sulfur,


38.00.001
100-240 Vac

Hardware

ELEMENTS mainframe

Included/Pre-installed

Installation kit*
System Testing Documentation
System Operation Manual
Quick guides (installation, maintenance, etc.)
Calibration Standards (part of installation kit)
Pyrolysis tube
Software (ELEMENTS for IRIS plugin License)
IRIS CORE License

Not included/To configure separately

PC
Monitor
Printers
Autosampler
Syringes

* Installation kit contains: Connection materials, Copper Tubing, set of tools, Oxygen/Carrier gas
filters, Calibration sets (Trace-Low-Mid-High), set of reference standards, vials/pre-slit caps, power
cord for installations, Ethernet cable for communication.

Page 12 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


ElemeNtS Nitrogen

Ordering No. Description

ElemeNtS Combustion Analyzer, Nitrogen,


38.00.002
100-240 Vac

Hardware

ELEMENTS mainframe

Included/Pre-installed

Installation kit*
System Testing Documentation
System Operation Manual
Quick guides (installation, maintenance, etc.)
Calibration Standards (part of installation kit)
Pyrolysis tube
Software (ELEMENTS for IRIS plugin License)
IRIS CORE License

Not included/To configure separately

PC
Monitor
Printers
Autosampler
Syringes

* Installation kit contains: Connection materials, Copper Tubing, set of tools, Oxygen/Carrier gas
filters, Calibration sets (Trace-Low-Mid-High), set of reference standards, vials/pre-slit caps, power
cord for installations, Ethernet cable for communication.

Page 13 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


ElemeNtS Sulfur & Nitrogen

Ordering No. Description

ElemeNtS Combustion Analyzer, Sulfur + Nitrogen,


38.00.003
100-240 Vac

Hardware

ELEMENTS mainframe

Included/Pre-installed

Installation kit*
System Testing Documentation
System Operation Manual
Quick guides (installation, maintenance, etc.)
Calibration Standards (part of installation kit)
Pyrolysis tube
Software (ELEMENTS for IRIS plugin License)
IRIS CORE License

Not included/To configure separately

PC
Monitor
Printers
Autosampler
Syringes

* Installation kit contains: Connection materials, Copper Tubing, set of tools, Oxygen/Carrier gas
filters, Calibration sets (Trace-Low-Mid-High), set of reference standards, vials/pre-slit caps, power
cord for installations, Ethernet cable for communication.

Page 14 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


6. Accessories / Options
Model 749 for Liquid Sampling

Ordering No. Description Pictures

749 Liquid autosampler for ElemeNtS, Including


12.80.200
Accessory Kit 100-240 Vac

Model 749 Robotic Liquid Autosampler with capacity of up to six trays, 324 total vial capacity/
standard 2 ml VT 54 tray; Microprocessor controlled sample and method storage. Smart and easy
controllable handheld terminal and display.

Hardware

Model 749 Robotic Liquid Autosampler

Included/Pre-installed

PALbase 534 X/Y unit with PAL Control Board


PAL LSI z-head for D7/57 syringes
PALterminal
Safety guard
PALpower and cables
PAL standalone support, 160 mm (2 pcs)
Large Wash Module 2x 100mL solvent vials and waste
Trayplate for 6x MT, DW VT54 or VT15 (2pcs)
Sample Rack for 54 vials of 2 ml (6 pcs)
Vials + Caps, Screw Top, 2 ml, 100 pc (PN:32057)
Assy, 25 ul Syringe + needle (PN:106289)

Includes Model 749 accessory kit for ELEMENTS, containing of Vials/Caps, 25 ul Syringe, mounting
materials, 749 manual and installation Quick guide

Page 15 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


Options 749 sampler

Ordering No. Description

PAL3-Trayplate Tray holder for VT 54 and VT15 vial trays, a maximum of 2


standard tray holders can be configured per 749.

PAL3-WashLarge Large Wash Module for 2x 100mL solvent vials and one waste
position

PAL3- Temperature controlled Peltier stacks for 2 trays, or plates


PeltierStack2DW
can be added to prevent degradation or to heat samples for
derivatization reactions or kinetic studies. 4 – 40°C. This
module will replace 1 tray holder.

NOTE: PAL3-LowNdlGdeRepl needed for operation, need to be


ordered separately.

PAL3-Liquidcooler Tray holder and sample tray for housing up to 32 vials of 10/20
ml in the temperature range between 4° C - 70° C. External
circulator bath for heating/cooling not included. This module
will replace 1 tray holder.

Note: Contact PAC for use of 2 ml vials.

PAL3-Vortexmixer The Vortex Mixer Module offers efficient mixing for dilution
/extraction.

• Standard vial sizes: 2 mL / 10 mL / 20 mL


• 1 additional slot for custom specific vials
• Provides efficient mixing with up to 2000 rpm

This module will replace 1 tray holder.

Page 16 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


Accura for Gas & LPG Compressed Gas Sampling

Ordering No. Description Pictures

36.00.003 Accura for ElemeNtS, 110-240 Vac

Hardware

Mainframe: ACCURA sampling Box


Connections: Sample in (1/8” Male Sulfinert)
From/To device (1/8” Male Sulfinert)
Vent out (1/8” Male)
He/Ar in (1/8” Male)
Vaporizer: Max 150°C
Sample inlet: Max 30 Bar pressure
Injection: 6 port Sampling Valve, with 10 ml std sample loop valve

Included/Pre-installed

Startup kit*
System Testing Documentation
System Operation Manual

Not included/To configure separately

Calibration Standards (Cylinders)


Regulators for calibration gasses

* Startup kit contains, connection materials to and from ElemeNtS, power cable to region, 0.5µm
Sulfinert treated sample inlet filter and communication cable.

Page 17 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


Single Shot Autosampler for Liquid Injections

Ordering No. Description Pictures

Single Shot Autosampler for ElemeNtS, Including


38.50.001
Accessory Kit

The Single Shot Autosampler (SSA) for ElemeNtS provides a cost-effective solution for injecting a
single replicate sample by syringe. The SSA uses a dual motor design to facilitate automatic needle
insertion through the septum which guarantees optimum repeatability. The SSA is fully
programmable and receives power and automatic control through the ElemeNtS.

Hardware

Single Shot Autosampler for Liquid Injections

Included/Pre-installed

Mounting Materials & Cable


Vials + Caps, Screw Top, 2 ml, 100 pc (PN:32057)
Assy, 25 ul Syringe + needle (PN:106289)
Manual for SSA, (PN: 23070.160)

Page 18 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


Nitrogen Interference kit for ElemeNtS

Ordering No. Description Pictures

23800.550 /
Nitrogen Interference kit for ElemeNtS
38.00.550

The analysis of total Sulfur in ULSD samples by oxidative combustion followed by UVF detection can
be hindered by interference from certain Nitrogen components. This effect is caused by a cross-
sensitivity to NO, which fluoresces in the same wavelength range as SO2. Depending on the Nitrogen
content of the sample, this can lead to false positive Sulfur results up to 1,7 mg/kg at 160 mg/kg
Nitrogen.
To increase the cetane index of diesel fuels, alkyl nitrates, more commonly 2-Ethylhexyl Nitrate
(2-EHN), is added to the diesel. The relative high Nitrogen concentration might cause a serious
positive bias in the Sulfur determination.
By means of an optimized Sulfur detector the ElemeNtS can efficiently reduce the interference of
these type of compounds without the need of software correction or added ozone supply.
The Nitrogen Interference kit is available as a separate option (23800.550) or build in and factory
tested in the ElemeNtS (38.00.550).

Note: System having the Nitrogen Interference kit will have an LOD of 60 ppbW for Sulfur.

Hardware

Nitrogen Interference kit

Included/Pre-installed

Optical Interference Filter


Filter Mounting Plate
Filter Mounting Plate Screws (4 pcs)
18 x 1.5 O-ring (2 pcs)
33 x 1.5 O-ring
Installation Procedure

* The Nitrogen Interference kit will be included in the factory test of the system when ordered
together with the ElemeNtS. On-site installation can only be performed by a PAC certified service
engineer.

Page 19 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


PC and Software

Ordering No. Description Pictures

12.41.061 ElemeNtS PC Bundle

The software provides complete automated operation including unlimited method storage,
hardware setting control, data handling, real-time peak display, complete multiple peak storage,
and remote operation capabilities. The data can be exported into a CSV file for import into a LIMS
system or printed in a report format.

Hardware

PC Bundle for ElemeNtS

Included/Pre-installed*

PC with Windows 10 operating system


Monitor
Mouse
Keyboard
IRIS for ElemeNtS

* The PC will be included in the factory test of the system when ordered together with the
ElemeNtS. All software will be installed (media included in the shipment) and test data is available
in the database of the IRIS software.

Page 20 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


7. Consumable & Spare part Kits

ElemeNtS Consumables kits

ELEMENTS, Consumables/PM Kit for Nitrogen/Sulfur configurations,


38.00.200
consisting of:

Carrier gas filter


Disposable In-line filter PTFE
Drier assembly
Membrane dryer
O-ring elbow connection pyrotube
O-ring kit Sulfur reaction chamber
O-ring Nitrogen reaction chamber
Oxygen gas filter
Scrubber assembly
Septa 12.7 mm Teflon-Backed S70 (100 pcs)
Septum cap for pyrotube, ElemeNtS
Snap cap (PE) and septum (Silicon/PTFE) for 10mL/100mL vials
Syringe 25R-RTC/RSH-GT-MS1 25 ul Diamond MS
Syringe Needle NMS10/50-P-RTC/RSH-57063(0.24)-C
Tubing, PTFE 1.5x3.2 (5m)

ELEMENTS, Consumables Kit for Sulfur only configurations, consisting


38.00.201
of:

Carrier gas filter


Disposable In-line filter PTFE
Drier assembly
Membrane dryer
O-ring elbow connection pyrotube
O-ring kit Sulfur reaction chamber
Oxygen gas filter
Scrubber assembly
Septa 12.7 mm Teflon-Backed S70 (100 pcs)
Septum cap for pyrotube, ElemeNtS
Snap cap (PE) and septum (Silicon/PTFE) for 10mL/100mL vials
Syringe 25R-RTC/RSH-GT-MS1 25 ul Diamond MS
Syringe Needle NMS10/50-P-RTC/RSH-57063(0.24)-C
Tubing, PTFE 1.5x3.2 (5m)

Page 21 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


ELEMENTS, Consumables/PM Kit for Nitrogen configurations, consisting
38.00.202
of:

Carrier gas filter


Disposable In-line filter PTFE
Drier assembly
Membrane dryer
O-ring elbow connection pyrotube
O-ring Nitrogen reaction chamber
Oxygen gas filter
Scrubber assembly
Septa 12.7 mm Teflon-Backed S70 (100 pcs)
Septum cap for pyrotube, ElemeNtS
Snap cap (PE) and septum (Silicon/PTFE) for 10mL/100mL vials
Syringe 25R-RTC/RSH-GT-MS1 25 ul Diamond MS
Syringe Needle NMS10/50-P-RTC/RSH-57063(0.24)-C
Tubing, PTFE 1.5x3.2 (5m)

Page 22 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


ElemeNtS Spares kit Sulfur

38.00.101 ELEMENTS, Spares Kit Sulfur, consisting of:

Valve NO, 2 port, 24V, 1/8" NPT connect


Valve NC, 2 port, 24V, 1/8" tube, straight
Connector straight, 1/8 tube to 10/32"
Reducing Union, 1/4" tube x 1/8"tube, PTFE/PVDF
Pressure regulator Pin 0-10bar, Pout 0-5bar, M5, non-relieving
O-ring reaction chamber 28x1.5
O-ring, 33 x 1.5
TUBE 1/16", .020"/.50MM 20CM LONG, 316SS
Tubing black, PTFE, 1/4"
Tubing black, PTFE, 1/8"
Fuse, T 16A, 5x20mm
Furnace Thermocouple
Pyro tube 90° tube connector incl Rotalex receiver
O-ring for 90° tube connector to Pyro tube
Clamp for 90° tube connector to Pyro tube
Pyro tube
Pyro tube septum cap
Sulfur Module ND filter
ZN UV lamp ElemeNtS, with 3-pin connector
O-RING 1.614" ID. x .070 S70
O-RING.799" ID x .103 S70
O-RING .864" ID. x .070 S70
Lens, 50mm fl, 25mm Dia., Uncoated Plano Convex
Filter UV Cs-7-54
Filter, Optical, 214nm Bandpass, 25mm Dia., Non-standard Thickness

Page 23 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


ElemeNtS Spares Kit Nitrogen

38.00.102 ELEMENTS, Spares Kit Nitrogen, consisting of:

Valve NO, 2 port, 24V, 1/8" NPT connect


Valve NC, 2 port, 24V, 1/8" tube, straight
Connector straight, 1/8 tube to 10/32"
Reducing Union, 1/4" tube x 1/8"tube, PTFE/PVDF
Pressure regulator Pin 0-10bar, Pout 0-5bar, M5, non-relieving,
O-ring reaction chamber 28x1.5
O-ring, FKM, 33 x 1.5
TUBE 1/16", .020"/.50MM 20CM LONG, 316SS
Pre-cut tube1500 mm, 1/16", 0.015"
Tubing black, PTFE, 1/4"
Tubing black, PTFE, 1/8"
Fuse, T 16A, 5x20mm
Ozone control assy
Fan 24V model 40 x 40x 20 assy
Furnace Thermocouple
Pyro tube 90° tube connector incl Rotalex receiver
O-ring for 90° tube connector to Pyro tube
Clamp for 90° tube connector to Pyro tube
Pyro tube
Pyro tube septum cap
Filter, Optical, Nitrogen Detector, 1.25 Diameter

Page 24 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


ElemeNtS Spares Kit Sulfur + Nitrogen

38.00.103 ELEMENTS, Spares Kit Sulfur + Nitrogen, consisting of:

Lens, 50mm fl, 25mm Dia., Uncoated Plano Convex


ZN UV lamp ElemeNtS, with 3-pin connector
Pyro tube septum cap
Pyro tube
Clamp for 90° tube connector to Pyro tube
O-ring for 90° tube connector to Pyro tube
Pyro tube 90° tube connector incl Rotalex receiver
Furnace Thermocouple
Fan 24V model 40 x 40x 20 assy
Ozone control assy
Fuse, T 16A, 5x20mm
Tubing black, PTFE, 1/8"
Tubing black, PTFE, 1/4"
Pre-cut tube1500 mm, 1/16", 0.015"
TUBE 1/16", .020"/.50MM 20CM LONG, 316SS
O-ring, FKM, 33 x 1.5
O-ring reaction chamber 28x1.5
Pressure regulator Pin 0-10bar, Pout 0-5bar, M5, non-relieving,
Reducing Union, 1/4" tube x 1/8"tube, PTFE/PVDF
Connector straight, 1/8 tube to 10/32"
Valve NC, 2 port, 24V, 1/8" tube, straight
Valve NO, 2 port, 24V, 1/8" NPT connect
Sulfur Module ND filter
Filter, Optical, 214nm Bandpass, 25mm Dia., Non-standard thickness
Filter UV Cs-7-54
O-RING .864" ID. x .070 S70
O-RING.799" ID x .103 S70
O-RING 1.614" ID. x .070 S70
Filter, Optical, Nitrogen Detector, 1.25 Diameter

Page 25 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


749 Autosampler Kits

PAL3-LSI-PM-Kit 749 Autosampler Preventive Maintenance Kit, consisting of:

Vial Teaching Kit


Test Label Barcode
Lubricant
Brush
Spare Screw for fixation of PALhead (3 pcs)
Kit Wash Vials
Soft Tissue for outer surface cleaning
Rough Tissue for degreasing X-,Y-,Z-Axes or Guide Rollers

38.00.500 ELEMENTS, Model 749 accessory kit, consisting of:

Vial, 2ml clear glass, with screw cap & PTFE/Silicone, Pre-slit, 100 pc
Manual, 749 ElemeNtS
SCREW, M5X20, TORX RAISED CHEESE HEAD, SS
Syringe Needle NMS10/50-P-RTC/RSH-57063(0.24)-C
Syringe 25R-RTC/RSH-GT-MS1 25 ul Diamond MS
Snap cap (PE) and septum (Silicon/PTFE) for 10mL/100mL vials

Page 26 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


SSA Kits

38.50.100 Spare parts kit for Single Shot Autosampler

Door assembly SSA


Plunger transporter assembly SSA
Syringe transporter assembly SSA

38.50.300 Refurbished exchange unit for Single Shot Autosampler

Exchange unit for an SSA requiring major repairs. A replacement unit will be
shipped, the defective unit must be returned to the factory using the same
packing material. A label and RMA number will be provided.

- Applicable for units ≤ 5 years of age.


- 38.50.001 will be invoiced if defective unit is not received within 1 month after
shipment of 38.50.300

Page 27 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


8. Consumables ElemeNtS & Accesories
ElemeNtS consumables

Ordering No. Description

69091 Septa, 12.7 mm, Teflon-Backed, S70 (100 pcs)


70002 Septa, 12.7 mm, Teflon-Backed, S70 (12 pcs)
70058 MEMBRANE DRYER
00.80.090 Vial, 2ml cl.glass, screw cap & PTFE/Sil Septa, Pre-slit (2)
38.09.100 Scrubber Assy
38.09.200 Drier Assy
38.10.100 Disposable In-line filter PTFE
71033 Carrier gas filter
10.30.065 Oxygen gas filter

749 Autosampler Consumables

Ordering No. Description

106289 Complete assembly, 25 ul SYRINGE, AUTOSAMPLER 749


106290 Complete assembly, 10 ul SYRINGE, AUTOSAMPLER 749
106291 Complete assembly, 50 ul SYRINGE, AUTOSAMPLER 749
106292 Complete assembly, 100 ul SYRINGE, AUTOSAMPLER 749
106285 25 ul SYRINGE BODY, AUTOSAMPLER 749
106286 10 ul SYRINGE BODY, AUTOSAMPLER 749
106287 50 ul SYRINGE BODY, AUTOSAMPLER 749
106288 100 ul SYRINGE BODY, AUTOSAMPLER 749
106284 Replaceable NEEDLE, CONE TIPPED, 57mm x .63mm OD AUTOSAMPLER 749
00.80.090 Vial, 2ml clear glass, screw cap & PTFE/Sil Septa, Pre-slit 100 / Pkg

Page 28 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


9. Calibration and Reference Standards
Liquid standards

▪ Chosen for purity and ASTM compliance


▪ Calibration for ASTM methods D4629, D5453 and D5762
▪ “Certificate of Analysis” supplied with each set
▪ Blank matrix included with each set
▪ Trace level standards for calibration of ElemeNtS analyzers: Nitrogen & Sulfur

Nitrogen Source (Isooctane matrix): Pyridine


Sulfur Source (Isooctane matrix): Dibenzothiophene

Part
Description Levels Quantity Matrix
number
20001.719 Sample box NS calibration, Full range Blank – 1000 mg/L 24 * 1.5 ml Iso-C8

20001.700 Sample box NS calibration, Blank 0 6 * 1.5 ml Iso-C8

20001.701 Sample box NS calibration, 0.05 0.05 mg/L 6 * 1.5 ml Iso-C8

20001.702 Sample box NS calibration, 0.1 0.10 mg/L 6 * 1.5 ml Iso-C8

20001.703 Sample box NS calibration, 0.25 0.25 mg/L 6 * 1.5 ml Iso-C8

20001.704 Sample box NS calibration, 0.5 0.50 mg/L 6 * 1.5 ml Iso-C8

20001.705 Sample box NS calibration, 0.75 0.75 mg/L 6 * 1.5 ml Iso-C8

20001.706 Sample box NS calibration, 1.0 1.0 mg/L 6 * 1.5 ml Iso-C8

20001.707 Sample box NS calibration, 2.5 2.5 mg/L 6 * 1.5 ml Iso-C8

20001.708 Sample box NS calibration, 5.0 5.0 mg/L 6 * 1.5 ml Iso-C8

20001.709 Sample box NS calibration, 7.5 7.5 mg/L 6 * 1.5 ml Iso-C8

20001.710 Sample box NS calibration, 10 10 mg/L 6 * 1.5 ml Iso-C8

20001.711 Sample box NS calibration, 25 25 mg/L 6 * 1.5 ml Iso-C8

20001.712 Sample box NS calibration, 50 50 mg/L 6 * 1.5 ml Iso-C8

20001.713 Sample box NS calibration, 75 75 mg/L 6 * 1.5 ml Iso-C8

20001.714 Sample box NS calibration, 100 100 mg/L 6 * 1.5 ml Iso-C8

20001.715 Sample box NS calibration, 250 250 mg/L 6 * 1.5 ml Iso-C8

20001.716 Sample box NS calibration, 500 500 mg/L 6 * 1.5 ml Iso-C8

20001.717 Sample box NS calibration, 750 750 mg/L 6 * 1.5 ml Iso-C8

20001.718 Sample box NS calibration, 1000 1000 mg/L 6 * 1.5 ml Iso-C8

Page 29 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


Part
Description Levels Quantity Matrix
number
Mix sample box Gasoline enhanced, BOB,
20002.076 Various 8 * 1.5 ml Various
Diesel B7, E85, Jet fuel, Heating oil
9.44 mg/kg S /
20001.661 Sample Box Diesel Low Sulfur 5 * 1.5 ml Diesel
61.7 mg/kg N
20001.662 Sample Box E85 Low Sulfur 1.72 mg/kg S 5 * 1.5 ml E85

20001.663 Sample Box Jet fuel 506.8 mg/kg S 5 * 1.5 ml Jet


4.42 mg/kg S/
20001.664 Sample Box Gasoline enhanced 5 * 1.5 ml Gasoline
1.35 mg/kg N
1.35 mg/kg S/
20001.665 Sample Box Gasoline BOB 5 * 1.5 ml Gasoline
1.29 mg/kg N
9.64 mg/kg S/
20001.666 Sample Box Diesel B7 5 * 1.5 ml Diesel
39.9 mg/kg N
20001.667 Sample Box Jet fuel 167.8 mg/kg S 5 * 1.5ml Jet
29.4 mg/kg S /
20001.668 Sample Box Heating oil 5 * 1.5ml Gasoil
55.6 mg/kg N

Page 30 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


Calibration Gasses

▪ Calibration for ASTM methods D6667, D7551 and UOP 936


▪ “Certificate of Analysis” supplied with each level
▪ Blank matrix available

- Nitrogen Source: Acetonitrile


- Sulfur Source: Dimethyl sulfide

Part
Description Levels Quantity Matrix
number

Calibration gas standards


Pressure regulator, DIN-5, inert for gas
99.10.014 - 1 -
calibration standards

Calibration Gas, 0 mg/kg Sulfur & Nitrogen in


99.10.050 Blank 10 L Ethane
Ethane matrix, 10l, DIN-5, without regulator

Calibration Gas, 5 mg/kg Sulfur & Nitrogen in


99.10.055 5 mg/kg 10 L Ethane
Ethane matrix, 10l, DIN-5, without regulator

Calibration Gas, 10 mg/kg Sulfur & Nitrogen in


99.10.051 10 mg/kg 10 L Ethane
Ethane matrix, 10l, DIN-5, without regulator

Calibration Gas, 30 mg/kg Sulfur & Nitrogen in


99.10.052 30 mg/kg 10 L Ethane
Ethane matrix, 10l, DIN-5, without regulator

Calibration Gas, 100 mg/kg Sulfur & Nitrogen in


99.10.053 100 mg/kg 10 L Ethane
Ethane matrix, 10l, DIN-5, without regulator

Calibration Gas, 200 mg/kg Sulfur & Nitrogen in


99.10.054 200 mg/kg 10 L Ethane
Ethane matrix, 10l, DIN-5, without regulator

Calibration LPG standards

99.12.002 Reduce fitting, DIN-1 (1/4 NPT) - 1 -

Calibration Liquid, 0 mg/kg Sulfur & Nitrogen in


99.10.060 Blank 10 L Butane
Butane matrix, 10l, DIN-1, without regulator

Calibration Liquid, 5 mg/kg Sulfur & Nitrogen in


99.10.065 5 mg/kg 10 L Butane
Butane matrix, 10l, DIN-1, without regulator

Calibration Liquid, 10 mg/kg Sulfur & Nitrogen in


99.10.061 10 mg/kg 10 L Butane
Butane matrix, 10l, DIN-1, without regulator

Calibration Liquid, 30 mg/kg Sulfur & Nitrogen in


99.10.062 30 mg/kg 10 L Butane
Butane matrix, 10l, DIN-1, without regulator

Calibration Liquid, 100 mg/kg Sulfur & Nitrogen


99.10.063 100 mg/kg 10 L Butane
in Butane matrix, 10l, DIN-1, without regulator

Calibration Liquid, 200 mg/kg Sulfur & Nitrogen


99.10.064 200 mg/kg 10 L Butane
in Butane matrix, 10l, DIN-1, without regulator

Page 31 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


10. Technical Requirements

Gas supply & connections

Carrier gas: Helium: > 200 mL/min

Argon or Helium > 5 bar (75 psig), grade 5.0 or better (99.999%)

Argon: > 200 mL/min

> 5 bar (75 psig), grade 5.0 or better (99.999%)

* Helium may be substituted for Argon as a carrier gas (carrier gas adjustment
is required)

Connections 1/8” Stainless Steel

Oxygen > 600 ml/min, >5 bar (75 psig), grade 5.0 or better (99.999%)

Connection 1/8” Stainless Steel

Gas Supply All gasses must be dry and free of particulate matter.

Filters, gas scrubbers, and dryers are available from PAC and are
recommended in all cases:

Moisture : 5 ppm maximum

CO2 : 400 ppm maximum

THC : 10 ppm maximum

All supply gasses should be delivered through clean, dry chromatographic


grade tubing.

All national pipe thread (NPT) connections or pipe threaded connections


should be assembled carefully with Teflon® tape to guard against
contamination and leaks.

Each supply gas cylinder requires a regulator to reduce the cylinder pressure
to the desired working pressure. PAC can supply the recommended regulators.
Care should be taken to ensure use of the proper regulator for each gas in use.

Page 32 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


Power requirements

Voltage range*/** 100 – 240 Vac, 50/60 Hz, 1550 VA

* Max fluctuation 10% on nominal voltage

** Only use the main supply cord supplied with the system

One electrical outlet required for basic system

Four additional outlets required for sample introduction & pc

NEMA-5R female receptacle REQUIRED for 115VAC

Power consumption Main Voltage 115VAC 230VAC

Max. Consumption 12.2A 5.9A

Main switch location At the back of the right-hand side of the ElemeNtS.
with circuit breaker

Physical specifications

System Recommended bench space

Weight: 41 kg (90 lb.) with CTC >53 kg (117 lb.)

Width: 52.5 cm (1.7 feet) 75 cm (2.5 feet) (excluding PC)

with Accura: 75 cm (2.5 feet) (when positioned under


749)

Depth: 55.5 cm (1.8 feet) 60 cm (2 feet)

Height: 51.5 cm (1.7 feet) with CTC 749: 125 cm (4.1 feet)

with SSA: 80 cm (2.6 feet)

Remarks ▪ Adequate air space should be provided around the instrument.

▪ Place the ElemeNtS in such a way that the rear side is always
accessible in case the unit needs to be disconnected.

▪ Lifting the instrument requires 2 people

Page 33 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


Environmental conditions

Altitude Up to 3000 m (9842 feet)

Environment Indoor use

Maximum relative 80 % for temperatures up to 31°C (88°F) decreasing linearly to 50 % relative


humidity humidity at 40°C (104°F), recommended 50 to 60%, non-condensing.

Operating 15 to 35°C (59-95°F)


Temperatures
For storage or shipping, the allowable temperature range is -40 to 70°C (-40-
158°F) and the allowable humidity range is 5-95%, non-condensing. After
exposing the GC to extremes of temperature or humidity, allow 2 hours for it
to return to the recommended ranges.

Power Connection Grounded outlet within 1.5 meter of the instrument with a grounded power
cable

Ventilation The ElemeNtS is cooled by an air flow which leaves the unit at the back side.
Requirements Do not obstruct or block the air flow.

Heat Dissipation The maximum additional heat dissipation from the ElemeNtS is 7681 BTU /
hour or 8,103,455 joules /hour. This measurement represents the heat given
off when all heated zones are set for maximum temperatures.

Page 34 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


Compliance & certifications

• Low Voltage Directive 2006/95/EC

• EN 61010-1: 2010 “Safety requirements for electrical equipment for


measurement, control and laboratory use”

• The ElemeNtS has been designed and tested according to 61010-01

(Safety requirements for electrical equipment for measurement,

control and laboratory use), tested by TUV Certification B.V., Leek,

the Netherlands

• Electrical safety tests in accordance with EN61010-1:2010 (Report

form J with the US and Canadian national deviations)

• Safety tests in accordance with EN61010-2-081:2015 (Report form F

version with the US and Canadian national deviations)


• Safety tests in accordance with EN61010-2-010:2014 (Safety

requirements for electrical equipment for measurement, control and

laboratory use)

• cTUVus certification, safety testing applicable to US and Canadian

standards

• EMC, EN55011:2016 (Industrial, scientific and medical equipment)

• FCC: 47 CFR Part 15 RF Devices

• Canada ICES-003 Issue 6 (Validation only in combination with FCC

Part 15B testing

• EMC test in accordance with IEC 61326-2014

Page 35 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com


Computer Requirements

▪ Minimum Intel Core i3 processor or equivalent (Core i5


recommended)
▪ 4 GB memory
▪ 20 GB free disk space (80 GB recommended)

▪ Windows 7 x64 (64 bit) with SP1 Professional, Business, or


Ultimate
▪ Windows 7 x86 (32 bit) with SP1 Professional, Business, or
Ultimate
▪ Windows 8.1 x64 (64 bit) Professional, Business, or Ultimate
▪ Windows 10 x64 (64 bit) Professional or Enterprise
▪ Windows 2012 Server x64 (64 bit)

▪ 1920 x 1080 or higher resolution and 256 Color Palette

▪ TCP/IP networking 100/1000 Mbps

▪ Yes, USB 2.0

An Ethernet switch is integrated in the ElemeNtS for easy connection to


the 749 autosampler & PC.

• Left: Connection of ELEMENTS/749 straight to PC


• Right: Connection of Elements/749 sampler trough lab network
infrastructure

Page 36 of 36 ElemeNtS Technical Offer 2020.5 www.paclp.com