05 Ultrasonic Test Procedure 5004

You might also like

Download as doc, pdf, or txt
Download as doc, pdf, or txt
You are on page 1of 11

DOC.

NO: EK 09
ULTRASONIC TEST PART.NO: 05
REV NO: 00
PROCEDURE

ULTRASONIC TEST PROCEDURE

PREPARED BY CHECKED BY APPROVED BY

TECHNICAL APPLICATION TECHNICAL APPLICATION GENERAL MANAGER


MANAGER MANAGER

FIRST ISSUE : 01.06.2007 ISSUED : 01.06.2007 Page 1 / 11


DOC.NO: EK 09
ULTRASONIC TEST PART.NO: 05
REV NO: 00
PROCEDURE

CONTENS

1. SCOPE

2. APPLICATION

3. APPLICABLE DOCUMENTS

4. RESPONSIBILITIES

5. TEST APPARATUS

6. CALIBRATION

7. EXAMINATION COVERAGE AND SCANNING


REQUIREMENTS

8. TEST

9. SIZING AND MEASURING METHODS

10. EVALUATIONS

11. ACCEPTANCE CRITERIA

12. NONCONFORMANCE

13. REPORTS / RECORDS

FIRST ISSUE : 01.06.2007 ISSUED : 01.06.2007 Page 2 / 11


DOC.NO: EK 09
ULTRASONIC TEST PART.NO: 05
REV NO: 00
PROCEDURE

1 SCOPE

1.1 This procedure defines the procedures and techniques of the ultrasonic test to detect cracks and other
discontinuities, in weldments when allowed within the provisions of
EN 583-1/2/3
EN 12223- EN 27963
EN 1714 EN 1713 EN 1712
ASME I PW-11.2 and
ASME B31.1 Table 136.4 (Note 3).

1.2 This procedure shall be used in conjunction with the applicable technical and construction specifications,
drawings, codes and standards.

2 APPLICATIONS

2.1 This procedure shall be applied to all ultrasonic test of the weldments fabricated for the subject project.

3 APPLICABLE DOCUMENTS

3.1 ASME V
3.2 ASME Sec I/ASME B31.1
3.3 EN 583-1/2/3
3.4 EN 12223- EN 27963
3.5 EN 1714 EN 1713 EN 1712

4 RESPONSIBILITIES

4.1 The N.D.E Level I shall conduct the test under the supervision of certified level - II (SNT- TC-1A) or EN 473
and he shall prepare the Ultrasonic Test Report/Record in accordance with this procedure, drawing and
specifications.

4.2 The N.D.E. Level II shall interpret or evaluate the test results.

5 TEST APPARATUS

5.1 Instruments

5.1.1 The ultrasonic instrument shall be of the pulse - echo type suitable for use with
transducers oscillating at frequencies between 1 to 6 MHz. The display will be "A"-
scan presentation.

5.1.2 The horizontal linearity shall be qualified over the full sound path distance to be used in testing.

5.1.3 The test instrument shall include internal stabilization so that after warm up, no variation in response greater
than 1 dB occurs with supply voltage change of 15 % nominal or in the case of a battery, the charge operating life.

5.1.4 The test instrument shall have a calibrated gain control adjustable in discrete 1or 2 dB steps over a range
of at least 60 dB. The accuracy of the calibrated gain control settings shall be within 1 dB.

FIRST ISSUE : 01.06.2007 ISSUED : 01.06.2007 Page 3 / 11


DOC.NO: EK 09
ULTRASONIC TEST PART.NO: 05
REV NO: 00
PROCEDURE

5.2 Search Units

5.2.1 Straight beam search units

1) The straight beam search unit transducers shall have an active area of not less than 9.6 mm dia., or 10x10
square, nor more than 25 mm dia., or 25x25 square.

2) Transducer frequency used shall be between 1 and 5 Mhz.

5.2.2 Angle Beam Search Units

1 ) The angle beam search units shall consist of a transducer and an angled wedge.

2 ) The transducer crystal size shall be 8x9 mm to 25x25 mm square or rectangular in shape.

3 ) The transducer frequency used shall be between 1 and 5 MHz.

4 ) Angle beam search units shall produce a sound beam in the material being tested within 2 degrees of the
stated angle, i.e.70 degree, 60 degree, or 45 degree

5.2.3 The search units both straight and angled beam shall be capable of resolving the three reflections on the V1
resolution block

5.2.4 Different size or frequency of search unit may be used for better results

5.3 Reference Blocks

5.3.1 Standard Block

1 ) The International Institute of Welding (IIW) ultrasonic standard block or V1 block shall be the standard
used for distance calibration, sensitivity calibration,check to exit point and actual fracted angle of angled beam
search units.

2 ) The miniature block (V2 block) shall be used for distance calibration instead of the IIW/V1 block in the
field.

3 ) Other suitable blocks can be used for the purposes of 5.3.1 (1) and (2)

5.3.2 Calibration Block

The calibration block(s), shall be used to establish sensitivity and distance amplitude correction (DAC) curve.

5.4 Couplants

5.4.1 A couplant material shall be used between the search unit and the test material, such as light oil, grease,
glycerin or glue mixed with water to a suitable consistency.

6 CALIBRATION

FIRST ISSUE : 01.06.2007 ISSUED : 01.06.2007 Page 4 / 11


DOC.NO: EK 09
ULTRASONIC TEST PART.NO: 05
REV NO: 00
PROCEDURE

6.1 Equipment Calibration

6.1.1 The horizontal linearity of the test instrument shall be calibrated after each 40 hours of instrument use. (EN
12668-1/2 )

6.1.2 The instrument gain control shall meet the requirements of 5.1.4 and shall be checked for correct calibration
at two month intervals.

6.1.3 Each angle beam search unit shall be checked after each eight hours of use to determine exit point and actual
refracted angle.

6.2 Calibration for Test

6.2.1 All calibrations and Tests shall be made with the reject (or suppression) control turn off.

6.2.2 Calibration for horizontal sweep, DAC construction and establishment of the sensitivity shall be made by the
ultrasonic operator just prior to and at the location of the testing of each location.

6.2.3 Horizontal Sweep Range Calibration

1 ) Straight Beam Calibration Position the search unit for the first back reflection from the IIW/Miniature
block or base material, and adjust the left edge of this indication to the desired position on the screen with the
delay control. Position the search unit for the second or latest back reflection from the above blocks or base
material, and adjust the left edge of this indication to desired position on the screen with the length control. Repeat
delay and length control adjustments until the first and second/latest back reflections from the IIW/Miniature block
or base material, and fix the correct position the horizontal sweep shall be adjusted for distance calibration to
present the equivalent of at least two plate thicknesses on the screen.

2 ) Angle Beam Calibration


Position the search unit for the first maximum reflection from the IIW/Miniature block, and adjust the left hand
edge of this indication to the desired position on the screen with the delay control. Repeat delay and length control
adjustments until the first second/latest back reflections from the IIW/Miniature block or base material, and fix the
correct position.

6.2.4 Construction of Distance Amplitude Correction (DAC) Curve for Angle Beam Testing.

The calibration block which will be use for DAC curve is determined acc. to EN 583-2 or ASME V Article-5, T-
542.2. and figure T-542.2.1

1 ) Position the search unit for maximum response from the 1/8, 2/8 and 3/8 vee path which gives the highest
amplitude.

2 ) Adjust the sensitivity control to provide 80 % of full screen height, and mark the peak on the screen.

3 ) Decrease 6dB of the above adjusted sensitivity control and mark the peak on the screen.

4 ) Position the search unit for maximum response from another vee path indication, and mark the peak on
the screen.
5 )Repeat the above step 3).

6 )Repeat steps 4) and 5) until 9/8 vee path is achieved.

FIRST ISSUE : 01.06.2007 ISSUED : 01.06.2007 Page 5 / 11


DOC.NO: EK 09
ULTRASONIC TEST PART.NO: 05
REV NO: 00
PROCEDURE

7 )Connect the screen marks to provide the DAC curve.

6.3 Establishing Sensitivity

6.3.1 Straight Beam

The sensitivity shall be adjusted at a location free of indication so that the first back reflection from the base
material shall be 50 to 75 % of full screen height.

6.3.2 Angle Beam

The sensitivity shall be adjusted at a location of primary reference sensitivity level (80 % of FSH) of the DAC
curve.

6.4 Re calibration

6.4.1 Re calibration shall be made after a change of operators at 30 minute maximum time intervals or when the
electrical circuitry is disturbed in any way including the following:

1 ) Transducer change

2 ) Battery change

3 ) Electrical outlet change

4 ) Coaxial cable change

5 ) Power cut (failure)

6.4.2 Re calibration procedures are the same as para, 6.2 and 6.3.

7 EXAMINATION COVERAGE AND SCANNING REQUIREMENTS

7.1The examination coverage shall be as required by EN 1714 or ASME V art. T-424.1.

7.2The volume of weld to be examined is the weld volume plus 0.5” (12.7 mm) on either side of the welds,
regarded as heat affected zone.

7.3The base metal on each side of the weldseam shall be scanned with a straight-beam search unit as required by
ASME V para. T-441.3.2.2.b.

7.4 The volume of weld to be examined shall be scanned for reflectors oriented parallel to the weld, with angle
beam probes, so that each point of the complete volume of weld to be examined shall be reached by two angle
beams, with a difference in angle of more than 10º. The surface breaking zone on the inside of vessel, shall in
addition be examined with another two angle beam probes, or in total with four different angles.

7.5 The volume of weld to be examined shall be scanned for reflectors oriented transverse to the weld, with angle
beam probes, oriented at ca.15° to the weld axes, from two opposite directions from either side of the
weldseam. The same series of search units used for the different scans for defects oriented parallel to the weld
shall at least be used.

FIRST ISSUE : 01.06.2007 ISSUED : 01.06.2007 Page 6 / 11


DOC.NO: EK 09
ULTRASONIC TEST PART.NO: 05
REV NO: 00
PROCEDURE

The surface breaking zone shall also be examined with a beam orientation of 0º to the weld axes.

7.6 Each scan of weld-volume and heat affected zone shall be divided into a top-half-scan (T), a bottom-half-scan
(B), a penetration (P) or root scan(R) and an inner surface breaking-scan (S).

8 TEST

8.1 Surface Preparation

8.1.1 All surfaces to which a search unit is applied shall be free of weld spatter, dirt, grease, oil (except couplant),
paint and loose scale and shall have a contour permitting adequate coupling.

8.1.2 The outside weld crown can be left on. 100 mm or the widht required for scanning adjacent to the weld, on
the shell and/or head, shall be smooth, with an andulation smaller than 0.3 mm over 25 mm. A roughness of 250
inch CLA or sand paper grain size of 80 is sufficent. The as painted conditions will normally satisfy these
requirements.

8.2 Identification Of The Scanning Surfaces

8.2.1 Scanning surfaces are the surfaces beside the weld from which the examination is performed.

8.2.2 The surface above the weld, when the direction is seen from left to right, is called face A, the other one side
face B.

8.2.3 This specification only applies to welds examined from the external surface of the vessel.

8.3 X-, Y-, Z-, Coordinates

8.3.1 The X- co-ordinate starts at the zero-point.

8.3.2 The Z- co-ordinates zero point is the inside surface of the vessel, with the positive direction into the weld.

8.3.3 The Y- co-ordinate zero point is :


The center of the weld for butt welds, with the positive direction towards scanning face A on which the zero point
is die stamped.
The plane of the nozzle’s external surface for the corner welds, with the positive direction into the weld.

8.4 Test Technique

8.4.1 Straight Beam

The entire base metal through which ultrasound will travel to test the weld shall be tested for laminar reflectors
using a straight beam search unit conforming to the requirements of 5.2.1 and calibrated in accordance with 6.2.3
and 6.3.1.

8.4.2 Angle Beam

Welds shall be tested using an angle beam search unit conforming to the requirements of 5.2.2 with the
calibration in accordance with 6.2.3, 6.2.4 and 6.3.2

8.5 Scanning

FIRST ISSUE : 01.06.2007 ISSUED : 01.06.2007 Page 7 / 11


DOC.NO: EK 09
ULTRASONIC TEST PART.NO: 05
REV NO: 00
PROCEDURE

8.5.1 Scanning Sensitivity

1) Scanning shall be performed at a gain setting of at least two times (6dB) the reference level.

2 ) During scanning only the gain controls may be adjusted. Any adjustments to the other controls shall
require re calibration.

Scanning Pattern

Scanning shall be performed as follows:

1) Longitudinal discontinuities
Movement A : Rotating angle A = 10 deg.
Movement B : Scanning distance B shall be such that the section of the weld being tested will be
covered.
Movement C : Progression distance C shall be approximately 1/2 the width of the transducer being used.
Movements A,B and C are combined in one scanning pattern.

2 ) Transverse discontinuities
Scanning pattern D :(When welds are ground flush.)
Scanning pattern E :(When welds are not ground flush.)
Scanning angle E = 15 deg. maximum.
The scanning pattern is to be such that the full weld section is covered.

Figure 1 : Plan view of UT scanning patterns


9 SIZING AND MEASURING METHODS

9.1 Any one of the known sizing methods can be used within their typical restriction;

9.1.1 20dBPB-(plotted beam) or 10dBPB-drop systems can be used for sizing in the VERTICAL PLANE and
LENGHT DETERMINATION for ALL INDICATIONS.

FIRST ISSUE : 01.06.2007 ISSUED : 01.06.2007 Page 8 / 11


DOC.NO: EK 09
ULTRASONIC TEST PART.NO: 05
REV NO: 00
PROCEDURE

9.1.2 DGS/AVG: shall only be used for VERY SMALL PORE-LIKE INDICATIONS.

9.1.3 MA: maximum amplitude sizing can be used for LENGHT DETERMINATION of LONGER
INDICATIONS.

9.1.4 6dBCB: 6dB-drop with the central beam (a good system for large lamination sizing) is NOT ACCEPTED for
sizing in this specification seen the small dimensions of the possible indications.

9.1.5 6dBPB: 6dB-drop with a plotted beam can be used for sizing in the VERTICAL AND HORIZONTAL
PLANE for SHORT INDICATIONS.

9.2 Attenuation Measurements

For each scan, attenuation shall be measured and compared with the one of the calibration block.
Attenuation shall be measured with two equal probes in the T-R mode, comparing the signals at skip distance
with DAC – curves.

9.3 Surface Transfer Measurements

The difference between the calibration block surface and the vessel surface shall be compensated by the surface
transfer losses.
This measurement shall be done at skip distance on the calibration block and the surface under test, at the same
moment as the attenuation measurement for each scan.

10 EVALUATION

10.1 Recording Criteria

10.1.1 Straight Beam

The following indications shall be recorded.

1 ) Indications that are equal to or greater than the original back reflection.

2 ) If any area of base metal exhibits total loss of back reflection.

10.1.2 Angle Beam

The indications greater than 20 % of the primary reference sensitivity level shall be investigated to the
extent.

10.2 Acceptance Criteria: EN 1712 / ASME Sec I (PW-52)/ASME B31.1 Chapter VI clause 136.4.6

11 ACCEPTANCE CRITERIA

11.1 As Per ASME Section I - PW-52

Acceptance standards for ultrasonic examination :

(a)Technique and standards for ultrasonic examination shall follow Section V, Article 5.

FIRST ISSUE : 01.06.2007 ISSUED : 01.06.2007 Page 9 / 11


DOC.NO: EK 09
ULTRASONIC TEST PART.NO: 05
REV NO: 00
PROCEDURE

(b)The Manufacturer’s report, as required by T-590 of Section V, shall be retained by the Manufacturer for a
minimum of 5 years.

(c)Acceptance-rejection standards. imperfections that cause an indication greater than 20% of the reference level
shall be investigated to the extent that the ultrasonic examination personnel can determine their shape, and
location, and evaluate them in terms of (a) and (b).

(d)Cracks, lack of fusion, or incomplete penetration are unacceptable regardless of length.

(e)Other imperfections are unacceptable if the indication exceeds the reference level and their length exceeds the
following:

(1) ¼ in. for t up to ¾ in.


(2) 1/3 t for t from ¾ in. to 2 ¼ in.
(3) ¾ in. for t over 2 ¼ in.

where t is the thickness of the weld being examined. If the weld joins two members having different thicknesses
at the weld, t is the thickness.

11.2 As Per ASME B31.1 Paragraph 136.4.6

When required by this Chapter, ultrasonic examination shall be performed in


accordance with Article 5 of Section V of the ASME boiler and Pressure Vessel Code.

(a)Acceptance Standards. Welds that are shown by ultrasonic examination to have discontinuities which produce an
indication greater than 20% of the reference level shall be investigated to the extent that ultrasonic examination
personnel can determine their shape, identity, and location so that they may evaluate each discontinuity for
acceptance in accordance with (1) and (2) below.

(1)Discontinuity evaluated as being cracks, lack of fusion, or incomplete penetration are unacceptable regardless of
length.

(2)Other discontinuities are unacceptable if the indication exceeds the reference level and their length exceeds the
following;

(a)¼ in. (6.0 mm) for t up to ¾ in. (19.0 mm);


(b)1/3 t for t from ¾ in. (19.0 mm) to 2 ¼ in. (57.0 mm)
(c)¾ in. (19.0 mm) for t over 2 ¼ in. (57.0 mm) where t is the thickness of the weld being examined. If the weld
joins two members having different thicknesses at the weld, is the thinner of these two thicknesses.

12 NON CONFORMANCE

12.1 Nonconformances identified by the ultrasonic test shall be reported in accordance with the project
documentation and this procedure.

12.2 Nacceptable defects shall be repaired, and re examined by the same ultrasonic technique and procedure used
originally.

13 REPORTS / RECORDS

FIRST ISSUE : 01.06.2007 ISSUED : 01.06.2007 Page 10 / 11


DOC.NO: EK 09
ULTRASONIC TEST PART.NO: 05
REV NO: 00
PROCEDURE

All indications shall be recorded to the ultrasonic test report/record by the examiner who carried out the Test.

FIRST ISSUE : 01.06.2007 ISSUED : 01.06.2007 Page 11 / 11

You might also like