Catalog
1. Installation ......................................................................................... 2
5. MP-900 Configure tools setting. (Note: please make sure the server is
7. Enter Center Station [iCIS], click below icon on the desktop. ............. 11
1
1. Installation
1.1 Software
3) Click [Install]
2
4) Click [Finish]
1.2.2 Install Server, Configure tool and CenterStation, the steps are
3
2) Click the [Generate], then it would generate computer code, you need to
copy this computer code,
code then send it to Medcaptain service engineer.
3) Medcaptain service engineer would send you the Auth code according your
computer code, you just copy the Auth
Auth code , then paste it to below menu,
click the [Register].
4
5) Enter MP-900 Server menu after select a valid IP.
2 Connect a wifi
5
4
5
6 7
6
2) Select the valid IP which is same with host IP.
3) Click [Set Department], Enter department name (for example: ICU), click [Add]
7
4) Click [Set Room]-[ICU], enter [Start Bed] number (for example :1) and [Bed number] (for
example : 10 beds). So there would be ten beds in the ICU rooms.
5) About [Set Net], do not change the it, just use the default IP: 127.0.0.1 and default server
port: 9900.
7) Modify password.
8
6. Pumps’ WLAM setting (UI version V1.3.X).
1) Press [MENU] button, click [InterConnect set];
2) Click [Info Channel] in [more] bar, choose right WLAN channel. If the pumps directly connect
the server, please select [Local WLAN] or [Local RS485]. If the pumps which installed in the
MP-80, and MP-80 connecting the server, please select [Station WLAN] or [Station RS485].
9
(B) For Station WLAN, Press [HOME], Select [InterConnect Set], Click [Station WLAN] in
[more] bar.
4) Enter WLAN information as below, include wifi address, IP address and server IP.
10
You can enter a new department, but it should be same with configure tool’s department
name. Also Bed No. should be setting.
11
You can all pumps’ which are connecting with MP-900 status. For more information about how to
use Center Station, please review MP-900 center station introduction file.