Calibration of Thermal Process Equipments
Calibration of Thermal Process Equipments
1.0 OBJECTIVE:
2.0 SCOPE:
4.0RESPONSIBILITY:
Calibration-engineer will perform the Calibration and note down the details and
measurements on the observation sheet.
Technical/Quality Manager or his designee will check the observation sheet and approve
the Calibration Certificate.
(a) Equipment should be in good physical condition and should not have any physical
damage.
(b) Electrical connection and wiring should be proper and should not have any risk of
electrical shock.
8.0 PROCEDURE:
8.1 Note Down the details of Equipment under calibration (UUT) in Observation Sheet.
8.2 Check out Freezer/ Oven/ Incubator/ Autoclave for its connection and visually
inspection.
8.3 For single position place the temperature sensor in center of the unloaded chamber.
8.4 Set the temperature of UUT (Unit under Calibration) for desired temperature.
8.5 Wait for the stabilization of temperature and hold it for 20-30 min.
8.6 Record 10 readings at 1 min. interval for single position at center.
8.7 Change the position of sensor without disturbing set point and record 10 more readings
at 1 min. interval after temperature stabilization.
8.8 Calculate the stability and uniformity of UUT as following :
UNIFORMITY (±˚C) = Avg. Temp. of center point – Max. ( avg. of other location from
center)
STABILITY (±˚C) = Max. Temp. of sensor (Centre) - Min. Temp. of sensor during
survey
8.9 Calculate the uncertainty of measurement by taking 10 repeated readings from the
recorded measurement (center location). Uniformity and Stability should be taken
during uncertainty calculation.
8.10 Affix the calibration sticker on the Equipment.
8.11 Prepare the Calibration certificate.
9.0 PRECAUTIONS:
(a) Always observe manufacturer safety precaution before starting calibration.
10.0 REFERENCE:
(a) DKDR5-7
(b) EURAMET cg-13
11.0 ABBREVIATIONS
CP: Calibration Procedure.
STD: Standard
UUT: Unit under Test