Professional Documents
Culture Documents
Uniformity & Stability Calculation Procedure
Uniformity & Stability Calculation Procedure
THERMAL SOURCES
1.0 OBJECTIVE:
To estimate the Stability and Uniformity of Dry Block /Low Temperature Bath &
Temperature/Humidity control cabinet etc.
2.0 SCOPE:
This procedure is applicable to covers the process of estimating stability & uniformity
for Dry Block/Low Temperature Bath & Temperature/Humidity control cabinet etc.
4.0 RESPONSIBILITY:
Calibration Engineer will perform the Stability and Uniformity of Dry Block/Low
Temperature Bath & Temperature/Humidity control cabinet etc. & note down the
data for the same.
Technical/Quality Manager will check the data and approve the Uniformity /Stability
records.
6.0 PROCEDURE:
6.1 Clean the equipment and check the equipment under test for any physical damage.
The equipment must be free from any visual defect.
6.2 STABILITY:
Now interchange the STD. and observe the temperature in the same manner with 10
sets of readings. This will be noted as
Let A2 is the average temperature of hole A with Temperature Sensor
Let B2 is the average temperature of hole B with Temperature Sensor
(A1-B1)- (B2-A2)
Now the Radial Uniformity is calculated as Uniformity = ------------------------
2
In the same manner calculate the uniformity at different temperatures as per
requirement.
(Only applicable if using 2 Nos. of sensors and option is available for two sensors)
NOTE :
For estimating the radial uniformity at high temperatures, do not remove the
sensors at high temperature while interchanging the sensors to avoid thermal shock
to the sensors. Instead repeat the experiment next day after interchanging the probe
when the bath/dry block with PRTs/TCs are at ambient or room temperature.
6.5 To calculate axial uniformity put two STD. in two different holes with same immersion
and take the deviation 1 after that move any one STD. by 2 cm take the reading and
note deviation 2 the difference between both deviations 1& 2 is called axial
uniformity.
In the same manner stability and uniformity of the humidity chamber at two
positions
can be defined by placing two different RH sensors/Indicators in the humidity
chamber.
7.0 PRECAUTIONS:
1. Make sure at high temperatures do not pull out sensor From Thermal Source.
2. Thermal Source should be cooled with power on position.
8.0 Record observations and analysis the data for stability and uniformity of sources as
given below equation :-
CALCULATION FOR RADIAL UNIFORMITY:
Ru=Avg.(T1-T2)
Where Ru=Redial Uniformity
Avg.=Average of Ten measured readings
T1= Temperature of Sensor Location 1
T2=Temperature of Sensor Location 2