OQ

OQ Checklist

for the ELx808


Absorbance Microplate Reader
All Models
OQ Checklist for ELx808, All Models

Model:      
Serial Number:      
Date:      
Location:      

# Step Result or Information


OQ-1 Verify/Record Software
Information
Record information:
Basecode Part Number/Version:      
Basecode Checksum:      
Assay Software Part
Number/Version:      
Correct/expected version of YES NO
basecode software installed? N/A      
Correct/expected version of assay YES NO
software installed? N/A      
OQ-2 Verify the Filter Table
Reader’s software filter table YES NO
matches the contents of the filter N/A      
wheel?
System Test was run if filter table YES NO
was updated? N/A      

Record information:
1       nm or plug
2       nm or plug
3       nm or plug
4       nm or plug
5       nm or plug
6       nm or plug

OQ-3 System Test


System Test pass? PASS FAIL
N/A      
Incubator test pass? PASS FAIL
N/A      
Incubator Zones 1-4 pass? PASS FAIL
N/A      
Results report attached? YES NO
N/A      
If NO, record location:      

ELx808 Qualification and Maintenance Procedures OQ Checklist


© 2011, BioTek Instruments, Inc. Page 1
OQ Checklist for ELx808, All Models

# Step Result or Information


OQ-4 Empty Carrier Test PASS FAIL
Empty Carrier Test pass? N/A      
OQ-5 Shake Test PASS FAIL
Shake Test pass? N/A      
Is the reader equipped with a barcode YES NO
scanner? If NO, skip OQ-6. N/A      
OQ-6 Barcode Scanner Test PASS FAIL
Barcode Scanner Test pass? N/A      
OQ-7 Absorbance Plate Test
Absorbance Plate Test pass? PASS FAIL
N/A      
Test Plate Serial Number:      
Results report attached? YES NO
N/A      
If NO, record location:      
OQ-8 Absorbance Liquid Test 1
Absorbance Liquid Test 1 pass? PASS FAIL
N/A      
Spreadsheet attached? YES NO
N/A      
If NO, record location:      
OQ-9 Absorbance Liquid Test 3
(optional, UV models only)
Absorbance Liquid Test 3 pass? PASS FAIL
N/A      
Spreadsheet attached? YES NO
N/A      
If NO, record location:      
OQ-10 Run Final System Test
(if required)
System Test pass? PASS FAIL
N/A      
Incubator test pass? PASS FAIL
N/A      
Incubator Zones 1-4 pass? PASS FAIL
N/A      
Results report attached? YES NO
N/A      
If NO, record location:      

ELx808 Qualification and Maintenance Procedures OQ Checklist


© 2011, BioTek Instruments, Inc. Page 2
OQ Checklist for ELx808, All Models

Comments:
     

Operational Qualification conducted by: Date:

PRINT Name:

If required, reviewed/approved by: N/A Date:


     

PRINT Name:      


.

ELx808 Qualification and Maintenance Procedures OQ Checklist


© 2011, BioTek Instruments, Inc. Page 3