OUT OF TOLERANCE REPORT

INSPECTION, MEASURING AND TEST EQUIPMENT

Equipment: RH/Temp. Logger Date: 10/27/20

Manufacturer: MadgeTech Model #: MicroRHTemp/IFC200

Serial #: N45248 EMC ID #: 30405

Cal Lab: Graftel Cal Date: 10/21/20 Previous Cal Date: 10/2/19

Nominal Accuracy: ±3% RH, ±0.9°F Measured Accuracy: See below


Details Of Out Of Tolerance Condition: At Nominal 20% RH, meter indicated 13.9% RH or -3.1%
RH lower than the lower limit (17% RH).

Previous History of Out of Tolerance Conditions?: 2019 – indicated 4.3% higher than higher limit

Impact On Product: None. This logger is used for monitoring ambient temperature for Service Level 1
(SL1) coating storage. Monitoring humidity is not required since all paint containers
are still factory sealed.
Note: This equipment is used for monitoring safety-related coating storage conditions.

Corrective Action Required: None.

Is Customer Notification Required? No Yes


Impact on safety-related equipment repairs–10CFR21 Notification N/A* No Yes
required?
* N/A is applicable specifically where equipment is for non-safety related use, e.g. Courtland & EPD shops.

10/27/20
Quality Assurance Manager Date Engineering Review By Date