6. Calibrations
Contents
Calibrations
Technical Facilities ..................................... 6:2
Before calibration ........................................ 6:2
A/D Converters
A/D Converters ........................................... 6:3
Transducers
Calibration of linear transducers ............... 6:4
Selection diagram .................................. 6:4
Venous Pressure Transducer Calibr ...... 6:5
Arterial Pressure Transducer Calibr .... 6:6
Priming Detector calibration ................. 6:7
Occlusion adjustment, blood pump ....... 6:7
Blood Leak Detector calibration ............ 6:8
Conductivity Transducer calibration .... 6:9
Dialysis Pressure Transd Calibr ......... 6:10
Degassing Pressure Transd Calibr ...... 6:11
pH transducer calibration ................... 6:12
Pressure Regulator Calibr, PR 1 ......... 6:14
Pressure Regulator Calibr, PR 2 ......... 6:14
HPG Pressure Transducer Calibr ........ 6:15
Variable flow calibration ...................... 6:16
Calibration AK 95 S
Warning! After completing the calibrations, the conductivity value must always be confirmed by the
hospital laboratory.
Ordering no:
UF - calibration:
UF-calibration box K21908002
AD:
Digital voltmeter
Priming detector:
A piece of blood tube used at the hospital.
Pressure
Reference instrument 300333001
Calibration kit K13983002
Special tool
For silicone tube connectors K12755001
Calibration kit
AD Converters
1. Start the AK 95 S, enter the Service menu 5. Set the “AD REF VOLTAGE” on the
and select Calibrations. display to the value shown on the voltme-
ter and press .
2. Choose AD-converter.
SELECT AD-CONVERTER
CPU X AD-REF VOLTAGE STORED
CPU-C CPU-P CPU-W MM95
TP 5 TP2
Z0VAD
CPU C A/D ref
TP 6 Buzzer
(TP1)
CPU W PROM (TP2)
TP4 R24
Z0VRW
CPU W
SafeRAM
pH plug
Transducers
Calibration of linear
transducers
1. Start the AK 95 S, enter the Service • Select Internal and the AK 95 S main-
menu and select Calibrations. tains a flow during the calibration.
Selection diagram
INSERT TUBE Use the gauge pins (stop/go) available in the kit
Press SELECT to confirm K40158001. The kit includes following three
different gauge pins:
5. Insert an empty tube in the priming
T = 1.1 (Stop pin = 1.1*2*0.7 = 1.5mm)
detector and press .
T = 1.6 (Stop pin = 1.6*2*0.7 = 2.2mm)
Use the normal tube from the blood line. T = 2.0 (Stop pin = 2.0*2*0.7 = 2.8mm)
6. The calibration is now done automatically T = wall thickness of the blood tubing
and the calibration factor is writen into
the EEPROM. Use following formula to calculate which gage
pin to use:
7. When the info. display shows “EEPROM 2 * wall thickness * 0,7 (mm)
STORAGE COMPLETED”, the calibration
is completed.
8. Press to exit.
= Measuring
START
12. Press UF
STOP
to store the calibration values.
pH transducer calibration
1. Offset calibration, with a buffer solution 8. Stir in order to achieve balance in the
(a solution with a very accurate and stable buffer solution. Do not let the pH sensor
pH-value) having a pH-value in the range lean against the bottom of the cup.
of pH 6.5-7.5, followed by: Wait for 2 minutes.
2. Slope calibration with a buffer solution 9.* Press to set the “REF SET” value on
having pH-value that differs at least 2 pH-
the AK 95 S and wait until the calibration
units from the buffer used for offset cali-
bration. display returns.
START
Calibration schedule: 10. Press UF
STOP
to load the calibration value
into the EEPROM.
An Offset calibration should be per-
formed after one month in operation and 11. Clean the pH-sensor in RO-water.
then every third month to verify the Remount the pH sensor into the
performance. pH sensorhousing.
Offset Calibration:
Complete calibration:
1. Make sure there is no flow or fluid in the 13. Stir in order to achieve balance in the
AK 95 S. The inlet water must be closed. buffer solution. The pH sensor must not
lean against the bottom of the cup.
2. Start the AK 95 S, enter the Service Wait for 2 minutes.
menu and select Calibrations.
14.* Press and wait.
3. Select Transducer, External and
pH TRANSDUCER from the selection The display shows: Entering sensitivity
diagram. calibration.
4. Press to start calibration routine. 15. Press to confirm that the calibration
shall continue.
5. Remove the pH sensor from the pH sensor
housing by loosening the nut holding the 16. When the calibration display returns,
pH sensor and rinse the sensor in distilled press
START
UF
STOP
to load the calibration values
water.
into the EEPROM.
6. Check that the “REF SET” on the AK 95 S
17. Remount the pH-sensor into the pH sensor
displays pH 7.
housing.
7. Put the pH sensor in a cup with pH 7
* If an error message is shown after this
buffer solution (ordering no. 100 810 001),
step, check the buffer solutions, level in
or whatever reference solution available,
the cup and repeat the procedure from
which must cover the glass membrane i.e.
step 1. If there still is an error after 2nd
the level in the cup should be at least
try, replace the pH-sensor.
30mm from the bottom.
NOTE! The pH-sensor should be changed every
8. Stir in order to achieve balance in the
eighteen month.
buffer solution. Do not let the pH sensor
lean against the bottom of the cup.
Wait for 2 minutes.
1. Connect a pressure gauge just before the 1. Disconnect tube no. 4 from the T-connector
Parker connector inside MM 95 according at INVA.
to this picture.
PR 1
Fluid connector
detection
Water
Inlet Pressure PR 2 INVA
regulator Flow
PR 1 °C switch
AIVA
Heat Suction
exchanger Pump
Drain
Outlet
Restrictor
UF Calibration
General
This section describes the calibration routine of The Complete calibration routine is done with
the UF-measuring unit. the mainflow 500 ml/min. It requires that the
external UF calibration equipment is connected
The UF-measuring unit consists of two flow to the AK 95 S, see fig.below. The equipment
measurement channels called channel 1 (CH 1) contains a reference volume, a flow directing
and (CH 2). CH 1 is the channel used to measure valve (UFVA) and a UF- calibration box. The
the main flow to the dialyser and CH 2 is used to UF calibration box controls the directing valve
measure the flow from the dialyzer, i.e. the sum by measuring low level and high level of the
of the main flow and the UF flow. The UF flow is reference volume and depending on the selected
the difference between CH1 and CH2 flows. UF UF box mode the valve will become closed or
flow is also called differential flow. opened when low or high level is detected.
High Level
AK
AK9590
S
V.ref.
UF
Low Level Cal.
Flow in/out Box
UFVA
Complete
3. Place B-pickup tube in the acetate-concentrate Then the display for Isolated UF Volume for 3
container. seconds:
10. Set Reference volume with the SET knob. The equipment is turned off in the
(The Referense volume is labeled on the UF- following order:
Calibration unit) 1 AK 95 S ON/OFF button.
2 AK 95 S main switch.
11. Press to confirm. 3 UF calibration box.
Intern
1. Start the AK 95 S with the main switch and When the internal UF calibration is ready:
the ON/OFF button. After the calibration, AK 95 S displays:
5. Select Transducers.
Press to save the calibrations values.
6. Select UF.
When the storage is completed, AK 95 S displays:
7. Select Intern.
EEPROM STORAGE COMPLETED
The internal UF calibration runs automatically,
until the calibration is finished.
Press HOLD to exit
Verify
Turn on the equipment on the following The information displays also shows the different
order: stages during the calibration:
• Fluid preparation
1. Plug in the UF calibration box. • UF stabilization
• Main flow offset calibration
Note! All electrical connections must be done • CH 1 flow calibration
before the power for AK 95 S is turned on. • Differential flow calibration
Be aware of that there is no main switch on the • Verifying calibration of UF
UF calibration box. • Verifying calibration of Isolated UF
Connect the UF calibration box to the When the verification of the UF cell
AK 95 S. calibration is ready:
When the verification is completed, AK95
Use the UF calibration cable, K1 4036, to displays the results:
connect the AK 95S with the UF calibration box.
The part of the cable that is labeled UF BOX First the display for UF Volume for 3 seconds:
should be connected to the UF calibration box.
UF VOL REF VOL VOL DIFF
Note! The verification will not run (but nothing 1006.5 1007.3 -0.28 %
will get destroyed) if the cable is reversed.
2. Start the AK 95 S with the main switch and Then the display for Isolated UF Volume for 3
the ON/OFF button. seconds:
3. Place B-pickup tube in the acetate-concentrate ISO VOL REF VOL VOL DIFF
container.
1007.5 1007.3 +0.28 %
4. Enter Service mode.
Then the previous display is shown again in
3 seconds and so on.
5. Connect the fluid tubes to the UF calibration
box.
When ready, press to exit the verification of
6. Select Calibration. the UF cell calibration.