Professional Documents
Culture Documents
Validation of A Portable Blood Glucose Testing Device in Measuring Masticatory Performance
Validation of A Portable Blood Glucose Testing Device in Measuring Masticatory Performance
ORIGINAL ARTICLE
Abstract
Introduction
Purpose: The purpose of this experiment was to in-
vestigate the possibility of objectively evaluating mas-
For evaluating masticatory function, especially
ticatory performance using a portable blood glucose masticatory performance, the gummy-jelly (a
testing device. kind of popularly available jelly candy) has re-
Methods: First, the glucose concentrations of four ceived much attention because the shape and in-
types of glucose solutions with known concentrations gredients are well-standardized, hygiene control
were measured using a spectrophotometer and portable is relatively easy, and it has been used as a test
blood glucose testing device. Next, 20 healthy subjects food to analyze daily masticatory perfor-
were asked to chew gummy-jelly for 20 seconds and the mance.1-3 Tanaka et al,1 who investigated the re-
amount of glucose extraction was measured. The re- lation between the masseter muscle activity, du-
sults for the two devices were compared.
ration of chewing, and the amount of glucose
Results: The glucose concentration according to the
extraction from chewing gummy-jelly as measu-
spectrophotometer was very close to the true concen-
tration. On the other hand, the glucose concentration reed with a spectrophotometer, concluded that
according to the blood glucose testing device was not masticatory performance could be quantitatively
close to the true value, although a statistically signifi- measured by analyzing the amount of glucose ex-
cant linearity was found. In regard to measurement of traction from chewing gummy-jelly. However, one
the amount of glucose extraction, there was a signifi- of the drawbacks of this measurement method is
cant difference between the results obtained with the that the spectrophotometer used for measuring
two devices. However, by using a regression line and the amount of glucose extraction is cumbersome
applying the corrections to the values obtained, the to use and expensive.
values of the blood glucose testing device could be On the other hand, with the increase in the
brought close to the values of the spectrophotometer
number of diabetic patients4,5 and the necessity
(y=1.036x−2.491, r =0.994, P <0.01).
to control the blood glucose levels in these pa-
Conclusion: From these results, it was concluded
that, after applying correction, a portable blood glu- tients, relatively inexpensive and small, portable
cose testing device could measure the amount of glu- blood glucose testing devices 6-9 have become
cose extraction from chewing gummy-jelly as accu- available in the market, using which patients
can easily measure their own blood glucose lev-
Corresponding to: Dr Hiroshi Shiga els. There is a possibility that this portable blood
Department of Partial and Complete Denture, glucose testing device could be used to measure
The Nippon Dental University, School of Dentistry at Tokyo, the amount of glucose extraction from chewing
1-9-20 Fujimi, Chiyoda-ku, Tokyo 102-8159, Japan gummy-jelly.
Tel: +81-3-3261-5729, FAX: +81-3-3261-8464
E-mail: h-shiga@tokyo.ndu.ac.jp
Therefore in this experiment, in order to inves-
tigate the possibility of objectively evaluating
Received on September 15, 2005 / Accepted on November 11, 2005 masticatory performance using a portable blood
15
Shiga et al., Prosthodont Res Pract 5 : 15-20, 2006
30~ 45 seconds
Placing on the blood glucose testing strip
Stir
glucose testing device through the comparison of tube and stirred. The two tubes were then incu-
the amount of glucose extraction from chewing bated at room temperature for 15 minutes and the
gummy-jelly measured using a spectrophotome- difference in the degree of color absorption be-
ter and a portable blood glucose testing device. tween the glucose solution and the distilled water
was measured with the spectrophotometer (Fig. 1).
The glucose concentration of the solution was cal-
culated using the glucose concentration vs. the de-
Materials and methods gree of color absorption graph.
Experiment 1 was performed for calibration of the In the blood glucose testing device method, the
portable blood glucose testing device and experi- glucose concentration was measured simply by
ment 2 was performed to investigate the valida- placing a few drops of the glucose solution using a
tion of the portable blood glucose testing device in micro-pipette on the blood glucose testing strip
measuring masticatory performance. and reading the valued automatically displayed
after 30 to 45 seconds (Fig. 1).
Experiment 1
Four graded of glucose solutions with concentra- Experiment 2
tions of 100 mg/dl, 200 mg/dl, 300 mg/dl, and 400 1. Subjects and test material
mg/dl were prepared. Each solution was measured Twenty subjects, all in their twenties, consisting
five times using a spectrophotometer (Model 200− of 10 males and 10 females, with an average age
20, Hitachi, Tokyo, Japan) and a portable blood of 24.6 years, who were aware of their habitual
glucose testing device (Accu-check® comfort, Roche chewing side and had neither systemic nor masti-
Diagnostic, Basel, Switzerland) with the immobi- catory abnormalities were selected for the study.
lized enzyme electrode method. The relationship Informed consent was obtained from each of the
between the known concentrations and the con- subjects after explaining to them the general na-
centrations as measured with the spectrophoto- ture of the study to be conducted. The test food
meter, and similarly, the relationship between the was one cylindrically-shaped gummy-jelly with a
known concentrations and the concentrations as diameter of 10 mm, height of 10 mm, and weight
measured with the blood glucose testing device of 2 g (Table 1).
were compared. The glucose concentration as
measured with the spectrophotometer was ob- 2. Recording and analyzing method
tained with the glucose oxidase method. First, 0.02 Subjects were asked to sit in a dental chair with
ml of the glucose solution was transferred to a test the Frankfort plane parallel to the floor and the
tube by a micro-pipette followed by the addition of head not being fixed but in a relaxed state, and
3.0 ml of the coloring agent and stirring. As a con- were asked to chew the gummy-jelly using their
trol, 0.02 ml of distilled ionized water and 3.0 ml habitual chewing side for 20 seconds. Thereafter,
of the coloring solution were added to another test they were asked to rinse their mouth with 10 ml
16
Validation of Blood Glucose Testing Device in Measuring Masticatory Performance
Table 1 Size, weight, hardness, and ingredients Table 2 The average and standard deviation of the glucose
of the gummy-jelly. concentrations of the 4 types of glucose solutions as mea-
sured with the spectrophotometer and as measured with the
Size ( mm ) φ10×10 blood glucose testing device.
(mg/dl)
Weight ( g ) 2
Glucose solution 100 200 300 400
Har dn e ss ( k gf ) 2.17
of distilled water and spit the extracted glucose as Table 3 The average and standard deviation of the glucose
well as the gummy-jelly into a cup containing a concentrations of the 4 types of glucose solutions as mea-
sured with the blood glucose testing device after correction.
filter. The filtrate served as the test material, and
the glucose concentration was measured with both (mg/dl)
the spectrophotometer and the blood glucose test-
ing device. The measurements for each filtrate Glucose solution 100 200 300 400
was conducted three times, and the average value
Glucose concentration 101.0 196.7 303.3 398.7
was taken for the analysis. The relationship be- after correction ± 1.99 ± 3.32 ± 3.57 ± 4.55
tween the glucose concentrations measured using
the spectrophotometer and those measured using
the blood glucose testing device was investigated.
17
Shiga et al., Prosthodont Res Pract 5 : 15-20, 2006
( mg/dl ) ( mg/dl )
400 400
(after correction)
200 200
Fig. 2 The relationship between the amount of glucose extraction as measured with
the spectrophotometer and that as measured with the blood glucose testing device.
18
Validation of Blood Glucose Testing Device in Measuring Masticatory Performance
a blood glucose testing device as with a spectro- ance through measurement of the amount of
photometer. glucose extraction.
When the blood glucose testing device was
used to measure the glucose concentration of the
solutions with known glucose concentrations, it
was found that lower values were obtained for Conclusion
the 100 mg/dl solution and higher values were In order to investigate the possibility of objectively
obtained for the 200 mg/dl, 300 mg/dl and 400 mg/ evaluating masticatory performance using a por-
dl solutions. Two reasons for this discrepancy table blood glucose testing device, the amount of
were contemplated. One is that the blood glucose glucose extraction from chewing gummy-jelly
testing device lacked accuracy when measuring measured using a spectrophotometer was com-
glucose in solution, which was probably due to pared with that measured with a portable blood
the fact that it was designed to measure glucose glucose testing device. Based on the results, it
concentration in arterial blood taken from a fin- was concluded that the amount of glucose extrac-
ger tip. A much larger testing devices in hospitals tion from chewing gummy-jelly measured using
was built for use in hospitals to measure the glu- the blood glucose testing device was as accurate
cose concentrations in blood of the veins of the as that measured using the spectrophotometer
forearm, so the smaller portable blood glucose when corrections were applied. In other words,
testing device had to change the glucose concen- the blood glucose testing device could quantita-
tration value of the finger tip to that of the fore- tively measure masticatory performance as accu-
arm.8 The other reason is that when diabetic pa- rately as the spectrophotometer, and may thus of-
tients use these devices to control their blood fer a simple and comparatively inexpensive
sugar levels, values lower than the actual can be alternative to more complex and expensive meth-
very dangerous ; For that reason, since the nor- ods.
mal value range for blood sugar is 65−110 mg/dl
on an empty stomach and 65−160 mg/dl right af-
ter a meal, the blood sugar measuring device may
have been preset to display higher than actual References
values for concentrations above 160 mg/dl. 1. Tanaka A, Shiga H, Kobayashi Y. Quantitative
As a quantitative method of evaluating the evaluation of mandibular movements and masti-
masticatory efficiency, the sieve method11,12 had catory muscular activities by analyzing the
been used for a long time in the field of dentistry, amount of glucose discharge during gummy-jelly
but it has its shortcomings, e.g., the procedure is chewing. J Jpn Prosthodont Soc 38: 1281-1294,
complex and time-consuming. On the other hand, 1994.
measuring the amount of glucose extraction from 2. Kuriyama S, Shiga H, Kobayashi Y. Property of
chewing gummy-jelly with a spectrophotometer test food for masticatory movement analysis, the
takes only 20 to 25 minutes for recording and hardness of gummy-jelly. J Jpn Prosthodont Soc
analysis, however, it is costly, special agents are 45: 592-601, 2001.
required for quantitative determination of color, 3. Okiyama S, Ikebe K, Nokubi T. Association be-
and procedure is relatively troublesome. In con- tween masticatory performance and maximal oc-
trast, the blood glucose testing device tried in clusal force in young men. J Oral Rehabil 30:
this experiment only requires the patient to rinse 278-282, 2003.
his/her mouth with 10 ml of distilled water and 4. Dec GW, DeSanctis RW. Cardiomyopathies. In:
place the test strip in the filtrate, and the glucose Dale DC, Federman DD, editors, Scientific Ameri-
concentration would be automatically displayed can Medicine vol. 1, Part 1, Chap. 14 New York:
after in 30 to 45 seconds It was found in this ex- Scientific American, 2000.
periment that although the glucose concentra- 5. Mandrup-Poulsen T. Recent advances-diabetes.
tions were not measured accurately with the glu- Br Med J 316: 1221-1225, 1998.
cose testing device, application of correction 6. Conrad PD, Sparks JW, Osberg I et al. Clinical
factors yielded results that were close to the ac- application of a new glucose analyzer in the neo-
curate results obtained using the spectrophotom- natal intensive care unit: comparison with other
eter, and that therefore, this method may be use- methods. J Pediatrics 114: 281-287, 1989.
ful for clinical evaluation of masticatory perform- 7. Edelman SV, Callahan P, Deeb LC. Multisite eval-
19
Shiga et al., Prosthodont Res Pract 5 : 15-20, 2006
uation of a new diabetes self-test for glucose and Advantage II portable blood glucose monitoring
glycated protein (fructosamine). Diabet Tech Ther system with immobilized enzyme electrode meth-
2: 233-238, 2000. od and it’s clinical usefulness. Medicine & Phar-
8. Tieszen KL, New JP. Alternate side blood glucose macology 45: 645-649, 2001.
testing: do patients refer it? Diabet Med 20: 11. Manly RS, Braley LC. Masticatory performance
325-328, 2003. and efficiency. J Dent Res 29: 448-462, 1950.
9. Riddle MC. Measuring fructosamine at home. 12. Kaper KK, Yurkstas AA. Test foods for measuring
Diabet Tech Ther 2: 239-240, 2000. masticatory performance of denture wearers. J
10. Hirata K, Ito C, Kataoka M et al. Evaluation of Prosthet Dent 14: 483-491, 1964.
20