Professional Documents
Culture Documents
Dseurope: Pitagora
Dseurope: Pitagora
MULTICHANNEL MICROPROCESSOR
SIGNAL CONDITIONER
SERIES Pitagora
Version 1.19
dated 31/01/07
INSTRUCTION
MANUAL
-2-
DOCUMENT REVISIONS
FOREWORD
1. This manual is integral part of the shipment and it is supplied with the product even if not listed
in the invoice.
This manual can also be sent during purchase negotiations on Buyer’s request provided that the
Buyer states end application of the Product.
Except as expressly requested by the Buyer, only one copy of this manual may be attached to
shipments consisting of more than one unit of the Product.
Additional or updated copies of this document may be found on our web site at the following
address: www.dseurope.com
2. The products, which are the subjects of this manual, are designed and manufactured for general
use; therefore, being not possible to specify limits of use for all applications, the Buyer shall
introduce safety measures in order to avoid personal injuries, damages to property or damages
due to plants stoppage etc.
In case of use that can cause risk of damages, the Buyer shall immediately a notice the Supplier
in order that It can suggest safety solutions or can decide to reject the order and not deliver the
Product.
3. Series Pitagora instruments are not certified instruments; it is Buyer’s responsibility to check
instrument suitability to the requirements of the specific application.
4. Series Pitagora instruments, even when connected to transducers or other equipment, are only
parts of more complex systems and plants; they are sold in thousands pieces. Every year, for
most different applications which shall meet as many different norms that are not always known
to the Supplier.
For these reasons, DS Europe is obliged to disclaim all responsibility concerning the use of the
instrument; however, a list of most elementary and basic precautions for the correct use of the
instrument is given in this manual.
Besides, the need of a complete and specific insurance is highlighted, especially when systems
and machinery are to be installed in Countries like United States or Canada.
5. Series Pitagora instruments are designed and manufactured by DS Europe conforming to the
highest quality standards; the instruments are rugged and designed to the highest possible
safety and reliability: limitations and precautions listed in this manual are aimed to draw
Buyer’s attention to the importance of avoiding damage.
For the above mentioned reasons, It is hereby certified that the delivered Product is
completely operating.
DS EUROPE S.r.l.
Via F. Russoli , 6
20143 - MILANO - ITALY
Tel. +39 02 8910142 Fax. +39 02 89124848 8910145
Internet : www.dseurope.com E-mail : dseurope@dseurope.com
INDEX
1 CE CONFORMITY DECLARATION .......................................... 6
2 Electrical connections ...................................................................... 7
2.1 Rear terminal boards .......................................................................................................7
2.2 Power supply connections................................................................................................7
2.2.1 Accepted power supply................................................................................................7
2.3 Connection of I1 single channel input card for 80 mV differential signal ......................8
2.4 Connection of I2 two channels input card for 80 mV differential signal ........................8
2.5 Connection of I4 single channel input card for 0-5 V and 0-10 V signals ......................8
2.6 Connection of I6 single channel input card for 4-20 mA and 0-20 mA (2 wire) signals 8
2.7 Connection of I10 single channel input card for Start-Stop digital signal......................9
2.8 Connection of output cards O1 = 0-5 V and O2 = 0-10 V...............................................9
2.9 Connection of output cards O3 = 4-20 mA and O4 = 0-20 mA .....................................9
2.10 Connection of D0 card, two relays alarm level..............................................................10
2.11 Connection of D1 card, four optically insulated alarm levels .......................................11
2.12 Connection of external contacts.....................................................................................12
2.13 Connection of grounding for CE regulation ..................................................................12
2.14 Connection of RS232 serial port (COM1) .....................................................................13
2.15 Connection of RS485 (RJ-11b connector) serial com port............................................13
2.15.1 Terminator Resistances ..............................................................................................14
3 Menu setting – general sequence .................................................. 14
4 Display............................................................................................. 15
5 Instrument setting .......................................................................... 16
6 AdC setting (I1 or I4 input cards)................................................... 18
7 dAC setting (O1, O2, O3, O4 analog output cards)....................... 21
8 Md setting (Modbus communication setting) ................................ 22
9 LEV alarm level setting .................................................................... 23
10 MAG setting (I10 Start Stop input card) ......................................... 24
11 LCD display setting........................................................................ 25
12 Return to Measure modality ......................................................... 25
13 Introduction to measure modes .................................................... 26
14 Two channel summing unit (MD=1)............................................. 27
15 Three channel summing unit (MD=2) .......................................... 28
16 Four channel summing unit (MD=3)............................................ 29
17 Four channel summing unit – analog output (MD=4) ................ 30
18 Four channel summing unit (MD=5)............................................ 31
19 Four channel summing unit – analog output (MD=6) ................ 32
Pitagora instruction manual V1 r 19 - 31/01/07
-5-
1 CE CONFORMITY DECLARATION
Conformity to standards:
EMC:
EN 61000-6-4 (2001)
EN 61000-6-2 (2001)
EN 55011 (1998)+ A1 (1999) + A2 (2002)
EN 55014-1 (2000) + A1 (2001) + A2 (2002)
EN 61000-4-2 (1995) + A1 (1998) + A2 (2002)
EN 61000-4-3 (1996) + A1 (1998) + A2 (2001)
EN 61000-4-4 (1995) + A1 (2001) + A2 (2002)
EN 61000-4-5 (1995) + A1 (2001)
EN 61000-4-6 (1996) + A1 (2001)
EN 61000-4-8 (1993) + A1 (2001)
EN 61000-4-11 (1994) + A1 (2001)
Model: Pitagora
The equipment has been tested in the typical installation configuration, as described by the
instruction manual of the product.
DS Europe Srl certify that the above defined equipment meets the requirements of above mentioned
EMC directives.
2 Electrical connections
2.1 Rear terminal boards
D C B A
1 1 1 1
4321
RJ-11A
CanBus 2 2 2 2
3 3 3 3
4 4 4 4
4321
RJ-11B 5 5 5 5
RS-485
6 6 6 6
RS-232 INPUT
5 4 3 2 1
9 8 7 6
EARTH
5 4 3 2 1
Terminal Description
Power supply generator must be in position to provide at least 10W power (depending on the
number and type of connected transducers).
On terminal connector is available the earth terminal for instrument grounding (see par. 2.11).
2.5 Connection of I4 single channel input card for 0-5 V and 0-10 V signals
Terminal Description Notes DS Europe colour code
1 + power supply to transducer + power supply for amplified transducer. RED
2 + signal input from transducer Example: ±10 V in put, maximum voltage WHITE
3 - signal input from transducer ± 18 V YELLOW or BLACK
Between terminal 1 and 6 It is available
6 - power supply to transducer the power supply fed to the instrument (i.e. YELLOW or BLACK
+24 V)
Terminals 4-5 are not connected
2.6 Connection of I6 single channel input card for 4-20 mA and 0-20 mA (2 wire)
signals
IMPORTANT NOTE: Power supply fed to the transducer, with current loop, is the same that is
provided to the Pitagora instrument.
2.7 Connection of I10 single channel input card for Start-Stop digital signal
Magnetostrictive transducer
Terminal Description
DS Europe series PCX-ss
1 + RTX Brown
SERIAL INPUT ( RS-422 )
2 - RTX Blue
3 - TXD Yellow
SERIAL INPUT ( RS-422 )
4 + TXD White
5 + 24 Volt ( + Power supply ) from external power
Red
supply generator
6 GND ( - Power supply ) Black
IMPORTANT NOTE: power supply fed to the magnetostrictive transducer is the same provided
to the instrument.
Power supply is to be:
• Not lower than 18 V, if connection cable to transducer is long no more than 20 mt.
• Not lower than 24 V, if connection cable to transducer is long no more than 100 mt.
• In case of longer distances, verify that at least 18 V are powering the connected
transducer.
Calibration and measuring functions of magnetostrictive transducers are like all other types of
transducers (load cells, pressure transducers etc.).
Note: under request this card can be supplied with optional voltage and current output, at the
same time, with the same setting.
Terminal Description
1 Relay 1 Normally Closed contact Relays technical specifications:
2 Relay 1 Common contact - Maximum Vdc: 30 Vdc.
- Maximum current: 2 A.
3 Relay 1 Normally Open contact
- Maximum power: 1W.
4 Relay 2 Normally Closed contact - Switching time: 2 mS.
5 Relay 2 Common contact
6 Relay 2 Normally Closed contact
Terminal Description
1 Optically insulated output 1 ( open collector) Opto insulator technical features
2 Optically insulated output 2 ( open collector) - Maximum Vdc: 70 Vdc.
- Maximum current: 50 mA.
3 Optically insulated output 3 ( open collector))
- Maximum power: 150 mW.
4 Optically insulated output 4 ( open collector) - Switching time: 18µSmax.
5 Emitter, common output - Insulation: 3000 Vrms
6 Not Connected
1 +24V
4
+24V
5
6
Slot A o B
+24V
OUTPUT INTERFACE
24V Common
Above figure shows the typical connection used for PLC “SINK” type connection.
It is to be considered that this is an “open collector” setting and that emitters are connected
together to terminal 5 of the instrument. Internal protection diodes are applied, for reverse
polarity protection.
AN 401 PLUS
1
RELAY-DPDT
I max 50 mA
2
4 V_Relè
6
Slot A o B
OUTPUT INTERFACE
External relays, connection to optically insulated outputs (same connection for the other 3).
INPUT
INPUT
1
+24Vdc 1
EXGND
1
EARTH
EXGND
2 KEY1
2 KEY1
3 KEY2
3 KEY2
4 KEY3
4 KEY3 +24Vdc
5 KEY4
5 KEY4
Above figures (Fig.1a and Fig.1b ) show the connection for push buttons use (same for relays,
TTL, transistors etc.).
A voltage, between 12 and 28 Vdc, is to be fed in order to activate the opto-insulator LED’s.
Instrument Computer
TX ⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯> RX
RX <⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯ TX
GND ⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯ GND
Usual connections for RS232 communication, between Instrument and computer are (standard straight serial cable):
▼ Down arrow: It allows to scroll all the possible choices of a menu or to decrease a
parameter’s value.
▲ Up arrow: It allows to scroll all the possible choices of a menu or to increase a
parameter’s value.
E Enter: It confirms modified parameter and stores the value on the instrument.
C ESC: It allows to exit from a menu without confirming any change to the parameter’s
value or to zero (dynamic zero) the value shown on display when Pitagora is in normal
working status.
These push buttons can have additional functions, depending on the chosen operating modality.
Push button can be damaged, if sharp objects are applied (screwdrivers or similar).
By keeping pressed the E button, when modifying a parameter, allows to show on display (prior to the
button release) the numeric index corresponding to the parameter (i.e.: P03501). The index can be used
to modify the parameter by means of digital serial communication.
5
2
4
4 Display
Pitagora instrument is has an LCD display, with backlight, able to show the following data:
1. Main 7 alphanumeric digits display, 8 segments, plus polarity able to display numbers up to ±
9999999 or words (capital or small letters).
2. Secondary display, with 14 alphanumeric digits, used to show information referable to the
measured value displayed on the main displayed or used to provide additional indications
during instrument customization.
3. Bar display, with central zero, to display the behaviour of the measured value.
4. MENU indication, active when Pitagora instrument is used during Customization procedure
with a numeric value that indicates the active personalization level.
5. Serial communication and data transfer activity.
When Pitagora instrument is switched on it will show, in sequence, the following information:
1. Pitagora Name of instrument
2. Pt 401 Hardware version
3. V 1.00 Installed firmware version
4. SN. X Serial number of instrument
5. MD 1 Active measurement modality
5 Instrument setting
Switch on the instrument, keeping pressed at the same time ▲ and C push buttons.
Instrument will allow to setup the card eventually installed into slot A.
SLSET will confirm on the display that personalization menu is active (SLot SETup).
“MENU 1-” indicates that setup procedure for the chosen card is on the topmost level of the main
personalization menu.
When push button are released display shows on upper alphanumeric line the type of menu (SLOT
SETUP); main display line indicates the name of the installed card to be customized.
Firmware uses the following syntaxes, when setting card parameters:
1. Slot reference: a letter, followed by a decimal point, to indicate the reference position of the
card installed on the instrument (real card) or to indicate virtual card reference.
2. Function/parameter reference: 3 or more letters that identify the function/parameter to be set.
If an A/D card (mod. I4) is installed in slot C, setup menu will refer to It as C.AdC (C. to indicate
slot number = C; AdC to indicate card type = A/D converter); B.AdC indicates that an A/D card
is installed on slot B.
Hereafter is indicated the sequence that allows to set the A/D parameters of an input channel.
A/D channels are identified by a numeric index, displayed as prefix of the parameter to be set.
• Channel A/D 0 (CH0) = setting of parameters 0.AdSP, 0.VLo, 0.VHI, 0.VtAr, 0,PSdP
Transducer connected to this channel is identified as Bridge 1
• Channel A/D 1 (CH1) = setting of parameters 1.AdSP, 1.VLo, 1.VHI, 1.VtAr, 1,PSdP
Transducer connected to this channel is identified as Bridge 2
When a new parameter is set and stored, by keeping pressed E push button, a numeric index is shown
prior the release of the push button (i.e. P03501), to indicate the parameter’s reference.
This index reference is useful when Modbus serial utility is used, as it allows an easy understanding of
the parameter’s dictionary to be used with a remote connection (serial connection) allowing an
advanced use of the unit.
The indicated A/D converter setup structure applies to channel 0 and channel 1 of a two channel A/D
input card.
A multi channel Pitagora unit, with two I4 input cards, will have the following parameters to be set:
Slot (example) Type of setting Available channels Settings
B B.AdC Ch0 0.AdSP
0.VLo
0.VHI
0.VTar
0.PSdP
CH1 1.AdSP
1.VLo
1.VHI
1.VTar
1.PSdP
C C.AdC Ch0 0.AdSP
0.VLo
0.VHI
0.VTar
0.PSdP
CH1 1.AdSP
1.VLo
1.VHI
1.VTar
1.PSdP
It is important to note that the AdSP settings for a summing unit are to be the same for all the channels,
so that the result of the sum is obtained using measurements that can be related each other on a time
basis.
Cont (Continuous) parameter allows to set the instrument in order to:
Cont = 0 Pitagora samples values from channels CH0 and CH1, from selected A/D converter.
Cont = 1 Pitagora samples data solely from CH0, from selected A/D converter.
Note: the analog output card, whenever installed, is active with all operating modes of Pitagora
instrument.
DMODE = 1 DMODE = 0
Measure 1. NET SUM (sum of 4 channel measures = C.ADC0 + C.ADC1 + D.ADC0 + D.ADC1)
2. PEAK MIN (Minimum peak value of Net measure NET SUM)
3. PEAK MAx (Maximum peak value of Net measure NET SUM)
4. P.C.0 Net Value (Measure from channel 1 = C.ADC0)
5. P.C.1 Net Value (Measure from channel 2 = C.ADC1)
6. P.D.0 Net Value (Measure from channel 3 = D.ADC0)
7. P.D.1 Net Value (Measure from channel 4 = D.ADC1)
Display • NET (GROSS – DYNAMIC ZERO) and PEAK, if activated
Push • C push • DYNAMIC ZERO = currently measured value
buttons button ( value not stored on permanent memory, lost when power supply is
switched off)
• E push • Press button to show measure 1 on display
button
• push • Press button to swap measures shown on display, from 1 to 7.
button
• Push • Alarm level settings
button
Inputs • Key 1 • TARE = currently measured value
• Key 2 • Hold of the measure
• Key 3 • PUSH BUTTONS blocked
• Key 4 • Activates the peak function. Previously stored positive and negative
peak values are deleted.
Relays • RL2 • 0 alarm level function
• RL3 • 1 alarm level function
Measure 1. NET SUM (sum of 4 channel measures = B.ADC0 + B.ADC1 + C.ADC0 + C.ADC1)
2. P.B.0 Net Value (Measure from channel 1 = B.ADC0)
3. P.B.1 Net Value (Measure from channel 2 = B.ADC1)
4. P.C.0 Net Value (Measure from channel 3 = C.ADC0)
5. P.C.1 Net Value (Measure from channel 4 = C.ADC1)
Display • NET (GROSS – DYNAMIC ZERO) and PEAK, if activated
Push • C push • DYNAMIC ZERO = currently measured value
buttons button ( value not stored on permanent memory, lost when power supply is
switched off)
• E push • Press button to show measure 1 on display
button
• push • Press button to swap measures shown on display, from 1 to 5
button
• push • Alarm level settings
button
Inputs • Key 1 • TARE = currently measured value
• Key 2 • Hold of the measure
• Key 3 • PUSH BUTTONS blocked
• Key 4 • Activates the peak function. Previously stored positive and negative
peak values are deleted .
Relays • RL2 • 0 alarm level function
• RL3 • 1 alarm level function
Measure 1. NET SUM (sum of 4 channel measures = C.ADC0 + C.ADC1 + D.ADC0 + D.ADC1)
2. P.C.0 Net Value (Measure from channel 1 = C.ADC0)
3. P.C.1 Net Value (Measure from channel 2 = C.ADC1)
4. P.D.0 Net Value (Measure from channel 3 = D.ADC0)
5. P.D.1 Net Value (Measure from channel 4 = D.ADC1)
Display • NET (GROSS – DYNAMIC ZERO) and PEAK, if activated
Push • C push • DYNAMIC ZERO = currently measured value
buttons button ( value not stored on permanent memory, lost when power supply is
switched off)
• E push • Press button to show measure 1 on display
button
• push • Press button to swap measures shown on display, from 1 to 5
button
• push • Alarm level settings
button
Inputs • Key 1 • TARE = currently measured value
• Key 2 • Hold of the measure
• Key 3 • PUSH BUTTONS blocked
• Key 4 • Activates the peak function. Previously stored positive and negative
peak values are deleted.
Relays • RL2 • 0 alarm level function
• RL3 • 1 alarm level function
Measure 1. NET SUM (sum of 4 channel measures = B.ADC0 + B.ADC1 + C.ADC0 + C.ADC1)
2. P.B.0 Net Value (Measure from channel 1 = B.ADC0)
3. P.B.1 Net Value (Measure from channel 2 = B.ADC1)
4. P.C.0 Net Value (Measure from channel 3 = C.ADC0)
5. P.C.1 Net Value (Measure from channel 4 = C.ADC1)
Display • NET (GROSS – DYNAMIC ZERO) and PEAK, if activated
Push • C push • DYNAMIC ZERO = measured value
buttons button ( value not stored on permanent memory, lost when power supply is
switched off )
• E push • Press button to show measure 1 on display
button
• push • Press button to swap measures shown on display, from 1 to 5
button
• push • Alarm level settings
button
Inputs • Key 1 • TARE = currently measured value
• Key 2 • Hold of the measure
• Key 3 • PUSH BUTTONS blocked
• Key 4 • Activates the peak function. Previously stored positive and negative
peak values are deleted.
Relays • RL2 • 0 alarm level function
• RL3 • 1 alarm level function
Press push button to enter the alarm level setting menu.
Tab. 1 – Alarm level setting menu
Parameter Description Default Allowed values
0.LEV 0 alarm level threshold 1000 ±999999
0.HIS 0 alarm level hysteresys 0 From 0 to 50
1.LEV 1 alarm level threshold 1000 ±999999
1.HIS 1 alarm level hysteresys 0 From 0 to 50
-12345
Main display It shows the tension /
compression force measured by
the DS Europe load cell.
It is possible to set Kg or
Newton engineering units.
Operating mode
Measure • Force (main or alternative engineering unit) and displacement.
• Force peak at sample breaking (main engineering unit or alternative)
• Displacement peak at sample breaking.
Display • NET force = (GROSS force – DYNAMIC ZERO force).
• NET displacement = (GROSS displacement – DYNAMIC ZERO displacement).
• Lev 1 indicator blinks, to indicate that test is running and relay RL2 is active.
• Lev 1 indicator is switched on, continuously, when test ends; displayed force and
displacement measurements, corresponding to sample breaking, have been stored.
RL2 is inactive.
• Lev 2 indicator, when switched on, indicates that measured force is ≥ to the alarm
level
LEV 1; RL3 is active.
Push C push • Stops the test cycle and RL2 is inactive.
buttons button On display Lev1 indicator is switched off, to indicate that test cycle is on
hold. Force and displacement measures are possible.
E push • Begins a new test cycle.
button Force value, shown on main display, is zeroed (=DYNAMIC ZERO).
Displacement value, on secondary display, is zeroed (=DYNAMIC
ZERO).
Relay RL2 is activated and Lev1 indicator blinks on display, till RL2 is
active.
• At breaking of sample (force reduction ≥ DELTA) display shows force
and displacement values when sample broke.
Lev1 indicator is switched on, continuously, to indicate that test end due
to sample breaking; RL2 is deactivated.
push • Allows to swap between all available measures performed by Pitagora
button (make reference to tab. 2, for types of available measures).
push • With KEY1 open, DELTA value of breaking can be set. (see Tab.3)
button • With KEY1 closed values of DELTA, UNIT, LEV1 and HYST can be
set (see tab. 4).
Inputs Key 1 • Refer to push button
Key 2 • Same functions as E push button.
Key 3 • Deactivates the front panel of Pitagora instrument (no pushbuttons
active).
Key 4 • No function
Relays RL2 • Refer to push button E
RL3 • Level 1 functions
Active when measured force (load cell) is greater then the alarm level 1.
Lev2 indicator is switched on, till measured force is greater or equal to
the value of Level 1.
Pressing
push button, with KEY1 open, activates SETUP sequence [Tab. 3]
Sequence Main display Description
DELTA Sets the DELTA breaking reference value.
Pitagora instrument stops the test when measured force decreases
1 of an amount greater or equal to the DELTA reference value,
having broken the sample. Delta value is to be set accordingly to
measurement’s resolution set for the main display.
Pressing push button, with KEY1 closed, activates extended SETUP sequence [Tab. 4]
Sequence Main display Description
DELTA Sets the DELTA breaking reference value.
Pitagora instrument stops the test when measured force decreases
1 of an amount greater or equal to the DELTA reference value,
having broken the sample. Delta value is to be set accordingly to
measurement’s resolution set for the main display.
UNIT Sets the engineering units of the measured force.
2 0 = Kg
1=N
3 LEV1 alarm level load, in Kg, for relay RL3
4 HYST1 Hysteresys value for relay RL3.
Press and to scroll between the listed parameters and to return to combined
force/displacement measurement, so to start a new test cycle.
Press C to come to the force / displacement measure modality.
Measure 1. NET SUM (sum of 4 channel measures = B.ADC0 + B.ADC1 + C.ADC0 + C.ADC1)
2. P.B.0 Net Value (Measure from channel 1 = B.ADC0)
3. P.C.0 Net Value (Measure from channel 3 = C.ADC0)
Measure 1. NET SUM (sum of 4 channel measures = B.ADC0 + B.ADC1 + C.ADC0 + C.ADC1)
2. P.b.0 Net Value (Measure from channel 1 = B.ADC0)
3. P.B.1 Net Value (Measure from channel 2 = B.ADC1)
4. P.C.0 Net Value (Measure from channel 3 = C.ADC0)
5. P.C.1 Net Value (Measure from channel 4 = C.ADC1)
6. P.D.0 Net Value (Measure from channel 5 = D.ADC0)
7. P.D.1 Net Value (Measure from channel 6 = D.ADC1)
Display • NET (GROSS – DYNAMIC ZERO) and PEAK, if activated
Push • C push • DYNAMIC ZERO = measured value
buttons button ( value not stored on permanent memory, lost when power supply is
switched off )
• E push • Press button to show measure 1 on display
button
• push • Press button to swap measures shown on display, from 1 to 7
button
• push • Alarm level settings
button
Inputs • Key 1 • TARE = currently measured value
• Key 2 • Hold of the measure
• Key 3 • PUSH BUTTONS blocked
• Key 4 • Activates the peak function. Previously stored positive and negative
peak values are deleted.
Relays • RL2 • Not active
• RL3 • Not active
39 Serial connection
Serial connection between Pitagora and remote computer allows to handle, graphically, the
measured data, to save measurements into the computer hard disk as TXT or XLS format files and
to print the graph or test reports.
A data acquisition software (see above as example) for computer can be ordered to DS Europe.
Additional information can be downloaded from web site www.dseurope.com.
Contact DS Europe to obtain indications about possible customization of this software tool to Your
measurement’s needs.
40 Sales conditions
(for all DS Europe products)
40.1 WARRANTY
Several electrical and mechanical tests carried out during the manufacturing process an the final test carried out on
each unit warrant that the delivered Product is free from defects in materials, workmanship and performance.
During the warranty period of six months from delivery, at no additional charge, Product rendered defective under
normal use will be repaired or replaced at DS Europe factory.
The normal Product shall be forwarded at Buyer’s expenses concerning shipping, insurance, customs duties or any
other charges associated with transportation of the product.
This Limited Warranty does not extend to any product that has been damaged or rendered defective as a result of
accident or misuse.
Moreover, in case of heavy non-reparable damages, the product may be rendered disassembled to the buyer if the
cost f re-assemble of the Product will not be paid.
The Product contains firmware that is provided on an “AS IS” basis: essentially, firmware is in accordance with
the description of the Product manual.
Except as expressly set forth in this warranty, DS Europe makes not other warranties, expressed or implied,
including any implied warranties of merchantability and fitness for a particle purposes, concerning hardware and
firmware of the Product and its relevant documentation.
This limited warranty does not extend to any semiconductors: integrated circuits, transistors, diodes,
microprocessors, memories and whatever else not covered by semincoductor Manufacturers’ warranty.
The Buyer shall check the delivered Product within 10 days from receipt; after this limit, the Product shall be
considered accepted.
DS Europe liability is limited to the above; DS Europe is not liable for any personal injuries, damages to property
or damages to stoppage of machinery or plants caused by the installation and use of supplied Product (including,
without any limits, any lost profits, lost savings, stoppage of activities, lost information or any other economic
losses).
NOTE:
Even if not expressly mentioned, these “Terms of Sale” are integral and complimentary part of any bulletin,
invoice or instruction manual.
DS EUROPE S.R.L.
Terms of Sale N°140998 of September 14th, 1998.