TI047DAE Promag 50 53 P TI
TI047DAE Promag 50 53 P TI
Information Electromagnetic
TI 047D/24/ae
Flow Measuring System
PROline promag 50/53 P
Flow measurement in
chemical or process applications
Ue = B · L · v
Q =A·v
Ue = induced voltage
B = magnetic induction (magnetic field)
L = electrode gap
v = flow velocity
Q = volume flow
A = pipe cross-section
I = current strength
Transmitter:
• Promag 50 (user interface with push buttons for operation, two-line display)
• Promag 53 (“Touch Control” without opening the housing, four-line display)
Sensor:
• Promag P, 1/2” to 24” (DN 15 to 600)
2 Endress+Hauser
Promag 50/53 P
Input
Measured variable Flow rate (proportional to induced voltage)
Measuring range Typically v = 0.033 to 33 ft/s (0.01 to 10 m/s) with the specified measuring accuracy
Output
Pulse/frequency output:
Passive, open collector, 30 V DC, 250 mA, galvanically isolated.
Frequency output: full scale frequency 2 to 1000 Hz (fmax = 1250 Hz), on/off ratio 1:1,
pulse width max. 10 s
Pulse output: pulse value and pulse polarity selectable, max. pulse width
configurable (0.5 ms to 2 s), maximum pulse frequency selectable
PROFIBUS-PA interface
• PROFIBUS-PA in accordance with EN 50170 Volume 2, IEC 61158-2 (MPB),
galvanically isolated
• Current consumption: 11 mA
• Permissible supply voltage: 9 to 32 V
• FDE (Fault Disconnection Electronic): 0 mA
• Data transmission rate, supported bauderate: 31.25 kBit/s
• Signal encoding: Manchester II
• Function blocks: 1 x Analog Input, 1 x Totalizer
• Output data: Volume flow, Totalizer
• Input data: Positive zero return (ON/OFF), Control totalizer, Value for local display
• Bus address adjustable via DIP switches at the measuring device
Endress+Hauser 3
Promag 50/53 P
Promag 53
Current output:
Active/passive selectable, galvanically isolated, time constant selectable
(0.01 to 100 s), full scale value selectable, temperature coefficient: typically
0.003% o.r./°F (0.005% o.r./°C); resolution: 0.5 µA
• active: 0/4 to 20 mA, RL < 700 Ω (HART®: RL ≥ 250 Ω)
• passive: 4 to 20 mA, max. 30 V DC, Ri ≤ 150 Ω
Pulse/frequency output:
Active/passive selectable, galvanically isolated (I.S. version: passive only)
• active: 24 V DC, 25 mA (max. 250 mA during 20 ms), RL > 100 Ω
• passive: open collector, 30 V DC, 250 mA
Frequency output: full scale frequency 2 to 10,000 Hz (fmax = 12,500 Hz),
on/off ratio 1:1, pulse width max. 10 s
Pulse output: pulse value and pulse polarity selectable, pulse width configurable
(0.05 ms to 2 s), on/off ratio is 1:1 as of a frequency of 1 / (2 x pulse width)
PROFIBUS-DP interface:
• PROFIBUS-DP/PA in accordance with EN 50170 Volume 2, IEC 61158-2 (MPB),
galvanically isolated
• Data transmission rate, supported baudrate: 9.6 kBaud to 12 MBaud
• Automatic data transmission rate recognition
• Signal encoding: NRZ-Code
• Function blocks: 2 x Analog Input, 3 x Totalizer
• Output data: Volume flow, Corrected volume flow, Totalizer 1 to 3
• Input data: Positive zero return (ON/OFF), Totalizer control, Value for local display
• Bus address adjustable via DIP switches at the measuring device
PROFIBUS-PA interface:
• PROFIBUS-PA in accordance with EN 50170 Volume 2, IEC 61158-2 (MPB),
galvanically isolated
• Current consumption: 11 mA
• Permissible supply voltage: 9 to 32 V
• Data transmission rate, supported baudrate: 31.25 kBit/s
• Error current FDE (Fault Disconnection Electronic): 0 mA
• Signal encoding: Manchester II
• Function blocks: 2 x Analog Input, 3 x Totalizer
• Output data: Volume flow, Corrected volume flow, Totalizer 1 to 3
• Input data: Positive zero return (ON/OFF), Totalizer control, Value for local display
• Bus address adjustable via DIP switches at the measuring device
4 Endress+Hauser
Promag 50/53 P
Low flow cutoff Switch points for low flow cutoff are selectable
Galvanic isolation All circuits for inputs, outputs, and power supply are galvanically isolated from each
other.
Endress+Hauser 5
Promag 50/53 P
Power supply
Electrical connection
−27
measuring unit +26
f b
b e −25
+24
−23 d
+22
−21
+20
a
N (L-) 2 c
g L1 (L+) 1
a
N (L-)
L1 (L+)
e
+ − + − + − + −
1 2 20 21 22 23 24 25 26 27
a b a c b d
a Cable for power supply: 85 to 260 VAC, 20 to 55 VAC, or 16 to 62 VDC; power consumption: 15 VA / 15 W
Terminal No. 1: L1 fo AC, L+ for DC
Terminal No. 2: N for AC, L- for DC
b Signal cable: terminals No. 20 to 27 ´® see Page 8
c Grounding terminal for protective conductor
d Grounding terminal for signal-cable shield
e Service connection for connecting service interface FXA 193 (FieldCheck, ToF Tool-Field Tool Package)
f Cover for wiring compartment
g Cover securing clamp
6 Endress+Hauser
Promag 50/53 P
Electrical connection f
Measuring unit (bus communication)
DP (A)/PA (–)/FF (–) 27
DP (B)/PA (+)/FF (+) 26
(DGND) 25 b
b (+5 V) 24
23 d
g e
22
21
20 g
a
N (L-) 2 c
h L1 (L+) 1
a
(+5 V)
N (L–) (DGND)
L1 (L+) DP (B)/PA (+)/FF (+)
DP (A)/PA (–)/FF (–)
e
1 2 20 21 22 23 24 25 26 27
Esc
- + E
f
a g b a c g d b
a Cable for power supply: 85 to 260 VAC, 20 to 55 VAC, or 16 to 62 VDC; power consumption:15 VA / 15 W
Terminal No. 1: L1 fo AC, L+ for DC
Terminal No. 2: N for AC, L- for DC
b Fieldbus cable:
Terminal No. 26: DP (B) / PA (+) / FF (+) (with polarity protection)
Termianl No. 27: DP (A) / PA (-) / FF (-) (with reverse polarity protection)
DP (A) = RxD/TxD-N; DP (B) = RxD/TxD-P
c Grounding terminal for protective conductor
d Grounding terminal for signal-cable shield
e Service connector for connecting service interface FXA 193 (FieldCheck, ToF Tool-Field Tool Package)
f Cover for connection compartment
g Cable for external termination (PROFIBUS only)
h Cover securing clamp
Endress+Hauser 7
Promag 50/53 P
Order variant 20 (+) / 21 (-) 22 (+) / 23 (-) 24 (+) / 25 (-) 26 (+) / 27 (-)
The inputs and outputs on the communication board can be either permanently
assigned or variable, depending on the version ordered (see table). Replacement
for modules which are defective or which have to be replaced can be ordered as
accessories.
Order variant 20 (+) / 21 (-) 22 (+) / 23 (-) 24 (+) / 25 (-) 26 (+) / 27 (-)
8 Endress+Hauser
Promag 50/53 P
Electrical connection
Electrode Circuit
remote version
Meas. signal Pipe EPD
GND
S1 E1 E2 S2 E S Coil circuit
Esc
- + E
6 5 7 8 4 37 36 42 41
yellow
brown
green
white
2
1
n.c. n.c. n.c.
5 7 4 37 42 41
E1 E2 E
GND
n.c. = cable shields to be isolated and not to be connected
Signal cable:
• 3 x 20 AWG (0.38 mm2) PVC cable with common, braided copper shield,
approximately 0.28” diameter (Ø approx. 7 mm) and individually shielded cores.
• With Empty Pipe Detection (EPD): 4 x 20 AWG (0.38 mm2) PVC cable with
common, braided copper shield, approximately 0.28” diameter (Ø approx. 7 mm)
and individually shielded cores.
• Conductor resistance: ≤ 0.015 Ω/ft (50 Ω/km)
• Capacitance: core/shield: ≤ 128 pF/ft (420 pF/m)
• Permanent operating temperature: -5° to +180°F (–20 to +80 °C)
• Cable cross-section: maximum 16 AWG (2.5 mm2)
1
2
3
4
5
6
7
a b
Endress+Hauser 9
Promag 50/53 P
Caution:
Grounding is by means of the ground terminals provided for that purpose inside the
connection housing. Keep the stripped and twisted lengths of cable shield to the
terminals as short as possible.
NOTE: For installation in metal pipes, it is advisable to connect the ground terminal
of the transmitter housing to the piping.
Caution:
For sensors without reference electrodes or without metal process terminals, carry
out potential matching as per the instructions for special cases described below.
These special measures are particularly important when standard grounding
practice cannot be ensured or extremely strong matching currents are expected.
10 Endress+Hauser
Promag 50/53 P
Note:
The grounding cable for flange-to-flange connections can be ordered separately as
an accessory from Endress+Hauser.
• ≤ 12”; the grounding cable is in direct connection with the conductive flange
coating and is secured by the flange screws.
• ≥ 14”, the grounding cable connects directly to the metal transport bracket.
9 AWG Cu
Caution:
Risk of damage by electrochemical corrosion. Note the electrochemical insulation
rating if the grounding rings and measuring electrodes are made of different materi-
als.
Observe the plant internal grounding guidelines.
9 AWG Cu
Endress+Hauser 11
Promag 50/53 P
b b
a
9 AWG Cu
Performance characteristics
Reference conditions To DIN 19200 and VDI/VDE 2641:
• Medium temperature: +28 °C ± 2 K
• Ambient temperature: +22 °C ± 2 K
• Warm-up time: 30 minutes
Installation:
• Inlet run >10 x Dia
• Outlet run > 5 x Dia
• Sensor and transmitter grounded.
• Sensor centered relative to the pipe.
Promag 53:
Pulse output: ± 0.2% o.r. ± 2 mm/s (o.r. = of reading)
Current output: plus typically ± 5 µA
[%]
2.5
2.0
0.5 %
1.5
0.2 %
1.0
0.5
0
0 3 6 13 20 26 33
(1) (2) (4) (6) (8) (10)
v ft/s (m/s)
12 Endress+Hauser
Promag 50/53 P
Operating conditions
Installation conditions
Installation instructions Mounting location
Correct measuring is possible only if the pipe is full. Avoid the following locations:
• Highest point in the run. Risk of air accumulating
• Directly upstream from an open pipe outlet in a down pipe
h2xD
Installation of pumps
Do not install the sensor on the inlet side of a pump. This precaution is to avoid low
pressure and the consequent risk of damage to the lining of the measuring tube.
It might be necessary to install pulse dampers in systems incorporating reciprocat-
ing, diaphragm or peristaltic pumps. The sensor is shock and vibration resistant for
acceleration up to 2 g in accordance with IEC 68-2-6.
Caution:
To avoid the risk of solids accumulating, do not install the sensor at the lowest point
in the drain. It is advisable to install a cleaning valve.
2xD
5xD
Endress+Hauser 13
Promag 50/53 P
Vertical pipes
Install a siphon (b) or a vent valve (a) downstream of the sensor in vertical pipes
longer than 15 feet (5 meters). This precaution is to avoid low pressure and the
consequent risk of damage to the lining of the measuring tube. These measures also
prevent the system losing prime, which could cause air inclusions.
a
> 15 ft
(5 m)
Orientation
An optimum orientation helps avoid gas and air accumulations and deposits in the
measuring tube. Promag, nevertheless, supplies a range of options and accessories
for correct measuring of problematic mediums:
• Electrode Cleaning Circuitry (ECC) to remove electrically conductive deposits in
the measuring tube, e.g. in accretive mediums
• Empty Pipe Detection (EPD) for recognition of partially filled measuring tubes, or
for degassing mediums or for applications with fluctuating process pressure.
Vertical orientation:
This orientation is ideal for self-emptying piping systems and for use in conjunction
with Empty Pipe Detection.
Horizontal orientation:
The measuring electrode-plane should be horizontal. This prevents brief insulation
of the two electrodes by entrained air bubbles.
Caution:
Empty Pipe Detection functions correctly only when the measuring device is installed
horizontally and the transmitter housing is facing upward. Otherwise, there is no
guarantee that Empty Pipe Detection will respond if the measuring tube is only
partially filled or empty.
A
1
2 2
14 Endress+Hauser
Promag 50/53 P
Vibrations
Secure the piping and the sensor if vibration is severe.
Caution:
It is advisable to install the sensor and transmitter separately if vibration is exces-
sively severe. The sensor is shock and vibration resistant for acceleration up to 2 g
in accordance with IEC 68-2-6.
> 30 feet
(> 10 m)
Foundations, supports
If the nominal diameter is 14” or larger, mount the sensor on a foundation of ad-
equate loadbearing strength.
Caution:
Do not allow the sensor casing to take the weight of the sensor. This would buckle
the casing and damage the internal magnetic coils.
Inlet and outlet runs If possible, install the sensor well clear of fittings such as valves, T-pieces, elbows,
etc. Compliance with the following requirements for the inlet and outlet runs is
necessary in order to ensure measuring accuracy:
• Inlet run ≥ 5 x pipe diameter
• Outlet run ≥ 2 x pipe diameter
5xD 2xD
Adapters With the help of the appropriate adapters (reducers and expanders) the sensor can
be mounted in a pipeline of a larger diameter. For slowly flowing fluids, the resulting
higher velocity increases the measuring accuracy.
max. 4°
d D
Endress+Hauser 15
Promag 50/53 P
Length of connecting cable Permissible cable length Lmax depends on the conductivity of the medium. A
minimum conductivity of 20 µS/cm is required for measuring demineralized water.
[S/cm]
Esc
- + E
200
L max 100
5
ft (m)
33 330 650
(10) (100) (200)
L max
In order to ensure measuring accuracy, comply with the following instructions when
installing the remote version:
• Secure the cable run or route the cable in a conduit. Movement of the cable
can falsify the measuring signal, particularly if the conductivity of the medium
is low.
• Route the cable well clear of electrical machines and switching elements.
• Ensure ground equalization between the sensor and transmitter, if necessary.
Environment
Ambient temperature Standard: -5° to +140°F (-20° to +60°C) for sensor and transmitter
Optional: -40° to +140°F (-40° to +60°C)
Storage temperature 15° to + 120°F (-10° to +50°C), 70°F (20°C) is the preferable storage temperature.
• The measuring device must be protected against direct sunlight during storage
in order to avoid unacceptably high surface temperatures.
• Choose a storage location where moisture does not collect in the measuring
device. This helps prevent mold, fungus and bacteria from damaging the liner.
• Do not remove the protective plates or caps on the process connections until the
device is ready to install.
Degree of protection • Standard: NEMA 4X (IP 67) for transmitter and sensor
• Optional: NEMA 6P (IP 68) for Promag P sensor, remote version
16 Endress+Hauser
Promag 50/53 P
Process conditions
Medium temperature range The permissible medium temperature depends on the measuring tube liner:
• PTFE, -40° to +265°F (-40° to + 130°C) for 1/2” to 24” sensors
• PFA, -5° to +355°F (-20° to +180°C) for 1” to 8” sensors
Note: refer to graphs below for restrictions
HT
265
160
140
120
105
100
PTFE
PFA
TA [°F]
80
60
40
32
20
0
-10 20 60 100 140 180 220 260 300 340 380
-5 230 355
TF [°F]
HT
265
160
140
120
PTFE
PFA
TA [°F]
80
40
32
0
-10 20 60 100 140 180 220 260 300 340 380
-5 TF [°F] 355
Note that in the case of the remote version, the minimum conductivity is also influ-
enced by the length of the connecting cable. Refer to “Length of connecting cable”.
Endress+Hauser 17
Promag 50/53 P
Pressure tightness (liner) Size Measuring Resistance to partial vacuum of measuring tube lining
tube limit values for abs. pressure (psia) at various
lining fluid temperatures
75°F 175°F 212°F 265°F 300°F 355°F
1/2” PTFE 0 0 0 1.5 - -
1” PTFE/PFA 0/0 0/0 0/0 1.5 / 0 -/0 -/0
1-1/2” PTFE/PFA 0/0 0/0 0/0 1.5 / 0 -/0 -/0
2” PTFE/PFA 0/0 0/0 0/0 1.5 / 0 -/0 -/0
3” PTFE/PFA 0/0 * 0.6 / 0 1.9 / 0 -/0 -/0
4” PTFE/PFA 0/0 * 2.0 / 0 2.5 / 0 -/0 -/0
6” PTFE/PFA 2.0 / 0 * 3.5 / 0 5.6 / 0 -/0 -/0
8” PTFE/PFA 2.9 / 0 * 4.2 / 0 5.9 / 0 -/0 -/0
10” PTFE 4.8 * 5.8 7.7 - -
12” PTFE 5.8 * 7.3 9.1 - -
14” PTFE 6.8 * 8.7 10.6 - -
16” PTFE 7.8 * 9.7 11.6 - -
18” PTFE
20” PTFE No vacuum is permissible!
24” PTFE
Flow limit As a rule, the pipeline diameter determines the sensor diameter. With a known
flowrate and the help of the Endress+Hauser Flow Applicator, the best size can be
selected. Generally, a velocity range of 6 to 10 ft/sec should be selected.
A necessary increase in velocity can be achieved through a reduction of the sensor
diameter. The higher installation expense is normally balanced by the lower sensor
cost.
The flow velocity should also be determined by the fluid’s physical properties:
• v < 6 ft/sec, for abrasive mediums such as potter’s clay, lime milk, ore slurry, etc.
• v > 6 ft/sec, for fluids causing build-up such as wastewater sludge, etc.
Pressure loss No pressure loss if the sensor is installed in a pipe of the same nominal diameter.
Pressure losses occur in configurations incorporating adapters (reducers, expand-
ers). Contact your Endress+Hauser representative or Endress+Hauser for informa-
tion.
18 Endress+Hauser
Promag 50/53 P
Mechanical construction
Dimensions / mounting details Wall mounted housing
Weight: 13 lbs (6 kg)
NOTE: A mounting kit (for panel or post mounting) is available from
Endress+Hauser. Please refer to Accessories.
6.27"
Esc
(159.5)
- + E
9.84"
(250)
3.56"
(90.5)
4.01"
(102)
2.08"
(53)
3.74"
3.20" (95)
(81.5)
6.57"
(167)
13.9"
(355)
7.40"
3.94" 5.24" (188)
(100) (133)
Æ 0.34"
3.94"
(8.6 M8)
(100)
4.84"
(123)
Endress+Hauser 19
Promag 50/53 P
6.30" 6.30"
Esc
(160) (160)
Esc
- + E - + E
B
A
B
C A
K L
E
* Blind option dimension K L
Dimensions in inches (mm)
Nominal Weight *
Size L A B C K E
inches mm in mm in mm in mm in mm in mm in mm lb kg
1/2 15 7.87 200 13.4 341 10.1 257 3.31 84 4.72 120 3.70 94 14 6.5
1 25 7.87 200 13.4 341 10.1 257 3.31 84 4.72 120 3.70 94 16 7.3
1-1/2 40 7.87 200 13.4 341 10.1 257 3.31 84 4.72 120 3.70 94 20 9.4
2 50 7.87 200 13.4 341 10.1 257 3.31 84 4.72 120 3.70 94 23 10.6
3 80 7.87 200 15.4 391 11.1 282 4.29 109 7.08 180 3.70 94 31 14
4 100 9.84 250 15.4 391 11.1 282 4.29 109 7.08 180 3.70 94 35 16
6 150 11.8 300 18.6 472 12.7 322 5.90 150 10.2 260 5.51 140 56 25.5
8 200 13.8 350 20.7 527 13.6 347 7.08 180 12.7 324 6.14 156 99 45
10 250 17.7 450 22.7 577 14.6 372 8.07 205 15.7 400 6.14 156 165 75
12 300 19.7 500 24.7 627 15.6 397 9.05 230 18.1 460 6.53 166 242 110
14 350 21.6 550 29.1 738.5 17.9 456.5 11.1 282 22.2 564 10.9 276 385 175
16 400 23.6 600 31.1 790.5 18.9 482.5 12.1 308 24.2 616 10.9 276 451 205
18 450 25.6 650 33.1 840.5 19.9 507.5 13.1 333 26.2 666 11.5 292 561 255
20 500 25.6 650 35.1 891.5 20.9 533 14.1 358.5 28.2 717 11.5 292 627 285
24 600 30.7 780 39.2 995.5 23.0 585 16.2 410.5 32.3 821 15.8 402 891 405
* Weight data valid for 150 lb ANSI version. For high temperature version, add 3 lb (1.5 kg).
Note: The fitting length (L) is always the same, regardless of the pressure rating.
Esc
- + E
B1
4.33"
(110)
A1
Measurement A1, B1: refer to compact version dimension chart above, add 4.33”
(110 mm) to dimensions A and B for high temperature version.
20 Endress+Hauser
Promag 50/53 P
4.01"
(102)
B
A
B
C
A
E
L K
L K
Nominal Weight *
Size L A B C K E
inches mm in mm in mm in mm in mm in mm in mm lb kg
1/2 15 7.87 200 11.2 286 7.95 202 3.31 84 4.72 120 3.70 94 12 4.5
1 25 7.87 200 11.2 286 7.95 202 3.31 84 4.72 120 3.70 94 12 5.3
1-1/2 40 7.87 200 11.2 286 7.95 202 3.31 84 4.72 120 3.70 94 16 7.4
2 50 7.87 200 11.2 286 7.95 202 3.31 84 4.72 120 3.70 94 19 8.6
3 80 7.87 200 13.2 336 8.94 227 4.29 109 7.08 180 3.70 94 24 12
4 100 9.84 250 13.2 336 8.94 227 4.29 109 7.08 180 3.70 94 31 14
6 150 11.8 300 16.4 417 10.5 267 5.90 150 10.2 260 5.51 140 52 23.5
8 200 13.8 350 18.6 472 11.5 292 7.08 180 12.7 324 6.14 156 95 43
10 250 17.7 450 20.5 522 12.5 317 8.07 205 15.7 400 6.14 156 161 73
12 300 19.7 500 22.5 572 13.5 342 9.05 230 18.1 460 6.53 166 238 108
14 350 21.6 550 26.9 683.5 15.8 401.5 11.1 282 22.2 564 10.9 276 381 173
16 400 23.6 600 29.0 735.5 16.8 427.5 12.1 308 24.2 616 10.9 276 447 203
18 450 25.6 650 30.9 785.5 17.8 452.5 13.1 333 26.2 666 11.5 292 557 253
20 500 25.6 650 32.9 836.5 18.8 478 14.1 358.5 28.2 717 11.5 292 623 283
24 600 30.7 780 37.0 940.5 20.9 530 16.2 410.5 32.3 821 15.8 402 887 403
* Weight data valid for 150 lb ANSI version. For high temperature version, add 3 lb (1.5 kg).
B1
4.33"
(110)
A1
Measurement A1, B1: refer to compact version dimension chart above, add 4.33”
(110 mm) to dimensions A and B for high temperature version.
Endress+Hauser 21
Promag 50/53 P
ØB
∅ 0.26"
(6.5)
0.59"
(15)
ØA
0.39"
(10)
ØD
0.08"
(2)
Nominal A B D H
size 1)
in mm in mm in mm in mm
22 Endress+Hauser
Promag 50/53 P
870 60
725 50
580 40
Class 300
435 30
290 20
Class 150
145 10
0 0
-76 -40 -4 32 68 104 140 176 212 248 284 320 [°F]
-60 -40 -20 0 20 40 60 80 100 120 140 160 [°C]
870 60
725 50
Class 300
580 40
435 30
290 20
Class 150
145 10
0 0
-76 -40 -4 32 68 104 140 176 212 248 284 320 [°F]
-60 -40 -20 0 20 40 60 80 100 120 140 160 [°C]
[psi] [bar]
360 25
290 20
215 15
145 10
72 5
0 0
-76 -40 -4 32 68 104 140 176 212 248 284 320 [°F]
-60 -40 -20 0 20 40 60 80 100 120 140 160 [°C]
Endress+Hauser 23
Promag 50/53 P
Human interface
Display elements • Liquid crystal display, backlit, two lines (Promag 50) or four lines (Promag 53)
with 16 characters per line
• Custom configurations for presenting different measured value and status
variables
• Flow rate: 5 digits with sign and units
• Totalizer:
Promag 50: 2 totalizers (7-digit plus 7-digit overflow with sign and units)
Promag 53: 3 totalizers (7-digit plus 7-digit overflow with sign and units)
CE mark By attaching the CE mark, Endress+Hauser confirms that the instrument fulfills all the
requirements of the relevant EC directives.
Pressure equipment directive Flow meters with a nominal diameter smaller or equal to 1” (DN 25) are covered by
Art. 3(3) of the European directive 97/23/EG (Pressure Equipment Directive) and are
designed according to sound engineering practice. For larger nominal diameters,
optional approvals according to Cat. III are available when required (depends on the
fluid and process pressure).
PROFIBUS-DP/PA certification The flow device has successfully passed all the test procedures carried out and is
certified and registered by the PNO (PROFIBUS User Organization). The device thus
meets all the requirements of the following specifications:
• Certified to PROFIBUS-PA, profile version 3.0 (device certification number on
request)
• The device can also be operated with certified devices of other manufacturers
(interoperability)
FOUNDATION Fieldbus certification The flow device has successfully passed all the test procedures carried out and is
certified and registered by the Fieldbus FOUNDATION. The device thus meets all the
requirements of the following specifications:
• Certified to FOUNDATION Fieldbus Specification
• The device meets all the specifications of the FOUNDATION Fieldbus H1
• Interoperability Test Kit (ITK), revision status 4.0 (device certification number on
request)
• The device can also be operated with certified devices of other manufacturers
• Physical Layer Conformance Test of the Fieldbus FOUNDATION
24 Endress+Hauser
Promag 50/53 P
Ordering information
Promag 50 P Electromagnetic
Flowmeter
1 2 3 4 5 6 7 8 9 10 11 12 13
Promag 50 P -
Nominal Diameter 8 Protection type / Version
1 15 1/2” A NEMA 4X / compact, aluminum field housing
25 1” C NEMA 4X / remote wall mounted housing (only for approvals A or R)
40 1-1/2” G NEMA 4X / remote aluminum field housing for nonhazardous areas
50 2” K NEMA 6P / remote, wall mounted housing (only for approvals A or R),
80 3” not for liner “B” (PFA high temp.)
1H 4” N NEMA 6P / remote aluminum field housing, nonhazardous, not for
1F 6” use with liner “B” (PFA high temp.)
2H 8” 1 -40°F (-40°C) ambient temperature, NEMA 4X / compact aluminum
2F 10” housing for transmitter (only for approvals A or R)
3H 12” 3 -40°F (-40°C) ambient temperature, NEMA 4X / compact SS housing
3F 14” for transmitter (only for approvals A or R)
4H 16” 5 -40°F (-40°C) ambient temperature, NEMA 4X / remote wall mount
4F 18” housing for transmitter (only for approvals A or R)
5H 20” 9 Other
6H 24” 9 Cable for Remote Version
2 Liner 0 Without cable
A PFA liner, standard (available for 1” to 8” sensors) 1 15 foot coil and signal cable
B PFA liner, high temperature version (available for 1” to 8” sensors, not 2 30 foot coil and signal cable
for approval N) 5 Coil and signal cable, specify length in feet, maximum 650 feet,
E PTFE liner depending on conductivity of process material
3 Process Connection 9 Other
L Class 150 ANSI B16.5 carbon steel A105 flanges 10 Cable Entries
M Class 300 ANSI B16.5 carbon steel A105 flanges B 1/2” NPT
R Class 150 ANSI B16.5, 316L SS flanges L Fieldbus connector (Outputs / Inputs H and Approval A only)
S Class 300 ANSI B16.5, 316L SS flanges (not for 8” an X Sensor only, without cable (compact version only)
larger sensors) 9 Other
9 Other 11 Display / Power Supply / Operation
4 Electrodes / Material 7 85 to 260 VAC, without display, remote configuration only (not for
0 Measuring, reference and EPD electrodes / 316L SS wall mount or SS housing)
1 Measuring, reference and EPD electrodes / Alloy C-22 8 20 to 55 VAC / 16 to 62 VDC, without display, remote configuration
2 Measuring, reference and EPD electrodes / tantalum only (not for wall mount or SS housing)
3 Measuring electrodes only / Platinum/Rhodium 80/20% A 85 to 260 VAC, with display, push button operation (language: EN,
4 Measuring and reference electrodes only / Platinum/Rhodium 80/20% ES, FR, IT, NL, PT, DE)
5 Measuring, reference and EPD electrodes / Platinum/Rhodium 80/20% B 20 to 55 VAC / 16 to 62 VDC, with display, push button operation
G Measuring, reference and EPD electrodes, bullet nose / 316L SS (language: EN, ES, FR, IT, NL, PT, DE)
H Measuring, reference and EPD electrodes, bullet nose / Alloy C-22 X Sensor only (without transmitter, only available up to 8”)
9 Other 12 Software
5 Calibration A Standard
A 3-point calibration, 0.5% C ECC electrode cleaning circuit (only for approvals A or R)
B 3-point calibration, 0.2% X Sensor only
D SCS/A2LA 3-point 0.5% calibration (ISO/IEC 17025) with certificate 13 Outputs / Inputs
traceable according to ISO 9000 (specify flow range) W Current / HART
E SCS/A2LA 3-point 0.2% calibration (ISO/IEC 17025) with certificate A Current / HART, frequency
traceable according to ISO 9000 (specify flow range) D Current / HART, frequency, status, status input
9 Other H PROFIBUS-PA (only for approvals A or R)
6 Certificates S Current HART active, frequency passive; I.S. (only for approval N)
1 Standard, no certificate T Current HART passive, frequency passive; I.S. (only for approval N)
2 3.1B material certificate for pipe and flanges X Sensor only
3 2.3 pressure test certificate (1.5 x PN, 3 minutes) for sensors
up to 12” only
4 3.1B material and 2.3 pressure test certificate for sensors
up to 12” only
5 CRN approval for ANSI flanges only
8 CRN approval, 3.1B material certificate and 2.3 pressure test
certificate (1.5 x PN) for ANSI flanges only
7 Approvals
A For use in nonhazardous areas
N FM explosion proof Class I, Div. 1 / CSA Class I, Div. 1 (only for
aluminum field housing, compact version, not for 14” and larger
sensors)
R FM non-incendive Class I, Div. 2 / CSA Class I, Div. 2
Endress+Hauser 25
Promag 50/53 P
Promag 53 P Electromagnetic
Flowmeter
1 2 3 4 5 6 7 8 9 10 11 12 13
Promag 53 P -
Nominal Diameter 9 Cable for Remote Version
1 15 1/2” 0 Without cable
25 1” 1 15 foot coil and signal cable
40 1-1/2” 2 30 foot coil and signal cable
50 2” 5 Coil and signal cable, specify length in feet, maximum 650 feet,
80 3” depending on conductivity of process material
1H 4” 9 Other
1F 6” 10 Cable Entries
2H 8” B 1/2” NPT
2F 10” L Fieldbus connector (Outputs / Inputs H or K, and Approval A only)
3H 12” X Sensor only, without cable (compact version only)
3F 14” 9 Other
4H 16” 11 Display / Power Supply / Operation
4F 18” 7 85 to 260 VAC, without display, remote configuration only (not for
5H 20” wall mount or SS housing)
6H 24” 8 20 to 55 VAC / 16 to 62 VDC, without display, remote configuration
2 Liner only (not for wall mount or SS housing)
A PFA liner, standard (available for 1” to 8” sensors) A 85 to 260 VAC, with display, touch control operation (language: EN,
B PFA liner, high temperature version (available for 1” to 8” sensors, not ES, FR, IT, NL, PT, DE)
for approval N) B 20 to 55 VAC / 16 to 62 VDC, with display, touch control operation
E PTFE liner (language: EN, ES, FR, IT, NL, PT, DE)
3 Process Connection X Sensor only (without transmitter, only available up to 8”)
L Class 150 ANSI B16.5 carbon steel A105 flanges 9 Other
M Class 300 ANSI B16.5 carbon steel A105 flanges 12 Software
R Class 150 ANSI B16.5, 316L SS flanges A Standard
S Class 300 ANSI B16.5, 316L SS flanges (not for 8” and larger C ECC electrode cleaning circuit (only for Approvals A or R)
sensors) J Batching
9 Other L Batching and ECC electrode cleaning (only for Approvals A or R)
4 Electrodes / Material X Sensor only
0 Measuring, reference and EPD electrodes / 316L SS 13 Outputs / Inputs
1 Measuring, reference and EPD electrodes / Alloy C-22 A Current / HART, frequency
2 Measuring, reference and EPD electrodes / tantalum B Current / HART, frequency and 2 relays
3 Measuring electrodes only / Platinum/Rhodium 80/20% C Current / HART, frequency, 2 relays (flexible module)
4 Measuring and reference electrodes only / Platinum/Rhodium 80/20% D Current / HART, frequency, relay, and status input (flexible module)
5 Measuring, reference and EPD electrodes / Platinum/Rhodium 80/20% F PROFIBUS PA IS (only for Approval N)
G Measuring, reference and EPD electrodes, bullet nose / 316L SS G FOUNDATION fieldbus IS (only for Approval N)
H Measuring, reference and EPD electrodes, bullet nose / Alloy C-22 H PROFIBUS-PA (only for Approvals A or R)
9 Other J PROFIBUS-DP
5 Calibration K FOUNDATION fieldbus
B 3-point calibration, 0.2% L Current / HART, 2 relays and status input/output (flexible module)
E SCS/A2LA 3-point 0.2% calibration (ISO/IEC 17025) with certificate M Current / HART, 2 frequency, status input/output (flexible module)
traceable according to ISO 9000 (specify flow range) S Current / HART (active), frequency; IS(only for Approval N)
9 Other T Current / HART (passive), frequency; IS (only for Approval N)
6 Certificates 2 Current / HART, frequency, current, relay (flexible module)
1 Standard, no certificate 4 Current / HART, frequency, current input, relay (flexible module)
2 3.1B material certificate for pipe and flanges 5 Current / HART, frequency, current input, status input (flexible module)
3 2.3 pressure test certificate (1.5 x PN, 3 minutes) for sensors X Sensor only
up to 12” only
4 3.1B material and 2.3 pressure test certificate for sensors
up to 12” only
5 CRN approval for ANSI flanges only
8 CRN approval, 3.1B material certificate and 2.3 pressure test
certificate (1.5 x PN) for ANSI flanges only
7 Approvals
A For use in non-hazardous areas
N FM explosion proof Class I, Div. 1 / CSA Class I, Div. 1 (only for
aluminum field housing, compact version, not for 14” and larger
sensors)
R FM non-incendive Class I, Div. 2 / CSA Class I, Div. 2
8 Protection type / Version
A NEMA 4X / compact, aluminum field housing
C NEMA 4X / remote wall mounted housing (only for approvals A or R)
K NEMA 6P remote, wall mounted housing (only for approvals A or R),
not for liner “B” (PFA high temp.)
N NEMA 6P / remote aluminum field housing, nonhazardous, not for
use with liner “B” (PFA high temp.)
1 -40°F (-40°C) ambient temperature, NEMA 4X / compact aluminum
housing for transmitter (only for approvals A or R)
3 -40°F (-40°C) ambient temperature, NEMA 4X / compact SS housing
for transmitter (only for approvals A or R)
5 -40°F (-40°C) ambient temperature, NEMA 4X / remote wall mount
housing for transmitter (only for approvals A or R)
9 Other
26 Endress+Hauser
Promag 50/53 P
Accessories
Various accessories, which can be ordered from Endress+Hauser, are available for
the transmitter and the sensor. Contact your local representative or Endress+Hauser
for detailed information.
DK8WM -
1 Transmitter type
A Panel/wall/pipe mount (2” pipe) hardware
B Wall/pipe mount for remote field housing
+0.02
8.27" -0.02
+0.5
(210 -0.5
) Ø 2"
(50)
+0.02
-0.02
9.65"+0.5
(245 -0.5 )
6.10"
(155)
4.33"
(110)
DK8WM - A
Mounting set for wall / panel enclosure, 3/4” to 2-3/4”
pipe
Includes:
4 mounting plates
4 threaded rods, M6 x 150 mm
4 washers
4 nuts, M6
Washer
Bolt
DK8WM - B
Mounting set for wall / pipe mount remote field enclosure, 3/4” to 3” pipe
Includes:
Mounting plate
4 bolts, M8 x 150 mm
4 washers
4 nuts, M8
Endress+Hauser 27
Supplemental documentation Promag System Information (SI 028D/24/ae)
Promag 50/53 W Technical Information (TI 046D/24/ae)
Promag 50/53 H Technical Information (TI 048D/24/ae)
Promag 50 Operating Instructions (BA 046D/06/en and BA 049D/06/en)
Promag 53 Operating Instructions (BA 047D/06/en and BA 048D/06/en)
Hazardous Area Approvals Documentation (FM, CSA, ATEX, etc.)
Endress+Hauser
The Power of Know How
TI 047D/24/ae/08.04
© 2004 Endress+Hauser, Inc.