Technical

Information Electromagnetic
TI 047D/24/ae
Flow Measuring System
PROline promag 50/53 P
Flow measurement in
chemical or process applications

Features and benefits • Interfaces for integration into all


• Nominal diameters 1/2” to 24” major process-control systems:
• PTFE or PFA lining – HART® interface as standard
• PFA for high-temperature applica- – Promag 50: PROFIBUS-PA
tions up to +355°F (+180°C) – Promag 53: PROFIBUS-PA/-DP,
• Fitting lengths to ISO and DVGW FOUNDATION fieldbus
• High measuring accuracy for • Explosion proof approval for
improved process control: installation in Division 1 (FM, CSA,
– Promag 50: ± 0.5% ATEX, etc.)
(option: ± 0.2%)
– Promag 53: ± 0.2% Application
• Robust field housing, NEMA 4X All fluids with a minimum conductivity of
(IP 67) ≥ 5 µS/cm can be measured:
• NEMA 4X (IP 67) wall-mount housing • acids and alkalis
for straight-forward installation of the • paints, lacquers
remote version • pastes, mashes
• Promag 53 with Touch Control: • water, wastewater, etc.
operation without opening the
housing - also for explosion proof A minimum conductivity of ≥ 20 µS/cm is
applications required for measuring demineralized
• Expandable software packages: water.
– for batching applications Liner specific applications
– electrode cleaning • PTFE lining for standard applications
– pulsating flow in chemical and process industries
• Quick setup menus for straight- • PFA lining for all applications in
forward commissioning in the field chemical and process industries;
especially for high process
temperatures and applications with
thermal shocks

The Power of Know How


Promag 50/53 P

Function and system design


Measuring principle Faraday’s law of induction states that a voltage is induced in a conductor moving in a
magnetic field.
In electromagnetic measuring, the flowing medium corresponds to the moving
conductor. The induced voltage is proportional to the flow velocity and is detected by
two measuring electrodes and transmitted to the amplifier. Flow volume is computed
on the basis of the pipe’s diameter. The constant magnetic field is generated by a
switched direct current of alternating polarity.

Ue = B · L · v
Q =A·v

Ue = induced voltage
B = magnetic induction (magnetic field)
L = electrode gap
v = flow velocity
Q = volume flow
A = pipe cross-section
I = current strength

Measuring system The measuring system consists of a transmitter and a sensor.


Two versions are available:
• Compact version: transmitter and sensor form a single mechanical unit.
• Remote version: transmitter and sensor are installed separately.

Transmitter:
• Promag 50 (user interface with push buttons for operation, two-line display)
• Promag 53 (“Touch Control” without opening the housing, four-line display)

Sensor:
• Promag P, 1/2” to 24” (DN 15 to 600)

2 Endress+Hauser
Promag 50/53 P

Input
Measured variable Flow rate (proportional to induced voltage)

Measuring range Typically v = 0.033 to 33 ft/s (0.01 to 10 m/s) with the specified measuring accuracy

Operable flow range Over 1000 : 1

Input signal Status input (auxiliary input):


U = 3 to 30 V DC, Ri = 5 kΩ, galvanically isolated.
Configurable for: totalizer(s) reset, measured-value suppression, error-message
reset
Current input (for Promag 53 only):
Active/passive selectable, galvanically isolated, full scale value selectable,
resolution: 3 µA
Temperature coefficient: typically 0.003% o.r./°F (0.005% o.r./°C) (o.r = of reading)
Active: 4 to 20 mA, Ri ≤ 150 Ω, Uout = 24 VDC, short-circuit proof
Passive: 0/4 to 20 mA, Ri ≤ 150 Ω, Umax = 30 VDC

Output

Output signal Promag 50


Current output:
Active/passive selectable, galvanically isolated, time constant selectable
(0.01 to 100 s), full scale value selectable, temperature coefficient: typically
0.003% o.r./°F (0.005% o.r./°C); resolution: 0.5 µA
• active: 0/4 to 20 mA, RL < 700 Ω (HART®: RL ≥ 250 Ω)
• passive: 4 to 20 mA, max. 30 V DC, Ri ≤ 150 Ω

Pulse/frequency output:
Passive, open collector, 30 V DC, 250 mA, galvanically isolated.
Frequency output: full scale frequency 2 to 1000 Hz (fmax = 1250 Hz), on/off ratio 1:1,
pulse width max. 10 s
Pulse output: pulse value and pulse polarity selectable, max. pulse width
configurable (0.5 ms to 2 s), maximum pulse frequency selectable

PROFIBUS-PA interface
• PROFIBUS-PA in accordance with EN 50170 Volume 2, IEC 61158-2 (MPB),
galvanically isolated
• Current consumption: 11 mA
• Permissible supply voltage: 9 to 32 V
• FDE (Fault Disconnection Electronic): 0 mA
• Data transmission rate, supported bauderate: 31.25 kBit/s
• Signal encoding: Manchester II
• Function blocks: 1 x Analog Input, 1 x Totalizer
• Output data: Volume flow, Totalizer
• Input data: Positive zero return (ON/OFF), Control totalizer, Value for local display
• Bus address adjustable via DIP switches at the measuring device

Endress+Hauser 3
Promag 50/53 P

Promag 53
Current output:
Active/passive selectable, galvanically isolated, time constant selectable
(0.01 to 100 s), full scale value selectable, temperature coefficient: typically
0.003% o.r./°F (0.005% o.r./°C); resolution: 0.5 µA
• active: 0/4 to 20 mA, RL < 700 Ω (HART®: RL ≥ 250 Ω)
• passive: 4 to 20 mA, max. 30 V DC, Ri ≤ 150 Ω

Pulse/frequency output:
Active/passive selectable, galvanically isolated (I.S. version: passive only)
• active: 24 V DC, 25 mA (max. 250 mA during 20 ms), RL > 100 Ω
• passive: open collector, 30 V DC, 250 mA
Frequency output: full scale frequency 2 to 10,000 Hz (fmax = 12,500 Hz),
on/off ratio 1:1, pulse width max. 10 s
Pulse output: pulse value and pulse polarity selectable, pulse width configurable
(0.05 ms to 2 s), on/off ratio is 1:1 as of a frequency of 1 / (2 x pulse width)

PROFIBUS-DP interface:
• PROFIBUS-DP/PA in accordance with EN 50170 Volume 2, IEC 61158-2 (MPB),
galvanically isolated
• Data transmission rate, supported baudrate: 9.6 kBaud to 12 MBaud
• Automatic data transmission rate recognition
• Signal encoding: NRZ-Code
• Function blocks: 2 x Analog Input, 3 x Totalizer
• Output data: Volume flow, Corrected volume flow, Totalizer 1 to 3
• Input data: Positive zero return (ON/OFF), Totalizer control, Value for local display
• Bus address adjustable via DIP switches at the measuring device

PROFIBUS-PA interface:
• PROFIBUS-PA in accordance with EN 50170 Volume 2, IEC 61158-2 (MPB),
galvanically isolated
• Current consumption: 11 mA
• Permissible supply voltage: 9 to 32 V
• Data transmission rate, supported baudrate: 31.25 kBit/s
• Error current FDE (Fault Disconnection Electronic): 0 mA
• Signal encoding: Manchester II
• Function blocks: 2 x Analog Input, 3 x Totalizer
• Output data: Volume flow, Corrected volume flow, Totalizer 1 to 3
• Input data: Positive zero return (ON/OFF), Totalizer control, Value for local display
• Bus address adjustable via DIP switches at the measuring device

FOUNDATION Fieldbus interface:


• FOUNDATION Fieldbus H1, IEC 61158-2 (MBP), galvanically isolated
• Current consumption: 12 mA
• Permissible supply voltage: 9 to 32 V
• Error current FDE (Fault Disconnection Electronic): 0 mA
• Data transmission rate, supported buadrate: 31.25 kBit/s
• Signal encoding: Manchester II
• Function blocks: 5 x Analog Input, 1 x Discrete Output, 1 x PID
• Output data: Volume flow, Corrected volume flow, Totalizer 1 to 3
• Input data: Positive zero return (ON/OFF), Reset totalizer
• Link Master (LM) functionality is supported

4 Endress+Hauser
Promag 50/53 P

Signal on alarm • Current output → failure response selectable


• Pulse/frequency output → failure response selectable
• Status output (Promag 50) → non-conductive by fault or power supply failure
• Relay output (Promag 53) → de-energized by fault or power supply failure

Load See “Output signal”

Switching output Status output (Promag 50):


Open collector, max. 30 V DC / 250 mA, galvanically isolated
Configurable for: error messages, Empty Pipe Detection (EPD), flow direction, limit
values

Relay output (Promag 53):


Normally closed (NC or break) or normally open (NO or make) contacts available
(default: relay 1 = NO, relay 2 = NC),
max. 30 V / 0.5 A AC; 60 V / 0.1 A DC, galvanically isolated.
Configurable for: error messages, Empty Pipe Detection (EPD), flow direction, limit
values, batching contacts

Low flow cutoff Switch points for low flow cutoff are selectable

Galvanic isolation All circuits for inputs, outputs, and power supply are galvanically isolated from each
other.

Endress+Hauser 5
Promag 50/53 P

Power supply
Electrical connection
−27
measuring unit +26
f b
b e −25
+24
−23 d
+22
−21
+20
a

N (L-) 2 c
g L1 (L+) 1
a

N (L-)
L1 (L+)

e
+ − + − + − + −
1 2 20 21 22 23 24 25 26 27

a b a c b d

Top: field housing


Bottom: wall-mount housing

a Cable for power supply: 85 to 260 VAC, 20 to 55 VAC, or 16 to 62 VDC; power consumption: 15 VA / 15 W
Terminal No. 1: L1 fo AC, L+ for DC
Terminal No. 2: N for AC, L- for DC
b Signal cable: terminals No. 20 to 27 ´® see Page 8
c Grounding terminal for protective conductor
d Grounding terminal for signal-cable shield
e Service connection for connecting service interface FXA 193 (FieldCheck, ToF Tool-Field Tool Package)
f Cover for wiring compartment
g Cover securing clamp

6 Endress+Hauser
Promag 50/53 P

Electrical connection f
Measuring unit (bus communication)
DP (A)/PA (–)/FF (–) 27
DP (B)/PA (+)/FF (+) 26
(DGND) 25 b
b (+5 V) 24
23 d
g e
22
21
20 g
a

N (L-) 2 c
h L1 (L+) 1
a

(+5 V)
N (L–) (DGND)
L1 (L+) DP (B)/PA (+)/FF (+)
DP (A)/PA (–)/FF (–)

e
1 2 20 21 22 23 24 25 26 27

Esc
- + E

f
a g b a c g d b

Top: Field housing


Bottom: Wall-mount housing

a Cable for power supply: 85 to 260 VAC, 20 to 55 VAC, or 16 to 62 VDC; power consumption:15 VA / 15 W
Terminal No. 1: L1 fo AC, L+ for DC
Terminal No. 2: N for AC, L- for DC
b Fieldbus cable:
Terminal No. 26: DP (B) / PA (+) / FF (+) (with polarity protection)
Termianl No. 27: DP (A) / PA (-) / FF (-) (with reverse polarity protection)
DP (A) = RxD/TxD-N; DP (B) = RxD/TxD-P
c Grounding terminal for protective conductor
d Grounding terminal for signal-cable shield
e Service connector for connecting service interface FXA 193 (FieldCheck, ToF Tool-Field Tool Package)
f Cover for connection compartment
g Cable for external termination (PROFIBUS only)
h Cover securing clamp

Endress+Hauser 7
Promag 50/53 P

Terminal assignment Promag 50

Terminal No. (inputs / outputs)

Order variant 20 (+) / 21 (-) 22 (+) / 23 (-) 24 (+) / 25 (-) 26 (+) / 27 (-)

50***-***********W - - - Current output


HART®
50***-***********A - - Frequency output Current output
HART®
50***-***********D Status input Status output Frequency output Current output
HART®
50***-***********S - - Frequency output Current output
IS, passive HART®
50***-***********T - - Frequency output Current output
Intrinsically Safe HART®

Ground connection, power supply → refer to illustration on top of page 6

Terminal assignment Promag 53

The inputs and outputs on the communication board can be either permanently
assigned or variable, depending on the version ordered (see table). Replacement
for modules which are defective or which have to be replaced can be ordered as
accessories.

Terminal No. (inputs / outputs)

Order variant 20 (+) / 21 (-) 22 (+) / 23 (-) 24 (+) / 25 (-) 26 (+) / 27 (-)

Fixed communication boards (fixed assignment)

53***-***********A - - Frequency output Current output


HART®
53***-***********B Relay output 2 Relay output 1 Frequency output Current output
HART®
53***-***********F - - - Profibus-PA
IS
53***-***********G - - - Foundation
Fieldbus IS
53***-***********H - - - Profibus-PA
53***-***********J - - - Profibus-DP
53***-***********K - - - Foundation
Fieldbus
53***-***********S - - Frequency output Current output
Intrinsically safe IS, active
HART®
53***-***********T - - Frequency output Current output
Intrinsically safe IS, passive
HART®

Flexible communication boards

53***-***********C Relay output 2 Relay output 1 Frequency output Current output


HART®
53***-***********D Status input Relay output Frequency output Current output
HART®
53***-***********L Status input Relay output 2 Relay output 1 Current output
HART®
53***-***********M Status input Frequency output 2 Frequency output 1 Current output
HART®
53***-***********2 Relay output Current output 2 Frequency output Current output
1 HART®
53***-***********4 Current input Relay output Frequency output Current output
HART®
53***-***********5 Status input Current input Frequency output Current output
HART®

Ground connection, power supply → refer to illustration on top of page 6

8 Endress+Hauser
Promag 50/53 P

Electrical connection
Electrode Circuit
remote version
Meas. signal Pipe EPD

GND
S1 E1 E2 S2 E S Coil circuit
Esc
- + E

6 5 7 8 4 37 36 42 41

yellow
brown

green
white

2
1
n.c. n.c. n.c.

5 7 4 37 42 41

E1 E2 E

GND
n.c. = cable shields to be isolated and not to be connected

Cable entry Power-supply and signal cables (inputs/outputs):


• 1/2" NPT

Connecting cable for remote version:


• 1/2" NPT

Cable specifications Coil cable:


remote version • 2 x 18 AWG (0.75 mm2) PVC cable with common, braided copper shield,
approximately 0.28” diameter (Ø approx. 7 mm)
• Conductor resistance: ≤ 0.011 Ω/ft (37 Ω/km)
• Capacitance: core/core, shield grounded: ≤ 36 pF/ft (120 pF/m)
• Permanent operating temperature: -5° to +180°F (–20 to +80°C)
• Cable cross-section: maximum 16 AWG (2.5 mm2)

Signal cable:
• 3 x 20 AWG (0.38 mm2) PVC cable with common, braided copper shield,
approximately 0.28” diameter (Ø approx. 7 mm) and individually shielded cores.
• With Empty Pipe Detection (EPD): 4 x 20 AWG (0.38 mm2) PVC cable with
common, braided copper shield, approximately 0.28” diameter (Ø approx. 7 mm)
and individually shielded cores.
• Conductor resistance: ≤ 0.015 Ω/ft (50 Ω/km)
• Capacitance: core/shield: ≤ 128 pF/ft (420 pF/m)
• Permanent operating temperature: -5° to +180°F (–20 to +80 °C)
• Cable cross-section: maximum 16 AWG (2.5 mm2)

1
2
3
4
5
6
7
a b

a = signal cable, b = coil current cable (maximum 16 AWG (2.5 mm2)


1 = Core 5 = Core strengthening
2 = Core insulation 6 = Cable shield
3 = Core shield 7 = Outer jacket
4 = Core jacket

Endress+Hauser 9
Promag 50/53 P

Endress+Hauser optionally supplies reinforced connecting cables with an additional


metal braid. We recommend such cables for the following cases:
• Underground laid cables
• Danger of rodent attack
• Device used with NEMA 6P (IP 68) ingress protection

Operation in areas of severe electrical interference:


The measuring device complies with the general safety requirements in accordance
with EN 61010, the EMC requirements of EN 61326, and NAMUR recommendation
NE 21.

Caution:
Grounding is by means of the ground terminals provided for that purpose inside the
connection housing. Keep the stripped and twisted lengths of cable shield to the
terminals as short as possible.

Supply voltage 85 to 260 VAC, 45 to 65 Hz


20 to 55 VAC, 45 to 65 Hz
16 to 62 V DC

PROFIBUS-PA and FOUNDATION Fieldbus


Nonhazardous: 9 to 32 VDC
Intrinsically safe: 9 to 24 VDC
Explosion proof: 9 to 32 VDC

Power consumption AC: <15 VA (including sensor)


DC: <15 W (including sensor)
Switch-on current:
• maximum 13.5 A (< 50 ms) at 24 V DC
• maximum 3 A (< 5 ms) at 260 V AC

Power supply failure Lasting minimum 1 power cycle:


• EEPROM or T-DAT™ (Promag 53 only) retain the measuring-system data in the
event of a power supply failure
• S-DAT™ = exchangeable data storage chip which stores the data of the sensor:
nominal diameter, serial number, calibration factor, zero point, etc.

Potential equalization Standard case


Perfect measurement is only ensured when the medium and the sensor have the
same electrical potential. Most Promag sensors have a standard installed reference
electrode which guarantees the required potential matching. This usually means that
additional potential matching measures are unnecessary.

NOTE: For installation in metal pipes, it is advisable to connect the ground terminal
of the transmitter housing to the piping.

Caution:
For sensors without reference electrodes or without metal process terminals, carry
out potential matching as per the instructions for special cases described below.
These special measures are particularly important when standard grounding
practice cannot be ensured or extremely strong matching currents are expected.

10 Endress+Hauser
Promag 50/53 P

Metallic, ungrounded pipes


In order to avoid errors in measurement, use grounding cables to connect each
sensor flange to its corresponding pipe flange and ground the flanges. Connect the
transmitter or sensor connection housing, as applicable, to ground by means of the
grounding terminal provided for that purpose.

Note:
The grounding cable for flange-to-flange connections can be ordered separately as
an accessory from Endress+Hauser.
• ≤ 12”; the grounding cable is in direct connection with the conductive flange
coating and is secured by the flange screws.
• ≥ 14”, the grounding cable connects directly to the metal transport bracket.

9 AWG Cu

Plastic pipes and lined pipes


Normally, potential is matched using the reference electrodes in the measuring tube.
However, in exceptional cases it is possible that, due to the grounding plan of a
system, large matching currents flow over the reference electrodes. This can lead to
destruction of the sensor, e.g. through electrochemical decomposition of the elec-
trodes. In such cases, e.g. for fiberglass or PVC piping, it is recommended that you
use additional ground disks for potential matching.
When using ground disks, not the following points:
• Ground disks (1/2” to 12”) can be ordered from Endress+Hauser as an accessory
• Ground disks (including seals) increase the installation length. Refer to the
disk dimensions on page 22.

Caution:
Risk of damage by electrochemical corrosion. Note the electrochemical insulation
rating if the grounding rings and measuring electrodes are made of different materi-
als.
Observe the plant internal grounding guidelines.

9 AWG Cu

Endress+Hauser 11
Promag 50/53 P

Pipes with cathodic protection


If the medium cannot be grounded for process-related reasons, the measuring
device must be installed in such a way as to be potential-free:
• When installing the measuring device, make sure that there is an electrical
connection between the two piping runs, 9 AWG copper wire (6 mm2).
• Make sure that the installation materials do not establish a conductive connection
to the measuring device and that the installation materials withstand the
tightening torques applied when the threaded fasteners are tightened.
• Also comply with the regulations applicable to potential-free installation.

b b
a

9 AWG Cu

a = isolating transformer, b = electrically insulated

Performance characteristics
Reference conditions To DIN 19200 and VDI/VDE 2641:
• Medium temperature: +28 °C ± 2 K
• Ambient temperature: +22 °C ± 2 K
• Warm-up time: 30 minutes

Installation:
• Inlet run >10 x Dia
• Outlet run > 5 x Dia
• Sensor and transmitter grounded.
• Sensor centered relative to the pipe.

Measured error Promag 50:


Pulse output: ± 0.5% o.r. ± 1 mm/s (o.r. = of reading)
Current output: plus typically ± 5 µA

Promag 53:
Pulse output: ± 0.2% o.r. ± 2 mm/s (o.r. = of reading)
Current output: plus typically ± 5 µA

Supply-voltage fluctuations have no effect within the specified range.

[%]
2.5

2.0
0.5 %
1.5
0.2 %
1.0

0.5

0
0 3 6 13 20 26 33
(1) (2) (4) (6) (8) (10)
v ft/s (m/s)

Measured error in [%] of reading

Repeatability ± 0.1% o.r. ± 0.5 mm/s (o.r. = of reading)

12 Endress+Hauser
Promag 50/53 P

Operating conditions
Installation conditions
Installation instructions Mounting location
Correct measuring is possible only if the pipe is full. Avoid the following locations:
• Highest point in the run. Risk of air accumulating
• Directly upstream from an open pipe outlet in a down pipe

h2xD

Installation of pumps
Do not install the sensor on the inlet side of a pump. This precaution is to avoid low
pressure and the consequent risk of damage to the lining of the measuring tube.
It might be necessary to install pulse dampers in systems incorporating reciprocat-
ing, diaphragm or peristaltic pumps. The sensor is shock and vibration resistant for
acceleration up to 2 g in accordance with IEC 68-2-6.

Partially filled pipes


Partially filled pipes with gradients necessitate a drain-type configuration. The Empty
Pipe Detection (EPD) function offers additional protection by detecting empty or
partially filled pipes.

Caution:
To avoid the risk of solids accumulating, do not install the sensor at the lowest point
in the drain. It is advisable to install a cleaning valve.

2xD

5xD

Endress+Hauser 13
Promag 50/53 P

Vertical pipes
Install a siphon (b) or a vent valve (a) downstream of the sensor in vertical pipes
longer than 15 feet (5 meters). This precaution is to avoid low pressure and the
consequent risk of damage to the lining of the measuring tube. These measures also
prevent the system losing prime, which could cause air inclusions.
a

> 15 ft
(5 m)

a = vent valve, b = siphon

Orientation
An optimum orientation helps avoid gas and air accumulations and deposits in the
measuring tube. Promag, nevertheless, supplies a range of options and accessories
for correct measuring of problematic mediums:
• Electrode Cleaning Circuitry (ECC) to remove electrically conductive deposits in
the measuring tube, e.g. in accretive mediums
• Empty Pipe Detection (EPD) for recognition of partially filled measuring tubes, or
for degassing mediums or for applications with fluctuating process pressure.

Vertical orientation:
This orientation is ideal for self-emptying piping systems and for use in conjunction
with Empty Pipe Detection.

Horizontal orientation:
The measuring electrode-plane should be horizontal. This prevents brief insulation
of the two electrodes by entrained air bubbles.

Caution:
Empty Pipe Detection functions correctly only when the measuring device is installed
horizontally and the transmitter housing is facing upward. Otherwise, there is no
guarantee that Empty Pipe Detection will respond if the measuring tube is only
partially filled or empty.

A
1

2 2

1 = EPD electrode (Empty Pipe Detection)


2 = Measuring electrodes (signal detection)
3 = Reference electrode (potential equalization)

14 Endress+Hauser
Promag 50/53 P

Vibrations
Secure the piping and the sensor if vibration is severe.

Caution:
It is advisable to install the sensor and transmitter separately if vibration is exces-
sively severe. The sensor is shock and vibration resistant for acceleration up to 2 g
in accordance with IEC 68-2-6.

> 30 feet
(> 10 m)

Foundations, supports
If the nominal diameter is 14” or larger, mount the sensor on a foundation of ad-
equate loadbearing strength.

Caution:
Do not allow the sensor casing to take the weight of the sensor. This would buckle
the casing and damage the internal magnetic coils.

Inlet and outlet runs If possible, install the sensor well clear of fittings such as valves, T-pieces, elbows,
etc. Compliance with the following requirements for the inlet and outlet runs is
necessary in order to ensure measuring accuracy:
• Inlet run ≥ 5 x pipe diameter
• Outlet run ≥ 2 x pipe diameter
5xD 2xD

Adapters With the help of the appropriate adapters (reducers and expanders) the sensor can
be mounted in a pipeline of a larger diameter. For slowly flowing fluids, the resulting
higher velocity increases the measuring accuracy.

max. 4°

d D

Endress+Hauser 15
Promag 50/53 P

Length of connecting cable Permissible cable length Lmax depends on the conductivity of the medium. A
minimum conductivity of 20 µS/cm is required for measuring demineralized water.

[S/cm]
Esc
- + E

200

L max 100

5
ft (m)
33 330 650
(10) (100) (200)
L max

Gray shaded area = permissible range for medium conductivity


Lmax = length of connecting cable in feet (m)
Medium conductivity in µS/cm

In order to ensure measuring accuracy, comply with the following instructions when
installing the remote version:
• Secure the cable run or route the cable in a conduit. Movement of the cable
can falsify the measuring signal, particularly if the conductivity of the medium
is low.
• Route the cable well clear of electrical machines and switching elements.
• Ensure ground equalization between the sensor and transmitter, if necessary.

Environment

Ambient temperature Standard: -5° to +140°F (-20° to +60°C) for sensor and transmitter
Optional: -40° to +140°F (-40° to +60°C)

Note the following points:


• Install the device at a shady location. Avoid direct sunlight, particularly in
warm climatic regions.
• If both medium and ambient temperatures are high, specify the remote
transmitter (refer to the “Medium temperature” ranges).
• At ambient temperatures below -5°F (-20°C), the readability of the display may
be impaired.

Storage temperature 15° to + 120°F (-10° to +50°C), 70°F (20°C) is the preferable storage temperature.
• The measuring device must be protected against direct sunlight during storage
in order to avoid unacceptably high surface temperatures.
• Choose a storage location where moisture does not collect in the measuring
device. This helps prevent mold, fungus and bacteria from damaging the liner.
• Do not remove the protective plates or caps on the process connections until the
device is ready to install.

Degree of protection • Standard: NEMA 4X (IP 67) for transmitter and sensor
• Optional: NEMA 6P (IP 68) for Promag P sensor, remote version

Shock and vibration resistance Acceleration up to 2 g in accordance with IEC 68-2-6


(high-temperature version: no data available)

Electromagnetic compatibility (EMC) To EN 61326 and NAMUR recommendation NE 21

16 Endress+Hauser
Promag 50/53 P

Process conditions

Medium temperature range The permissible medium temperature depends on the measuring tube liner:
• PTFE, -40° to +265°F (-40° to + 130°C) for 1/2” to 24” sensors
• PFA, -5° to +355°F (-20° to +180°C) for 1” to 8” sensors
Note: refer to graphs below for restrictions

Compact version (PFA and PTFE lining):


TA = ambient temperature, TF = fluid temperature
HT = High Temperature version with insulation

HT

265

160
140
120
105
100

PTFE

PFA
TA [°F]
80
60
40
32
20
0
-10 20 60 100 140 180 220 260 300 340 380
-5 230 355
TF [°F]

Remote version (PFA and PTFE lining):


TA = ambient temperature, TF = fluid temperature
HT = High Temperature version with insulation

HT
265

160
140

120
PTFE

PFA
TA [°F]

80

40
32

0
-10 20 60 100 140 180 220 260 300 340 380
-5 TF [°F] 355

Conductivity Minimum conductivity:


≥ 5µS/cm → for fluids generally
≥ 20µS/cm → for demineralized water

Note that in the case of the remote version, the minimum conductivity is also influ-
enced by the length of the connecting cable. Refer to “Length of connecting cable”.

Medium pressure range ANSI B 16.5:


(nominal pressure) Class 150, 1/2” to 24”
Class 300, 1/2” to 6”

Endress+Hauser 17
Promag 50/53 P

Pressure tightness (liner) Size Measuring Resistance to partial vacuum of measuring tube lining
tube limit values for abs. pressure (psia) at various
lining fluid temperatures
75°F 175°F 212°F 265°F 300°F 355°F
1/2” PTFE 0 0 0 1.5 - -
1” PTFE/PFA 0/0 0/0 0/0 1.5 / 0 -/0 -/0
1-1/2” PTFE/PFA 0/0 0/0 0/0 1.5 / 0 -/0 -/0
2” PTFE/PFA 0/0 0/0 0/0 1.5 / 0 -/0 -/0
3” PTFE/PFA 0/0 * 0.6 / 0 1.9 / 0 -/0 -/0
4” PTFE/PFA 0/0 * 2.0 / 0 2.5 / 0 -/0 -/0
6” PTFE/PFA 2.0 / 0 * 3.5 / 0 5.6 / 0 -/0 -/0
8” PTFE/PFA 2.9 / 0 * 4.2 / 0 5.9 / 0 -/0 -/0
10” PTFE 4.8 * 5.8 7.7 - -
12” PTFE 5.8 * 7.3 9.1 - -
14” PTFE 6.8 * 8.7 10.6 - -
16” PTFE 7.8 * 9.7 11.6 - -
18” PTFE
20” PTFE No vacuum is permissible!
24” PTFE

* No value can be specified.

Flow limit As a rule, the pipeline diameter determines the sensor diameter. With a known
flowrate and the help of the Endress+Hauser Flow Applicator, the best size can be
selected. Generally, a velocity range of 6 to 10 ft/sec should be selected.
A necessary increase in velocity can be achieved through a reduction of the sensor
diameter. The higher installation expense is normally balanced by the lower sensor
cost.
The flow velocity should also be determined by the fluid’s physical properties:
• v < 6 ft/sec, for abrasive mediums such as potter’s clay, lime milk, ore slurry, etc.
• v > 6 ft/sec, for fluids causing build-up such as wastewater sludge, etc.

Nominal Recommended Factory settings


diameter flow rate
Min./max. full scale value Full scale value Pulse weighting Creepage
inch mm (v ~ 1.0 or 33 ft/s) (v ~ 8 ft/s) (~ 2 pulses/s) (~ 0.1 ft/s)
1/2 15 1.0 to 27 gpm 6 gpm 0.05 gal 0.10 gpm
1 25 2.5 to 80 gpm 18 gpm 0.20 gal 0.25 gpm
1-1/2 40 7 to 190 gpm 50 gpm 0.50 gal 0.75 gpm
2 50 10 to 300 gpm 75 gpm 0.5 gal 1.25 gpm
3 80 24 to 800 gpm 200 gpm 2 gal 2.5 gpm
4 100 40 to 1250 gpm 300 gpm 2 gal 4.0 gpm
6 150 90 to 2650 gpm 600 gpm 5 gal 12 gpm
8 200 155 to 4850 gpm 1200 gpm 10 gal 15 gpm
10 250 250 to 7500 gpm 1500 gpm 15 gal 30 gpm
12 300 350 to 10,600 gpm 2400 gpm 25 gal 45 gpm
14 350 500 to 15,000 gpm 3600 gpm 30 gal 60 gpm
16 400 600 to 19,000 gpm 4800 gpm 50 gal 60 gpm
18 450 800 to 24,000 gpm 6000 gpm 50 gal 90 gpm
20 500 1000 to 30,000 gpm 7500 gpm 75 gal 120 gpm
24 600 1400 to 44,000 gpm 10,500 gpm 100 gal 180 gpm

Pressure loss No pressure loss if the sensor is installed in a pipe of the same nominal diameter.
Pressure losses occur in configurations incorporating adapters (reducers, expand-
ers). Contact your Endress+Hauser representative or Endress+Hauser for informa-
tion.

18 Endress+Hauser
Promag 50/53 P

Mechanical construction
Dimensions / mounting details Wall mounted housing
Weight: 13 lbs (6 kg)
NOTE: A mounting kit (for panel or post mounting) is available from
Endress+Hauser. Please refer to Accessories.

6.27"
Esc
(159.5)
- + E

9.84"
(250)

3.56"
(90.5)

8.46" 3.54" 1.77"


(215) (90) (45)
5.31"
(135)

3.18" 2.08" 3.18"


> 1.97" (81) (53) (81)
(50)

4.01"
(102)

2.08"
(53)

3.74"
3.20" (95)
(81.5)

0.45" 7.55" 0.45"


(11.5) (192) (11.5)

Dimensions in inches (mm)

Remote field housing


10.4"
(nonhazardous areas only U.S.) (265)
9.53"
(242)
Blind version

6.57"
(167)

13.9"
(355)

7.40"
3.94" 5.24" (188)
(100) (133)

Æ 0.34"
3.94"
(8.6 M8)
(100)
4.84"
(123)

Endress+Hauser 19
Promag 50/53 P

Promag P (compact version)

Promag P 1/2" to 12" Promag P 14" to 24"


Compact Version Compact Version
8.94" 7.68" 8.94" 7.68"
(227) (195) (227) (195)
6.61" 8.15" * 6.61"
8.15" *
(168) (207) (168)
(207)

6.30" 6.30"
Esc

(160) (160)
Esc
- + E - + E

B
A
B

C A

K L

E
* Blind option dimension K L
Dimensions in inches (mm)
Nominal Weight *
Size L A B C K E

inches mm in mm in mm in mm in mm in mm in mm lb kg

1/2 15 7.87 200 13.4 341 10.1 257 3.31 84 4.72 120 3.70 94 14 6.5
1 25 7.87 200 13.4 341 10.1 257 3.31 84 4.72 120 3.70 94 16 7.3
1-1/2 40 7.87 200 13.4 341 10.1 257 3.31 84 4.72 120 3.70 94 20 9.4
2 50 7.87 200 13.4 341 10.1 257 3.31 84 4.72 120 3.70 94 23 10.6
3 80 7.87 200 15.4 391 11.1 282 4.29 109 7.08 180 3.70 94 31 14
4 100 9.84 250 15.4 391 11.1 282 4.29 109 7.08 180 3.70 94 35 16
6 150 11.8 300 18.6 472 12.7 322 5.90 150 10.2 260 5.51 140 56 25.5
8 200 13.8 350 20.7 527 13.6 347 7.08 180 12.7 324 6.14 156 99 45
10 250 17.7 450 22.7 577 14.6 372 8.07 205 15.7 400 6.14 156 165 75
12 300 19.7 500 24.7 627 15.6 397 9.05 230 18.1 460 6.53 166 242 110
14 350 21.6 550 29.1 738.5 17.9 456.5 11.1 282 22.2 564 10.9 276 385 175
16 400 23.6 600 31.1 790.5 18.9 482.5 12.1 308 24.2 616 10.9 276 451 205
18 450 25.6 650 33.1 840.5 19.9 507.5 13.1 333 26.2 666 11.5 292 561 255
20 500 25.6 650 35.1 891.5 20.9 533 14.1 358.5 28.2 717 11.5 292 627 285
24 600 30.7 780 39.2 995.5 23.0 585 16.2 410.5 32.3 821 15.8 402 891 405

* Weight data valid for 150 lb ANSI version. For high temperature version, add 3 lb (1.5 kg).
Note: The fitting length (L) is always the same, regardless of the pressure rating.

Promag P high temperature version, 1/2” to 12”, compact

Esc
- + E

B1

4.33"
(110)
A1

Measurement A1, B1: refer to compact version dimension chart above, add 4.33”
(110 mm) to dimensions A and B for high temperature version.

20 Endress+Hauser
Promag 50/53 P

Promag P (remote version)

Promag P 1/2" to 12" Promag P 14" to 24""


Remote Version Remote Version
6.73"
6.73" 5.08"
(171)
(171) (129)
5.63" 5.63" 5.08"
(143) (143) (129)

4.01"
(102)
B

A
B

C
A
E
L K

L K

Dimensions in inches (mm)

Nominal Weight *
Size L A B C K E

inches mm in mm in mm in mm in mm in mm in mm lb kg

1/2 15 7.87 200 11.2 286 7.95 202 3.31 84 4.72 120 3.70 94 12 4.5
1 25 7.87 200 11.2 286 7.95 202 3.31 84 4.72 120 3.70 94 12 5.3
1-1/2 40 7.87 200 11.2 286 7.95 202 3.31 84 4.72 120 3.70 94 16 7.4
2 50 7.87 200 11.2 286 7.95 202 3.31 84 4.72 120 3.70 94 19 8.6
3 80 7.87 200 13.2 336 8.94 227 4.29 109 7.08 180 3.70 94 24 12
4 100 9.84 250 13.2 336 8.94 227 4.29 109 7.08 180 3.70 94 31 14
6 150 11.8 300 16.4 417 10.5 267 5.90 150 10.2 260 5.51 140 52 23.5
8 200 13.8 350 18.6 472 11.5 292 7.08 180 12.7 324 6.14 156 95 43
10 250 17.7 450 20.5 522 12.5 317 8.07 205 15.7 400 6.14 156 161 73
12 300 19.7 500 22.5 572 13.5 342 9.05 230 18.1 460 6.53 166 238 108
14 350 21.6 550 26.9 683.5 15.8 401.5 11.1 282 22.2 564 10.9 276 381 173
16 400 23.6 600 29.0 735.5 16.8 427.5 12.1 308 24.2 616 10.9 276 447 203
18 450 25.6 650 30.9 785.5 17.8 452.5 13.1 333 26.2 666 11.5 292 557 253
20 500 25.6 650 32.9 836.5 18.8 478 14.1 358.5 28.2 717 11.5 292 623 283
24 600 30.7 780 37.0 940.5 20.9 530 16.2 410.5 32.3 821 15.8 402 887 403

* Weight data valid for 150 lb ANSI version. For high temperature version, add 3 lb (1.5 kg).

Promag P high temperature version, 1/2” to 12”, remote

B1
4.33"
(110)

A1

Measurement A1, B1: refer to compact version dimension chart above, add 4.33”
(110 mm) to dimensions A and B for high temperature version.

Endress+Hauser 21
Promag 50/53 P

Ground disk, 1/2” to 12” H

ØB

∅ 0.26"
(6.5)
0.59"
(15)
ØA

0.39"
(10)

ØD

0.08"
(2)
Nominal A B D H
size 1)
in mm in mm in mm in mm

1/2 0.75 19 1.69 43 2.42 61.5 2.87 73


1 1.18 30 2.44 62 3.05 77.5 3.44 87.5
1-1/2 1.75 44.5 3.23 82 3.98 101 4.06 103
2 2.22 56.5 3.98 101 4.55 115.5 4.25 108
3 3.35 85 5.16 131 6.08 154.5 5.31 135
4 4.33 110 6.14 156 7.34 186.5 6.02 153
6 6.42 163 8.54 217 10.1 256 7.24 184
8 8.29 210.5 10.5 267 11.3 288 8.07 205
10 10.4 265 12.9 328 14.1 359 9.45 240
12 2) 12.5 317 14.8 375 16.3 413 10.7 273
1)
Ground disks, apart from those for 12", can be used for all
supplied flange standards / pressure ratings
2)
Class 150

Weight Refer to Pages 20 to 21.

Materials Transmitter housing:


• Compact housing: powder coated die-cast aluminum
• Remote field housing: powder coated die-cast aluminum
• Wall mounted housing: powder coated die-cast aluminum
Sensor housing:
• 1/2” to 12”: powder coated die-cast aluminum
• 14” to 24”: painted steel (Amerlock 400)
Measuring tube:
• < 14”: 304 SS or 304L SS with Al/Zn protective coating when supplied with
non-stainless flange material
• > 12”: 304 SS with Amerlock 400 paint when supplied with non-stainless flange
material
Flanges:
• ANSI: A105 CS (< 14” with Al/Zn protective coating; > 12” with
Amerlock 400 paint) or 316L SS
Electrodes:
• 316L SS; optional: platinum/rhodium 80/20, Alloy C-22, tantalum
Seal:
• Seals to DIN 2690
Ground disks (accessory):
• 316L SS or Alloy C-22

22 Endress+Hauser
Promag 50/53 P

Material load diagrams Caution!


The following material load diagrams are refering to a temperature range of -40° to
+356°F (-40° to +180°C) reference curves. The maximum permissible fluid tempera-
ture always depends on the lining material of the sensor and/or the seal material.

Flange, 316L SS, ANSI B16.5


[psi] [bar]

870 60

725 50

580 40
Class 300

435 30

290 20
Class 150
145 10

0 0
-76 -40 -4 32 68 104 140 176 212 248 284 320 [°F]
-60 -40 -20 0 20 40 60 80 100 120 140 160 [°C]

Flange, A105 CS, ANSI B16.5


[psi] [bar]

870 60

725 50
Class 300
580 40

435 30

290 20
Class 150
145 10

0 0
-76 -40 -4 32 68 104 140 176 212 248 284 320 [°F]
-60 -40 -20 0 20 40 60 80 100 120 140 160 [°C]

Flange, A105 CS, to AS2129 Table E

[psi] [bar]
360 25

290 20

215 15

145 10

72 5

0 0
-76 -40 -4 32 68 104 140 176 212 248 284 320 [°F]
-60 -40 -20 0 20 40 60 80 100 120 140 160 [°C]

Fitted electrodes Measuring, reference and Empty Pipe Detection electrodes:


• Standard with: 316L SS, Alloy C-22 or tantalum
• Optional: measuring electrodes with EPD and/or reference electrodes made of
platinum/rhodium 80/20

Process connections Flange connections: ANSI

Surface roughness • PFA liner: ≤ 16 µin (0.4 µm)


• Electrodes:
- 316L SS, Alloy C-22: 12 to 20 µin (0.3 to 0.5 µm)
- Tantalum, platinum/rhodium: 12 to 20 µin (0.3 to 0.5 µm)
All data relates to wetted parts.

Endress+Hauser 23
Promag 50/53 P

Human interface
Display elements • Liquid crystal display, backlit, two lines (Promag 50) or four lines (Promag 53)
with 16 characters per line
• Custom configurations for presenting different measured value and status
variables
• Flow rate: 5 digits with sign and units
• Totalizer:
Promag 50: 2 totalizers (7-digit plus 7-digit overflow with sign and units)
Promag 53: 3 totalizers (7-digit plus 7-digit overflow with sign and units)

Operating elements Unified operation concept for both types of transmitter:


Promag 50:
• Local operation with three push buttons (-, +, E)
• Quick Setup menus for straightforward commissioning
Promag 53:
• Local operation with Touch Control (-, +, E)
• Application specific Quick Setup menus for straightforward commissioning

Remote operation Promag 50; remote operation via HART®, PROFIBUS-PA


Promag 53; remote operation via HART®, PROFIBUS-PA/DP, FOUNDATION fieldbus

Certificates and approvals


Hazardous area approvals Information on the currently available hazardous area rated versions
(FM, CSA, ATEX, etc.) is available on request from Endress+Hauser. All information
relevant to explosion protection is available in separate documentation.

CE mark By attaching the CE mark, Endress+Hauser confirms that the instrument fulfills all the
requirements of the relevant EC directives.

Pressure equipment directive Flow meters with a nominal diameter smaller or equal to 1” (DN 25) are covered by
Art. 3(3) of the European directive 97/23/EG (Pressure Equipment Directive) and are
designed according to sound engineering practice. For larger nominal diameters,
optional approvals according to Cat. III are available when required (depends on the
fluid and process pressure).

PROFIBUS-DP/PA certification The flow device has successfully passed all the test procedures carried out and is
certified and registered by the PNO (PROFIBUS User Organization). The device thus
meets all the requirements of the following specifications:
• Certified to PROFIBUS-PA, profile version 3.0 (device certification number on
request)
• The device can also be operated with certified devices of other manufacturers
(interoperability)

FOUNDATION Fieldbus certification The flow device has successfully passed all the test procedures carried out and is
certified and registered by the Fieldbus FOUNDATION. The device thus meets all the
requirements of the following specifications:
• Certified to FOUNDATION Fieldbus Specification
• The device meets all the specifications of the FOUNDATION Fieldbus H1
• Interoperability Test Kit (ITK), revision status 4.0 (device certification number on
request)
• The device can also be operated with certified devices of other manufacturers
• Physical Layer Conformance Test of the Fieldbus FOUNDATION

Other standards, guidelines EN 60529: Degrees of protection by housing (IP Code)


EN 61010: Protection Measures for Electrical Equipment for Measurement, Control,
Regulation and Laboratory Procedures
EN 61326/A1 (IEC 1326): Electromagnetic compatibility (EMC requirements)
NAMUR 21: Electromagnetic compatibility (EMC) of industrial process and laboratory
control equipment
NAMUR 43: Standarization of the signal level for the breakdown information of digital
transmitters with analog output signals

24 Endress+Hauser
Promag 50/53 P

Ordering information
Promag 50 P Electromagnetic
Flowmeter
1 2 3 4 5 6 7 8 9 10 11 12 13
Promag 50 P -
Nominal Diameter 8 Protection type / Version
1 15 1/2” A NEMA 4X / compact, aluminum field housing
25 1” C NEMA 4X / remote wall mounted housing (only for approvals A or R)
40 1-1/2” G NEMA 4X / remote aluminum field housing for nonhazardous areas
50 2” K NEMA 6P / remote, wall mounted housing (only for approvals A or R),
80 3” not for liner “B” (PFA high temp.)
1H 4” N NEMA 6P / remote aluminum field housing, nonhazardous, not for
1F 6” use with liner “B” (PFA high temp.)
2H 8” 1 -40°F (-40°C) ambient temperature, NEMA 4X / compact aluminum
2F 10” housing for transmitter (only for approvals A or R)
3H 12” 3 -40°F (-40°C) ambient temperature, NEMA 4X / compact SS housing
3F 14” for transmitter (only for approvals A or R)
4H 16” 5 -40°F (-40°C) ambient temperature, NEMA 4X / remote wall mount
4F 18” housing for transmitter (only for approvals A or R)
5H 20” 9 Other
6H 24” 9 Cable for Remote Version
2 Liner 0 Without cable
A PFA liner, standard (available for 1” to 8” sensors) 1 15 foot coil and signal cable
B PFA liner, high temperature version (available for 1” to 8” sensors, not 2 30 foot coil and signal cable
for approval N) 5 Coil and signal cable, specify length in feet, maximum 650 feet,
E PTFE liner depending on conductivity of process material
3 Process Connection 9 Other
L Class 150 ANSI B16.5 carbon steel A105 flanges 10 Cable Entries
M Class 300 ANSI B16.5 carbon steel A105 flanges B 1/2” NPT
R Class 150 ANSI B16.5, 316L SS flanges L Fieldbus connector (Outputs / Inputs H and Approval A only)
S Class 300 ANSI B16.5, 316L SS flanges (not for 8” an X Sensor only, without cable (compact version only)
larger sensors) 9 Other
9 Other 11 Display / Power Supply / Operation
4 Electrodes / Material 7 85 to 260 VAC, without display, remote configuration only (not for
0 Measuring, reference and EPD electrodes / 316L SS wall mount or SS housing)
1 Measuring, reference and EPD electrodes / Alloy C-22 8 20 to 55 VAC / 16 to 62 VDC, without display, remote configuration
2 Measuring, reference and EPD electrodes / tantalum only (not for wall mount or SS housing)
3 Measuring electrodes only / Platinum/Rhodium 80/20% A 85 to 260 VAC, with display, push button operation (language: EN,
4 Measuring and reference electrodes only / Platinum/Rhodium 80/20% ES, FR, IT, NL, PT, DE)
5 Measuring, reference and EPD electrodes / Platinum/Rhodium 80/20% B 20 to 55 VAC / 16 to 62 VDC, with display, push button operation
G Measuring, reference and EPD electrodes, bullet nose / 316L SS (language: EN, ES, FR, IT, NL, PT, DE)
H Measuring, reference and EPD electrodes, bullet nose / Alloy C-22 X Sensor only (without transmitter, only available up to 8”)
9 Other 12 Software
5 Calibration A Standard
A 3-point calibration, 0.5% C ECC electrode cleaning circuit (only for approvals A or R)
B 3-point calibration, 0.2% X Sensor only
D SCS/A2LA 3-point 0.5% calibration (ISO/IEC 17025) with certificate 13 Outputs / Inputs
traceable according to ISO 9000 (specify flow range) W Current / HART
E SCS/A2LA 3-point 0.2% calibration (ISO/IEC 17025) with certificate A Current / HART, frequency
traceable according to ISO 9000 (specify flow range) D Current / HART, frequency, status, status input
9 Other H PROFIBUS-PA (only for approvals A or R)
6 Certificates S Current HART active, frequency passive; I.S. (only for approval N)
1 Standard, no certificate T Current HART passive, frequency passive; I.S. (only for approval N)
2 3.1B material certificate for pipe and flanges X Sensor only
3 2.3 pressure test certificate (1.5 x PN, 3 minutes) for sensors
up to 12” only
4 3.1B material and 2.3 pressure test certificate for sensors
up to 12” only
5 CRN approval for ANSI flanges only
8 CRN approval, 3.1B material certificate and 2.3 pressure test
certificate (1.5 x PN) for ANSI flanges only
7 Approvals
A For use in nonhazardous areas
N FM explosion proof Class I, Div. 1 / CSA Class I, Div. 1 (only for
aluminum field housing, compact version, not for 14” and larger
sensors)
R FM non-incendive Class I, Div. 2 / CSA Class I, Div. 2

Endress+Hauser 25
Promag 50/53 P

Promag 53 P Electromagnetic
Flowmeter
1 2 3 4 5 6 7 8 9 10 11 12 13
Promag 53 P -
Nominal Diameter 9 Cable for Remote Version
1 15 1/2” 0 Without cable
25 1” 1 15 foot coil and signal cable
40 1-1/2” 2 30 foot coil and signal cable
50 2” 5 Coil and signal cable, specify length in feet, maximum 650 feet,
80 3” depending on conductivity of process material
1H 4” 9 Other
1F 6” 10 Cable Entries
2H 8” B 1/2” NPT
2F 10” L Fieldbus connector (Outputs / Inputs H or K, and Approval A only)
3H 12” X Sensor only, without cable (compact version only)
3F 14” 9 Other
4H 16” 11 Display / Power Supply / Operation
4F 18” 7 85 to 260 VAC, without display, remote configuration only (not for
5H 20” wall mount or SS housing)
6H 24” 8 20 to 55 VAC / 16 to 62 VDC, without display, remote configuration
2 Liner only (not for wall mount or SS housing)
A PFA liner, standard (available for 1” to 8” sensors) A 85 to 260 VAC, with display, touch control operation (language: EN,
B PFA liner, high temperature version (available for 1” to 8” sensors, not ES, FR, IT, NL, PT, DE)
for approval N) B 20 to 55 VAC / 16 to 62 VDC, with display, touch control operation
E PTFE liner (language: EN, ES, FR, IT, NL, PT, DE)
3 Process Connection X Sensor only (without transmitter, only available up to 8”)
L Class 150 ANSI B16.5 carbon steel A105 flanges 9 Other
M Class 300 ANSI B16.5 carbon steel A105 flanges 12 Software
R Class 150 ANSI B16.5, 316L SS flanges A Standard
S Class 300 ANSI B16.5, 316L SS flanges (not for 8” and larger C ECC electrode cleaning circuit (only for Approvals A or R)
sensors) J Batching
9 Other L Batching and ECC electrode cleaning (only for Approvals A or R)
4 Electrodes / Material X Sensor only
0 Measuring, reference and EPD electrodes / 316L SS 13 Outputs / Inputs
1 Measuring, reference and EPD electrodes / Alloy C-22 A Current / HART, frequency
2 Measuring, reference and EPD electrodes / tantalum B Current / HART, frequency and 2 relays
3 Measuring electrodes only / Platinum/Rhodium 80/20% C Current / HART, frequency, 2 relays (flexible module)
4 Measuring and reference electrodes only / Platinum/Rhodium 80/20% D Current / HART, frequency, relay, and status input (flexible module)
5 Measuring, reference and EPD electrodes / Platinum/Rhodium 80/20% F PROFIBUS PA IS (only for Approval N)
G Measuring, reference and EPD electrodes, bullet nose / 316L SS G FOUNDATION fieldbus IS (only for Approval N)
H Measuring, reference and EPD electrodes, bullet nose / Alloy C-22 H PROFIBUS-PA (only for Approvals A or R)
9 Other J PROFIBUS-DP
5 Calibration K FOUNDATION fieldbus
B 3-point calibration, 0.2% L Current / HART, 2 relays and status input/output (flexible module)
E SCS/A2LA 3-point 0.2% calibration (ISO/IEC 17025) with certificate M Current / HART, 2 frequency, status input/output (flexible module)
traceable according to ISO 9000 (specify flow range) S Current / HART (active), frequency; IS(only for Approval N)
9 Other T Current / HART (passive), frequency; IS (only for Approval N)
6 Certificates 2 Current / HART, frequency, current, relay (flexible module)
1 Standard, no certificate 4 Current / HART, frequency, current input, relay (flexible module)
2 3.1B material certificate for pipe and flanges 5 Current / HART, frequency, current input, status input (flexible module)
3 2.3 pressure test certificate (1.5 x PN, 3 minutes) for sensors X Sensor only
up to 12” only
4 3.1B material and 2.3 pressure test certificate for sensors
up to 12” only
5 CRN approval for ANSI flanges only
8 CRN approval, 3.1B material certificate and 2.3 pressure test
certificate (1.5 x PN) for ANSI flanges only
7 Approvals
A For use in non-hazardous areas
N FM explosion proof Class I, Div. 1 / CSA Class I, Div. 1 (only for
aluminum field housing, compact version, not for 14” and larger
sensors)
R FM non-incendive Class I, Div. 2 / CSA Class I, Div. 2
8 Protection type / Version
A NEMA 4X / compact, aluminum field housing
C NEMA 4X / remote wall mounted housing (only for approvals A or R)
K NEMA 6P remote, wall mounted housing (only for approvals A or R),
not for liner “B” (PFA high temp.)
N NEMA 6P / remote aluminum field housing, nonhazardous, not for
use with liner “B” (PFA high temp.)
1 -40°F (-40°C) ambient temperature, NEMA 4X / compact aluminum
housing for transmitter (only for approvals A or R)
3 -40°F (-40°C) ambient temperature, NEMA 4X / compact SS housing
for transmitter (only for approvals A or R)
5 -40°F (-40°C) ambient temperature, NEMA 4X / remote wall mount
housing for transmitter (only for approvals A or R)
9 Other

26 Endress+Hauser
Promag 50/53 P

Accessories
Various accessories, which can be ordered from Endress+Hauser, are available for
the transmitter and the sensor. Contact your local representative or Endress+Hauser
for detailed information.

Promag 50/53 field housing mounting kit

DK8WM -

1 Transmitter type
A Panel/wall/pipe mount (2” pipe) hardware
B Wall/pipe mount for remote field housing

+0.02
8.27" -0.02
+0.5
(210 -0.5
) Ø 2"
(50)

+0.02
-0.02
9.65"+0.5
(245 -0.5 )
6.10"
(155)
4.33"
(110)

DK8WM - A
Mounting set for wall / panel enclosure, 3/4” to 2-3/4”
pipe
Includes:
4 mounting plates
4 threaded rods, M6 x 150 mm
4 washers
4 nuts, M6

Remote Transmitter Housing

Washer
Bolt

Mounting Plate Nut


3/4" to 3" Pipe

DK8WM - B
Mounting set for wall / pipe mount remote field enclosure, 3/4” to 3” pipe
Includes:
Mounting plate
4 bolts, M8 x 150 mm
4 washers
4 nuts, M8

Endress+Hauser 27
Supplemental documentation Promag System Information (SI 028D/24/ae)
Promag 50/53 W Technical Information (TI 046D/24/ae)
Promag 50/53 H Technical Information (TI 048D/24/ae)
Promag 50 Operating Instructions (BA 046D/06/en and BA 049D/06/en)
Promag 53 Operating Instructions (BA 047D/06/en and BA 048D/06/en)
Hazardous Area Approvals Documentation (FM, CSA, ATEX, etc.)

For application and selection assistance,


in the U.S. call 888-ENDRESS

For total support of your installed base, 24 hours


a day, in the U.S. call 800-642-8737

Visit us on our web site, www.us.endress.com

United States Canada Mexico

Endress+Hauser, Inc. Endress+Hauser Canada Endress+Hauser


2350 Endress Place 1075 Sutton Drive Paseo del Pedregal No. 610
Greenwood, IN 46143 Burlington, ON L7L 5Z8 Col. Jardines del Pedregal
Phone: (317) 535-7138 Phone: (905) 681-9292 01900, Mexico D.F.
888-ENDRESS 800-668-3199 Mexico
FAX: (317) 535-8498 FAX: (905) 681-9444 Phone: (525) 568-2405
FAX: (525) 568-7459

Endress+Hauser
The Power of Know How

TI 047D/24/ae/08.04
© 2004 Endress+Hauser, Inc.