This document reports the results of a pipette calibration test. It lists the pipette's serial number, brand name, volume range and report date. It then details the test performed at 3 volume levels, reporting the number of measurements, mean weight, mean volume, inaccuracy, imprecision and whether the pipette passed or failed the calibration based on specified targets. The pipette passed calibration according to the specified ISO standard.
This document reports the results of a pipette calibration test. It lists the pipette's serial number, brand name, volume range and report date. It then details the test performed at 3 volume levels, reporting the number of measurements, mean weight, mean volume, inaccuracy, imprecision and whether the pipette passed or failed the calibration based on specified targets. The pipette passed calibration according to the specified ISO standard.
This document reports the results of a pipette calibration test. It lists the pipette's serial number, brand name, volume range and report date. It then details the test performed at 3 volume levels, reporting the number of measurements, mean weight, mean volume, inaccuracy, imprecision and whether the pipette passed or failed the calibration based on specified targets. The pipette passed calibration according to the specified ISO standard.
Statu Test Volume SD Actual Target Status Actual Target s 10.00 µl ± 3.00 ± 1.50 25.00 µl ± 1.00 ± 0.50 50.00 µl ± 0.80 ± 0.15 STATUS : PASSED Calibration Method according to………………. EN ISO 8655-6