Professional Documents
Culture Documents
211-12.11.18-User Manual PHOENIX NCC-3300 V15.05
211-12.11.18-User Manual PHOENIX NCC-3300 V15.05
PHOENIX NCC-3300
Automated Hematology Analyzer
User Manual
V15.05 eng
NOTE:
CONTENTS
GUIDANCE ............................................................................................................................................... 9
CHAPTER 1 SYSTEM DESCRIPTION ........................................................................................................ 10
1.1 Overview ..................................................................................................................................... 10
1.1.1 Function ............................................................................................................................... 10
1.1.2 Intended Use ........................................................................................................................ 10
1.1.3 Front Panel ........................................................................................................................... 11
1.1.4 Rear Panel ................................................................................................................................ 15
1.2 Parameters .................................................................................................................................. 16
1.3 Structure ..................................................................................................................................... 17
1.3.1 Flow System ......................................................................................................................... 17
1.3.2 Electrical System .................................................................................................................. 17
1.3.2.1 FPGA Board ....................................................................................................................... 17
1.3.2.2 WBC Metering Assembly .................................................................................................. 17
1.3.2.3 RBC/PLT Metering Assembly ............................................................................................ 18
1.3.3 Display .................................................................................................................................. 18
1.4 Accessory .................................................................................................................................... 18
1.5 Sample Volume ........................................................................................................................... 18
1.6 Reagent Volume for Single sample ............................................................................................. 19
1.7 Test Speed ................................................................................................................................... 19
1.8 Storage ........................................................................................................................................ 19
1.9 Background Test .......................................................................................................................... 19
1.10 Carryover ................................................................................................................................... 19
1.11 Accuracy .................................................................................................................................... 19
1.12 Precision .................................................................................................................................... 20
1.13 Linearity .................................................................................................................................... 20
1.14 Transport and Storage Specifications ....................................................................................... 20
1.15 Environment Requirement ....................................................................................................... 20
1.16 Electrical Requirement .............................................................................................................. 21
1.17 Reagent ..................................................................................................................................... 21
1.17.1 Diluent ................................................................................................................................ 21
1.17.2 Lyse .................................................................................................................................... 21
1.17.3 Detergent ........................................................................................................................... 22
page 2
1.17.4 Probe Detergent ................................................................................................................ 22
1.17.5 Note of Reagent Use .......................................................................................................... 22
1.17.8 Reagent Storage ................................................................................................................. 23
CHAPTER 2 PRINCIPLES OF OPERATION ................................................................................................ 23
2.1 PRINCIPLES OF MEASURING ........................................................................................................... 23
2.1.1 Electrical Impedance Method .................................................................................................. 23
2.1.2 HGB Colorimetric Method ....................................................................................................... 25
2.2 REAGENTS FUNCTION ..................................................................................................................... 25
2.3 CALCULATION OF PARAMETERS ..................................................................................................... 25
CHAPTER 3 INSTALLATION AND SPECIMEN ANALYSIS ......................................................................... 27
3.1 UNPACKING AND INSPECTION ........................................................................................................ 27
3.2 INSTALLATION REQUIREMENTS ...................................................................................................... 27
3.3 POWER SUPPLY INSPECTION........................................................................................................... 28
3.4 TUBING INSTALLATION ................................................................................................................... 28
3.4.1 LYSE Tubing Installation ........................................................................................................... 29
3.4.2 DILUENT Tubing Installation .................................................................................................... 29
3.4.3 WASTE Tubing Installation ....................................................................................................... 29
3.4.4 DETERGENT Tubing Installation ............................................................................................... 29
3.5 PRINTER INSTALLATION (OPTIONAL) .............................................................................................. 30
3.6 KEYBOARD AND MOUSE INSTALLATION ......................................................................................... 30
3.7 POWER CONNECTION ..................................................................................................................... 30
3.8 STARTUP .......................................................................................................................................... 30
3.9 BACKGROUND TEST ........................................................................................................................ 31
3.10 QUALITY CONTROL ........................................................................................................................ 32
3.11 CALIBRATION ................................................................................................................................. 32
3.12 COLLECTION OF BLOOD SAMPLE .................................................................................................. 32
3.12.1 Venous Blood Collection ........................................................................................................ 33
3.12.2 Peripheral Blood Collection ............................................................................................... 33
3.13 MODE SWITCH .............................................................................................................................. 33
3.14 SAMPLE COUNTING AND ANALYSIS .............................................................................................. 33
3.14.1 Information Input ................................................................................................................... 33
3.14.2 Counting and Analysis ............................................................................................................ 35
3.14.3 Special Function ..................................................................................................................... 36
3.14.3.1 Parameter Alarm ............................................................................................................. 36
page 3
3.14.3.2 Histogram Alarm ............................................................................................................. 36
3.16 REPORT OUTPUT ........................................................................................................................... 38
3.17 RESULT MODIFICATION................................................................................................................. 38
3.18 Shutdown .................................................................................................................................. 40
3.19 Data Query ................................................................................................................................ 40
3.19.1 Selection, Browse, Modification and Output of Data ........................................................ 42
3.19.2 Data Deletion ..................................................................................................................... 43
3.19.3 Workload Statistics ............................................................................................................ 44
3.20 Special Function ........................................................................................................................ 45
3.20.1 Precision Counting ............................................................................................................. 45
3.20.1.1 Select Sample Results ..................................................................................................... 45
3.20.1.2 Check Precision ............................................................................................................... 45
3.20.2 Trend Graph ....................................................................................................................... 46
CHAPTER 4 SOFT KEYBOARD ................................................................................................................. 50
4.1 Soft Keyboard Introduction ......................................................................................................... 50
4.2 Function ...................................................................................................................................... 51
4.2.1 Data Edit ............................................................................................................................... 51
4.2.2 Condition Query ................................................................................................................... 52
4.2.3 parameter Limit setting ....................................................................................................... 53
4.2.4 L‐J QC Edit ............................................................................................................................ 54
4.2.5 X‐B QC Edit ........................................................................................................................... 55
4.2.6 X QC Edit............................................................................................................................... 56
4.2.7 Manual Calibration ............................................................................................................... 57
4.2.8 Automatic Calibration .......................................................................................................... 58
4.2.9 System Setting...................................................................................................................... 59
4.2.10 Service ................................................................................................................................ 60
CHAPTER 5 SYSTEM SETTING ................................................................................................................ 62
5.1 System Maintenance .................................................................................................................. 62
5.2 Transfer Setting ........................................................................................................................... 63
5.3 Print Setting ................................................................................................................................ 64
5.4 Parameter Setting ....................................................................................................................... 65
5.5 Date/Time Setting ....................................................................................................................... 66
5.6 System Version ............................................................................................................................ 67
CHAPTER 6 QUALITY CONTROL ............................................................................................................. 68
page 4
6.1 Quality Control Options .............................................................................................................. 68
6.2 QC Operation .............................................................................................................................. 69
6.2.1 QC Mode Select .................................................................................................................... 69
6.2.2 L‐J QC .................................................................................................................................... 70
6.2.2.1 L‐J QC Edit ......................................................................................................................... 71
6.2.2.2 L‐J QC Run ......................................................................................................................... 72
6.2.2.3 LJ QC Review ..................................................................................................................... 73
6.2.3 X QC ...................................................................................................................................... 75
6.2.3.1 X QC Edit ........................................................................................................................... 75
6.2.3.2 X QC Run ........................................................................................................................... 76
6.2.3.3 QC Review ........................................................................................................................ 77
6.2.4 X‐R QC .................................................................................................................................. 79
6.2.4.1 X‐R QC Run ........................................................................................................................ 79
6.2.4.2 X ‐R QC Review .................................................................................................................. 80
6.2.5 X‐B QC .................................................................................................................................. 83
6.2.5.1 X‐B QC Edit ........................................................................................................................ 83
6.2.5.2 X‐B QC Run ........................................................................................................................ 84
6.2.5.3 X‐B QC Review ................................................................................................................... 85
CHAPTER 7 CALIBRATION ...................................................................................................................... 88
7.1 Preparation for Calibration ......................................................................................................... 89
7.3 Auto Calibration .......................................................................................................................... 91
CHAPTER 8 PARAMETER LIMIT ............................................................................................................. 93
8.1 Limit Review ................................................................................................................................ 93
8.2 Limit Modification ....................................................................................................................... 95
8.3 Print ............................................................................................................................................. 95
CHAPTER 9 MAINTENANCE ................................................................................................................... 96
9.1 Daily Maintenance ...................................................................................................................... 96
9.2 Weekly Maintenance .................................................................................................................. 97
9.2.1 Surface Maintenance ........................................................................................................... 97
9.3 Monthly Maintenance ................................................................................................................ 98
9.4 System Maintenance .................................................................................................................. 98
9.4.1 Cauterize Aperture ............................................................................................................... 99
9.4.2 Flush Aperture...................................................................................................................... 99
9.4.3 Drain Cups ............................................................................................................................ 99
page 5
9.4.4 Rinse Cups .......................................................................................................................... 100
9.4.5 Rinse Fluidics ...................................................................................................................... 100
9.4.6 Prime Lyse .......................................................................................................................... 101
9.5 Maintenance before Shipping ................................................................................................... 102
CHAPTER 10 SERVICE .......................................................................................................................... 104
10.1 System Check .......................................................................................................................... 104
10.1.1 System Status Check ............................................................................................................ 104
10.1.2 Valve Check ...................................................................................................................... 105
10.1.3 Motor Check..................................................................................................................... 107
10.2 System Log .............................................................................................................................. 108
10.2.1 System Log Query Mode .................................................................................................. 109
CHAPTER 11 TROUBLESHOOTING ....................................................................................................... 112
11.1 Troubleshooting Guidance ...................................................................................................... 112
11.2 Obtaining Technical Assistance ............................................................................................... 113
11.3 Troubleshooting ...................................................................................................................... 113
11.4.1 Faults Related to Reagents .............................................................................................. 114
11.3.2 Faults Relate to Vacuum .................................................................................................. 114
11.3.3 Faults Relate to 5V Voltage .............................................................................................. 115
11.3.4 Faults Relate to Test Results ............................................................................................ 115
11.3.5 Faults Relate to Hardware ............................................................................................... 116
11.3.6 Faults Relate to Temperature .......................................................................................... 116
CHAPTER 12 PRECAUTINOS, LIMITATIONS AND HAZARDS ................................................................. 117
12.1 Limitations ............................................................................................................................... 117
12.2 Location Limitations ................................................................................................................ 117
12.3 Personal Protection and Infection Control ............................................................................. 118
APPENDIX A: INSTRUMENT SPECIFICATIONS ...................................................................................... 119
APPENDIX B: INSTRUMENT ICONS AND SYMBOLS ............................................................................. 122
APPENDIX C: COMMUNICATION ......................................................................................................... 123
1. HEXADECIMAL FORMAT COMMUNICATION .................................................................................. 123
1.1 Data Link MAC Sublayer Parameters Convention ..................................................................... 123
1.2 Data Link Layer Frame Format .................................................................................................. 123
1.2.1 Frame Format ..................................................................................................................... 123
1.2.2 Meaning of Fields or Control Fields ................................................................................... 123
1.2.3 Convention ......................................................................................................................... 124
page 6
1.3 Message Field Structure ............................................................................................................ 124
1.3.1 Message Structure ............................................................................................................. 124
1.3.2 DATA Field Definition ......................................................................................................... 124
2. ASCII FORMAT COMMUNICATION .................................................................................................. 125
2.1 Message Transfer Format ......................................................................................................... 125
2.2 Massage Grammar .................................................................................................................... 125
2.3 Data Type .................................................................................................................................. 125
2.4 Message Type ........................................................................................................................... 126
page 7
Copyright and Declaration
Declaration:
All contents in this manual were strictly compiled according to related laws and
regulations in Serbia, as well as the specific condition of PHOENIX NCC-3300 Auto
hematology Analyzer, covering all the updated information before printing.
NeoMedica DOO is fully responsible for the revision and explanation of the manual,
and reserves the right to renovate the relevant contents without separate notification.
Some of the demonstration pictures are for reference and subject to real object if any
differences.
All the information included is protected by copyright. No part of this document may
be reproduced, stored or transmitted in any form or by any means unless written
authorization by NeoMedica DOO
All instructions must be followed strictly in operation. In no event should NeoMedica
DOO be responsible for failures, errors and other liabilities resulting from user's
noncompliance with the procedures and precautions outlined herein.
Limited Responsibility for Quality Warranty:
The manual for PHOENIX NCC-3300 Auto Hematology Analyzer, defines the rights
and obligations between the NEOMEDICA and the customers about the
responsibility for quality warranty and after-sale service, also the related agreements
on commencement and termination.
NEOMEDICA warrants the PHOENIX NCC-3300 sold by the NEOMEDICA and its
authorized agents to be free from defects in workmanship and materials during
normal use by the original purchaser. This warranty shall continue for a period of one
year since the date of installation. The analyzer life is ten years.
NEOMEDICA assumes no liability in the following situations even during the period
of warranty:
a) Failure due to abuse the analyzer or neglect the maintenance.
b) Use reagents and accessories other than manufactured
or recommended by NEOMEDICA.
c) Failure due to operation not under the instructions described in the manual.
d) Replace accessories not specified by NEOMEDICA, or after maintenance or
repair by a service agent not approved or authorized by NEOMEDICA.
page 8
CAUTION:
NeoMedica DOO
Tel: +381 (18) 573-820, +381 (18) 573-606, +381 (18) 533-935
Web: www.NeoMedica.rs
Email: info@NeoMedica.rs
Version : V15.05
Page 9
GUIDANCE
General information for the operation of the analyzer is contained in this manual, which
covers the best guidance for a new operator to master the characteristics of the ainstrument
and operation methods, as well as for daily inquiry. Do peruse before first operation.
WARNING: Denotes a hazard which, if not avoided, could result in moderate to serious
injury.
CAUTION: Denotes potential hazards that could result in a minor injury, also used for
condition or activities which could interfere with proper function of the analyzer.
Page 10
CHAPTER 1 SYSTEM DESCRIPTION
1.1 Overview
This is multi-parameters, automated hematology analyzer designed for in vitro
diagnostic use. It gives accurate test data of human blood cells as the necessary
reference of clinical diagnosis.
1.1.1 Function
The automated hematology analyzer adopt Coulter electrical impedance and
colorimetry methods to obtain test data of WBC, RBC, PLT, HGB and other
parameters. Meanwhile analyzer gives 3 differentials of WBC and provides histogram
informations.
1.1.2 Intended Use
The automated hematology analyzer is appropriate to the qualitative and quantitative
analysis of the visible components in human beings blood.
Page 11
1.1.3 Front Panel
3 2
1. Aspiration Probe
Aspirate samples.
2. RUN Key
Press the RUN key to startup the aspiration probe and then analyze specimen
only in the screens of main menu or Quality Control. At other screens, the RUN
key is invalid.
3. Recorder
Print the test result.
4. Touch Screen
10.4 inch LCD. The screen is divided into 5 areas as shown in Figure 1-2.
Page 12
Figure 1-2 Screen
• Prompt Information
Display the prompt information.
• Mode
Display the work mode: whole blood mode or pre-diluent mode.
• System Time
Display the system date and time.
• Test Result
Display the test result.
• Menu
Display functional menus which fall into two categories.
First category menu is displayed across the bottom of the main
menu screen as shown in Figure 1-3.
Page 13
Figure 1-6 Rear Panel
1.COMI and COM2
Connect to the standard RS-232 network.
2. PRINTER
Connect to the printers
3.USB Port
Connect to USB equipment.
4. PS2 port
Connect to the keyboard and mouse.
5. Grounding Terminal
It's used to ground the analyzer.
6.Fan
It is for the heat dissipation of power supply.
7.Power Receptacle
Connect to the main power cord to the analyzer.
8.Power Switch
Turn the power supply on or off.
9.SENSOR
Connect to the waste sensor
10. DETERGENT
Detergent port connect to the detergent inlet tube.
11. WASTE
Waste port connect to the waste outlet tube
Page 16
12. LYSE
Lyse port connect to the lyse inlet tube.
13. DILUENT
Diluent port connect to the diluent inlet tube.
1.2 Parameters
Analyzer can automatically analyze the sample data, differentiates WBCs into 3
differentials and display 21 parameters and 3 histograms of WBC, RBC and PLT. Refer
to Table 1-1 for details of 21 parameters.
1.3 Structure
The automated hematology analyzer consists of a host, accessories and an external
printer (optional). A host mainly includes FPGA board, WBC metering assembly,
RBC/PLT metering assembly, flow system and a touch screen etc. Accessories include
power cord and grounding cord.
1.3.1 Flow System
The flow system is composed of solenoid valves, a vacuum pump, a force pumpe, a vacuum
chamber and plastic tubes.
Solenoid Valve --- These contact two-way or tree-way solenoid valves control the flow of
reagent.
Vacuum Pump --- Pump the waste generated in the processing out to the analyzer and
produce negative pressure.
Force Pump --- Provide positive pressure for reverse clean and lyse mixture.
Vacuum Chamber --- Generate negative pressure and play the role of temporary waste
reservoir.
Plastic Tube --- Reagent and waste flow in the plastic tube.
1.3.2 Electrical System
1.3.2.1 FPGA Board
FPGA board is the control center of the analyzer, it controls the following components and
their movement:
• All the valves open and close, reagent aspiration, rinse and waste discharge.
• Run force pump and vacuum pump to offer power to mix reagent, eliminate clog,
aspirate and discharge reagents.
• Control the A/D conversion of WBC, RBC/PLT and HGB; provide previous service for
the computer’s data processing; check all optical and electrical switch movements.
1.3.2.2 WBC Metering Assembly
WBC metering assembly is composed of a signal collection board, electrodes, a micro-
aperture sensor and flow system etc..
Page 18
• Signal Collection Board --- It provides constant current for electrodes. Then
amplifies and deals with the collected pulse signals for mainboard.
• Electrode --- There are two electrodes in WBC metering assembly. One is located
within WBC cup, and the other one is outer. Both electrodes are submerged in the
conductive liquid, creating an electrical pathway trough the micro-aperture.
• Flow System --- The flow system uses negative pressure to aspirate diluent,
detergent and sample from each container into metering tube, and discharge waste
at the end of the processing. The step motor controlled by mainboard runs to add
specific volume lyse into WBC cup, then mix it by the air created by force pump.
1.3.2.3 RBC/PLT Metering Assembly
RBC/PLT Metering Assembly is composed of a signal collection board, electrodes, a micro-
aperture sensor and flow system etc..
• Signal Collection Board --- It provides constant current for electrodes. Then
amplifies and deals with the collected pulse signals for mainboard.
• Electrode --- There are two electrodes in RBC/PLT metering assembly. One is
located within RBC/PLT cup, and the other one is outer. Both electrodes are
submerged in the conductive liquid, creating an electrical pathway trough the micro-
aperture.
• Flow System --- The flow system uses negative pressure to aspirate diluent,
detergent and sample from each container into metering tube, and discharge waste
at the end of the processing.
1.3.3 Display
The automated hematology analyzer uses a 10.4-inch LCD which can display 21 parameters
and 3 histograms.
1.4 Accessory
The accessory of the analyzer include power cord, grounding cord, external printer
(optional), etc...
1.5 Sample Volume
Whole Blood Mode for Venous Blood: Venous Blood 10µL
1.6 Reagent Volume for Single sample
Diluent: 31mL
Detergent: 8mL
Lyse: 0.7mL
1.7 Test Speed
The automated hematology analyzer is able to process at least 60 samples per hour.
1.8 Storage
The automated hematology analyzer is equipped with a memory which can store
100.000 samples data.
1.9 Background Test
WBC≤0.2x109/L; RBC≤0.02x109/L; HGB≤1g/L; PLT≤10x109/L
1.10 Carryover
WBC≤0.5%; RBC≤0.5%; HGB≤0.5%; HCT≤0.5%; PLT≤0.5%
1.11 Accuracy
The accuracy of analyzer should be complied with Table 1-2
WBC ≤±2.0%
RBC ≤±1.5%
HGB ≤±1.5%
MCV ≤±0.5%
HCT ≤±1.0%
PLT ≤±4.0%
Page 20
1.12 Precision
Table 1-3 Precision
1.13 Linearity
1.14 Transport and Storage Specifications
1.15 Environment Requirement
1) Temperature: 15 ~35°C
2) Relative Humidity: ≤90%RH
3) Barometric Pressure: 60kPa~106kPa
Page 21
1.16 Electrical Requirement
1.17 Reagent
The reagent is configured specifically for the PHOENIX NCC-3300 flow systems in order to
provide optimal system performance. Each PHOENIX NCC-3300 is checked at the factory
using the specified reagents and all performance claims were generated using these
reagents. Thus non-NEOMEDICA reagents will lead to defects in the performance of the
analyzer and serious mistakes, even accidents.
NOTE
1.17.1 Diluent
Diluent is a kind of reliable isotonic diluent to meet the requirements as follows:
(1) Dilute WBC, RBC, PLT, HGB.
(2) Keep the shape of cells during test process.
(3) Clean WBC and RBC micro-aperture and tubes.
(4) Provide a conductive environment for counting
Storage and service life after opening: Keep the diluent under 5-35 ℃, after opened, it can
be used to the validity period on the label. Once opened (connected to the instrument), the
product shelf life is only 60 days.
1.17.2 Lyse
Lyse is a new reagent without azide (NaN3 complex) and cyanide and meets the
requirements as follows:
Page 22
(1) Dissolve RBC instantly with minimum ground substance complex.
(2) Transform the membrane of the WBC to diffuse the cytoplasm. At the same time, the
membrane will shrink centre on nucleus. As a result, WBC is present in granular shape.
(3) Transform the hemoglobin to the hemo-compound which is suitable for the measurement
in the condition of 540nm wavelength.
(4) Avoid the serious pollution to human body and environment that caused by cyanide.
Storage and service life after opening: Keep the lyse under 5-35 ℃, after opened, it can be
used to the validity period on the label. Once opened (connected to the instrument), the
product shelf life is only 60 days.
1.17.3 Detergent
Detergent contains the active enzyme to clean the agglomerated proteins in the WBC, RBC
probes, the WBC, RBC cups and measurment devices.
Storage and service life after opening: Please store it in a cool dry place under 5-35℃. Away
from direct sunlight, or ingredients of detergent will be invalid as the exposure time goes on.
Once opened (connected to the instrument), the product shelf life is only 60 days.
1.17.4 Probe Detergent
Probe detergent contains effective oxide to clean the stubbornly-blocked apertures on the
WBC, RBC probes.
1.17.5 Note of Reagent Use
CAUTION
¾ Detergent and probe detergent is alkali cleaning agent
Page 23
(1) Prevent skin and eyes from contacting the reagent.
(2) Once contact with skin, rinse with water.
(3) Once contact with eyes, rinse with water and seek medical treatment immediately.
(4) If ingested, induce vomiting and seek medical treatment immediately.
1.17.8 Reagent Storage
CHAPTER 2 PRINCIPLES OF OPERATION
2.1 Principles of Measuring
The measurement is mainly on the quantity, volume of blood cells and HGB.
2.1.1 Electrical Impedance Method
The cells are counted and sized by the electrical impedance method. As Figure
2-1 shows, this method is based on the measurement of changes in electrical
current which are produced by a particle, suspended in a conductive liquid, as
it passes through an aperture of known c:limensions. An electrode is submerged in the
liquid on either side of the aperture in order to create an electrical pathway through it.
As each particle passes through the aperture, a transitory change in the resistance between
the electrodes is produced. This change produces a measurable electrical pulse. The
number of pulses generated is indicative of the number of particles that traversed the
aperture. The amplitude of each pulse is essentially proportional to the volume of the
particle that produced it.
Page 24
Each pulse is amplified and compared to internal reference voltage channels. These
channels are delineated by calibrated size discriminators to accept only pulses of a certain
amplitude. Thus, the pulses are sorted into various size channels according to their
amplitude.
Cell-flow
Direction
Iner electrode
Outer electrode
The size channels are basically divided into three categories by a pre-set
classification program in the analyzer as follows:
WBC 35 450 fL
RBC 30-110 fL
PLT 2-30 fL
According to the volume, WBCs handled by lyse can be subdivided into three
Categories: Lymphocyte (LYM), Monocyte (MID) and Granulocyte (GRAN).
LYM 35-98 fL
MID 99-135 fL
GRAN 136-450 fL
Page 25
2.1.2 HGB Colorimetric Method
Lyse added into the blood sample will crack the membrane of red blood cells promptly
and transfer into a kind of compound which can absorb the wavelength of 540 nm.
Through the comparison of the absorbance between the pure diluent and the sample,
the concentration of sample hemoglobin is calculated.
2.2 Reagents Function
In automated hematology analyzer, counting system has a high sensitivity of the cell
volume. Cells which are suspended in conducting liquid should be protected from physical
condense and adhesion. Control the osmotic pressure of conducting liquid (mainly diluent)
and keep the structure of cells so as to minimize the volume change. Lyse can dissolve the
RBC membrane fleetly and keep the structure of WBC so that the instrument can count and
classify cells.
2.3 Calculation of Parameters
• HCT ( % ) = RBCxMCV/10
• MCH ( pg ) = HGB/RBC
• MCHC ( g/L ) = 100xHGB/HCT
• PCT ( % ) = PLTxMPV/ 10000
• LYM ( % ) = 100xAL / ( AL+AM+AG )
• MID ( % ) = 100xAM / ( AL+AM+AG )
• GRAN ( % ) = 100xAG/ ( AL+AM+AG)
Page 26
12fL
Figure 2-3 P-LCR
CHAPTER 3 Installation and Specimen Analysis
3.1 Unpacking and Inspection
Carefully remove the analyzer and accessories from shipping carton, keep the
kit stored for further transport or storage. Check the following:
1) Quantity of accessories according to the packing list.
2) Leakage or soakage.
3) Mechanical damage.
4) Bare lead, inserts and accessories.
3.2 Installation Requirements
Details please refer to chapter 12 Precautions, Limitations and Hazards
3.3 Power Supply Inspection
Be sure that the system is located at the desired site before attempting any connections.
See Table 3-1 for details.
WARNING: A grounded power outlet is required to connect directly with the grounding
terminal on the rear panel. Be sure to guarantee the security of the work site.
CAUTION: A fluctuated voltage would impair performance and reliability of the
analyzer. Proper action such as the installation of E.C manostat should be taken before
operation.
CAUTION: Frequent power failure will seriously decrease the performance and
reliability of the analyzer. Proper action such as the installation of UPS should be taken
before operation.
3.4 Tubing Installation
There are four tube-connectors on the rear panel: LYSE, DILUENT, DETERGENT and
WASTE, each of which is wrapped with a cap to avoid contamination before shipment.
Uncover and set the caps aside carefully for further use on initial installation.
Page 29
3.4.1 LYSE Tubing Installation
Remove the lyse tube with red faucet from reagent kit and attach it to LYSE connector
on the rear panel, place the other end into the lyse container. Twist the cap until secure.
Place the container on the same level as the analyzer.
3.4.2 DILUENT Tubing Installation
Remove diluent tube with blue faucet from reagent kit and attach it to DILUENT connector
on rear panel. Place the other end into diluent container. Twist cap until secure. Place the
container on the same level as the analyzer.
3.4.3 WASTE Tubing Installation
Remove the waste tube with black faucet from reagent kit and attach it to WASTE connector
on the rear panel, connect BNC plug with the socket marked "SENSOR" on the rear panel.
Twist the tube's cap clockwise onto the waste container until secure. Place the container on
the level at least 50cm lower than the analyzer.
3.4.4 DETERGENT Tubing Installation
Remove the detergent tube with yellow faucet from reagent kit and attach it to DETERGENT
connector on the rear panel. Place the other end into the detergent container. Twist the cap until
secure. Place the container on the same level as the analyzer.
CAUTION: keep the tube in loose condition after installation, no distortion or folding.
CAUTION: All the tubes should be installed manually. Do NOT utilize any tool.
CAUTION: If any damage or leakage occurs in the reagent container, or the reagents have
exceeded expiry date, contacts Customer Support Centre for replacement.
WARNNING: The waste must be handled with biochemical or chemical methods before disposal,
or it will cause contamination to the environment. Users have obligation to follow the local and
national environmental regulations.
Page 30
3.5 Printer Installation (optional)
Take out the printer from the shipping carton. Inspect the printer carefully according to its manual
and Section 3.1 and perform the following procedures:
1) Find a suitable location adjacent to the analyzer. Location of at least 30cm away from analyzer
on its right side is recommended.
2) Assemble the printer as directed in the printer manual.
3) Connect the printer and analyzer with printer cable which plug into PRINTER or US8
on rear panel of the analyzer according to the type of printer.
4) Be sure that the printer power switch is OFF; plug the end of power cord to power
socket.
5)install printing paper as directed in the manual.
3.6 Keyboard and Mouse Installation
Remove keyboard, mouse and mouse pad from the shipping carton, and insert the plugs of
keyboard and mouse into the two connector of the line, then connect to the rear panel with
"PS/2" port. It is recommended to place the keyboard beneath the display.
3.7 Power Connection
Make sure the power switch is OFF (O) and the grounding terminal on the rear panel is well
grounded firstly, then connect the analyzer to the main power with the power cable.
3.8 Startup
Turn on the power switch on the rear panel. The analyzer will start system check, self-
checking and after system is up, the analyzer starts with fluidic initialization and
automatically prime the diluent, detergent and lyse, then rinse the tubing.
The main menu screen is displayed after self-checking (See Figure 3-1).
Page 31
3.9 Background Test
Background test should be performed after startup and before blood sample test, operate as
follow:
1) Put the clean empty tube under the aspiration probe. At main menu screen, click the
mode switch button on the top of the screen, current mode will be switched to “Pre-diluent”
mode, then click “Drain to discharge the diluents into the tube.
2) At main menu screen, click “Info”, and then modify ID to 0, click “OK” back to save it.
3) Click “Pre-diluent” to switch to “Whole Blood” mode, put the tube containing diluents
beneath aspiration probe and ensure the probe touch the bottom of tube.
4) Press RUN key on the front panel, move away the tube after the beep sounds. Then
analyzer starts to count and measure automatically.
5) Counting time of RBC, WBC will be displayed at the lower right corner of screen during
counting. Analyzer will alarm and display the error at top left corner if counting time is too
long or too short. Refer to Chapter 11 for problem correction.
If the background result is out of acceptable range, repeat the above procedures until reach the
acceptable results.
NOTE: ID number of background test is set to be 0 by the software to make the result not
memorized in the analyzer.
NOTE: The ID number of blood sample test can’t be set to 0.
3.10 Quality Control
Quality control should be performed before daily test or on the initial installation. Refer to
Chapter 6.
3.11 Calibration
On the initial installation, if the backgroun results and quality control are normal,
recalibration is not necessary. If not and there are shifts or trends in some parameters,
recalibrate the analyzer referring to Chapter 7.
3.12 Collection of Blood Sample
CAUTION: Consider all the clinical specimens, controls and calibrators etc that contain
human blood or serum as being potentially infectious, wear lab coats, gloves and safety
glasses and follow required laboratory or clinical procedures when handling these
materials.
CAUTION: Blood collection and disposal should be performed according to the local and
national environmental regulations or laboratory's requirements.
Page 33
CAUTION: Be sure the blood collection clean and contamination-free. All specimens must
be properly collected in tubes containing the EDTA (EDTA-K2'2H20) anticoagulant used by
the laboratory.
3.12.1 Venous Blood Collection
Collecting whole blood sample through vein-puncture and store in a clean sample tube
with EDTA-K2·2H20, which can keep the configuration of WBC, RBC and avoid
platelets aggregation. Gently shake the tube 5~10 times to make it well mixed.
3.12.2 Peripheral Blood Collection
Capillary blood is usually collected from finger tip. The volume of sample tube is set to
be 20µL.
CAUTION: Never over-press the finger avoiding collecting tissue liquid into sample
tube, tissue liquid will cause error in results.
3.13 Mode Switch
In the screen as Figure 3-1 shows, click the mode switch button on the top to switch among
Whole Blood Mode for Venous Blood, Pre-diluent Mode for Peripheral Blood and Whole
Blood Mode for Peripheral Blood. Corresponding sign on screen will indicate current
operating mode..
3.14 Sample Counting and Analysis
3.14.1 Information Input
Click "info" at main menu screen, the Info edit window present (shown in Figure 3-2), input
or select data. Click "OK" to save the input data and return to the main menu. Click "Cancel"
Page 34
to cancel the input data and return to the main menu.
NOTE: The ID number is set to 0 only under background test. The blood sample ID CAN
NOT be 0.
3.14.2 Counting and Analysis
Counting and analysis should be performed within 3~5 minutes after blood
collection.
Pre-diluent Mode for Peripheral Blood
1) Present the empty sample tube under the aspiration probe. At main menu screen, click "Drain";
the diluent will be dispensed into the tube.
2) Remove the tube, add 20uL of the blood sample to the tube, and gently shake the tube
to make them well mixed.
3) Present the well-mixed sample under the aspiration probe; make sure the probe touches
the tube bottom slightly.
4) Press RUN key on the front panel and remove the sample after hearing beep sound.
1) Gently shake the tube to well mix the blood sample, then place the sample tube beneath
the probe, make sure the probe touches tube bottom slightly.
2) Press RUN key and remove the sample after hearing beep sound.
3) Process of analysis will take some time, please wait a moment.
Page 36
Whole Blood Mode for Peripheral Blood
1) Gently shake the tube to well mix the blood sample, then place the sample
tube beneath the probe,
2) Press RUN key and remove the sample tube after hearing beep sound.
3) Process of analysis will take some time, please wait a moment.
Test results and histograms of WBC, RBC and PLT will be displayed at main menu screen
after counting and analysis (see Figure 3-1).
If Auto Rec or Auto Print is ON (set in "system setting" screen), the test results will be
printed out automatically.
If problems like clog or bubbles occur during the counting and analysis procedures, the
analyzer will alarm and give indication at the top left corner of the screen. The test results
are invalid. Refer to Chapter 11 for solution.
3.14.3 Special Function
3.14.3.1 Parameter Alarm
"H" or "L" present on the right side of the parameter means the result is out of
the range of reference value.
"***" means the result is invalid or out of display range.
3.14.3.2 Histogram Alarm
If the WBC Histogram is abnormal, R1, R2, R3, R4, RM will be displayed on the right side of
the histogram.
R1 indicates there is abnormality in the left side of LYM wave peak, which probably caused
by incomplete hemolysis of RBC, platelet clump, giant platelet, plasmodium, nucleated
RBC, abnormal lymphocyte, proteinic or fat granule.
R2 indicates there is abnormality in the area between LYM wave peak and MID wave, which
probably caused by pathologic lymphocyte, plasmocyte, atypia lymphocyte, original cell or
an increase in eosinophil and basophilia,
Page 37
R3 indicates there is abnormality in the area between MID wave and GRAN wave peak,
which probably caused by immature granulocyte, abnormal cell subpopulation,
eosinophilia.
R4 indicates there is abnormality in the right side of GRAN wave peak, which probably
caused by an absolute increase in granulocyte.
When the histogram of PLT has abnormalities, PM alarm will be shown in the right side.
PM indicates there is ill-defined boundary between PLT and RBC, which probably
caused by the present of giant platelet, platelet clump, small RBC, cell debris or fibrin.
NOTE: WBC differential may be incorrect if WBC is lower than 0.5x109/L. Microscope
examination is recommended.
3.16 Report Output
The automated hematology analyzer offers recorder and printer which are optional
according to customer needs. After blood sample analysis completed, if Auto Print is ON,
test report will be printed automatically by recorder or printer; if the Auto Trans is ON, test
results will be transmitted to network automatically.
The recorder, printer and transmit are set up at Settings window. Refer to Chapter 5 for
details.
3.17 Result Modification
If the auto-classification of floating limit for WBC, RBC and PLT do not reach clinical or
laboratory requirements on special samples, manual classification is feasible.
CAUTION: Unnecessary or incorrect manual classification will cause unreliable test results.
It is recommended to microscopic exam before clinical
2) Once the diagram parameter need to be modified is selected, click “Class” to select the
desired classification, then the classified line will change from white line to red line.
3) Click "Left" or "Right" to move the classified line, and the value of classified line will
be indicated at the lower right of the screen.
4) Click "Back" after modification, the dialog box as shown in Figure 3-4 will display;
click "NO" to cancel the modification, while click "YES" to save the modified
results.
3.18 Shutdown
Shutdown procedure is performed after daily operation and before turning the analyzer off.
Daily maintenance and tubing-clean avoid protein aggregation during non-working and keep
system clean. Shutdown procedure is as follows:
1) At main menu screen, click "Exit", shutdown information will appear (see Figure 3-5).
Figure 3-5
If turn off the instrument, click "Yes" After finishing maintenance, cleaning and
shutdown procedures, "Thank you, and now turn off power" will appear to instruct
operator to turn off the power switch on rear panel.
CAUTION: May lead to data loses if turn off the analyzer against procedures.
3.19 Data Query
The information, parameters and histograms of test results can be reviewed and printed
out by recorder or printer.
At main menu screen, click "Rev" to enter query screen as shown in Figure 3-6.
Click "Condi" at query screen, then condition query screen will appear as shown in Figure 3-
7. Operator can query the results according to Date, Name, Barcode, Dept, Sender and
Checker. In Figure 3-7, enter the time interval, click "OK", and then the samples in this time
interval will be shown in the query list. As shown in Figure 3-8.
Page 41
3.19.1 Selection, Browse, Modification and Output of Data
At main menu screen, click "Func", and then click "Rev" to enter query screen.
Data of today will be displayed in the list box as Figure 3-6.
Select data in the list, and then click "Detail", that the analyzer enters into detail
inquiry window.
Condi: Query data that are compliant with specific criteria in certain period.
Detail: Select a data in the list, click "Detail", the parameters result and
histograms of selected data will be displayed.
Pgprv/Pgnex: If the data is too much to display in one page, the system will display the
data in more pages. Click "Pgpre" or "Pgnex" to view more information.
If the sample quantity reaches a certain amount and take up a large save space, operator
can delete the data termly if necessary. Data deletion is divided into “Delete” and “Delete
All”.
1) Delete All
Click “DelAll”, a dialog box as Figure 3-9 will present, input password 9999, then the system
will prompt to ask whether you want to delete all, see Figure 3-10, if click “Yes”, then system
will perform all delete operations to current page.
2) Delete Single
In the interface as shown in Figure 3-8, select the data and then click “Del”, the dialog box as
Figure 3-11 will display.
3.19.3 Workload Statistics
At main menu screen, click "Func"~ "Stast" to enter Workload statistics window. See Figure
3-12. Operation procedure is as follows:
1) At “From” and “To”, select starting date and ending date in pop-up calendar, then press
“OK”.
2) Select one statistic type on left side of the Workload Statistics screen and then all items
will be displayed in the middle box.
3) Select item needed (multi-select is allowed), click “Stast”, and then the desired data will
be displayed in the list on right.
5) Choose one sender, and click “Print”, then all items will be print.
Page 45
3.20 Special Function
3.20.1 Precision Counting
3.20.1.1 Select Sample Results
Aer data review from condition Que click one result, then press space bar (click "Seect"
if using touch screen), the result will be selected and in rd letters as Figure 3-13.
3.20.1.2 Check Precision
After selecting one sample result as preceding method, press F9 (click „CV” if using touch
screen) to enter the CV data screen like as Figure 3-14.
„Mean” indicates the parameter average value of selected samples. „CV” indicates the
Coefficient of Variance of corresponding parameter.
NOTE: The system can only calculate CV automatically when more then one sample results
are selected.
Page 46
NOTE: If only one sample result is selected, the „Mean” indicates the sample result itself.
3.20.2 Trend Graph
After data review from condition query, click one result, and press space bar (click „Select” if
using touch screen), the result will be selected to be in red letters like Figure 3-15.
Page 47
After selecting one sample result as preceding method, press F8 *click „Trans” if using touch
screen) to enter the trend screen like Figure 3-16.
Page 48
Left: Click „Left”, then the chart pole will shift one grid to the left. See Figure 3-17.
Right: Click „Right”, then the chart pole will shift one grid to the right. See Figure 3-18.
Page 49
The automated hematology analyzer adopt t-touch interactive mode. Systems provide soft
keyboard input mode to facilate the data input and other input operations. Soft keyboard is
as Figure 4-1 shown.
The default soft keyboard is in English lowercase mode, click „CAP” key to switch to English
uppercase mode, as shown in Figure 4-2.
4.1 Soft Keyboard Introduction
1. Layout
Ther are 43 keys in soft keyboard: 10 number keys, 26 alphabet keys, backspace key, slash
key, space key, CAP key.
First click the target edit box, at this time the top of the target of the target interface is blue,
then click soft keyboard, right now the top of the soft keyboard changes from gray to blue.
User can switch between uppercase and lowercase by clickin „CAP” and then click the
keyboard input.
3. Delete Character
First move the cursor behind the character that needs to be deleted, click „Å” to delete the
character before the cursor.
4.2 Function
The soft keyboard is displayed in the following screens: data edit, condition query,
parameters limit setting, control edit, system calibration, system setting and service.
4.2.1 Data Edit
The procedure of inputting information in edit box is as follows:
1) Click target edit box, move the cursor on the edit box;
2) Click the soft keyboard, then user can input data through soft keyboard;
3) After data input, click data edit screen to move the cursor to it. Click „OK” to save the
operations, click „No” to cancel and exit the data edit screen. See Figure 4-3.
Page 52
4.2.2 Condition Query
Click „Condi” at query screen, condition query and soft keyboard will pop up at the same
time, and soft keyboard is warded off by condition query screen.
1) Click target edit box, move the cursor on the edit box;
2) Click the soft keyboard, this moment soft keyboard is warded off by condition query
screen, click soft keyboard to input data;
3) After data input, click condition query screen to move the cursor on this screen, click „OK”
to query according to input conditions, click „No” to cancel and exit query. See Figure 4-4.
Page 53
4.2.3 parameter Limit setting
Enter the parameter setting screen, click „Keyboard”, soft keyboard will pop up, if click
„Keyboard” again, soft keyboard will be hidden.
1) Click target edit box, move the cursor on the edtit box;
2) Click soft keyboard, user can input data through soft keyboard;
3) After data input, click parameter limit setting screen to move the cursor on this screen.
This moment soft keyboard will be hidden. If click „Keyboard” again, the soft keyboard will
pop up;
4) When the focus is on parameter limit setting screen, click „OK” to save, click „Back” to
cancel and exit. See Figure 4-5.
Page 54
Figure 4-5 Parameter Limit Setting
4.2.4 LJ QC Edit
Enter L.-J QC Edit screen, click "Keyboard", then soft keyboard will pop up, if
click "Keyboard" again, soft keyboard will be hidden.
The procedure of inputting information in edit box is as follows:
1) Click target edit box first, move the cursor on the edit box;
2) Click soft keyboard, user can input data through soft keyboard;
3) After data input, click L-J QC edit screen to move the cursor on this screen.
This moment soft keyboard will be hidden. If click "Keyboard" again, the
soft keyboard will pop up;
4) When the focus is on L-J QC edit screen, click "OK" to save, click "Back" to
cancel and exit. See Figure 4-6.
Page 55
4.2.5 XB QC Edit
Enter X-B QC Edit screen, click „Keyboard”, then soft keyboard will pop up. If click
„Keyboard” again, soft keyboard will be hidden.
1) Click target edit box first, move the cursor on edit box;
2) Click soft keyboard, then user can input data through soft keyboard;
3) After data input, click X-B QC edit screen to move the cursor on this screen, this moment
the soft keyboard will be hidden. If click „Keyboard”again, the soft keyboard will pop up;
4) When the focus is on X-B QC edit screen, click „OK” to save, click „Back” to cancel and
exit. See Figure 4-7.
Page 56
Figure 4-7 X-B QC Edit
4.2.6 X QC Edit
Enter X QC Edit screen, click "Keyboard", then the soft keyboard will pop up, if
click "Keyboard" aglain, soft keyboard will be hidden.
The procedure of inputting information in edit box is as follows:
1) Click target edit box first, move the cursor on the edit box;
2) Click soft keyboard, then user can input data through soft keyboard;
3) After data input, click X QC edit screen to move the cursor on this screen,
this moment soft keyboard will be hidden. If click "Keyboard" again, the
soft keyboard will pop up;
4) When the focus is on X QC edit screen, click "OK" to save, click "Back" to
cancel and exit. See Figure 4-8.
Page 57
4.2.7 Manual Calibration
Enter manual calibration screen, click „Keyboard”, then soft keyboard will pop up, if click
„Keyboard”again, soft keyboard will be hidden.
1) Click target edit box, move the cursor on the edit box;
2) Click soft keyboard, then user can input data through soft keyboardl
3) After data input, click manual calibration screen to move the cursor on this screen, this
moment soft keyboard will be hidden. If click „Keyboard” again, soft keyboard will pop up;
4) When the focus is on manual calibration screen, click „OK” to save, click „Back” to cancel
and exit. See Figure 4-9.
Page 58
4.2.8 Automatic Calibration
Enter automatic calibration screen, click "Keyboard", soft keyboard will pop up.
If click "Keyboard" again, soft keyboard will be hidden.
The procedure of inputting information in edit box is as follows:
1) Click target edit box, move the cursor on the edit box;
2) Click soft keyboard, then user can input data through soft keyboard;
3) After data input, click automatic calibration screen to move the cursor on
this screen, this moment soft keyboard will be hidden. If click "Keyboard"
again, soft keyboard will pop up;
4) When the focus is on automatic calibration screen, click "OK" to save, click
"Back" to cancel and exit. See Figure 4-10.
Page 59
4.2.9 System Setting
Enter system setting screen, click "Keyboard", soft keyboard will pop up.
If click "Keyboard" again, soft keyboard will be hidden.
The procedure of inputting information in edit box is as follows:
5) Click target edit box, move the cursor on the edit box;
6) Click soft keyboard, then user can input jets through soft keyboard;
7) After data input, click system setting on screen to move the cursor on
this screen, this moment soft keyboard will be hidden. If click "Keyboard"
again, soft keyboard will pop up;
8) When the focus is on system settin screen, click "OK" to save, click
"Back" to cancel and exit. See Figure 4-11.
Page 60
4.2.10 Service
Enter service screen, click "Keyboard", then soft keyboard will pop up, if click
"Keyboard" again, soft keyboard will be hidden.
The procedure of inputting information in edit box is as follows:
1) Click target edit box, move the cursor on edit box;
2) Click soft keyboard, then user can input data through soft keyboard.
3) After data input, click service screen to move the cursor on this screen, this
moment the soft keyboard will be hidden. If click "Keyboard" again, the soft
keyboard will pop up.
4) When the focus is on service screen, click "OK" to save, click "Back" to
cancel and exit. See Figure 4-12.
Page 61
This machine has 4 options to satisfy different requirements of laboratory and clinical
diagnostics. Operator can choose different operating modes according to actual need.
At main menu screen, click "Func" And then click "Setup". The Setup menu will be displayed
as Figure 5-1:
5.1 System Maintenance
5.2 Transfer Setting
In transfer setting as Figure 5-2, operator can setup the port number, boud rate, data bit,
stop bit and parity bit of the communication port. The communication parameters are set
before delivery. User should not modify them; otherwise the data can not be transmitted.
Auto trans: the test result will be transmitted from the communication port automatically.
In Print Setting as Figure 5-3, operator can select printer type, print format, auto print, and
input hospital name in „print title”.
In Parameter Setting as Figure 5-4, operator can setup the unit of WBC, RBC, PLT, HGB
and MCHC, as well as the language and other parameters.
In Date/Time Setting as Figure 5-5, there are 3 formats of date: YYYY-MM-DD, MM-DD-
YYYY, and DD-MM-YYYY. Y indicates Year, M indicates Month, D indicates Day. The
selected date format will be displayed.
The resetting of the data format will affect the display format of data on blood cell analysis
screen.
User can check the software version, printer software version printer template version,
FPGA version, kernel version and library version. If there are problems, the user can provide
this information to the service engineers as Figure 5-6.
NOTE: Above system setting have been set up before delivery. User does not need to reset
them generally. If needed, all the operations should be under guidance of the engineer.
Page 68
CHAPTER 6 Quality Control
Quality control is needed for maintaining the analyzer precision and eliminating system
errors. The automated hematology analyzer offers four quality control options: L-J QC,
QC, -R QC and X.-B QC. In following conditions, perform quality control with control
materials recommended by us.
o After daily start-up procedures completed
o The reagent lot number changed
o After calibration
o After maintenance, or component replacement
o In accordance with the laboratory or clinical QC protocol
CAUTION: Consider all clinical specimens, controls and calibrators etc. that
contain human blood or serum as potentially infectious. Wear lab coats, gloves
and safety glasses and follow required laboratorial or clinical procedures when
handling these materials.
6.1 Quality Control Options
(1) L-J QC
L.-J QC (Levey-Jennings graph) is a simple and visual QC method. Operator can draw QC
value directly on graph after getting the Mean, SD and CV which derived from following
formulas:
Page 69
(2) X-R QC
In X-R QC method, indicates mean value, R indicates range of value. graph is mainly used
to judge that if the mean value falls in required level. R graph is mainly used to judge that if
the range of value falls in required level.
(3) X QC
QC is the variation of -R QC; they have the same basic principle. The difference is that the
control dot in graph indicates the mean value of two values other than a single value. On
this foundation, analyzer calculates the
Mean, SD and CV.
(4) X-B QC
X-B QC is a moving average method promoted in 1970s. Its based on the principle that RBC
count is varied due to the concentration of dilution, human blood pathology and technical
factor, but the hemoglobin content in specific unit is hardly interfered by those preceding
factors. According to this characteristic, X-B QC is done by detecting the value of MCV,
MCH, and MCHC.
6.2 QC Operation
6.2.1 QC Mode Select
In main menu screen, click “Func” and then select “QC”, dialog box as Figure 6-1 will pop
up.
Page 70
The automated hematology analyzer offers 4 quality control options: L-.J QC, X-B QC, -x -R
QC and QC. Select QC mode and click ":OK" to enter corresponding QC interface.
6.2.2 LJ QC
Select L-J QC mode and click "OK" to enter corresponding screen as Figure 6-2.
In L-J QC screen click “Edit”, enter QC Edit screen as figure 6-3. There are 3 different
groups and each group has 3 (low, normal and high) levels. Input control lot No., expiry date,
assay and limit according to the control description.
NOTE: Assay is a standard value which is the reference of quality control. Limit indicates the
allowable biased range, but the limit should not be more than 40% of assay, or it cannot be
saved in database.
NOTE: The expiry date format should be MM-DD-YYYY.
At QC Run screen, place the control tube under aspiration probe, press RUN key, the
analyzer will start to process control sample. L-J QC needs control material. If run a
background QC, the system will alarm QC result is invalid. Each time runs the L-J QC, "Run
Time" on upper right corner of the Run screen will be updated correspondingly. Lot No. and
expiry date can be input in Edit screen.
The automated hematology analyzer offers 2 ways to review: QC Graph and QC Data.
Figure 6- S LJ QC Graph
In L.-J QC screen, there are low, normal and high graphs. If select group 1 and
low level to run QC, the control dot will present in low 1 graph. It is also true for
other types of QC.
There are function buttons at the bottom of L-.J QC screen. Click "Group" to
change the group. Click "Paramr to change current displayed parameter, for
instance, change from WBC to RBC. Click "Level" to shift the classification line
in the same group. Click "Left" or "Right" to shift the classification line in same
Page 74
QC graph. Click "Print" to print the current data.
QC results are arranged in graphs according to storage time. The latest is on the left
side and its serial number is 1.
QC graph instruction:
1. Graph abscissa indicates QC times, ordinate indicates QC results.
2.QC graph can display 31 dots for each parameter.
3. Every parameter graph's upper transverse line means assay plus limit.
4.Every parameter graph's lower transverse line means assay subtract limit.
5. The 3 values on the left side of each parameter's graph mean:
o upper limit -- assay plus limit
o middle line -- assay
o lower limit -- assay subtract limit
If the control dot falls in the area between upper and lower lines of the
corresponding graph, it means the dot is in the control range; if not, the dot is
not in the control range. .
Figure 6-6 L-J QC Data
Page 75
In this screen, click “Group” to change group, click “Left” or “Right” to switch page. Operator
could review 31 items at most. Click “Del All” to delete all data. The assay and limit can be
input and changed in QC Edit screen.
The QC data would be updated after running a new control.
6.2.3 X QC
6.2.3.1 X QC Edit
Select X QC mode in the dialog box as Figure 6-1 shows and then click “OK” to enter
corresponding screen. Click “Edit”, enter X QC Edit screen as Figure 6-7.
In X QC Editscreen, click |Group” to switch group; click “Del” to delete current assay and
limit; click “OK” to save the current assay and limit; click “Back” to exit X QC Edit screen.
NOTE: The same as L-J QC, the limit should not be more than 40% of assay or it cannot be
saved in database.
Page 76
NOTE: The expiry date format should be MM-DD-YYYY
6.2.3.2 X QC Run
At this screen, system displays two control results and calculate the mean value
automatically. The assay is input in X QC Edit screen. Click “Group” to swith group; click
“Back” to exit.
In X QC Run screen, place the control tube under aspiration probe, press RUN key, the
analyzer will start to process control sample. If the current group assay is empty, the system
will display “No QC reference data, cannot perform QC running”. At this time , operator
should back to the Edit screen to input assay and limit.
X QC needs control material. If run a background QC, the system will alarm QC result is
invalid.
Page 77
6.2.3.3 QC Review
The automated hematology analyzer offers 2 ways to review: QC Graph and
QC Data.
(1) X QC Graph
In QC screen, there are low, normal and high graphs. If select group 1 and low level to run
QC, the control dot will present in low 1 graph. It is also true for other types of QC.
There are function buttons at the bottom of QC screen. Click "Group" to change the group.
Click "Param" to change current displayed parameter, for instance, change from WBC to
RBC. Click "Level" to shift the classification line in the same group. Click "Left" or "Right" to
shift the classification line in same QC graph. Click "Print" to print the current data.
QC results are arranged in graphs according to storage time. The latest is on the left side
Page 78
and its serial number is 1.
QC graph instruction:
1.Graph abscissa indicates QC run times, ordinate indicates QC results.
2. QC graph can display 31 dots for each parameter.
3. Every parameter graph's upper transverse line means assay Plus limit.
4. Every parameter graph's lower transverse line means assay subtract limit.
In L-,J QC graph screen (see Figure 6-9), click "Data", operator can review QC data with 12
parameters as Figure 6-10 shows.
In this screen, click “Group” to change group, click “Left” or “Right” to switch page. Operator
could review 31 items at most. Click “Del All” to delete all data. The assay and limit can be
input and changed in QC Edit screen.
The QC data would be updated after running QC twice. At the same time, the mean value
will be displayed..
Page 79
6.2.4 XR QC
6.2.4.1 XR QC Run
At this screen, system display two control result and calculate the mean value and range
automatically. Click “Group” to switch group, click “Back”to exit.
X-R QC needs material. If run a background QC, the system will alarm QC result is invalid.
Page 80
6.2.4.2 X R QC Review
The automated hematology analyzer offers two ways to review: QC Graph and QC Data.
(1) X -R QC Graph
Click "Back" on QC Run screen or select corresponding QC mode in QC Mode dialog box,
enter the X-R QC screen as Figure 6-12 shows. Operator can review QC results with 12
parameters. The dot on X-R QC Graph indicates mean value or range of 2 QC results. The
system cannot display low, normal and high control graphs simultaneously in one screen,
please click "Group" to switch.
X-R QC graph is divided into 2 parts: X graph and R graph. X graph displays mean value
dots while the R graph displays range dots.
There are function buttons at the bottom of X-R QC screen. Click "Group" to change the
group. Click "Param" to change current displayed parameter, for instance, change from
WBC to RBC. Click "Level" to shift the classification line between and R graphs. Click "Left"
or "Right" to shift the classification line in X or R graph. Click “Print” to print current data.
Page 81
QC results are arranged in graphs according to storage time. The latest is on the left side
and its serial number is 1.
X QC graph instruction:
1. Graph abscissa indicates QC run times, ordinate indicates QC results.
2. QC graph can display 31 dots for each parameter.
3. Every parameter graph's middle transverse line indicates X – the mean value of QC
results.
4. Every parameter graph's upper transverse line means X upper limit = X+AxR
5. Every parameter graph's lower transverse line means X lower limit = X-AxR
3. Every parameter graph's middle transverse line indicates R-the mean value of
QC result range.
4. Every parameter graph's upper transverse line means R upper limit=BxR.
5. Every parameter graph's lower transverse line means R lower limit=CxR.
6. The 3 values on the left side of parameter graph mean:
o upper limit -- R upper limit= BxR
o middle line -- R
o lower limit -- R lower limit- CxR
If the control dot falls in the area between upper and lower lines of the
corresponding graph, it means the dot is in the control range; if not, the dot is
not in the control range.
In X-R QC Graph screen (See Figure 6-12), click "Data", operator can review QC data
with 12 parameters as Figure 6-13 shows.
On this screen, click "Group" to change group, click "Left" or "Right" to switch page.
Operator could review 31 items at most. Click "DelAlI" to delete all the data.
X-R QC data screen a only display three control results, and each one contains mean
and range. The first two lists on this screen are total mean and average range. See
Figure 6-13.
The QC data would be p ated after running new controls twice.
Page 83
6.2.5 XB QC
6.2.5.1 XB QC Edit
X-B QC Edit is different to others, with which the system edits only three parameters: MCV,
MCH and MCHC.
Select X-B QC mode in the dialog box as Figure 6-14 shows and then click “OK” to enter the
X-B QC screen, then click “Edit” to enter X-B QC Edit screen. See figure 6-14.
At Edit screen, click “Del” to delete current assay and limit; click “OK| to save them; click
“Back” to exit.
NOTE: Assay is a standard value which is the reference of quality control. Limit indicate the
allowable biased range, but limit should not be ore then 40% of assay or it cannot be saved
in database.
NOTE: The expiry date format should be MM-DD-YYYY.
Page 84
6.2.5.2 XB QC Run
Click "Run" on the X-B QC screen to enter the X-.B QC Run screen as Figure 6-15. "X-BQC
Run" is for selecting to run X-.B QC or not. "Sample num" is for selecting the quantity of
samples for each group. For example, if "X-BQC Run" is ON and "Sample num" is 20, the
subsequent 20 counts will be X-B QC.
Click "OK" to save the current selections.
After 20 times of count completed, back to the Run screen and click "COUNT", system will
calculate the QC results and display them in QC Graph and QC Data.
Page 85
6.2.5.3 XB QC Review
The automated hematology analyzer offer two ways to review QC result: QC graph and QC
data.
Figure 6-16 X-B QC Graph
In the screen as Figure 6-16, click "Data", operator can review 3 parameters
QC data as Figure 6-17 shows. Click "Left" or "Right" to switch page, operator
could review 31 items at most. Click "DelAII" to delete all the data. The Assay
Page 87
And Limit can be input and changed in QC Edit screen.
The QC data would be updated after running a new control.
To ensure analyzer s precision and obtain reliable test results, the parameters (WBC, RBC
PLT, HGB, and MCV) should be calibrated in the following situations:
MCV, HCT are relative parameters to each other, thus one can be obtained from given value
of the other. Only MCV will be calibrated by the analyzer. Usually the manufacturer artificial
blood material will give reference value for MCV, HCT at the same time.
CAUTION: Consider all clinical specimens, control and calibrators etc. that contain human
blood or serum as potentially infectious. Wear lab coat, gloves and safety grasses and follow
required laboratorial or clinical procedure when handling thes materials.
CAUTION: Only calibrators recommended by NeoMedica can be used to accomplish the
calibration.
CAUTION: Follow the recommendations provided by manufacturer to store the calibrators
CAUTION: Check if the container is broken or cracked before using the calibrator.
CAUTION: Make sure the calibrators are whitin expiry date.
CAUTION: Make sure the analyzator has no problem before calibration.
CAUTION: Never apply test data to laboratory or clinic use unless all parameters are
accurately calibrated.
Page 89
7.1 Preparation for Calibration
2) Run a background test and make sure the results are qualified.
4) Verify the precision of the analyzer. At main menu screen, run normal control 11 times,
query the results from second to eleventh in Query screen, make sure the CVs are within the
prescribed limit in Table 7-1.
5) Running with high control in triplicate, then using diluents instead of high control to test 3
times continuously, Carryover(%) is calculated from the folloing formula. Results must
conform to Table 7-2.
Parameter Result
WBC ≤0.5%
RBC ≤0.5%
HGB ≤0.5%
PLT ≤0.5%
Page 90
NOTE: If whole blood and capillary blood are both used in daily work, calibration
should be done after confirming the sampling mode.
NOTE: After confirming the mode, all test should be done in the same mode.
NOTE: If any malfunction occurs during measurement, the test results are invalid. Repeat
the measurement after troubleshooting.
At main menu screen, click "Cal" to enter System Calibration screen. Choose "manual Cal",
click "OK" to enter manual calibration interface.
Input assay and values, then click "New Cal" button, the system will calculate the new
calibrated value automatically and the date will be updated simultaneously. See Figure 7-1:
NOTE: Calibration coefficient is allowed in the range of 70%~130%, if the test value exceed
the limit, this critical value will be selected as the new coefficient for calibration.
NOTE: Analyzer can calibrate a certain parameter or all parameters of WBC, RBC,
HGB, MCV, MPV, RDW_CV, RDW_SD, PLT and PDW.
CAUTION: Data will be lost if exit without pressing "OK" to save.
7.3 Auto Calibration
At main menu screen, click "Cal" to enter System Calibration screen, Choose "Auto Cal',
click "OK" to enter auto calibration interface. See Figure 7-2.
Figure 7-2 Automatic Calibration
Page 92
At Auto Calibration Mode, input assay then place the calibrator tube under the aspiration
probe, press RUN key, the analyzer starts to count and then displays the results in Value 1
to Value 4 according to the sequence of 4 counts.
Analyzer cannot count or display the test value in following conditions:
1) After counting for 5 times, press RUN key, analyzer will prompt that there is no space to
process calibration count.
2) If the precision of test result is abnormal, analyzer will prompt "data is abnormal, please
re-counting"
3) After each counting, analyzer will calculate a new calibration value according to
reference value and test result and update the calibration date.
Click "Print" to print the new calibration values.
To monitor abnormal blood sample measurement, it is essential for the operator to setup
normal ranges of the parameter according to laboratorial or clinical requirement. Prompt or
indication will be given if the test values exceed the range. The limits of 21 parameters are
discussed in this chapter, any results exceeding the range will be marked with H (High) or L
(Low). H means the results are higher than the upper limits, while L means the results are
lower than the lower limits.
CAUTION: The shift in parameter limit may cause changes in abnormal
indication of hematology index. Please confirm the necessity for changing.
8.1 Limit Review
At Limit Setting screen, operator can input proper parameter limits or use default limits.
Default limits are different depending on patient group. Figure 8-1 depicts General group
limits. Figure 8-2 depicts User 1 group limits
Click "Def”, the system prompts the operator to decide whether to recover all the default
limits. Select "Yes" to recover parameters of all groups to default limits; select "No" to exit.
See Figure 8.-3.
Click "OK" to save current default limits which will be displayed when operator enter Limit
Setting screen again.
Page 95
8.2 Limit Modification
1) At main menu screen, click “Func”, then click “Limits” to enter limit setting screen
2) Click “Group”, the screen displays the lower and upper limits of parameters in current
group.
3) Select the lower or upper limit of the parameter that operator wanted to modify,
delete the former limit by Backspace key on keyboard, then input the new lower or upper
limit.
4) Click “OK”, and then the dialog box as Figure 8-4 shows will pop up, select “Yes” or
“No” to save the modification or not.
8.3 Print
Click “Print”, then system will print out limits of all groups in list form automatically.
Page 96
CHAPTER 9 Maintenance
Routine care and regular maintenance are essential to keep optimal performance, minimize
system problems and prolong the life span. Procedures and instructions of preventive
maintenance are discussed in this chapter. More information is available at the Customer
Support Centre.
Preventive maintenance should be performed periodically. Pertinent maintenance is also
included in this Chapter according to actual requirement.
9.1 Daily Maintenance
9.2 Weekly Maintenance
9.2.1 Surface Maintenance
Clean the smudge on the surface of analyzer, especially the spilled blood on the aspiration
probe and surrounding, to remove the protein buildup or debris and reduce the possibility of
a blockage. Wipe the outside of the probe and surrounding with gauze soaked by litmusless
Page 98
detergent before cleaning other parts.
CAUTION: Never use corrosive acids, alkali or volatile organic solvent (such as acetone,
aether and chloroforms etc.) to clean the outside of the analyzer. Only litmusless detergent
is allowed.
9.3 Monthly Maintenance
9.4 System Maintenance
At main menu screen select "Func", then select "Maint" to enter the screen as
Figure 9-2.
9.4.1 Cauterize Aperture
Cauterize Aperture helps to prevent and remove aperture clog. Procedure is as follows:
1. Select "Cauterize Aperture" at Maintain screen.
2. The analyzer starts to perform the function and display the progress bar at the
bottom of the screen.
3. After completing, system will back to the Maintain screen.
9.4.2 Flush Aperture
Flush Aperture helps to prevent and remove aperture clog associating with Cauterize
Aperture. The procedure is as follows:
1 Select "Flush Aperture" in Maintain screen.
2. The analyzer starts to perform the function and display the progress bar at the
bottom of the screen.
3. After completing, system will back to the Maintain screen.
9.4.3 Drain Cups
Perform this operation to drain diluent out of WBC and RBC cups.
Page 100
9.4.4 Rinse Cups
Perform this operation to rinse the aperture to prevent blockage when counting time is too
long. The procedure is as follows:
2. The analyzer starts to perform the function and display the progress
bar at the bottom of the screen.
3. After completing, system will back to the Maintain screen.
9.4.5 Rinse Fluidics
CAUTION: Consider all clinical specimens, controls and callibrators etc. that contain human
blood or serum as potentially infectious. Wear lab coats, gloves and safety glasses and
follow required laboratorial or clinical procedures when handling these materials.
CAUTION: As probe detergent is corrosive, operator should wear lab coats, gloves and
follow required laboratory operation procedures.
Probe detergent is a kind of alkalescence detergent. Prime Fluidics is to rinse WBC and
RBC cups as well as related tubings with probe detergent. If the analyzer keep) s on working
day by day, perform Prime Fluidics every 3 days; If not, perform this operation every week.
1) Place the probe detergent container under the aspiration probe. Select "Prime Fluidics"
at Maintain screen, then the dialogue box as Figure 9-3 will pop up, select "Yes" to
aspirate the detergent, select "No" to back to Maintenance screen.
2) Remove the detergent after the probe retracting back. Analyzer starts to perform the
function and display progress bar at the bottom of the screen.
Page 101
3) After several seconds, the dialogue box as Figure 9-4 will pop up, put the probe
detergent container under the aspiration probe again then click “OK”.
9.4.6 Prime Lyse
CAUTION: Consider all clinical specimens, control and calibrator etc. that contain human
blood or serum as potentially infectious. Wear lab coats, gloves, and safety glasses and
follow required laboratorial or clinical procedures when handling thes materials
NOTE: Keep the lyse still for a certain time to ensure it stable
NOTE: After replacing diluents, detergent or lyse, perform background test to make sure the
background values are in a acceptable range.
• Replacement of lyse.
2) Analyzer starts to perform and display the progress bar at the bottom of the screen.
If the analyzer is left unused for more than 3 months or being shipped, please perform
following operations for maintenance:
1) Take out the diluent inlet tube which is connecting with diluent port on the rear panel
from waste container, discharge the remaining diluient in tube.
2) Take out the lyse inlet tube which is connecting with lyse port on the rear panel from
waste container, discharge the remaining lyse in tube.
3) Take out the detergent inlet tube which is connecting with detergent port on the rear
panel from waste container, discharge the remaining detergent in tube.
4) Store the remaining reagents according to instructions and avoid supercooling and
overheating. Operator should establish and conform to effective storage measures to
prevent reagent from degeneration, misusage or misdrinking.
5) Keep the diluent, lyse and detergent inlet tubes hanging in the air. At main menu
screen, click "Prime" several times until the top left corner of the screen present "No
Diluent", "No Lyse", "No Detergent", then Click "Prime" once again.
6) Insert diluent, lyse and detergent tubes into distilled water. At main menu screen, click
"Func"Æ"Maint"Æ"Prepare Shipping" to inter the screen as Figure 9 6 shows.
Page 103
1) After completed, take out the diluent, lyse and detergent tubes from distilled
water and click "Prepare Shipping" again to drain the reagent in tubes.
2) At main menu screen, click "Exit", "Thank you, now turn off power" will appear to
instruct the operator to turn off the power switch on the rear panel.
3) Pull out outlet tube from the rear panel, clean it with distilled water and save it with
plastic bag after dry by airing.
4) Cover the connectors of DILUENT, LYSE, DETERGENT and WASTE on the rear
panel with caps which are taken out at initial installation.
5) Disconnect the power cord of analyzer and save it in plastic bag. Place the analyzer
and components in plastic bags into the shipping box.
Page 104
CHAPTER 10 Service
Routine care and regular maintenance are essential to keep optimal performance, minimize
system problems and prolong the life span.
This chapter introduces the Service function, with which operator may monitor the system
status, valve and motor status etc.. More information is available at the Customer Support
Centre.
CAUTION: Incorrect maintenance may lead to impairment of analyzer. Please maintain the
analyzer according to this manual.
NOTE: If there is any problem which Is not discussed in the manual, please contact the the
Customer Support Centre.
10.1 System Check
Click "Func" at main menu screen, select "Sev", input "2006" in the pop-up dialog box to
enter the System Check screen.
10.1.1 System Status Check
The System Status Check screen presents the current status information for
example temperature, constant-current source voltage, 5V voltage, HGB zero voltage,
HGB background voltage, WBC aperture voltage, RBC aperture voltage etc.. See
Figure 10 -1.
Page 105
NOTE: At System Status Check screen, operator can view the temperatun vacuum etc.
that mention above, but cannot modify.
10.1.2 Valve Check
At Valve Check scre(en (see Figure 10-2), operator can check if the valves are in normal
condition.
Page 106
At Valve check screen, click valve buttons, the corresponding result will be displayed in
result list. See Fugure 10-3.
At motor Check screen, operator can check if the motors are in normal condition. At the
screen, click motor buttons, the corresponding result will be displayed in result list. See
Figure 10.4
Click “Func” at main menu screen, select “Sev”, input “6666” in the pop-up dialog box to
enter the System Log screen as Figure 10-5 shows:
Data Query:
Choose the begin date and end date on System Log Screen, then click “Rev”, the query
results will be displayed in the list box. As shown in Figure 10-6.
Event Query
On System Log screen, cancel Date optio in check box in top left corner, then check Event
query mode, select intended event in dropdown options. Click “Rev”, result will be displayed
in list as Figure 10-7 and Figure 10-8.
Page 110
1) The number of total pages and current page number will be automatically displayed
on the list box
2) If query are in a large quantity and cannot be displayed in one page, pess Pgprv and
Pgnex buttons to view the results in previous or next page.
11.1 Troubleshooting Guidance
The Troubleshooting Guidance is designed to assist the operator to identify and resolve
analyzer faults. It also gives instructions on obtaining technical assistance from the
Customer Support Centre. The first step is to understand the normal operation and
preventive maintenance of analyzer. Rich experience is essential for troub eshooting Logical
troubleshooting can be divided into 3 steps:
Hardware and software faults can only be corrected by an authorized engineer. While the
test fauIts related to sample analyses can be corr ected by operator with assistance from
the engineers.
Corrective Action means taking appropriate action to correct the fault. If operator is able
to correct faults with or without technical assistance from the engineer, time spent can
be reduced greatly.
Page 113
11.2 Obtaining Technical Assistance
If technical assistance is needed, please contact the Customer Support Centre. Refer to
Copyright and Declaration for Tel. number and fax number User should provide detailed and
clear faults description. Requirements are as follows:
a) Model;
b) Serial number and version number;
c) Description of fault and operation environment (for example, the fault happened in which
screen status);
d) Lot numbers of reagents (lyse, diluents and detergent etc.);
e) Related data and report
Familiar faults and corrective actions are also given in this Chapter. Operator can identify the
fault cause according to warning information and correct the fault follow Troubleshooting
Guide.
11.3 Troubleshooting
Familiar faults corrective action are listed as follows. If faults still cannot be corrected
according to this chapter, or more technical assistance is needed, please contact with
NeoMedica Customer Support Centre.
Page 114
11.4.1 Faults Related to Reagents
11.3.2 Faults Relate to Vacuum
11.3.4 Faults Relate to Test Results
Diluent or
detergent run
1. Check if the diluent or detergent has been run out.
out or deficient
WBC bubble 2. Check the reagent tubing connection, prevent leakage.
Reagent tubing
or RBC bubble 3. Perform Rinse Fluidic in Maintain;.
loose leads to
4. If the fault still occurs, please contact with NEOMEDICA.
leakage
Page 116
11.3.5 Faults Relate to Hardware
1.Moto connecting
wire is loose;
1. Click “Sev”, input password “2006” to enter System Check
2.Travel Optocoupler
Motor sounds screen, ensure the motor items are in a normal condition;
problem.
is abnormal 2 If fault still occurs, turn off the power, and contact with
3.Motor problem;
NEOMEDICA.
4.Moto drive circuit
problem.
11.3.6 Faults Relate to Temperature
Improper operation will lead to errors, event damage to operator or other peaple. Therfore, a
perfect criterion should be designed to the service conditions to reach the optimal
performance.
12.1 Limitations
2) All the relevant personnel for operation, shipment, installation or maintenance etc. should
strictly follow the requirements in this manual, otherwise non-standard operation may lead to
fault and user will lose the right of free service.
3) Using reagents, controls and calibrators which is not specified by NeoMedica, may lead to
faults even accident. User cannot get free service from NeoMedica in this condition.
4) Repair can be done only with permition of NeoMedica. Please, use components specified
by NeoMedica for replacement. For the problems derived from illegal operation, we will not
offer free service.
12.2 Location Limitations
2) Place the analyzer on a stable and level operating platform. Please pay attention to the
following:
• Keep away instrument from the air outlet to avoid temperature extremes.
• Keep away instrument from drying oven, centrifuge, x-ray equipment, copiers or
ultrasonic cleaner.
12.3 Personal Protection and Infection Control
1) Follow required laboratorial or clinical procedures during daily operation or
maintenance. Wear gloves, lab clothing and safety glasses to avoid direct contact with
the samples.
2) Consider all clinical specimens, controls and calibrators etc. that contain
human blood or serum as potentially infectious. Wear standard laboratorial clothing,
gloves and safety glasses and follow required laboratorial or
clinical procedures when handling these materials. Do not smoke, eat or
drink in working area. Do not suck or blow the tubing.
3) Blood samples and waste have potential biological and chemical hazard,
thus operator should handle with care. Follow relevant local regulations to
clean, dispose and discharge the waste.
4) Follow directions to store reagent, calibrators and controls. Reagents
should be kept away from temperature extremes. Customer should set up
and execute effective safekeeping measurement to prevent expired use,
deterioration, misapplication or ingestion.
CAUTION: Reagent will freeze if being stored below 0°C . Discard the reagent if iit was
frozen.
CAUTION: Keep away from direct sunlight. Seal the cap of the container and minimize
the pore size to avoid evaporation and contamination.
Page 119
APPENDIX A: Instrument Specifications
Dimension and Weight Environmental Requirements
Dimension: Temperature: 15°C~35°C
Scope of Application
Venous blood, peripheral blood of human being
Appearance Specifications
Display: 10.4-inch LCD
Language: English
Parameter: 21 parameters and 3 histograms
Indicator: Status Indicators/Work Mode Indicator!
System Alert: Alert message/Alert beep
Ports: Power Receptacle
Printer Ports
RS-232 Port
PS/2 Port
USBPorts
Recorder Specifications
Recorder Width: 4Bmm
Paper width: 57.5mm
Paper Roll Diameter: 53mm
Print Speed: 25mm/S
Page 120
Sample Volume
Whole Blood Mode for Venous Blood: Venous Blood 10 µL
Pre-diluent Mode for Peripheral Blood: Capillary Blood 20 µL
Whole Blood Mode for Peripheral Blood: Capillary Blood 10 µL
Background Results
WBC≤0.2x109/L ; RBC≤02x1012/L ; HGB≤1g/L ; PLT≤10x109/L
Carryover
WBCS≤.0.5%; RBC≤0.5%; HGB≤0.5%; HCT≤0.5%; PLT≤0.5%
Accuracy
WBC ≦±2.0%
RBC ≦±1.5%
HGB ≦±1.5%
MCV ≦±0.5%
HCT ≦±1.0%
PLT ≦±4.0%
Page 121
Precision
Linearity
Caution
Equipotentiality
Lot number
Serial number
Use by
Production Date
Manufacturer
Page 123
APPENDIX C: Communication
The system transfers sample data and analyzer information to outer computer through RS-232
COM. Communication can be done automatically after analysis or manually when the analyzer is
in idle mode. This appendix explains the settings of communication parameters and data
communication formatter for easy operation.
Before communication, please ensure the analyzer has connected with outer computer through
appropriate COM.
1. Hexadecimal Format Communication
1.1 Data Link MAC Sublayer Parameters Convention
1.2 Data Link Layer Frame Format
1.2.1 Frame Format
STX LENGTH Messege ETX LRC
1.2.2 Meaning of Fields or Control Fields
Page 124
1.2.3 Convention
Comply with Big-Enddian format, high byte is prior when transferring.
1.3 Message Field Structure
1.3.1 Message Structure
TYPE DATA
Field Definition:
Field Lenght
1 TYPE 1
2 DATA XX
TYPE Value:
Type Value
TRANS_CONDITON 0x42
1.3.2 DATA Field Definition
DATA Type (1Byte) DATA Content (depends on specific DATA Type)
If TYPE value of DATA field is TRANS_CON and the opposite party can receive 0x01
message which sent by us, it means online is normal.
Page 125
2. ASCII Format Communication
2.1 Message Transfer Format
<SB> means the start of massage and its corresponding ASCII sign is <VT>, namely
0x0B;
<EB> means the end of message and its corresponding ASCII sigh is <FS>, namely
0x1C;
<CR> means the confirmation of termination and the field mark of different
message,namely 0x0D;
ddddd is the actual transfer content. It includes several fields, each field will
end with <CR>, namely 0x0D.
2.2 Massage Grammar
| Field mark
^ Component mark
& Child component mark
~ Repeat mark
I Escape character
2.3 Data Type
2.4 Message Type
}
Page 127
Definition of MSH (Message Header):
1. Field mark ST 1
2. Encoding chars ST 4
8. Security ST 40
9. MessageType CM 7
11. Processing ID PT 3
13. Keep
14. Keep
15. Keep
16. Keep
17. Keep
Example:
MSH|^~\&|UT|UT-3O2O|LIS|PC|20100930100436||ORU^ROI|UT...I3LD|P/23.1||||||UNICODE
Page 128
Definition of PID(Patient Data) Field:
2. Patient ID EI 20
4. Alternate Patient ID CX 20
5. PatientName XPN 48
7. Date/Time of Birth TS 26
8. Sex IS 1 M or F
Example: PID|1|1010051|A1123145/15/Jame||19811011|M
Page 129
OBR Field:
5. Priority ID 2
7. ObservatiomDateTime TS 26
OBR/1/1010051|OOOOO1|UT^UT-3020|/20101010093020|2010101009350 0||||||
Jaundice|/BLD/Tom//011
Page 130
OBX Field:
4. Observation SubID ST 20
6. Units CE 90
7. References Range ST 90
15. Producer ID
ampleTime]|[StartTime]/|/|//[Symptom]||[SanpleType]/[SendDOCName]|/[Send
DP]<CR>
OBX|[ResultType]|[ValueType]/[ltemlD]/[ltemName]/[TestResult]/[Unit]/[Co
nsultValue]|[FIag]//|F|///[DocDP]/[DOCName]|<CR>
OBX/1/NM/[ItemlD]^LeftLine//[TestResult]//|/|/F////[DocDP]|[DOCName]|<C
R>
OBX|1|NM/[ltemlD]^RightLine/|[TestResult]|||//|F/|||[DOCDP]/[DOCName]/
<CR>
OBX\1/ED/[ ltemlD]//[lnstrlD]^Histogram^32Byte^HEX^[TestResult]|/||/|F/||
/[DOCDP]|[DOCName]/<CR>
<EB>
3 Communication Operations
If choose hexadecimal as transmission mode, the system will send data in hexadecimal
format. Likewise, choose ASCII, the system will send data in ASCII format.
If automatic transmission is on, then after finishing each analysis, the system will transmit
data through COM automatically. If you do not need, please choose off in setting interface.
Users can press "Trans." in main menu screen to transmit data.