"Injector not ready" displayed on X-ray equipment.
6. Operator is pressing the buttons too rapidly; system can't keep up.
2. MS plate likely jammed forward; manually reverse plate then restart injector.
3. Reseat head cable and extension cable.
4. Check MS Plate alignment (make sure it's square on the two gears).
2. MS plate likely jammed forward; manually reverse plate then restart injector.
8. Replace MS Motor.
9. Replace PL pot.
2. MS plate likely jammed forward; manually reverse plate then restart injector.
9. Replace MS motor.
5. MS plate likely jammed forward; manually reverse plate then restart injector.
6. Replace MS motor.
3. MS plate likely jammed forward; manually reverse plate then restart injector.
4. Reset PPI memory and enter correct head type and syringe size.
6. Clean and re-oil MS worm gears; ensure proper MS plate alignment on two worm gears.
3. MS plate likely jammed forward; manually reverse plate then restart injector.
4. Reset PPI memory and enter correct head type and syringe size.
1. MS plate likely jammed forward; manually reverse plate then restart injector.
2. Clean and re-oil MS worm gears; ensure proper MS plate alignment on two worm gears.
5. Replace MS motor.
3. Replace PL pot.
2. MS plate likely jammed forward; manually reverse plate then restart injector.
7. Replace MS motor.
2. Reset PPI memory and enter correct head type and syringe size.
1. Check PPI battery voltage; should be 3.0 VDC. Replace battery if below 2.7 VDC.
2. Reset PPI memory and enter correct head type and syringe size.
1. Manual motor knob rubbing on Armed Light lens cap; re-seat manual knob.
Head display will not change from 150 ml head to 200 ml head.
1. Operator not answering all the questions on the Sentinel Display before arming.
Injector not holding syringe value after being turned off; keeps resetting to 150ml syringe.
1. Check PPI battery voltage; should be 3.0 VDC. Replace battery if below 2.7 VDC.
2. Reset PPI memory and enter correct head type and syringe size.
1. Move the jumper on the Main Unit ISI card to the Z5 option.
Injector will inject from imaging system but not from injector hand switch.
1. Move the jumper on the Main Unit ISI card to the Z5 option.
2. MS plate likely jammed forward; manually reverse plate then restart injector.
5. Replace MS motor.
3. Move the jumper on the Main Unit ISI card to the Z5 option.
1. Check MS Plate alignment (make sure it's square on the two gears).
2. Clean and re-oil MS worm gears; ensure proper MS plate alignment on two worm gears.
2. Replace MS motor.
No piston movement.
6. Replace motor.
No power to display.
No power to injector.
1. MS plate likely jammed forward; manually reverse plate then restart injector.
2. Replace MS motor.
1. Replace MS motor.
2. Replace heater.
1. Check PPI battery voltage; should be 3.0 VDC. Replace battery if below 2.7 VDC.
Volume filled into the syringe does not match volume on display.