Download as pdf or txt
Download as pdf or txt
You are on page 1of 4

Heated and humidified nasal CPAP on neonatal transport_layout 25/04/2012 15:00 Page 82

AUDIT © 2012 SNL All rights reserved

Heated and humidified nasal CPAP on


neonatal transport
It is widely accepted that neonatal patients who require airway support should receive heated
and humidified gases. This can be a challenge when providing short-prong nasal continuous
positive airway pressure (CPAP) in the transport environment. This article describes how heated
and humidified gases can be provided during transport in this group of patients and describes
the results of an audit into its effectiveness.

Ian Braithwaite
RN(CH), BEng
Senior Transport Nurse, Embrace
T he importance of heating and
humidifying airway gases administered
via a mechanical ventilator has been well
transport incubator temperature or using a
chemical gel mattress. Despite these efforts
43% of babies included in the audit arrived
ian.braithwaite@sch.nhs.uk established. If inspired gases are not at their destination with an axilla
adequately humidified, the respiratory temperature of <36.6°C (FIGURE 1). After
Karen Spinks system can be compromised through the 2010 audit, the investigators sought to
RGN, RSCN, MSc Advanced Clinical Practice thickened secretions and damage to cilia obtain the necessary equipment to actively
Advanced Neonatal Nurse Practitioner,
and mucus membranes1. humidify ventilator gas during transport
Embrace
In the transport setting every effort is and to assess this change in practice.
Claire Davidson made to maintain the standards of
intensive care that the baby receives in the Current practice
MBChB, MRCPCH
Transport Registrar, Embrace nursery. Practices are constantly assessed There is a wide variety in practice
and reviewed to identify problematic areas. throughout the UK in terms of humidity
Cath Harrison An internal audit in 2010 showed that and how it can be provided in the
BMedSci, BM, BS, DTM&H, FRCPCH transferring neonates on continuous transport setting. In a survey carried out by
Lead Neonatologist for Embrace positive airway pressure (CPAP) led to a the Peninsula Transport service in 2011,
decrease in axilla temperature for some Madar found that 18 out of 21 UK
Yorkshire and Humber Infant and Children patients2. At the time Embrace had no neonatal transport teams were able to
Transport Service, Capitol Business Park,
means to actively humidify ventilator gas provide either actively or passively
Barnsley
on transport. The audit showed a trend humidified ventilator gases. Of the ten
towards a longer stabilisation time as teams services utilising the Babypac (Smiths
attempted to optimise the temperature Medical, UK) ventilator, two did not
before departure by increasing the humidify gases, five used only heat

80%

70%
Keywords
continuous positive airway pressure; 60%
Percentage of transports

humidification; neonatal transport;


50%
ventilation

Key points 40%

Braithwaite I., Spinks K., Davidson C., 30%


Harrison C. Heated and humidified nasal
CPAP on neonatal transport. Infant 2012; 20%
8(3): 82-85.
10%
1. Humidification of ventilated gases is
possible to achieve safely and
0%
effectively on transport.
Referral baby temperature Arrival baby temperature
2. Patient temperature data demonstrates
<36.6 36.6-37.2 >37.2 degrees Celsius
the benefit of using active
humidification with transport CPAP.
FIGURE 1 Mean axilla temperatures for patients on non-humidified CPAP, 2010 audit.

82 VOLUME 8 ISSU E 3 2012 infant


Heated and humidified nasal CPAP on neonatal transport_layout 25/04/2012 15:00 Page 83

AUDIT

moisture exchangers (HME) and three also Humid-Vent Mini Fisher Paykel MR850 Westmed Neo-pod T
had access to active humidity3. HME Humidifier Humidifier
Embrace use Babypac ventilators to
Weight (excluding 4 grams 2,800 grams 200 grams
provide ventilation for neonatal patients.
circuit)
The babies are nursed in Ti-500 incubators
Power Supply N/A AC DC or AC
(Draeger Medical). Additional warmth, if
required, is provided by single-use Temperature and Humidity 30mg Humidity >33mg No humidity data in
‘Transwarmer’ chemical gel mattresses humidity H2O/L at a tidal H2O/L product information.
(CooperSurgical, USA). (manufacturer’s volume of 20mL Chamber 35.5-42°C Chamber 30-38°C
For intubated patients a non-humidified data)
Airway 35-40°C
circuit (Intersurgical, UK) is used, together
with a Humid-Vent Mini (Gibeck/Teleflex, Additional circuit 2.4mL N/A N/A
Ireland) HME. An HME is a passive dead-space
humidification device which traps expired TABLE 1 Comparison of equipment for humidity.
heat and moisture in a membrane. The gas
passes back over the membrane on requires a mains power supply. The expiratory port of the ventilator was noted.
inspiration allowing some of the heat and investigators chose to use this humidifier as The volume of water was dependent on
moisture to be recaptured. HMEs in they were readily available within the base ambient conditions and incubator
neonates are not as effective as active hospital, and the circuit lengths were temperature. The experience of other
humidification, but are considered appropriate. The dedicated Embrace transport services gave reassurance that
acceptable for short-term use4. A recent ambulances have 240v AC inverters so this water does not infiltrate the ventilator
Cochrane Review showed that body power could be supplied to the humidifiers during routine use.
temperature was significantly decreased during the journey. For added fixation security, Paraid
when an HME was used compared with The Babypac ventilator is not commonly (Birmingham, UK) developed a secure
heated humidifiers5. used with humidified circuits. Work was bracket (FIGURE 2) to attach onto the
Short-prong nasal CPAP is provided undertaken with both Fisher Paykel and transport trolley that meets European
using the BabyFlow system (Draeger Smiths Medical to establish that the incubator transport standards7.
Medical, UK). These are short nasal prongs combination would be safe and effective.
which attach to the ventilator using simple The humidifier was positioned below both Completion of the 2011 audit
unheated tubing. The Humid-Vent Mini the ventilator and the patient. The Transport teams were asked to fill in an
HME is designed to attach to an chamber auto-fill line was not used to audit form for patients transferred in an
endotracheal tube and so cannot be used remove the risk associated with hanging a incubator on short-prong nasal CPAP.
in this circuit. bag of water. The ventilator was used in Planned, unplanned and time-critical
constant-flow mode. A calibrated transfers were included. Transfers from an
Equipment reference-grade pressure monitor was used operating theatre or X-ray department
There are a variety of options available to to quantify the pressure drop across the were excluded as the procedure or
transport services to actively heat and circuit due to the added flow resistance and investigation may have impacted the
humidify ventilator gases. Some ventilators this was taken into account clinically. patient’s temperature. The data collection
have an integrated humidification system, Variations in humidifier performance ran for a six-and-a-half month period
eg the Reanimator F-120 (Stephan, related to external factors such as from June 17th to December 31st 2011.
Germany). Using new ventilators for the environmental temperature have been One out of the four Embrace transport
study was not feasible so two potential described6. Despite the use of dual-limb incubator trolleys was set up with the
stand-alone humidifiers were reviewed heated circuits, condensation at the Fisher Paykel humidifier. If the humidifier
(TABLE 1).
The Neo-pod T (Westmed, UK) is a
humidifier specifically designed for the
transport environment. The water
chamber sits inside the incubator, with an
external control panel and power supply.
The equipment is light-weight and simple
to operate. The ventilator circuits were felt
to be too short for use in the study and
although the manufacturer can provide
longer circuits, they were not available
within the audit time frame.
The MR850 (Fisher Paykel, UK) is a
humidifier commonly used in nurseries in
the region and Embrace staff are familiar
and experienced with the product. It is
however bulky, relatively heavy and FIGURE 2 MR850 humidifier in bracket.

infant VOLUME 8 ISSU E 3 2012 83


Heated and humidified nasal CPAP on neonatal transport_layout 25/04/2012 15:00 Page 84

AUDIT

trolley was available at the time of referral


37
it was used for the CPAP patient, otherwise
another trolley was used and non-
humidified CPAP was provided. 36
Embrace use axilla and abdominal skin
temperature measurements as core 35

Degrees C
temperature surrogates for patients who
require a level of care consistent with
34
receiving CPAP. The skin temperature is
monitored continuously during the
journey via the incubator skin probe which 33
is fixed to the skin over the patient’s liver
with a heat-reflective sticker. An internal 32
audit found a good correlation between Referring inc temp Departure transport inc temp Arrival transport inc temp
skin temperature and axilla temperature8. Non-humidity (n=36) Humidity (n=26)
Embrace aspire towards precise
temperature control on transport and FIGURE 3 Mean incubator temperatures for patients transported on CPAP, mean values +/-
consider 36.6 to 37.2°C to be an acceptable standard error of the mean.
range of axilla temperatures. These
parameters are tighter than some incubator set-temperature on departure addition of humidification almost
published guidelines9, but provide a factor from the referring unit and the incubator eradicated hypothermia in patients
of safety and allow for measurement error. set-temperature on arrival at the transported on CPAP; only 11% had skin
Data were collected on incubator destination unit. Ten of the total number temperatures that fell below 36.6°C at any
temperature, Transwarmer use, of patients (16%) were nursed under point during the journey (TABLE 3), and no
stabilisation time and the total time the radiant warmers at the referring hospital patient arrived at the destination with an
patient spent on transport CPAP. Transport and did not count towards the initial mean axilla temperature below 36.6°C (FIGURE 5).
teams were asked to indicate if there were values. Initial incubator and patient FIGURE 3 shows that without humidity,
any adverse events, including delays related temperatures at the referring hospitals transport incubator temperatures on
to thermoregulation or equipment. were similar for both groups, as were arrival were on average around 3°C higher
stabilisation time and total time on than unit incubator temperatures (33°C).
Results transport CPAP (TABLE 2). With humidity they were around 1°C
There were 94 CPAP transports that met Most temperature-related interventions, higher. The incubator temperature for
the audit inclusion criteria and 64 audit issues and delays occurred in the non- patients who received humidity was
forms were completed (68%). Two humidified group (TABLE 3). When significantly reduced during the journey in
patients were excluded; one had missing analysed in detail the interventions were response to the patient’s skin temperature
data and one deteriorated en-route and almost entirely related to patients <1300 (Student’s T-Test p=0.03).
the team returned to the referral hospital. grams or <31 weeks’ corrected gestation.
The remaining 62 transports were There was no clear correlation with age or Discussion
analysed and split into humidified and gestation at birth. Since the disappointing results of the
non-humidified groups. FIGURES 4 and 5 show axilla temperatures original audit there has been a greater
There were similar demographics in measured at either end of the transport emphasis on achieving appropriate
both groups, although the non-humidity episode: the arrival of the transport team at temperature control within the service and
group had a slightly lower gestational age the patient’s cot-side in the referring transport teams usually set a higher
at time of transfer (TABLE 2). hospital, and at handover of the patient at transport incubator temperature for
FIGURE 3 compares incubator the destination hospital. FIGURE 4 illustrates patients receiving CPAP in expectation of
temperatures on the arrival of the an improvement in temperature control temperature-related challenges. However
transport team at the patient’s cot-side in since the original audit (FIGURE 1), even FIGURE 5 shows that 23% of patients in the
the referring hospital, the transport without humidified CPAP. However the humidified group arrived with an axilla
temperature of greater than 37.2°C and the
Non-humidity (n=36) Humidity (n=26) downwards trend of transport incubator
temperatures in this group suggests that
Gestation, mean completed weeks 28 (24-40) 28 (24-40) they were initially set too high for the
Corrected gestation, mean completed weeks 31 (26-40) 32 (27-44) patient. With the introduction of
humidified CPAP and in the light of the
Age, mean days 17 (1-52) 28 (1-58)
audit results, initial transport incubator
Weight at transport, mean grams 1360 (730-3550) 1350 (630-3600) temperature settings need to be
Stabilisation time, mean minutes 79 (34-215) 73 (30-175) reconsidered. The investigators
acknowledge that there is a group of
Time on transport CPAP, mean minutes 103 (43-185) 117 (55-215)
mature infants for whom temperature
TABLE 2 Demographics and times for stabilisation and transport, range in brackets. control is difficult because they are only

84 VOLUME 8 ISSU E 3 2012 infant


Heated and humidified nasal CPAP on neonatal transport_layout 25/04/2012 15:00 Page 85

AUDIT

Non-humidity (n=36) Humidity (n=26)


Transwarmer use 36% 8%
Skin temperature dropping below 36.6°C at any point during the journey, even transiently 42% 11%
Temperature control-related delays to transport 14% 0%

TABLE 3 A comparison of three markers of temperature control improvement.

80% 80%

70% 70%

60% 60%

Percentage of transports
Percentage of transports

50% 50%

40% 40%

30% 30%

20% 20%

10% 10%

0% 0%
Referral baby temperature Arrival baby temperature Humidity referral baby temperature Humidity arrival baby temperature
<36.6 36.6-37.2 >37.2 degrees Celsius <36.6 36.6-37.2 >37.2 degrees Celsius

FIGURE 4 Mean axilla temperatures for patients on non-humidified FIGURE 5 Mean axilla temperatures for patients on humidified CPAP,
CPAP, 2011 audit. 2011 audit.

being nursed in an incubator due to the Fewer Transwarmer mattresses were Declarations of interest
needs of the transfer process10. needed in this group yet there were no None
There were several limitations associated instances of babies arriving at the referral
with this project. The small sample size unit with sub-optimal temperatures. There References
and lack of randomisation may have were subjective reports of increased patient 1. Burton J.D. Effects of dry anaesthetic gases on resp-
reduced the quality of the results, although comfort and improved tolerance of CPAP. iratory mucous membrane. Lancet 1962;1:235-38.
2. Babarao S. Neonatal CPAP transfers in Yorkshire &
probably not the general conclusions. The The patient temperature data, together
Humber (unpublished internal audit). Embrace.
recording of temperature-related with the unseen protection given to 2010.
interventions has now been made manda- delicate airways, demonstrates the benefit 3. Madar J. Ventilator humidity on transport
tory in the Embrace transport paperwork, of using humidification during CPAP with (unpublished survey, personal communication).
and compliance will be audited. neonates during transport. 2011.
Safety was of primary importance 4. Fassassi M., Michel F., Thomachot L. et al. Airway
The results of this study showed that a
humidification with a heat and moisture exchanger
throughout the project. There were no greater number of babies in the humidified in mechanically ventilated neonates: a preliminary
adverse incidents related to the equipment. group became too warm (axilla evaluation. Intensive Care Med 2007;33:336-43.
The weight of the humidifier was offset by temperature >37.2°C) compared with the 5. Kelly M., Gillies D., Todd D.A., Lockwood C. Heated
the removal of two syringe pumps from non-humidified group. Episodes of humidification versus heat and moisture
the incubator trolley. Staff found the exchangers for ventilated adults and children.
hyperthermia may be detrimental to the
Cochrane Collaboration. 2010;1:1-90.
humidification set-up easy to use, and the neonate and should be avoided11. It is 6. Todd D.A., Boyd J., Lloyd J., John E. Inspired gas
introduction of extra equipment did not anticipated that improved experience and humidity during mechanical ventilation: effects of
impact on stabilisation times. confidence may reduce the incidence of humidification chamber, airway temperature probe
Adding new equipment to a transport environmental hyperthermia. position and environmental conditions. J Paediatr
service has a potential financial impact in Child Health. 200;37:489-94.
Extending the quality of care provided in
7. CEN Standards. Rescue systems-Transportation of
terms of consumable use. Humidifier the intensive care nursery to the transport incubators-Part 2: System requirements (draft). EN
compatible ventilator circuits are more setting would suggest that humidifiers 13976-2:2011.
expensive than comparable non-heated should also be used for ventilated patients 8. Spinks K. A comparison of skin temperature probes
versions, but this additional cost was offset as well as all infants who receive CPAP. The and axilla temperature (unpublished internal audit).
by a dramatic reduction in Transwarmer Embrace. 2009.
use of active humidification for intubated
9. Waldron S., Mackinnon R. Neonatal thermo-
mattress usage. Without considering the patients will be developed and there will be regulation. Infant 2007;3:101-04.
initial capital equipment costs, the practice audit of this practice to compare it with 10. Bowman E. Control of temperature during newborn
change was cost-neutral. passive humidification using an HME. transport: An old problem with new difficulties.
J Paediatr Child Health. 1997;33:398-401.
Conclusions Acknowledgements 11. Department of Reproductive Health and Research
(RHR), World Health Organisation. Thermal
Babies on humidified CPAP had better The authors would like to acknowledge
protection of the newborn: a practical guide
temperature control during transport with Dr John Madar and Mark Mulcahy from (WHO/RHT/MSM/97.2). Geneva: World Health
no temperature-related stabilisation delays. the Peninsula Neonatal Transport Service. Organisation. 1997.

infant VOLUME 8 ISSU E 3 2012 85

You might also like