Copia de Diag
5 Maker Maintenance
This menu is used by authorized service personnel for service, inspection, and repair work
performed on this system.
Caution
The [Maker Maintenance] submenu should be used according to the instructions provided by
Olympus Technical Services or by authorized service personnel. This submenu is used for
service, inspection, and repair work.
Help Exit
ALARMCLEAR
Description:
3. After viewing the screen, press function key F2 (exit) to return to the main menu.
Help Exit
ALARMCLEAR
Description:
This menu is used by authorized service personnel for service, inspection, and repair work
performed on this system.
2. Use your mouse to select the item to check. Explanations of each function follow this
procedure.
3. Press the STAT ROTATION/DIAG Switch on the analyzer to perform the selection.
Sample Probe
A/Reset: Aligns the sample probe in its home position.
B/Up/Down Step: Moves the sample probe 1 step vertically (up/down).
C/Rotation Step: Moves one step between the wash station and items selected. Select
the following: From (rack, cuvette, STAT table, or wash/STD table) and To (cuvette,or
ISE).
D/Up-Down Pulse: Pulses the sample probe the number of steps specified from 1 to
999.
E/Rotation Pulse: Pulses the sample probe clockwise or counterclockwise the number of
steps specified from 1 to 999.
F/Sequence: Performs a sequence according to the selection.
"Rack" samples from the rack.
"STAT" samples from the STAT table.
"Wash/STD" Samples from the wash standard position. This option can only be
selected if "To" is set to ISE.
"Cuv." samples from the STAT table.
"ISE" samples from the STAT table.
Sample Syringe
G/Reset: Aligns the sample syringe in its home position.
H/Up and Down: Moves the sample syringe 1 step vertically (up/down).
Reagent Probe
I/Reset: Aligns the reagent probe in its home position.
J/Up and Down: Moves the reagent probe 1 step vertically (up/down).
K/Rotation: Moves the reagent probe 1 step horizontally. Select the following: From
(inner, outer, or dilution bottle) and To (cuvette).
L/Up-Down Pulse: Pulses the reagent probe the number of steps specified from 1 to
999.
M/Rotation Pulse: Pulses the reagent probe clockwise or counterclockwise the number
of steps specified from 1 to 999.
N/Sequence: Performs a complete sequence including movement from the wash well,
reagent compartment to cuvette wheel.
Reagent Syringe
O/Reset: Aligns the reagent syringe in its home position.
P/Up and Down: Moves the reagent syringe 1 step (up/down).
Refrigerator (Reagent)
Q/Reset: Resets the reagent compartment.
R/Rotation: Rotates the compartment according to user specifications (CW, CCW, Step
1-76. For complete rotation select 76.)
S/Pulse: Pulses the reagent compartment clockwise or counterclockwise the number of
steps specified from 1 to 999.
All Pumps
T/On: Activates pumps.
U/Off: Deactivates pumps.
Mix
A/Reset: Aligns the mix bar unit in its home position.
B/Up and Down Step: Moves the mix unit 1 step vertically (up/down).
C/Rotation Step: Moves the mix unit horizontally to the next position.
D/Up-Down Pulse: Pulses the mix unit the number of steps specified from 1 to 999.
E/Rotation Pulse: Pulses the mix unit clockwise or counterclockwise the number of
steps specified from 1 to 999.
F/Mix Rotation: Spins the mix bars.
G/Sequence: Performs a complete sequence from mixing to moving to the next
position.
Wash Nozzle
H/Reset: Aligns the wash nozzle in its home position.
I/Up and Down Step: Moves the wash nozzle unit 1 step vertically (up/down).
J/Up-Down Pulse: Pulses the wash nozzle unit the number of steps specified from 1 to
999.
K/Sequence: Performs a wash nozzle sequence.
Rack Receiver
L/Reset: Resets the rack receiver unit.
M/Push-Pull: Rack bar moves forward and backward.
N/Stopper On-Off: Moves the stop bar forward and backward each time the STAT
ROTATION/DIAG Switch is pressed.
Cuvette Wheel
O/Reset: Aligns the cuvette wheel in its home position.
P/Rotation: Rotates the cuvette wheel according to user specifications. (Clockwise,
counterclockwise, step 1 to 88)
Q/Pulse: Pulses the cuvette wheel the number of steps specified from 1 to 999.
R/Magnet Off: Removes power to the pulse motor. To reset select "O/Reset."
Rack Collection
S/Reset: Aligns CC with CB assembly.
T/Forward: Moves forward one pitch feed at a time.
U/Back: Aligns CC with CD assembly.
ID Rotation
D/Reset: Resets the unit.
E/Step 1 (R<->S): Resets the pitch move motor at the cup position.
F/Step 2 (S<->S'): Resets the pitch move motor at the sample position.
G/Pulse: Pulses the ID rotation unit the number of steps specified from 1 to 999.
STAT
K/Reset: Aligns the STAT table in its home position.
L/Rotation: Rotates the STAT table according to user specs. (CW, CCW, step 1 to 18)
M/Pulse: Pulses the STAT table the number of steps specified from 1 to 999.
Rack Feed
N/Normal: Feeds rack(s) onto the analyzer.
O/Auto Repeat Run: Feeds repeat rack(s) onto the analyzer.
Rack Feeder
P/Reset: Resets the rack feeder.
Q/Forward: Moves rack(s) forward.
R/Back: Moves rack(s) backward.
Whole
E/Reset: Aligns all mechanical parts to their home position.
F/Magnet Off: Removes all power from pulse motors. To reset select "E/Reset."
Barcode Read
G/S. ID: Reads sample ID at cup position.
H/S. ID Movement: Performs a complete rack sequence and reads S. ID data.
I/Rack ID: The rack located in the "rack ID read position" is read.
J/Rack ID Movement: The barcode reader is activated.
K/STAT: The STAT cup data is read.
L/STAT Movement: Performs a STAT rotation and reads the S. ID data.
M/Reagent ID: Reads reagent ID in the reagent compartment.
N/Reagent ID Movement: Reads the reagent ID while moving in the reagent
compartment.
O/Rack Kind: Identifies rack type in rack ID position.
E/Drain-Bypass R/Sequence 2
D. Pot / Electrode Reset S/Sequence 3
M/Reset
F/Mix
G/Overflow Sensor
H/Thermistor
Help Exit
ALARMCLEAR
Description:
This menu is used by authorized service personnel for service, inspection, and repair work
performed on this system.
2. Using your mouse, select the item to check. Explanations of each function follow
procedures.
3. Press the STAT ROTATION/DIAG Switch on the analyzer to perform the selection.
3. The calibration begins as soon as one of the following is selected: I/Calibration for
serum, J/Calibration for urine, K/Calibration for serum +urine.
Automated Chemistry Analyzer
D-158 Chapter D Software AU400 User's Guide
August, 1998
4 . View the printout.
If acceptable ranges are exceeded, and * will appear on the right of the data.
<<SERUM>>
Mid Std. H Mid Std. L Mid
Slope Factor
Na 59.5 0.965 -37.9 -81.6 -91.9 -82.0
K 55.2 0.997 133.5 35.6 -0.8 30.7
Cl -49.9 0.947 1452.9 1504.2 1533.6 1510.2
* Voltage (Mid)
Former Latter The acceptable range between the former and latter is +/- 10
Na -83.4 -84.1
K 31.3 30.2
Cl 1507.1 1506.9
*Voltage (Std. H)
1st 2nd 3rd 4th
The acceptable range between the 3rd and 4th value is +/- 5 for all electrolytes except urine K,
which is +/- 10
*Voltage (Std. L)
See Also
For information on setting up a selectivity check, refer to the "ISE Maintenance" section
located in the Maintenance Chapter.
2. Put the rack on the transport belt so the front of the rack is up to the pitch feeder in the
sample aspiration position.
4. Analyze the data printout. The results of the 10 sequential samples will print with the
average, max, min, range, and CV results. Verify the CV is within the acceptable range.
Test Acceptable % CV
Na 0.6
K 2.0
C1 0.7
See Also
For procedures refer to the Troubleshooting Chapter "Sequential Sample Measure."
Explanation of functions:
Prime
A/Total Prime: Primes all solutions.
B/Buffer Prime: A one cycle buffer solution prime is performed.
C/Mid/Ref Prime: A one cycle standard solution prime is performed.
D/Drain-Electrode: Liquid is drained from the sample pot through the electrode for
electrode replacement. Use this function when replacing bypass tubing.
E/Drain-Bypass: Liquid is drained from the sample pot through the bypass tubing.
D. Pot/Electrode
F/Mix: Mix rod is moved for 3 seconds.
G/Overflow Sensor: Checks the overflow sensor and thermistor.
H/Thermistor: Checks thermistor voltage.
Calibration Selectivity
I/Calibration for Serum: Calibrates serum.
J/Calibration for Urine: Calibrates urine.
K/Calibration for Serum & Urine: Calibrates serum and urine.
L/Selectivity: Performs a selectivity check.
Reset
M/Reset: Aligns all ISE mechanical parts in their home positions.
Measurement
N/Rack: A single sample from a rack.
O/STAT: A single sample from position #11 on the STAT table.
P/Sequence 1 (Rack): Samples 10 times from the same sample in the rack.
Q/ Sequence 1 (STAT): Samples 10 times from the same sample in STAT table
position #11.
R/Sequence 2: Pulls mid-standard solution through the flowcell 10 times & calculates
statistics.
S/Sequence 3: Mid-standard is pulled through the flowcell 1 time & measures 10 times.
Automated Chemistry Analyzer
D-160 Chapter D Software AU400 User's Guide
August, 1998
5.5.4 Alarm Log
(M)Maintenance
(M)Maker Maintenance
(L)Alarm Log
STANDBY
? 1998/08/03 10:10
Alarm Log
ALARMCLEAR
Description:
This screen displays an alarm list of previous alarms and screens accessed on the system up
to 2,000 events. Use this menu to determine recent or common alarms (past alarms).
If an alarm is occurring at this time, corrective actions are located in the alarm help function.
To start alarm help, select the icon located at the top of the screen with the triangle and
exclamation point.
2. Use the pull-down menu or << >> buttons to select a date on which you wish to view
the alarm log.
2. Choose "Displayed Date" and "All Date." Displayed date only prints the date selected.
All date prints alarm logs for all dates.
(F)File Management
Operation Files
Description:
Instruction
If a floppy is removed from the drive while data is being saved, the data being saved or
stored in the floppy may be lost. Floppy disks for parameter storage have an “AU400
PARAM” volume label. They can be formatted using the [FD Data Management] screen of
the [Maintenance] menu.
2. Under the "Operation" column, select either "Save Files to FD" or "Load Files from FD."
4. Select function key F5 (execute) and insert the floppy. Start? Yes.
5. The selected files will revert to the background color when the operation is complete.