i-Prime series

Patient Monitoring System

VP-1000/1200 Service Manual (Ver.


1.3)
Service Manual VP-1200/1000

Characteristics and Principles of VP-1200/1000


Thank you for purchasing the VP-1200/1000 (Multi Parameter Patient Monitor)
VP-1200/1000 provides a variety of information about the patient by numeric
values and waveforms.
VP-1200/1000 has features as follows

※ Basic functions of Patient measuring


- ECG lead (I, II, III, aVR, aVL, aVF, V) select and Respiration measurement
- Arrhythmia measurement, ST Level analysis and Pacemaker Detect function
- SpO2 and Pulse Rate measurement
- NIBP, 2 IBP measurements
- Body Temperature (2 Temp)

※ Optional functions of Patient measuring


- ECG 12 lead (I, II, III, aVR, aVL, aVF, V1 ~ V6) select
- ICO (Cadiac Output)
- Blood Pressure (4 IBP)
- Body Temperature (4 Temp)
- End-tidal partial pressure respiration (EtCO2)
- Multi Gas measurement
- BIS (Bispectal Index) measurement

※ Etc
- Trend Data Saving for 7 days
- 20 Event Data Saving
- VGA (Display Output) Function
- Built-in 3 channel Printer. (Option)
- Touch Screen (Option)
- Trend Data-saving using SD card (Option)
- Drug Dose Calculation Function
- ECG Recall Function
- 2hr Mini Trend View

2
Service Manual VP-1200/1000

- Color change (Wave & Measurement value)


- NIBP STAT Function
- NIBP VENOUS STAT Function
- OXY-CRG View
- Patient information
- Temp TD

If you buy SD Package which is supported by VOTEM, you can upgrade it


easily and you can get more services.

3
Service Manual VP-1200/1000

Contents

◆ Composition of VP-1200/1000 ------ 5 ➢ MAIN ----------------------- 47

◆ How to set VP-1200/1000 --------- 6 ➢ ECG ---------------------- 51

◆ Overview of VP-1200/1000 -------- 7 ➢ SpO2 ---------------------- 55

◆ Components of VP-1200/1000 ------ 8 ➢ NIBP ---------------------- 59

➢ Front & Operation panel --------- 8 ➢ POWER -------------------- 62

➢ Right side -------------------- 9 ◆ Appendix ---------------------- 65

➢ Left side --------------------- 10 ➢ I. 3CDaemon Installation Method -- 65

➢ Back side -------------------- 11 ➢ II. Tera Term Installation Method ----- 69

◆ Description of product symbols ------ 12 ➢ III. Boot Loader Upgrade method ---- 73

◆ Explanation of screen ------------ 13 ➢ IV. F/W Upgrade by RS-232C with PC 85

◆ Product Specifications ------------ 16 ➢ V. F/W Upgrade by SD CARD ------- 95

◆ Safety ------------------------- 23 ➢ Ⅵ. PRINTER Supplement----------- 99

◆ System Diagram----------------- 29 ➢ Ⅶ. ADAPTER Replacement --------- 109

➢ Overall System Block Diagram ----- 29 ➢ Ⅷ. ECG, SpO2 B/D Replacement---- 118

➢ Main B/D ------------------- 30 ➢ Ⅸ. Main Board Replacement-------- 129

➢ ECG B/D ------------------- 31 ➢ Ⅹ. EtCO2 Replacement---------- 133

➢ SpO2 B/D ------------------- 32

➢ NIBP B/D ------------------- 33

➢ POWER B/D ----------------- 34

◆ TRUBLE SHOOING -------------- 35

➢ ECG ------------------------ 36

➢ RESP ---------------------- 38

➢ SpO2 ----------------------- 39

➢ IBP ------------------------ 41

➢ TEMP --------------------- 43

➢ NIBP ----------------------- 44

➢ POWER --------------------- 45

◆ PART LIST -------------------- 47

4
Service Manual VP-1200/1000

Kindly check all the items inside


Composition of VP-1200/1000 the package before installation.

Standard Accessories

ECG Cable
Main 1 set
5 lead type (1EA)

SpO2 Finger Probe 3M Electrode


(1EA) (5 PCS/1set)

Operation Manual
NIBP Hose (1EA)
(1EA)

Cuff for adult (1EA)

Power Cord (1EA)

⚫ Optional Accessories

- ECG Cable 3 lead type - TROLLEY (=CART)


- EKG Cable 5 lead type - Wall Mount
- SpO2 Extension Cable - Disposable SpO2 Sensor
- Press Measuring Kit - CUFF for Child and Neonate
- Printer Module - Temperature Sensor
- Roll Paper(18M) - Ground Cable
- ICO Cable - Multi Gas Kit set
- EtCO2 Kit set - Multi Gas Extension Cable
- (EtCO2 when operating) - BIS Extension Cable
- 3M Electrode (50 PCS/1SET) - EEG Cable
- Battery

5
Service Manual VP-1200/1000

How to set VP-1200/1000

◆ Precautions for Settings


Pay attention when setting VP-1200/1000 for the following points
➢ Use VP-1200/1000 under conditions of 10℃∼35℃ of circumstantial
temperature and 30%∼85% humidity.
➢ Check the connection status of the power cord.
➢ Do not connect multiple cords to the power supply.
➢ Place the main body on a flat area.
➢ If noise occurs, ground the equipment further.
➢ Do no use any electric cords generating connecting noise.
➢ Take care since the product can be broken by mechanical shock.
➢ Get rid of dust or inflammable material near the product when setting it up.

◆ Check points Before Use


Check the following points before measuring the patient’s condition.
- Make sure there is no mechanical risk.
- Check the lead, power plug, and accessories connected on the outside.

◆ Maintenance and Cleaning After Use


➢ Keep the product clean by cleaning it using a soft cloth at least once a
month. Do not use thinner, ethylene, oxidant or lacquer since these may
damage the product.
➢ Keep the accessories clean of dust or foreign materials. For cleaning, use a
soft cloth and clinical alcohol.
➢ Do not soak the accessories in liquid or detergents. Keep the product and
accessories out of any liquid.

6
Service Manual VP-1200/1000

Overview of VP-1200/1000

◆ VP-1200 (12.1 inch)

◆ VP-1000 (10.4 inch)

7
Service Manual VP-1200/1000

Components of VP-1200/1000

◆ Front & Operation Panel

● 2 ●
1 ● 3 ●
4 ●
6 ● 8 ●
7 ● 9

5

(1) Power Button (6) FREEZE Button


(2) ALARM ON/OFF Button (7) NIBP START/STOP Button
(3) EVENT SAVE Button (8) PRINT START/STOP Button
(4) SCREEN Button (9) EXIT Button
(5) MENU Button

8
Service Manual VP-1200/1000

Components of VP-1200/1000

◆ Right Side

1 ●
● 2 ●
3


4

(1) PRINT FEED BUTTON (3) PRINT POWER LED


(2) PRINT ERROR LED (4) SPEAKER

9
Service Manual VP-1200/1000

Components of VP-1200/1000

◆ Left Side


16

<Microstream EtCo2> < EtCO2 / Multi Gas>

(1) NIBP Connection Terminal (9) CO Connection Terminal (Option)

(2) SpO2 Connection Terminal (10) CO2 Output Connection Terminal (Option)

(3) EKG Connection Terminal(Option) (11) CO2 Input Connection Terminal (Option)

(4) ECG/RESP Connection Terminal (12) TEMP 4 Connection Terminal (Option)

(5) IBP2 Connection Terminal (13) TEMP 3 Connection Terminal (Option)

(6) IBP1 Connection Terminal (14) TEMP 2 Connection Terminal

(7) IBP3 Connection Terminal (Option) (15) TEMP 1 Connection Terminal


(16) Sidestream & Mainstream CO2 Input or
(8) IBP4 Connection Terminal (Option)
Multi Gas Input Connection Terminal (Option)

10
Service Manual VP-1200/1000

Components of VP-1200/1000

◆ Back Side


1


3 ●
4 ●
5 ●
6 ●
7 ●
8

9


10


2 ●
11

(1) Portable Handle (7) LAN Port


(2) MAIN Power Switch (8) Input DC12V, 6A (Option)
(3) Ground Port (9) IBP Signal Output Port (Option)
(4) AC Power Input Port (10) ECG Output Port (Option)
(5) VGA Port (Option) (11) SD Card Slot
(6) RS232 Port

11
Service Manual VP-1200/1000

Description of Product Symbols

Symbol Description Symbol Description

Type CF Product,
Power Switch
Defibrillation protected

Type BF Product,
Alarm On/Off
Defibrillation protected

Type BF Product Alarm Event

Attention, Warning consult


Screen Change(Screen)
Accompanying documents

IPX 7 Liquid Protection Grade Stop Waveforms (Freeze)

Non Invasive Blood Pressure


TEMP 1~4 Temperature
Manual Measurement(NIBP)

CO2 Input
Gas Input Start Printing (Printer)

CO2 Exhaust
Gas output Initial Screen (Exit)

IBP 1~4 Invasive Blood Pressure Ground

ECG/RESP Electrocardiograph / Respiration COM1 Use for Program Upgrade

EKG 12 leads Electrocardiograph VGA Screen Transfer

NIBP Non Invasive Blood Pressure LAN Use for Central Monitoring

CO Cadiac Output ECG Out

12
Service Manual VP-1200/1000

Explanation of screen (EtCo2 Optional screen)

<EtCo2 Optional Screen>


: EtCO2 Expiratory : NIBP PR Measurement
1 : Bed Number 11 21
Measurement
: EtCO2 Inspiratory : NIBP Systolic
2 : Function Icon 12 22
Measurement
3 : Default set 13 : Respiration Measurement 23 : NIBP Diastolic

4 : Date & Time 14 : IBP PR Measurement 24 : NIBP Mean

5 : Heart Rate Measurement 15 : IBP Mean 25 : Temp Measurement

6 : Heart Rate Alarm Limits 16 : IBP Systolic 26 : EtCO2 Wave


: Arrhythmia, : IBP Diastolic : RESP Wave
7 17 27
Pacemaker Detect
: NIBP before
8 : ST Segments Detect 18 28 : IBP Wave
Measurement

9 : SpO2 Measurement 19 : NIBP Auto Interval 29 : SpO2 Wave


: NIBP Measurement end
10 : SpO2 Alarm Limits 20 30 : ECG Wave
time

13
Service Manual VP-1200/1000

Explanation of screen (Multi Gas Optional screen)

<Multi Gas Optional Screen>


1 : Bed Number 12 : EtCO2 Measurement 23 : NIBP Mean

2 : Function Icon 13 : Multi Gas Measurement 24 : Temp Measurement

3 : Default set 14 : IBP PR Measurement 25 : IBP 1 Wave

4 : Date & Time 15 : IBP Mean 26 : IBP 2 Wave

5 : Heart Rate Measurement 16 : IBP Systolic 27 : EtCo2 Wave

6 : Heart Rate Alarm Limits 17 : IBP Diastolic 28 : Resp Wave


: Arrhythmia, : NIBP Auto Interval time : Multi Gas 2 or O2 Wave
7 18 29
Pacemaker Detect
: NIBP Measurement end
8 : ST Segments Detect 19 30 : Multi Gas 1 Wave
time

9 : SpO2 Measurement 20 : NIBP PR Measurement 31 : SpO2 Wave

10 : SpO2 Alarm Limits 21 : NIBP Systolic 32 : ECG Wave

11 : Respiration Measurement 22 : NIBP Diastolic

14
Service Manual VP-1200/1000

Explanation of screen -BIS(Bispectal Index) Optional screen

<BIS Optional Screen>


1 : Bed Number 12 : EtCO2 Measurement 23 : NIBP Mean

2 : Function Icon 13 : Multi Gas Measurement 24 : Temp Measurement

3 : Default set 14 : BIS Measurement 25 : BIS Measurement Wave

4 : Date & Time 15 : IBP Mean 26 : EEG Measurement Wave

5 : Heart Rate Measurement 16 : IBP Systolic 27 : IBP 2 Wave

6 : Heart Rate Alarm Limits 17 : IBP Diastolic 28 : IBP 1 Wave


: Arrhythmia, : NIBP Auto Interval time : Multi Gas 2 or O2 Wave
7 18 29
Pacemaker Detect
8 : ST Segments Detect 19 : NIBP Measurement end time 30 : Multi Gas 1 Wave

9 : SpO2 Measurement 20 : NIBP PR Measurement 31 : EtCo2 Wave

10 : SpO2 Alarm Limits 21 : NIBP Systolic 32 : ECG/SpO2/Resp Wave

11 : Respiration Measurement 22 : NIBP Diastolic

15
Service Manual VP-1200/1000

You can check setting information on the


Explanation of screen LCD screen.

◆ VP-1200/1000 provides 6 different screens to use “Screen” button.

1. Main screen 2. ECG 7 leads Screen


(Use for ECG 5 leads cable)

3. ESN Screen 4. Numeric Screen

5. Trend Screen 6. Event Screen

◼ Main Screen: Display Full parameter


◼ ECG 7 leads Screen: Display 7 Lead (I, II, III, aVR, aVL, aVF, V) waveform
◼ ESN Screen: Display ECG, SpO2, NIBP, Resp, Temp parameter
NOTE ◼ Trend Screen: Display Trend Dave Data
◼ Event Screen: Display event save 20 Data of Alarm

16
Service Manual VP-1200/1000

You can check setting information on the


Explanation of screen LCD screen.

ICON Meaning How to display

ECG Lead Fault & Lead off Flickers at 1-second interval

Arrhythmia Detect Flickers at 1-second interval

ST Detect Fixed

Pacemaker Detect Flickers at 1-second interval

No Finger & No Sensor Flickers at 1-second interval

NIBP Measuring Flickers at 1-second interval

E rr
NIBP Measurement Error Flickers at 1-second interval

O VP
NIBP Over Pressure Flickers at 1-second interval

NIBP Auto pressurization Fixed

IBP1 Lead Fault Flickers at 1-second interval

IBP2 Lead Fault Flickers at 1-second interval

Temp1 Lead Fault Flickers at 1-second interval

Temp2 Lead Fault Flickers at 1-second interval

During the save : Fixed


EVENT Save If the saving data is existed : Flickers at 1-
second interval

Alarm OFF Fixed when Alarm OFF is set.

Alarm Suspend : Automatically


Flickers at 1-second interval
cancelled after 1~5 minutes.

Printing Flickers at 1-second interval

Waveform Stop Flickers at 1-second interval

AC in use Fixed

Battery in use Fixed

17
Service Manual VP-1200/1000

Product Specifications

◼ Measurement parameters
- 3ch ECG, NIBP, SpO2, Respiration, 4 IBP, 4 Temp, EtCO2, ICO
- PROCESSING PARAMETERS:
HR, NIBP(Systolic/Diastolic/Mean pressure), SpO2 Saturation,
Pulse rate, Respiration, IBP, Temp, EtCO2, ICO, Multi Gas, BIS

◼ Monitor Performances specifications


- TFT Color LCD Display
- Waveform Display Method: Scroll
- Display Size: VP-1200(12.1 inch) / VP-1000(10.4 inch)
- Resolution : VGA (800Х600)
- Sweep Speed : 6.25, 12.5, 25, 50mm/sec
- Display waveform Time: 2.5 sec
- Waveform Freeze: Function of waveform freeze
- Waveform Display: 3ch ECG, SpO2, 4 IBP, Respiration or EtCO2
- Waveform Traces: 9 Traces
- Measurement Value Display: Heart Rate, SpO2, Pulse Rate, NIBP,
4 IBP, 4 Temp, EtCO2, ICO, Multi Gas, BIS
◼ ECG Measurement
- ECG Wave: 3 leads (STD: I, II, III) – 1 Ch
5 leads (OPT: I, II, III, aVR, aVL, aVF, V) – 3 Ch / 7 Ch
10 leads (OPT: I, II, III, aVR, aVL, aVF, V1~V6) – 12 Ch
- Input: Installed the circuit for protecting from the high frequency
electric potential of ESU, Defibrillator and so on.
- Frequency Filter : High frequency noise remove filter / Install the
50/60Hz power noise remove filter
- Input Impedance: Greater than 5 M ohm(50/60)
- Input for Initial Electric currency: Below 50 μA
- CMRR: Greater than 120dB
- Frequency Bandwidth:
0.5 ~ 40 Hz, 0.5Hz ~ 80Hz, 0.05Hz ~ 40Hz, 0.05Hz ~ 80Hz
- Gain: 2.5, 5, 10, 20, 30, 40mm/mV, Auto
- Calibration Waveform: 1mVp-p old waveform, ±10%
- Input Range: Maximum Input Range: ± 5mV

18
Service Manual VP-1200/1000

- DC OFFSET: -300mV to 300mV


- ARRHYHMIA Detect (TAC, BRD, PVC, VTAC, ASY, BGM, TGM, VENT,
VFIB, CPT, TPT, MIB, R on T)
- ST Analysis range: -9.9mm ~ +9.9mm, Resolution: 0.1mm
- HR Calculation: 4 ~ 16 wave
- Defibrilator Sync
- QRS WIDTH: 100, 200, 300, 400, 600, 800ms, AUTO

◼ Heart Rate
- Measurement Range: 0 ~ 300bpm
- Accuracy: ± 2bpm (Measurement value)
- Alarm Range : High alarm limit : 0 ~ 300, OFF
Low alarm limit : OFF, 0 ~ 300

◼ SpO2 (Pulse Oximetery)


- Measurement Range: 0 ~ 100%
- Accuracy : 70 ~ 100% SpO2 : ± 2%, 50 ~ 69% SpO2 : ± 3%
20 ~ 49% SpO2 : Unspecified
- Alarm Range : High limit - 1 ~ 100% OFF
Low limit - OFF, 0 ~ 99%
- Pulse Range : 0 ~ 300bpm
- Pulse Measurement Accuracy : 0 ~ 240bpm ±2 bpm
241 ~ 300bpm ±3 bpm

◼ NIBP (Non-invasive Blood Pressure Measurement)


- Measurement Method: Oscillometric, Automatic measurement
- Display Value: Systolic, Diastolic, Mean Pressure
- Measurement Range: Adult – 0 to 300mmHg
Neonate – 0 to 150mmHg
- Accuracy : ±8 mmHg(SYSTOLIC, DIASTOLIC), ±5 mmHg (MEAN)
- Inflation pressure MAX : Adult – 300mmHg + 5 mmHg
Neonate – 150mmHg + 5 mmHg
- Alarm Range
Display High Low
NIBP Systolic
[mmHg] Diastolic 5 ~ 275, OFF OFF, 0 ~ 275
Mean

19
Service Manual VP-1200/1000

Auto Pressure : Manual, Automatic (1, 2, 3, 4, 5, 10, 15, 20, 30 min., 1, 2, 4,


8 ,12 hours interval)
- NIBP STAT : 5 ~ 15 min
- NIBP VENOUS STAT : 50 ~ 200 mmHg
- NIBP MAX LIMIT : 70 ~ 300 mmHg

◼ Respiration
- Measurement Range: 0 ~ 200 bpm
- Accuracy: ±2% or ±2 bpm
- Sweep Speed: 6.25, 12.5, 25, 50 mm/sec
- Waveform Amplitude: x0.25, x0.5, x1, x2, x3, x4, AUTO
- Apnea Alarm Range: OFF, 10 ~ 40 sec

◼ IBP (Invasive Blood Pressure) - Optional


- Measurement Range: -50 ~ 350mmHg
- Channel: 2 channels (Option: 4 channels)
- Accuracy: ± 1%mmHg
- Auto zero balancing range: ±200mmHg
- Alarm Range
Display High Low
IBP 1/2 Systolic
[mmHg] Diastolic -50 ~ 350, OFF OFF, -50 ~ 350
Mean

◼ TEMP - Optional
- Measurement Range: 0 ~ 50℃
- Accuracy: 0℃ ~ 24.9℃ ± 0.2℃, 25.0℃ ~ 50.0℃ ± 0.1℃
- Alarm Range: High alarm limit - 0.1℃ ~ 50℃, OFF
Low alarm limit - OFF, 0.1℃ ~ 50℃

◼ EtCO2 (End-Tidal CO2) – Optional


⚫ Microstream
- Measurement Mode: Microstream
- Range: 0 ~ 99mmHg
- Temperature: 0~50℃

20
Service Manual VP-1200/1000

- Accuracy:
Time * CO2 Density ** Accuracy
0 ~ 20 0~38mmHg ±4mmHg
min 39~99mmHg ±12%
0~38mmHg ±2mHg
After 20 ±5% of reading + 0.08% for
in 39~99mmHg every 1 mmHg(above
38mmHg)

* (760 mmHg) level


** It is accuracy for greater than 80bpm of respiration. In case of greater
than 80bpm, EN864/ISO9918 (4mmHg or ±12%) standard
accuracy is same as greater than 18mmHg of EtCO2 same.
- Respiration Measurement Range: 0 ~ 150bpm
- Respiration Accuracy: 0 ~ 70bpm ±1bpm
71 ~ 120bpm ±2bpm
121 ~ 150bpm ±3bpm
- Initial Time: Self test for 30 seconds after turning on the device.
- Sampling Value: 50ml/min ±7.5ml/min
- Period of calibration: 1 year after operating or operating for 4,000 hours
- Response Time: 2.9 seconds after measurement (including rising time
and delay time)
; Rising Time – 190m sec , Delay Time – 2.7 sec

⚫ Mainstream
- Initial time : less than 15 seconds (25℃)
- CO2 measurement range : 0 ~ 150mmHg, 0 ~ 19.7%,
0 ~ 20 kPa(at 760mmHg)
- CO2 resolution : 0 ~ 69 mmHg → 0.1mmHg
70 ~ 150 mmHg → 0.25mmHg
- CO2 accuracy : 0 – 40 mmHg → ±2 mmHg
41 – 70 mmHg → ±5% of reading
71 – 100 mmHg → ±8% of reading
101 – 150 mmHg → ±10% of reading
- Respiration Measurement Range : 0 ~ 150 BPM
- Respiration Accuracy : ±1 BPM
- Required voltage : +0.5 VDC, 1.125 W (Max 1.5 W)
- Temperature/Humidity: operating environment → 0 ~ 45℃, 10 ~ 90% RH.
custody environment → -40 ~ 70℃, < 90% RH.
21
Service Manual VP-1200/1000

- Waterproof : IPX4 – Splash-proof (sensor head only)

⚫ Sidestream
- Initial time : less than 20 seconds (25℃)
- CO2 measurement range : 0 ~ 150mmHg, 0 ~ 19.7%,
0 ~ 20 kPa (at 760mmHg)
- CO2 resolution : 0 ~ 69 mmHg → 0.1mmHg
70 ~ 150 mmHg → 0.25mmHg
- CO2 accuracy : 0 – 40 mmHg → ±2 mmHg
41 – 70 mmHg → ±5% of reading
71 – 100 mmHg → ±8% of reading
101 – 150 mmHg → ±10% of reading
More than 80 BPM ±12%
- Respiration Measurement Range : 2 ~ 150 BPM
- Respiration Accuracy : ±1 BPM
- Required voltage : +0.5 VDC ±5%
- Temperature/Humidity: operating environment → 0 ~ 40℃, 10 to 90% RH.
custody environment → -40 ~ 70℃, < 90% RH.
- Waterproof : IPX4 – Splash-proof (When sample cell is inserted in sample
cell receptacle)

◼ Cardiac Output - Optional


- Measurement Method : Thermodilution
- Measurement range : 0.1 ~ 20 L/min
- Repeatability : 2% or 0.2 L/min
- Injectate temperature range : 0 ~ 27 ℃
- Blood temperature range : 30 ~ 40 ℃
- Injectate volume: 3, 5, 10cc
- Catheter : Swan-Ganz Standard Thermodilution
Pulmonary Artery Catheter
(Edward’s Life Sciences)

22
Service Manual VP-1200/1000

◼ Multi-Gas Concentration Measurement (Multi-Gas Agent) - Optional

⚫ Common Specification (Mainstream and Sidestream)


- Operation Environment : at. pressure 525~1200hPa, temp 10~40℃,
humidity 10~95%RH
- Custody Environment : at. pressure 500~1200hPa, temp -20~75℃,
humidity 5~100%RH
- Measurement range and accuracy : (standard : 22±5℃, 1013±40hPa)
type range accuracy
0 ~ 10% ±(0.2 vol% + 2% of reading)
CO2 10 ~ 15% ±(0.3 vol% + 2% of reading)
15 ~ 25% Unspecified
N2O 0 ~ 100% ±(2 vol% + 2% of reading)
0 ~ 8% ±(0.15 vol% + 5% of reading)
HAL, ISO, ENF
8 ~ 25% Unspecified
0 ~ 10% ±(0.15 vol% + 5% of reading)
SEV
10 ~ 25% Unspecified
0 ~ 22% ±(0.15 vol% + 5% of reading)
DES
22 ~ 25% Unspecified
O2
0 ~ 100% ±(1 vol% + 2% of reading)
(Sidestream Only)
- Respiration measurement range : 0 ~ 150bpm
- Respiration accuracy : 0 ~ 70bpm ±1bpm
71 ~ 120bpm ±2bpm
121 ~ 150bpm ±3bpm
- Sampling volume : 50ml/min ±7.5ml/min
- Period of Calibration : one year after operating or operating for 4,000
hours.
- Response time : within 2.9 seconds after measurement ( including rising
time and delay time)
; rising time – 190 sec , delay time – 2.7 sec
- interface : RS-232 serial interface(9600bps)

⚫ Mainstream
- Size(WⅹDⅹH) : 38ⅹ37ⅹ34 mm
- Weight : 25g( not including cable )
- Initial time : within 20 seconds (25℃)
- Surface temp : 50℃ max(at ambient temp. 23℃)

23
Service Manual VP-1200/1000

- Required voltage : 5±0.5 VDC, 1.4 W max


- Airway Adaptors
; Disposable Adult/Pediatric :
Adds less than 6ml deadspace
Pressure drop less than 0.3cm H2O @ 30 LPM
; Disposable Infant :
Adds less than 1ml deadspace
Pressure drop less than 1.3cm H2O @ 10 LPM

⚫ Sidestream
- Size(WⅹDⅹH) : 33ⅹ78ⅹ49 mm
- Weight : 130g( including cable )
- Initial time : within 20 seconds (25℃)
- Required voltage : 5±0.5 VDC, 1.6 W(2.0 W peak)
- Water Handling : Sampling line with proprietary water removal tubing
- Sampling Lines : 2 ± 0.1 m and 3 ± 0.1 m versions
- Sampling Flow Rate : 50 ± 10 ml/mm

◼ Printer - Optional
- Print Method: Thermal
- Type: Internal
- Print Mode
a. BOTH (Max 3 Wave) + Trend
b. Wave (Max 3 Wave)
c. Trend
- Paper Width: 57mm
- Valid Width: 120 ~ 300mm (Change to 25mm step)

◼ TREND
- Parameter: Heart Rate or Pulse Rate, SpO2, NIBP, IBP, Temp,
ST Segment, EtCO2, Multi gas, BIS, ICO
- 7 Days trend data save

◼ Interface
- RS-232C port
- LAN Port (Central Monitor)
- VGA Port (Display Transferring)
- SD CARD
- ECG Output port

24
Service Manual VP-1200/1000

◼ Physical Specifications
- Size: 290 x 175 x 280 mm (W x D x H)
- Weight: 7.0kg (with battery)
- Power Requirement: 1.2A, 100-240V~, 50/60Hz
- Power consumption: 80VA (MAX)
- BATTERY (1 Cell): 11.1V 4400mAh
; Recharge Time – 9 hours, Discharge Time – 120minutes

◼ Operating Environment
- Temperature: (Operating: 10°C ~ 35°C / Storage: -10°C ~ 50°C)
- Humidity: (Operating: 30 ~ 85% / Storage: 20 ~ 95%)
- Atmospheric Pressure: 700 to 1060 hPa

◼ Standard Accessories
- Patient Cable (5 leads) 1 ea
- NIBP cuff for adult 1 ea
- NIBP Hose 1 ea
- Power Code 1 ea
- Electrode (5pcs/set) 1 set
- SpO2 Sensor (Reusable) 1 ea
- Operation Manual 1 ea

◼ Optional Accessories
- ECG 3 leads Cable - Disposable SpO2 Sensor
- EKG 5 leads Cable - BIS Extension Cable
- SpO2 Extension Cable - EEG Cable
- Press Measuring Kit
- Printer Module
- Roll Paper(18M)
- Ground Cable
- NIBP cuff for neonate, Child
- Temperature sensor
- EtCO2 Kit set (EtCO2 when operating)
- Multi Gas Kit set
- Multi Gas Extension Cable
- ICO Cable
- 3M Electrode (50 PCS/1SET)
- Battery 11.1V, 4400mAh
- Trolley (=Cart)
- Wall Mount
25
Service Manual VP-1200/1000

Safety

◼ Use only accessories supplied or recommended by Votem. Use accessories


according to the manufacture’s directions for use and your facility’s standards. The
use of accessories, transducers, and cables other than those specified may result
WARNING
in increased emission or decreased immunity of the VP-1200/1000 monitor.

◼ In case where EMI is causing disturbances, it may be necessary to relocate this

CAUTION system

◼ EMI

Although this system has been manufactured in compliance with existing


EMI(Electromagnetic Interface) requirements, use of this system in the
presence of an electromagnetic field can cause momentary degradation of the
vital Signal.
If this occurs often, Votem suggests a review of the environment in which the
system is begin use, to identity possible sources of radiated emissions. These
emissions could be from other electrical devices used within the same room or
an adjacent room. Communication devices such as cellular phones and pagers
can cause these emissions. The existence of radios, TVs, or microwave
transmission equipment nearby can also cause interface.

◼ EMC

A The testing for EMC(Electromagnetic Compatibility) of this system has been


performed according to the international standard for EMC with medical
device(IEC60601-1-2). This IEC standard was adopted in Europe as the
European norm(EN60601-1-2).

26
Service Manual VP-1200/1000

Safety

The VP-1200/1000 monitor is intended for use in the specified electromagnetic


environment. The customer or the user of the VP-1200/1000 should assure that it is
used in such an environment as described below;

⚫ Electromagnetic emission
Emissions Compliance Electromagnetic environment - guidance
The VP-1200/1000 uses RF energy only for its internal
RF emissions function. Therefore, its RF emissions are very low and
Group 1
CISPR 11 are not likely to cause any interference in nearby
electronic equipment.

RF emissions
Class B
CISPR 11

Harmonic emissions
Class A
IEC 61000-3-2

Voltage fluctuations /
Flicker emissions Complies
IEC 61000-3-3

⚫ Electromagnetic immunity
IEC 60601 Electromagnetic
Immunity test Compliance level
Test level environment-guidance
Electrostatic ±6kV Contact ±6kV Contact Floors should be wood,
Discharge (ESD) concrete or ceramic tile. If floors
±8kV air ±8kV air are covered with synthetic
IEC 61000-4-2 material, the relative humidity
should be at least 30%.

Electrical fast ±2kV for power ±2kV for power Mains power quality should be
Transient/burst supply lines supply lines that of a typical commercial or
±1kV for ±1kV for hospital environment.
IEC 61000-4-4 Input/output lines Input/output lines

27
Service Manual VP-1200/1000

IEC 60601 Electromagnetic


Immunity test Compliance level
Test level environment-guidance
Surge ±1kV differential ±1kV differential Mains power quality should be
mode mode that of a typical commercial or
IEC 61000-4-5 ±2kV common hospital environment.
mode
Voltage dips, short <5% UT <5% UT Mains power quality should be
Interruptions and (>95% dip in UT) (>95% dip in UT) that of a typical commercial or
Voltage variations for 0.5cycle for 0.5cycle hospital environment. If the user
On power supply of the VP-1200/1000 monitor
Input lines 40% UT 40% UT image intensifier requires
(60% dip in UT) (60% dip in UT) continued operation during
IEC 61000-4-11 for 5 cycle for 5 cycle power mains interruptions, it is
recommended that the VP-
70% UT 70% UT 1200/1000 monitor image
(30% dip in UT) (30% dip in UT) intensifier be powered from an
for 25 cycle for 25 cycle uninterruptible power supply or a
battery.
<5% UT <5% UT
(<95% dip in UT) (<95% dip in UT)
for 5s for 5s
Power frequency 3A/m 3A/m Power frequency magnetic fields
(50/60 Hz) should be at levels
magnetic field characteristic of a typical
location in a typical commercial
IEC 61000-4-8 or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.

28
Service Manual VP-1200/1000

IEC 60601 Compliance


Immunity test Electromagnetic environment-guidance
Test level level
Conducted RF 3 Vrms 3 Vrms Portable and mobile RF communications equipment
IEC 61000-4-6 150kHz to 150kHz to should be used no closer to any part or the VP-
80MHz 80MHz 1200/1000, including cables, than the recommended
separation distance calculated form the equation
applicable to the frequency of the transmitter.

Recommended separation distance


Radiated RF 3V/m 3V/m
IEC 61000-4-3 80MHz to 80MHz to
2.5GHz 2.5GHz
80 MHz to 800 MHz

3V/m 3V/m
80MHz to 80MHz to 800 MHz to 2.5 GHz
2.5GHz 2.5GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).

Field strengths from fixed RF transmitters, as deter-


a
mined by an electromagnetic site survey, should be
less than the compliance level in each frequency
b
range.

Interference may occur in the vicinity of equipment


marked with the following symbol:

NOTE 1) At 80MHz and 800 MHz, the higher frequency range applies.
NOTE 2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by
Absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/ cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the VP-1200/1000 is used exceeds the applicable RF compliance
level above, the VP-1200/1000 should be observed to verify normal operation. If abnormal perform-
ance is observed, additional measures may be necessary, such as re-orienting or relocating the
VP-1200/1000.
b Over the frequency range 150kHz to 80 MHz, field strengths should be less than [V 1] V/m.

29
Service Manual VP-1200/1000

⚫ Recommended Separation Distance


Recommended separation distance between portable and mobile RF communications equipment and the
VP-1200/1000 monitor.
This VP-1200/1000 monitor is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the VP-1200/1000 monitor can help Prevent
electromagnetic interface by maintaining a minimum distance between portable and mobile RF
communication equipment (transmitters) and the VP-1200/1000 monitor as recommended below,
according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter
Rated Maximum
[m]
Output Power of
150kHz to 80 MHz 80MHz to 800 MHz 800MHz to 2.5GHz
transmitter
[W]
V1=3 Vrms E1=3V/m E1=20V/m E1=3V/m E1=20V/m
0.01 0.12 0.11 0.02 0.23 0.04
0.1 0.37 0.36 0.06 0.73 0.11
1 1.17 1.16 0.18 2.33 0.35
10 3.69 3.68 0.55 7.37 1.11
100 11.66 11.66 1.75 23.33 3.50
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m)can be estimated using the equation applicable to the frequency of the
transmitter, where p is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1) At 80 MHz and 800MHz, the separation distance for the higher frequency range applies.
NOTE 2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.

⚫ Immunity and Compliance Level


Frequency of Interest
Immunity Test Actual Immunity Level
[MHz]
Conducted RF
0.15 MHz to 80 MHz 3 Vrms
IEC 61000-4-6

80 MHz to 2.5 GHz 3 V/m


Radiated RF

IEC 61000-4-3
80 MHZ to 2.5 GHz 20 V/m

30
Service Manual VP-1200/1000

⚫ Electromagnetic immunity
IEC 60601 Compliance
Immunity test Electromagnetic environment-guidance
Test level level
The VP-1200/1000 monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of the VP-1200/1000 monitor should assure that it is used in such an electro
-magnetic environment.
Conducted RF 3 Vrms 3 Vrms The VP-1200/1000 monitor must be used only in a
IEC 61000-4-6 150kHz to 150kHz to shielded location with a minimum RF shielding
80MHz 80MHz effectiveness and, for each cable that enters the
shielded location with a minimum RF shielding
effectiveness and, for each cable that enters the
shielded location.

Radiated RF 3V/m 3V/m


IEC 61000-4-3 80MHz to 80MHz to Field strengths outside the shielded location from
2.5GHz 2.5GHz fixed RF transmitters, as determined by an
electromagnetic site survey, should be less than
a
3V/m 3V/m 3 V/m.
80MHz to 80MHz to
2.5GHz 2.5GHz Interference may occur in the vicinity of equipment
marked with the following symbol:

NOTE 1) These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
NOTE 2) It is essential that the actual shielding effectiveness and filter attenuation of the shielded
location be verified to assure that they meet the minimum specification.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/ cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength outside the location in which the VP-1200/1000 is used exceeds 3V/m, the VP-1200/1000
should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as relocating the VP-1200/1000 or using a shielded location
with a higher RF shielding effectiveness and filter attenuation.

31
Service Manual VP-1200/1000

System Diagram

◆ Overall System Block Diagram

32
Service Manual VP-1200/1000

◆ Main B/D
▶ BOARD PICTURE

▶ BLOCK DIAGRAM

When booting, processor activates as program of system FLASHROM dumps to


working SDRAM. The screen information from each CPU and the screen menu for
user interface are saved in video RAM through data interface. DSC displays menu
needed on the screen, and enables CPU to work with delivering instruction of each
CPU by KEY information inputted from CPU of MAIN part

33
Service Manual VP-1200/1000

◆ ECG B/D
▶ BOARD PICTURE
A. ECG B/D B. EKG B/D

▶ BLOCK DIAGRAM

The diagram shows that ECG signal processing. ECG signal sources from the
ECG lead are Pacemaker detection. The CPU process these sources and then
34
Service Manual VP-1200/1000

produces HR detections and Arrhythmia detection, so on.

◆ SpO2 B/D
▶ BOARD PICTURE

▶ BLOCK DIAGRAM

Above SpO2 block reads saturation rate of O2 in blood and pulse rates
And 4IBP, 4TEMP, ICO Mesurement.

35
Service Manual VP-1200/1000

◆ NIBP B/D
▶ BOARD PICTURE

▶ BLOCK DIAGRAM

Above the CPU operates motor & pump and closes valve 1, 2 and operates.
During pressuring, it reads pulse from artery delivered through hose & cuff, and
measures systolic/diastolic/average blood pressure. The result of measurement is
delivered through serial I/O. NIBP module is designed to read again or stop
reading when cuff is not in the right position or air leakage, and so on.

36
Service Manual VP-1200/1000

◆ POWER B/D
▶ BOARD PICTURE

▶ BLOCK DIAGRAM

As above, when connect AC power cord and turn on the main power switch,
the main power supplies the systems through the charger board. There are
+15V main power source and external battery source in the charger board.

37
Service Manual VP-1200/1000

Trouble Shooting

Referred
NO B/D Trouble Description
PAGE

1 ECG B/D ECG Unable To Measure. 23

2 ECG B/D Respiration Unable To Measure. 25

3 SpO2 B/D SpO2 Unable To Measure. 27

4 SpO2 B/D IBP Unable To Measure. 29

5 NIBP B/D NIBP Unable To Measure. 31

POWER
6 Unable To Turn On The Device. 34
B/D

7 MAIN B/D Boot Phase Failures. 36

8 MAIN B/D Audio Does Not Sound. 37

LCD &
9 LCD Display Is Blank. 38
LCD B/D
LCD &
10 Backlight Display Not Functioning. 39
LCD B/D
BATTERY &
11 Battery Charging Problem. 40
CHARGER B/D

38
Service Manual VP-1200/1000

◆ 1 . ECG Unable To Measure.

39
Service Manual VP-1200/1000

40
Service Manual VP-1200/1000

◆ 2. Respiration Unable To Measure.

41
Service Manual VP-1200/1000

42
Service Manual VP-1200/1000

◆ 3. SpO2 Unable To Measure.

43
Service Manual VP-1200/1000

44
Service Manual VP-1200/1000

◆ 4. IBP Unable To Measure.

45
Service Manual VP-1200/1000

46
Service Manual VP-1200/1000

◆ 5. Unable to measure NIBP.

47
Service Manual VP-1200/1000

48
Service Manual VP-1200/1000

49
Service Manual VP-1200/1000

◆ 6. Unable To Turn On The Device.

50
Service Manual VP-1200/1000

51
Service Manual VP-1200/1000

◆ 7. Boot Phase Failures.

52
Service Manual VP-1200/1000

◆ 8. Audio Does Not Sound.

53
Service Manual VP-1200/1000

◆ 9. LCD Display Is Blank.

54
Service Manual VP-1200/1000

◆ 10. LCD Display Is Blank.

55
Service Manual VP-1200/1000

◆ 11. Battery Charging Problem.

56
Service Manual VP-1200/1000

57
Service Manual VP-1200/1000

Appendix

I. 3CDaemon Installation Method

1. Open the 3cdv2r10 zip folder and open 3cdv2r10 folder. Execute SETUP file.

2. After LOADING, Click Next button as below screen.

58
Service Manual VP-1200/1000

3. Click Yes button as below screen.

4. Click Next button as below screen.

59
Service Manual VP-1200/1000

5. Select PROGRAM FLODERS and Click Next button as below screen.

6. Click OK button as below screen.

7. Click OK button as below screen.

60
Service Manual VP-1200/1000

8. Click Finish button as below screen.

** Installation Finish **

61
Service Manual VP-1200/1000

II. TERA TERM installation method


1. Open ttermp23 folder and execute setup.exe file.

2. Click Continue as below.

62
Service Manual VP-1200/1000

3. Click Continue as below.

4. Click Continue as below. (If you want to change installation directory,


appoint the directory which you want.)

63
Service Manual VP-1200/1000

5. If you finish installation as below, click the icon and execute the program.

6. Execute program, and select Serial as below and select COM port.
(Normally COM1)

64
Service Manual VP-1200/1000

7. Click Setup and click Serial port.

8. On Serial port setup screen, setting as below and click OK.

PORT : COM1 (Select same PORT which you set up)


Baud rate : 115200
Data : 8 bit
Parity : none
Stop : 1 bit
Flow control : none

** The end **
65
Service Manual VP-1200/1000

III. BOOT Loader Upgrade method


* TERA TERM Program shall be installed. (Refer to the Tera Term installation
guide.)
* 3CDaemon Program shall be installed. (Refer to the 3CDaemon installation guide)

Connect monitor and computer with RS-232C CABLE and LAN CABLE(CLOSS) as
above picture..

1. Execute Tera Term on PC.

<RS-232C CABLE>

<LAN CABLE>

66
Service Manual VP-1200/1000

2. Connect PC and monitor with RS-232C Cable. Execute Tera Term program
and booting monitor and press space bar immediately. Then below screen
will be displayed on Tera Term.

It’s normal when cursor blink.

(If you didn’t press space bar, monitor will be turned on because booting process
will be normally processed.)

3. Input SET on prompt and press Enter.

“SET” ➔ ENTER

67
Service Manual VP-1200/1000

4. Choose Select >>2 and setting local ip address as below, then press Enter.
(Newvalue : 192.168.10.21)
➔ IP Address can be changed but first three digit(192.168.10.***) shall be
same for all the time.

5. Choose Select >>3 and setting host ip address as below, then press Enter.
(Newvalue : 192.168.10.20)
➔ IP Address can be changed but first three digit(192.168.10) shall be same
for all the time. Host IP Address have to be same with IP Address of PC which
you are using.

68
Service Manual VP-1200/1000

6. Choose Select >>4 and setting gateway ip address as below, then press
Enter. (Newvalue : 192.168.10.1)
➔ Gateway IP Address have to be same with Gateway IP Address of PC which
you are using.

7. Choose Select >>5 and setting netmask ip address as below, then press
Enter. (Newvalue : 255.255.255.0)
➔Netmask IP Address have to be same with Netmask IP Address of PC which
you are using.

69
Service Manual VP-1200/1000

8. After checking everything is changed, type S on prompt and press Enter.(save)


After finishing Save, type P on prompt and press Enter.(apply & exit)

9. Click the right button of your mouse on the network folder at PC screen and
select R. Click the right button of your mouse on local are connect icon and select
R.

70
Service Manual VP-1200/1000

10. Double click the internet protocol (TCP/IP).

71
Service Manual VP-1200/1000

11. Select next IP address use(S) and setting IP address, servenet mask,
Gateway
IP address(I): 192.168.10.20
Servenet mask(U) : 255.255.255.0
Basic gateway(D) : 192.168.10.1

12. Click Ok.

72
Service Manual VP-1200/1000

13. Find 3CDaemon program as below and execute.


Start → All program → PROGRAMS → 3CDaemon → 3CDaemon

73
Service Manual VP-1200/1000

14. When executing 3CDaemon program, below screen will be displayed.


{Check address(192.168.10.20) at the below screen.}

Please check as below if address is not displayed as above.


① Check if LAN CABLE(CLOSS) is connected between PC and LAN CABLE(CLOSS).
② At TERA TERM, after vBoot> SET, check the setting value of LOCAL IP, HOST IP,
servenet mask, gateway.
③ Check PC Network IP SETTING value.
④ After checking ①~③, OFF/ON the POWER S/W at the rear part of VP-700.
⑤ Open Tera Term screen, turn ON SOFT S/W press PC KEY B/D SPACE BAR 1~3
times at the same time and execute 3CDaemon program.

15. At 3CDaemon program File, select Configure selected service.

74
Service Manual VP-1200/1000

16. At Upload/Download directory of 3CDaemon Configuration, select icon as below


to find the folder.

Click

17. Find the folder which includes boot.bin file and click ok.
(Find and go to BOOTLOADER FILE.)

75
Service Manual VP-1200/1000

18. Click at 3CDaemon Configuration as below.

19. Click yes as below.

20. On TERA TERM screen type vBoot> TFB.

Boot Loader Upgrade


“ TFB “

76
Service Manual VP-1200/1000

21. After Bootloader is copied as below, type “rst” at prompt and press ENTER key
to re-boot the monitor.

22. Completed if monitor is booted properly.

77
Service Manual VP-1200/1000

IV. F/W Upgrade by RS-232C with PC

* TERA TERM Program shall be installed. (Refer to Tera Term installation


guide)

1. Connect PC with monitor by RS-232C Cable as below picture.

<RS-232C CABLE>

2. Execute Tera Term on PC.

78
Service Manual VP-1200/1000

3. After connecting PC and monitor with RS-232C Cable, and executing Tera
Term program, booting monitor and press space bar immediately. Then
below screen will be displayed on Tera Term.

It’s normal if cursor is blinking

(If you don’t press space bar, monitor will be turned-on.)

4. Input SET on prompt and press Enter.

“SET” ➔ ENTER

79
Service Manual VP-1200/1000

* Explanation of each funtion

1 MAC Address of monitor


2 IP Address (Use when using Central Monitoring System) of monitor.
3 IP Address (Use when using Central Monitoring System) of PC.
4 Gateway IP Address(Use when using Central Monitoring System) of PC.
5 Netmask IP Address(Use when using Central Monitoring System) of PC.
6 Select application’s route which will be executed when booting.
7 Select MODE
8 Select route of Logo which is displayed on screen when booting the monitor.
9 Select BOOT MODE
A Select LCD TYPE

L BOOT initialization
P Apply & Exit
S Set value Save
0 Exit

80
Service Manual VP-1200/1000

5. Check if “6” in above picture is right.

Ex) When you are upgrading VP-1000. ➔ 1:/bin/vp1000s.bin


When you are upgrading VP-700. ➔ 1:/bin/vp700s.bin

6. Input S on prompt and press Enter.(save)


Input P on prompt and press Enter.(apply & exit)

81
Service Manual VP-1200/1000

7. Insert SD CARD to the monitor as below.


Upgrade file shall be inside of SD Card, and files are as below.
(Refer to “File name of Firmware and how to distinguish Version” for more
detail.)

MAIN F/W SMC.IMA (File name is same for all model)


ECG VM10 XXX “XXX” means Version and it can
Module F/W SpO2 VM20 XXX be changed.
NIBP VM30 XXX
*Without distinguishing Main, and Module, you can Upgrade each
separately.

8. Input on prompt as below and press ENTER. Then each F/W Upgrade will
be executed.
MAIN F/W SAN 2:SMC.IMA (SMC.IMA is file name of MAIN F/W)
ECG MU 2:VM10XXX (VM10XXX is file name of ECG F/W)
Module SpO2 MU 2:VM20XXX (VM20XXX is file name of SpO2 F/W)
NIBP MU 2:VM30XXX (VM30XXX is file name of NIBP F/W)
Note) Press space bar between SAN 2: or MU 2: .
Each file name shall be same with which is in SD Card. If file name is different,
Upgrade will not be executed properly.

82
Service Manual VP-1200/1000

9. Ex) When you upgrade MAIN F/W of VP-700,

Check route of application which will execute when booting the monitor.
(1:/BIN/VP700S.BIN)
MAIN F/W (SMC.IMA) of VP-700 shall be in the SD Card.

Input “SAN 2:SMC.IMA”.


Note: san 2  there is blank between san and 2.

10. F/W Upgrade will be executed as below.


Wait until upgrade is completed.

83
Service Manual VP-1200/1000

84
Service Manual VP-1200/1000

11. After completing F/W Upgrade and displaying below screen, input “rst” on
prompt and re-booting the monitor.

12. After re-booting the monitor, check if Upgrade is executed properly by


checking MODULE and MAIN S/W Version.

85
Service Manual VP-1200/1000

MAIN Version Check

MAIN F/W Version Check

When Module F/W is upgraded, check Module Version as below.


MODULE Version check

86
Service Manual VP-1200/1000

13. After checking Version, execute default set (MENU ➔ SETUP ➔ DEFAULT
SET ➔ ADULT).
(Note, When you are using NEONATE MODE, execute default set as NEONATE.)

** The END **

87
Service Manual VP-1200/1000

V. F/W Upgrade by SD CARD.

1. Insert SD Card at the SD Card port at the back of the monitor.


Upgrade file shall be in the SD Card, and files are as below.
(Refer to the File name of Firmware and how to distinguish Version for more detail.)
NOTE: Although you only want to upgrade Module F/W, MAIN F/W shall be always in
the SD Card, Upgrade will be executed together.

MAIN F/W SMC.IMA(File name is same for all


model)
ECG VM10 XXX “XXX” means
Module F/W SpO2 VM20 XXX version and it can
NIBP VM30 XXX be changed.

1. Turn-on the monitor and upgrade Software as below steps.

MENU ➔ SETUP ➔ S/W UPGRADE ➔ SET ➔


EXECUTE
* Refer to the below pictures.

88
Service Manual VP-1200/1000

(* Software Upgrade)


1 ●
2


3 ●
4


5

Note : Do not turn-off during Upgrading and wait until monitor is turned off
by itself after proper upgrading.

89
Service Manual VP-1200/1000

After re-booting the monitor, use screen change KEY as above picture to check
Version of MODULE and MAIN S/W.

MODULE Version Check ECG

SpO2

NIBP

90
Service Manual VP-1200/1000

91
Service Manual VP-1200/1000

MAIN Version Check

MAIN
After checking Version of MAIN & Module, execute “DEFAULT SET” as below steps.

“DEFAULT SET”

MENU ➔ SETUP ➔ DEFAULT SET ➔ ADULT


(Note, When you are using NEONATE MODE, execute default set as NEONATE.)

92
Service Manual VP-1200/1000

VI. PRINTER Supplement


1) Unfasten SCREW(M3*18 2EA) at the rear part.

2) Unfasten SCREW(M3*12 2EA) at the bottom part.

M3*12

3) Separate REAR ASS’Y and FRONT ASS’Y to the narrow direction.

93
Service Manual VP-1200/1000

4) Detach each CABLE.

POWER S/W CABLE ADAPTOR CABLE

ECG POWER
CHARGER 1 CABLE
CABLE

SPEAKER CABLE

NIBP POWER CABLE

SpO2 POWER

CABLE

5) Remove FRONT ASS’Y.


6) Puts PRINTER POWER B/D on the FRONT ASS’Y as picture.
7) Fix PRINTER POWER B/D with SCREW(M3*6S 2EA).
8) Connect POWER TO PRINTER POWER CABLE.

94
Service Manual VP-1200/1000

9) Remove PRINTER CLOSE COVER on REAR ASS’Y by using rubber hammer.


10) Remove foreign substance from PRINTER CLOSE COVER by using cutter to the
outside of REAR CASE.

Hit

11) Insert POWER TO PRINTER CABLE to the PRINTER MODULE.

12) Put the PRINTER MODULE on the place which removed PRINTER CLOSE COVER.

13) Open the PRINTER DOOR, take out PAPER.

14) Push PRINTER MODULE to inside of REAR CASE, push left and right side of

PRINTER

LOCK until sound “CLACK”

15) Insert PAPER to the PRINTER MODULE, push the DOOR until sound “CLACK”

Pus
hH
Clack

POWER TO

PRINTER

CABLE
95
Service Manual VP-1200/1000

16) Place REAR ASS’Y and FRONT ASS’Y as picture.

17) Fix each CABLE.


a. Connect POWER S/W CABLE and ADAPTOR CABLE to POWER B/D CON 1, 2.
(CON 1 → POWER S/W CABLE, CON 2 → ADAPTOR CABLE)
b. Connect each POWER CABLE to POWER B/D.
(CON 3 → CHARGER 1 CABLE, CON 7 → ECG POWER CABLE,
CON 8 → SpO2 POWER CABLE, CON 9 → NIBP POWER CABLE
CON 11 → POWER TO PRINTER CABLE)
c. Connect SPEAKER CABLE to MAIN B/D CON 6.
d. Connect MAIN TO REAR BOARD CABLE.

POWER S/W CABLE ADAPTOR CABLE


ECG POWER

CABLE

CHARGER 1 CABLE

MAIN TO REAR CABLE

POWER TO
PRINTER CABLE

SPEAKER CABLE SpO2 POWER


NIBP POWER CABLE
CABLE

96
Service Manual VP-1200/1000

18) Assemble REAR ASS’Y and FRONT ASS’Y to the arrow direction.

19) Fasten with SCREW(M3*12 2EA) at the bottom side.

20) Fasten with SCREW(M3*18 2EA) at the rear side.

97
Service Manual VP-1200/1000

21) After System booting, execute “ DEFAULT SET → SET → ADULT” at SETUP by using
ENCODER.

Screen change KEY Encoder knob

98
Service Manual VP-1200/1000

22) Use screen change KEY as above picture to check Version of MODULE and MAIN
S/W.

➢ Module version check(ECG, SpO2, NIBP screen)

ECG Version

SpO2 Version

NIBP Version

➢ Main version check(TREND screen)

MAIN Version

99
Service Manual VP-1200/1000

VII. ADAPTER Replacement

1. Unfasten SCREW at the rear part

2. Unfasten SCREW at the bottom part

100
Service Manual VP-1200/1000

3. Separate REAR ASS’Y and FRONT ASS’Y to the narrow direction

MAIN TO REAR BOARE CABLE POWER S/W CABLE ADAPTOR CABLE

CHARGER 1 CABLE

NIBP POWER CABLE

SPEAKER CABLE ECG POWER CABLE SpO2 POWER CABLE

4. Detach each CABLE

5. Detach GROUND CABLE which is fixed at FRONT ASS’Y with SCREW


Remove FRONT ASS’Y

101
Service Manual VP-1200/1000

6. Detach each CABLE which is connected with SIDE ASS’Y

7. Unfasten SCREW at the bottom side of REAR ASS’Y

102
Service Manual VP-1200/1000

8. At REAR ASS’Y, Pull REAR BRACKT ASS’Y to the arrow direction


9. Unfasten the SCREW as above picture

103
Service Manual VP-1200/1000

10. Separate ADAPTER as above picture

ADAPTER GROUND GROUND POLE


CABLE

11. Separate CABLE which is connected with GROUND POLE inside of REAR CASE
12. Assemble new ADAPTER from GROUND CABLE with opposite steps as above

104
Service Manual VP-1200/1000

13. Push REAR BRACKET ASS’Y which ECG BOARD is fixed to the arrow
direction(Inside of REAR CASE).
- {Each CABLE(SPEAKER CABLE, SIDE TO ECG, SIDE TO IBP, SIDE TO TEMP,
POWER TO ECG, POWER TO SpO2 CABLE, GROUND CABLE shall not be
oppressed between REAR CASE and REAR BRACKET A ASS’Y.}
- LAN JACK of REAR BOARD shall be fixed at HOLE of REAR CASE when fixing
REAR BRACKET ASS’Y with REAR CASE

105
Service Manual VP-1200/1000

14. At rear side of REAR CASE, place CONNECTOR of REAR B/D as picture.
Fix with SCREW and electric screw driver at NO.
1 at the bottom side of REAR CASE
When fixing NO. 2, push REAR BRACKET ASS’Y to arrow direction until you can
see the SCREW HOLE. And fix with SCREW and electric screw driver.
Fix with SCREW and electric screw driver at NO. 3, and 4

106
Service Manual VP-1200/1000

SIDE TO SpO2

CABLE

SIDE TO ECG

CABLE

SIDE TO TEMP

CABLE

SIDE TO IBP CABLE

15. Fix each CABLE (which is connect with ECG, SpO2 BOARD) to SIDE ASS’Y

16. Place REAR ASS’Y and FRONT ASS’Y as picture


Fix GROUND CABLE at HOLE of POWER BOARD with SCREW

107
Service Manual VP-1200/1000

MAIN TO REAR BOARE CABLE POWER S/W CABLE ADAPTOR CABLE

CHARGER 1 CABLE

NIBP POWER CABLE

SPEAKER CABLE ECG POWER CABLE SpO2 POWER CABLE

17. Fix each CABLE.


- Connect POWER S/W CABLE and ADAPTOR CABLE to POWER B/D CON 1, 2.
(CON 1 -> POWER S/W CABLE, CON 2 -> ADAPTOR CABLE)
- Connect each POWER CABLE to POWER B/D.
(CON 3 -> CHARGER 1 CABLE, CON 7 -> ECG POWER CABLE,
CON 8 -> SpO2 POWER CABLE, CON 9 -> NIBP POWER CABLE)
- Connect SPEAKER CABLE to MAIN B/D CON 6.
- Connect MAIN TO REAR BOARD CABLE

108
Service Manual VP-1200/1000

18. Assemble REAR ASS’Y and FRONT ASS’Y to the arrow direction

19. Fasten with SCREW at the bottom side

20. Fasten with SCREW at the rear side


Check if monitor is operating properly after you finish assembling

109
Service Manual VP-1200/1000

VIII. ECG, SpO2 B/D Replacement

110
Service Manual VP-1200/1000

111
Service Manual VP-1200/1000

112
Service Manual VP-1200/1000

113
Service Manual VP-1200/1000

114
Service Manual VP-1200/1000

115
Service Manual VP-1200/1000

116
Service Manual VP-1200/1000

117
Service Manual VP-1200/1000

118
Service Manual VP-1200/1000

119
Service Manual VP-1200/1000

IX. Main Board Replacement

120
Service Manual VP-1200/1000

121
Service Manual VP-1200/1000

122
Service Manual VP-1200/1000

123
Service Manual VP-1200/1000

X. EtCO2 (RESPIRONICS) Replacement

124
Service Manual VP-1200/1000

125
Service Manual VP-1200/1000

126
Service Manual VP-1200/1000

127
Service Manual VP-1200/1000

128
Service Manual VP-1200/1000

XI. BATTERY Replacement Manual.

* Unfasten SCREW(M3*18 2EA) at the rear part.

* Unfasten SCREW(M3*12 2EA) at the bottom part.

* Separate REAR ASS’Y and FRONT ASS’Y to the narrow direction.

129
Service Manual VP-1200/1000

MAIN TO REAR BOARE CABLE POWER S/W CABLE ADAPTOR CABLE

CHARGER 1 CABLE

NIBP POWER CABLE

SPEAKER CABLE ECG POWER CABLE SpO2 POWER CABLE

* Detach each CABLE.

* Detach GROUND CABLE which is fixed at FRONT ASS’Y with SCREW(M3*8 1EA).
* Remove FRONT ASS’Y.

130
Service Manual VP-1200/1000

* Detach BATTERY CABLE which is connected with charger board.

* Unfasten SCREW fix battery upper bracket


use screw driver to the arrow direction.
*At middle bracket, Pull battery from upper bracket to the arrow direction.

131
Service Manual VP-1200/1000

* Detach BATTERY CABLE which is connected with battery upper bracket.

NEW BATTERY

* Fix NEW BATTERY at the Battery upper bracket following arrow direction as picture.

* Push the BATTERY CABLE through MIDDLE BRACKET as the direction of picture,
take the position of back side of MIDDLE BRACKET.

132
Service Manual VP-1200/1000

* Fix in the BATTERY CABLE at the CHARGER BOARD as pictures.

* Push Battery upper bracket as arrow direction. Fix it completely by using


screwdriver.

133
Service Manual VP-1200/1000

* Place REAR ASS’Y and FRONT ASS’Y as picture.


* Fix GROUND CABLE at HOLE of POWER BOARD with SCREW(M3*8S 1EA).

134
Service Manual VP-1200/1000

MAIN TO REAR BOARE CABLE POWER S/W CABLE ADAPTOR CABLE

CHARGER 1 CABLE

NIBP POWER CABLE

SPEAKER CABLE ECG POWER CABLE SpO2 POWER CABLE

* Fix each CABLE.


1. Connect POWER S/W CABLE and ADAPTOR CABLE to POWER B/D CON 1, 2.
(CON 1 -> POWER S/W CABLE, CON 2 -> ADAPTOR CABLE)
2. Connect each POWER CABLE to POWER B/D.
(CON 3 -> CHARGER 1 CABLE, CON 7 -> ECG POWER CABLE,
CON 8 -> SpO2 POWER CABLE, CON 9 -> NIBP POWER CABLE)
3. Connect SPEAKER CABLE to MAIN B/D CON 6.
4. Connect MAIN TO REAR BOARD CABLE.

* Assemble REAR ASS’Y and FRONT ASS’Y to the arrow direction.

135
Service Manual VP-1200/1000

* Fasten with SCREW(M3*12 2EA) at the bottom side.

* Fasten with SCREW(M3*18 2EA) at the rear side.

136
Service Manual VP-1200/1000

* Check if monitor is operating properly after you finish assembling.

137
Service Manual VP-1200/1000

XII.BATTERY 2-PACK ADDITION Manual.


1) Unfasten SCREW(M3*18 2EA) at the rear part.

2) Unfasten SCREW(M3*12 2EA) and SCREW(M3*10F 4EA) at the bottom part.

M3*1

M3*1

0F

3) Separate REAR ASS’Y and FRONT ASS’Y to the narrow direction.

138
Service Manual VP-1200/1000

4) Detach each CABLE.

POWER S/W CABLE ADAPTOR CABLE

ECG POWER CABLE


CHARGER 1 CABLE

SPEAKER CABLE
NIBP POWER CABLE

SpO2 POWER CABLE

5) Detach each CABLE which is connected with SIDE ASS’Y.(ECG, SpO2,


TEMP, IBP)
6) Detach BATTERY CABLE which is connected with charger board.

ECG CABLE

SpO2 CABLE

TEMP CABLE

IBP CABLE

139
Service Manual VP-1200/1000

7) At REAR ASS’Y, Pull REAR BRACKT ASS’Y to the arrow direction.

8) Unfasten SCREW fix battery upper bracket use screw driver to the arrow
direction.
9) At middle bracket, Pull battery from upper bracket to the arrow direction .

140
Service Manual VP-1200/1000

10) Detach BATTERY CABLE which is connected with battery upper bracket .

11) Fix BATTERY and NEW BATTERY at the BATTERY UPPER BRACKET
following arrow direction as picture.

141
Service Manual VP-1200/1000

12) Place MIAN TO REAR CABEL as picture.

13) Push the BATTERY CABLE through MIDDLE BRACKET as the direction of
picture, take the position of back side of MIDDLE BRACKET.
14) Push Battery upper bracket as arrow( ) direction. Fix it completely by using
screwdriver.

BATTER BATTER
Y2 Y1

142
Service Manual VP-1200/1000

15) Unfasten SCREW(M3*6S 2EA) at the CHARGER 1 BOARD.


16) Fix CHARGER 1 BOARD(STEEL SUPPORT by using long nose.)
17) Fix in the BATTERY 1 CABLE at the CHARGER 1 BOARD as pictures.

M3*6

S
BATTERY 1 CABLE

18) Put CHARGER 2 BOARD on support.


19) Fix CHARGER 2 BOARD with SCREW(M3*6S 2EA).
20) Connect in the BATTERY 2 CABLE at the CHARGER 2 BOARD as pictures.
21) Connect POWER TO CHARGER CABLE.

BATTERY 2 CABLE

M3*6S

POWER TO CHARGER

143
Service Manual VP-1200/1000

22) Push REAR BRACKET ASS’Y which ECG BOARD is fixed to the arrow
direction(Inside of REAR CASE).
1.{Each CABLE(SPEAKER CABLE, SIDE TO ECG, SIDE TO IBP, SIDE TO
TEMP, POWER TO ECG, POWER TO SpO2 CABLE, GROUND CABLE)
shall not be oppressed between REAR CASE and REAR BRACKET
ASS’Y.}
2. LAN JACK of REAR BOARD shall be fixed at HOLE of REAR CASE when
fixing REAR BRACKET ASS’Y with REAR CASE.

LAN JACK shall be assembled at

HOLE of REAR CASE

23) Fix each CABLE (ECG TO SIDE, IBP 1, 2 TO SIDE, SpO2 TO SIDE, TEMP 1,
2 TO SIDE)
to SIDE ASS’Y

ECG CABLE

SpO2 CABLE

TEMP CABLE

IBP CABLE

144
Service Manual VP-1200/1000

24) At rear side of REAR CASE, place CONNECTOR of REAR B/D as picture.
Fix with SCREW M3*10F 1EA and electric screw driver at NO. 1 at the
bottom side of REAR CASE When fixing NO. 2, push REAR BRACKET ASS’Y
to arrow direction until you can see the SCREW HOLE.
And fix with SCREW M3*10F 1EA and electric screw driver.
Fix with SCREW M3*10F 2EA and electric screw driver at NO. 3, and 4.

when you work

procedure No2

3 2

1 M3*10
4
F
Rear
Each CONNECTOR shall be on right place

25) Place REAR ASS’Y and FRONT ASS’Y as picture.


26) Fix GROUND CABLE at HOLE of POWER BOARD with SCREW(M3*6S 1EA).

145
Service Manual VP-1200/1000

27) Fix each CABLE.


a. Connect POWER S/W CABLE and ADAPTOR CABLE to POWER B/D CON 1, 2.
(CON 1 -> POWER S/W CABLE, CON 2 -> ADAPTOR CABLE)
b. Connect each POWER CABLE to POWER B/D.
(CON 3 -> CHARGER 1 CABLE, CON 4 -> CHARGER 2 CABLE
CON 7 -> ECG POWER CABLE, CON 8 -> SpO2 POWER CABLE,
CON 9 -> NIBP POWER CABLE)
c. Connect SPEAKER CABLE to MAIN B/D CON 6.
d. Connect MAIN TO REAR BOARD CABLE.

MAIN TO REAR CABLE

ADAPTOR CABLE
POWER S/W CABLE

ECG POWER CHARGER 1 CABLE

CABLE
SPEAKER CABLE

CHARGER 2 CABLE

SpO2 POWER CABLE NIBP POWER CABLE

28) Assemble REAR ASS’Y and FRONT ASS’Y to the arrow direction.

146
Service Manual VP-1200/1000

29) Fasten with SCREW(M3*12 2EA) at the bottom side.

30) Fasten with SCREW(M3*18 2EA) at the rear side.

31) After System booting, execute “DEFAULT SET → SET → ADULT”


at SETUP by using ENCODER.

Screen change KEY Encoder knob

147
Service Manual VP-1200/1000

32) Check up the display of ECG, SpO2, NIBP version by pushing SCREEN
BUTTON two times.(If there is no display of each version, it may be due
to the wrong assembly. Please check the procedure and do the
assembly again.)

148
Service Manual VP-1200/1000

33) Check the MAIN VERSION at the left bottom side of TREND Screen.

149
Service Manual VP-1200/1000

XIII. LCD(TIANMA) REPLACEMENT Manual.


1) Unfasten SCREW(M3*18 2EA) at the rear part.

2) Unfasten SCREW(M3*12 2EA) at the bottom part.

M3*12

3) Separate REAR ASS’Y and FRONT ASS’Y to the narrow direction.

150
Service Manual VP-1200/1000

4) Detach each CABLE.

POWER S/W CABLE ADAPTOR CABLE

ECG POWER CABLE


CHARGER 1 CABLE

SPEAKER CABLE
NIBP POWER CABLE

SpO2 POWER

CABLE

5) Detach GROUND CABLE which is fixed at FRONT ASS’Y with SCREW(M3*6 1EA).

6) Remove FRONT ASS’Y.

151
Service Manual VP-1200/1000

7) Detach LCD INVERTER CABLE which is connected with Inverter board.


8) Detach KEY CABLE which is connected with key board.

LCD INVERTER

CABLE

KEY CABLE

9) Unfasten SCREW(M3*12S 6EA) which is fixed at LCD BRACKET.

152
Service Manual VP-1200/1000

10) Detach the LCD BRACKET ASS’Y from FRONT CASE ASS’Y.

LCD BRACKET ASS’Y

11) Detach MAIN TO LCD CONTROL CABLE which is connected with main board.
12) Detach LCD CABLE which is connected with driver board
① Lower the LOCK(at tge LCD cable connector on the driver board) device following
the arrow( ) direction. Handle it with care not to break the LOCK.
② Detach the LCD CABLE like the photo

MAIN TO LCD
CONTROL CABLE

LCD CABLE

153
Service Manual VP-1200/1000

13) Unfasten SCREW(M2*4S 4EA) which is fixed at LCD BRACKET.

14) Detach the LCD MODULE from LCD BRACKET ASS’Y.

15) Insert LCD CABLE of NEW LCD MODULE to the LCD BRACKET as picture.
16) Insert LCD INVERTER CABLE to the LCD BRACKET as picture.

NEW LCD MODULE LCD INVERTER

CABLE

154
Service Manual VP-1200/1000

17) Fasten with SCREW(M2*4S 4EA) at the LCD BRACKET.

18) Connect LCD INVERTER CABLE.


MAIN TO LCD CONTROL CABLE

LCD INVERTER CABLE

LCD CABLE

155
Service Manual VP-1200/1000

19) Connect LCD CABLE


① Fix the LCD cable tightly to the LCD cable connector on the driver board.
② Push back the LOCK device completely following the arrow( ) direction to the end.
20) Connect MAIN TO LCD CONTROL CABLE.
21) Lay the LCD BRACKET ASS’Y onto the FRONT CASE, following the arrow( )
direction.

LCD BRACKET ASS’Y

22) Fasten with SCREW(M3*12S 6EA) at the LCD BRACKET.


KEY CABLE
23) Connect KEY CABLE.

156
Service Manual VP-1200/1000

24) Place REAR ASS’Y and FRONT ASS’Y as picture.


25) Fix GROUND CABLE at HOLE of POWER BOARD with SCREW(M3*6S 1EA).

26) Fix each CABLE.


a. Connect POWER S/W CABLE and ADAPTOR CABLE to POWER B/D CON 1, 2.
(CON 1 -> POWER S/W CABLE, CON 2 -> ADAPTOR CABLE)
b. Connect each POWER CABLE to POWER B/D.
(CON 3 -> CHARGER 1 CABLE, CON 7 -> ECG POWER CABLE,
CON 8 -> SpO2 POWER CABLE, CON 9 -> NIBP POWER CABLE)
c. Connect SPEAKER CABLE to MAIN B/D CON 6.
d. Connect MAIN TO REAR BOARD CABLE.

POWER S/W CABLE ADAPTOR CABLE

ECG POWER
CHARGER 1 CABLE
CABLE

SPEAKER CABLE
NIBP POWER CABLE

SpO2 POWER CABLE

157
Service Manual VP-1200/1000

27) Assemble REAR ASS’Y and FRONT ASS’Y to the arrow direction.

28) Fasten with SCREW(M3*12 2EA) at the bottom side.

29) Fasten with SCREW(M3*18 2EA) at the rear side.

158
Service Manual VP-1200/1000

30) Turn on power of Monitor.

Screen change KEY Encoder knob

31) After System booting, execute “ DEFAULT SET → SET → ADULT” at SETUP by using
ENCODER.

159
Service Manual VP-1200/1000

32) Use screen change KEY as above picture n order to of MODULE and MAIN S/W

➢ Module version check(ECG, SpO2, NIBP screen)

ECG Version

SpO2 Version

NIBP Version

➢ Main version check(TREND screen)

MAIN Version

160
Service Manual VP-1200/1000

161
Service Manual VP-1200/1000

CMC Medical Devices & Drugs S.L.


<< EC Authorized Representative >>
C/Horacio Lengo Nº 18 CP 29006, Málaga-Spain

Tel: +34951214054 Fax: +34952330100

E-Mail : Info@cmcmedicaldevices.com

VOTEM CO., LTD


<< Manufacturer / Head Office >>
27, Geodudanji 1-gil, Dongnae-myeon, Chuncheon-si, Gangwon-do, Korea
(zip. 24398) Tel: +82-33-910-0701, Fax: +82-33-911-0701

162