VP-1200, 1000 (Eng) - SVC - Manual Ver 1.4 19.01.07
VP-1200, 1000 (Eng) - SVC - Manual Ver 1.4 19.01.07
※ Etc
- Trend Data Saving for 7 days
- 20 Event Data Saving
- VGA (Display Output) Function
- Built-in 3 channel Printer. (Option)
- Touch Screen (Option)
- Trend Data-saving using SD card (Option)
- Drug Dose Calculation Function
- ECG Recall Function
- 2hr Mini Trend View
2
Service Manual VP-1200/1000
3
Service Manual VP-1200/1000
Contents
◆ Description of product symbols ------ 12 ➢ III. Boot Loader Upgrade method ---- 73
➢ Overall System Block Diagram ----- 29 ➢ Ⅷ. ECG, SpO2 B/D Replacement---- 118
➢ ECG ------------------------ 36
➢ RESP ---------------------- 38
➢ SpO2 ----------------------- 39
➢ IBP ------------------------ 41
➢ TEMP --------------------- 43
➢ NIBP ----------------------- 44
➢ POWER --------------------- 45
4
Service Manual VP-1200/1000
Standard Accessories
ECG Cable
Main 1 set
5 lead type (1EA)
Operation Manual
NIBP Hose (1EA)
(1EA)
⚫ Optional Accessories
5
Service Manual VP-1200/1000
6
Service Manual VP-1200/1000
Overview of VP-1200/1000
7
Service Manual VP-1200/1000
Components of VP-1200/1000
● 2 ●
1 ● 3 ●
4 ●
6 ● 8 ●
7 ● 9
●
5
8
Service Manual VP-1200/1000
Components of VP-1200/1000
◆ Right Side
1 ●
● 2 ●
3
●
4
9
Service Manual VP-1200/1000
Components of VP-1200/1000
◆ Left Side
●
16
(2) SpO2 Connection Terminal (10) CO2 Output Connection Terminal (Option)
(3) EKG Connection Terminal(Option) (11) CO2 Input Connection Terminal (Option)
10
Service Manual VP-1200/1000
Components of VP-1200/1000
◆ Back Side
●
1
●
3 ●
4 ●
5 ●
6 ●
7 ●
8
●
9
●
10
●
2 ●
11
11
Service Manual VP-1200/1000
Type CF Product,
Power Switch
Defibrillation protected
Type BF Product,
Alarm On/Off
Defibrillation protected
CO2 Input
Gas Input Start Printing (Printer)
CO2 Exhaust
Gas output Initial Screen (Exit)
NIBP Non Invasive Blood Pressure LAN Use for Central Monitoring
12
Service Manual VP-1200/1000
13
Service Manual VP-1200/1000
14
Service Manual VP-1200/1000
15
Service Manual VP-1200/1000
16
Service Manual VP-1200/1000
ST Detect Fixed
E rr
NIBP Measurement Error Flickers at 1-second interval
O VP
NIBP Over Pressure Flickers at 1-second interval
AC in use Fixed
17
Service Manual VP-1200/1000
Product Specifications
◼ Measurement parameters
- 3ch ECG, NIBP, SpO2, Respiration, 4 IBP, 4 Temp, EtCO2, ICO
- PROCESSING PARAMETERS:
HR, NIBP(Systolic/Diastolic/Mean pressure), SpO2 Saturation,
Pulse rate, Respiration, IBP, Temp, EtCO2, ICO, Multi Gas, BIS
18
Service Manual VP-1200/1000
◼ Heart Rate
- Measurement Range: 0 ~ 300bpm
- Accuracy: ± 2bpm (Measurement value)
- Alarm Range : High alarm limit : 0 ~ 300, OFF
Low alarm limit : OFF, 0 ~ 300
19
Service Manual VP-1200/1000
◼ Respiration
- Measurement Range: 0 ~ 200 bpm
- Accuracy: ±2% or ±2 bpm
- Sweep Speed: 6.25, 12.5, 25, 50 mm/sec
- Waveform Amplitude: x0.25, x0.5, x1, x2, x3, x4, AUTO
- Apnea Alarm Range: OFF, 10 ~ 40 sec
◼ TEMP - Optional
- Measurement Range: 0 ~ 50℃
- Accuracy: 0℃ ~ 24.9℃ ± 0.2℃, 25.0℃ ~ 50.0℃ ± 0.1℃
- Alarm Range: High alarm limit - 0.1℃ ~ 50℃, OFF
Low alarm limit - OFF, 0.1℃ ~ 50℃
20
Service Manual VP-1200/1000
- Accuracy:
Time * CO2 Density ** Accuracy
0 ~ 20 0~38mmHg ±4mmHg
min 39~99mmHg ±12%
0~38mmHg ±2mHg
After 20 ±5% of reading + 0.08% for
in 39~99mmHg every 1 mmHg(above
38mmHg)
⚫ Mainstream
- Initial time : less than 15 seconds (25℃)
- CO2 measurement range : 0 ~ 150mmHg, 0 ~ 19.7%,
0 ~ 20 kPa(at 760mmHg)
- CO2 resolution : 0 ~ 69 mmHg → 0.1mmHg
70 ~ 150 mmHg → 0.25mmHg
- CO2 accuracy : 0 – 40 mmHg → ±2 mmHg
41 – 70 mmHg → ±5% of reading
71 – 100 mmHg → ±8% of reading
101 – 150 mmHg → ±10% of reading
- Respiration Measurement Range : 0 ~ 150 BPM
- Respiration Accuracy : ±1 BPM
- Required voltage : +0.5 VDC, 1.125 W (Max 1.5 W)
- Temperature/Humidity: operating environment → 0 ~ 45℃, 10 ~ 90% RH.
custody environment → -40 ~ 70℃, < 90% RH.
21
Service Manual VP-1200/1000
⚫ Sidestream
- Initial time : less than 20 seconds (25℃)
- CO2 measurement range : 0 ~ 150mmHg, 0 ~ 19.7%,
0 ~ 20 kPa (at 760mmHg)
- CO2 resolution : 0 ~ 69 mmHg → 0.1mmHg
70 ~ 150 mmHg → 0.25mmHg
- CO2 accuracy : 0 – 40 mmHg → ±2 mmHg
41 – 70 mmHg → ±5% of reading
71 – 100 mmHg → ±8% of reading
101 – 150 mmHg → ±10% of reading
More than 80 BPM ±12%
- Respiration Measurement Range : 2 ~ 150 BPM
- Respiration Accuracy : ±1 BPM
- Required voltage : +0.5 VDC ±5%
- Temperature/Humidity: operating environment → 0 ~ 40℃, 10 to 90% RH.
custody environment → -40 ~ 70℃, < 90% RH.
- Waterproof : IPX4 – Splash-proof (When sample cell is inserted in sample
cell receptacle)
22
Service Manual VP-1200/1000
⚫ Mainstream
- Size(WⅹDⅹH) : 38ⅹ37ⅹ34 mm
- Weight : 25g( not including cable )
- Initial time : within 20 seconds (25℃)
- Surface temp : 50℃ max(at ambient temp. 23℃)
23
Service Manual VP-1200/1000
⚫ Sidestream
- Size(WⅹDⅹH) : 33ⅹ78ⅹ49 mm
- Weight : 130g( including cable )
- Initial time : within 20 seconds (25℃)
- Required voltage : 5±0.5 VDC, 1.6 W(2.0 W peak)
- Water Handling : Sampling line with proprietary water removal tubing
- Sampling Lines : 2 ± 0.1 m and 3 ± 0.1 m versions
- Sampling Flow Rate : 50 ± 10 ml/mm
◼ Printer - Optional
- Print Method: Thermal
- Type: Internal
- Print Mode
a. BOTH (Max 3 Wave) + Trend
b. Wave (Max 3 Wave)
c. Trend
- Paper Width: 57mm
- Valid Width: 120 ~ 300mm (Change to 25mm step)
◼ TREND
- Parameter: Heart Rate or Pulse Rate, SpO2, NIBP, IBP, Temp,
ST Segment, EtCO2, Multi gas, BIS, ICO
- 7 Days trend data save
◼ Interface
- RS-232C port
- LAN Port (Central Monitor)
- VGA Port (Display Transferring)
- SD CARD
- ECG Output port
24
Service Manual VP-1200/1000
◼ Physical Specifications
- Size: 290 x 175 x 280 mm (W x D x H)
- Weight: 7.0kg (with battery)
- Power Requirement: 1.2A, 100-240V~, 50/60Hz
- Power consumption: 80VA (MAX)
- BATTERY (1 Cell): 11.1V 4400mAh
; Recharge Time – 9 hours, Discharge Time – 120minutes
◼ Operating Environment
- Temperature: (Operating: 10°C ~ 35°C / Storage: -10°C ~ 50°C)
- Humidity: (Operating: 30 ~ 85% / Storage: 20 ~ 95%)
- Atmospheric Pressure: 700 to 1060 hPa
◼ Standard Accessories
- Patient Cable (5 leads) 1 ea
- NIBP cuff for adult 1 ea
- NIBP Hose 1 ea
- Power Code 1 ea
- Electrode (5pcs/set) 1 set
- SpO2 Sensor (Reusable) 1 ea
- Operation Manual 1 ea
◼ Optional Accessories
- ECG 3 leads Cable - Disposable SpO2 Sensor
- EKG 5 leads Cable - BIS Extension Cable
- SpO2 Extension Cable - EEG Cable
- Press Measuring Kit
- Printer Module
- Roll Paper(18M)
- Ground Cable
- NIBP cuff for neonate, Child
- Temperature sensor
- EtCO2 Kit set (EtCO2 when operating)
- Multi Gas Kit set
- Multi Gas Extension Cable
- ICO Cable
- 3M Electrode (50 PCS/1SET)
- Battery 11.1V, 4400mAh
- Trolley (=Cart)
- Wall Mount
25
Service Manual VP-1200/1000
Safety
CAUTION system
◼ EMI
◼ EMC
26
Service Manual VP-1200/1000
Safety
⚫ Electromagnetic emission
Emissions Compliance Electromagnetic environment - guidance
The VP-1200/1000 uses RF energy only for its internal
RF emissions function. Therefore, its RF emissions are very low and
Group 1
CISPR 11 are not likely to cause any interference in nearby
electronic equipment.
RF emissions
Class B
CISPR 11
Harmonic emissions
Class A
IEC 61000-3-2
Voltage fluctuations /
Flicker emissions Complies
IEC 61000-3-3
⚫ Electromagnetic immunity
IEC 60601 Electromagnetic
Immunity test Compliance level
Test level environment-guidance
Electrostatic ±6kV Contact ±6kV Contact Floors should be wood,
Discharge (ESD) concrete or ceramic tile. If floors
±8kV air ±8kV air are covered with synthetic
IEC 61000-4-2 material, the relative humidity
should be at least 30%.
Electrical fast ±2kV for power ±2kV for power Mains power quality should be
Transient/burst supply lines supply lines that of a typical commercial or
±1kV for ±1kV for hospital environment.
IEC 61000-4-4 Input/output lines Input/output lines
27
Service Manual VP-1200/1000
28
Service Manual VP-1200/1000
3V/m 3V/m
80MHz to 80MHz to 800 MHz to 2.5 GHz
2.5GHz 2.5GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
NOTE 1) At 80MHz and 800 MHz, the higher frequency range applies.
NOTE 2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by
Absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/ cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the VP-1200/1000 is used exceeds the applicable RF compliance
level above, the VP-1200/1000 should be observed to verify normal operation. If abnormal perform-
ance is observed, additional measures may be necessary, such as re-orienting or relocating the
VP-1200/1000.
b Over the frequency range 150kHz to 80 MHz, field strengths should be less than [V 1] V/m.
29
Service Manual VP-1200/1000
IEC 61000-4-3
80 MHZ to 2.5 GHz 20 V/m
30
Service Manual VP-1200/1000
⚫ Electromagnetic immunity
IEC 60601 Compliance
Immunity test Electromagnetic environment-guidance
Test level level
The VP-1200/1000 monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of the VP-1200/1000 monitor should assure that it is used in such an electro
-magnetic environment.
Conducted RF 3 Vrms 3 Vrms The VP-1200/1000 monitor must be used only in a
IEC 61000-4-6 150kHz to 150kHz to shielded location with a minimum RF shielding
80MHz 80MHz effectiveness and, for each cable that enters the
shielded location with a minimum RF shielding
effectiveness and, for each cable that enters the
shielded location.
NOTE 1) These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
NOTE 2) It is essential that the actual shielding effectiveness and filter attenuation of the shielded
location be verified to assure that they meet the minimum specification.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/ cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength outside the location in which the VP-1200/1000 is used exceeds 3V/m, the VP-1200/1000
should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as relocating the VP-1200/1000 or using a shielded location
with a higher RF shielding effectiveness and filter attenuation.
31
Service Manual VP-1200/1000
System Diagram
32
Service Manual VP-1200/1000
◆ Main B/D
▶ BOARD PICTURE
▶ BLOCK DIAGRAM
33
Service Manual VP-1200/1000
◆ ECG B/D
▶ BOARD PICTURE
A. ECG B/D B. EKG B/D
▶ BLOCK DIAGRAM
The diagram shows that ECG signal processing. ECG signal sources from the
ECG lead are Pacemaker detection. The CPU process these sources and then
34
Service Manual VP-1200/1000
◆ SpO2 B/D
▶ BOARD PICTURE
▶ BLOCK DIAGRAM
Above SpO2 block reads saturation rate of O2 in blood and pulse rates
And 4IBP, 4TEMP, ICO Mesurement.
35
Service Manual VP-1200/1000
◆ NIBP B/D
▶ BOARD PICTURE
▶ BLOCK DIAGRAM
Above the CPU operates motor & pump and closes valve 1, 2 and operates.
During pressuring, it reads pulse from artery delivered through hose & cuff, and
measures systolic/diastolic/average blood pressure. The result of measurement is
delivered through serial I/O. NIBP module is designed to read again or stop
reading when cuff is not in the right position or air leakage, and so on.
36
Service Manual VP-1200/1000
◆ POWER B/D
▶ BOARD PICTURE
▶ BLOCK DIAGRAM
As above, when connect AC power cord and turn on the main power switch,
the main power supplies the systems through the charger board. There are
+15V main power source and external battery source in the charger board.
37
Service Manual VP-1200/1000
Trouble Shooting
Referred
NO B/D Trouble Description
PAGE
POWER
6 Unable To Turn On The Device. 34
B/D
LCD &
9 LCD Display Is Blank. 38
LCD B/D
LCD &
10 Backlight Display Not Functioning. 39
LCD B/D
BATTERY &
11 Battery Charging Problem. 40
CHARGER B/D
38
Service Manual VP-1200/1000
39
Service Manual VP-1200/1000
40
Service Manual VP-1200/1000
41
Service Manual VP-1200/1000
42
Service Manual VP-1200/1000
43
Service Manual VP-1200/1000
44
Service Manual VP-1200/1000
45
Service Manual VP-1200/1000
46
Service Manual VP-1200/1000
47
Service Manual VP-1200/1000
48
Service Manual VP-1200/1000
49
Service Manual VP-1200/1000
50
Service Manual VP-1200/1000
51
Service Manual VP-1200/1000
52
Service Manual VP-1200/1000
53
Service Manual VP-1200/1000
54
Service Manual VP-1200/1000
55
Service Manual VP-1200/1000
56
Service Manual VP-1200/1000
57
Service Manual VP-1200/1000
Appendix
1. Open the 3cdv2r10 zip folder and open 3cdv2r10 folder. Execute SETUP file.
58
Service Manual VP-1200/1000
59
Service Manual VP-1200/1000
60
Service Manual VP-1200/1000
** Installation Finish **
61
Service Manual VP-1200/1000
62
Service Manual VP-1200/1000
63
Service Manual VP-1200/1000
5. If you finish installation as below, click the icon and execute the program.
6. Execute program, and select Serial as below and select COM port.
(Normally COM1)
64
Service Manual VP-1200/1000
** The end **
65
Service Manual VP-1200/1000
Connect monitor and computer with RS-232C CABLE and LAN CABLE(CLOSS) as
above picture..
<RS-232C CABLE>
<LAN CABLE>
66
Service Manual VP-1200/1000
2. Connect PC and monitor with RS-232C Cable. Execute Tera Term program
and booting monitor and press space bar immediately. Then below screen
will be displayed on Tera Term.
(If you didn’t press space bar, monitor will be turned on because booting process
will be normally processed.)
“SET” ➔ ENTER
67
Service Manual VP-1200/1000
4. Choose Select >>2 and setting local ip address as below, then press Enter.
(Newvalue : 192.168.10.21)
➔ IP Address can be changed but first three digit(192.168.10.***) shall be
same for all the time.
5. Choose Select >>3 and setting host ip address as below, then press Enter.
(Newvalue : 192.168.10.20)
➔ IP Address can be changed but first three digit(192.168.10) shall be same
for all the time. Host IP Address have to be same with IP Address of PC which
you are using.
68
Service Manual VP-1200/1000
6. Choose Select >>4 and setting gateway ip address as below, then press
Enter. (Newvalue : 192.168.10.1)
➔ Gateway IP Address have to be same with Gateway IP Address of PC which
you are using.
7. Choose Select >>5 and setting netmask ip address as below, then press
Enter. (Newvalue : 255.255.255.0)
➔Netmask IP Address have to be same with Netmask IP Address of PC which
you are using.
69
Service Manual VP-1200/1000
9. Click the right button of your mouse on the network folder at PC screen and
select R. Click the right button of your mouse on local are connect icon and select
R.
70
Service Manual VP-1200/1000
71
Service Manual VP-1200/1000
11. Select next IP address use(S) and setting IP address, servenet mask,
Gateway
IP address(I): 192.168.10.20
Servenet mask(U) : 255.255.255.0
Basic gateway(D) : 192.168.10.1
72
Service Manual VP-1200/1000
73
Service Manual VP-1200/1000
74
Service Manual VP-1200/1000
Click
17. Find the folder which includes boot.bin file and click ok.
(Find and go to BOOTLOADER FILE.)
75
Service Manual VP-1200/1000
76
Service Manual VP-1200/1000
21. After Bootloader is copied as below, type “rst” at prompt and press ENTER key
to re-boot the monitor.
77
Service Manual VP-1200/1000
<RS-232C CABLE>
78
Service Manual VP-1200/1000
3. After connecting PC and monitor with RS-232C Cable, and executing Tera
Term program, booting monitor and press space bar immediately. Then
below screen will be displayed on Tera Term.
“SET” ➔ ENTER
79
Service Manual VP-1200/1000
L BOOT initialization
P Apply & Exit
S Set value Save
0 Exit
80
Service Manual VP-1200/1000
81
Service Manual VP-1200/1000
8. Input on prompt as below and press ENTER. Then each F/W Upgrade will
be executed.
MAIN F/W SAN 2:SMC.IMA (SMC.IMA is file name of MAIN F/W)
ECG MU 2:VM10XXX (VM10XXX is file name of ECG F/W)
Module SpO2 MU 2:VM20XXX (VM20XXX is file name of SpO2 F/W)
NIBP MU 2:VM30XXX (VM30XXX is file name of NIBP F/W)
Note) Press space bar between SAN 2: or MU 2: .
Each file name shall be same with which is in SD Card. If file name is different,
Upgrade will not be executed properly.
82
Service Manual VP-1200/1000
Check route of application which will execute when booting the monitor.
(1:/BIN/VP700S.BIN)
MAIN F/W (SMC.IMA) of VP-700 shall be in the SD Card.
83
Service Manual VP-1200/1000
84
Service Manual VP-1200/1000
11. After completing F/W Upgrade and displaying below screen, input “rst” on
prompt and re-booting the monitor.
85
Service Manual VP-1200/1000
86
Service Manual VP-1200/1000
13. After checking Version, execute default set (MENU ➔ SETUP ➔ DEFAULT
SET ➔ ADULT).
(Note, When you are using NEONATE MODE, execute default set as NEONATE.)
** The END **
87
Service Manual VP-1200/1000
88
Service Manual VP-1200/1000
(* Software Upgrade)
●
1 ●
2
●
3 ●
4
●
5
Note : Do not turn-off during Upgrading and wait until monitor is turned off
by itself after proper upgrading.
89
Service Manual VP-1200/1000
After re-booting the monitor, use screen change KEY as above picture to check
Version of MODULE and MAIN S/W.
SpO2
NIBP
90
Service Manual VP-1200/1000
91
Service Manual VP-1200/1000
MAIN
After checking Version of MAIN & Module, execute “DEFAULT SET” as below steps.
“DEFAULT SET”
92
Service Manual VP-1200/1000
M3*12
93
Service Manual VP-1200/1000
ECG POWER
CHARGER 1 CABLE
CABLE
SPEAKER CABLE
SpO2 POWER
CABLE
94
Service Manual VP-1200/1000
Hit
12) Put the PRINTER MODULE on the place which removed PRINTER CLOSE COVER.
14) Push PRINTER MODULE to inside of REAR CASE, push left and right side of
PRINTER
15) Insert PAPER to the PRINTER MODULE, push the DOOR until sound “CLACK”
Pus
hH
Clack
POWER TO
PRINTER
CABLE
95
Service Manual VP-1200/1000
CABLE
CHARGER 1 CABLE
POWER TO
PRINTER CABLE
96
Service Manual VP-1200/1000
18) Assemble REAR ASS’Y and FRONT ASS’Y to the arrow direction.
97
Service Manual VP-1200/1000
21) After System booting, execute “ DEFAULT SET → SET → ADULT” at SETUP by using
ENCODER.
98
Service Manual VP-1200/1000
22) Use screen change KEY as above picture to check Version of MODULE and MAIN
S/W.
ECG Version
SpO2 Version
NIBP Version
MAIN Version
99
Service Manual VP-1200/1000
100
Service Manual VP-1200/1000
CHARGER 1 CABLE
101
Service Manual VP-1200/1000
102
Service Manual VP-1200/1000
103
Service Manual VP-1200/1000
11. Separate CABLE which is connected with GROUND POLE inside of REAR CASE
12. Assemble new ADAPTER from GROUND CABLE with opposite steps as above
104
Service Manual VP-1200/1000
13. Push REAR BRACKET ASS’Y which ECG BOARD is fixed to the arrow
direction(Inside of REAR CASE).
- {Each CABLE(SPEAKER CABLE, SIDE TO ECG, SIDE TO IBP, SIDE TO TEMP,
POWER TO ECG, POWER TO SpO2 CABLE, GROUND CABLE shall not be
oppressed between REAR CASE and REAR BRACKET A ASS’Y.}
- LAN JACK of REAR BOARD shall be fixed at HOLE of REAR CASE when fixing
REAR BRACKET ASS’Y with REAR CASE
105
Service Manual VP-1200/1000
14. At rear side of REAR CASE, place CONNECTOR of REAR B/D as picture.
Fix with SCREW and electric screw driver at NO.
1 at the bottom side of REAR CASE
When fixing NO. 2, push REAR BRACKET ASS’Y to arrow direction until you can
see the SCREW HOLE. And fix with SCREW and electric screw driver.
Fix with SCREW and electric screw driver at NO. 3, and 4
106
Service Manual VP-1200/1000
SIDE TO SpO2
CABLE
SIDE TO ECG
CABLE
SIDE TO TEMP
CABLE
15. Fix each CABLE (which is connect with ECG, SpO2 BOARD) to SIDE ASS’Y
107
Service Manual VP-1200/1000
CHARGER 1 CABLE
108
Service Manual VP-1200/1000
18. Assemble REAR ASS’Y and FRONT ASS’Y to the arrow direction
109
Service Manual VP-1200/1000
110
Service Manual VP-1200/1000
111
Service Manual VP-1200/1000
112
Service Manual VP-1200/1000
113
Service Manual VP-1200/1000
114
Service Manual VP-1200/1000
115
Service Manual VP-1200/1000
116
Service Manual VP-1200/1000
117
Service Manual VP-1200/1000
118
Service Manual VP-1200/1000
119
Service Manual VP-1200/1000
120
Service Manual VP-1200/1000
121
Service Manual VP-1200/1000
122
Service Manual VP-1200/1000
123
Service Manual VP-1200/1000
124
Service Manual VP-1200/1000
125
Service Manual VP-1200/1000
126
Service Manual VP-1200/1000
127
Service Manual VP-1200/1000
128
Service Manual VP-1200/1000
129
Service Manual VP-1200/1000
CHARGER 1 CABLE
* Detach GROUND CABLE which is fixed at FRONT ASS’Y with SCREW(M3*8 1EA).
* Remove FRONT ASS’Y.
130
Service Manual VP-1200/1000
131
Service Manual VP-1200/1000
NEW BATTERY
* Fix NEW BATTERY at the Battery upper bracket following arrow direction as picture.
* Push the BATTERY CABLE through MIDDLE BRACKET as the direction of picture,
take the position of back side of MIDDLE BRACKET.
132
Service Manual VP-1200/1000
133
Service Manual VP-1200/1000
134
Service Manual VP-1200/1000
CHARGER 1 CABLE
135
Service Manual VP-1200/1000
136
Service Manual VP-1200/1000
137
Service Manual VP-1200/1000
M3*1
M3*1
0F
138
Service Manual VP-1200/1000
SPEAKER CABLE
NIBP POWER CABLE
ECG CABLE
SpO2 CABLE
TEMP CABLE
IBP CABLE
139
Service Manual VP-1200/1000
8) Unfasten SCREW fix battery upper bracket use screw driver to the arrow
direction.
9) At middle bracket, Pull battery from upper bracket to the arrow direction .
140
Service Manual VP-1200/1000
10) Detach BATTERY CABLE which is connected with battery upper bracket .
11) Fix BATTERY and NEW BATTERY at the BATTERY UPPER BRACKET
following arrow direction as picture.
141
Service Manual VP-1200/1000
13) Push the BATTERY CABLE through MIDDLE BRACKET as the direction of
picture, take the position of back side of MIDDLE BRACKET.
14) Push Battery upper bracket as arrow( ) direction. Fix it completely by using
screwdriver.
BATTER BATTER
Y2 Y1
142
Service Manual VP-1200/1000
M3*6
S
BATTERY 1 CABLE
BATTERY 2 CABLE
M3*6S
POWER TO CHARGER
143
Service Manual VP-1200/1000
22) Push REAR BRACKET ASS’Y which ECG BOARD is fixed to the arrow
direction(Inside of REAR CASE).
1.{Each CABLE(SPEAKER CABLE, SIDE TO ECG, SIDE TO IBP, SIDE TO
TEMP, POWER TO ECG, POWER TO SpO2 CABLE, GROUND CABLE)
shall not be oppressed between REAR CASE and REAR BRACKET
ASS’Y.}
2. LAN JACK of REAR BOARD shall be fixed at HOLE of REAR CASE when
fixing REAR BRACKET ASS’Y with REAR CASE.
23) Fix each CABLE (ECG TO SIDE, IBP 1, 2 TO SIDE, SpO2 TO SIDE, TEMP 1,
2 TO SIDE)
to SIDE ASS’Y
ECG CABLE
SpO2 CABLE
TEMP CABLE
IBP CABLE
144
Service Manual VP-1200/1000
24) At rear side of REAR CASE, place CONNECTOR of REAR B/D as picture.
Fix with SCREW M3*10F 1EA and electric screw driver at NO. 1 at the
bottom side of REAR CASE When fixing NO. 2, push REAR BRACKET ASS’Y
to arrow direction until you can see the SCREW HOLE.
And fix with SCREW M3*10F 1EA and electric screw driver.
Fix with SCREW M3*10F 2EA and electric screw driver at NO. 3, and 4.
procedure No2
3 2
1 M3*10
4
F
Rear
Each CONNECTOR shall be on right place
145
Service Manual VP-1200/1000
ADAPTOR CABLE
POWER S/W CABLE
CABLE
SPEAKER CABLE
CHARGER 2 CABLE
28) Assemble REAR ASS’Y and FRONT ASS’Y to the arrow direction.
146
Service Manual VP-1200/1000
147
Service Manual VP-1200/1000
32) Check up the display of ECG, SpO2, NIBP version by pushing SCREEN
BUTTON two times.(If there is no display of each version, it may be due
to the wrong assembly. Please check the procedure and do the
assembly again.)
148
Service Manual VP-1200/1000
33) Check the MAIN VERSION at the left bottom side of TREND Screen.
149
Service Manual VP-1200/1000
M3*12
150
Service Manual VP-1200/1000
SPEAKER CABLE
NIBP POWER CABLE
SpO2 POWER
CABLE
5) Detach GROUND CABLE which is fixed at FRONT ASS’Y with SCREW(M3*6 1EA).
151
Service Manual VP-1200/1000
LCD INVERTER
CABLE
KEY CABLE
152
Service Manual VP-1200/1000
10) Detach the LCD BRACKET ASS’Y from FRONT CASE ASS’Y.
11) Detach MAIN TO LCD CONTROL CABLE which is connected with main board.
12) Detach LCD CABLE which is connected with driver board
① Lower the LOCK(at tge LCD cable connector on the driver board) device following
the arrow( ) direction. Handle it with care not to break the LOCK.
② Detach the LCD CABLE like the photo
MAIN TO LCD
CONTROL CABLE
LCD CABLE
153
Service Manual VP-1200/1000
15) Insert LCD CABLE of NEW LCD MODULE to the LCD BRACKET as picture.
16) Insert LCD INVERTER CABLE to the LCD BRACKET as picture.
CABLE
154
Service Manual VP-1200/1000
LCD CABLE
155
Service Manual VP-1200/1000
156
Service Manual VP-1200/1000
ECG POWER
CHARGER 1 CABLE
CABLE
SPEAKER CABLE
NIBP POWER CABLE
157
Service Manual VP-1200/1000
27) Assemble REAR ASS’Y and FRONT ASS’Y to the arrow direction.
158
Service Manual VP-1200/1000
31) After System booting, execute “ DEFAULT SET → SET → ADULT” at SETUP by using
ENCODER.
159
Service Manual VP-1200/1000
32) Use screen change KEY as above picture n order to of MODULE and MAIN S/W
ECG Version
SpO2 Version
NIBP Version
MAIN Version
160
Service Manual VP-1200/1000
161
Service Manual VP-1200/1000