Professional Documents
Culture Documents
VM15-Service Manual-ENG
VM15-Service Manual-ENG
VM15-Service Manual-ENG
※ Etc
- Trend Data (7 days)
- Event Data (20 event)
- VGA (Display Output) Function
- Built-in 3 channel Printer. (Option)
- Touch Screen
- Trend Data-saving using USB
- Drug Dose Calculation Function
- ECG Multiview arrhythmia Function
- ECG OUT Function (Option)
- Mini Trend View
- Color change (Wave & Measurement value)
- NIBP STAT Function
- NIBP VENOUS STAT Function
- OXY-CRG View
- Patient information
2
Service Manual Characteristics and Principles of VM15
- Temp TD
- USB Host for S/W Upgrade
3
Service Manual Contents
Table of Contents
1. Composition of VM15 ..................................................................................................................................................... 9
4
Service Manual Contents
TREND .............................................................................................................................................................. 37
Interface .......................................................................................................................................................... 37
Standard Accessories................................................................................................................................. 38
8. Safety.................................................................................................................................................................................... 40
EMI ............................................................................................................................................................................. 40
EMC ............................................................................................................................................................................ 40
5
Service Manual Contents
NIBP B/D(VOTEM)................................................................................................................................................ 53
General ............................................................................................................................................................. 54
Display Trouble................................................................................................................. 55
ECG / RESPIRATION Measurement Trouble .......................................................... 55
SpO2 Measurement Trouble (VOTEM) .................................................................... 55
SpO2 Measurement Trouble (MASIMO – Option) .............................................. 56
NIBP Measurement Trouble (VOTEM) ..................................................................... 56
NIBP Measurement Trouble (SunTech - Option) ................................................. 57
IBP Measurement Trouble ............................................................................................ 58
CO2 Measurement Trouble ......................................................................................... 58
qCON Measurement Trouble ...................................................................................... 58
TEMP Measurement Trouble ....................................................................................... 58
PRINTER Trouble (OPTION) ......................................................................................... 59
11. Appendix............................................................................................................................................................................. 60
6
Service Manual Contents
7
Service Manual Contents
8
Service Manual Composition of VM15
1. Composition of VM15
Table 1-1 Standard Item.
9
Service Manual Composition of VM15
IBP Kit Set (Utah) à 1 per channel (2ch : 2ea, 4ch : 4ea)
Note: Please contact the technical supporting team or sales team for the more information on
optional items.
10
Service Manual How to Set VM15
11
Service Manual Overview of VM15
3. Overview of VM15
12
Service Manual Description of Product Symbols
4. Components of VM15
13
Service Manual Description of Product Symbols
14
Service Manual Description of Product Symbols
Note: You can put the option box at any layer of the track.
Note: The standard module box with ECG, SpO2, NIBP, 1TEMP module and 4.3" LCD is called
the Transport Module, and is referred to as TM in this document.
15
Service Manual Description of Product Symbols
13 Optional module box Port that connects the option module box Option
CAUTION: Please do not on or off the power supply when the main monitor (front panel) and
Transport Module (rear panel) is separated. It may cause the malfunction.
16
Service Manual Description of Product Symbols
Defibrillation-Proof
Power switch
Type CF Applied part
Defibrillation-Proof
Alarm Paused
Type BF Applied part
Waring or Cautions
Change screen
(Refer to the manual)
(Freeze)
IPX Waterproof rating
Freeze the Wave
ECG Electrocardiograph /
NBP Start/Stop
Respiration
SpO2 (Printer)
SpO2
, Start Print
T1 ~ T4 (Exit)
Temperature Measurement
IPX7 Initial Screen
CO RS-232 communication
Cardiac Output COM1
terminal
CO2
EtCO2 Module port VGA VGA OUT port
17
Service Manual Description of Product Symbols
MULTI GAS
Multi-Gas Module port LAN LAN port
qCON
qCON index Ground
18
Service Manual Explanation of screen
6. Explanation of screen
Display of Main
The picture below is the basic screen of VM15 that can be seen after the booting screen passes
when the product is first turned on.
The main screen is consist of ECG, Respiration, SpO2, NIBP, 4ch IBP(Option), CO2(Option),
Multi-Gas(Option), qCON(Option), 1TEMP(Standard), 3TEMP(Option), Smart Key.
19
Service Manual Explanation of screen
20
Service Manual Explanation of screen
1 Alarm Pause
Screen change/popup
3
function
NIBP measurement
4
start/stop
5 Screen freeze
7 Print function
8 Setup menu
21
Service Manual Explanation of screen
AC in use Fixed
22
Service Manual Product Specifications
7. Product Specifications
Measurement Parameters
⚫ ECG or EKG, NIBP, SpO2, Respiration, 4ch IBP, ICO, 4ch Temp, EtCO2 or Multi Gas, qCON
⚫ Processing Parameters
HR (Heart Rate)
NIBP (Systolic / Diastolic / Mean pressure)
SpO2 Saturation
PR (Pulse rate)
IBP (Systolic / Diastolic / Mean pressure)
ICO (Cardiac Output)
Temperature
Resp (Respiration)
EtCO2 (End-Tidal CO2) or Multi Gas
qCON
23
Service Manual Product Specifications
24
Service Manual Product Specifications
⚫ UI (User Interface)
◼ Key buttons: Power, Alarm Snooze, NIBP start/stop
⚫ Parameters
◼ ECG
◼ Resp (Respiration)
◼ SpO2
◼ NIBP
◼ 1 Channel Temp
⚫ Sound
◼ Beep, Alarm (Low / Middle / High), Information
◼ Volume: N.A (Fixed)
25
Service Manual Product Specifications
⚫ Heart Rate
◼ Measurement Range: 0, 15 ~ 300 bpm (± 1 bpm or ± 1%, which is greater)
◼ HR Calculation: 4, 6, 8, 10, 12, 14, 16 waves
(Set the usage of HR calculation with combining number of Wave)
⚫ Pacemaker Detect
⚫ Arrhythmia Detect
◼ ASYS (Asystole) - Asystole or electromotive disassociation.
◼ VFIB (ventricular flutter/fibrillation) - Irregular ventricular waveform
◼ PVC (premature ventricular contraction) - Premature ventricular contraction
◼ VTAC (ventricular tachycardia) - PVCs at 180 BPM
◼ BGM (ventricular bigeminy) - Normal beat followed by a PVC, repeated
◼ TGM (ventricular trigeminy) - Two normal beats followed by a PVC, repeated
◼ CPT (Couplet) - Two PVCs together in a series with normal beats, repeated
◼ SBRD (sinus bradycardia) - Normal sinus rhythm below 56 BPM (HR≤55)
◼ ATAC (atrial tachycardia) - Normal sinus rhythm over 135 BPM (HR≥136)
◼ SVT (supra-ventricular tachycardia) - Normal sinus rhythm over 195 BPM (HR≥196)
◼ VENT - Ventricular rhythm
◼ TPT (Triplet) - Three PVCs together in a series with normal beats, repeated
◼ R on T (R on T)
⚫ ECG Output
◼ Gain: 1 mm/mV
◼ DC Offset Voltage: 10 mV
◼ Time Delay: 112ms from real ECG waveform
⚫ Defibrillator Sync
◼ Pulse Amplitude: 3 V when heart beat detected
◼ Pulse Width: 100ms
◼ Time Delay: 188~424ms from real ECG waveform
(Variable with Pacemaker ON/OFF, Notch ON/OFF)
◼ 0.05 % up to
⚫ Setting Time
◼ Wave out Time: Max. 2 sec
◼ SpO2 Percentage Display: Max. 10 sec
⚫ Measurement Range
◼ SpO2
◆ Display Range: 0 ~ 100 %
◆ Calibration Range: 70 ~ 100 %
◼ Pulse Rate
◆ Display Range: 25 ~ 240 bpm
◆ Calibration Range: 25 ~ 240 bpm
◼ Perfusion Index
◆ Display Range: 0.02 ~ 20.0 %
◆ Calibration Range: 0.10 ~ 20.0 %
⚫ Accuracy
◼ SpO2
◆ No Motion: ≤ 2.0 %
◆ Motion(RMS): ≤ 3.0 %
◼ Pulse Rate
◆ No Motion: ≤ 3.0 bpm
◆ Motion(RMS): ≤ 5.0 bpm
⚫ Resolution
◼ SpO2: ≤ 0.1 %
◼ Pulse Rate: ≤ 1 bpm
⚫ Time to Display
◼ SpO2: ≤8, ≤12 seconds
◼ Pulse Rate: ≤8, ≤12 seconds
◼ Perfusion Index: ≤8, ≤12 seconds
⚫ Asystole Detection Time
◼ SpO2: ≤8 seconds
◼ Pulse Rate: ≤8 seconds
◼ Perfusion Index: ≤8 seconds
⚫ Delay
◼ SpO2: ≤10 seconds
◼ Pulse Rate: ≤10 seconds
◼ Perfusion Index: ≤10 seconds
⚫ Response Time
◼ SpO2: ≤20 seconds
◼ Pulse Rate: ≤20 seconds
◼ Perfusion Index: ≤20 seconds
⚫ Setting
◼ Sensitivity: Maximum, Normal, APOD
◼ Averaging Time (seconds): 2~4, 4~6, 8, 10, 12, 14, 16 seconds
◼ Fastsat : ON/OFF
◼ Smart Tone: ON/OFF
◼ SIQ: ON/OFF
27
Service Manual Product Specifications
⚫ Alarm Range
Display High Low
NIBP Systolic
[mmHg] Diastolic 0 ~ 275, OFF OFF, 0 ~ 275
Mean
⚫ Pressure Modes
◼ Manual
◼ Automatic: 2, 3, 4, 5, 10, 15, 30 min, 1, 2, 4, 8 ,12 hours intervals
◼ Stat: 5 ~ 15 min intervals
◼ Venous Stasis
⚫ Error Messages
◼ NIBP ERR. CUFF/HOSE
(Cuff or Hose err)
28
Service Manual Product Specifications
◆ 7 seconds
◼ Patient Safety
◆ Internal operating software ensures that:
Maximum cuff inflation time is limited to 75 seconds
Duration of blood pressure reading is limited to
130 seconds (Adult mode)
120 seconds (Adult Motion Tolerant mode)
90 seconds (Pediatric mode)
75 seconds (Neonate mode)
◆ Additional redundant safety circuitry oversees normal operation and will override
to abort a reading if:
cuff pressure exceeds 300 mmHg (Adult & Pediatric modes) or 150 mmHg
(Neonate mode) at any time
the cuff has been inflated for 180 seconds (Adult & Pediatric modes) or
90 seconds (Neonate mode)
The Module meets all relevant parts of the following
Safety Standards
IEC60601-1:1997
IEC/EN60601-2-30:1999/2000
AAMI SP10:1992/2002
EN1060-1:1996
EN1060-3:1997
30
Service Manual Product Specifications
⚫ Label
◼ ABP (Arterial Blood Pressure)
◆ Artery group throughout the whole body (systemic)
◼ CVP (Central Venous Pressure)
◆ Central vein / atrium group
◼ LAP (Left Arterial Pressure)
◆ Intracranial group
◼ PAP (Pulmonary Arterial Pressure)
◆ Pulmonary artery group
⚫ Catheter
◼ Utah (Optional)
◼ Abbot (Not provided)
◼ Edward (Not provided)
◼ Bd (Not provided)
◼ Medex (Not provided)
◼ Philips (Not provided)
⚫ Hemodynamic Parameters
◼ Input/Measured Data from VM15
◆ CO: Cardiac Output (L/min)
◆ HR: Heart Rate (bpm)
◆ MAP: Mean Arterial Pressure (mmHg)
◆ CVP: Central Venous Pressure (mmHg)
◆ PAM: Pulmonary Artery Mean Pressure (mmHg)
◆ PAW: Pulmonary Artery Wedge Pressure (mmHg)
◆ PAD: Pulmonary Artery Diastolic Pressure (mmHg)
◆ LA: Left Atrium Pressure (mmHg)
◆ WEIGHT: Weight (kg)
◆ HEIGHT: height (cm)
◼ Calculated Data
◆ BSA: Body Surface Area (m2)
◆ CI: Cardiac Index (L/min/m2)
◆ SV: Stroke Volume (mL/beat)
◆ SVR: Systemic Vascular Resistance (dny*sec*cm-2)
◆ SVRI: Systemic Vascular Resistance Index (dny*sec*cm-5*m2)
◆ PVR: Pulmonary Vascular Resistance (dny*sec*cm-5)
◆ PVRI: Pulmonary Vascular Resistance Index (dny*sec*cm-5*m2)
31
Service Manual Product Specifications
⚫ Mainstream (EtCO2)
◼ Initialization Time
◆ Capnogram, displayed in less than 15 seconds (at 25℃)
◆ Full specifications within 2 minutes
◼ Respiration Rate Range: 0 ~ 150 BPM (Breaths Per Minute)
◼ Respiration Rate Accuracy: ± 1 breath
◼ CO2 Measurement Range
◆ 0 ~ 150 mmHg
◆ 0 ~ 19.7%
◆ 0 ~ 20 kPa (at 760 mmHg)
◆ Barometric Pressure supplied by host
◼ Rise Time
◆ Less than 60 ms - Adult Reusable or Single-Patient-Use Airway Adapter
◆ Less than 60 ms - Infant Reusable or Single-Patient-Use Airway Adapter
◼ CO2 Resolution
◆ 0 ~ 69 mmHg → 0.1 mmHg
◆ 70 ~ 150 mmHg → 0.25 mmHg
◼ CO2 Accuracy
◆ 0 ~ 40 mmHg → ± 2 mmHg
◆ 41 ~ 70 mmHg → ± 5% of reading
◆ 71 ~ 100 mmHg → ± 8% of reading
◆ 01 ~ 150 mmHg → ± 10% of reading
◼ CO2 Stability
◆ Short Term Drift: Drift over four hours shall not exceed 0.8 mmHg max
◆ Long Term Drift: Accuracy specification will be maintained over a 120-hour period
◼ Calibration
32
Service Manual Product Specifications
⚫ Sidestream (EtCO2)
◼ Initialization Time
◆ Capnogram, displayed in less than 20 seconds (at 25℃)
◆ Full specifications within 2 minutes
◼ Respiration Measurement Range: 2 ~ 150 BPM (Breaths Per Minute)
◼ Respiration Rate Accuracy: ± 1 breath
◼ CO2 Measurement Range
◆ 0 ~ 150 mmHg
◆ 0 ~ 19.7%
◆ 0 ~ 20 kPa (at 760mmHg)
◆ Barometric Pressure supplied by host
◼ CO2 Resolution
◆ 0 ~ 69 mmHg → 0.1 mmHg
◆ 70 ~ 150 mmHg → 0.25 mmHg
◼ CO2 Accuracy
◆ 0 ~ 40 mmHg → ± 2 mmHg
◆ 1 ~ 70 mmHg → ± 5 % of reading
◆ 71 ~ 100 mmHg → ± 8 % of reading
◆ 101 ~ 150 mmHg → ± 10 % of reading
◆ Above 80 bpm ± 12 % of reading
◼ CO2 Stability
◆ Short Term Drift: Drift over four hours shall not exceed 0.8 mm Hg max
◆ Long Term Drift: Accuracy specification will be maintained over a 120-hour period
◼ Calibration
◆ No routine user calibration required.
◼ Sample Kit Hours of Use
◆ Nasal Cannula (all styles) – up to 12 hours
◆ On-Airway Adapter Kits without dehumidification tubing – up to 12 hours
◆ On-Airway Adapter Kits with dehumidification tubing – up to 120 hours
◼ Temperature/Humidity
◆ Operating environment: 0 ~ 40℃, 10 to 90% RH, non-condensing
◆ Storage environment: -40 ~ 70℃, <90% RH, non-condensing
◼ Waterproof
◆ IPX4 / Splash-proof (When sample cell is inserted in sample cell receptacle)
◼ Required Voltage
◆ +5.0 VDC, 5%
33
Service Manual Product Specifications
◼ Power Rating
◆ Rated input: 1.3 Watts typical. Steady State
◆ Less than 2.0 Watts maximum on power up (warm up)
◼ Operating Temperature
◆ IRMA AX+: 5 to 50 °C (41 to 122 °F)
◼ Operating Humidity
◆ < 40 hPa H2O (non-condensing) (95 %RH at 30 °C)
◼ Operating Atmospheric Pressure
◆ 525–1200 hPa
(525 hPa corresponding to an altitude of 4,572 m / 15,000 feet)
◼ Power Supply
◆ IRMA AX+: 4.5-5.5 VDC, max 1.4 W
(power on surge @ 5 V less than 350 mA during 200 ms)
◼ Breath Detection
◆ Adaptive threshold, minimum 1 vol% change in CO2 concentration
◼ Respiration Rate
◆ 0 ~ 150 bpm, ± 1 bpm
◼ Warm-up Time
◆ IRMA AX+: < 20 seconds (agent identification enabled and full accuracy)
◼ Rise Time
◆ CO2 ≤ 90 ms
◆ N2O ≤ 300 ms
◆ HAL, ISO, ENF, SEV, DES ≤ 300 ms
◼ Total System Response Time
◆ < 1 second
◼ Accuracy - Standard Conditions
(valid for dry single gases at 22 ± 5 °C and 1013 ± 40 hPa)
◆ CO2:
0 to 15 vol% ± (0.2 vol% + 2% of reading
◆ N2O:
0 to 100 vol% ± (2 vol% + 2% of reading)
◆ SEV:
0 to 10 vol% ± (0.15 vol% + 5% of reading)
◆ DES:
0 to 22 vol% ± (0.15 vol% + 5% of reading)
34
Service Manual Product Specifications
◼ Operating Temperature
◆ ISA OR+/AX+: 5 to 50 °C (41 to 122 °F)
◼ Operating Humidity
◆ < 40 hPa H2O (non-condensing) (95 %RH at 30 °C)
◼ Operating Atmospheric Pressure
◆ 525–1200 hPa
(525 hPa corresponding to an altitude of 4,572 m / 15,000 feet)
◼ Power Supply
◆ 4.5 to 5.5 VDC
◆ ISA AX+: < 1.6 W (normal op.), < 2.0 W (peak 5 VDC)
◆ ISA OR+: < 2.0 W (normal op.), < 2.4 W (peak @ 5 VDC)
◼ Breath Detection
◆ Adaptive threshold, minimum 1 vol% change in CO2 concentration
◼ Respiration Rate
◆ 0 ~ 150 bpm, ± 1 bpm
◼ Warm-up Time
◆ ISA OR+/AX+: < 20 seconds
(Concentrations reported, automatic agent identification enabled and full accuracy)
◆ CO2:
0 to 15 vol% ± (0.2 vol% + 2% of reading)
15 to 25 vol% (Unspecified)
◆ N2O:
0 to 100 vol% ± (2 vol% + 2% of reading)
◆ SEV:
0 to 10 vol% ± (0.15 vol% + 5% of reading)
10 to 25 vol% (Unspecified)
◆ DES:
0 to 22 vol% ± (0.15 vol% + 5% of reading)
35
Service Manual Product Specifications
22 to 25 vol% (Unspecified)
8 to 25 vol% (Unspecified)
qCON - Optional
Printer - Optional
TREND
⚫ Popup Screen
◼ Text Trend
◼ Graph Trend
⚫ Interval Time: 1, 5, 10 minutes
⚫ Parameter
◼ HR (Heart Rate), SpO2, Resp (Respiration), NIBP (Sys, Dia, Mean),
◼ 4ch IBP, 4ch Temp, ST Segment, EtCO2, Multi Gas, qCON.
⚫ 7 Days Trend Data Save
Interface
⚫ RS-232C port (for service port only)
⚫ LAN Port (10/100/1000Mbps)
⚫ USB Port (for Software Update)
⚫ VGAOUT (Optional)
⚫ ECG Output (Optional)
Physical Specifications
⚫ Size: 283 x 414 x 229 mm (W x D x H)
⚫ Weight: 8.5kg (with VM15 Standard)
⚫ Power Requirement: 1.2A, 100-240V~, 50/60Hz
⚫ Power consumption: 90VA (MAX)
⚫ Battery
◼ Recharge Time: 12 hours
◼ Storage Temperature (Only Battery)
◆ 1 year: -20 to 25 ℃
◆ 3 months: -20 to 45 ℃
◆ 1 month: -20 to 60 ℃
◼ VM15 Discharge Time
◆ Performance Specification (1 Pack): 10.8VDC, 4400mAh
◆ 1 Pack: Basic monitoring configuration 2 Hours (Standard)
◆ 2 Pack: Basic monitoring configuration 4 Hours (Optional)
◼ TM (Transport Module) Part Discharge Time – Used to TM only
◆ Performance Specification (1 Pack): 10.8VDC, 2200mAh
◆ 1 Pack: Basic monitoring configuration 1 Hours (Standard)
37
Service Manual Product Specifications
Operating Environment
⚫ Temperature
◼ Operating: 10°C ~ 35°C
◼ Storage: -10°C ~ 50°C
⚫ Humidity
◼ Operating: 30 ~ 85%
◼ Storage: 20 ~ 95%
⚫ Atmospheric Pressure: 700 to 1060 hPa
Standard Accessories
⚫ ECG 5 Leads Cable (Mini-Candal Type) 1 ea
⚫ NIBP Cuff (for Adult) 1 ea
⚫ NIBP Hose 1 ea
⚫ Adult Finger Clip SpO2 Sensor (Reusable) 1 ea
⚫ Temperature Sensor (Skin Type) 1 ea
⚫ Power Code 1 ea
⚫ Electrode (5 pcs/set) 1 set
⚫ Quick Manual 1 ea
Optional Accessories
⚫ ECG 3 leads Cable (Mini-Candal Type)
⚫ 12Ch EKG Cable (Mini-Candal Type)
⚫ SpO2 Extension Cable
⚫ Disposable SpO2 Sensor
⚫ Masimo Set uSpO2 Pulse Oximetry Cable (with Masimo)
⚫ Masimo Adult Reusable Sensor (with Masimo)
⚫ NIBP Cuff for Neonate
⚫ NIBP Cuff for Child
⚫ NIBP Hose (with Suntech)
⚫ NIBP Cuff for Adult (with Suntech)
⚫ Temperature Sensor (Skin Type)
⚫ Temperature Sensor (Rectal Type)
⚫ IBP-Y Extension Cable (Mini-Candal Type)
⚫ IBP Extension Cable (Abbot) -- Only Extension cable / (IBP Kit Set : Not provided)
⚫ IBP Extension Cable (Edward) -- Only Extension cable / (IBP Kit Set : Not provided)
⚫ IBP Extension Cable (Bd) -- Only Extension cable / (IBP Kit Set : Not provided)
⚫ IBP Extension Cable (Medex) -- Only Extension cable / (IBP Kit Set : Not provided)
⚫ IBP Extension Cable (Philips) -- Only Extension cable / (IBP Kit Set : Not provided)
⚫ IBP Extension Cable (Utah)
⚫ IBP Kit Set (Utah)
⚫ ICO Extension Cable (Mini-Candal Type)
⚫ EtCO2 Extension Cable (Mini-Candal Type)
⚫ EtCO2 Kit Set (Mainstream or Sidestream)
⚫ Multi-Gas Extension Cable (Mini-Candal Type)
⚫ Multi-Gas Kit Set (Mainstream or Sidestream)
38
Service Manual Product Specifications
39
Service Manual Safety
8. Safety
◼ Use only accessories supplied or recommended by Votem. Use accessories according to the
manufacture’s directions for use and your facility’s standards. The use of accessories,
transducers, and cables other than those specified may result in increased emission or
WARNING
decreased immunity of the VM15 monitor.
◼ In case where EMI is causing disturbances, it may be necessary to relocate this system
CAUTION
EMI
Although this system has been manufactured in compliance with existing EMI(Electromagnetic
Interface) requirements, use of this system in the presence of an electromagnetic field can cause
momentary degradation of the vital Signal.
If this occurs often, Votem suggests a review of the environment in which the system is begin use, to
identity possible sources of radiated emissions. These emissions could be from other electrical devices
used within the same room or an adjacent room. Communication devices such as cellular phones and
pagers can cause these emissions. The existence of radios, TVs, or microwave transmission equipment
nearby can also cause interface.
EMC
A The testing for EMC(Electromagnetic Compatibility) of this system has been performed according
to the international standard for EMC with medical device(IEC60601-1-2). This IEC standard was
adopted in Europe as the European norm(EN60601-1-2).
40
Service Manual Safety
The VM15 monitor is intended for use in the specified electromagnetic environment. The customer or
the user of the VM15 should assure that it is used in such an environment as described below;
Electromagnetic emission
Electromagnetic immunity
41
Service Manual Safety
± 1 ㎸ line(s) to
Surge ± 1 ㎸ line(s) to line(s) line(s) Mains power quality should be that of a
IEC 61000-4-5 ± 2 ㎸ line(s) to earth ± 2 ㎸ line(s) to typical commercial or hospital environment.
earth
<5 % UT <5 % UT
(>95 % dip in UT) (>95 % dip in UT)
for 0,5 cycle for 0,5 cycle
Mains power quality should be that of a
Voltage dips, short 40 % UT 40 % UT
typical commercial or hospital environment.
Interruptions and (60 % dip in UT) (60 % dip in UT)
If the user of the VM15 requires continued
Voltage variations for 5 cycles for 5 cycles
operation during power mains interruptions,
On power supply 70 % UT 70 % UT
it is recommended that the VM15 be
Input lines (30 % dip in UT) (30 % dip in UT)
powered from an uninterruptible power
IEC 61000-4-11 for 25 cycles for 25 cycles
supply or a battery.
<5 % UT <5 % UT
(>95 % dip in UT) (>95 % dip in UT)
for 5 s for 5 s
NOTE UT is the a.c. mains voltage prior to application of the test level.
42
Service Manual Safety
NOTE 1) At 80MHz and 800 MHz, the higher frequency range applies.
NOTE 2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by
Absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To access the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the VM15 is used exceeds
the applicable RF compliance level above, the VM15 should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such
as re- orienting or relocating the VM15.
b Over the frequency range 150 ㎑ to 80 ㎒, field strengths should be less than 3 V/m & 20 V/m
43
Service Manual Safety
Recommended separation distance between portable and mobile RF communications equipment and the
VM15.
The VM15 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the VM15 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the VM15 as recommended below,
according to the maximum output power of the communications
[W]
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
metres (m)can be estimated using the equation applicable to the frequency of the transmitter, where p is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1) At 80 MHz and 800MHz, the separation distance for the higher frequency range applies.
NOTE 2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
44
Service Manual System Diagram
9. System Diagram
45
Service Manual System Diagram
MCU B/D
▶ BOARD PICTURE
Figure 9.2 1 Solo MCU B/D(TM). Figure 9.2 2 Dual MCU B/D(MAIN).
▶ BLOCK DIAGRAM
HDMI Controller
USB2.0
USB HUB
UART 1 TX1/RX1 Controller
UART / CH1 USB_OTG
(DEBUG)
Audio
Controller
PCIe
Expansion
UART
DC/DC
5V VDD
15->5v
DC/DC Core
5 -> 1.38V 1.38V
PWM4
DC/DC Core
PWM Buzzer
5 -> SNVS SNVS(3.3V)
DC/DC Core
SNVS -> 3.3V 3.3V
DC/DC Core
SNVS -> 2.5V 2.5V
MCU receives data from Main b/d and Base b/d. Received data processes and sends feedback from
MCU. MCU controls Touch, LCD, and LCD backlight. MCU receives and processes data such as Debug,
Transport Module, and Printer through UART.
46
Service Manual System Diagram
Main B/D
▶ BOARD PICTURE
▶ BLOCK DIAGRAM
As shown above, the power supply(DC 15V) is divided through the regulator. Power and battery
information is sent to the MCU through the 4ch ADC. TOUCH SCREEN, USB, SOUND is controlled
through MCU. EtCO2, MultiGas, IBP/ICO, qCON,3-TEMP, PRINTER, RS232 is controlled through UART of
MCU. KEY, RESET information is transmitted to MCU. It provides HDMI and LAN information to the MCU.
It transmits VGA OUT, LCD, LCD BACKLIGHT information provided by MCU.
47
Service Manual System Diagram
BASE B/D
▶ BOARD PICTURE
▶ BLOCK DIAGRAM
BASE B/D
MAIN B/D
USB
DC15V REGULATOR USB
HUB/CONTROLLER
BATTERY DEBUG
CHARGER 9V 5V 3.3V UART
NIBP
Li-ion
10.8V 2.2Ah FUSE ECG
125V/5A
SpO2
4ch ADC Micro SD
MCU
RTC B/D KEY
4.3'' LCD
SPEAKER
Figure 9.4 2 BASE Block Diagram.
As shown above, the power supply(DC 15V, Li-ion battery-10.8V) is divided through the regulator. Power
and battery information is sent to the MCU through the 4ch ADC. USB information is sent to the MCU
through USB hub/controller. DEBUG, NIBP, ECG, SpO2 information is sent to MCU by UART. KEY, RESET
information is sent to MCU. MCU send LCD and SPEAKER information.
48
Service Manual System Diagram
PWR B/D
▶ BOARD PICTURE
▶ BLOCK DIAGRAM
As above, when connect AC power cord and turn on the main power switch, the main power supplies
the systems through the Power board. There are +15V main power source and external battery source
in the Power board.
49
Service Manual System Diagram
ECG B/D
▶ BOARD PICTURE
Figure 9.6 1 ECG B/D(with MCU B/D). Figure 9.6 2 ECG B/D(without MCU B/D).
▶ BLOCK DIAGRAM
RA
LEAD
LA SELECT
24-Bit MCU
LPF B/D
AFE
LL LPF
V1 COMMON
FEEDBACK
RL SELECT
RESPIRATION
LEAD CIRCUIT
LPF
SELECT
TEMP
SENSOR
TEMPERATURE
CIRCUIT
The diagram shows that ECG signal processing. ECG signal sources from the ECG lead are Pacemaker
detection. The MCU process these sources and then produces HR detections and Arrhythmia detection,
so on. In addition, the MCU processes the TEMP signal passing through the TEMPERATURE CIRCUIT and
the RESP signal passing through the RESPIRATION CIRCUIT.
50
Service Manual System Diagram
SpO2 B/D(VOTEM)
▶ BOARD PICTURE
▶ BLOCK DIAGRAM
Above SpO2 block reads saturation rate of O2 in blood and pulse rates
51
Service Manual System Diagram
IBP/ICO B/D
▶ BOARD PICTURE
▶ BLOCK DIAGRAM
Above IBP block reads invasive blood pressure. Also ICO block reads invasive cardiac output.
52
Service Manual System Diagram
NIBP B/D(VOTEM)
▶ BOARD PICTURE
▶ BLOCK DIAGRAM
Above the CPU operates motor & pump and closes valve 1, 2 and operates. During pressuring, it reads
pulse from artery delivered through hose & cuff, and measures systolic/diastolic/average blood pressure.
The result of measurement is delivered through serial I/O. NIBP module is designed to read again or
stop reading when cuff is not in the right position or air leakage, and so on.
53
Service Manual Trouble Shooting
WARNING: The disassembled monitor may cause any damage to the patient
when it is touched.
Collect or repair the patient monitor only by the authorized persons.
WARNING: If the result is suspicious, please check the status of the patient
monitor after checking patient with other medical devices.
General
If you have any trouble with the monitor, please check with the below table. Check again before
contacting to the A/S center.
After self-checking, please contact Votem Co., Ltd or the regional distributor if the troubles are
not solved
How to Solve
54
Service Manual Trouble Shooting
Display Trouble
The sudden power off during 1. Check the battery if it is completely discharged.
operation. 2. Check AC power source.
1. Check ‘power off” message on the screen after pressing the power button
When Power off does for 1~ 2 seconds.
not work. 2. Turn off the monitor with force => press the hole of Power Off with sharp
instrument like ball point pen. (RESET hole is also in Transport Module)
1. Check the equipments with high magnetic field near the monitor.
Abnormal Waves 2. Check the cable connection.
3. Check the ground.
Symptoms Description
No display “LEAD FAULT” 1. Check ECG cable and the gel at electrode.
Message 2. Check with other ECG cable
55
Service Manual Trouble Shooting
Improper sensor type or application. Check and see if blood flow to the site is restricted.
LOW SIGNAL Excessive motion relative to perfusion. Check the placement of the sensor.
QUALITY Sensor or cable is damaged or not Re-apply sensor or move to a different site.
functioning. Replace sensor or cable.
56
Service Manual Trouble Shooting
Symptoms Description
Check that the cuff is in the correct position.
Check the patient.
WEAK PULSE Check that the cuff is properly tightened.
Check that there is no excessive clothing between the arm and the cuff.
Check that the correct size cuff is being applied.
Check that the hose is connected to the system and the cuff.
Check that the cuff is properly tightened.
Check that the cuff is in the correct position.
INFLATE TIMEOUT
Check that the correct size cuff is being applied.
Check that the cuff is not leaking air.
Check that the hose connections are not damaged or loose.
Check that Vpump and Vlogic are within the voltage specifications.
Check the data/power input connection.
HARDWARE ERROR
If applicable, check that sleep mode operation is correct.
Service may be required. Call a SunTech representative.
57
Service Manual Trouble Shooting
Symptoms Description
1. Separate the tube and remove the air bubble and make the tubing length
Unstable BP Measurement
shorter.
Symptoms Description
FILTERLINE
Connect Filter line.
NOT CONNECTION
Check the sampling line and filter. Clean them. Change if necessary. Disconnect
No Wave and value of EtCO2
the filter line and connect again.
Check that the probe adapter is properly connected to the acquisition module.
Check that the probe is properly connected to the probe adapter.
Make sure you are using the correct probe for the monitored anatomical
location.
Temperature is not measured
Use a probe that is compatible with your system.
If the sensor is damaged, try using a good probe.
Make sure the acquisition module is properly connected to the monitor.
Make sure there are no two identical measurement modules in the system.
58
Service Manual Trouble Shooting
Symptoms Description
59
Service Manual Appendix
11. Appendix
Version Check
1) Turn on VM15
2) Touch SETUP button at the bottom of the screen to display the SETUP window.
60
Service Manual Appendix
61
Service Manual Appendix
62
Service Manual Appendix
3) Prepare the USB. Copy the SWU file to the specified path of USB.(see below).
A. Specified path of USB : USB root directory (ex. F: , G: , H:).
63
Service Manual Appendix
6) Turn on VM15.
7) Touch SETUP button at the bottom of the screen to display the SETUP window.
8) Touch ADMIN OPTION button at the bottom of the screen to display the SETUP window.
64
Service Manual Appendix
10) Touch MAIN OS update button at the bottom of the screen to display the ADMIN OPTION
window.
65
Service Manual Appendix
18) Compare the recorded kernel version with the current kernel version. If the kernel version
has been changed, the kernel swu update has been successfully performed.
66
Service Manual Appendix
3) Prepare the USB. Copy the S/W file to the specified path of USB.(see below)
B. Specified path of USB : USB root directory (ex. F: , G: , H:)
67
Service Manual Appendix
10) Touch SETUP button at the bottom of the screen to display the SETUP window.
68
Service Manual Appendix
69
Service Manual Appendix
70
Service Manual Appendix
19) Press the Reset switch to turn off the VM15 TM power.
71
Service Manual Appendix
72
Service Manual Appendix
26) Touch SETUP button at the bottom of the screen to display the SETUP window.
73
Service Manual Appendix
29) Compare the recorded kernel version with the current kernel version. If the kernel version
has been changed, the kernel swu update has been successfully performed.
74
Service Manual Appendix
3) Prepare the USB. Copy the S/W file to the specified path of USB.(see below)
A. Specified path of USB : USB root directory (ex. F: , G: , H:)
7) Touch SETUP button at the bottom of the screen to display the SETUP window.
76
Service Manual Appendix
77
Service Manual Appendix
78
Service Manual Appendix
13) After completing the S/W update, check that the VM15 power is off.
14) Turn on the VM15.
15) Touch SETUP button at the bottom of the screen to display the SETUP window.
79
Service Manual Appendix
80
Service Manual Appendix
18) Compare the recorded F/W version with the current F/W version. If the F/W version has
been changed, the S/W Update has been successfully performed.
81
Service Manual Appendix
TM S/W Update(Application)
1) Record the current TM F/W version.(*)
* Refer to 11.1 Version Check
3) Prepare the USB. Copy the S/W file to the specified path of USB.(see below)
A. Specified path of USB : USB root directory (ex. F: , G: , H:)
7) Touch SETUP button at the bottom of the screen to display the SETUP window.
83
Service Manual Appendix
84
Service Manual Appendix
85
Service Manual Appendix
13) After completing the S/W update, check that the VM15 power is off.
86
Service Manual Appendix
15) Touch SETUP button at the bottom of the screen to display the SETUP window.
87
Service Manual Appendix
18) Compare the recorded F/W version with the current F/W version. If the F/W version has
been changed, the S/W Update has been successfully performed.
88
Service Manual Appendix
3) Prepare the USB. Copy the S/W file to the specified path of USB.(see below)
B. Specified path of USB : USB root directory (ex. F: , G: , H:)
89
Service Manual Appendix
7) Touch SETUP button at the bottom of the screen to display the SETUP window.
90
Service Manual Appendix
9) Touch ‘S/W UPDATE’. Check the S/W UPDATE window. S/W UPDATE
91
Service Manual Appendix
12) Check that the module S/W update proceeds in order.(see below)
A. Order of update progress : SpO2 -> NIBP -> TEMP -> IBP
B. In the case of 2-IBP update, ‘2IBP OK’ is displayed on the screen.
92
Service Manual Appendix
13) After completing the S/W update, check that the VM15 power is off.
15) Touch SETUP button at the bottom of the screen to display the SETUP window.
93
Service Manual Appendix
94
Service Manual Appendix
18) Compare the recorded F/W version with the current F/W version. If the F/W version has
been changed, the S/W Update has been successfully performed.
95
Service Manual Appendix
3) Prepare the USB. Copy the S/W file to the specified path of USB.(see below).
B. Specified path of USB : USB root directory (ex. F: , G: , H:).
96
Service Manual Appendix
6) Turn on VM15.
9) Turn on VM15.
97
Service Manual Appendix
14) Compare the recorded kernel version with the current kernel version. If the kernel version
has been changed, the kernel recovery has been successfully performed.
98
Service Manual Appendix
3) Prepare the USB. Copy the S/W file to the specified path of USB.(see below)
A. Specified path of USB : USB root directory (ex. F: , G: , H:)
99
Service Manual Appendix
100
Service Manual Appendix
13) Press the Reset switch to turn off the VM15 TM power.
101
Service Manual Appendix
102
Service Manual Appendix
20) Touch SETUP button at the bottom of the screen to display the SETUP window.
103
Service Manual Appendix
23) Compare the recorded kernel version with the current kernel version. If the kernel version
has been changed, the kernel recovery has been successfully performed.
104
Service Manual Appendix
3) Prepare the USB. Copy the S/W file to the specified path of USB.(see below).
C. Specified path of USB : USB root directory (ex. F: , G: , H:)
105
Service Manual Appendix
6) Turn on VM15
8) Remove USB.
9) Turn on VM15
10) Compare the recorded F/W version with the current F/W version. If the F/W version has
been changed, the S/W recovery has been successfully performed.
106
Service Manual Appendix
3) Prepare the USB. Copy the S/W file to the specified path of USB.(see below)
A. Specified path of USB : USB root directory (ex. F: , G: , H:)
107
Service Manual Appendix
108
Service Manual Appendix
109
Service Manual Appendix
19) Touch SETUP button at the bottom of the screen to display the SETUP window.
110
Service Manual Appendix
111
Service Manual Appendix
22) Compare the recorded F/W version with the current F/W version. If the F/W version has
been changed, the S/W recovery has been successfully performed.
112
Service Manual Appendix
113
Service Manual Appendix
5) Push PRINTER MODULE to inside of SIDE CASE, push left and right side of PRINTER (until
sound “CLACK”).
6) Turn on.
114
Service Manual Appendix
115
Service Manual Appendix
7) Insert battery(3S2P).
8) Close the battery cover and fasten with countersunk head screw(M3x6mm-2ea).
116
Service Manual Appendix
3) Unfasten the countersunk head screw(M3x6mm-2ea) on the ESN BATTERY COVER of VM15
TM.
117
Service Manual Appendix
4) Lightly push the battery(3S1P) from inside the VM15 TM. Make sure the ESN BATTERY
COVER is removed.
118
Service Manual Appendix
Figure 11.13 7 How to remove the battery(3S1P) from the ESN BATTERY COVER.
119
Service Manual Appendix
Figure 11.13 8 How to combine the battery(3S1P) in the ESN BATTERY COVER.
120
Service Manual Appendix
11) Press the ESN BATTERY COVER 'part' of the battery cover and push the ESN BATTERY COVER
into the ESN CASE. Be careful not to pinch the battery connector wire between the
cases.
121
Service Manual Appendix
12) Fasten the countersunk head screw(M3x6mm-2ea) on the ESN BATTERY COVER of VM15
TM.
14) Place the washer(M2-4ea) on the BOTTOM COVER of VM15 TM and fasten the countersunk
head screw(M2x6mm-4ea).
122
Service Manual Appendix
123
Service Manual Appendix
6) Close the battery cover and fasten with countersunk head screw(M3x6mm-2ea)
124
Service Manual Appendix
125
Service Manual Appendix
126
Service Manual Appendix
127
Service Manual Appendix
10) Detach each Cable(Careful of connector damage when removing the FFC Cable)
128
Service Manual Appendix
129
Service Manual Appendix
17) Assemble REAR CASE and FRONT ASS’Y to the arrow direction.
18) Fasten with countersunk head screw (M3x8mm-13ea) at the bottom side.
130
Service Manual Appendix
131
Service Manual Appendix
21) Close the battery cover and fasten with countersunk head screw(M3x6mm-2ea).
132
Service Manual Appendix
133
Service Manual Appendix
134
Service Manual Appendix
135
Service Manual Appendix
12) Fasten with round head screw (M3x6mm-4ea) at the FRONT BRACKET.
13) Lay the FRONT BRACKET onto the 15LCD ASS’Y CASE.
Figure 11.16 10 Combine the FRONT BRACKET and 15LCD ASS’Y CASE.
14) Fasten with round head screw (M3x6mm-4ea) at the FRONT BRACKET.
136
Service Manual Appendix
15) Lay the FRONT BODY CASE onto the FRONT BRACKET.
Figure 11.16 12 Combine the FRONT BODY CASE and FRONT BRACKET.
137
Service Manual Appendix
138
Service Manual Appendix
2) If the installed SpO2 B/D is VOTEM SpO2, Proceed to 11.23-2) of SpO2 B/D Replacement
Manual.
If the installed SpO2 B/D is Masimo SpO2, Proceed to 11.24-2) of SpO2 B/D Replacement
Manual.
4) Remove the 1Temp Cable, ECG Mini-Candal Cable, EKG Mini-Candal Cable, ECG Power Cable,
ECG Out Cable on the ECG B/D.
139
Service Manual Appendix
5) Unfasten the sem's screw with spring washer(M3x6mm-1ea) and PCB metal
support(M3x20mm-2ea) from ECG B/D
Figure 11.17 3 Position to unfasten the screw and support on ECG B/D
Figure 11.17 5 Position to unfasten the support on Heat Sink Fix Bracket
140
Service Manual Appendix
Figure 11.17 6 How to remove Heatsink and Heat Sink Fix Bracket
9) Unfasten PCB metal support(M3x10mm-2ea) on the MCU B/D.
141
Service Manual Appendix
10) Pull the trigger. Check that the MCU B/D is floating.
142
Service Manual Appendix
12) Remove Speaker Cable, ECG Power Cable, SPO2 Power Cable, ECG Out Cable, 1Temp Cable,
Masimo Side Cable, Membrane Cable, 4.3 LCD FFC Cable
Figure 11.17 10 How to remove the Cable from the Base B/D(detailed 4.3 LCD FFC Cable)
13) Unfasten the countersunk head screw(M3x6mm) on the SIDE STD COVER
143
Service Manual Appendix
15) Unfasten the sem's screw with spring washer(M3x6mm-6ea) and the PCB plastic
support(M3x20mm-1ea) on the Base B/D
Figure 11.17 13 Position to unfasten the screws and the support on Base B/D
144
Service Manual Appendix
Figure 11.17 17 Separate the ESN CASE and the Base Bracket
20) Unfasten the sem's screw with spring washer(M3x6mm-4ea) on the VM15 LCD B/D
Figure 11.17 18 Position to unfasten the screws on the VM15 LCD B/D
145
Service Manual Appendix
21) Separate 4.3inch LCD connected to VM15 LCD B/D from base bracket.
Figure 11.17 20 How to disconnect 4.3inch LCD and VM15 LCD B/D
23) Connect the replaced 4.3inch LCD and VM15 LCD B/D
Figure 11.17 21 How to connect New 4.3inch LCD and VM15 LCD B/D
146
Service Manual Appendix
25) Fasten the sem's screw with spring washer(M3x6mm-4ea) on the VM15 LCD B/D
Figure 11.17 23 Position to fasten the screws on the VM15 LCD B/D
Figure 11.17 24 How to place the Base Bracket inside the ESN CASE
147
Service Manual Appendix
29) Fasten the sem's screw with spring washer(M3x6mm-6ea) and the PCB plastic
support(M3x20mm-1ea) on the Base B/D
Figure 11.17 27 Position to fasten the screws and the support on Base B/D
148
Service Manual Appendix
31) Fasten the countersunk head screw(M3x6mm-2ea) on the SIDE STD COVER
32) Connect Speaker Cable, ECG Power Cable, SPO2 Power Cable, ECG Out Cable, 1Temp Cable,
Masimo Side Cable, Membrane Cable, 4.3 LCD FFC Cable
149
Service Manual Appendix
34) Push MCU B/D and confirm that the trigger is in place.
150
Service Manual Appendix
36) Place Heatsink to cover the CPU chip of MCU B/D. Place the Heat Sink Fix Bracket to fit the
screw hole of the PCB support.
Figure 11.17 35 Position to fasten the support on Heat Sink Fix Bracket
151
Service Manual Appendix
39) Fasten the sem's screw with spring washer(M3x6mm-1ea) and PCB metal support(M3x20mm-
2ea) from ECG B/D.
Figure 11.16 37 Position to fasten the screws and support on ECG B/D
152
Service Manual Appendix
40) Connect the 1Temp Cable, ECG Mini-Candal Cable, EKG Mini-Candal Cable, ECG Power Cable,
ECG Out Cable on the ECG B/D.
41) If the installed SpO2 B/D is VOTEM SpO2, Proceed to 11.23-5) of SpO2 B/D Replacement
Manual. 67
If the installed SpO2 B/D is Masimo SpO2, Proceed to 11.24-5) of SpO2 B/D Replacement
Manual.
153
Service Manual Appendix
154
Service Manual Appendix
155
Service Manual Appendix
156
Service Manual Appendix
157
Service Manual Appendix
14) Lay the PWR B/D onto the INNER ASS’Y and Fasten sem's screw with spring washer(M3x6mm-
4ea).
158
Service Manual Appendix
159
Service Manual Appendix
160
Service Manual Appendix
161
Service Manual Appendix
162
Service Manual Appendix
3) Unfasten SUPPORT.
163
Service Manual Appendix
Figure 11.19 7 Position to fasten the screw on Heat Sink Fix Bracket.
164
Service Manual Appendix
2) If the installed SpO2 B/D is VOTEM SpO2, Proceed to 11.23-2) of SpO2 B/D Replacement
Manual.
If the installed SpO2 B/D is Masimo SpO2, Proceed to 11.24-2) of SpO2 B/D Replacement
Manual.
4) Remove the 1Temp cable, EKG cable, ECG cable, ECG-PWR cable, ECGOUT cable on the ECG
B/D.
165
Service Manual Appendix
5) Unfasten the sem's screw with spring washer(M3x6mm-1ea) and PCB metal support
(M3x20mm-2ea) from ECG B/D
Figure 11.20 3 Position to unfasten the screw and support on ECG B/D
Figure 11.20 5 Position to unfasten the support on Heat Sink Fix Bracket
166
Service Manual Appendix
Figure 11.20 6 How to remove Heatsink and Heat Sink Fix Bracket
167
Service Manual Appendix
10) Pull the trigger. Check that the MCU B/D is floating.
168
Service Manual Appendix
169
Service Manual Appendix
13) Push MCU B/D and confirm that the trigger is in place.
170
Service Manual Appendix
15) Place Heatsink to cover the CPU chip of MCU B/D. Place the Heat Sink Fix Bracket to fit the
screw hole of the PCB support.
Figure 11.20 14 Position to fasten the support on Heat Sink Fix Bracket
171
Service Manual Appendix
18) Fasten the sem's screw with spring washer(M3x6mm) and the PCB metal support(M3x20mm-
2ea) from ECG B/D.
Figure 11.20 16 Position to fasten the screws and support on ECG B/D
19) Connect the 1Temp Cable, ECG Mini-Candal Cable, EKG Mini-Candal Cable, ECG Power Cable,
ECG Out Cable on the ECG B/D.
20) If the installed SpO2 B/D is VOTEM SpO2, Proceed to 11.23-5) of SpO2 B/D Replacement
Manual. 67
If the installed SpO2 B/D is Masimo SpO2, Proceed to 11.24-5) of SpO2 B/D Replacement
Manual.
172
Service Manual Appendix
3) Detach each cable(Careful of connector damage when removing the FFC Cable)
173
Service Manual Appendix
174
Service Manual Appendix
6) Fixed MAIN B/D at Front ASS’Y with sem's screw with spring washer(M3x6mm-4ea).
Figure 11.21 4 Replace the Main B/D and fix the screws.
7) Fix each Cable (Careful of connector damage when connecting the FFC Cable)
175
Service Manual Appendix
176
Service Manual Appendix
2) If the installed SpO2 B/D is VOTEM SpO2, Proceed to 11.23-2) of SpO2 B/D Replacement
Manual.
If the installed SpO2 B/D is Masimo SpO2, Proceed to 11.24-2) of SpO2 B/D Replacement
Manual.
4) Remove the 1Temp cable, EKG cable, ECG cable, ECG-PWR cable, ECGOUT cable on the ECG
B/D.
177
Service Manual Appendix
5) Unfasten the sem's screw with spring washer(M3x6mm-1ea) and PCB metal
support(M3x20mm-2ea) from ECG B/D
Figure 11.22 3 Position to unfasten the screw and support on ECG B/D
Figure 11.22 5 Position to unfasten the support on Heat Sink Fix Bracket
178
Service Manual Appendix
Figure 11.22 6 How to remove Heatsink and Heat Sink Fix Bracket
179
Service Manual Appendix
10) Pull the trigger. Check that the MCU B/D is floating.
180
Service Manual Appendix
12) Remove Speaker Cable, ECG Power Cable, SPO2 Power Cable, ECG Out Cable, 1Temp Cable,
Masimo Side Cable, Membrane Cable, 4.3 LCD FFC Cable
Figure 11.22 10 How to remove the Cable from the Base B/D(detailed 4.3 LCD FFC Cable)
181
Service Manual Appendix
13) Unfasten the countersunk head screw(M3x6mm-2ea) on the SIDE STD COVER
15) Unfasten the sem's screw with spring washer(M3x6mm-6ea) and the PCB plastic
support(M3x20mm-1ea) on the Base B/D
Figure 11.22 13 Position to unfasten the screws and the support on Base B/D
182
Service Manual Appendix
17) Place Base B/D to fit the screw hole of the Base Bracket
18) Fasten the sem's screw with spring washer(M3x6mm-6ea) and the PCB plastic
support(M3x20mm-1ea) on the Base B/D
Figure 11.22 16 Position to fasten the screws and the support on Base B/D
183
Service Manual Appendix
20) Fasten the countersunk head screw(M3x6mm-2ea) on the SIDE STD COVER
21) Connect Speaker Cable, ECG Power Cable, SPO2 Power Cable, ECG Out Cable, 1Temp Cable,
Masimo Side Cable, Membrane Cable, 4.3 LCD FFC Cable
184
Service Manual Appendix
23) Push MCU B/D and confirm that the trigger is in place.
185
Service Manual Appendix
25) Place Heatsink to cover the CPU chip of MCU B/D. Place the Heat Sink Fix Bracket to fit the
screw hole of the PCB support.
Figure 11.22 24 Position to fasten the support on Heat Sink Fix Bracket
186
Service Manual Appendix
28) Fasten the sem's screw with spring washer(M3x6mm-1ea) and PCB metal support(M3x20mm-
2ea) from ECG B/D.
Figure 11.22 26 Position to fasten the screws and support on ECG B/D
29) Connect the 1Temp Cable, ECG Mini-Candal Cable, EKG Mini-Candal Cable, ECG Power Cable,
ECG Out Cable on the ECG B/D.
187
Service Manual Appendix
30) If the installed SpO2 B/D is VOTEM SpO2, Proceed to 11.23-5) of SpO2 B/D Replacement
Manual. 67
If the installed SpO2 B/D is Masimo SpO2, Proceed to 11.24-5) of SpO2 B/D Replacement
Manual.
188
Service Manual Appendix
2) Remove SpO2 Connector Cable and SPO2 Power Cable of SpO2 B/D.
189
Service Manual Appendix
190
Service Manual Appendix
7) Connect SpO2 Connector Cable and SPO2 Power Cable to SpO2 B/D
191
Service Manual Appendix
2) Remove Masimo Side Cable and SPO2 Power Cable of IPWR B/D.
192
Service Manual Appendix
7) Connect Masimo Side Cable and SPO2 Power Cable to IPWR B/D
193
Service Manual Appendix
2) If the installed SpO2 B/D is VOTEM SpO2, Proceed to 11.23-2) of SpO2 B/D Replacement
Manual.
If the installed SpO2 B/D is Masimo SpO2, Proceed to 11.24-2) of SpO2 B/D Replacement
Manual.
Figure 11.25 1 How to remove the NIBP Power Cable and NIBP Hose.
194
Service Manual Appendix
195
Service Manual Appendix
7) Separate NIBP B/D and NIBP FIX BKT. Replace NIBP B/D.
8) Place the replaced NIBP B/D on the NIBP FIX BKT according to the direction. (VOTEM NIBP)
9) Fasten the sem's screw with spring washer(M3x6mm-4ea) on the NIBP B/D
196
Service Manual Appendix
11) Place the NIBP FIX BKT on the Base Bracket according to screw hole.
197
Service Manual Appendix
12) Fasten the countersunk head screw(M3x8mm-2ea) on the NIBP FIX BKT.
Figure 11.25 11 How to fix the NIBP Power Cable and NIBP Hose.
14) If the installed SpO2 B/D is VOTEM SpO2, Proceed to 11.23-5) of SpO2 B/D Replacement
Manual.
If the installed SpO2 B/D is Masimo SpO2, Proceed to 11.24-5) of SpO2 B/D Replacement
Manual.
198
Service Manual Appendix
2) If the installed SpO2 B/D is VOTEM SpO2, Proceed to 11.23-2) of SpO2 B/D Replacement
Manual.
If the installed SpO2 B/D is Masimo SpO2, Proceed to 11.24-2) of SpO2 B/D Replacement
Manual.
Figure 11.26 1 How to remove the Suntech Power Cable and NIBP Hose.
199
Service Manual Appendix
7) Separate NIBP B/D and NIBP FIX BKT-Suntech. Replace NIBP B/D.
200
Service Manual Appendix
8) Place the replaced NIBP B/D on the NIBP FIX BKT-Suntech according to the direction.
201
Service Manual Appendix
11) Place the NIBP FIX BKT-Suntech on the Base Bracket according to screw hole.
12) Fasten the countersunk head screw(M3x8mm-2ea) on the NIBP FIX BKT-Suntech.
202
Service Manual Appendix
Figure 11.26 11 How to fix the Suntech Power Cable and NIBP Hose.
14) If the installed SpO2 B/D is VOTEM SpO2, Proceed to 11.23-5) of SpO2 B/D Replacement
Manual. 67
If the installed SpO2 B/D is Masimo SpO2, Proceed to 11.24-5) of SpO2 B/D Replacement
Manual.
203
Service Manual Appendix
3) If the installed SpO2 B/D is VOTEM SpO2, Proceed to 11.23-2) of SpO2 B/D Replacement
Manual.
If the installed SpO2 B/D is Masimo SpO2, Proceed to 11.24-2) of SpO2 B/D Replacement
Manual.
.
204
Service Manual Appendix
5) Remove the 1Temp cable, EKG cable, ECG cable, ECG-PWR cable, ECGOUT cable on the ECG
B/D.
6) Unfasten the sem's screw with spring washer(M3x6mm-1ea) and PCB metal
support(M3x20mm-2ea) from ECG B/D
Figure 11.27 5 Position to unfasten the screws and support on ECG B/D
205
Service Manual Appendix
9) Fasten the sem's screw with spring washer(M3x6mm-1ea) and PCB metal support(M3x20mm-
2ea) from ECG B/D.
Figure 11.27 8 Position to fasten the screws and support on ECG B/D
206
Service Manual Appendix
10) Connect the 1Temp Cable, ECG Mini-Candal Cable, EKG Mini-Candal Cable, ECG Power Cable,
ECG Out Cable on the ECG B/D.
11) If the installed SpO2 B/D is VOTEM SpO2, Proceed to 11.23-5) of SpO2 B/D Replacement
Manual.
If the installed SpO2 B/D is Masimo SpO2, Proceed to 11.24-5) of SpO2 B/D Replacement
Manual.
207
Service Manual Appendix
208
Service Manual Appendix
4) Fixed new 3-TEMP B/D at MODULE CASE with sem's screw with spring washer(M3x6mm-4ea).
209
Service Manual Appendix
210
Service Manual Appendix
211
Service Manual Appendix
4) Fixed new IBP/ICO B/D at MODULE CASE with sem's screw with spring washer(M3x6mm-4ea).
Figure 11.29 4 Replace the IBP/ICO B/D and Fasten the screws.
212
Service Manual Appendix
213
Service Manual Appendix
Figure 11.30 3 Position to unfasten the screws on qCON B/D and How to detach cable
214
Service Manual Appendix
4) Fixed cable and new qCON B/D at MODULE CASE with sem's screw with spring
washer(M3x6mm-4ea).
Figure 11.30 4 How to fix the qCON Power Cable and Position to fix the screws
215
Service Manual Appendix
216
Service Manual Appendix
217
Service Manual Appendix
5) Detach cable.
Figure 11.31 5 Replace the KEY B/D and How to connect the connector
218
Service Manual Appendix
219
Service Manual Appendix
220
Service Manual Appendix
5) Place new EXT B/D and Fasten the sem's screw with spring washer(M3x6mm-4ea).
Figure 11.32 4 Position the new EXT B/D and fasten the screws.
221
Service Manual Appendix
Figure 11.32 5 How to fix the EXT FFC cable in EXT B/D
222
Service Manual Appendix
2) Remove USB cable and Unfasten the sem's screw with spring washer(M3x6mm-4ea).
Figure 11.33 1 Position of the USB cable and screws to be removed from the USB B/D.
223
Service Manual Appendix
4) Connect USB cable and Fasten the sem's screw with spring washer(M3x6mm-4ea).
224
Service Manual Appendix
225
Service Manual Appendix
Figure 11.34 4 How to fix the 15’’ Touch FFC cable in TCH B/D
6) Fix TCH B/D at Front ASS’Y with sem’s screw with spring washer(M3x6mm-2ea)..
Figure 11.34 6 How to fix the Touch FFC cable in TCH B/D
226
Service Manual Appendix
227
Service Manual Appendix
228
Service Manual Appendix
4) Fixed new BLT B/D with sem's screw with spring washer(M3x6mm-4ea).
229
Service Manual Appendix
230
Service Manual Appendix
231
Service Manual Appendix
5) Fixed VGA B/D at Front ASS’Y with sem’s screw with spring washer(M3x6mm-4ea)
232
Service Manual Appendix
233
Wall Mount Setup
1) Place wall mount fix bracket on the VM15
2) Fasten the pan washer head screw(M3x12mm-4ea) of wall mount fix bracket screw hole on the
bottom of VM15.
Figure 11.37 2 Position to fix the screws on Wall mount fix bracket(VM15 bottom)
3) Fasten the countersunk head screw(M3x8mm-2ea) of wall mount fix bracket screw hole on the
rear of VM15
Figure 11.37 3 Position to fix the screws on Wall mount fix bracket(VM15 rear)
234
4) Place the wall mount slide bracket and Fasten the round head truss screw(M4x9mm-4ea).
Figure 11.37 4 Wall mount slide bracket position and screw fixing position.
5) Depending on the wall type or the user's convenience, the wall mount is fixed to the wall using
screw anchor and M4 tapping screws.
235
6) Slide Wall mount slide bracket into the wall mount.
236
11.38 Patient Monitor Cleaning
Warning! Caution!
• Cleaning is a mixture of warm water and detergent. The ratio is 1L(water) : 15ml(detergent)
Then dampen a clean cloth with the mixed solution, wipe the contaminated surface and dry
it. After cleaning, please dry the product.
• After cleaning clean the mixture with a cloth dampened with clean water.
237
CMC Medical Devices & Drugs S.L.
E-Mail : Info@cmcmedicaldevices.com
238