Professional Documents
Culture Documents
Anthem Generator: Installation, Operation, and Service Manual
Anthem Generator: Installation, Operation, and Service Manual
Table of
Contents
Safety
4 Acceptance
Testing
5 Operation
6 Periodic
Maintenance
7 Trouble-
shooting
8 Component-
Replacement
9 Electrical
Schematics
P/N 8000-ANTHEM-U
Revision G, June 25, 2018 10 Illustrated
Parts List
Copyright © 2018, Del Medical Inc. All rights reserved.
This document is the property of Del Medical Inc. and contains confidential and proprietary information
owned by Del Medical Inc. Any unauthorized copying, use or disclosure of this document without the prior
written permission of Del Medical Inc. is strictly prohibited.
Revision History
Chapter 1 — Introduction
1.1 Document Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1
1.2 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1
1.3 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–2
1.3.1 General Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–2
1.3.2 Anthem Generator Models and Options . . . . . . . . . . . . . . . . . . . . . . 1–3
1.3.3 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–3
1.3.4 Architecture . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–4
1.3.5 X-Ray Tube . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–4
1.3.6 Power Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–4
1.3.7 In-Table Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–4
1.4 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–5
1.5 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–7
1.5.1 Power Specifications, by Anthem Model. . . . . . . . . . . . . . . . . . . . . . 1–7
1.5.2 General Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–8
1.5.3 AEC Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–8
1.5.4 System Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–9
1.6 X-Ray Tube Compatibility Listing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–10
1.7 Identification Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–10
1.7.1 Generator Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–11
1.7.2 Equipment Label Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–11
1.8 Optional Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–13
1.9 Functional Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–14
1.9.1 Prep and Ready States . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–15
Chapter 2 — Installation
2.1 Pre-Installation Planning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–1
2.1.1 X-ray Room Planning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–1
2.1.2 Space Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–2
2.1.3 Power Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–3
2.1.4 Wiring Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–4
2.1.5 Shipment Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–5
2.1.6 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–5
2.1.7 In-Table Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–6
2.1.8 Installation Tools Required: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–6
2.2 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–7
2.2.1 External Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–7
2.2.2 Power Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–9
2.2.3 Connecting Components to the Generator . . . . . . . . . . . . . . . . . . . 2–16
2.2.4 Wall Bucky Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–19
2.2.5 Wallstand Grid Cabinet Connections. . . . . . . . . . . . . . . . . . . . . . . . 2–20
2.2.6 Table Bucky Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–21
2.2.7 Table Grid Cabinet Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–22
2.2.8 Tube Rotor Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–22
2.2.9 Collimator Power Connections (If Required) . . . . . . . . . . . . . . . . . 2–23
2.2.10 Table/Wallstand Lock and Other DC Connections . . . . . . . . . . . 2–24
2.2.11 Collimator Interlock and Door Interlock Connections . . . . . . . . 2–24
2.2.12 Digital Imaging Receptor Connections . . . . . . . . . . . . . . . . . . . . . 2–25
2.2.13 Room-In-Use Indicator Light Connection . . . . . . . . . . . . . . . . . . . 2–28
2.2.14 Control Console Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–29
2.2.15 High Tension (HT) Cable Connections . . . . . . . . . . . . . . . . . . . . . 2–29
2.3 Initial Testing and Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–32
2.3.1 Test Equipment Required. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–33
2.3.2 Inspection of the HT Tank . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–33
2.3.3 Inspection of Power Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–33
2.3.4 Inspection of Control Console . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–34
2.3.5 Operational Testing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–35
2.3.6 Verify the Door Interlock Safety Function. . . . . . . . . . . . . . . . . . . . 2–40
2.3.7 The Service Mode Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–40
2.3.8 Installing the Del APR Utility Software . . . . . . . . . . . . . . . . . . . . . . 2–47
2.3.9 Initial Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–48
iv Table of Contents
Anthem Installation, Operation, & Service Manual
Chapter 5 — Operation
5.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–1
5.2 Safety Precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–1
5.3 Operator Control Console . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–2
5.3.1 Control Console Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–2
5.3.2 Display Parameters and Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–3
5.4 Power On/Off Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–6
5.4.1 Powering Up The Anthem Generator . . . . . . . . . . . . . . . . . . . . . . . . 5–6
5.4.2 Powering Down the Anthem Generator . . . . . . . . . . . . . . . . . . . . . . 5–6
5.5 Operating Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–6
5.5.1 Manual Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–6
5.5.2 APR (Anatomically Programmable Radiography). . . . . . . . . . . . . . 5–8
Table of Contents v
Anthem Installation, Operation, & Service Manual
Chapter 7 — Troubleshooting
7.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–1
7.2 Discharge Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–1
7.3 Troubleshooting Tables and Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–1
7.4 Fault Conditions and Symptoms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–2
7.5 Console Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–3
7.5.1 Interlock Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–41
7.6 Frequently Asked Questions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–43
7.7 Test Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–45
7.7.1 Filament Board Test Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–45
7.7.2 Power Relay Board Test Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–47
7.7.3 Main Control Board Test Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–50
7.7.4 Console Board Test Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–56
7.7.5 IPM Driver Board Test Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–57
7.8 LED Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–57
vi Table of Contents
Anthem Installation, Operation, & Service Manual
Introduction
Del Medical X-ray equipment meets high standards of performance and
reliability. We enforce strict quality control techniques to eliminate the
potential for defects and hazards in our products.
This equipment provides an adjustable X-ray source for acquiring images of
parts of a patient’s anatomy. Use of this equipment in any other fashion may
lead to serious personal injury.
These safety guidelines are intended to educate the operator on safety issues
to operate and maintain the Anthem Generator in a safe manner.
Warning
Read and fully understand this manual before operating this
equipment. Failure to follow operating instructions can result in
serious injury or death.
Statement of Liability
To prevent excess radiation exposure (primary or secondary) to patients or
staff, this equipment must be operated and serviced by trained personnel
knowledgeable in the required safety precautions. While this equipment is
designed for safe use, improper operation can cause serious injury or
equipment damage. The manufacturer or its agents and representatives
assume no responsibility for the following:
1. Injury or danger to any person from X-ray exposure.
2. Overexposure of patients or staff to X-radiation due to improper
techniques or procedures.
3. Injury or danger from improper use of the adjustable column or tube head
assembly.
4. Problems or hazards resulting from failure to maintain the equipment as
specified in this manual.
Warning Statements
• indicate hazards or unsafe practices that can cause severe injury or death.
• appear in bold type.
• have a triangular symbol enclosing an exclamation point, with the word
Warning.
• precede the step or piece of information to which they refer.
• appear as the following example:
Warning
This text describes special safety precautions to avoid unsafe practices
that can result in severe injury or death.
Caution Statements
• indicate hazards or unsafe practices that can cause minor injury, or
product or property damage.
• appear in bold type.
• have a triangular symbol enclosing an exclamation point, with the word
Caution.
• precede the step or piece of information to which they refer.
• appear as the following example:
Caution
This text describes special safety precautions to avoid unsafe practices
that can result in personal injury or product or property damage.
Warning
To avoid serious electric shock, disconnect all generator power and
follow the discharge procedure before servicing, disconnecting high-
tension cables, or removing access panels. Avoid dropping any tools
or objects in the unit.
Warning
To avoid the dangerous and cumulative effects of X-rays, minimize
radiation exposure to any and all patients and staff.
Radiation Protection
Serious detrimental health effects can result from exposure to both short-
term/high-level and long-term/low-level X-rays. Become familiar with the
effects of both types of radiation exposure, and take appropriate measures to
minimize exposure. Some effects of X-radiation are cumulative over a period
of months or years. Most importantly, avoid exposure to the primary beam at
all times.
X-ray facilities typically use lead to shield the radiologist's workstation. The
typical workstation partition is generally effective in reducing residual
radiation to a safe level. Windows in the partition are typically glazed with
lead glass and provide effective protection. Do not leave the protected area
during exposure.
Any object in the path of the primary beam produces scattered radiation,
which can result in a substantial radiation dose to unprotected persons. To
minimize exposure to scattered radiation, use movable lead screens, and
lead-impregnated gloves and aprons. These protective devices must contain a
lead (or equivalent) thickness of at least 0.25 mm. Use such protective devices
for all operators, observers, and/or servicing personnel exposed to radiation
fields of five or more milli-Roentgens per hour.
The X-ray system produces X-rays only when high voltage is applied to the
X-ray tube. When the high voltage is removed, X-ray emission ceases
immediately.
When the distance between the X-ray tube focal spot and the patient’s skin is
decreased, the radiation dose to the patient increases.
Ensure audio and visual communication between operator and patient even
during exposure. This may require video cameras or an intercom.
To minimize risk, ensure that all operating personnel are educated in the
hazards of radiation and understand the safety requirements for X-ray
operation. Review the manuals for each component in the system.
Note: This system may include an anti-scatter grid. For pediactric applications, the
grid can be removed by a qualified service technician. Refer to the service manual for
the pertinent table or wall stand.
Warning
Ensure exposure parameters are set to within safety limits.
Caution
Incorrectly positioning the X-ray tube and collimator can cause the
X-ray field to be misaligned, resulting in an unacceptable image and
the need to re-expose.
Pediatric Use
Exercise special care when imaging patients outside the typical adult size
range, especially smaller pediatric patients whose size does not overlap the
adult size range (e.g., patients less than 110 lb [50 kg] in weight and 59 in [150
cm] in height).
Ionizing radiation exposure is of particular concern in pediatric patients because:
• For certain organs and tumor types, younger patients are more
radiosensitive than adults (i.e., the cancer risk per unit dose of ionizing
radiation is higher for younger patients).
• Use of equipment and exposure settings designed for adults of average
size can result in excessive and unnecessary radiation of smaller patients.
• Younger patients have a longer expected lifetime over which the effects of
radiation exposure may manifest as cancer.
To help reduce the risk of excessive radiation exposure, follow the ALARA (As
Low As Reasonably Achievable) principle and seek to reduce radiation doses to
only the amount necessary for clinically adequate images.
If available on your system, utilize the following features (and their pediatric-
appropriate settings):
• filters
• grid or Bucky
• protocols or exposure indices
• variable focal spot size
• Automatic Exposure Control (AEC)
• DAP meter
Also, ensure that all testing and calibration procedures include measurements
and adjustments for pediatric settings.
For additional guidance, refer to these resources:
• The Image Gently Campaign: http://www.imagegently.org.
• FDA resource on Pediatric X-ray Imaging: http://www.fda.gov/
Radiation-EmittingProducts/RadiationEmittingProductsandProcedures/
MedicalImaging/ucm298899.htm
Restrictions on Use
The hardware specified for intended applications of the Anthem Generator
has been selected, tested, and verified by Del Medical to meet applicable
regulatory agency requirements for those countries where it is offered for sale
with respect to its intended applications.
Warning
Do not install components or accessories not intended for use with
this system. Failure to comply can cause injury or equipment damage.
You are responsible for ensuring that the application and use of this
equipment does not compromise the patient contact rating of any equipment
used in the vicinity of or in conjunction with the system.
Grounding
Provide protective earth grounding for all components. All associated
components must be grounded to a common bus according to national and
local electrical codes. Ensure that the earth grounding connections between
this equipment and the host system are maintained at all times.
Hazardous Materials
The X-ray tube, PCBs, and collimator used with this equipment contain lead
and oil. Refer to the equipment manuals or contact Del Medical for additional
information regarding use and disposal of hazardous materials.
Introduction
1
1.1 Document Information
This manual provides installation, operation, calibration and service
information for the Anthem Generator. For specific information regarding the
operation and service of the tube, collimator, stands, tables, and other
components, see their respective manuals.
1.2 Introduction
The Anthem generator is a high-frequency X-ray generator designed for
general radiographic applications. Features include a user-friendly operator’s
Control Console, programmable APR operation, and optional support of
Digital Radiography, Automatic Exposure Control (AEC), and veterinary
APR techniques.
All persons operating or servicing this equipment must read this manual
before performing tasks. You must have a thorough understanding of the
Anthem generator and its proper use before making radiographic exposures.
1 Introduction 1-1
Anthem Installation, Operation, & Service Manual
1.3 Description
The DR-ready Anthem generator is intended for use as part of radiographic
stationary X-ray systems. As part of a typical system, the generator provides
power and control for other parts of the system, as shown in Figure 1-2.
X-ray
tube
high collimator
voltage
supply voltage ion image
Anthem chamber receptor
generator (AEC)
table/wallstand,
interlock functions, or
Control Bucky/DR prep/expose
Console interface
PC or DR
Workstation
Handswitch or
Footswitch
Figure 1-2. Anthem generator integrated into an X-ray system
1-2 1 Introduction
Anthem Installation, Operation, & Service Manual
1.3.3 Operation
The Anthem generator operates in one of four modes:
• APR (Anatomically Programmed Radiography). Select from pre-
programmed techniques based on body region, examination type, and
patient size. You can then modify mA, kV, and/or time parameters and
save these customized settings. Veterinary techniques are available for
veterinary-configured systems.
• Manual 2 point. Set mAs and kV parameters. The Anthem software
automatically chooses the highest mA and shortest time to reduce motion
and deliver high quality images.
• Manual 3 point. The traditional 3 point selection of kV, mA and time
accommodates the needs of specialized views and the preference of some
technologists.
• DR (Digital Radiography). This optional mode replaces the APR mode.
Use the DR mode when interfacing a DR workstation and DR detector
devices. Technique parameters on the DR workstation also display at the
Control Console in real time.
The Anthem generator Control Console can also be placed in the following
support modes:
• Administrator Mode. Use this mode to set default parameter values.
• Service Mode. For use by service personnel only. Use this mode to set
options, configure the system, and perform calibration.
1 Introduction 1-3
Anthem Installation, Operation, & Service Manual
1.3.4 Architecture
The X-ray generator consists of an upper main power cabinet, a lower cabinet,
and a control console.
The upper main power cabinet contains the inverter, the filament drivers, the
power supply, the Main Control board, and interface connections to the room
equipment.
The lower cabinet contains the HT Tank, and, for 3-phase systems, the line
match transformers.
The control console allows the operator to select/program technique factors,
configure features, and prep/initiate an X-ray exposure via handswitch,
footswitch, or console panel buttons.
1-4 1 Introduction
Anthem Installation, Operation, & Service Manual
1.4 Dimensions
17.5”
444 mm
28”
711 mm
19.5”
495 mm
27”
686 mm
For in-table generators, the lower cabinet is replaced with 2 enclosures of the
following dimensions: .
1 Introduction 1-5
Anthem Installation, Operation, & Service Manual
Ø 0.281” (x4)
Ø 7 mm (x4)
2.75”
70 mm
4.0”
102 mm
Figure 1-4. Optional control console wall bracket mounting hole dimensions
36.5”
930 mm
8.3”
211 mm
Anti-Tip
11.0” Bracket
230 mm
1-6 1 Introduction
Anthem Installation, Operation, & Service Manual
8.5”
216 mm
21.5”
546 mm
15.0”
381 mm
1.5 Specifications
1 Introduction 1-7
Anthem Installation, Operation, & Service Manual
1-8 1 Introduction
Anthem Installation, Operation, & Service Manual
D.H.H.S. Compliance
The Anthem High Frequency x ray generator and its associated accessories
are certified components conforming to all applicable radiation performance
standards as outlined by the Food and Drug Administration, Code of Federal
Regulations, Title 21, Subchapter J, at the time of manufacture.
Regulatory Conformance
Complies with FDA radiation performance standards under 21 CFR
Subchapter J, in effect at the date of manufacture.
1 Introduction 1-9
Anthem Installation, Operation, & Service Manual
1-10 1 Introduction
Anthem Installation, Operation, & Service Manual
1 Introduction 1-11
Anthem Installation, Operation, & Service Manual
Alternating Current
1-12 1 Introduction
Anthem Installation, Operation, & Service Manual
The product is RoHS compliant. RoHS marks can vary greatly from
one product to the next.
1 Introduction 1-13
Anthem Installation, Operation, & Service Manual
1-14 1 Introduction
Anthem Installation, Operation, & Service Manual
Control of kV, mA, and Filament Preheat levels begins with analog reference
voltages, which are decoded by D/A converters on the Main Control board,
and then sent out to the kV Control Circuit, mA Control Circuit, and
Filament/Rotor control circuit.
1 Introduction 1-15
Anthem Installation, Operation, & Service Manual
is sent to Filament board to start the rotor boost. After 1.5 seconds, rotor run
pulses are sent to the Filament board to turn the rotor and maintain the speed
at 3400 RPM.
1.9.5 Interlocks
The Anthem generator has 4 interlocks that prohibit an exposure when
external devices are not ready, as indicated by closed contacts. These
interlocks are:
• door
• collimator
• wall Bucky
• table Bucky
The door and collimator interlocks cannot be overridden or disabled in the
console.
1.10 Abbreviations
AEC Automatic Exposure Control
AP Anteroposterior
APR Anatomically Programmable Radiography
AWG American Wire Gauge
BTU British Thermal Unit
CAN Controller Area Network
CDRH Center for Devices and Radiological Health
CE Communautés Européennes
DAC Digital to Analog Converter
DFMTS Digital Floor Mounted Tubestand
DR Digital Radiography
ga Gauge
hPa Hecto Pascal
HT High Tension
ICRP International Commission on Radiological Protection
inHg Inches Mercury
IPM Intelligent Power Module
kN kiloNewton
kV kilovolt
kVp peak kilovoltage
kW kilowatt
LAT lateral
mA milliampere
mAs milliampere second
MeV mega-electron Volt
1-16 1 Introduction
Anthem Installation, Operation, & Service Manual
1.11 Glossary
Automatic Exposure Control (AEC) — AEC uses an ionization chamber or a
solid-state detector within the Bucky to command the X-ray generator to stop
the exposure when the correct image density is reached.
AEC field — Commonly known as ionizing vector, the AEC field measures the
dose of X-ray in the interested region.
Angle of tube rotation — This is the angle of deviation from the vertical
direction when the tube rotates, and ranges from -180 to 180 degrees.
Bucky/Grid Cabinet — The component that includes a film cassette, AEC,
moving or stationary grid, and related components. In a Digital Radiology
system, the Bucky contains a Digital Radiology detector instead of a film
cassette.
Collimator — Device that regulates the size and shape of the X-ray beam to
accurately localize the area of interest on the patient, while reducing overall
patient irradiation exposure.
Free Mode — Exposing position independent of a table or wallstand.
Generator — Device that supplies power to and controls the X-ray tube.
N — Newton, the amount of the net force required to accelerate a mass of one
kilogram at a rate of one meter per second.
kVp — Peak kilo-volts, the highest energy of X-rays emitted by an X-ray tube
(equal to the peak applied tube voltage).
LAT — Lateral view position for X-ray exposure.
Longitudinal — Head-to-foot positioning of focal spot.
mAs — Milliampere-seconds. Combined with kVp, it indicates the dose of
X-rays.
OBL — Oblique view position for X-ray exposure.
Portrait Position — The long dimension parallel to the patient axis.
1 Introduction 1-17
Anthem Installation, Operation, & Service Manual
SID — Source to Image Distance; the distance from the source of the X-rays to
the film or detector plane.
Table Mode — Exposing position where the patient lays on a table
Wall Mode — Exposing position where the patient stands in front of a
wallstand.
Technique Parameter — Any of the parameters describing the properties of
an X-ray beam exposure, including beam energy (kVp), beam intensity (mA),
duration (seconds), time-intensity factor (mAs), and at times, Source to Image
Distance (SID).
Transverse — Cross table (front-to-rear) travel of focal spot.
1-18 1 Introduction
2 Installation
Installation
2
2.1 Pre-Installation Planning
The Anthem generator is shipped partially assembled. Review all installation,
assembly and calibration instructions before beginning the installation. Save
all packing materials until after the installation is complete. Check that the
physical location, conduit arrangement and junction boxes are in readiness for
permanent installation of this equipment.
Before you begin installation, perform the following tasks:
• Consult with the client and/or client physical plant superintendent to
verify that the site has sufficient power, utility, and network access.
• Determine site-specific requirements, client requests, and room workflow
considerations, by reviewing the following documents, as available:
- the shipping order
- statement of work
- job orders
- client room drawings
- special written or unwritten requirements
2 Installation 2-1
Anthem Installation, Operation, & Service Manual
Caution
If the floor condition is unsuitable for a plumb and secure
installation, take corrective action with floor repair or modification.
Note: To fully open the front of the power module, an additional 12 3/4” of space is
required in front of the generator at the height of the upper cabinet.
21 1/2”
12 3/4”
27 1/2”
2-2 2 Installation
Anthem Installation, Operation, & Service Manual
21 1/2”
23”
27 1/2”
31 1/2”
Caution
Do not drill equipment mounting holes in concrete floors before
determining the location of tensioned wire ropes. Consult with
customer or responsible project engineer to locate and avoid drilling
through wire rope, which. can cause severe structural damage.
Warning
Lethal voltages may be present! Never work on energized circuits
during this installation unless specifically instructed to do so.
2 Installation 2-3
Anthem Installation, Operation, & Service Manual
Note: The minimum wire sizes listed above have been selected for the maximum wire
resistance allowed at maximum momentary full load current. To meet National
Electric Code Specifications, select a fuse or circuit breaker with an amperage rating of
at least 50% of the generator’s peak current demand.
2-4 2 Installation
Anthem Installation, Operation, & Service Manual
Caution
Attempts to attach accessory items other than those described in this manual
may damage circuitry. Do not use aluminum wire. All wiring MUST be
copper.
Warning
These instructions are intended for use by qualified installers only.
2.1.6 Unpacking
1. Move the pallet to the approximate position where generator will be
installed.
2. Cut the plastic straps. Remove the straps and the top cover.
3. Lift off the outer box shell.
4. Remove shrink wrapping from around the generator.
5. Slightly tilt the generator and gently “walk” the generator off of the pallet
onto the floor at its final location.
2 Installation 2-5
Anthem Installation, Operation, & Service Manual
Power Module
HT Tank
For 3-phase systems, install the line match transformer enclosure outside the
table within about 8 feet.
Input power connections (“Power Connections” on page 2-9) are made at the
rear of the Power Module. Make these connections before sliding the Power
Module into the table.
2-6 2 Installation
Anthem Installation, Operation, & Service Manual
2.2 Installation
Before starting the installation, study the “External Cables” section to become
familiar with Anthem connections to other equipment, and to verify that you
have all the required cables. Connections to specific Anthem connectors and
terminal blocks are identified in the installation procedures that follow.
Observe wire size and grounding requirements per local codes.
4 Aux 24Vdc/24Vac
5 In-Use Lamp
Power
Module 6 Room Interlock
Wallstand
Tube
Collimator
Ion Chamber
Receptor
12
Bucky
Power Mains
(1-Phase) 1
7
10
11
8
10 Ion Chamber
11 Receptor Table
9 Bucky
20
DR Sync Box
DR
2 3 13 14 15 Workstation
19
17
Power Mains Line Match
(3-Phase) 1 Transformers and HT Tank 16 18
Terminal Blocks
(3-Phase)
Handswitch/
Lower Cabinet Console Footswitch
2 Installation 2-7
Anthem Installation, Operation, & Service Manual
2-8 2 Installation
Anthem Installation, Operation, & Service Manual
Warning
Make sure that the power to the main power cable is turned off.
Route Main
Power Cable
Through This
Clamp
Figure 2-5. Terminal TB1 and ground terminal on rear of power module
2 Installation 2-9
Anthem Installation, Operation, & Service Manual
7. Route the main power cable (126-5317G1) through the center cable clamp
(Figure 2-5). Rotate the cable so that the ground wire runs flat along the
power module panel.
8. Connect the green ground wire from the main power cable to the ground
terminal (Figure 2-6), and reconnect the other previously disconnected
ground wires to the ground terminal. Secure the ground wires with the
nut.
Line 1
(White)
Ground
(Green)
Line 2
(Black)
9. Connect the main power cable white wire (Line 1) to the 0 VOLTS terminal.
10. Connect the main power cable black wire (Line 2) to the matching 208 or
240 Volt terminal. Refer to Table 2-3 for correct voltage connections.
208-220 VAC, or
TB1 nominal 220 VAC 230-240 VAC
1 (0 VOLTS) Line 1 Line 1
2 (208 VOLTS) Line 2 -
3 (240 VOLTS) - Line 2
4 (290 VOLTS) - -
5 (P1) - -
6 (P2) - -
Table 2-3. Incoming single phase main power TB1 connection
11. Tighten the main power cable clamp.
12. Reinstall the rear panel.
3-Phase Generators
1. Ensure that the power to the main power cable is turned off.
2. Route the main power cable (126-5318G1) to the generator location.
2-10 2 Installation
Anthem Installation, Operation, & Service Manual
4. Feed the main power cable (126-5318G1) to through the cable clamp and
hole on the back of the lower cabinet. See Figure 2-8.
Cable Clamp
Figure 2-8. Main power cable routed through the back of the generator.
5. Connect the main power cable’s green/yellow ground wire to the ground
stud on the bottom of the lower cabinet (Figure 2-9). There are two nuts on
the ground bolt. If there are 3 or more ground wires, separate them
between nuts.
2 Installation 2-11
Anthem Installation, Operation, & Service Manual
TB1, TB2,
and TB3
2-12 2 Installation
Anthem Installation, Operation, & Service Manual
Note: For 220 VAC, use the 208V terminal. For 230 VAC, use the 240V terminal.
7. Connect the 126-5318G1 main power cable as follows. See Figure 2-10.
Color Connector
green/yellow ground stud on base plate
black TB1, voltage verified in step 6
white TB2, voltage verified in step 6
red TB3, voltage verified in step 6
2 Installation 2-13
Anthem Installation, Operation, & Service Manual
3. At the rear of the power module, locate the 2 power cables that are coiled
and cable-tied to the rear panel:
• 126-5291G2 (thicker, black)
• 126-5292G2 (thinner, gray)
Cut the cable tie, uncoil the cables, and route them to the line match
transformer enclosure location.
4. On the line match transformer enclosure, remove the 4 cover mounting
screws (Figure 2-11), and remove the cover.
Cover Mounting
Screws (4)
Power Mains In
To Power Module
5. At the line match transformer enclosure, feed both cables through the
right-side (as viewed from the outside) cable clamp and hole, and into the
enclosure.
6. Connect the 126-5291G2 (thicker, black) cable as follows:
Color Connector
green/yellow ground stud on base plate
black TB1-280
white or tan TB2-280
red TB3-280
Label Connector
GND STUD (green/yellow) ground stud on base plate
LM 240V TB2-3 (across from TB2-240)
LM 240V RTN TB3-3 (across from TB3-240)
TMSW-1 TB1-TSW1
TMSW-2 TB3-TSW2
2-14 2 Installation
Anthem Installation, Operation, & Service Manual
Mains
Power In
Ground Stud
Figure 2-12. 3-phase line match transformer separate enclosure; view from top
with cover removed
Note: For 220 VAC, use the 208V terminal. For 230 VAC, use the 240V terminal.
9. Route the main power cable (126-5318G2) to the line match transformer
enclosure location.
10. Feed the main power cable to the line match transformer enclosure and
through the left cable clamp and hole on the back of the enclosure.
11. Connect the 126-5318G2 main power cable as follows:
Color Connector
green/yellow ground stud on base plate
black TB1, voltage verified in step 10 on page 2-10
white TB2, voltage verified in step 10 on page 2-10
red TB3, voltage verified in step 10 on page 2-10
2 Installation 2-15
Anthem Installation, Operation, & Service Manual
Side Screws
Front Screws
2. At the side of the power module, route the cables from the tube and image
receptor(s) through the cable routing bracket holes (Figure 2-14).
Figure 2-14. Cable routing through the power module’s cable routing bracket
2-16 2 Installation
Anthem Installation, Operation, & Service Manual
3. At the front of the power module, route the cables through the cable
clamps (Figure 2-15).
Cable Clamps
Figure 2-15. Cable routing through the power module’s front cable clamps
4. Connect the following cables to the Room Interface board per Table 2-4.
2 Installation 2-17
Anthem Installation, Operation, & Service Manual
TB1
TB6
TB3
TB4
TB5
TB2 TB8
Warning
Be sure to connect ground cables to the ground screw above the board
(Figure 2-17).
Do not connect any loads to this supply that require more than 4A.
Heavier loads can damage the power supply.
2-18 2 Installation
Anthem Installation, Operation, & Service Manual
Ground Connection
2 Installation 2-19
Anthem Installation, Operation, & Service Manual
Note: For the Villa Bucky, 120 VAC operation, install a jumper between terminals 1
and 6 at the Bucky.
Note: If the wallstand is not equipped with a Bucky, see “Wallstand Grid
Cabinet Connections” below.
2-20 2 Installation
Anthem Installation, Operation, & Service Manual
Note: For the Villa Bucky, 120 VAC operation, install a jumper between terminals 1
and 6 at the Bucky.
Note: If the table is not equipped with a Bucky, see “Table Grid Cabinet
Connections” on page 2-22.
2 Installation 2-21
Anthem Installation, Operation, & Service Manual
Figure 2-22. Tube rotor connections to terminal block TB3 (colors may vary)
Connection Signal
TB3-1 PHASE Stator Phase or AUX
TB3-2 MAIN Stator Main
TB3-3 COMMON Stator Common
TB3-4 THERMAL SW2 Thermal Switch *
TB3-5 THERMAL SW1 Thermal Switch Common *
*An isolated, normally closed (NC) set of contacts, mounted on the tube
housing, which opens if the housing exceeds normal operating temperatures.
If no housing thermal switch is present, install a jumper across TB3-4 and
TB3-5.
2-22 2 Installation
Anthem Installation, Operation, & Service Manual
Connection Signal
TB6-1 24 VAC return
TB6-2 24 VAC
Ground Green/Yellow
Table 2-7. Collimator power supply connections.
2 Installation 2-23
Anthem Installation, Operation, & Service Manual
Connection Signal
TB6-3 Marked +24
TB6-4 Marked -24
Ground Green/Yellow
Table 2-8. Table/wallstand lock 24 VDC power supply connections
Note: If the required jumpers are not installed, exposures are inhibited.
Connect the collimator interlock and/or door open interlock to terminal block
TB1 on the Room Interface board as shown in Figure 2-25.
2-24 2 Installation
Anthem Installation, Operation, & Service Manual
J2 J1
Figure 2-25. Jumper settings for system with no collimator interlock and/or door open
interlocks.
Connect the door interlock cable to terminals TB1-1 and TB1-3 on the Room
Interface board (Figure 2-26). At the other end, a switch closes when the door
closes.
2 Installation 2-25
Anthem Installation, Operation, & Service Manual
Connect digital imaging receptors to terminal blocks TB2 and TB8 on the
Room Interface board as shown in Figure 2-27.
TB2 TB8
Figure 2-27. DR ready and exposure signal connections
Connection Signal
TB2-1 TABLE READY +
TB2-2 TABLE READY -
TB2-3 WALL READY +
TB2-4 WALL READY -
TB8-1 WALL EXPOSURE REQUEST +
TB8-2 WALL EXPOSURE REQUEST -
TB8-3 TABLE EXPOSURE REQUEST +
TB8-4 TABLE EXPOSURE REQUEST -
TB8-5 PREP +
TB8-6 PREP -
Table 2-10. DR Ready and Exposure signal connections
2-26 2 Installation
Anthem Installation, Operation, & Service Manual
DR
Anthem detector
generator
room interface
sync box
DB9
Control
Console
DR workstation
2 Installation 2-27
Anthem Installation, Operation, & Service Manual
Preferred Method
TB1-6
15 A
24 VAC 120 VAC
TB1-7
Anthem Customer
Contactor X-Ray Warning
Room Supplied,
Light
Interface Not Included
Board With Anthem
Alternate Method
TB1-6
VAC or VDC
Note: Always refer to local electrical codes for correct wiring methods.
2-28 2 Installation
Anthem Installation, Operation, & Service Manual
Foot/Hand
Communication Switch Cable
Cable Connector J5 Connector J11
Ground
Connections
Strain Relief Brackets
Figure 2-30. Communication and footswitch/handswitch cable connection
2 Installation 2-29
Anthem Installation, Operation, & Service Manual
Caution
Failure to apply silicone paste can cause tube damage due to arcing.
8. Remove the protective plugs from the high-tension cable sockets in the
X-ray tube.
9. Important. Ensure the set screw on the threaded ring (Figure 2-32) is
backed out several turns so that it does not interfere with the tightening of
the cable or cause damage.
2-30 2 Installation
Anthem Installation, Operation, & Service Manual
10. Connect the HT cable connectors to the tube sockets. Tighten the ring
firmly by hand.
Set Screw
Caution
Connecting a high tension cable from the X-ray tube to the wrong
socket on the generator will destroy the tube. The cables marked “+”
must go to the socket marked “+”, and the cable marked “-” must go to
the socket marked “-.”
11. Remove the protective plugs from the HT Tank inside the generator.
12. Connect the HT cable going to the cathode (-) socket on the tube to the
CATHODE (-) socket on the HT Tank (Figure 2-33).
2 Installation 2-31
Anthem Installation, Operation, & Service Manual
13. Connect the HT cable going to the anode (+) socket on the tube to the
ANODE (+) socket on the HT Tank.
14. Tighten the rings (Figure 2-34) to secure the connections.
Rings
Caution
Connecting an HT cable to the wrong socket on the generator will
destroy the tube. Verify that cathode (-) on the generator connects to
cathode (-) on tube and that anode (+) on the generator connects to
anode (+) on tube.
Warning
Lethal voltages are present on the DC bus capacitors in this unit.
Exercise extreme caution when working near the long copper bus bars
attached to the two large capacitors in the base of the power module.
Always verify with a DC voltmeter that the capacitors are discharged
before attempting any service in this area.
2-32 2 Installation
Anthem Installation, Operation, & Service Manual
Warning
The large commutating capacitors mounted to the right on the rear
panel of the power module remain charged even when the power to
the unit is turned off.
Oil Level
Window
2 Installation 2-33
Anthem Installation, Operation, & Service Manual
2-34 2 Installation
Anthem Installation, Operation, & Service Manual
Filament Board
Figure 2-36. Board locations
2 Installation 2-35
Anthem Installation, Operation, & Service Manual
TP1 - TP3
TP4 - TP10
2-36 2 Installation
Anthem Installation, Operation, & Service Manual
TP2
TP3
TP7
TP8
TP9
TP1 TP6
TP4
TP5
2 Installation 2-37
Anthem Installation, Operation, & Service Manual
TP6A
TP22A
TP20 - TP22
TP47 - TP52
TP22B
TP47A
TP305
TP306
TP15 - TP17
TP23
TP53 - TP55
TP300
TP18, TP19 TP301
TP304
TP38 - TP42
TP31 - TP37
TP22C
TP26A TP302
2-38 2 Installation
Anthem Installation, Operation, & Service Manual
2 Installation 2-39
Anthem Installation, Operation, & Service Manual
If the control console does not display DOOR OPEN, check the following:
- at the Room Interface board, the interlock connections
- at the door, the switch wiring
3. While DOOR OPEN is still displayed, press the prep button. Verify that the
green READY light does not turn on. If the READY light does turn on,
contact Technical Support.
2-40 2 Installation
Anthem Installation, Operation, & Service Manual
Caution
Write down your password and store it in a safe place. If you forget
your password, you must send the Control Console back to the factory
for reprogramming.
Power
Switch
When the Control Console boots up, you are prompted for a password. It
shows the default password of 0000.
2 Installation 2-41
Anthem Installation, Operation, & Service Manual
1 2
3
Figure 2-42. Service Mode password
2-42 2 Installation
Anthem Installation, Operation, & Service Manual
2 Installation 2-43
Anthem Installation, Operation, & Service Manual
2-44 2 Installation
Anthem Installation, Operation, & Service Manual
2 Installation 2-45
Anthem Installation, Operation, & Service Manual
b. Select MONTH and use the data knob to select the correct month. Repeat
for DAY and YEAR. The console saves the changes automatically.
c. Select HOUR and use the data knob to select the correct hour. Repeat for
MIN and SEC. The console saves the changes automatically.
4. Press EXIT. The generator reboots to User Mode.
2-46 2 Installation
Anthem Installation, Operation, & Service Manual
you might see ADJUSTING APR. After a few seconds you are returned to
the Service Mode screen.
7. Press EXIT. The generator reboots to User Mode, with the new changes in
effect.
Required Hardware
• Laptop or desktop PC running Windows XP or later.
• Internet connection.
Installation
1. Navigate a web browser to http://support.delmedical.com and log in to your
Del Medical customer support account.
2. Go to Product Information > Generators > Anthem. Under Software,
click on ANTHEM APR Utility v. 4.5 and download the file for this link,
which is filename APR_Utility45.zip.
3. Unzip the Del APR Utility files to a temporary folder on your hard drive
(e.g., ...\MyDocuments\APRtemp)
4. Exit all other applications.
5. Go into folder APR_Utility_4.5 and run (double click) setup.exe.
6. Verify that the default destination of the Del APR Utility Software is as
shown below (C:\Program files\DelAprUtility, or for a 64-bit computer,
C:\Program Files (x86)).
2 Installation 2-47
Anthem Installation, Operation, & Service Manual
2-48 2 Installation
3 Calibration and Adjustments
Calibration and
Adjustments 3
3.1 Introduction
This chapter provides calibration and adjustment procedures for the
generator.
Warning
All service and maintenance, including the procedures described
within this chapter, are to be performed by qualified service
personnel only.
3.2.1 Preparations
Caution
The following procedures produces X-rays. Take precautions to
ensure the safety of all personnel in the area. Minimum precautions
are as follows:
- Wear lead aprons.
- Personnel remaining in the X-ray room during exposure must
be behind a lead shield.
- Minimize radiation scatter through doorways, walls and floor.
TP22A
Channel 1
TP6A
TP47A
Channel 2
TP22B
Channel 1 Channel 2
Vertical 2 V/div. Vertical 1 V/div.
Horizontal 10 ms/div. Horizontal 10 ms/div.
Mode Normal Sweep Mode Normal Sweep
Trigger Channel 1 - Positive
Note: If you are calibrating with only a scope, read the values of kV and mA as
follows:
2. On the top of the HT Tank, unscrew the 2 feedback board cover mounting
screws (Figure 3-3). Remove the cover.
Feedback Board
Cover Mounting
Screws
Binding Posts
Figure 3-4. Binding posts that hold copper jumper J2 on the feedback board.
Jumper
5. Connect the mAs meter, black wire to black plug and red wire to red plug
(Figure 3-6).
Black
Red
Initial Exposure
Warning
Lethal voltages are present on the large filter capacitors in this unit.
Exercise extreme caution when working near the long copper bus bars
attached to the large capacitors in the power module. Never perform
any service work inside the generator without first performing the
“Discharge Procedure” on page 6-2. Always verify with a DC
voltmeter that all three capacitors (two mounted to the back wall and
one on the bottom of the power module) are discharged before
attempting any service in this area.
Warning
Do not touch any part of the mAs meter during exposures.
Dangerously high voltages may be present.
Mode 2 Points
kV 40 kV
mAs 1 mAs
3.2.3 mA Calibration
1. From the Service Mode menu, press CALIBRATE. You see the Main
Calibration Screen.
2. Press mA. You see STEP 1 in the upper left corner, and the following values
displayed in the center:
LARGE
100mA, 75kV, 50ms
Stabilized at 1V
4. Select MA DAC, and use the data knob to adjust the mA DAC value to
achieve the 1.0 V target level (STEP 1 in Table 3-1).
3.2.4 kV Calibration
1. From the Main Calibration screen, select KV. You see STEP 1 in the upper
left corner, and the following values displayed in the center:
200mA, 40kV, 50ms
2. Make an exposure and observe the waveform on oscilloscope channel 1. If
the flat top of the wave (see Figure 3-8) is at 2.0 V, go to Step 5. If not,
proceed to the next step.
3. Select KV DAC, and use the data knob to adjust the kV DAC value to
achieve the 2.0 V target level (per Table 3-2).
Fil DAC too high Fil DAC too low Fil DAC good
Figure 3-9. oscilloscope screens for high, low, and good Fil DAC settings.
4. Select FIL DAC, and use the data knob to adjust the Fil DAC value to
achieve the target shape.
5. If needed, repeat the previous two steps.
6. When you have achieved the target waveform shape on the oscilloscope,
press SAVE.
7. Press NEXT PREHEAT. You see 2 appear above NEXT PREHEAT.
8. Repeat 3 through 7 above for the remaining pre-heat calibration steps.
Always press SAVE for each step.
9. When you press NEXT PREHEAT and you again see 1 appear above NEXT
PREHEAT, press GLOBAL ADJUST. In the center, you see ARE YOU SURE?
Press YES.
Required Equipment
• Laptop computer, running Windows 98 or later.
• SD memory card.
• SD memory card adapter (if PC or laptop has no SD card slot).
• Interface cable:
- USB to DB9 active-circuit adapter cable, (Digi-Key part number 768-
1013-ND or equivalent), or
- DB9 female to DB9 female, wired pin-to-pin (Digi-Key part AE1016-ND
or equivalent). This is NOT a Null Modem or RS232 cable.
SD card
USB-DB9 adapter
DB9 cable
card reader
Caution
Do not perform any database optimization until you have selected the
X-ray tube installed with your generator. Incorrect tube selection may
cause permanent damage to the X-ray tube and generator.
1
Figure 3-11. DB9 connector at the rear of the Control Console
4. Enter Service Mode (“Entering Service Mode” on page 3-5). You see the
Service Mode menu.
2. Enter Service Mode (“Entering Service Mode” on page 3-5). You see the
Service Mode menu.
The APR Utility user interface screen consists of the following parts:
Generator Parameters. These unchangeable parameters show the Anthem
generator configuration. They must match the connected generator or else the
APR Utility cannot exchange settings with the console.
Region. Click a radio button to view/edit 1 of 8 regions. These selections
correspond to the console APR first-level screen.
View. Click a radio button to view/edit 1 of 12 views. These selections
correspond to the console APR second-level screen.
Index. This displays the individual parameter values for each of 20 indexes
(density values in cm) for the selected Region/View. (Use the scrollbar at the
bottom to view Index 11-20.) This defines each of the 20 values that are
selectable from the console APR Thickness button ( ). To change a value,
__ cm
25cm
Figure 3-14. New regional views can be created in SPARE View fields.
When creating a new view, you must enter exposure technique data for all 20
densities. After entering the kV, mA, and time for all available selections in
the new view, you must optimize the view for your X-ray tube, using the
View selection within the Global Adjustments section on APR Utility screen,
and selecting a mAs multiplier of 1.0.
You must enter the value for patient thickness (in cm) for all thicknesses
allowed within this view. To limit the available patient thickness of any
view, enter a value of 0 for all non-required selections.
Did the column header turn red when you changed a parameter?
Technique changes that fall outside the tube protection profile for the selected
X-ray tube are not allowed, and cause the column header to turn red.
Changing the excessive value of mA, Time, or kV clears this error.
Note: The database downloaded from an Anthem generator contains the specific
signature of that generator’s power limit, tube type, and derating status. Because of
this, you can upload a database to only be a generator equipped with the same tube,
generator power limit, and derating status.
1. Plug the adapter’s USB end or the cable’s DB9 end into the laptop.
2. Plug the other (DB9) end of the adapter or cable into the DB9 connector at
the rear of the Control Console.
3. Enter Service Mode (“Entering Service Mode” on page 3-5). You see the
Service Mode menu.
The Control Console is now loaded with the APR configuration data from the
PC.
Note: The database downloaded from any generator contains the specific signature of
that generator’s power limit, tube type, and derating status. As a result, the database
may only be uploaded to a generator equipped with the same tube, generator power
limit, and derating status.
1. When the APR data has been modified in the Del APR Utility software
chose File -> Save in the Menu and save the new modified data to the SD
memory card. Name the file APRUSER.csv. DO NOT name the file
anything else.
2. Remove the SD card from the PC/laptop. With the SD memory card
contacts facing upward, insert the card in the SD slot at the rear of the
Control Console.
3. Enter Service Mode (“Entering Service Mode” on page 3-5). You see the
Service Mode menu.
Warning
The following procedure produces X-rays. Take precautions to ensure
your safety and the safety of others in close proximity. Minimum
precautions are as follows:
a. Wear lead aprons.
b. Persons in the room during exposure must be behind a lead shield.
c. Minimize radiation scatter through doorways, walls and floor.
Warning
Ensure that main power circuit breaker for the generator is off and
console power switch is OFF before performing the following steps.
Otherwise, you may get severely shocked.
2. Connect the wallstand ion chamber to J21 at the front of the Anthem.
JP300
Figure 3-17. JP300 jumper connections
Warning
Use extreme caution when working around or adjusting the generator
when it is turned on and covers are off. Do not touch any circuits or
components. Circuits can carry lethal voltages which can severely
burn or electrocute you.
4. Enter Service Mode (“Entering Service Mode” on page 3-5). You see the
Service Mode menu.
12. Press the FILM SCREEN button (5) until the desired film screen appears.
Note: The Anthem stores separate AEC calibration data for each film screen speed at
the desired OD. This lets you calibrate multiple film speeds.
3 4
2
5
1 6
Note: On the Main Control board, P1 and P2 provide gain adjustment for the
table and wallstand chambers, respectively. However, do not adjust these unless
directed to do so.
19. If needed, repeat step 16 and step 17. Do not re-adjust gain for the
remainder of the calibration.
20. Press the NEXT STEP button. You see the STEP 3 screen, as shown in Figure
3-20 on page 3-24 with the ion chamber’s right field highlighted (1).
21. Insert an empty (no film) cassette into the table.
Note: Be sure to collimate the beam to cover the upper two cells or all three cells.
24. Press the NEXT STEP button (3) to proceed to balance the left cell.
25. You see the STEP 4 screen, as shown in Figure 3-21, with the ion chamber’s
left field (1) highlighted. Repeat step 22 and step 23 to balance the left
field.
26. Press the NEXT STEP. You see the STEP 5 screen as shown in Figure 3-22..
Place 6” of acrylic or water in a bucket over the center cell. Collimate the
beam over the entire cell.
27. Place film in the cassette.
28. Take an exposure and develop the film.
29. Check the optical density of the film with a densitometer.
30. If the optical density is within 5% of DESIRED OD (1 in Figure 3-22), press
the NEXT STEP button (2) and skip to step 32 of this procedure.
31. If optical density is not within 5% of desired OD, adjust the AEC DAC value
(3) as follows:
a. Press the AEC DAC button (4) and use the data knob (5) to increase or
decrease the value. A clockwise turn increases the DAC value, thus
increasing OD and elapsed mAs.
b. When optical density is within 5% of the desired OD, press the SAVE
button (6).
c. Press the NEXT STEP button (2).
Note: You can calculate the new AEC DAC with the following formula:
4 2
6
4 2
6
a. Press the AEC DAC button (4) and use the data knob (5) to increase or
decrease the DAC value. A clockwise turn increases the DAC value,
thus increasing OD and elapsed mAs.
b. When optical density is within 5% of the desired OD, press the SAVE
button (6).
c. Press the NEXT STEP button (2).
Note: You can calculate the new AEC DAC with the following formula:
4 2
6
Acceptance Testing
4
4.1 Acceptance Testing for Regulatory Compliance
The following sections identify required safety- and compliance-related
installation actions, which include the following activities:
• installation inspection per documented criteria
• installation approval
• creation and delivery of an acceptance report document
or, QARA@delmedical.com
Purchaser Information
Address
Contact Name
Contact Phone
Contact Email
Date of sale/award/lease
System Information
System Seller
System Installer
System ID
System Brand
System Components
Generator
Table
Wallstand
Collimator
X-Ray Tube
Workstation
Detector #1
Detector #2
Other
Other
Generator Tests
Subtract
Measured Mean Squared
Set 1 Verification Task (Kn) (Dn = Kn - K) (Dn2)
Dose, exposure #3
Dose, exposure #9
Subtract
Measured Mean Squared
Set 2 Verification Task (Kn) (Dn = Kn - K) (Dn2)
Dose, exposure #3
Dose, exposure #9
Minimum and maximum Zero body mass; max mAs = 5 Pass Fail
mAs limits for AEC (Auto-
matic Exposure Control) Maximum body mass; max mAs =
(if equipped). Set kVp to 600 Pass Fail
51.
Detector Evaluation
Acceptance
Return of Form
Mail the entire filled-out Acceptance Testing Check Sheet form to the
following address:
Del Medical, Inc.
Technical Support
241 Covington Dr.
Bloomingdale, IL 60108
b. Square each difference value (Dn2). Record these values in the Squared
column.
8. Calculate the sum of all the squared differences. Record this on the
worksheet.
9. Divide the sum of the squared differences by 9. Then calculate the square
root of that result. This is standard deviation SD.
10. Divide SD by K. This is coefficient of variation CV. This value must be less
than or equal to 0.045 (4.5%). Record this value on the worksheet, and
record whether or not the test passes.
11. Repeat all prior steps for Set 2, using different exposure settings per
facility standards.
For the Exposure Linearity Test, use the dose measurements previously
obtained.
12. Transfer the Reproducibility values from Set 1 and Set 2 to the Exposure
Linearity Tests worksheet.
13. Calculate the difference of the set 1/set 2 reproducibility results, divided
by their sum. This value must be less than or equal to 0.095 (9.5%).
Subtract
Measured Mean Squared
Set 1 Verification Task (Kn) (Dn = Kn - K) (Dn2)
3 7a 7b
Repeated dose measure- Dose, exposure #1 1.321 0.0006 0.00000036
ments for a performance
deviation calculation Dose, exposure #2 1.321 0.0006 0.00000036
Operation
5
5.1 Introduction
The Anthem generator operator’s Control Console features an easy-to-use
menu-driven, multifunction button user interface.
Caution
Warning
Do not allow foreign objects that can attenuate or scatter the X-ray
beam between the X-ray tube and table top during exposure. Failure to
comply may result in serious injury.
5 Operation 5-1
Anthem Installation, Operation, & Service Manual
Expose Button
Display
Power Switch
Figure 5-1. Control Console
Multifunction Buttons
Use the multifunction buttons to select parameters. There are two methods
available to change technique parameters:
• To change the values for settings such as kV, mAs, time, or mA, press the
button adjacent to that setting. The current value is then highlighted on
the screen. Change this value by turning the data knob clockwise to
increment or counterclockwise to decrement the value.
• For other parameters, such as the selection of focal spots, film receptors, or
modes, press the multifunction button repeatedly to step through the
possible values.
5-2 5 Operation
Anthem Installation, Operation, & Service Manual
Data Knob
Use the data knob in conjunction with the kV, mAs, time, and centimeter
thickness buttons. When you press one of those buttons, the associated
parameter highlights the screen in reverse video. Change the values by
turning the knob.
READY EXPOSE
FP10-0014.00 FP10-0013.00
prep expose
These buttons perform the same functions as the 2-position handswitch and
footswitch. Depressing the handswitch or footswitch to its first position is
equivalent to pressing the prep button. Depressing the handswitch or
footswitch to its second position is equivalent to simultaneously holding to
the prep button and pressing the expose button.
5 Operation 5-3
Anthem Installation, Operation, & Service Manual
Symbol/Text Description
kV Tube voltage value. This parameter typically controls con-
trast. Use the data knob to change the kV value in single
digit increments.
mAs (2-Point only) mAs value. mAs is an optimized composite
of tube current and exposure time values. Use the data
knob to change the mAs value in pre-determined steps.
mA (3-Point only) Tube current value. This parameter typically
controls image intensity, or darkness. Use the data knob
to change the mA value from 25 mA to 650 mA.
ms (3-Point only) Exposure time. Use the data knob to
change the ms value. Although you can select time val-
ues, the actual exposure duration is controlled by the mAs
timer. You can select from the following values:.010, .012,
.015, .020, .025, .030, .035, .040, .050, .060, .070, .080,
.09, 0.1, 0.12, 0.15, 0.25, 0.3, 0.35, 0.4, 0.5, 0.6, 0.75,
0.9, 1.0, 1.25, 1.5, 1.75, 2.0, 2.5, 3.0, 4.0, 5.0, and 6.0.
Toggle between 2 point and 3 point technique modes.
In 2-point mode, you select only the kV and mAs of an
2 POINT exposure. The generator calculates the mA value to pro-
duce the shortest possible exposure duration for the
3 POINT selected technique, while protecting the X-ray tube from
overloads.
In the 3-point mode, you select the kV, mA, and ms of the
exposure. Selections are limited by the X-ray tube over-
load circuit. If you see an Exceeds Tube Limits mes-
2 POINT sage, readjust either the kV, mA, or ms of the technique
3 POINT before making an exposure.
5-4 5 Operation
Anthem Installation, Operation, & Service Manual
Symbol/Text Description
Image Receptor Selection
• Table - The receptor is in the table and is not DR.
• Table Top (available with DR only) - The receptor is
on the table top, or is DR.
• Wall - The receptor is in a wallstand.
• NO BUCKY - The receptor is not in a table or
NO BUCKY wallstand.
Focal Spot Size
The focal spot size is pre-programmed to optimize the
image quality. The left symbol indicates large focal spot
and the right symbol indicates small focal spot.
Thickness
(APR only) For patient size, use either this button or the
__ cm
25cm Patient Size button. When you the select the thickness in
FP10-0022.0 centimeters (with the data knob), each cm value sets a
pre-programmed kV/mA/time combination. There are 20
combinations to select from, and the set of combinations
can vary with each Region/View. You program each of
these combinations using the APR Utility program (“Edit-
ing the APR Database” on page 3-14).
Patient Size
(APR only) For patient size, use either this button or the
Thickness button. This button toggles between small,
medium, and large patient size. Each size is pre-pro-
FP 0 002 00
grammed to select one of the Thickness values above,
and the selections can vary with each Region/View.
Symbol/Text Description
APR Operating Mode
When in Manual Mode, pressing this button switches to
APR mode. If already in APR mode, this button takes you
back to a previous APR-mode screen.
Manual Operating Mode
Displayed only in APR Mode, this button switches to
Manual mode.
Go back
This symbol appears when an error message occurs.
This button takes you back to the previous screen.
AEC
Toggle on or off the AEC (Automatic Exposure Control)
feature.
5 Operation 5-5
Anthem Installation, Operation, & Service Manual
When self test finishes, the display defaults to the mode of operation selected
at the time of installation. The system is now ready for use.
5-6 5 Operation
Anthem Installation, Operation, & Service Manual
2 POINT
80kV 12mAs 3 POINT
NO BUCKY
NO BUCKY
5 Operation 5-7
Anthem Installation, Operation, & Service Manual
UPPER CERVICAL
EXTREMITY SPINE
AP MODE
LOWER LUMBAR
EXTREMITY SPINE
SELECT
ANATOMICAL REGION
PELVIS CHEST &
& HIP THORAX
5-8 5 Operation
Anthem Installation, Operation, & Service Manual
After selecting a region, you see the available views for that region. Select a
view.
T-SPINE STERNUM
LAT LAT
After selecting a view, you see the technique parameters for that region. Edit
these parameter values as you would in 3-point manual mode.
SAVE
125 kV 350 mA 70 ms TECH
24 mAs
5 Operation 5-9
Anthem Installation, Operation, & Service Manual
To save modified APR techniques, use the SAVE TECH button in the upper
right corner. (Note: You can save APR techniques only if the feature “ALLOW APR
UPDATES” was enabled during installation.)
Anatomi-
cal
region View of region
Upper Finger Finger Hand/ Hand/ Forearm Elbow Elbow Humerus Shoulder Trans
Extremity AP/OBL LAT Wrist AP/ Wrist LAT AP/LAT AP/OBL LAT AP/LAT AP/LAT Thoractic
OBL
Lower Toe AP/ Foot AP/ Foot LAT Ankle Ankle TIB/FIB TIB/FIB Knee LAT Knee AP/ Femur
Extremity LAT LAT AP/OBL LAT AP LAT OBL AP
Femur
LAT
Pelvis & Hip Pelvis AP Hip AP Sacrum Sacrum
Coccyx/ Coccyx/
AP LT
Abdomen Abdo- Abdo- KUB AP/
KUB men AP men OBL
Upright
Cervical AP Odontoid Lateral/ Swimmer
Spine Oblique
Lumbar Spine AP Oblique Lateral L5-S1
Spot
Chest & Tho- Chest AP Chest T-Spine T-Spine Rib Rib Sternum Sternum
rax LAT AP LAT Above Below RAO LAT
Dia- Dia-
phragm phragm
Skull & Facial AP/PA or Skull Skull Skull LAT Facial Waters Facial
Caldwell Towne SMU AP/PA LAT
Base
5-10 5 Operation
Anthem Installation, Operation, & Service Manual
To activate AEC, press the AEC ( ) button. You see new parameters in the
right-side column.
2 POINT
80kV 12mAs 3 POINT
375ms
BKUP
NO BUCKY 1.2
600
For exposure procedures, see “Making an Exposure” on page 5-13. When the
exposure is complete, the display shows the actual mAs and elapsed time of
the exposure.
5 Operation 5-11
Anthem Installation, Operation, & Service Manual
Symbol/Text Description
Back Up
375ms This button lets you set the time backup value with the data
BKUP knob. Set this to approximately 2 to 3 times the expected mAs
for the selected kV.
Film Density
This button lets you set the optical density on the finished
1.2
radiograph. You can select changes in density, ranging from
0.8 to 2.0, with the data knob.
When you select a film speed value (below), this field shows
the value that was used during calibration, and is your “zero”
point.
Ion Chamber Fields
This button lets you select chamber field combinations for a 3-
field ionization chamber by toggling this button.
Film/Screen Selection
This button gives you a choice of different film-screen combi-
600
nation speeds, or CR (Computed Radiography). The film/
screen speeds available may vary by site and appear only if
calibrated during installation.
When you select a value, the Film Density display shows the
value that was used during calibration, and is your “zero”
point.
5-12 5 Operation
Anthem Installation, Operation, & Service Manual
2PT KV
DEFAULT
40
2PT MAS IMAGE
DEFAULT DEFAULT
1.0 NONE
SET AN OPTION
DEFAULT
CONTRAST
70
FLASH SET
OPTIONS CLOCK
EXIT
2. You can now program default settings for the parameters shown in
Figure 5-9 and described in “Operator Control Console” on page 5-2.
To exit this screen and save your changes, press EXIT. The generator completes
its power-up sequence and reboots to the new default settings.
Note: There is no default setting for the FLASH OPTIONS, which also appears in the
Service Mode menu. You use this option to transfer APR user settings to or from a
PC. If APR is not installed, you do not see FLASH OPTIONS.
5 Operation 5-13
Anthem Installation, Operation, & Service Manual
Note: If you release either of the two buttons during an exposure, the exposure
terminates and you see a BUTTON FAULT error message.
Fault Indication
If a fault occurs in your system, you see a fault condition message on the
screen. See Chapter 7 “Troubleshooting”.
5-14 5 Operation
Anthem Installation, Operation, & Service Manual
4. Handswitch: Press the upper button until it is flush with the lower button.*
Footswitch: Step down until you hear a click and/or feel some resistance.*
Console: Press and hold down the prep button.
5. When the green READY light turns on:
Handswitch: Pres the button (both levels) completely in.*
Footswitch: Step down completely.*
Console: Continue pressing the prep button and press and hold the expose
button.
When you start the exposure, the yellow EXPOSE light goes on and you hear a
tone for the duration of the exposure.
6. Continue to hold down the switch or both buttons until the EXPOSE light
goes off and the tone stops.
* When using a handswitch or footswitch with the Anthem, you can press all the way
into the second position and wait for the exposure to complete. A generator automatic
timer waits the proper period before entering EXPOSE mode.
Note: If you release either of the two buttons during an exposure, the exposure
terminates and you see a BUTTON FAULT error message.
Fault Indication
If a fault occurs in your system, you see a fault condition message on the
screen. See Chapter 7 “Troubleshooting”.
1. Select the APR Mode by pressing the button. If the APR symbol is not
showing, APR mode is already selected.
2. Change or verify thickness by pressing the respective button and use the
data knob to adjust the thickness.
3. Change or verify patient size by repeatedly pressing the respective button
to toggle between small, medium or large patient.
4. Handswitch: Press the upper button until it is flush with the lower button.*
Footswitch: Step down until you hear a click and/or feel some resistance.*
Console: Press and hold down the prep button.
5. When the green READY light turns on:
Handswitch: Pres the button (both levels) completely in.*
Footswitch: Step down completely.*
5 Operation 5-15
Anthem Installation, Operation, & Service Manual
Console: Continue pressing the prep button and press and hold the expose
button.
When you start the exposure, the yellow EXPOSE light goes on and you hear a
tone for the duration of the exposure.
6. Continue to hold down the switch or both buttons until the EXPOSE light
goes off and the tone stops.
* When using a handswitch or footswitch with the Anthem, you can press all the way
into the second position and wait for the exposure to complete. A generator automatic
timer waits the proper period before entering EXPOSE mode.
Note: If you release either of the two buttons during an exposure, the exposure
terminates and you see a BUTTON FAULT error message.
Fault Indication
If a fault occurs in your system, you see a fault condition message on the
screen. See Chapter 7 “Troubleshooting”.
Note: If you release either of the two buttons during an exposure, the exposure
terminates and you see a BUTTON FAULT error message.
Fault Indication
If a fault occurs in your system, you see a fault condition message on the
screen. See Chapter 7 “Troubleshooting”.
5-16 5 Operation
Anthem Installation, Operation, & Service Manual
5 Operation 5-17
5-18
TIME mA Selected
ms 25 50 75 100 150 200 250 300 350 400 450 500 550 600 650 700
10 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00
11 1.10 1.65 2.20 2.75 3.30 3.85 4.40 4.95 5.50 6.05 6.60 7.15 7.70
12 1.20 1.80 2.40 3.00 3.60 4.20 4.80 5.40 6.00 6.60 7.20 7.80 8.40
5 Operation
13 1.30 1.95 2.60 3.25 3.90 4.55 5.20 5.85 6.50 7.15 7.80 8.45 9.10
15 1.13 1.50 2.25 3.00 3.75 4.50 5.25 6.00 6.75 7.50 8.25 9.00 9.75 10.50
17 1.28 1.70 2.55 3.40 4.25 5.10 5.95 6.80 7.65 8.50 9.35 10.20 11.05 11.90
20 1.00 1.50 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00 14.00
22 1.10 1.65 2.20 3.30 4.40 5.50 6.60 7.70 8.80 9.90 11.00 12.10 13.20 14.30 15.40
25 1.25 1.88 2.50 3.75 5.00 6.25 7.50 8.75 10.00 11.25 12.50 13.75 15.00 16.25 17.50
27 1.35 2.03 2.70 4.05 5.40 6.75 8.10 9.45 10.80 12.15 13.50 14.85 16.20 17.55 18.90
30 1.50 2.25 3.00 4.50 6.00 7.50 9.00 10.50 12.00 13.50 15.00 16.50 18.00 19.50 21.00
32 1.60 2.40 3.20 4.80 6.40 8.00 9.60 11.20 12.80 14.40 16.00 17.60 19.20 20.80 22.40
35 1.75 2.63 3.50 5.25 7.00 8.75 10.50 12.25 14.00 15.75 17.50 19.25 21.00 22.75 24.50
37 1.85 2.78 3.70 5.55 7.40 9.25 11.10 12.95 14.80 16.65 18.50 20.35 22.20 24.05 25.90
Anthem Installation, Operation, & Service Manual
40 1.00 2.00 3.00 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00 24.00 26.00 28.00
45 1.13 2.25 3.38 4.50 6.75 9.00 11.25 13.50 15.75 18.00 20.25 22.50 24.75 27.00 29.25 31.50
50 1.25 2.50 3.75 5.00 7.50 10.00 12.50 15.00 17.50 20.00 22.50 25.00 27.50 30.00 32.50 35.00
55 1.38 2.75 4.13 5.50 8.25 11.00 13.75 16.50 19.25 22.00 24.75 27.50 30.25 33.00 35.75 38.50
60 1.50 3.00 4.50 6.00 9.00 12.00 15.00 18.00 21.00 24.00 27.00 30.00 33.00 36.00 39.00 42.00
65 1.63 3.25 4.88 6.50 9.75 13.00 16.25 19.50 22.75 26.00 29.25 32.50 35.75 39.00 42.25 45.50
70 1.75 3.50 5.25 7.00 10.50 14.00 17.50 21.00 24.50 28.00 31.50 35.00 38.50 42.00 45.50 49.00
75 1.88 3.75 5.63 7.50 11.25 15.00 18.75 22.50 26.25 30.00 33.75 37.50 41.25 45.00 48.75 52.50
80 2.00 4.00 6.00 8.00 12.00 16.00 20.00 24.00 28.00 32.00 36.00 40.00 44.00 48.00 52.00 56.00
85 2.13 4.25 6.38 8.50 12.75 17.00 21.25 25.50 29.75 34.00 38.25 42.50 46.75 51.00 55.25 59.50
90 2.25 4.50 6.75 9.00 13.50 18.00 22.50 27.00 31.50 36.00 40.50 45.00 49.50 54.00 58.50 63.00
Table 5-2: mAs vs. mA and time selected (10 to 135 ms)
95 2.38 4.75 7.13 9.50 14.25 19.00 23.75 28.50 33.25 38.00 42.75 47.50 52.25 57.00 61.75 66.50
100 2.50 5.00 7.50 10.00 15.00 20.00 25.00 30.00 35.00 40.00 45.00 50.00 55.00 60.00 65.00 70.00
110 2.75 5.50 8.25 11.00 16.50 22.00 27.50 33.00 38.50 44.00 49.50 55.00 60.50 66.00 71.50 77.00
120 3.00 6.00 9.00 12.00 18.00 24.00 30.00 36.00 42.00 48.00 54.00 60.00 66.00 72.00 78.00 84.00
125 3.13 6.25 9.38 12.50 18.75 25.00 31.25 37.50 43.75 50.00 56.25 62.50 68.75 75.00 81.25 87.50
130 3.25 6.50 9.75 13.00 19.50 26.00 32.50 39.00 45.50 52.00 58.50 65.00 71.50 78.00 84.50 91.00
135 3.38 6.75 10.13 13.50 20.25 27.00 33.75 40.50 47.25 54.00 60.75 67.50 74.25 81.00 87.75 94.50
Table 5-3: mAs vs. mA and time selected (140 to 1660 ms)
1000 25.0 50.0 75.0 100.0 150.0 200.0 250.0 300.0 350.0 400.0 450.0 500.0 550.0 600.0
1120 28.0 56.0 84.0 112.0 168.0 224.0 280.0 336.0 392.0 448.0 504.0 560.0
1250 31.3 62.5 93.8 125.0 187.5 250.0 312.5 375.0 437.5 500.0 562.5
1370 34.3 68.5 102.8 137.0 205.5 274.0 342.5 411.0 479.5 548.0
1500 37.5 75.0 112.5 150.0 225.0 300.0 375.0 450.0 525.0 600.0
1660 41.5 83.0 124.5 166.0 249.0 332.0 415.0 498.0 581.0
5 Operation
5-19
Anthem Installation, Operation, & Service Manual
5-20
5 Operation
TIME mA Selected
ms 25 50 75 100 150 200 250 300 350 400 450 500 550 600 650 700
1750 43.8 87.5 131.3 175.0 262.5 350.0 437.5 525.0
1875 46.9 93.8 140.6 187.5 281.3 375.0 468.8 562.5
2000 50.0 100.0 150.0 200.0 300.0 400.0 500.0 600.0
2250 56.3 112.5 168.8 225.0 337.5 450.0 562.5
2500 62.5 125.0 187.5 250.0 375.0 500.0
Anthem Installation, Operation, & Service Manual
Table 5-4: mAs vs. mA and time selected (1750 to 6000 ms)
6 Periodic Maintenance
Periodic Maintenance
6
6.1 Periodic Maintenance Schedule
Maintenance procedures for the Anthem X-ray generator, performed by
qualified service personnel, are required within 30 days after installation and
every 6 months thereafter. Service personnel must be specifically trained and
experienced with medical X-ray apparatus. Applicable preventive
maintenance or any repair service should be performed only by these skilled
individuals. These procedures are contained in the Anthem Generator
Installation, Operation and Service Manual (this manual).
Refer to “Periodic Maintenance Schedule” on page 6-1 for information on
when to perform periodic maintenance on the generator.
Warning
The main capacitor bank (located on the left side of the Power
Module) contains a very high charge when power is applied. This
charge is a fatal shock hazard. Never perform any service or
maintenance inside the generator without first performing the
“Discharge Procedure” on page 6-2. Check the capacitor bank for zero
volts with a DC voltmeter before working on any internal circuitry.
Warning
If maintenance procedures require power to be applied to the unit,
exercise extreme care to insure the safety of service and others in the
area. Always verify that the equipment is properly grounded before
attempting any electrical operation or adjustment.
Warning
Certain tests require the production of X-rays. Take precautions to
ensure your safety and the safety of others in close proximity. Observe
the following minimum precautions:
- wear lead aprons
- personnel remaining in the X-ray room during exposure must
be behind a lead shield
- minimize radiation scatter through doorways, walls and floor
Note: Due to varying operating conditions, the following procedures might need to
be performed at greater or lesser intervals. Set these intervals according to your
generator’s performance.
5. Take five exposures. (Note: This step sends voltage to the tube, but does not
recharge the capacitors.)
6. Open the upper cabinet and measure the voltage across the terminals (1 in
Figure 6-3) at capacitor C1 (Figure 6-3). If voltage is 10 VDC or higher, take
two more exposures and then verify voltage across the bus again.
Tools Required
• cleaning wipes
• non-abrasive, hospital-grade cleaner
Use cleaning wipes and non-abrasive, hospital-grade cleaner to clean external
surfaces of the generator.
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
This equipment is NOT classified as anaesthetic-proof and may ignite
inflammable anesthetics. Flammable agents used for skin cleaning or
disinfecting may also produce an explosion hazard.
• Ensure the power has been disconnected and that the emergency switches
have been activated, before starting any cleaning operation.
• Ensure no liquid gets into the unit.
• Do not immerse the equipment, including any components or accessories,
in liquid.
• Do not autoclave the equipment, including any component or accessories.
• Do not use water. Water can short-circuit the electrical installation and
cause corrosion to mechanical parts.
• Do not use acid or abrasive products.
• Use only a dry cloth to clean chrome-plated parts.
• Only the surface areas of unit parts, including accessories and connection
cables, should be disinfected using a gaseous disinfectant.
• Clean painted parts with a cloth and products appropriate for cleaning
plastic materials; after cleaning wipe the surfaces with a clean, dry cloth.
• Do not spray cleaning or disinfection solution directly on the equipment.
To disinfect, moisten a cloth with a 70% Isopropyl alcohol solution or
equivalent and wipe the surface of the equipment.
• When disinfecting the examination room, ensure the unit is covered with
plastic sheets.
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
Never perform any service or maintenance inside the generator
without first performing the “Discharge Procedure” on page 6-2.
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
Never perform any service or maintenance inside the generator
without first performing the “Discharge Procedure” on page 6-2.
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
Never perform any service or maintenance inside the generator
without first performing the “Discharge Procedure” on page 6-2.
1. Verify that insert of tube matches tube insert selected in the system
installation file. If it does not, call Del Medical Technical Support.
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
Never perform any service or maintenance inside the generator
without first performing the “Discharge Procedure” on page 6-2.
Caution
Always use an anti-static wrist strap when working on electrostatic
sensitive devices.
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
Never perform any service or maintenance inside the generator
without first performing the “Discharge Procedure” on page 6-2.
Oscilloscope settings
3. Set the oscilloscope as follows:
Channel #1
Vertical 2 V/div.
Horizontal 10 ms/div.
Mode Normal Sweep
Trigger Channel 1 - Positive
Channel #2
Vertical 1 V/div.
Horizontal 10 ms/div.
Mode Normal Sweep
Note: If you are performing calibration using only a scope, read the values of kV
and mA as follows:
Warning
When taking exposures for test and/or verification purposes, follow
the site’s specific safety routines and regulations.
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
Never perform any service or maintenance inside the generator
without first performing the “Discharge Procedure” on page 6-2.
Caution
Always use an anti-static wrist strap when working on electrostatic
sensitive devices.
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
Never perform any maintenance, inside the generator without first
performing the “Discharge Procedure” on page 6-2.
1. Check the transformer oil level. Oil is at proper level when it reaches the
horizontal level pin (Figure 6-4) in the circular window. To replenish
transformer oil, unscrew the four screws holding the circular window, and
fill with Shell Diala Oil AX only. The oil level must not be above the top of
the horizontal level pin.
Oil Level
Window
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
When taking exposures for test and/or verification purposes make
sure to follow the specific safety routines and regulations on site.
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
Never perform any maintenance inside the generator without first
performing the “Discharge Procedure” on page 6-2.
Troubleshooting
7
7.1 Introduction
This chapter contains troubleshooting tables and charts that guide you
through most of the problems that may be related to the generator. Included
are also status/error messages along with possible solution(s) to the error.
Because of the complex nature of some error conditions, part replacements
recommended in this chapter are not guaranteed to be the singular solution to
a problem.
Warning
All service and maintenance including the procedures in this chapter,
must be performed by qualified service personnel only.
Warning
Never perform any service or maintenance inside the generator
without first performing the “Discharge Procedure” on page 6-2.
7 Troubleshooting 7-1
Anthem Installation, Operation, & Service Manual
Error Message
Location
7-2 7 Troubleshooting
Anthem Installation, Operation, & Service Manual
5. (If the message occurs for both table and wall chambers) Measure the AEC
reference voltage on the Main Control board:
a. Select Wall AEC
b. Connect a DVM to TP302, and reference TP22C.
c. Increase the kVp selection, and verify that the reference voltage
decreases. You typically see approximately 2 VDC at 40 kV and
approximately 0.8 VDC at 125 kV. If this voltage is considerably out of
range, not present, or does not change with changes in kVp, replace the
Main Control board (“Replacing the Main Control Board” on
page 8-15).
d. Select Table AEC.
e. Connect a DVM to TP303, and reference TP22C.
a. Increase the kVp selection, and verify that the reference voltage
decreases. You typically see approximately 2 VDC at 40 kV and
approximately 0.8 VDC at 125 kV. If this voltage is considerably out of
range, not present, or does not change with changes in kVp, replace the
Main Control board (“Replacing the Main Control Board” on
page 8-15).
6. To measure wallstand ion chamber output ramp performance, connect an
oscilloscope and set parameters as follows:
Note: If possible remove the grid, or be aware that the grid will have an effect on
the angle of the ramp and the elapsed mAs. Select a technique that will result in at
least 10 – 20 mAs
a. Set up channel 1 (ion chamber output ramp): 1V/div, 10ms/div, probe
on Main Control board JP300, pin 1 (see Figure 3-17 on page 3-20),
referenced to TP22C.
b. Set up channel 2 (attenuated ramp): 1V/div, 10ms/div, probe on Main
Control board TP300 (see Figure 3-17 on page 3-20), referenced to
TP22C
c. Make an AEC exposure.
Ensure both ramps are linear and free of any noise. Verify that the
attenuated ramp is not be more than 50% of the ramp signal coming
from the ion chamber.
7. To measure table ion chamber output ramp performance, connect an
oscilloscope and set parameters as follows:
Note: If possible remove the grid, or be aware that the grid will have an effect on
the angle of the ramp and the elapsed mAs. Select a technique that will result in at
least 10 – 20 mAs
a. Set up channel 1 (ion chamber output ramp): 1V/div, 10ms/div, probe
on Main Control board JP300, pin 4 (see Figure 3-17 on page 3-20),
referenced to TP22C.
7 Troubleshooting 7-3
Anthem Installation, Operation, & Service Manual
Ensure both ramps are linear and free of any noise. Verify that the
attenuated ramp is not be more than 50% of the ramp signal coming
from the ion chamber.
8. If the ramp signals are present, clean and linear, perform the procedure
“AEC Calibration Procedure” on page 3-18.
9. If any attenuated ramp signals are more than 50% of the ramp signal
coming from the ion chamber, perform the procedure “AEC Calibration
Procedure” on page 3-18.
If you suspect gain pots P1 and/or P2 on the Main Control board must be
reset, perform the following procedures for centering these pots.
Centering the P2 Gain Potentiometer
1. With power OFF, move the jumper on JP300 to pins 1 and 2 (to the right
side; see Figure 7-2).
JP300
4 3 2 1
7-4 7 Troubleshooting
Anthem Installation, Operation, & Service Manual
4. Adjust P1 until the voltage between at TP301 equals the voltage recorded
at TP304.
5. P1 pot is now centered.
6. Set the jumper on JP300 back to pins 2 and 3 (center position).
Button Fault
Problem Description
This message can occur with early releases of any of the "in use" exposure
switches, e.g., console switch, hand switch or foot switch. When Button Fault
appears, an elapsed mAs value is also displayed. This is the total accumulated
output mAs read by the generator before the Button Fault occurred.
Note: A Button Fault can also occur if the Bucky interlock signal is not seen by the
generator within 7 seconds after the exposure switch is pressed. See also: “Table
Bucky Not Rdy” on page 7-38 and “Wall Bucky Not Rdy” on page 7-38.
Corrective Actions
1. Inspect the handswitch or footswitch connection at the rear of the console,
J11. See Figure 2-30 on page 2-29, and Figure 7-3 / Figure 7-4.
2. Inspect the internal console connection to the panel PREP and EXPOSE
switches, J10. See Figure 7-4, or the schematic on page 9-12
Note: See also “Table Bucky Not Rdy” on page 7-38 or “Wall Bucky Not Rdy” on
page 7-38.
PREP/EXPOSE Handswitch/Footswitch
Button Connector J10 Connector J11
7 Troubleshooting 7-5
Anthem Installation, Operation, & Service Manual
Handswitch
Footswitch
Collimator Error
Problem Description
The collimator interlock is not satisfied.
Corrective Action
Inspect the collimator connections at the collimator and at the Room Interface
board. See “Collimator Interlock and Door Interlock Connections” on
page 2-24.
If a manual collimator is installed, ensure a jumper is placed in the proper
position.
If an automatic collimator is installed, ensure the wiring, setup, and
calibration are complete and correct
Comm. Error
Problem Description
Communication between the Power Module and the console has been lost.
Possible causes include:
• Console cable disconnected or failed
• Main Control board failure
• Power Relay board failure
7-6 7 Troubleshooting
Anthem Installation, Operation, & Service Manual
7 Troubleshooting 7-7
Anthem Installation, Operation, & Service Manual
If LED 4 becomes brighter, replace the IPM Driver board (see “Replacing
the IPM Driver Board” on page 8-17).
If LED 4 remains dimly lit, go to Step 5.
5. (LED 4 fully lit) Measure the voltage between TP23 and TP15. It should be
5 VDC.
a. If the TP23-to-TP15 voltage is not 5 VDC, measure the voltage between
TP21 and TP22. It should be 12 VDC.
b. If the TP21-to-TP22 voltage is not 12 VDC, replace the Main Control
board (see “Replacing the Main Control Board” on page 8-15).
If the voltage is 12 VDC, go to Step 6.
6. If all of the previous steps has been completed and no failures were found
on any of the 3 boards listed, swap out the Console Control board.
Door Open
Problem Description
The door interlock is not satisfied.
Corrective Action
If a door switch is installed in the X-Ray room, ensure the door is closed prior
to making an exposure.
7-8 7 Troubleshooting
Anthem Installation, Operation, & Service Manual
If the door is closed, inspect the door connections. See “Collimator Interlock
and Door Interlock Connections” on page 2-24.
7 Troubleshooting 7-9
Anthem Installation, Operation, & Service Manual
Exceeds Limit
Description
The currently selected technique parameter, such kV, mA or time/mAs, is
adjusted to the minimum or maximum of its range, but combined settings are
still within the generator or tube limits. You see this informational message if
you attempt to adjust the parameter beyond its range.
Fil Hi Fault
Corrective Action
Contact your service representative.
7-10 7 Troubleshooting
Anthem Installation, Operation, & Service Manual
Invalid Keys
Problem Description
An invalid option key/code was entered at the control console.
Corrective Action
1. Verify the 16 digit option key code provided by Del Medical and retry the
key entry.
2. Contact Del Medical technical support or email support@delmedical.com.
Be ready to provide the following information:
7 Troubleshooting 7-11
Anthem Installation, Operation, & Service Manual
7-12 7 Troubleshooting
Anthem Installation, Operation, & Service Manual
IPM Fault
Problem Description
One or both of the IPMs (Intelligent Power Modules) have detected a failure
or threatening condition. A voltage or current parameter has been exceeded or
a condition is detected that is potentially harmful to the IPM. Fault conditions
are not specifically specified as defined errors; all fault conditions are reported
as an IPM Fault.
Possible Causes
Over Temperature. The IPMs contain a thermal detection circuit that shuts
itself down if the internal temperature exceeds 275°F.
If this message appears during or after a heavy patient load, or at the end of
the day, over temperature is the probable cause. Turn the generator off and
allow it to cool. Automatic reset occurs around 70°F.
Investigate user technique selection for high kW settings or excessive
exposure times.
Shorted Junction on the IPM. If a short occurs between the any of the supply
terminals to the output terminals of the IPM, short-circuit protection activates.
This also typically results in a “no prep” or “no ready” condition.
Corrective Action
Warning
Ensure the DC bus is fully discharged before attempting to disconnect
or perform any maintenance on the inverter section. Bodily harm and/
or damage to inverter components may occur.
1. On each IPM, disconnect the wires on output terminals E2 and C2E1 (the 2
lower terminals). See Figure 7-6 on page 7-14.
2. On the left-side IPM, measure the resistance between upper terminal C1
and bottom output terminal C2E1. Confirm that this resistance is greater
than 1 MΩ. If it is less, replace the IPM (“Replacing an IPM Block” on
page 8-19).
7 Troubleshooting 7-13
Anthem Installation, Operation, & Service Manual
C1
E2
C2E1
7-14 7 Troubleshooting
Anthem Installation, Operation, & Service Manual
2. If +12 VDC is present on TP6, but any of the +15 VDC readings are low or
missing, replace the IPM Driver board.
3. On the Power Relay board, measure the voltage across TP1 (reference) and
TP5 (+12 VDC).
4. If the +12 VDC supply is low or missing, disconnect the cable between the
IPM Driver board and the Master Control board, and measure the voltage
again.
5. If the+12 VDC supply recovers, replace the IPM Driver board.
KV DAC Fault
Problem Description
Normally, the DAC (Digital to Analog Converter) receives a digital input
from the Main Control CPU, and outputs a corresponding analog voltage.
This DAC voltage is used to establish precision reference signals for kVp
regulation, mA output, filament levels, and AEC reference.
7 Troubleshooting 7-15
Anthem Installation, Operation, & Service Manual
A DAC fault occurs when the CPU's the A-to-D converter finds the analog
signal produced by the DAC manager to be out of tolerance. This is an
internal failure within the Main Control board.
Corrective Action
Replace the Main Control board (see “Replacing the Main Control Board” on
page 8-15).
KV FB Fault
Problem Description
The kV feedback detected at microprocessor U30 on the Main Control board is
not within the acceptable range of the kV value displayed on the console.
When this fault occurs, observe the console display and note whether this is
the only fault or if the display alternates between this fault and Ma FB Fault
(on page 7-25).
Corrective Action
Select one of the following procedures:
• “If KV FB Fault is the only fault” on page 7-16
• “If You See Alternating KV FB Fault and Ma FB Fault” on page 7-17
If KV FB Fault is the only fault
Note: An oscilloscope is required to troubleshoot this fault.
1. Set the oscilloscope scale to 1V = 20kV, and connect the probe to Main
Control board TP6A, and the ground lead to TP22A.
2. Make an exposure and verify that the kV feedback signal is similar to that
in Figure 7-11. In particular, verify the correct amplitude of the signal.
(Waveforms may vary in time from what is shown in the figure. Because
an error is being produced and the exposure is being terminated, the total
time displayed for each waveform will probably be much shorter than the
example.)
7-16 7 Troubleshooting
Anthem Installation, Operation, & Service Manual
Warning
Use extreme caution when testing the DC rail. Harmful or lethal
voltages may be present.
7 Troubleshooting 7-17
Anthem Installation, Operation, & Service Manual
Warning
Before proceeding, ensure that the main power is off and the DC rail
is FULLY discharged. If it will not discharge using the capacitor
discharge procedure, allow it to dissipate on its own.
3. Disconnect the cable from the bottom terminal (C2E1) for each IPM. See
Figure 7-9.
7-18 7 Troubleshooting
Anthem Installation, Operation, & Service Manual
C1
E2
C2E1
KV Hi Fault
Problem Description
The kV feedback signal detected by the Main Control board differs by ±10% or
more from the kV Ref signal produced by the Main Control board. A leading
cause of this fault is when the generator is out of calibration.
Other possible causes are a failed Main Control board or a failed HT Tank.
Corrective Action
Note: An oscilloscope is required to troubleshoot this fault.
1. Perform the entire kV, mA, and preheat procedures in “Calibration
Procedure” on page 3-1 3. Do not skip this step.
2. Set the oscilloscope scale to 1V = 20kV, and connect the probe to Main
Control board TP6A, and the ground lead to TP22A.
7 Troubleshooting 7-19
Anthem Installation, Operation, & Service Manual
3. Make an exposure and verify that the kV feedback signal is similar to that
in Figure 7-11.
7. If the waveforms measured on the oscilloscope are the correct voltage but
the waveform in Step 3 is not, replace the Main Control board.
If the KV feedback is not at the correct level or the waveform is abnormal
at TP2 / TP3, continue to the next step to check the KV drive pulses being
output from the Main Control board to the IPM Driver board.
7-20 7 Troubleshooting
Anthem Installation, Operation, & Service Manual
Warning
Before connecting oscilloscope test probes to any test points on the
IPM Driver board, the oscilloscope MUST be isolated from Earth
ground. Failure to isolate the scope will result in damage to the
generator and / or the oscilloscope.
8. Connect the oscilloscope probes to the IPM Driver board, TP6A and TP6B,
and connect the ground leads to TP5.
9. Make an exposure and verify that the kV drive pulses are similar to those
in Figure 7-13.
Figure 7-13. Normal kV drive waveform at TP6A / TP6B of the IPM Driver,
60kV exposure
KV Imbalance Fault
Problem Description
The difference in amplitude between the POSITIVE and NEGATIVE kV
feedback signals is greater than 8kV. It is a function of the HT Tank to
distribute the primary input evenly between the (+) anode and (-) cathode
sides of the transformer.
However, impedance differences in the HT cables can also influence the
individual kV feedback signals. When this occurs, the feedback turn-off time
can be longer on one side or the other, thus causing this error message.
7 Troubleshooting 7-21
Anthem Installation, Operation, & Service Manual
Corrective Action
1. Verify the error by observing the positive and negative kV feedback
signals directly from the HT Tank.
Note: An oscilloscope is required to troubleshoot this fault.
Connect the oscilloscope probes to Main Control board TP2 (kV+) and TP3
(kV-), with ground leads connected to TP1. Set the oscilloscope scale for
both channels to 1V = 10kV
2. Enter the calibration utility and make an exposure.
Verify that the anode and cathode waveforms are equal in amplitude, and
have the same turn-off times, as shown in Figure 7-14.
3. If the kV waveforms are not equal or if an entire side is missing, check the
resistance of the feedback circuit on the HV transformer.
a. Turn the generator off and turn off the breaker.
b. Important: Perform the “Discharge Procedure” on page 6-2.
Warning
The DC bus capacitors hold a charge long after the generator and
breaker are turned off. Never perform any work inside the generator
without first performing the “Discharge Procedure” on page 6-2.
7-22 7 Troubleshooting
Anthem Installation, Operation, & Service Manual
e. Place an ohmmeter across J1-1 (+kV A) and J1-2 (kV GND). See
Figure 7-15. Confirm the measured resistance is about 35.5kΩ.
J1-3
J1-2
J1-1
7 Troubleshooting 7-23
Anthem Installation, Operation, & Service Manual
a. In Figure 7-13 notice the longer discharge time on the positive or anode
side of this waveform. To isolate a possible problem with one of the HT
cables, turn off the generator, and swap the anode and cathode cables.
Ensure all 4 HT cable connections are tight and secure.
b. Turn the generator on. Make another exposure with the same
oscilloscope and exposure settings and observe the waveform. If the
long trailing edge is on the opposite side, replace the HT cable set.
If the slow discharge remains on the same side, try another known
good X-ray tube, if possible (though it is uncommon for a tube to cause
this condition).
c. If the error message persists, replace the HT Tank.
KV LO Fault
Problem Description
The kV feedback signal detected by the Main Control board differs by ±10% or
more from the kV Ref signal produced by the Main Control board. A leading
cause of this fault is when the generator is out of calibration.
Corrective Action
For troubleshooting, see the corrective action for “KV Hi Fault” on page 7-19.
MA DAC Fault
Problem Description
Normally, the DAC (Digital to Analog Converter) receives a digital input
from the Main Control CPU, and outputs a corresponding analog voltage.
This DAC voltage is used to establish precision reference signals for kVp
regulation, mA output, filament levels, and AEC reference.
7-24 7 Troubleshooting
Anthem Installation, Operation, & Service Manual
A DAC fault occurs when the CPU's the A-to-D converter finds the analog
signal produced by the DAC manager to be out of tolerance. This is an
internal failure within the Main Control board.
Corrective Action
Replace the Main Control board (see “Replacing the Main Control Board” on
page 8-15).
Ma FB Fault
Problem Description
The mA feedback detected at microprocessor U30 on the Main Control board
is not within the acceptable range of the mA value displayed on the console.
When this fault occurs, observe the console display and note whether this is
the only fault or if the display alternates between this fault and KV FB Fault
(on page 7-16).
Corrective Action
Select one of the following procedures:
• “If Ma FB Fault is the only fault” on page 7-25
• “If You See Alternating KV FB Fault and Ma FB Fault” on page 7-26
If Ma FB Fault is the only fault
Note: An oscilloscope is required to troubleshoot this fault.
1. Set the oscilloscope scale to 1V = 20mA (small focus) or 1V = 100mA (large
focus), and connect the probe to Main Control board TP47A, and the
ground lead to TP22A.
2. Make an exposure and verify that the mA feedback signal is similar to that
in Figure 7-17. In particular, verify the correct amplitude of the signal.
(Waveforms may vary in time from what is shown in the figure. Because
an error is being produced and the exposure is being terminated, the total
time displayed for each waveform will probably be much shorter than the
example.)
7 Troubleshooting 7-25
Anthem Installation, Operation, & Service Manual
R1 C1
Figure 7-18. Feedback circuit resistor R1 and capacitor C1 locations
5. To replace C1:
d. Cut the leads of the old capacitor as close to the body as possible,
leaving the leads attached to the board.
e. Tac solder the new capacitor onto the existing leads.
Note: You cannot remove the Feedback board from the HT Tank.
6. Verify that feedback cable 126-0177G2 is properly connected to J14 of the
Anthem power module and J1 of the Feedback board. Follow the labeling
on the cable to verify that the connections are correct on the HV tank.
7. Use an Ohmmeter to test the continuity of each cable conductor. If you
measure a high resistance on any of the conductors (or shorts between
them), replace the cable.
8. If the cable passes the connection and continuity tests, but the feedback
signal from the HT Tank is still not present, the HT Tank probably needs
replacement.
Note: The Feedback board, mounted on top of the HT Tank, is not individually
replaceable. If a component on the Feedback board has failed or is in question,
contact your technical support representative.
If You See Alternating KV FB Fault and Ma FB Fault
Go to “If You See Alternating KV FB Fault and Ma FB Fault” on page 7-17 and
perform the steps in that procedure.
7-26 7 Troubleshooting
Anthem Installation, Operation, & Service Manual
mA Hi Fault
Problem Description
The mA feedback signal detected by the Main Control board differs by ±10%
or more from the mA Ref signal produced by the Main Control board. A
leading cause of this fault is when the generator is out of calibration.
Another possible cause is a failed capacitor on the kVmA Feedback board.
On the HT Tank Feedback board, capacitor C1 (Figure 7-19) might begin to
fail and leak DC current to ground, causing the mA feedback signal to
deteriorate.
Corrective Action
1. Perform the entire kV, mA, and preheat procedures in “Calibration
Procedure” on page 3-1 3. Do not skip this step.
2. If the error message persists, remove the lower cabinet front cover and
remove the feedback board cover.
3. To test capacitor C1, measure the resistance across resistor R1. R1 is a 100Ω
resistor with a tolerance of 1%. If the resistance is out of this spec
(typically, a failed C1 will cause a reading of around 70Ω), then replace C1
(part number 611-5027P1). See Figure 7-20.
R1 C1
Figure 7-20. Feedback circuit resistor R1 and capacitor C1 locations
7 Troubleshooting 7-27
Anthem Installation, Operation, & Service Manual
4. To replace C1:
f. Cut the leads of the old capacitor as close to the body as possible,
leaving the leads attached to the board.
g. Tac solder the new capacitor onto the existing leads.
Note: You cannot remove the Feedback board from the HT Tank.
5. If the error message persists, replace the Main Control board.
mA LO Fault
Problem Description
The mA feedback signal detected by the Main Control board differs by ±10%
or more from the mA Ref signal produced by the Main Control board. A
leading cause of this fault is when the generator is out of calibration.
Corrective Action
For troubleshooting, see the corrective action for “mA Hi Fault” on page 7-27.
No AEC on Bucky
Problem Description
AEC is not active on the selected receptor.
Corrective Action
See “AEC Calibration Procedure” on page 3-18, which includes steps for
activating AEC.
No Feedback Cable
Problem Description
The feedback cable between the HT Tank and the Power Module has been
disconnected.
Corrective Action
Inspect the connection between the HT Tank and J14 on the Power Module.
No Fil Fault
Problem Description
Primary filament current feedback is not being detected by the CPU on the
Main Control board. Possible causes include:
• an open connection in the primary filament circuit
• low or no filament power supply
7-28 7 Troubleshooting
Anthem Installation, Operation, & Service Manual
TB1 on
Filament Board
2. The other end of cable 126-5294G1 connects to the top of the HT Tank, J2.
Ensure this connection is secure. If the cable itself is suspect, disconnect it
at J2 and test it with an ohmmeter for continuity and shorts.
7 Troubleshooting 7-29
Anthem Installation, Operation, & Service Manual
3. At the Filament board, ensure the cable length jumper (which may be on
J4, J5, or J6) is secure. See Figure 7-22.
Long
Short
Medium
4. Test for an open circuit on the primary side of the filament transformer:
a. At the top of the HT Tank, disconnect the cable from J2 connector from
the top of the high voltage transformer.
b. Measure and confirm no or very low (and the same) resistance between
these pins (see Figure 7-23 on page 7-30):
• XC to XS
• XC to XL
J2-3 XC
J2-1 XL
J2-2 XS
Secondary Primary
HV -
7-30 7 Troubleshooting
Anthem Installation, Operation, & Service Manual
b. Verify 50 VDC on the Power Relay board at TP7 (+50 VDC) TP9 (-50
VDC) referenced to TP8 filament common.
c. Verify 50 VDC on the Filament board at TP3 (+50 VDC) TP2 (-50 VDC)
referenced to TP1 filament common.
d. Verify that Filament board LED 2 and LED 3 are lit, indicating that
voltage is present at TP3 and TP2.
e. If 50 VDC is not present at TP3 and TP2 or is not the same as measured
in Step b, inspect the board-to-board connection between Power Relay
board TB8 and Filament board J8.
6. Verify the filament reference and filament drive signals. Based on the
selected mA station, the Main Control board generates a filament
reference signal to the PWM IC (U13), which determines the pulse width
of the filament drive signal. During exposure, the reference signal changes
up or down slightly, as it's compared to the actual mA feedback to
maintain a stable mA output.
a. Verify the FIL/mA Reference signal at TP54 on the Main Control
board. You should read approximately 2.1 -2.5 VDC in standby, and
depending on the mA station selected, read between 3.8 -4.5 VDC
during Prep. If these voltages are not present, replace the Main Control
board
b. Connect 2 channels of an oscilloscope to the Filament board TP4 (FIL
DR1) and TP5 (FIL DR2). connect the ground lead(s) to TP9.
c. Make an exposure and verify that the traces look like those in
Figure 7-24.
Idle stby Normal Condition Prep Normal Condition Filament Drive during the No Fil Fault
d. If these signals are not present, or if one signal is missing, ensure the
board-to-board ribbon cables are making good connections.
e. Verify if the FIL DR1 and FIL DR2 signals are present on the Main
Control board at U13 pin 8 or U13 pin 11. If these signals are not
present, or if one signal is missing on the Main Control board, replace
the Main Control board.
7. If the above measurements have been verified and the observed
waveforms are correct, replace the Filament board.
7 Troubleshooting 7-31
Anthem Installation, Operation, & Service Manual
7-32 7 Troubleshooting
Anthem Installation, Operation, & Service Manual
SD Card Error
Problem Description
An error has occurred while reading or attempting to read the inserted SD
card.
Note: Earlier Anthem generators manufactured between 2009 and 2013 had a 2GB
limit on the SD card capacity. Anthems manufactured after 2013 can read SD cards
with a capacity of up to 32GB.
Corrective Action
• Use only SD form-fit cards. Do not use micro SD cards with an adapter.
• Do not insert an SD card into the console while the console is turned on.
Insert and eject the SD card only when the console is off.
SD Card WP
Problem Description
The write protection switch, located on the left side of the SD card is in the
locked position. In this condition, the SD card does not allow data saving.
7 Troubleshooting 7-33
Anthem Installation, Operation, & Service Manual
Corrective Action
1. Power off the Anthem.
2. Remove the SD card from the back on the Anthem console. Locate the WP
switch on the left side of the SD card. Push the switch upward towards the
top of the SD card. This position unlocks the SD card.
Warning
The following procedure exposes you to high voltage connections.
Take all necessary precautions and always assume the DC rail
capacitors are holding charge (until discharge is confirmed with a
meter), even if the power is turned off.
1. Remove the Power Module front cover and open the door.
7-34 7 Troubleshooting
Anthem Installation, Operation, & Service Manual
K12 QC10
QC11 QC12
QC9
b. Initiate the Prep state and verify that the voltage goes from 0 VAC to
about 280 VAC.
c. Connect the AC voltmeter to connectors QC10 and QC12.
d. Again, initiate the Prep state and verify that the voltage goes from 0
VAC to about 280 VAC.
e. If the previous steps pass, go to “If K12 is Properly Closing” on
page 7-35.
7 Troubleshooting 7-35
Anthem Installation, Operation, & Service Manual
2. Locate the soft-charge resistor (inside the power module, to the right; see
Figure 7-27 on page 7-36), and measure its resistance. Verify the resistance
is between 47.5Ω and 52.5Ω. If it is not, replace the resistor.
IPM Driver
Board
Soft-Charge
Resistor
Diode
Packs (3)
Front DC
Storage
Capacitor
7-36 7 Troubleshooting
Anthem Installation, Operation, & Service Manual
T1 T2 T3
Figure 7-28. Diode pack test points
U3
U4
2. Put the Anthem into the Prep state and verify that the voltage on U4-13
goes low.
1. If, during Prep, this voltage goes low and the 280 VAC is present at QC9
and QC11 but not at QC10 and QC12, replace the Power Relay board.
2. If the voltage on U4-13 remains high (+12 VDC) during the Prep state,
verify the voltage at U3-5 on the Power Relay board:
• during Prep: low
• during standby: The voltage here should be +5VDC in standby and go
low during Prep.
7 Troubleshooting 7-37
Anthem Installation, Operation, & Service Manual
3. If U3-5 goes low during Prep but U4-13 remains at +12 VDC, replace the
Power Relay board.
If U3-5 remains high during Prep, swap the ribbon cable between Power
Relay board TB4 and Main Control board J4 with another cable to see if
the condition changes. If this is still true after the cable swap, replace the
Main Control board.
Warning
This condition is potentially dangerous. Due to the excessive heat
built up in the X-ray tube, higher than normal pressures are also
present. If the thermal switch fails, is tampered with, or is electrically
bypassed, this condition can be an explosion hazard.
7-38 7 Troubleshooting
Anthem Installation, Operation, & Service Manual
Corrective Action
Warning
Do not attempt any prep or exposure commands while this error is
present.
1. Allow the X-ray tube to cool completely. Refer to the tube manual for
thermal characteristics and cooling data.
2. Monitor the thermal switch in the tube.
a. Inspect the condition of the thermal switch wires, and inspect the
connections at the Room Interface board TB3-4 and TB3-5.
b. Disconnect the two thermal switch wires from the Room Interface
board TB3-4 and TB3-5.
c. Measure the resistance of the thermal switch and wires.
If the switch is open, this indicates an overheat condition.
If the switch is closed, this indicates a normal condition.
The thermal switch automatically closes when the tube is cool enough
for safe use.
d. If this error message reoccurs or persists, contact your service
representative
Q. Can I change the default APR program from Medical to Vet on my Anthem
Console?
A. Yes. Follow these steps:
1. Power off the console.
2. On the bottom row of buttons, below the display, press and hold the
second button from the left.
3. While holding this button, turn the console on, then release the buttons
7 Troubleshooting 7-39
Anthem Installation, Operation, & Service Manual
4. Enter the password (default 0000) and press OK. You enter Service Mode.
5. Select FLASH OPTIONS
6. Select DFLT APR DATA. You see “Are you sure” After a few seconds the
console beeps and you see OK
7. Press Exit, and then press Exit again. You see VET.
You can use the same procedure to go from Vet to Medical.
Q. My room layout has changed and now the console is further than 45’ from the
Power Cabinet. Is a longer console cable available?
A. There are longer console cables available: 65’ Part# 126-5174G65 and 75’
Part# 126-5174G75
7-40 7 Troubleshooting
Anthem Installation, Operation, & Service Manual
Prep:
Prep:
7 Troubleshooting 7-41
Anthem Installation, Operation, & Service Manual
Ready:
7-42 7 Troubleshooting
Anthem Installation, Operation, & Service Manual
7 Troubleshooting 7-43
Anthem Installation, Operation, & Service Manual
7-44 7 Troubleshooting
Anthem Installation, Operation, & Service Manual
Ready:
7 Troubleshooting 7-45
Anthem Installation, Operation, & Service Manual
7-46 7 Troubleshooting
Anthem Installation, Operation, & Service Manual
7 Troubleshooting 7-47
Anthem Installation, Operation, & Service Manual
7-48 7 Troubleshooting
Anthem Installation, Operation, & Service Manual
Prep:
7 Troubleshooting 7-49
Anthem Installation, Operation, & Service Manual
7-50 7 Troubleshooting
Anthem Installation, Operation, & Service Manual
Rail Charged:
7 Troubleshooting 7-51
Anthem Installation, Operation, & Service Manual
7-52 7 Troubleshooting
Anthem Installation, Operation, & Service Manual
7 Troubleshooting 7-53
Anthem Installation, Operation, & Service Manual
7-54 7 Troubleshooting
8 Component Replacement
Component
Replacement 8
8.1 Introduction
This chapter provides instructions for replacing most of the major assemblies
on the generator.
Warning
All service and maintenance including the procedures described
within this chapter, are to be performed by qualified service
personnel only.
Tools Required
• standard hand tool kit (screwdrivers, wrenches, etc.)
Warning
Turn off all electrical power to the generator. Also, make sure that
power source is locked out and tagged “Equipment Being Serviced”
before servicing the generator. Otherwise you can be seriously
injured.
Warning
Never perform any service work inside the generator without first
performing the “Discharge Procedure” on page 6-2.
Tools Required
• standard hand tool kit (screwdrivers, wrenches, etc.)
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Never perform any service work inside the generator without first
performing the “Discharge Procedure” on page 6-2.
Side Screws
Front Screws
2. Unscrew the two thumb screws (Figure 8-3) to swing open the power
module door.
Thumbscrews
Tools Required
• Standard hand tool kit (screwdrivers, wrenches, etc.)
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Never perform any service work inside the generator without first
performing the “Discharge Procedure” on page 6-2.
Caution
Always use an anti-static wrist strap when working on electrostatic
sensitive devices.
Caution
Replace fuses only with same type and rating of fuse or machine may
get damaged.
F4
10 A, 250 V
Caution
Make sure there are no tools or loose hardware left inside the
generator before remounting the cover.
Tools Required
• standard hand tool kit (screwdrivers, wrenches, etc.)
• needle-nose pliers
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Never perform any service work inside the generator without first
performing the “Discharge Procedure” on page 6-2.
5. At the open front of the power module, pull the disconnected transformer
wires through the rear panel opening.
6. Unscrew the mounting nut from the bolt that secures the line match
transformer to the mounting bracket (Figure 8-7). Remove the line match
transformer.
Line Match
Transformer
Mounting Nut
Caution
Make sure there are no tools or loose hardware left inside the
generator before remounting the cover.
4. Disconnect all wires from the isolation transformer on TB6 and TB7 (see
Figure 8-7) on the Power Relay board.
5. Unscrew and retain the bolt holding the isolation transformer in place and
remove the isolation transformer.
6. Mount new isolation transformer to the bottom of the power module
using the bolt retained in step 4.
7. Connect all wires to the Power Relay board. Each wire from the isolation
transformer is labeled with the corresponding terminal number on the
Power Relay board TB6 and TB7 (see Figure 8-7).
Caution
Make sure there are no tools or loose hardware left inside the
generator before remounting the cover.
Tools Required
• standard hand tool kit (screwdrivers, wrenches, etc.)
• needle-nose pliers
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Never perform any service work inside the generator without first
performing the “Discharge Procedure” on page 6-2.
The three line match transformers are located on the bottom of the lower
cabinet (Figure 8-8). Each line match transformer is connected to its
corresponding terminal block above the transformers.
4. At the applicable terminal block, disconnect the wires from the line match
transformer(s) you are about to replace.
5. Unscrew and retain the bolt(s) holding the line match transformer(s) in
place and remove the line match transformer(s).
Note: When replacing the line match transformer in the rear, you must
remove the terminal block bracket.
6. Connect the line match transformer(s) power wires to its corresponding
terminal block on the terminal block bracket. Each wire from the line
match transformer is labeled with the corresponding socket number on
the terminal block.
Caution
Make sure there are no tools or loose hardware left inside the
generator before remounting the cover.
Mounting Bolt
Isolation
Transformer
TB6 TB7
Figure 8-9. Location of isolation transformer and connections (3-phase)
5. Unscrew and retain the transformer mounting bolt, and remove the
isolation transformer.
6. Mount the replacement isolation transformer to the bottom of the power
module using the bolt retained in Step 5.
7. Connect all wires to the Power Relay board. Each wire from the isolation
transformer is labeled with the corresponding terminal number on the
Power Relay board TB6 and TB7 (see Figure 8-9).
Caution
Make sure there are no tools or loose hardware left inside the
generator before remounting the cover.
Filament Board
Figure 8-10. Board locations
Caution
Always use an anti-static wrist strap when working on electrostatic
sensitive devices.
Tools Required
• standard hand tool kit (screwdrivers, wrenches, etc.)
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Never perform any service work inside the generator without first
performing the “Discharge Procedure” on page 6-2.
Caution
Always use an anti-static wrist strap when working on electrostatic
sensitive devices.
Tools Required
• standard hand tool kit (screwdrivers, wrenches, etc.)
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Never perform any service work inside the generator without first
performing the “Discharge Procedure” on page 6-2.
Caution
Always use an anti-static wrist strap when working on electrostatic
sensitive devices.
13. Reconnect the ribbon cable connectors at TB2, TB3, TB4, and TB5.
OC1/2
TB13A TB1
TB8
TB4 OC9-12
TB2
TB5
TB3
14. Reconnect the wires at TB6 and TB7, carefully observing the labels on the
wires.
15. Reconnect the wires at OC1/2/9/10/11.
16. Reattach the upper cabinet cover and reconnect power.
Tools Required
• standard hand tool kit (screwdrivers, wrenches, etc.)
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Never perform any service work inside the generator without first
performing the “Discharge Procedure” on page 6-2.
Caution
Always use an anti-static wrist strap when working on electrostatic
sensitive devices.
Part Number
124-5130G1
or
124-5144G1
Tools Required
• standard hand tool kit (screwdrivers, wrenches, etc.)
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Never perform any service work inside the generator without first
performing the “Discharge Procedure” on page 6-2.
Caution
Always use an anti-static wrist strap when working on electrostatic
sensitive devices.
IPM Connectors
and Guide Pins
IPM Blocks
5. Unscrew and retain the 4 mounting screws. The IPM Driver board is
connected to the two IPMs on the lower part of the board. Gently pull the
board straight out to disconnect it from the IPM blocks.
6. Carefully connect the replacement IPM Driver board to the IPMs,
directing the board’s guide pin holes over the guide pins and ensuring
that the pins on each connector mate correctly.
7. Secure the IPM Driver board with the 4 mounting screws.
8. Reconnect all cable connectors to the IPM Driver board.
9. Reinstall the cover and reapply power.
654-5001P1
(Silver Baseplate)
654-5001P2
(Copper Baseplate)
Tools Required
• standard hand tool kit (screwdrivers, wrenches, etc.)
• thermal grease
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Never perform any service work inside the generator without first
performing the “Discharge Procedure” on page 6-2.
Caution
Always use an anti-static wrist strap when working on electrostatic
sensitive devices.
Gently pull the board straight out to disconnect it from the IPM blocks
(Figure 8-13).
IPM Driver Board
IPM Connectors
and Guide Pins
IPM Blocks
Copper Rails
Tools Required
• standard hand tool kit (screwdrivers, wrenches, etc.)
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Never perform any service work inside the generator without first
performing the “Discharge Procedure” on page 6-2.
Caution
Always use an anti-static wrist strap when working on electrostatic
sensitive devices.
Electrical Schematics
9
9.1 List of Electrical Schematics
Current
Drawing Number Description Revision
034-5089 Wiring Diagram, single-phase 5
034-5090 Wiring diagram, 3-phase 6
110-5105 HV Tank 1
124-5162 Room Interface 0
Power Supply Distribution
Line Monitor
Rotor Circuit
Filament Circuit
kV and mA Feedback, kV Drive
n/a n/a
Soft Charge, Main Contactor, and DC
Bus
AEC Circuit
Console
Room Interface
Table 9-1. List of Schematics
BR3
+12V1 CB1 20VAC TB6-11
TP6 +
TP23 From Toroidal
Transformer
0V TB6-10
PS1 VR3 -
+5VC
VR2
PS1
VREF
TP24 LED4
VR3
120VAC 120VAC COM
+12VAEC
-12V TB7-42 TB7-41 TB7-34 TB7-33 TB7-27 TB7-28
TP20 -12VAEC PS2
TP22
J13A-3 TB13A-3
-15V
VR4
J7-42 J7-41 J7-34 J7-33 J7-27 J7-28 Filament Board
+5V
TP13
J13A-4 TB13A-4
+12VAEC
TP12 TB8-9
120VAC COM J8-4
J13A-5 TB13A-5
-12VAEC TP1
120VAC TB8-10 J8-3
LED1
TB7-1
In
Toroidal
Transformer
Thermal Switch +12VS
TB7-2 TP10
+12VS
VR4 Line monitor /
TB1-8 TB7-3 comparator circuit
3 Phase
External Line Match
Transformer
Thermal Switches TB3-9 TB7-4
+12VS
TP10B
TB7-3
+12VSB
Single Phase VR4B
Power Module Line
Match Transformer
Thermal Switch TB7-4
TP11
In Out
240VAC TB7-5
TB1-3
Single Phase
From Power Module
TB1 0VAC TB7-6
TB1-1
BR5
240VAC TB7-5 240VAC T1 V-A
3 Phase TB2-3 +
From External Line
Match Transformer 240VAC TB7-6 240VAC COM
TB3-3 - +12VS
D5
CB4 TB6-1
To Toroidal
TB6-2 Transformer
K11
Prep
+12V1
To Console
TB1-10 J1-3
In In
+3.3VA
TP1 Out
In
+5V
+3.3V
TP37 +12VDC
U1-9
Rotor Current
VREF Rotor FB Sensing Circuit
Data BUS and In Out In
Microprocessor
TP41 LED4 U100-1
Out
U27-6
U1-13 U4-16 U4-6
U7-3 RN2-12
Main Control Board Diagram – Filament / mA Drive and Feedback Circuit Filament Board
Del Medical
Anthem Generator TB1-2
X-SM
2018.05.08
HV Cable length
To HV Tank TB1-1 X-LG compensation
jumpers J4, J5, J6
+12VDC
K1
TP52
From Toroidal
X-Select
TP2
X-Out
FIL HI Fault Transformer
Out
R81 12
TP3
TP50 Out
MA ACT 8
FIL DR1 BR4
16
CB2 TB8-7 J8-6
36VAC
+
FIL PCB Fault Out TP54
U13 LED3
CB3 TB8-5 J8-8
36VAC - FIL -50V
U20-5 R84 MA / Fil Ref In
Out 15 11
FIL DR2 TP1
7 TB8-6 J8-7
FIL EN In
FIL COM
FIL COM
TP44 TP4
TP45 TP5
TP1
TP9
TP22
+12V +5V
Buffer +12VDC
Circuit
Out
U18-12 U7-6 OC5-4 TP6
RN2 LED1
J4-4 TB4-4 FIL Fault TB3-4 J3-4
In FIL Fault In Out
6 11 FIL Fault
Continued on
diagram Filament,
mA, and KV Drive
Out
Circuit
U42 Out Out P
1 5 -15V In
In
Prep 2 4
TP1
TP9
LED2
+15V
TP11
+12V
TP2
2 1 Out
+15V In
In In In
TP13 KV+ U5 TP12 J2
In Drive
Out
PS2 IPM1
mA 4 5 Circuit
FIL HI Fault SC Complete In N
In In Out -15V In
In
TP3
TP12
In
KV-
MA HI Fault IPM Fault
In In Out
R22 TP3
TP6
BUS Monitor J3-7 J5-7 BUS Monitor TP13
In +12V
FIL PCB Fault In In KV FB Out
U6-1 TP4 +15V Out
In
TP8 TP14 J3
SC Complete J3-6 J5-6 Drive
Out
Data BUS and Main Board In SC Complete PS3 IPM2
Circuit
FIL EN Microprocessor KV EN I/O P
Out Out In -15V In
U6-5 TP7 In
TP10
IPM Fault J3-5 J5-5
In IPM Fault
FIL REF In KV IMB Fault Out
Out
U5-11 TP6B
TP11
KV DR2 J3-4 J5-4 KV DR2 TP15
Out +12V
MA REF KV HT Fault
Out In Out
U5-8 TP6A +15V Out
In
TP7
J3-3 J5-3 TP16 J4
KV DR1 KV DR1 Drive
Out
Out PS4 IPM2
FIL FB KV REF Circuit
In Out In
-15V N
In
+12V In
TP22A TP6A
1V=20KV
J3-1 J5-1
TP47A
SM1V=20mA
TP22B LG1V=100mA TP6
+15V
TP5
PS5
5 1
24VAC TB6-17
In In
K1
U2-16
K1 TB6-18
+12V1
Out +12V1
QC12
+5V IPM2
+12V1
U4-13
QC9
TP42 U3-5
Buffer
Circuit
To IPM Driver
Board J6
To HT Tank
J21-6
TP22C
R303 U301 10
J21-3
In
W AEC FB 8
9 +5V +12VAEC
J21-2
U305-1 U305-15
U305-3 U305-13
U306-1 U306-15
U305-5 U305-11
TP300 +12VAEC
U303
4 12 J19-2
14
13
11
+12VAEC P1
J19-3
8 2
T AEC Stop 7 -12VAEC
In T AEC REF
3
1 R311 6 R313
R310 9 7 2 J19-6
U302 T AEC Ramp 8 5 1
10 3
U303
TP303 U304 U303
J19-5
AEC REF R305 3
U301
Out 1
2 To Table Chamber
J19-4
J19-7
U301 -12VAEC
R308 5
In T AEC FB 7
6
J19-8
+12VAEC
J19-9
Diagram – Console
Del Medical Prep / Expose Board
Console Board J10-1
Anthem Generator +5V
2018.05.08 Prep SW J10-5 Prep
J13-11 J5-14
CAN LO
+5V1
TP15
LCD Backlight
HV +300VAC J2-1
Data BUS and J13-1 J5-10 Console Data
CAN GND
Microprocessor BUS and
Microprocessor PS2
HV RTN J2-4
LED1
Heartbeat +3.3V
J1
+3.3V
+3.3V
Membrane
+3.3V Switches
8/9
J13 SD
Card
3/6
TP15 TP22 TP5
+5VC +5V1
C2+ C4B+
VR1A
PS1
+3.3V +5V
TP2 C4+ C3+
TB4-5
120 VAC COM
TP1
TP22
TB2-1
TB2-2
TB2-3
+5V
TB2-4
TB8-3
120 VAC
K3 K2
K5 TB8-4
Current
Drawing Number Description Revision
110-5105 HV tank 1
110-5115 Single phase power module 9
110-5115 Single phase power module, in table 9
110-5116 3-phase power module 9
110-5116 3-phase power module, in table 9
112-5469 Base power module 2
112-5471 Door power module 7
112-5472 Inverter power module 1
112-5484 lower cabinet 4
112-5485 Line Match Transformer Ext Assembly 4
112-5465 Control Console Assembly 4
Table 10-1. List of assembly drawings
110-5142G1-G10
35235,(7$5<$1'&21),'(17,$/
7+(,1)250$7,21&217$,1(',17+,6
'5$:,1*,67+(62/(3523(57<
$60+97$1.
2)'(/0(',&$/$1<5(352'8&7,21,1
3$5725$6$:+2/(:,7+2877+(:5,77(1 5(9
3(50,66,212)'(/0(',&$/,6352+,%,7('
%/220,1*'$/(,/
8000uF
8000uF
8000uF
8000uF
112-5700G1/G2/G3
112-5700G1/G2/G3
$
& &
* $60/2:(5&$%,1(73+
* $60/2:(5&$%,1(73+
$
6(&7,21$$
,7(012 3$57180%(5 '(6&5,37,21 *47< *47<
3 (48,30(17&$%,1(7
187+(;.(36
3 /$%(/($57+*5281'
6&5(:++06[
3 /$%(/6(5,$//2:(5&$%,1(73+
* +$51(66/2:(5&$%,1(7
6&5(:331+066(06[
3 75$16)250(53+$6(/,1(0$7&+
3 %5$&.(77(50,1$/675,3
3 7(50,1$/%/2&.326
3 /$%(/7(50,1$/675,33+;)05
3 /$%(/7%3+;)057%
3 /$%(/7%3+;)057%
3 /$%(/7%3+;)057%
187+(;.(36
:$6+(5)/$7
:$6+(563/,7/.67'
6&5(:++06;
3 /$%(/6(5,$//2:(5&$%,1(73+
3 7$*<(//2:&$87,21
%/220,1*'$/(,/