Professional Documents
Culture Documents
Installation: (C) 2005 Koninklijke Philips Electronics N.V. All Rights Reserved
Installation: (C) 2005 Koninklijke Philips Electronics N.V. All Rights Reserved
Installation: (C) 2005 Koninklijke Philips Electronics N.V. All Rights Reserved
LIST OF PAGES & DRAWINGS MODULE CODE NUMBER 4522 981 44684
1 ... 54 (05.1)
INSTALLATION
Contents
1 INTRODUCTION .....................................................................................................................................3
1.1 HaemoSphere (Hemodynamic System) ..................................................................................................3
1.2 HaemoSphere Compact (Hemodynamic System)...................................................................................4
1.3 Module version.........................................................................................................................................4
1.4 Tools ........................................................................................................................................................4
1.5 Required manpower ................................................................................................................................4
1.6 IEC 60601-1-1..........................................................................................................................................4
2 QUICK CHECK........................................................................................................................................5
3 UNPACK..................................................................................................................................................5
1 INTRODUCTION
The purpose of a Hemodynamic System is:
• Locate and identify irregularities within the heart and its valves.
• Define the size (severity) of lesions.
• Locate and identify electrical conduction abnormalities.
• Monitor patient condition during catheterization.
• Examination Room:
- Patient Connection Box (mounted on AD5 connection box)
- Slave Monitor (mounted in Monitor Ceiling Suspension, but fully insulated from the X-ray System)
- Option: Multi Measurement Server (mounted on the AD5, if IntelliVue MP90 is present)
1.4 TOOLS
For the mechanical installation of the Hemodynamic System the engineer will need:
• Service toolkit
• Soldering iron
• Wireman's snake (or rope) for adding cables in the 's cable harness
• Only for CRT monitors in the Monitor Ceiling Suspension:
Power drill and metal reamer (or 8mm HSS drill) required for enlarging holes in monitor mounting bracket
(to enable mounting the Monitor insulation set).
NOTE
No insulation transformer (MPSO) is required, insulation transformer already inside Hemodynamic system.
________________
2 QUICK CHECK
For an efficient installation check if the following is present:
• Table in control room to position the LCD's and Function Keyboard on.
Table is not required for HaemoSphere Compact with trolley !
• Space for the Main System (or trolley) in the control room (next to the table)
• 2 free wall sockets (hospital mains).
• 1 protective earth connection block/socket (PCB). The Hemodynamic system requires a separate
connection for protective earth (preferable connected to the PCB).
• Cable duct to route cables from:
- the Main System <--- to ---> the Monitor Ceiling Suspension in the Examination Room
- the Main System <--- to ---> the Patient Connection Box to be mounted on the AD5 Connection Box.
CAUTION
3 UNPACK
• Check if the shipping carton/container(s) is undamaged and note any signs of severe stress as an
indication of rough handling in transit. This may be necessary to support claims for hidden damage that
may only become apparent during subsequent testing.
• Open the shipping carton/container(s) and examine each part for visible damage.
5.1 INTRODUCTION
The HaemoSphere has 6 program keys. Each program has a channel configuration. The program keys
enable easy switching between different types of examinations. Normally channels 1, 2 and 3 are configured
for ECG I, ECG II and ECG III. The remaining channels can be used for pressure (up to 4 channels) or
additional ECG signals (up to 12 leads).
For more information about channel configuration see chapter 10.14.
On the X-ray system usually only one ECG signal is used. Preferred signal is ECG II (best signal).
The ECG trigger output can be used for ECG triggered injections and/or ECG controlled exposure.
The Physio channel outputs can be used to display a channel on the viewing monitor of the X-ray system.
The outputs are available on a 25 pins D-connector labelled 'TRIGGER' (=PDC1-BU1), see table Table 5-1.
The channels of the output board of the HaemoSphere cannot be configured separatly; they follow the
channel configuration of the program keys. As described above, ECG II will be available on channel 2 in most
cases.
Create the required interface cable using the information in the following chapters.
The wiring of this interface cable is dependent on the type of X-ray system and the user requirements.
• For ECG triggered injections and/or ECG controlled exposure connect "ECG-TRIG".
• If only ECG II has to be displayed, only that signal has to be connected. If so, make sure that all
HaemoSphere program keys (1 .. 6) have channel 2 configured for ECG II (see chapter 10.14).
The Physio interface of the X-ray system is prepared for two ways to connect the HaemoSphere:
• Directly to the M-Cabinet (MDY-X24) (normally used)
• Via the Table Base Connection Board (TBCB) (used for mobile Hemo equipment).
Cable MDY-X24 is delivered with the X-ray system (supplied with the Control Room Connection Box).
This cable has at one end (labeled MDY-X24) a 15 pins D-connector for the M-cabinet, see Table 5-1.
The other end (labeled ECG-Physio) has no connector. To this end a 25 pin male D-connector for the
HaemoSphere unit must be attached (delivered with the HaemoSphere system, see also Table 5-1).
NOTE
The cable between the HaemoSphere system and the TBCB must be made locally.
This cable must at one end have a Burndy connector for the TBCB, see Table 5-4.
To the other end a 25 pin male D-connector for the HaemoSphere unit must be attached (delivered with the
HaemoSphere system, see also Table 5-1).
The cable from the TBCB to the M-Cabinet (connector MDY-X24) is already present when the X-ray system
has been installed correctly.
SD-X(ECG) connector
signal pins cable colors
screen E screen
ECG-TRIG_GND L
ECG-TRIG M
PHY1_GND W
PHY1 K
PHY2_GND H
PHY2 J
PHY3_GND V
PHY3 G
PHY4_GND U
PHY4 F
NOTE
Do NOT connect the screen (MDY-X24 pin 1) to the HaemoSphere connector PDC1-BU1
________________
To display Physio channels on the Integris system, an optional Physio board (VCPHY) is required inside the
VISUB cabinet. VCPHY provides an 8-channel Physio interface. The Physio channels must be connected via
connector WDE-X402 on the VISUB coding field.
Integris systems have separate input boards for ECG trigger and for Physio channels. The ECG trigger input
is located inside the SYSCO cabinet on connector AG26-X8. This ECG-TRIG input of SYSCO can only be
used for ECG triggered exposure. See Table 5-7.
If ECG triggered injection is required, an additional cable towards the ECG input of the injector is required.
See Table 5-8.
NOTE
• Place a soft cloth on the floor or on a table (this will be the place where the monitor will be
positioned on).
• Place the monitor with the screen facing down on the cloth.
• Carefully place the monitor frame on top of the insulation material (make sure it does not move).
• Use the old screws (without the spring washer) to fix the monitor frame.
• Fit the warning label on the monitor (choose a place where it will not get damaged or fall off).
NOTE
• Add the Protective Earth cable and Power cable for the Slave Monitor into the hose of the Monitor Ceiling
Suspension.
• Route the 5 video cables, the protective earth cable and the power cable to the HaemoSphere Main Unit.
CAUTION
Under NO circumstances exceed maximum cable length (30m for slave monitor).
DO NOT route HaemoSphere cables close to the High Tension cables.
________________
• Fix the ECG cable with cable binders to the MMS cable.
• Xper systems:
Mount the bracket of the Patient Connection
Box on the cover of Table Base Connection
Box (no picture available).
CAUTION
• Route the female connector of the Fiber Optic Cable (FOC) through the AD5 base.
− Please note that there is a male and a female connector on the Fiber Optic Cable:
The female connector with a normal plastic housing is for the Patient Connection Box,
The male connector with a silver colored isolated housing is for the Main Unit.
• Connect the female connector of the Fiber Optic Cable (FOC) to the Patient Connection Box.
• Route the Fibre Optic Cable (FOC) and the long MMS cable to the Main Unit in the control room.
CAUTION
NOTE
When the Remote Alarm device is placed inside the Main Unit,
take care that the button cannot be pressed accidentally.
This button switches off the IntelliVue MP90!
________________
− Mount bracket to
Remote Alarm Device.
− Mount Clamp Wall
− Hook Remote Alarm
Device to Clamp Wall
CAUTION
• Connect the BNC cables of the Slave Monitor (in the examination room) to the BNC-VGA adapter cable
(using the BNC-BNC connectors).
• Connect the BNC-VGA adapter cable to the HaemoSphere unit, and store the BNC-BNC connectors in
the HaemoSphere Unit.
• Install the LCD monitor(s) on the table and connect them to the HaemoSphere Unit (video and power).
• Install the laser printer (option), and connect it to the HaemoSphere Unit.
• Connect the power supply of the laser printer to the hospital mains wall socket.
• Check if the Hemodynamic equipment is electrically insulated from the X-ray equipment.
• Connect the PE of the Hemodynamic equipment to the PE bushbar, on which the X-ray equipment has
been connected.
10 CONFIGURE HAEMOSPHERE
10.1 SWITCH ON
• Check that the correct monitor driver has been installed (Display options), if not:
Install driver on Hemodynamic PC.
The input locale must be set in accordance to the used keyboard type:
• Click on OK
NOTE:
NOTE
This is only required if the Hemo system has the IntelliVue MP90 option!
________________
NOTE
The second network adap[ter is only used for a local network to communicate with the IntelliVue MP90.
The network address of this local network must be different from the network address of the hospital network.
________________
NOTE
If more that one Hemodynamic system is connected to the hospital network, make sure that each
Hemodynamic PC has a different Computer name (e.g. hemo-pc-1, hemo-pc-2, ...)
________________
If you have the IntelliVue MP90 option, check with the network administrator that there is no conflict with the
default network settings for the connection to IntelliVue MP90 (see chapter 10.7.1). The network address of
local IntelliVue MP90 network must be different from the network address of the hospital network.
Item Value
Computer Name hemo-pc / ...
If a remote SQL server on the hospital network is used, contact the network administrator for the remote SQL
server name and the path to the folder on the remote SQL server where the patient data is stored. Use the
table below to write down the values. (These values are used in chapter 10.7.7.)
Item Value
Remote SQL server name ...
NOTE
This chapter is only required if your system has the IntelliVue MP90 hardware and the hospital network
administrator asked you to change these settings (see chapter 10.7.1 and 10.7.2).
________________
NOTE
This chapter is only required if your system has the IntelliVue MP90 hardware and the hospital network
administrator asked you to change these settings (see chapter 10.7.1 and 10.7.2).
________________
The IntelliVue MP90 is connected to the Hemo PC via a second network adapter in the Hemo PC.
To provide the IntelliVue MP90 with an IP-Address, the program SimpleBootP is used.
• Log-on as Administrator
− Click Exit
• Rename file "clients.cfg" into "clients.cfg.old" (this file contains the old IP address info)
• Start SimpleBootp:
− Start Menu > Programs > Startup > SimpleBootp.exe
The SimpleBootp window will appear automatically.
NOTE
This chapter is only required if you changed the Computer name (e.g. to hemo-pc-2)
or
if you will use a remote SQL server on the hospital network
________________
• Click Next
• In the text box: Path on server for the files containing the signal data
type the path to the folder on the remote SQL server where the patient curve data is stored
(see chapter 10.7.2)
• Logon with:
− User = service
− Password = ser98schwarzer
• Select the
Language of the applications
• Click on Accept
NOTE
• Click OK
• (If the icon is not visible, you must start the TrayComP program
manually: D:\Hemodynamic_V3\TrayComP.exe)
The section Send to OnlineViewer allows the customization of the display of vital signs on the Live Signal
Screen. The reserved area on the screen is divided into 4 data blocks. Each Data block has 3 lines.
This gives 4 x 3 = 12 areas in which text and/or signal values can be displayed. Each are can display up to 7
characters.
• To display a value:
− Enter the value code between $-characters, using the format: $[value code].[number of decimals]$
− For the [value code] see the table in chapter 10.13
− The [number of decimals] can be 0, 1 or 2
− See data block 1 in the example
The section Send to protocol allows the customization of the vital signs to be registered in the chronological
report.
The field Send has to be configured as described above (Send to OnlineViewer). In field "every" you can set
the time in seconds how often the value has to be registered. Select "every new value" if you don’t want a
regularly registration but a registration of changing values.
You can store up to 10 different configurations. They are easily changeable by right clicking the icon:
(right corner of the Taskbar)
NOTE
If you will use a Physio Interface to display channels on an X-ray System, the channel configuration must be
in line with the way you have created the Physio Interface cable, see chapter 5.
________________
• To add a channel:
− Click on New Channel... in the channel list
The Channel selection window is displayed
• To remove a channel:
− Click on the channel in the channel list you want to remove
− The Channel selection window is displayed
− Click on Off
Click on OK
− The serial number (00xx) is given by Schwarzer and must not be changed!
− The port number (3100) has to agree with the x-ray system and can be changed (see next page).
example
X-Ray System
Port Number 0-65535 65535 Agreed between Philips and Schwarzer
X-Ray System Given by the hospital or
Network Address ddd.ddd.ddd.ddd 160.0.2.10 agreed between Philips and Schwarzer
X-Ray System Given by the hospital or
Host Name logical name RIS agreed between Philips and Schwarzer
X-Ray System
Application Entity Title AE name VISUBRIS_AE Agreed between Philips and Schwarzer
Information System (IS) Defined in wlmscpdb.bat or
Port Number 0-65535 3100 agreed between Philips and Schwarzer
Information System (IS) Given by the hospital or
Network Address ddd.ddd.ddd.ddd 160.0.2.15 agreed between Philips and Schwarzer
Information System (IS) Given by the hospital or
Host Name logical name CL1 agreed between Philips and Schwarzer
Information System (IS)
Application Entity Title AE name AE_CL1 Agreed between Philips and Schwarzer
(ddd=0...255)
NOTE
If there is more than one modality in the network, different port numbers and AE-Titles have to be defined.
For every system (X-Ray + Hemo) one RIS.bat must be started, configured with the corresponding port
number (e.g. RIS1.bat, RIS2.bat,…).
________________
For more detailed information see the DICOM Conformance Statement on the RIS Interface CD.
11.3 WORKFLOW
After filling in the patient basic data click on: Transmit patient’s data to x ray
NOTE
Registration is only required if the HaemoSphere system is equipped with Dell PC's.
________________
13.1 EXPLANATION
When you received a Dell workstation, the registration must be done at the Dell website:
http://support.dell.com/us/en/kb/document.asp?dn=1011182#2ndstep
CAUTION
If you do not register the workstation, you will NOT receive any support from Dell (hardware issues).
________________
This procedure is used to transfer ownership of and service and support on your system to your local service
organization.
It is the policy of Dell Computer Corporation and its subsidiaries to export and/or re-export products, services
and technology in accordance with the regulations and licensing requirements set forth in the Export
Administration Regulations as governed by the United States Department of Commerce, and the European
Union Regulations.
Unless Dell has prior express permission from the U.S. Government, Dell is prohibited by U.S. law from
engaging in any type of transaction, including sales, transfers, consignments, replacements, or downloads of
product, software, technology or service, with the following countries: Afghanistan, Cuba, Iran, Iraq, Libya,
North Korea, Syria, or Sudan. This prohibition applies equally to any company or person that is from, in, or
representing any of these countries.
This form should be completed and submitted to Dell whenever a Dell unit is being physically transferred from
one country to another. Transfers between countries not listed are not currently possible.
13.2 PROCEDURE
1. Go to: http://support.dell.com/us/en/kb/document.asp?dn=1011182#2ndstep
2. Select registration for a system Inside the United States or Outside the United States
The information below will help you to fill in the form (all data, which is bold, is fixed!):
1)
The Service Tag can be found on the front of the PC,
behind the USB cover.
Or:
• Right-click on My Computer
• Select Properties
• Click on Support Information
NOTE
NOTE
• Log-on as CL1:
− User Name = CL1
− Password = cl
• Check connection to SQL server (message in title bar of Select examination window)
• Check measurement:
− pressure
− pull-back
− gradient
− etc.
• Check Reporting
− Switch to IntelliVue
• Open and view patient to check that examination has been saved
• Log-on as Admin:
− User Name = Admin
− Password = Admin