(Dosage) The Importance of Irradiance and Area in Neonatal Phototherapy
(Dosage) The Importance of Irradiance and Area in Neonatal Phototherapy
ORIGINAL ARTICLE
Background: Despite a long period of development, there are still considerable variations in the spectral
output, the levels of irradiance, and irradiated area provided by commercial phototherapy systems. These
See end of article for
variations depend on the types and output of the lamps used to produce the phototherapy, along with the
authors’ affiliations design of the systems, and principally on whether the phototherapy is provided from overhead or
....................... underneath.
Objective: To see whether commercially available phototherapy systems produce sufficient irradiance over
Correspondence to:
Mr Hart, Medical Physics the surface area of the neonate.
Department, Bradford Methods: Surface plots of the output irradiance were made on a number of systems and used to calculate
Royal Infirmary, the effective irradiance on the surface of a premature or term baby, using mapped outlines.
Duckworth Lane, Bradford
BD9 6RJ, UK;
Results: A 10-fold difference in peak central irradiances was found between the systems tested, with a
graham_hart_1953@ fourfold to fivefold difference in effective irradiance to the baby surfaces. Although work published over 20
yahoo.com years ago showed that levels of irradiance should reach 2 mW/cm2 to achieve optimal effectiveness,
some of the commercial systems tested do not appear to achieve this level.
Accepted 2 March 2005
Published Online First Conclusion: Purchasers of neonatal phototherapy systems need to take into account whether the systems
4 May 2005 will produce sufficient irradiance over the area to ensure maximal effect, to keep the treatment time to a
....................... minimum.
A
recent report from the American Academy of Pediatrics effectively double the treatment area by irradiating the
on neonatal hyperbilirubinaemia confirms the efficacy neonate from below as well as above.
of phototherapy as a treatment.1 Despite the long
period of development since the original paper of Cremer et
MATERIALS AND METHODS
al,2 there are still considerable variations in the spectral
The effective irradiance of a number of phototherapy systems
output, levels of irradiance, and irradiated area provided by
was assessed by taking a series of irradiance measurements
commercial phototherapy systems. This was noted in the
at 2 cm intervals in the illuminated field. Measurements
American Academy of Pediatrics report.
were made using an International Light IL1700 radiometer
It is clear that the irradiance produced by phototherapy
with an SED033 detector, a ‘‘BR’’ bilirubin filter, and a ‘‘W’’
systems is an important factor. Many authors have shown
cosine diffuser. The radiometer and detector were calibrated
that increasing irradiance produces a faster rate of fall of
serum bilirubin.3–8 Tan9 showed over 20 years ago that against a secondary standard spectroradiometer, using a
irradiances of 2 mW/cm2 (,40 mW/cm2/nm) are needed Philips TL52 phototherapy lamp as the light source.
before a possible plateau response in the rate of bilirubin All measurements for the overhead phototherapy systems
degradation is reached. Nevertheless, some current commer- were made with a source to measurement distance of 45 cm
cial phototherapy systems do not produce these irradiances. at the centre of the field, representing a typical treatment
Bringing the lights closer could increase the irradiance of distance for neonates in incubators. For the fixed or under-
overhead phototherapy systems. As this cannot easily be neath systems, measurements were made on the surface of
achieved for neonates in incubators, one solution is to the treatment area for that system, which included a blue-
provide phototherapy from underneath, as the neonate thus green gel mattress (JLJ Healthcare, Halesowen, West
lies closer to the lights. This has been shown to be more Midlands, UK). Measurements were made until the output
effective than conventional overhead phototherapy.10 had fallen to about 10% of the peak value or until the sides of
Another solution is to provide phototherapy from both above the crib had been reached.
and below. This also seems to show a greater level of efficacy The method used for the irradiance mapping was similar to
than overhead phototherapy alone.11 that used by Dicken et al12 and Vreman et al.13 The data were
The other parameter that affects the effectiveness of entered into Excel spreadsheets and then linearly interp-
phototherapy is the area of the neonate irradiated by the olated from the 2 cm grid used during the measurement
system. Maisels8 highlighted the need to consider this issue, process into a 1 cm grid to aid the mapping process. Outlines
and it was also used by Dicken et al,12 Vreman et al,14 and of both a premature and a term baby, similar to that of
Wentworth14 in the assessment of the overall performance of Vreman et al, were then mapped on to other spreadsheets.
phototherapy systems. The only significant difference from the approach of Vreman
The above appear to confirm the idea that many conven- et al is the use of an ellipsoidal model to simulate the curved
tional phototherapy systems do not produce sufficient edges of the body. This model suggested that the effective
irradiance to maximise the degradation of bilirubin. irradiance within 1 cm of the edge of the head and body
If overhead phototherapy does indeed reach Tan’s plateau should be reduced from 100% to 80%, and to 50% for the
value of optimum irradiance, logic suggests that the way to arms and legs, which reduces the overall effective irradiance
increase the effectiveness of phototherapy would be to by ,14%. Figure 1 shows the outline for the premature baby.
www.archdischild.com
F438 Hart, Cameron
A B
11.00 11.00
10.00 10.00
9.00 9.00
Irradiance (mW/cm )
Irradiance (mW/cm2)
2
8.00 8.00
7.00 7.00
6.00 6.00
5.00 5.00
4.00 4.00
3.00 3.00
2.00 2.00
1.00 1.00
0.00 0.00
C D
11.00 11.00
10.00 10.00
9.00 9.00
Irradiance (mW/cm2)
Irradiance (mW/cm )
2
8.00 8.00
7.00 7.00
6.00 6.00
5.00 5.00
4.00 4.00
3.00 3.00
2.00 2.00
1.00 1.00
0.00 0.00
Figure 2 Surface plots for (A) Microlite, (B) Draeger, (C) Duo, and (D) neoBLUE phototherapy systems. The plots show a three dimensional
representation of the irradiance produced by the phototherapy systems over the area of the crib. Irradiance on the vertical axis is measured in mW/
cm2. The horizontal axes represent distance on the crib surface.
www.archdischild.com
Irradiance and area in neonatal phototherapy F439
A B
11.00
11.00
10.00 10.00
9.00 9.00
Irradiance (mW/cm2)
Irradiance (mW/cm2)
8.00 8.00
7.00 7.00
6.00 6.00
5.00 5.00
4.00 4.00
3.00 3.00
2.00 2.00
1.00 1.00
0.00 0.00
C D
11.00
11.00
10.00
10.00
9.00
9.00
Irradiance (mW/cm2)
Irradiance (mW/cm2)
8.00
8.00
7.00
7.00
6.00
6.00
5.00
5.00
4.00 4.00
3.00 3.00
2.00 2.00
1.00 1.00
0.00 0.00
Figure 3 Surface plots for (A) IV2 (overhead), (B) Bilibed, (C) IV2 (under), and (D) LED phototherapy systems. The plots show a three dimensional
representation of the irradiance produced by the phototherapy systems over the area of the crib. Irradiance on the vertical axis is measured in mW/
cm2. The horizontal axes represent distance on the crib surface.
www.archdischild.com
F440 Hart, Cameron
Table 1 Peak and integrated irradiances of the systems under test, together with the ratios of the total irradiance when the
outlines were moved 4 cm and 8 cm
Central peak Central peak Term baby total Prem baby total
irradiance irradiance irradiance at irradiance at Ratio Ratio
(mW/cm2) (mW/cm2/nm) centre (mW) centre (mW) 4 cm:centre 8 cm:centre
www.archdischild.com