INSTALLATION PLAN

STEAM STERILIZER
STERIVAP HP 446-1
STERIVAP HP IL 446-1
STERIVAP HP E 446-1
● The given data are dedicated for dimension of the power supplies and capacities
of the sources of individual media. They take into consideration the optimal instruments
operation also in the adverse conditions and variants.

A Standard-drain Drain max. 15 l/min


Temperature: max. Floor drain DN 65
100 °C
DL Compressed air Connection ends with Connection DN 15 3
5 - 10 bar valve and internal thread Line capacity 15 Nm 3/h
(G1/2“) Consumption approx 0,4 Nm /h
E Electric supply for unit Length of the free cable Connecting terminals Input 2 kW
without steam generator end from the point E-3 m 16 mm2 Protection 10 A
3PE~50/60 Hz ± 5 %, Main switch installed close 1 pc. flexible power Consumption/h approx. 0,5 kWh
400 V ± 10 % to the unit supply cable
FD Steam 2.5 bar ± 10 % Connection ends with Connection DN 25
(overpressure) valve and internal thread Line capacity 110 kg/h
(G3/4“) Consumption approx 12 kg/h
KW Feed water max.15 °C Hardness 1 - 13 °d Connection ends with Connection DN 153
3 - 5 bar valve and internal thread Line capacity 0,9 m /h
3
(G1/2“) Consumption approx 0,13 m /h
Heat radiated by Cooling air flows via Space of unit approx. 1330 W
sterilizer openings in the support Front side approx. 270 W
Sterilized goods approx. 630 W
EK Electric supply for unit Length of the free cable Connecting terminals Input 24,5 kW
with steam generator end from the point E-3 m 16 mm2 Protection 63 A
3PE~50/60 Hz ± 5 %, Main switch installed close 1 pc. flexible power Consumption/h approx. 7 kWh
400 V ± 10 % to the unit supply cable
3
DW Demineralized water Max. 35 cm soluble gases Connection ends with Connection DN 15 3
for the steam generator /l valve and internal thread Line capacity 0,05 m3/h
<10-15 μS/cm 1 - 5 bar (G1/2“) Consumption approx 0,01 m /h
Heat radiated by steam Cooling air flows via Space of the unit 600 W
generator openings in the support
Only when the safety valve blows off out of the unit!
ABBL Valid also for the steam Connection ends with Connection DN 25
generator. Drain for valve and internal thread
safety valve blow-off. (G1“)
Only when dirty steam for heating of the jacket is used!
FDT Steam 2.5 bar ± 10 % Connection ends with Connection DN 20
(overpressure) valve and internal thread Line capacity 80 kg/h
(G3/4“) Consumption approx 8 kg/h
Only for sterilizer with jacket cooling!
EW Feed water For jacket cooling Connection ends with Connection DN 153
Hardness ≤ 0,1°d valve and internal thread Line capacity 0,8 m3/h
3-5 bar (G1/2“) Consumption approx 0,1 m /h
General data:
Dimensions of the unit Width 1200 mm
Height 1918 mm
Depth 970 mm
Transport dimensions (Dimensions of transport (Turning width in a Height 2000 mm
cross-section) corridor: 1700 mm) Width 1150 mm
Specific floor load The value in brackets (+50 kg) approx. 2
Weight holds for the steam 10000 N/m
generator approx. 670 kg
Conditions of the milieu: Environmental +5 to +40 °C
temperature
Max. relative humidity 85 % at 31 °C
Floors – antistatic
treatment
● Validity according to the model and equipment KV_164

STERIVAP HP_446-1_IP_en 1212_V1.02 1/3


INSTALLATION PLAN
STEAM STERILIZER
STERIVAP HP 446-1 L
STERIVAP HP IL 446-1 L
STERIVAP HP E 446-1 L

2/3 STERIVAP HP_446-1_IP_en 1212_V1.02


INSTALLATION PLAN
STEAM STERILIZER
STERIVAP HP 446-1 R
STERIVAP HP IL 446-1 R
STERIVAP HP E 446-1 R

STERIVAP HP_446-1_IP_en 1212_V1.02 3/3