Clinical Systems Microlab

THE SMARTEST CHOICE


FOR LOWVOLUME
LABORATORIES

Cresce
cum fide

Selectra Solutions
Clinical Systems

SPECIFICATIONS
OPTICAL MEASUREMENT PROCEDURES CALIBRATION
• Photometric range: -0.1 to 2.3 Absorbance • Kinetic, with linearity check • Factor, one-point, two-point and multi-
• Wavelengths: automatic by 12 positions • Kinetic, with linearity check and sample point
filter wheel slope blank • Automatic on 1 standard (linear mode)
• 7 Standard interference filters: 340, 405, • Two point kinetic, with or without reagent • Automatic on up to 10 standards (non-
505, 546, 578, 620 and 660 nm blank linear mode)
• 5 positions for optional filters • End point, with or without reagent blank
QUALITY CONTROL
• Detector: Photo diode (320-1000 nm) • Bichromatic end point, with or without
• Two controls per test
• Light source: quartz-halogen lamp reagent blank
• QC survey of last 30 control
12V-20W • End point, with sample blank and with or
measurements
without reagent blank
TEMPERATURE CONTROL • Levey Jenning plot
• Blanking: automatic zero setting
• Peltier system • High/Low flags
• Fixed temperature at 37°C MULTIPLE TESTING
STANDARDS AND REGULATIONS
• Up to 9 replicates
FLUIDIC SYSTEM • CE - IVDD
• Means, SD and CV%
• Internal bellows pump, driven by stepper • USA FDA listed
motor MEASURING TIME • CB
• Back panel connection for waste • Programmable, 2 to 998 seconds for end
POWER REQUIREMENTS
• Aspiration volume programmable point, kinetic and two point assays
• 100-240 Vac nominal, 50/60 Hz.
FLOWCELL DELAY TIME • Battery back-up to retain data
• Metal, with quartz windows • Programmable, 0 to 999 seconds
DIMENSIONS & WEIGHT
• Measuring volume 30 µL
METHOD PARAMETER SETTINGS • 40 cm (16 in) x 17 cm (7 in) x 36.5 cm
OPERATOR INTERFACE • Method name (14 in) (W x H x D)
• Membrane keyboard, for direct function • Measurement mode • 8.5 kg (19 lbs).
and alpha-numeric entry • Wavelength 1 and 2
• Optional external keyboard • Aspiration volume
• High contrast graphical LCD display • Measurement delay
• Measurement time
PRINTER
• Factor
• Internal matrix printer
• Concentration standards
• Normal paper
• Reagent blank y/n
• External printer port available
• Sample blank y/n
SIGNAL INTERFACE • Units for results
• Centronics type parallel port • Levels for flagging
• RS 232 type serial port • Curve fit y/n
• PS 2 type port for external keyboard • Linearity check
• Real time clock, 24 hours system

LANGUAGES
• English, Spanish, French, German and
Portuguese.
• Other languages on request

Order code: 6002-310


WORLDWIDE OFFICES
Please contact your sales representative for terms,
conditions and product availability in your country.

Headquarters +33 1 41 45 07 10
Australia 1800 815 098 New Zealand 0800 555 611
600231049001 EN

BELUX +32 92 820 531 Serbia +381 11 2467119


Brazil +55 27 3025 1415 The Netherlands +31 313 430 581
France +33 4 83 36 10 82 UK +44 1442 869320
Italy +39 011 97 61 91 United States +1 800 453 2725 www.elitechgroup.com
info@elitechgroup.com

You might also like