760 Ventilator System

Simplicity Outside,
Sophistication Inside

The Puritan Bennett ® 760 ™

Ventilator System provides

comprehensive, affordable

respiratory care in the Intensive

Care Unit. This easy-to-operate

ventilator offers all the

capabilities of the 740 ™

Ventilator, plus added features

such as pressure control

ventilation to support pediatric

and adult patients with a wide

variety of clinical conditions.

Puritan Bennett is Exceptional Value—Easy to operate and service, the affordable 760 Ventilator is
one of the world’s ideal for both metropolitan hospitals and remote medical facilities.
premier producers
Technological Sophistication—High performance gas delivery system, combined with
of respiratory-related
frictionless piston and flow triggering, make the ventilator extremely responsive to
medical products
patients’ breathing efforts without the need for a compressed air source.
serving patients
from the hospital Powerful System—The 760 Ventilator delivers mandatory breaths in both volume
to the home. control ventilation and pressure control ventilation.
Standard Accessories Catalog Number Selection Required Catalog Number

Collector Vial Assembly G-061668-00 Hose Assembly, oxygen


D/X7 Expiratory Bacteria Filter (case of 12)

G-060526-00 DISS female x female (United States) 4-001474-00
Re/X 700 ™ Expiratory Bacteria Filter, reusable, 22-mm ISO DISS female x male (Canada) G-061269-00
connectors; one comes standard with every ventilator G-060525-00 NIST (United Kingdom) G-061201-00
D/Flex Inspiratory Bacteria Filter, single-patient use,

Air Liquide (France) G-061190-00
22-mm ISO connectors; one comes standard with every
Australia G-061198-00
ventilator (carton of 12) 4-074601-00
Dräger (Germany) 5-029059-00
Test Lung 4-000612-00
Power Cord
Flex Arm, brushed aluminum 4-032006-00
North America and Japan G-061241-00
Australia G-061242-00
Continental Europe G-061243-00
Denmark G-061244-00
India/South Africa (old British-style plug with round prongs) G-061247-00
Israel G-061248-00
Italy G-061245-00
Switzerland G-061246-00
United Kingdom G-060135-00
Power Supply Kit
English 110 V G-062390-00
English 220 V G-062391-00
French Canadian 110 V G-062386-00
French 220 V G-062387-00
Spanish 110 V G-062394-00
Spanish 220 V G-062395-00
Portuguese 110 V G-062397-00
Portuguese 220 V G-062398-00
Japan 110 V G-062401-00
Japan 220 V G-062402-00
German 220 V G-062405-00
Italian 220 V G-062407-00
Russian 220 V G-062409-00
Poland 220 V G-062411-00
Manual, Operator’s
Czech G-061988-00
English G-061874-00
French G-061981-00
German G-061980-00
Italian G-061982-00
Japanese G-061987-00
Polish G-061985-00
Portuguese G-061984-00
Russian G-061986-00
Spanish G-061983-00
760 Ventilator System
SIMPLICITY OUTSIDE, SOPHISTICATION INSIDE

Modes of Ventilation Operator-Set Alarm Limits


• Assist/Control (A/C) Alarms designed to ISO and
• Synchronous intermittent manda- European standards, and are
tory ventilation (SIMV) classified as high or medium
priority, with distinct audible and
• Spontaneous (SPONT)
visual presentations.
Breath Types High rate: 3 to 100 breaths/min
• Volume controlled ventilation Low insp pressure: 3 to 60 cmH2O
(VCV), available in A/C or (3 to 60 hPa)
SIMV modes High pressure: 10 to 90 cmH2O
• Pressure control ventilation (PCV), (10 to 90 hPa)
available in A/C or SIMV modes Low minute volume: 0 to 50 L
• Pressure support ventilation (PSV), High tidal volume: 20 to 6,000 mL
available in SPONT or SIMV
Low tidal volume: 0 to 2,000 mL
modes
Ventilator Status Indicators
Mandatory (VCV) Settings
ALARM
Respiratory rate: 1 to 70 breaths/min
CAUTION
Tidal volume: 40 to 2,000 mL, NORMAL
compliance-compensated and VENT INOP
corrected to body temperature SAFETY VALVE OPEN
and pressure, saturated (BTPS) ON AC/BATTERY CHARGING
Peak flow: 3 to 150 L/min in A/C ON INTERNAL BATTERY
and SIMV modes (mandatory ON EXTERNAL BATTERY
breaths), 300 L/min for INTERNAL BATTERY LEVEL
spontaneous breaths
Other Keys and Indicators
Plateau: 0.0 to 2.0 seconds
MESSAGE WINDOW: Shows up to
Flow pattern: Square or four lines, 20 characters each.
decelerating ramp 100% O2: Delivers 100% O2 to the
Mandatory (PCV) Settings patient for 2 minutes, then returns
to the current %O2 setting.
Respiratory rate: 3 to 70 breaths/min
MENU: View active and reset
Inspiratory pressure: 5 to 80 cmH2O
alarms; run self tests; enter stand-
Inspiratory time: 0.2 to 8 seconds by mode; adjust settings including
I/E Ratio: 1:99 to 4:1 alarm volume, apnea interval, VCV
Rise time factor: 5 to 100 flow type, ET tube size, humidifier
type and date/time, PCV timing
Spontaneous (PSV) Settings setting, and volume LED bar);
Support pressure: 0 to 70 cmH2O access oxygen sensor functions
(0 to 70 hPa) (calibrate, enable/disable %O2
Rise time factor: 5 to 100 alarm limits, enable/disable display
of sensor readings); display bat-
Exhalation sensitivity: 1% to 80%
tery, software revision, and service
Common Settings information.
PEEP/CPAP: 0 to 35 cmH2O SPEAKING VALVE MODE:
(0 to 35 hPa) Allows clinician to ventilate
Trigger sensitivity: 1 to 20 L/min patients on speaking valve while
% O2: 21% to 100% monitoring delivered volume.
APNEA PARAMS: Allows clinician
to select apnea ventilation
settings (respiratory rate, tidal
volume, peak flow). Apnea
interval is programmable from
10 to 60 seconds.
MANUAL INSP: Delivers one
mandatory breath.
KNOB: Adjusts values or selects menu ALARM SILENCE: Silences audible alarm Pneumatic Gas Sources
options. A setting or alarm pulses in its for 2 minutes. Air: Gas delivery system based on frictionless
display window to indicate that the knob is ALARM RESET: Clears all alarm indicators. piston design eliminates the need for a
linked to the setting. Alarm reactivates if alarm condition persists. compressor or hospital-grade wall air.
CURRENT: Lights to indicate that displayed CLEAR: Cancels proposed settings. Oxygen: Must be supplied at 40 to 90 psi
settings are in effect. (275 to 620 kPa) and at least 160 L/min
ACCEPT: Makes proposed settings effective.
PROPOSED: Lights to indicate that displayed at standard temperature and pressure,
INSPIRATORY PAUSE: Measures inspiratory
mode, breath type, and settings are being dry (STPD).
resistance and static compliance.
proposed.
EXPIRATORY PAUSE: Measures intrinsic and
total PEEP.

For more information, call 1-800-635-5267 or visit www.puritanbennett.com.


Environmental Specifications Compliance and Approvals
Temperature CE marked, indicating compliance to the
Medical Device Directive (93/42/EEC).
Operating: 41 to 113°F (5 to 45°C) at 10% to
95% relative humidity IEC 601-1 classification: Protection class I,
Type B, internally powered, drip-proof
Storage: -40 to 140°F (-40 to 60°C) at 10% to
equipment, continuous operation.
95% relative humidity
Complies with these International and
Atmospheric pressure European standards:
Operating: 8.7 to 16.0 psi (600 to 1,100 hPa) IEC 601-1, EN 60601-1, IEC 601-1-2,
Storage: 7.3 to 16.0 psi (500 to 1,100 hPa) EN 60601-1-2, EN 794-1.
Approved to the type test requirements of
Altitude Annex III of the Medical Device Directive by
Operating: 4,570 m (15,000 ft) maximum the notified body, TUV Rheinland.
Storage: 15,240 m (50,000 ft) maximum Certified by these test agencies:
Power TUV Rheinland: EC Type Test Certificate to
the requirements of Annex III of the Medical Mounting
Power supply input range: 100 to 120 V AC Cart or shelf mount allows maximum flexibility
Device Directive.
nominal, 220 to 240 V AC nominal; 50/60 Hz;
CSA: C22.2 No. 601-1, CSA C22. 2 No. 601-2- Shelf requirements: Able to hold 220 lb
2.9 A (100 to 120 V units) or 1.3 A (220 to
12, UL 2601-1, and NRTL certification. (100 kg) (to accommodate ventilator,
240 V units); 290 VA.
humidifier, collector vial, external battery,
Earth leakage current (including ventilator, Physical Characteristics and brackets); 12'' (300 mm) deep minimum
power cord, and internal and external for access to fan filter; 40'' (1,000 mm)
batteries): Less than 300 µA at 100-120 V AC Weight
clearance above shelf; 20'' (500 mm)
operation; less than 500 µA at 220-240 V AC Ventilator only: 66 lb (30 kg) clearance below shelf; 36'' (900 mm)
operation. Cart only: 45.5 lb (20.7 kg) width (room for external battery)
Enclosure leakage current: Less than 50 µA in External battery: 28 lb (12.75 kg)
Normal Condition (all units); less than 300 µA Oxygen sensor life
in Single Fault Condition at 100-120 V AC Shipping weight 2 years, nominal (actual life depends on
operation; less than 500 µA in Single Fault Ventilator only: 118 lb (53.5 kg) operating environment)
Condition at 220-240 V AC operation. Cart only: 57 lb (23.6 kg) Alarm volume
Internal battery Dimensions 74 to 88 dBA at 1 m (user-adjustable)
Operating time: Approximately 2.5 hours under Ventilator only: Communications
nominal conditions. Approximately 2 hours 14.9'' H x 20.3'' W x 14.6'' D
under extreme conditions.* An optional communications panel allows
(378 mm H x 515 mm W x 370 mm D)
the 760 Ventilator to interface with a number
Recharge time: 2.5 to 3 hours in ventilator. Ventilator and cart: of external devices including a stand-alone
Charges automatically while ventilator is 47.0'' H x 26.0'' W x 23.5'' D remote alarm, nurse call systems, central
connected to AC power and power switch (1,194 mm H x 660 mm W x 597 mm D) stations and information systems, such
is ON (including standby mode).
Shipping dimensions as CliniVision ® Clinical Informatin System.
Charge level: Indicated on keyboard.
Ventilator only:
External battery 26.8'' H x 27.6'' W x 31.5'' D
Operating time: Approximately 7 hours under (680 mm H x 700 mm W x 800 mm D)
nominal conditions. Approximately 5.5 hours Cart only:
under extreme conditions.* 36.0'' H x 26.0'' W x 12.0'' D
Recharge time: 7.5 to 8 hours in ventilator, 3 to (915 mm H x 660 mm W x 305 mm D)
4 hours using optional battery charger.
Charges automatically while ventilator is con- Connectors
nected to AC power and the power switch is Inspiratory limb connector: ISO 22-mm
ON (including standby mode). Charge level conical male
available under menu function. Expiratory limb connector: ISO 22-mm
* Nominal conditions: tidal volume 0.6 L, respiratory conical male to collector vial
rate 15/min, PEEP/CPAP 5 cmH2O, peak flow Oxygen inlet: DISS male, DISS female, NIST,
60 L/min, plateau 0 seconds, peak pressure 30 Air Liquide, Dräger, or Australian fitting
cmH2O, mean pressure 8 cmH2O.
Gas exhaust port: ISO 30-mm conical
Extreme conditions: tidal volume 1.2 L, respiratory
rate 15/min, PEEP/CPAP 15 cmH2O, peak flow
60 L/min, plateau 0 seconds, peak pressure 64
cmH2O, mean pressure 24 cmH2O. Battery life
specifications are approximate and are for new,
fully charged batteries at 68˚F (20˚C) and sea level.
To ensure maximum battery life, maintain full charge
and minimize the number of complete discharges.
760 Ventilator System
UNIVERSAL RESERVOIR

Optional Accessories Catalog Number

Ventilator Breathing Circuits


Reusable, with water traps, adult G-061208-00
Reusable, with water traps, pediatric G-061223-00
Reusable, with heated wire, for Fisher & Paykel Humidifiers, adult G-061235-00
Reusable, with heated wire, for Fisher & Paykel Humidifiers, pediatric G-061237-00
Water Trap, in-line G-061213-00
D/X7 ™ Expiratory Bacteria Filter, single-patient use (carton of 12) G-060526-00
4280 Hacienda Drive
Re/Flex ™ Inspiratory Bacteria Filter, reusable, 22-mm ISO connectors (each) 4-074600-00
Pleasanton, CA 94588
Humidifier Bracket, adaptable G-062612-00 Tel 925.463.4000
Shelf-Mount Kit, adaptable to most rail systems; includes brackets Toll Free 1.800.635.5267
for mounting collector vial and humidifier for use with Fisher & Paykel Humidifiers G-061279-00 www.puritanbennett.com

for use with Hudson RCI ConchaTherm Humidifiers G-061601-00 Tyco Healthcare UK LTD.
External Battery Kit, designed for easy mounting on ventilator cart but can be used 154 Fareham Road
without cart; includes battery and cover to protect battery terminals G-061140-00 Gosport, UK PO13 0AS
Tel +44.1329.224000
External Battery Charger
North America and Japan G-061260-00
Australia G-061261-00 Specifications subject to
change without notice.
Continental Europe G-061500-00
© 2003 Nellcor Puritan
Denmark G-061501-00
Bennett Inc. All rights
India/South Africa (old British-style plug with round prongs) G-061504-00 reserved.

Israel G-061505-00 O.f 1213-0603


VE05100
Italy G-061502-00
Switzerland G-061503-00
United Kingdom G-061499-00
Filter, air intake, package of 6, to be replaced every 1,000 hours of operation;
one comes standard with every ventilator G-061262-00
Filter, main fan, package of 6, to be replaced every 250 hours or as needed;
one comes standard with every ventilator G-061263-00
Standard Cart G-062611-00
Standard Cart Kit designed for in-house transport; includes dual oxygen cylinder
mounting kit and adaptable humidifier bracket. Order humidifier mount separately. G-062616-00
Dual E Cylinder Holder Bracket G-062613-00
Linkage Arm Kit, used to connect non-standard wheelchair to EasyCart G-061845-00
Humidifier Mounts
For use with Fisher & Paykel Humidifiers G-061227-00
For use with Hudson RCI ConchaTherm Humidifiers G-061228-00
Manual, Service
English G-061875-00
Preventive Maintenance Kits
15,000-hour G-061166-00
30,000-hour G-061167-00 Standard Cart
Keyboard cleaning spray G-061576-00
O2 Regulator, for E cylinder (2 x DISS & 0-25 L/min) 6-900003-00